+ All Categories
Home > Documents > Bipolar Radiofrequency Ablation Using Dual Internally Cooled Wet … · 2015-09-17 · Bipolar...

Bipolar Radiofrequency Ablation Using Dual Internally Cooled Wet … · 2015-09-17 · Bipolar...

Date post: 05-Apr-2020
Category:
Upload: others
View: 2 times
Download: 0 times
Share this document with a friend
7
submit.radiology.or.kr J Korean Soc Radiol 2012;67(6):425-431 425 INTRODUCTION ere are two kinds of radiofrequency (RF) systems, one is monopolar and the other is bipolar. Monopolar RF system uses one electrode with large dispersive pads. Bipolar RF system usually uses dual electrodes without dispersive pads, but some- times uses only one single electrode with 2 poles, separated by insulation. Currently, monopolar RF system is widely and more popularly used, however, several experimental reports revealed that bipolar RF system induced larger volumes of coagulation, compared with monopolar RF system (1, 2). Bipolar RF system has the theoretical advantage that a localized circumscribed electrical field has a low risk of unwanted distribution of the electric field, resulting in a possibility of heating metallic im- planted devices of pacemaker, surgical clips or stents (3). Bipo- lar RF system also doesn’t need a dispersive pad. As such, pa- tient preparation is simpler than monopolar RF system, and there is no risk of skin burn at this ground pad. e purpose of this study is to determine the optimized protocols of bipolar RF ablation with electrodes dual internally cooled wet (ICW) elec- trode. MATERIALS AND METHODS Radiofrequency Systems Our bipolar radiofrequency ablation (RFA) was comprised with the RF generator (M-2004, RF medical Co., Seoul, Korea), providing a maximum power of 200 W, two ICW electrodes with 3 cm active tip (Jet-tip, RF medical, Seoul, Korea) and two peristaltic cooling pumps (RFP-100, variable-speed pump; RF medical Co., Seoul, Korea). is ICW electrode has one coaxial inflow lumen and outer outflow tracts with a micro-hole locat- ed in the middle of the active tip (Fig. 1A). One electrode was connected to the generator radiofrequency output, and the oth- er was connected to the generator ground output for bipolar ra- diofrequency circuit without a dispersive pad of the monopolar Original Article pISSN 1738-2637 J Korean Soc Radiol 2012;67(6):425-431 Received July 17, 2012; Accepted September 22, 2012 Corresponding author: Jae Young Byun, MD Department of Radiology, Seoul St. Mary’s Hospital, The Catholic University of Korea, College of Medicine, 222 Banpo-daero, Seocho-gu, Seoul 137-701, Korea. Tel. 82-2-2258-6235 Fax. 82-2-599-6771 E-mail: [email protected] Copyrights © 2012 The Korean Society of Radiology Purpose: To determine the optimized protocol for bipolar radiofrequency ablation (RFA), using dual internally cooled wet (ICW) electrodes in the ex vivo bovine liver. Materials and Methods: RFA was applied to the explanted bovine liver, using two 3 cm active tip electrodes with 3.5 cm spacing. A total of 25 ablation zones were created by five groups; group A: 70 W-20 minute (min), group B: 70 W-25 min, group C: 90 W-15 min, group D: 90 W-20 min, and group E: 90 W-25 min. We measured the total energy and size of ablation zones with a color of grey or pink. Statistical analysis was done using Kruskal Wallis test and Mann Whitney U-test. Results: The mean energy, mean volume of ablation zone with grey and pink color of groups A to E were 16.7, 23.9, 16.7, 21.8, 29.2 kcal, 25.7, 34.3, 29.5, 36.2, 45.2 cm 3 , and 60.0, 88.0, 71.5, 87.4, 104.5 cm 3 , respectively. Those were significantly dif- ferent ( p < 0.05). The volume of ablation zone of group E with grey color was larger than groups A, B and C ( p < 0.05). Conclusion: Bipolar RFA, using dual ICW electrodes, can produce a large ablation zone with the protocol of 90 W-25 min. Index terms Radiofrequency Ablation Bipolar Liver Ex Vivo Experiment Study Bipolar Radiofrequency Ablation Using Dual Internally Cooled Wet Electrodes: Experimental Study in Ex Vivo Bovine Liver 두 개의 내부냉각습식 전극을 사용한 양극성 고주파열치료: 체외 소간 실험 연구 Young Joon Lee, MD, Jae Young Byun, MD Department of Radiology, Seoul St. Mary’s Hospital, The Catholic University of Korea, College of Medicine, Seoul, Korea
Transcript

