Eutech 700 Series Bench Meters • pH • ORP • Ion • Conductivity • TDS • Dissolved Oxygen • Temperature
Transcript
Page 1: Bench 700 Series-1

Eutech 700 Series

Bench Meters• pH • ORP • Ion • Conductivity • TDS • Dissolved Oxygen • Temperature

Page 2: Bench 700 Series-1

Features •SimpleOperation•DualMeasurementIncludes TemperatureDisplayWithAllReadings•100PointMemory•DetachableElectrodeHolder

•CompactFootprintSavesValuable BenchSpace•CustomSetupOptions•HoldFunctionAndStabilityIndicator•3-YearWarranty

page | 02

Bench700Series

Comparison TableMeter

Features

pH700 Ion700 CON700 DO700 PC700

pH & mV 3 3 3Ion 3Temperature 3 3 3 3 3Conductivity & TDS 3 3Dissolved Oxygen 3Memory 100 data sets

Power Requirements 9 V DC adapter, 3.3 A (100/240 VAC, SMPS)

Dimension ; Weight Meter: 17.5 (L) x 15.5 (W) x 6.9 (H) cm ; 650 g ; Boxed: 30.8 (L) x 23.5 (W) x 12.4 (H) cm ; 1800 g

Multi- Electrode

Holder More

organised, less space occupied

Non-SkidPadsHold meter steady during operation

Slide-OutQuick ReferenceGuide

For quick, convenient referencing

GenerousDisplay with large digits

Splashproof Keypad protects against incidental spills

ElectrodeArmcan be used

on either side

Page 3: Bench 700 Series-1

page | 03

Ion 700pH/Ion/mV-Redox-ORP/TemperatureIon-selective electrode measurement has never been this easy – or economical! The Ion 700 measures and records up to 100 data points with 0.01 resolution.

• MeteracceptspH,RedoxorIonselectiveelectrodes–it’slikehavingthree meters in one!

• Large,easy-to-readdisplay

• DirectIonconcentrationmeasurement–usewithmanydifferent Ion-selectiveelectrodes

• Upto5-pointcalibrationwithauto-bufferrecognition

• Non-volatilememoryholdsupto100datapoints

• Integralelectrodeholder

Applications: • Educational: Useful for most laboratory, ecological studies and other applications. • Laboratory: Environmental studies, chemical labs, titrations and quality assurance testing. Use in all types of food processing.

pH 700pH/mV-Redox-ORP/TemperatureEconomical, user-friendly and accurate, the Eutech pH 700 is your ideal choice for routine applications in laboratories, productions plants and schools.

• Large,easy-to-readdisplay

• Readyindicatorletyouknowwhenreadingsarestable

• Upto5-pointcalibrationwithauto-bufferrecognition

• Non-volatilememoryholdsupto100datapoints

• Integralelectrodeholder

Applications: • Educational: Useful for most laboratory, ecological studies and other applications. • Laboratory: Environmental studies, chemical labs, titrations and quality assurance testing. Use in all types of food processing.

Eutech pH 700 & Ion 700

Page 4: Bench 700 Series-1

page | 04

Eutech CON 700 & DO 700

CON700 Conductivity/TotalDissolvedSolids/TemperatureThe economy Eutech CON 700 offers years of reliable, accurate and consistent performance. User-friendly features make the instrument an ideal research partner in laboratories, productions plants and schools.

• LargescreenthatdisplaysConductivityorTDSreadingswithTemperature in ºC or ºF

• Readyindicatorletyouknowwhenreadingsarestable

• Selectablecellconstant

• Auto-rangingacross5ConductivityandTDSranges

• Upto5-pointpushbuttoncalibration

• Non-volatilememoryholdsupto100datapoints

• Integralelectrodeholder

Applications: • Environmental studies • Chemical laboratories • Quality assurance testing • Food science • Ecological studies • Education institution

DO 700DissolvedOxygen/TemperatureEconomical, user-friendly and accurate, the Eutech DO 700 is your ideal choice for routine applications in laboratories, productions plants and schools.

