Date post: | 04-Dec-2014 |
Category: |
Documents |
Upload: | nash-anderson |
View: | 50 times |
Download: | 0 times |
Biomimetic Design of a Heart Valve
Nash Anderson
Cal Polytechnic State University
Spring 2011
Abstract
In order to test the theory of modifying the material of bio-prosthetic heart valves, an experiment
was conducted to determine whether elastomer films could be used as replacements for the
porcine tissue in bio-prosthetic heart valves. The elastic moduli and fatigue resistance of the
elastomers were tested to determine if they would be acceptable mechanically. The results were
found to be inconclusive due to improper testing methods and small sample size.
Background/Project Overview
Valvular heart disease (VHD) is characterized by the presence of a defect or damage to one of
the four heart valves. The defect or damage may be congenital or acquired. The damaged valve
either becomes too narrow to open fully, preventing normal blood flow; or unable to close,
completely resulting in back flow. About 5 million Americans are diagnosed with valvular heart
disease each year[1]. In mild cases valvular heart disease can be treated with medication, but in
most cases the valve must be replaced or repaired.
Valvular repair is the best possible solution to VHD, but when a patient’s heart valve is severely
damaged, repair is not an option. In these cases the patient’s valve must be replaced with a
either a mechanical or a bio-prosthetic valve.
Mechanical heart valves are made out of pyrolitic carbon and can last an entire lifetime if the
patient takes anticoagulant medication such as Warfarin on a daily basis. Without a frequent
regimen of anticoagulant medication the valve will clot at the mechanical hinges of the valve
resulting in failure. The valve hinges tend to clot due to their shape which causes turbidity of
flow.
Bio-prosthetic valves are the most common valvular replacement, being used in about 80% of
patients today[2]. These valves are made out of pericardium tissue (often from a pig), and closely
mimic the shape of actual heart valves. Because of their unique shape bio-prosthetic valves do
not require anticoagulant medication, which is why they are so popular amongst patients. The
drawback to these valves is that they need to be replaced every 10-15 years due to degradation
and calcification of the pericardium tissue. The pericardium tissue degrades and calcifies
because it is foreign tissue to the body, even though it is treated with chemicals to decrease this
effect.
Problem Statement
All replacement heart valves either have a limited lifespan or require the patient to take
anticoagulant medication for the remainder of their lives. The patient must choose which type of
valve is less inconvenient for them.
User Needs/ Current Solutions
User needs a prosthetic heart valve that will not require long term medication regimens and that
will have a lifespan longer than 25 years. The valve must last long enough so that it never has
to be replaced in the majority of patients.
Design Requirements
● Bio Compatible - not incurring a toxic or detrimental immunological response.
● Resist blood coagulation without use of anticoagulant medication
● Undergo elastic shear deformation
● Maintain a seal that does not permit back flow
● Maintain elastic properties under cyclic shear load (>1.05 billion cycles)
Proposed Solution
Theory Behind Solution
The design of the bio-prosthetic valve has preferred flow characteristics in comparison to
all mechanical valves. The only problem with the bio-prosthetic valve is the tendency of
the pericardium tissue leaflets to degrade and calcify over time. A synthetic material
would not degrade, therefore if a synthetic material can be found which replicates the
mechanical behavior of the pericardium tissue, a valve could be designed with the flow
characteristics of a bio-prosthetic valve and the durability and lifetime of a mechanical
valve.
How Solution Meets Design Requirements and user needs
The proposed heart valve will not require the patient to be put on an anti-coagulant
regimen, and will out last bio-prosthetic heart valves.
Explanation of design
The pericardium tissue used in bio-prosthetic heart valves will be replaced with an
artificial elastomer. The proposed heart valve will maintain the design and function of the
bio-prosthetic valve, but will last longer because the elastomer will not break down and
calcify over time.
Materials Science
Hypothesis
Elastomers closely resemble the mechanical properties of pericardial tissue. They
should therefore be a suitable material to mimic the function of a healthy heart valve.
Elastomers are chemical compounds whose molecules consist of several thousand
smaller molecules called monomers linked together by covalent bonds. These
monomers repeat and are linked together to form long chains. These chains have a
backbone most often made up of carbon bonds, either (C-C) or (C=C). These long
carbon chains are highly flexible, disordered and intertwined. The chains are flexible
because rotation around (C-C) bonds allows the molecules to take up many different
configurations. [3]
In elastomers normal state they are highly disordered with a high entropy. This is the
preferred state of the elastomer. When elastomers are put under tensile stress, the
molecular chains are pulled into alignment and often take on aspects of a crystalline
arrangement. When the chains are lined up under a load, they are at a lower disorder
with a lower entropy. Upon release they spontaneously return to their naturally
disordered, entangled state allowing the polymer to maintain its shape.[3] Both the
deformation and the subsequent recovery are time-dependent, suggesting that some
part of their behaviour is viscous. Elastomers show a combination of elastic and viscous
behavior known as viscoelasticity. The degree of viscoelasticity is strongly dependent
upon temperature and the rate of deformation, as well as such structural variables as
degree of crystallinity, crosslinking, and molecular weight.