submit.radiology.or.kr J Korean Soc Radiol 2012;67(6):425-431 425

INTRODUCTION

There are two kinds of radiofrequency (RF) systems, one is monopolar and the other is bipolar. Monopolar RF system uses one electrode with large dispersive pads. Bipolar RF system usually uses dual electrodes without dispersive pads, but some-times uses only one single electrode with 2 poles, separated by insulation. Currently, monopolar RF system is widely and more popularly used, however, several experimental reports revealed that bipolar RF system induced larger volumes of coagulation, compared with monopolar RF system (1, 2). Bipolar RF system has the theoretical advantage that a localized circumscribed electrical field has a low risk of unwanted distribution of the electric field, resulting in a possibility of heating metallic im-planted devices of pacemaker, surgical clips or stents (3). Bipo-lar RF system also doesn’t need a dispersive pad. As such, pa-tient preparation is simpler than monopolar RF system, and there is no risk of skin burn at this ground pad. The purpose of

this study is to determine the optimized protocols of bipolar RF ablation with electrodes dual internally cooled wet (ICW) elec-trode.

MATERIALS AND METHODS

Radiofrequency Systems

Our bipolar radiofrequency ablation (RFA) was comprised with the RF generator (M-2004, RF medical Co., Seoul, Korea), providing a maximum power of 200 W, two ICW electrodes with 3 cm active tip (Jet-tip, RF medical, Seoul, Korea) and two peristaltic cooling pumps (RFP-100, variable-speed pump; RF medical Co., Seoul, Korea). This ICW electrode has one coaxial inflow lumen and outer outflow tracts with a micro-hole locat-ed in the middle of the active tip (Fig. 1A). One electrode was connected to the generator radiofrequency output, and the oth-er was connected to the generator ground output for bipolar ra-diofrequency circuit without a dispersive pad of the monopolar

Original ArticlepISSN 1738-2637J Korean Soc Radiol 2012;67(6):425-431

Received July 17, 2012; Accepted September 22, 2012Corresponding author: Jae Young Byun, MDDepartment of Radiology, Seoul St. Mary’s Hospital, The Catholic University of Korea, College of Medicine, 222 Banpo-daero, Seocho-gu, Seoul 137-701, Korea. Tel. 82-2-2258-6235 Fax. 82-2-599-6771E-mail: [email protected]

Copyrights © 2012 The Korean Society of Radiology

Purpose: To determine the optimized protocol for bipolar radiofrequency ablation (RFA), using dual internally cooled wet (ICW) electrodes in the ex vivo bovine liver.Materials and Methods: RFA was applied to the explanted bovine liver, using two 3 cm active tip electrodes with 3.5 cm spacing. A total of 25 ablation zones were created by five groups; group A: 70 W-20 minute (min), group B: 70 W-25 min, group C: 90 W-15 min, group D: 90 W-20 min, and group E: 90 W-25 min. We measured the total energy and size of ablation zones with a color of grey or pink. Statistical analysis was done using Kruskal Wallis test and Mann Whitney U-test.Results: The mean energy, mean volume of ablation zone with grey and pink color of groups A to E were 16.7, 23.9, 16.7, 21.8, 29.2 kcal, 25.7, 34.3, 29.5, 36.2, 45.2 cm3, and 60.0, 88.0, 71.5, 87.4, 104.5 cm3, respectively. Those were significantly dif-ferent (p < 0.05). The volume of ablation zone of group E with grey color was larger than groups A, B and C (p < 0.05).Conclusion: Bipolar RFA, using dual ICW electrodes, can produce a large ablation zone with the protocol of 90 W-25 min.

