s medical Data AXIOM Artis U The little big one for universal angiography applications
Transcript
Page 1: Axiom Art is u Data Sheet

smedical

Data

AXIOM Artis U

The little big one for universal angiography applications

Page 2: Axiom Art is u Data Sheet

2

The AXIOM Artis U has the power, strength and image quality of an angiographic suite with the footprint and maneuverability of a portable C-arm for a dedicated interventional room.

The system is intended to support a wide field of applications, e.g., angiographic interventions, cardiac interventions and GI procedures.

The AXIOM Artis U provides a state-of-the-art imaging system that can be configured from single exposure acquisition up to 30 f/s for cardiac applications. The imaging system provides exceptional image management to meet the needs of any IT depart-ment for seamless integration.

The AXIOM Artis concept is based on four main pillars:

• Easy handling• Outstanding image quality• Maximized radiation protection• Optimal networking

These four elements enable the user to optimize the clinical workflow in the field of cardiology, angiography, neuroangiography and various non-vascular procedures.

The flexible configuration of the system makes it possible to control the system either in the examination room or outside of the control room*.

System Highlights

Gantry System:

• Movable C-arm system

Operation:

• Complete ergonomic system control in the examination room with integrated generator and imaging system control

• syngo® operation in the examination room and in the control room

Image acquisition system:

• Choice between two high-resolution image intensifier sizes:

23 cm I.I. with three modes, 33 cm I.I. with four modes for improved detail

recognition

• Full-size CCD TV system with digital imaging chain

Image processing system:

• Digital imaging system with FLUOROSPOT Compact

• Standardized platform with syngo®-based user interface

• Imaging modes ranging from single frame to series acquisition with frame rates from 0.5 to

8 frames/second in 15 f/s with 1k2 or 30 f/s with 1024 x 512/10-bit matrix

• Online DDO* for fluoroscopy and acquisition

• Pulsed fluoroscopy

• Expanded CARE radiation protection

• Comprehensive DICOM 3 functionality• Integrated CD burner*

X-ray generation:

• High frequency generator with CAREMATIC

– Choice of 65 kW or 80 kW*

– Automatic exposure control IONTOMAT PN

– Operation of the generator integrated into the imaging system

• Dual-focus, high performance X-ray tube assembly with low-noise air-cooling

• High performance focal spot for optimal image quality

Display and display carrier:

• One or two* 18“ (46 cm diagonal) TFT high-contrast black/white flat display for flicker-free image on a movable display trolley

• In the trolley: integrated drawer with keyboard and mouse**

Control room*:

• Additional 18“ TFT flat-screen monochrome display

• Keyboard and mouse

• 2nd footswitch for X-ray release

** Option** Without control room option

AXIOM Artis U

The little big one for universal angiography applications

Page 3: Axiom Art is u Data Sheet

2

Page 4: Axiom Art is u Data Sheet

4

AXIOM Artis U

Technical Data

5

Stand

Orbital movement + 90° to – 45°

Swivel range ± 10°

Immersion depth 87 cm

Horizontal travel 20 cm

BV 33 cm 23 cm

Angulation ± 145° ± 190°

Vertical travel 38 cm 28 cm

Source I.I. distance 110 cm 100 cm

Imaging system

High resolution, digital imaging system with innovative image display, DICOM network connection and syngo® -based user interface

Tableside control for system operation of all important imaging system functions, fluoro and acquisition parameters, filters, collimators and selection of organ programs. Can be mounted directly at the accessory rail of the patient table or put on the monitor trolley

Reference image*: Storage and direct access to reference images on a second flat display in the examination room

DDO (dynamic density optimization) for real-time harmonization of fluoroscopy, native series and single images*

Real-time edge enhancement, positive/negative image display, windowing, contrast/brightness, electronic shuttering, image shift (roaming), vertical and horizontal image reversal, image rotation, zoom functions

Fluoro Loop*: Storage of fluoroscopy images and scenes up to 512 images. Stores fluoroscopy images during fluoroscopy as well

Fast, direct access to all series, individual images, and photo files via MULTIMAP, in both the examination room and the control room*

Image storage capacity:40,000 images in 512 matrix or 10,000 images in 1k matrix

Standard video output for video recorder* or video printer*

Paper printer output for printing images (not for diagnostics), examination protocols and patient data via PostScript printer level 2*.In the U.S., paper printouts should not be used for diagnosis, unless the PostScript printer has specifically received 510 (k) clearance for this purpose.