• DissolvedOxygenmeasurementsinppm,mg/Lor%saturation

• AccuratereadingsinvaryingconditionswithTemperature,Salinityand Barometric Pressure Compensation

• Large,easy-to-readdisplay

• Readyindicatorletyouknowwhenreadingsarestable

• Non-volatilememoryholdsupto100datapoints

• Integralelectrodeholder

Applications: • Environmental studies • Wastewater and water treatment • Ecological studies • Education institution

Page 5: Bench 700 Series-1

page | 05

PC 700pH/mV-Redox-ORP/Conductivity/TotalDissolvedSolids/TemperatureThe multi-parameter Eutech PC 700 measures pH, ORP, conductivity and temperature with a single meter. User-friendly with advanced features and a large custom-size display – perfect for research work on a budget!

• Largescreeninacompactfootprint

• MeteracceptspHandRedoxelectrodes,inadditiontoConductivity/TDS– it’slikehavingfourmetersinone!

• Upto5-pointpushbuttoncalibration

• Selectablecellconstant

• Auto-rangingacross5Conductivityranges

• Easyinterchangeabilityofprobes

• Non-volatilememoryholdsupto100datapoints

• Integralelectrodeholder

Applications: • Water quality testing • Checking metal finishes • Solutions • Cooling tower water • Printing fountain solutions • Boiler water • Brines • Drilling mud • Rinse tanks • Ponds • Pollution control • Recirculating systems • Waste water • Industrial process systems • General use in laboratories • Titration • Quality assurance • Ecological studies • Food-processing

Ordering InformationModel Order Code Part No. DescriptionpH ECPH70042S 01X541605 EutechpH700MeterWithpHElectrode (ECFC7252101B),ATCProbe(ECPHTEMPB01P), IntegralElectrodeHolder&100/240VACAdapter

pH ECPH70040S 01X541603 EutechpH700MeterWithIntegral ElectrodeHolder&100/240VACAdapter

pH ECPH70040 01X541601 EutechpH700Meter100/240VACAdapter

Ion ECION7004S 01X541609 EutechIon700MeterWithIntegral ElectrodeHolder&100/240VACAdapter

CON ECCON70043S 01X543401 EutechCON700MeterWithConductivity/ ATCProbe(CONSEN9501D),Integral ElectrodeHolder&100/240VACAdapter

DO ECDO70042S 01X543501 EutechDO700MeterWithDissolved Oxygen/ATCProbe(01X233913),Integral ElectrodeHolder&100/240VACAdapter

PC ECPC70043S 01X543601 EutechPC700MeterWithpHElectrode (ECFC7252101B),Conductivity/ATCProbe (CONSEN9501D),IntegralElectrodeHolder &100/240VACAdapter

Meters

Model Order Code Part No. DescriptionOthers 60X030130 60X030130 SMPS, 100/240 VAC, 9 V, 6 W Power Adapter

Accessories

Model Order Code Part No. DescriptionpH ECBU1BT 01X211211 pH1.68BufferSolutionpH ECBU4BT 01X211201 pH4.01BufferSolutionpH ECBU686BT 01X211242 pH6.86BufferSolutionpH ECBU7BT 01X211202 pH7.00BufferSolutionpH ECBU918BT 01X211271 pH9.18BufferSolution

pH ECBU10BT 01X211203 pH10.01BufferSolutionpH ECBU12BT 01X211212 pH12.45BufferSolutionpH ECDPCBT 01X211216 ProteinCleaningSolutionForpHElectrodepH ECRE005 01X211205 StorageSolutionForpHElectrodeORP ECORPPRE 01X081500 ORPPre-TreatmentSolution(475mV)

ORP ECORPQUIN 01X062700 ORPQuinhydrone255 (255 mV, ±15 mV @ 25 ºC) Solution

ORP ECORPQUIN086 01X211215 ORPQuinhydrone86(86mV)Solution

CON ECCON84BT 01X211241 84 μS Conductivity Standard Solution

CON ECCON1413BT 01X211207 1413 μS Conductivity Standard Solution

CON ECCON1288BT 01X211210 12.88 mS Conductivity Standard Solution

CON ECCON1118BT 01X211244 111.8 mS Conductivity Standard Solution

TDS EC442300BT 01X109102 300 ppm 442 Standard Calibration Solution

TDS EC4421000BT 01X109104 1000 ppm 442 Standard Calibration Solution

TDS EC4423000BT 01X109101 3000 ppm 442 Standard Calibration Solution

Solutions (480 ml Per Bottle)