Index termsRadiofrequency AblationBipolarLiverEx VivoExperiment Study

Bipolar Radiofrequency Ablation Using Dual Internally Cooled Wet Electrodes: Experimental Study in Ex Vivo Bovine Liver두 개의 내부냉각습식 전극을 사용한 양극성 고주파열치료: 체외 소간 실험 연구 Young Joon Lee, MD, Jae Young Byun, MDDepartment of Radiology, Seoul St. Mary’s Hospital, The Catholic University of Korea, College of Medicine, Seoul, Korea

Bipolar Radiofrequency Ablation Using Dual Internally Cooled Wet Electrodes

submit.radiology.or.krJ Korean Soc Radiol 2012;67(6):425-431426

Radiofrequency Protocol

RFA was performed in the fresh explanted bovine livers. Two ICW electrodes with 3 cm active tip were placed parallel into the liver vertically with 3.5 cm spacing at room temperature (Fig. 1B). Prior to placing the electrode, we checked the path of a large vessel in the explanted bovine liver with a long screw driver for avoiding inadequate ablation by leakage of perfused normal saline. As a pilot study, RFA was done with two proto-cols of 60-200 W (Power automatically increase 20 W every minute) for 15 min and 70 W for 15 min. The results of the pi-lot study showed an incomplete ablation between the two elec-trodes. We also found that the power could not reach more than 120 W because the impedance quickly elevated during 120 W power. We decided five groups of different power out-put and duration of RFA. Group A was 70 W power for 20 min duration of RFA. Group B was 70 W power for 25 min dura-tion. Group C was 90 W power for 15 min duration. Group D was 90 W power for 20 min duration. Group D was 90 W pow-er for 25 min duration. We performed five RF ablations in each groups. Total numbers of RFA were 25.

Measurement of Ablation Size

RF ablated lesions were sliced along the electrode insertion axis and then cut perpendicular plane at the middle, between the two electrodes. We measured both the grey colored central area of the ablation zone only, including the marginal pink col-ored area.

Dt is the long transverse diameter between the two elec-trodes. Ds is the short transverse diameter of the perpendicular plane of the two electrodes. Dv is the longitudinal diameter, along the axis of electrodes (Fig. 2). Three dimensional volume measurement of ablation size was done by Dt × Ds × Dv × π/6.

Statistical Analysis

Statistical analysis of the difference in size of ablation and ap-plied total energy among the groups was done using Kruskal Wallis test and Mann Whitney U-test. Spearman correlation analysis is used for the correlation between the total energy and the volume of the ablation zone. A p-value of less than 0.05 was considered as statistically significant. SPSS software (Version 13.0, SPSS Inc., Chicago, IL, USA) was used for all statistical analysis.

system. When the impedance of the electrodes rises 30% more than the baseline, the generator automatically shuts down the power for 15 seconds. A 0.9% normal saline was used for the internal cooling and perfused through the micro-hole. The tis-sue infusion rate was 1.2 mL/min. We monitored the power output, impedance and temperature of the electrode with a computer software that was programmed by RF medical Co. Applied energy (kcal) was calculated by 0.000239 × Watt (W) × second (s).

Fig. 1. Schematic diagram. A. Internal structure of internally cooled wet tip electrode with cool-ing and perfused flow (dotted arrows). B. Bipolar radiofrequency ablation system with single generator and two infusion pumps. Note.-RF = radiofrequency

A

B

A

BFig. 2. Measurement of the ablation with grey color. A. Dt is the long transverse diameter between the two electrodes and Dv (long dotted arrow) is the longitudinal diameter along the axis of electrodes. B. Ds (short dotted arrow) is the short transverse diameter of perpen-dicular plane at the middle between the two electrodes.

Perfused flow

Outflow

Outflow

RF generator

3.5 cm

Pump

Pump

Saline inflow

Micro-hole

0.9%NaCI

Young Joon Lee, et al

submit.radiology.or.kr J Korean Soc Radiol 2012;67(6):425-431 427

total energy of each group were seen in Table 1. All these pa-rameters of the ablation zone, volume and total energy were significantly different among the group by Kruskal Wallis test (p < 0.05). The Dt of the ablation zone with grey color of group B and E was longer than group A and C (p < 0.05), but there

RESULTS

The mean applied energy in group A to E was 16.7, 23.9, 16.7, 21.8 and 29.2 kcal, respectively. The average Dt, Ds, Dv of the ablation zone and volume with grey color and pink color and

A

D

B

E

C

Fig. 3. Cut surfaces of coagulated necrosis along the axis of electrodes in the five groups. (A) group A; 70 W-20 min, (B) group B; 70 W-25 min, (C) group C; 90 W-15 min, (D) group D; 90 W-20 min, (E) group E; 90 W-25 min: Long transverse diameter between the two electrodes (arrows) of necrosis of 25 min groups (group B and E) is longer than that of group A and C. Longitudinal diameter along the axis of electrode of necrosis of group E is longer than that of group A, B, and C.