** Option** Scheduled for later release

Page 5: Axiom Art is u Data Sheet

4

AXIOM Artis U

Technical Data

5

Operating modes

Digitally pulsed fluoroscopy with 3, 7.5 or 15 p/s and continuous fluoroscopy with 30 f/s in 1k/10 bit and digital real-time filtering, noise reduction through sliding weighted averaging

DR

Digital Radiography with 1k/10-bit matrix, digital filtering, single image and series native acquisitions with 0.5, 1, 2, 3, 4, 6 or 8 f/s

DSA*

Digital subtraction angiography with pixel shift, remasking, roadmapping, peak opacification (max. and min. opac.), landmarking in 1k-matrix with real-time filtering and 0.5, 1, 2, 3, 4 or 6 f/s*

Cardiac acquisition*

Cardiac acquisition with 15 f/s, acquisition and storage in 1k2 matrix

Cardiac acquisition with 30 f/s, acquisition and storage in 1024 x 512 matrix

Vascular Analysis

Measurement program integrated into the image system for objective, exact, and reproducible vascular analysis. With automatic contour recognition, stenosis quantification, distance and angle measurement and different calibration methods*

DICOM 3 Network Interfaces

DICOM Send/StC*

Network interface in DICOM 3 standard for DICOM-3-compatible image communication network

Feedback from the image archive (StC = Storage Commitment)

DICOM Print*

For connection to a network laser camera or printer

DICOM Query/Retrieve*

Retrieve images from a patient image archive

DICOM Get Worklist/MPPS*

Get Worklist function for importing patient data from a data management system (RIS/HIS)

Modality Performed Procedure Step (MPPS): send examination statistics and dose information to a patient data management system

CD Recorder*

CD drive for digital image storage on a CD-ROM for offline data exchange in DICOM 3, TIFF and AVI formats

Page 6: Axiom Art is u Data Sheet

6

AXIOM Artis U

Technical Data

7

CARE

CAREMATIC

CAREMATIC automatic X-ray control system for fully automatic calculation and optimization of exposure data based on fluoroscopy values

CAREFILTER

Three-level adaptive Cu-prefiltration for reducing skin dose; configured in organ program or automatically guided selection depending on the absorption of the object

Filtration levels 0.1, 0.2, 0.3 mm Cu

CAREVISION

Pulsed fluoroscopy with reduced pulse frequencies from 3.0, 7.5, 15 p/s. Pulse frequency adjusted to the requirements of each application for significant reduction of radiation exposure, particularly during interventions

CAREPROFILE*

Radiation-free positioning of primary and semi-transparent collimators via graphic display in LIH-image on the image monitor

CAREWATCH*

Electronic unit with measurement chamber integrated into the collimator housing for acquisition of area dose product or skin dose displayed on the image system monitor

Different displays can be configured for fluoroscopy and for fluoro pause:

During fluoro Skin dose rate

During fluoro pause Accumulated skin dose, or area dose product or percentage of a configurable dose limit value (amount of fluoro and acquisition)

* Option

Page 7: Axiom Art is u Data Sheet

6

AXIOM Artis U

Technical Data

7

X-ray TV system

High-resolution digital X-ray TV system with maintenance-free 1k CCD-camera chip, max. 69 dB

Operation

In the examination room

Moveable cart with four casters and brake

One or two high-contrast monochrome flat displays, drawer with keyboard and mouse

Acoustic alarm and two radiation displays

Ergonomic foot switch for triggering fluoroscopy and acquisition

TSC (Tableside control) for all important functions of the image system, acquisition and fluoroscopy parameters incl. organ programs and collimator control. The TSC can be mounted at the table or on the trolley.

In the control room*

Control room with additional high-contrast monochrome flat display

Extender for keyboard and mouse use in control room:

Input of patient data and post-processing can be performed outside the examination room

Ergonomic foot switch for triggering fluoroscopy and acquisition:

2nd X-ray release and live-image display is possible without dose burden for the examiner

Data display DSB 1803-DC

18“ TFT high-contrast black/white flat display for flicker-free, distortion-free live and reference image display in X-ray diagnostics as well as for interventional, therapeutic procedures

Light weight, high luminosity and contrast values

Diagonal screen measurement 18“ (46 cm)

Image display 1280 x 1024

Typical brightness 450 cd/m2

Maximum brightness 600 cd/m2

Contrast ratio 400 : 1

Horizontal viewing area 160º

Page 8: Axiom Art is u Data Sheet

8

AXIOM Artis U

Technical Data

9

X-ray image intensifiers

Depending on the diagnostic and therapeutic focus, 23 cm or 33 cm I.I. are available with high dynamics for super-selective and survey exposures

The high contrast and excellent modulation properties allow display of the finest details

Cesium iodide input screen for high quantum absorption to generate minimal quantum noise