In order to be useful for various applications, elastomers
must be strengthened by cross-linking the polymer chains.
With a low frequency of the branching cross links, a soft
rubbery material is produced. Silicones and polyurethanes
can be cast this way, using low-viscosity liquid precursors
with reactive end groups.
If an elastomer is stretched, as shown in Figure 1, energy
is stored in it. Just as in the application of a slingshot, the
elastomer used in the propulsion mechanism will snap
back into place after being stretched. The energy stored
per unit volume in an elastically strained material:
Materials
To pick the materials for testing a CES plot from the biomaterials database was created
looking for low Young’s Modulus (.8-12MPa) and high Fatigue Strength. Fatigue
Strength will be one of the most important factors in the decision because the material
will have to withstand over 1 billion cycles. Young’s Modulus was chosen because it is
closely related to Shear Modulus through the equation:
Shear Modulus: G = E / [2(1+v)]
v = poisson’s ratio = - εt / ε = (lateral or transverse strain) / (axial strain)
For elastomers v = ~ ½ ( G = ~ .333333E)
After research on availability of materials and consideration of the CES plot (Figure 2),
Thermo Polyurethane and PDMS were chosen for testing.
Figure 1. Elastomer molecular chain representation of an unloaded elastomer (A) and loaded elastomer (B).1
Thermo Polyurethane:
(B)
Thermo Polyurethane (Figure 3) polymers are formed through step-growth
polymerization.
Step-growth polymerization refers to a type of polymerization mechanism in which bi-
functional or multifunctional monomers react to form first dimers, then trimers, longer
oligomers and eventually long chain polymers. Many of these are naturally occurring
polymers but some synthetic polymers exist such as polyesters, polyamides,
polyurethanes, and many more. Due to the nature of the polymerization mechanism, a
high extent of reaction is required to achieve high molecular weight.
Figure 2. Biomaterials database CES plot comparing Fatigue Strength v. Young’s Modulus.
Figure 3. Monomer of a polyurethane molecule. 5
The polyurethane chains are complex structures. Due to the presence of benzene rings,
the structure has areas of hard and soft areas within the chains. This results in a
structure that will organize into stronger, less flexible areas and areas that are weak and
elastic. The stiffer areas are the result of the benzene rings from multiple chains lining up
and stacking on top of each other. The weaker areas in the material are the the areas
where the benzene rings have not lined up, and they form a regular disordered
elastomer structure. These softer areas will stretch and result in the elastic properties of
polyurethane, where the benzene alignment is the result in the material’s strength. When
stretched, the soft areas that contain a double bonded oxygen molecule form hydrogen
bonds with methyl’s from other chains within the structure.
Polydimethylsiloxane (PDMS):
(C)
Silicones are inert synthetic compounds, formed through chain growth polymerization.
Chain growth polymerization is when unsaturated monomer molecules add onto a
growing polymer chain one at a time. The structure consists of an -O-Si-O-Si-
“backbone” replacing the common -C-C-C-C- in carbon-based elastomers. This results
in a linear polymer with lower bond angles than the common carbon backbone. This
structure results in a viscous polymer that requires the polymer chains to be crosslink in
order to form silicone rubber and useful in engineering applications. In crosslinking,
methyl groups are substituted by vinyl groups to form crosslinking sites between
entangled chains. They have good stability of rubber properties over large temperature
range of -50C to 200C. Silicones are chemically resistant and have good sealing
capability. They are commonly used in biomedical applications for seals and o-rings.
Figure 4. Monomer of PDMS. 6
Silicone chains have a simple structure with low bond angularity and evenly bonded
methyl-groups that surround the chains. These molecular properties result linear
molecules. Silicone molecules will thus slide past each other very easily when for
example a tensile load is applied. This is why the material results in a lower elastic
modulus than that of polyurethane.
Testing
Objective
The goal of our testing procedures was to obtain values of elastic modulus for
elastomers PDMS and TPU and compare these values to those recommended for
elastomers being used in this application. For elastomers, the shear modulus can be
approximated to be one third of the elastic modulus. It was also an objective to observe
if these values for elastic modulus would be changed after putting the samples through
multiple cycles of fatigue.