Table 1. Results of Bipolar RFA with Dual Internally Cooled Wet Tip Electrodes in Five Groups

GroupA, 70 W-20 min B, 70 W-25 min C, 90 W-15 min D, 90 W-20 min E, 90 W-25 min

p ValueMean (SD) Mean (SD) Mean (SD) Mean (SD) Mean (SD)

Dt Grey 4.4 (0.2) 4.8 (0.2) 4.8 (0.3) 4.8 (0.3) 5.2 (0.3) p < 0.05*Pink 5.9 (0.2) 6.3 (0.3) 6.2 (0.3) 6.2 (0.4) 6.6 (0.3)

Ds Grey 3.2 (0.2) 3.6 (0.2) 3.1 (0.1) 3.6 (0.1) 3.8 (0.3) p < 0.05†

Pink 4.3 (0.3) 5.2 (0.4) 4.4 (0.2) 4.9 (0.3) 5.2 (0.4)Dv Grey 3.5 (0.3) 3.8 (0.1) 3.7 (0.4) 4.0 (0.4) 4.3 (0.2) p < 0.05‡

Pink 4.6 (0.3) 5.2 (0.2) 5.0 (0.4) 5.5 (0.5) 5.8 (0.3)Vol Grey 25.7 (2.4) 34.3 (3.0) 29.5 (5.3) 36.2 (5.8) 45.2 (6.8) p < 0.05§

Pink 60.0 (8.0) 88.0 (10.8) 71.5 (10.2) 87.4 (13.9) 104.5 (14.5)TE kcal 16.7 (1.4) 23.9 (2.2) 16.7 (2.2) 21.8 (2.4) 29.2 (3.1) p < 0.05||

Note.-*Difference in the groups (A, C) and (B, E).†Difference in the groups (A, C) and (B, D, E).‡Difference in the (A, B, C) and (E).§Difference in the (A, B, C) and (E) with grey color and in the (A, C) and (B, D, E) with pink color.||Difference in the (A, B, C, D) and (E).min = minutes, RFA = radiofrequency ablation, SD = standard deviation, TE = total energy, Vol = volume

Bipolar Radiofrequency Ablation Using Dual Internally Cooled Wet Electrodes

submit.radiology.or.krJ Korean Soc Radiol 2012;67(6):425-431428

DISCUSSION

During the last 10-year period, local tumor ablation methods for treatment of hepatocellular carcinoma (HCC) were used more and more. Among the various local tumor ablation meth-ods, those percutaneous ethanol injection (PEI) and thermal based RFA are recognized as a curative treatment of less than three and less than 3 cm sized early stage HCC (4). It is impor-tant to increase the volume of ablation for reducing local tumor recurrence and treating intermediate, 3 to 5 cm sized HCC. Volume of coagulation can be increased with the methods of overlapping (5, 6) or combined therapy with transarterial che-moembolization, PEI, and chemotherapeutic agent (7-10). There have been many studies to increase the ablation size with various types of electrodes, including internally cooled and perfused electrodes (11-14), clustered electrode or multiple electrodes (3, 15-20).

At this time, as far as we know, only one bipolar RF system is commercially available and that is composed of three internally cooled bipolar electrodes (CelonProSurge; Celon Medical In-

was no significant difference between group B and group E (Fig. 3). The Ds of the ablation zone with grey color was signifi-cantly longer in groups B, D and E than in groups A and C (p < 0.05) (Fig. 4). There was also no significant difference in Ds among the groups B, D and E. Dv of the ablation zone with grey color was longer in group E than in groups A to C (p < 0.05) (Fig. 3). The mean volume of the ablation zone with grey color was 25.7, 34.3, 29.5, 36.2 and 45.2 cm3 in groups A to E. The mean volume of the ablation zone with pinkish color in each group was 60.0, 88.0, 71.5, 87.4 and 104.5 cm3, respective-ly. There was strong positive linear correlation between the to-tal energy and the volume of the ablation zone with grey (rho = 0.877, p < 0.05) and pink (rho = 0.88, p < 0.05) color. The vol-ume of the ablation zone of group E with grey color was larger than groups A, B and C (p < 0.05), but not than group D. Vol-ume of the ablation zone of groups A and C with pink color was smaller to groups B, D and E (p < 0.05). Total energy of group E was significantly higher than the others (p < 0.05). To-tal energy was also significantly higher in the longer time group with the same power output.