Precision electronic optics with minimal image distortion and consistent high resolution across the entire image field

Anti-reflective coating and glare trap

Semiconductor measurement fields for dose control

Beam quality according to IEC 61262-5

DQE (ICE 61262-5) > 65%

I.I. laser light localizer for radiation-free positioning*

Image intensifier 23 cm

I.I. formats 23 cm 16 cm 12 cm

Minimum resolution (lp/mm) 4.6 5.2 5.8

Removable grid Pb 15/80

Image intensifier 33 cm

I.I. formats 33 cm 22 cm 16 cm 13 cm

Minimum resolution (lp/mm) 3.8 4.6 5.0 5.8

Removable grid Pb 17/70

Collimator

Angio collimator with iris and square collimator, wedge-shaped diaphragm blades for DSA and cardiology applications and graduated finger filter

Independent rotation and shift of the filter blades

Card collimator for configuration with 23 cm I.I. with a rectangular blade and with a wedge-shaped filter diaphragm (scheduled for later release)

Tube OPTITOP 150/40/80 HC-100

Nominal voltage (IEC 60613) 150 kV

Focal spot (IEC 60613) 0.6 1.0

Output (IEC 60613) 40 kW 80 kW

Cold output (OW) 52 kW 103 kW

Anode angle (IEC 60788) 12°

Anode heat storage capacity 580,000 J (783,000 HU)

Anode drive 150 Hz (3-phase current)

Total filtration (IEC 60601-1-3) ≥ 2.5 mm Al

* Option

Page 9: Axiom Art is u Data Sheet

8

AXIOM Artis U

Technical Data

9

Generator POLYDOROS SX 65 or POLYDOROS SX 80

Microprocessor controlled HF-generator with automatic dose regulation for radiography and fluoroscopy

Generator control is completely integrated into the image system

Up to 25 organ programs for storage of acquisition parameters

POLYDOROS SX 65

Power rating 65 kW (650 mA at 100 kV) acc. to IEC 601-2-7 40 kV to 150 kV

Fluoroscopy 0.1 mA to 100 mA

Acquisition up to 800 mA

Minimal pulse width 4 ms

POLYDOROS SX 80*

Power rating 80 kW (800 mA at 100 kV) acc. to IEC 601-2-7 40 kV to 150 kV

Fluoroscopy 0.1 mA to 100 mA

Acquisition up to 1,000 mA

Minimal pulse width 4 ms

Remote Service

Preparation for Siemens Remote Service SRS:

Hardware and software remote diagnostics

Remote system configuration, e.g., addition of a DICOM node, early warning system for ensuring system operation

Installation Data

Rated voltage 3/N/PE ~ 400/440/480 V (± 10%) at 50Hz/60Hz; 440/480 V (± 10%) possible with optional power transformer*

Power consumption max. 145 kVA

Room size min. 400 x 450 cm

Room height min. 250 cm

Ambient conditions (operation)

Examination and control room Temperature range: + 10°C to + 35°C Relative humidity: 15 – 75% below dew point Air pressure: 700 hPa to 1060 hPa

System weight

AXIOM Artis U approx. 410 kg

FLUOROSPOT Compact approx. 80 kg

Trolley approx. 110 kg

POLYDOROS SX 65/80 approx. 290 kg

Page 10: Axiom Art is u Data Sheet

10

AXIOM Artis U

Technical Data

11

Room layout (mm)

����

��

��

����

��

��

Page 11: Axiom Art is u Data Sheet

10

AXIOM Artis U

Technical Data

11

System view (mm)

��������

1100

/ ¹/

1000

²

¹ BV 33 / ² BV 23

145°¹/190° 145°/190°²

Page 12: Axiom Art is u Data Sheet

© 2005 Siemens Medical SolutionsOrder No. A91001-M1400-G936-3-7600Printed in GermanyAX CRM NA 03053

CAUTIONLASER RADIATIONDO NOT STAREINTO BEAM

PEAK POWER < 1 mW

WAVE LENGTH 540 - 700 mm

CLASS II LASER PRODUCT

(For U.S. and Canada)

Siemens Medical Solutions that help

The information in this document contains general descriptions of the technical options available, and may not always apply in individual cases. The required features should therefore be specified in each individual case at completion of contract.

Siemens reserves the right to modify the design and specifications contained herein without prior notice. Please contact your local Siemens sales representative for the most current information.

Original images always lose a certain amount of detail when reproduced.

In the interest of complying with legal requirements concerning the environmental compatibility of our products (protection of natural resources and waste conservation), we recycle certain components. Using the same extensive quality assurance measures as for factory-new components, we guarantee the quality of these recycled components.