Model Order Code Part No. DescriptionpH ECFC7252101B 01X099412 GeneralPurposePlastic-BodySingle JunctionGel-FilledpHCombination Electrode,90x12mm,BNC Connector, 1 m Cable

CON CONSEN9501D 01X466602 2StainlessSteelRingsUltem-Body ConductivityElectrode,ATC, Cell Constant K=1.0, 144 x 16 mm, 8-PinConnector,1mCable

DO ECDO6HANDY 01X233913 GalvanicDissolvedOxygenElectrode WithATC,2AssembledMembrane CapHousings,1RefillingElectrolyte &1ScouringPad,0.9mCable

TEMP PH5TEMB01P 01X210303 ATCProbe,StainlessSteel, 117 x 3 mm Shaft, 1 m Cable

Electrodes

Eutech PC 700

Page 6: Bench 700 Series-1

|Warranty| EutechInstrumentsprovidesthree years of warranty against manufacturing defects for our meters, and six months for our electrodes.

Eutech reserves the right to make changes, improvements and modifications to products shown.

Bench700Series

Specification Tables

<<Wastewater&WaterTreatment

>>EcologicalStudies

pH/IonBenchMeters

pH700 Ion700

pH

Range -1.99to16.00pH 0.00 to 14.00 pH

Resolution 0.01 pH

Accuracy ±0.002

Cal. Points Upto5points

Buffer Sets USA,NIST

ORP

Range ±1999 mV

Rel. mV Range ±1999 mV

Resolution 0.1 mV (±199.9 mV) / 1 mV (beyond)

Accuracy ±0.2 mV (±199.9 mV) / ±2 mV (beyond)

Ion

Concentration – 0.01 to 2000 ppm

Resolution (ppm) – 0.01 (<1); 0.1 (1.0 to 199.9); 1 (>200)

Accuracy – ±0.5%fullscale(monovalent)±1.0%fullscale(divalent)

Cal. Points – From 2 to 5 consecutive points; 0.1, 1.0, 10.0, 100.0 or 1000.0 ppm

Temperature

Range 0.0 to 100.0 ºC / 32.0 to 212.0 ºF

Resolution 0.1 ºC / 0.1 ºF

Accuracy ±0.5 ºC / ±0.9 ºF

Compensation Automatic or Manual (0 to 100 ºC / 32 to 212 ºF)

Calibration Offsetin0.1ºincrements;offsetrange:±5ºC/9ºF

ConductivityBenchMeter

CON700

Conductivity

Range .. µS/cm to 200.0 mS/cm

Resolution 0.01 μS; 0.1 μS ; 1 μS ; 0.01 mS ; 0.1 mS

Accuracy ±1%fullscale

Cal. Points Automatic (4 points); maximum 1 per range; manual (5 points); maximum 1 per range

Cell Constant 0.1, 1.0, 10.0 (selectable)

TDS

Range [email protected]([email protected])

Resolution 0.01 ppm; 0.1 ppm ; 1 ppm ; 0.01 ppt ; 0.1 ppt

Accuracy ±1%fullscale

Cal. Points Upto5points

TDS Factor 0.40 to 1.00 (adjustable)

Temperature

Range (Meter) 0.0 to 100.0 ºC / 32.0 to 212.0 ºF

Resolution 0.1 ºC

Accuracy ± 0.5 ºC / ± 0.9 ºF

Coefficient (Per ºC) 0.00to10.00%

Normalization 15.0 to 30.0 ºC (adjustable)

Compensation Type & Range

Automatic with supplied cell or manual 0.0 to 100 ºC / 32.0 to 212.0 ºF, 0 to 80 ºC / 32.0 to 176.0 ºF with supplied cell

Calibration Offsetin0.1ºincrements;offsetrange:±5ºC/9ºF

page | 06

Page 7: Bench 700 Series-1

<< Pharmaceutical

>>CoolingTower

page | 07

Multi-ParameterBenchMeter

PC700

pH

Range -1.99to16.00pH

Resolution 0.01 pH

Accuracy ±0.002

Cal. Points Upto5points

Buffer Sets USA,NIST

ORP

Range ±1999 mV

Rel. mV Range ±1999 mV

Resolution 0.1 mV (±199.9 mV) / 1 mV (beyond)

Accuracy ±0.2 mV (±199.9 mV) / ±2 mV (beyond)