A

D

B

E

C

Fig. 4. Cut surfaces of coagulated necrosis, perpendicular plane at the middle between the two electrodes in the five groups. (A) group A; 70 W-20 min, (B) group B; 70 W-25 min, (C) group C; 90 W-15 min, (D) group D; 90 W-20 min, (E) group E; 90 W-25 min: Short transverse diam-eter of necrosis of 15 min duration groups (group A and C) is shorter than that of 20 and 25 min duration groups (group B, D, and E).

Young Joon Lee, et al

submit.radiology.or.kr J Korean Soc Radiol 2012;67(6):425-431 429

charring more efficiently than that of internal cooling only. In our results, total ablation volume with pink color was at

least more than two times than that with grey color. This large ablation volume with pink color could have a chance to com-plete necrosis with the combination of lyso-thermosensitive li-posomal doxorubicin, which consists of the heat-enhanced cy-totoxic doxorubicin within a heat-activated liposome (10), which has been evaluated in randomized, double-blind, dum-my-controlled trial (23).

There are several limitations in this experimental study. First, number of ablated ex vivo bovine liver in this study was small. Second, ablated volume of ex vivo is usually larger than that of in vivo because the heat sink effect, which is induced by an ad-jacent blood flow in the large vessels (24). Third, variable size or shape of coagulation can be achieved in vivo (25). Fourth, electrical conductivity or impedance of tumor tissue could be different from that of normal tissue.

In conclusion, even though there are several limitations, bipo-lar RFA using dual ICW electrodes can produce large ablation zone with the protocol of 90 W for 25 minutes among several preselected fixed power output and duration. Further evaluation and optimization of the bipolar RFA using dual ICW electrodes in vivo are needed.

REFERENCES

1.HaemmerichD,StaelinST,TungjitkusolmunS,LeeFTJr,

MahviDM,WebsterJG.Hepaticbipolarradio-frequency

ablationbetweenseparatedmultiprongelectrodes.IEEE

TransBiomedEng2001;48:1145-1152

2.LeeJM,HanJK,KimSH,SohnKL,LeeKH,AhSK,etal.A

comparativeexperimentalstudyofthein-vitroefficiencyof

hypertonicsaline-enhancedhepaticbipolarandmonopolar

radiofrequencyablation.KoreanJRadiol2003;4:163-169

3.ClasenS,SchmidtD,DietzK,BossA,KröberSM,Schraml

C,etal.Bipolarradiofrequencyablationusinginternally

cooledelectrodesinexvivobovineliver:predictionofco-

agulationvolumefromappliedenergy.InvestRadiol2007;

42:29-36

4.LlovetJM,BurroughsA,BruixJ.Hepatocellularcarcinoma.

Lancet2003;362:1907-1917

5.DoddGD3rd,FrankMS,AribandiM,ChopraS,Chintapalli

struments, Telow, Germany). Seror et al. (21) reported multi-polar RF ablation with three internally cooled bipolar elec-trodes could ablate completely and safely HCCs larger than 5.0 cm, even those that were infiltrative and involved a segmental portal vein. Clasen et al. (22) evaluated the relationship be-tween parameters of bipolar RFA, using internally cooled elec-trodes and reported that prolonged RFA at lower power out-puts produce larger volumes of coagulation, and faster RFA at higher power outputs produces smaller volume of coagulation and also revealed that volume of coagulation was directly pro-portional to the duration and the amount of energy application (3). In our study, maximum power output was not investigated because high power, especially more than 120 W power, induced rapid increase of tissue impedance, resulting in less duration of energy application and incomplete fusion of the coagulation zone between the two electrodes by our pilot experiment. In our results, the amount of energy application proportionally corresponded to the volume size of coagulation, and the total energy with 90 W power in 25 minutes was significantly larger than 70 W power in the same minutes, and the volume was also significantly different between those two groups (p = 0.008). We thought that there should be a threshold of power output, which induces rapid increase of tissue impedance and the 90 W power output in our experiment must be lower than the threshold inducing rapid increase of tissue impedance.

