RELIABLE
CONVENIENT
EFFICIENT
• Sensitivity: 94,4%• Specificity: 97,3%
• Sample collected by supplied sterile swab• 2 step-test with results in 10 mn
• Shelf-life 24 months• Room temperature storage
BIOSYNEX® STREP A
Immunochromatographic rapid test for qualitative detection of Group A
Streptococcus antigens in throat swab specimens
BACTER IOLOGY
Relative sensitivity: 94.4%Relative specificity: 97.3%The analytical sensitivity of the assay is 1x 105 bacteria / swab.
Kit content• BIOSYNEX® Strep A test devices• Extraction solution 1 (2M NaNO2)• Extraction solution 2 (0.027M citric acid)• Positive control (non-viable Strep A ; 0.09% NaN3)• Negative control (non-viable Strep C ; 0.09% NaN3)
Streptococcus pyogenes are non-motile gram-positive cocci, which can colonize human organs and cause serious infections. Beta-haemolytic Group A Streptococci (Streptococcus pyogenes) are the main cause for infections of the upper respiratory tracts, especially angina. An early diagnosis and treatment of Group A Streptococcal angina reduces symptoms severity and further complications, such as rheumatic fever and glomerulonephritis.
Performances
Procedure & interpretation
POSITIVE: 2 lines appear. Any shade of colour in the test area should be considered as positive.
NEGATIVE: One line appears in the control line area; no line appears in the test line area.
INVALID: control line fails to appear.
THE SOLUTION
BIOSYNEX SA22 Boulevard Sébastien Brant
67400 ILLKIRCH-GRAFFENSTADEN - FranceTél. : +33 3 88 77 57 52 - Fax : +33 3 88 78 76 78
E-mail : [email protected]
Carefully read the instructions on thelabeling and / or reagent notices.
EXP-
v1
- 01
/201
8
Reference
• Extraction tubes with dropper caps• Sterile swabs• Reagent holder
• BIOSYNEX Strep A 20 tests Ref 1030005 BAC T E R I O LOGY
4 x 4 x
2. Deposit 4 drops of extraction reagent 1 in the extraction tube
3. Add 4 drops of extraction reagent 2 and mix
4. Add the swab into the extraction tube and shake vigorously, incubate for 1 to 5 mn
5. Collect the volume with the dropper then add 3 drops in the well S of the cassette
6. Read the results after 5 mn, do not read after 10 mn
1. Collect the throat sample using the swab provided with the kit