Conductivity

Range .. µS/cm to 200.0 mS/cm

Resolution 0.01 μS; 0.1 μS ; 1 μS ; 0.01 mS ; 0.1 mS

Accuracy ±1%fullscale

Cal. Points Automatic (4 points); maximum 1 per range; manual (5 points); maximum 1 per range

Cell Constant 0.1, 1.0, 10.0 (selectable)

TDS

Range [email protected]([email protected])

Resolution 0.01 ppm; 0.1 ppm ; 1 ppm ; 0.01 ppt ; 0.1 ppt

Accuracy ±1%fullscale

Cal. Points Upto5points

TDS Factor 0.40 to 1.00 (adjustable)

Temperature

Range (Meter) 0.0 to 100.0 ºC / 32.0 to 212.0 ºF

Resolution 0.1 ºC / 0.1 ºF

Accuracy ± 0.5 ºC / ± 0.9 ºF

Coefficient (Per ºC) 0.00to10.00%

Normalization 15.0 to 30.0 ºC (adjustable)

Compensation Type & Range

Automatic with supplied cell or manual 0.0 to 100 ºC / 32.0 to 212.0 ºF, 0 to 80 ºC / 32.0 to 176.0 ºF with supplied cell

Calibration Offsetin0.1ºincrements;offsetrange:±5ºC/9ºF

DissolvedOxygenBenchMeter

DO700

Dissolved OxygenRange 0 to 30.0 mg/L

Resolution 0.01 mg/L

Accuracy ±0.5%fullscale

% Saturation Of Oxygen

Range 0to199.9%;300%

Resolution 0.1%/1%

Accuracy ±0.5%fullscale

Salinity CorrectionRange 0 to 50 ppt

Resolution 0.1

Barometric Pressure Correction

Range 450 to 825 mmHg

Resolution 1

Accuracy Manual input

Method Galvanic

TemperatureRange 0.0 to 100.0 ºC / 32.0 to 212.0 ºF

Resolution 0.1

Accuracy ± 0.5 ºC / ± 0.9 ºF

Page 8: Bench 700 Series-1

BEN

08X

2338

09 0

2/10

Rev

0

Eutech Instrumentsmanufactures a comprehensive rangeof laboratory andfield instrumentation forwater analysis. Eutech Instruments alsoproduces on-lineprocessinstrumentsforwaterqualitymonitoringandcontrol.InnovativeproductsfromEutechinclude:

|World-WideSalesandServiceSupport | Eutechprovidesworld-wide sales and service support for all its products through its network of international distributors. Contact us at [email protected] for details of your nearest distributor.

Dual Display Testers | Economy + Series Handhelds |Waterproof

Colorimeters and Turbidimeters | CyberScan Handhelds | Bench

Meters | CyberScanTouchscreenMeters | HandheldThermometers

Part of Thermo Fisher Scientific

Distributed by:

Global OfficesAsiaPacific(HeadOffice) | EutechInstrumentsPteLtdBlk55,AyerRajahCrescent,#04-16/24,Singapore139949Tel:(65)6778-6876•Fax:(65)6773-0836•www.eutechinst.com

North&SouthAmerica | OAKTONInstruments625EBunkerCt,VernonHills,IL60061,USATel:tollfree1-888-4OAKTON(1-888-462-5866)•Fax:(1)847-247-2984•www.4oakton.com

Europe | EutechInstrumentsEuropeB.V.P.O.Box254,3860AGNijkerkNetherlandsWallerstraat125K,3862 CNNijkerkNetherlandsTel:(31)033-2463887•Fax:(31)033-2460832•www.eutech.nl

Malaysia | EutechInstrumentsSdnBhdNo.1,LengkokKeluli2,BukitRajaPrimeIndustrialPark,41050BukitRaja,KlangSelangorD.E.,MalaysiaTel:(603)3342-8533•Fax:(603)3342-6520•www.eutechinst.com

China | ThermoFisherScientificBuilding6,No.27,XinJinqiaoRoad, Shanghai 201206, ChinaTel:(86)21-6865-4588•Fax(86)21-6445-7909•www.eutech.cn

India | ThermoFisherScientificPlotNo.C-327,TTCIndustrialArea,Pawane, NaviMumbai-400705,IndiaTel:(91)22-4175-8800•Fax:(91)22-4175-8801•www.eutechinst.com

Awarded in 2001 & 2003

2006 & 2007


Recommended