Lee et al. (17) reported serially that saline enhanced bipolar RFA was more efficient in making larger volume of thermal ab-lation than conventional and simultaneous monopolar RFA and bipolar RFA with the cooled-wet electrode also produce larger tissue coagulation than with the open-perfused electrode (18). They had used a prototype cooled-wet electrode with a coaxial saline perfusion system, but that type of electrode has not been commercially available now. We used ICW electrode, which induced significantly larger volume of coagulation than the internally cooled electrode in the monopolar RF system (14). In our bipolar RFA system with ICW electrode, perfused normal saline with simultaneous internally cooling effect made the tissue impedance low, which made longer duration of ener-gy application than the internally cooled bipolar electrodes in the higher power output contrast to the study of Clasen et al. (22). Another function of interstitially perfused cool saline around the electrode prevented active tip of electrode from

Bipolar Radiofrequency Ablation Using Dual Internally Cooled Wet Electrodes

submit.radiology.or.krJ Korean Soc Radiol 2012;67(6):425-431430

porcinelivers:comparisonoftheuseofinternallycooled

electrodesandinternallycooledwetelectrodes.Cardio-

vascInterventRadiol2009;32:1235-1240

15.LeeJM,HanJK,KimHC,ChoiYH,KimSH,ChoiJY,etal.

Switchingmonopolarradiofrequencyablationtechnique

usingmultiple, internallycooledelectrodesandamulti-

channelgenerator:exvivoandinvivopilotstudy.Invest

Radiol2007;42:163-171

16.LeeJM,HanJK,LeeJY,KimSH,ChoiJY,LeeMW,etal.He-

paticradiofrequencyablationusingmultipleprobes:ex

vivoandinvivocomparativestudiesofmonopolarversus

multipolarmodes.KoreanJRadiol2006;7:106-117

17.LeeJM,KimSH,HanJK,SohnKL,ChoiBI.Exvivoexperi-

mentofsaline-enhancedhepaticbipolarradiofrequency

ablationwithaperfusedneedleelectrode:comparison

withconventionalmonopolarandsimultaneousmonopo-

larmodes.CardiovascInterventRadiol2005;28:338-345

18.LeeJM,HanJK,KimSH,SohnKL,ChoiSH,ChoiBI.Bipolar

radiofrequencyablationinexvivobovineliverwiththe

open-perfusedsystemversusthecooled-wetsystem.Eur

Radiol2005;15:759-764

19.HaemmerichD,LeeFTJr,SchuttDJ,SampsonLA,Webster

JG,FineJP,etal.Large-volumeradiofrequencyablationof

exvivobovineliverwithmultiplecooledclusterelectrodes.

Radiology2005;234:563-568

20.LeeJM,HanJK,KimSH,LeeJY,ChoiSH,ChoiBI.Hepatic

bipolar radiofrequencyablationusingperfused-cooled

electrodes:acomparativestudyintheexvivobovineliver.

BrJRadiol2004;77:944-949

21.SerorO,N’KontchouG,IbraheemM,AjavonY,BarrucandC,

GanneN,etal.Large(>or=5.0-cm)HCCs:multipolarRFab-

lationwiththreeinternallycooledbipolarelectrodes--ini-

tialexperiencein26patients.Radiology2008;248:288-296

22.ClasenS,SchmidtD,BossA,DietzK,KröberSM,Claussen

CD,etal.Multipolarradiofrequencyablationwithinter-

nallycooledelectrodes:experimentalstudyinexvivobo-

vineliverwithmathematicmodeling.Radiology2006;238:

881-890

23.PoonRT,BorysN.Lyso-thermosensitiveliposomaldoxorubi-

cin:anadjuvanttoincreasethecurerateofradiofrequency

ablationinlivercancer.FutureOncol2011;7:937-945

24.BitschRG,DüxM,HelmbergerT,LubienskiA.Effectsof

KN.Radiofrequencythermalablation:computeranalysis

ofthesizeofthethermalinjurycreatedbyoverlapping

ablations.AJRAmJRoentgenol2001;177:777-782

6.ChoiD,LimHK,KimMJ,LeeJ,KimSK,KimEY,etal.Over-

lappingablationusingacoaxialradiofrequencyelectrode

andmultiplecannulaesystem:experimentalstudyinex-

vivobovineliver.KoreanJRadiol2003;4:117-123

7.PoggiG,GattiC,TeragniC,DelmonteA,BernardoG.Ra-

diofrequencyablationcombinedwithpercutaneousetha-

nolinjectioninthetreatmentofhepatocellularcarcinoma

andportalveinneoplastic thrombosis.AnticancerRes

2004;24:2419-2421

8.VeltriA,MorettoP,DoriguzziA,PaganoE,CarraraG,Gan-

diniG.Radiofrequencythermalablation(RFA)aftertrans-

arterialchemoembolization(TACE)asacombinedtherapy

forunresectablenon-earlyhepatocellular carcinoma

(HCC).EurRadiol2006;16:661-669

9.MorimotoM,NumataK,KondouM,NozakiA,MoritaS,

TanakaK.Midtermoutcomesinpatientswithintermedi-

ate-sizedhepatocellularcarcinoma:arandomizedcon-

trolledtrialfordeterminingtheefficacyofradiofrequency

ablationcombinedwithtranscatheterarterialchemoem-

bolization.Cancer2010;116:5452-5460

10.PoonRT,BorysN.Lyso-thermosensitiveliposomaldoxoru-

bicin:anovelapproachtoenhanceefficacyofthermal

ablationoflivercancer.ExpertOpinPharmacother2009;

10:333-343

11.ZlottaAR,WildschutzT,RavivG,PenyMO,vanGansbeke

D,NoelJC,etal.Radiofrequencyinterstitialtumorabla-

tion(RITA)isapossiblenewmodalityfortreatmentofre-

nalcancer:exvivoand invivoexperience.JEndourol

1997;11:251-258

12.MiaoY,NiY,YuJ,ZhangH,BaertA,MarchalG.Anexvivo

studyonradiofrequencytissueablation:increasedlesion

sizebyusingan“expandable-wet”electrode.EurRadiol

2001;11:1841-1847

13.LeeJM,HanJK,KimSH,LeeJY,ShinKS,HanCJ,etal.Op-

timizationofwetradiofrequencyablationusingaper-

fused-cooledelectrode:acomparativestudy inexvivo

bovinelivers.KoreanJRadiol2004;5:250-257

14.ChaJ,ChoiD,LeeMW,RhimH,KimYS,LimHK,etal.Ra-

diofrequencyablationzonesinexvivobovineandinvivo

Young Joon Lee, et al

submit.radiology.or.kr J Korean Soc Radiol 2012;67(6):425-431 431

HölscherAH,BeckurtsKT.Variabilityofsizeandshapeof

necrosisinducedbyradiofrequencyablationinhumanlivers:

avolumetricevaluation.AnnSurgOncol2004;11:420-425

vascularperfusiononcoagulationsizeinradiofrequency

ablationofexvivoperfusedbovinelivers.InvestRadiol

2006;41:422-427

25.StippelDL,BrochhagenHG,ArenjaM,Hunkemöller J,

두 개의 내부냉각습식 전극을 사용한 양극성 고주파열치료: 체외 소간 실험 연구

이영준 · 변재영

목적: 체외 소간에서 두 개의 내부냉각습식 전극을 사용한 양극성 고주파열치료의 적절한 프로토콜을 정하고자 하였다.

대상과 방법: 생리식염수를 주입한 내부냉각습식 전극을 사용하여 체외 소간에 두 개의 3 cm 활성전극을 3.5 cm 간격

으로 위치한 후 고주파열치료를 시행하였다. 다음과 같이 다섯 개의 그룹으로 나누어 총 25개의 고주파열치료 절제구역

을 만들었다; A 그룹: 70 W-20분, B 그룹: 70 W-25분, C 그룹: 90 W-15분, D 그룹: 90 W-20분, E 그룹: 90

W-25분. 회색 변성이 있는 절제구역의 크기와 분홍색 변성을 포함한 절제구역의 크기 그리고 총 에너지를 측정하였다.

통계분석은 Kruskal Wallis 검정과 Mann Whitney U 검정으로 하였다.

결과: A 그룹에서 E 그룹의 평균 에너지, 회색 변성의 절제구역과 분홍색 변성을 포함한 절제구역의 평균 크기는 각각

16.7, 23.9, 16.7, 21.8, 29.2 kcal, 25.7, 34.3, 29.5, 36.2, 45.2 cm3, 그리고 60.0, 88.0, 71.5, 87.4, 104.5 cm3였

다. 평균 에너지와 절제구역의 크기는 그룹 간에 유의한 차이가 있었다(p < 0.05). E 그룹의 회색 변성의 절제구역의 크

기는 A, B, C 그룹보다는 유의하게 높았다(p < 0.05).

결론: 두 개의 내부냉각습식 전극을 사용한 양극성 고주파열치료는 90 W-25분의 치료에서 큰 절제구역을 생성할 수

있다.

가톨릭대학교 의과대학 서울성모병원 영상의학과


Recommended