Design of Experiment
Values for elastic modulus would be obtained using an Instron Tester. An instron tester
that was more sensitive to strain would have been ideal but was unavailable. The
variables in the experiment were chosen to be materials, amount of cyclic fatigue, and
thickness of elastomer film. Controls included shape of the sample, temperature, and
rate of deformation. These values can be seen in:
Input Variables
Factors: Levels
Materials: PDMS and TPU
Thickness: .01 in and .02 in
Fatigue: 0 cycles and 8,000 cycles
Fatigued Samples
Samples subjected to shear bending for 8,000 cycles, a cycle being one shear
bend to 90 degrees and back to 0 degrees.
Controls
Rate of deformation: 500mm/min
Geometry of sample: Gage length: 1 inch Width: 1 inch
Temperature: 25 C
Response Variable
Elastic modulus
Expected Results
For a material to be considered for the heart valve application, it must have no
difference in elastic modulus between fatigued and unfatigued samples. It is
expected that the PDMS elastomer will have no significant change in elastic
modulus due to its ease of chains sliding past one another at low stresses, and
that it will out-perform the TPU samples..
Results of Test
A statistical analysis of our results in terms of main effects and interactions of variables
can be seen below.
Both of these plots show that the only factor that had a significant effect on our data was
the material of the samples. When the values we obtained were compared to the
recommended values for elastomers in this application, both of our materials fell short.
Obtained E Values:
PDMS = 0.003 MPa
TPU = 0.15 MPa
Recommended elastic modulus values[F] for heart valve leaflet:
> .8 MPa
< 12 MPa
Discussion
Our results did show, as expected, that PDMS has a lower modulus than TPU, but that
is as far as we can go with our own conlusions due to lack of power in our experiment.
The obtained values for the elastic modulus of the polymers did not compare closely to
the recommended values for elastomeric elastic modulus[F], nor the pericardial tissue
values. This result is likely due to the location of the obtained slope on the stress strain
diagram. Figure X depics one of our obtained stress strain diagrams for TPU. As shown
in the figure, the difference the location of the line results in a very different slope. This
difference results in our obtained values of elastic modulus being much lower than the
recommended. There are few to little experiments on elastomers that depict an elastic
modulus from the initial slope (approx. <25% strain) which is the area in question for our
application.
Figure The figure above depicts one of the trials of tensile testing TPU. Line A portrays the slope that was used to
obtain our value of the elastic modulus. Line B is the line that was likely used to obtain the recommended values for
elastic modulus, as well as the E values found in CES.
Because of this uncertainty and our low sample size in testing, there is considerable
room for improvement in our methods and results. A promising place to begin improving
would be to establish a better foundational knowledge of he high-cycle fatigue behavior
of elastomers. There is very little information in this area however because the vast
majority of elastomers are not used in applications where cyclic fatigue is a factor.
It is also possible that we could have obtained more valid results by changing our test
method. This would include testing more samples of each material and testing them
differently. As opposed to simple tensile tests, generation of hysterisis curves for the
elastomers would yield useful information concerning cyclic elastic loading.
Conclusions
Neither PDMS nor TPU can be accepted for use as heart valve leaflets based on our
findings. However, these materials should not necessarily be ruled out either. With more
accurate testing tools and methods, valid elastic modulus values could have been
attained, which could yield a definitive answer as to whether our samples of PDMS and
TPU would exhibit acceptable mechanical properties for use in heart valves.
We suggest further mechanical testing, and once a material is found with sufficient
mechanical properties we suggest testing on coagulation properties.
Sources
1Chemical Composition and Structure of Elastomers." Elastomer Chemistry. 19 Feb. 2011.
<http://www.standard-gasket.com/tech_specs/elastomer_chemistry.htm>.
2“Elastomeric Sheet Materials for Heart Valve and Other Prosthetic Implants.” US Patent, July
20, 1982.
3"Heart Disease: Heart Valve Disease." MedicineNet. 22 Feb. 2011.
<http://www.medicinenet.com/heart_valve_disease/article.htm>.
4 Interview with Dr. Luke Faber on February 3, 2011
5"Polyurethane." Wikipedia The Free Encyclopedia. 3 Mar. 2011. Wikipedia.
<http://en.wikipedia.org/wiki/Polyurethane>.
6"Silicone Rubber." Caojunbang. 11 Feb. 2011. <http://caojunbang.centerblog.net/5-107-rtv-
silicone-rubber>.