+ All Categories
Home > Documents > Bomba Cal 1266

Bomba Cal 1266

Date post: 03-Jun-2018
Category:
Upload: ronal-opazo-salazar
View: 249 times
Download: 4 times
Share this document with a friend

of 113

Transcript
  • 8/12/2019 Bomba Cal 1266

    1/113

    1266Isoperibol Bomb Calorimeter

    No. 367MOperating Instruction Manual

  • 8/12/2019 Bomba Cal 1266

    2/113i-i

    1266 Calorimeter Operating Instruction Manual

    TABLE OF CONTENTS

    Preface

    Scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I

    Getting Started. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II

    Concept of Operation (Chapter 1) 1-1

    Isoperibol Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

    Dynamic Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

    Full Microprocessor Based Process Control . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

    Full Microprocessor Based Data Acquisition and Handling . . . . . . . . . . . . . 1-2

    Flexible Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

    Installing the Calorimeter (Chapter 2) 2-1

    Unpack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

    Install the Water Handling System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1Installing the Water Cooler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2

    Install the Printer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5

    The Balance Port Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5

    Smart Link Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5

    Fill the Water Jacket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5

    Installing the Oxygen Filling System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6

    Standardization. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-7

    Instrument Description (Chapter 3) 3-1

    Main Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1

    Page and Line No. Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1Types of Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1

    Keyboard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2

    Program Installation & Control (Chapter 4) 4-1

    Software Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1

    Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1

    Revised Default Settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1

    Fig. 4-1 Factory Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2

    Operating Instructions (Chapter 5) 5-1

    Operating the 1108 Oxygen Bomb. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1Operating the Filling Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1

    Operating the Calorimeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2

    Operating Suggestions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4

    Combustion Aids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5

    Combustion Capsules. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5

    Coarse Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6

    Explosives and High Energy Fuels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6

    Volatile Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6

    Poor Combustion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8

    Oxygen Charging Pressure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8

  • 8/12/2019 Bomba Cal 1266

    3/113

    1266 Calorimeter Operating Instruction Manual

    i-ii

    TABLE OF CONTENTS

    Corrections & Final Reports (Chapter 6) 6-1

    Entering Corrections and Obtaining the Final Report . . . . . . . . . . . . . . . . . . 6-1

    Manual Entry. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1Fixed Corrections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2

    Automatic Data Transfers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2

    Reporting Instructions (Chapter 7) 7-1

    Report Option Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1

    Report Destination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1

    Report Generation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2

    Memory Management (Chapter 8) 8-1

    Clearing Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1

    Editing Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1

    Maintenance and Trouble Shooting (Chapter 9) 9-1

    Oxygen Bomb. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1

    Installing Support Rod Mounting Plates . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2

    1266 Calorimeter Controller and Keyboard Display Removal . . . . . . . . . . . . 9-3

    Keyboard/Display Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3

    Removing the Entire Controller from the Calorimeter. . . . . . . . . . . . . . . . . . 9-3

    1266/1356 Series Calorimeter Firmware Update Procedure . . . . . . . . . . . . . 9-4

    Acquire the Program Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4

    Connect the Calorimeter Controller to the PC . . . . . . . . . . . . . . . . . . . . . . . . 9-4

    Update the Calorimeter Controller Program . . . . . . . . . . . . . . . . . . . . . . . . . 9-4Troubleshooting the 1266 Calorimeter Controller . . . . . . . . . . . . . . . . . . . . . 9-5

    Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5

    Keyboard Inoperative . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8

    Temperature Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8

    Bomb Firing Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9

    Low Oxygen Pressure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-9

    Bucket Thermistor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-10

    Instructions for Adjusting Cover. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-10

    How to Check the Motor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11

    Motor Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11

    Cooling Coil Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-11

    Pump Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12

    EE Standard Deviation Exceeds Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-12

    Preperiod or Postperiod Time Limit Violation . . . . . . . . . . . . . . . . . . . . . . . 9-12

    Jacket Over Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13

    1266 Calorimeter Error List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-13

  • 8/12/2019 Bomba Cal 1266

    4/113i-iii

    1266 Calorimeter Operating Instruction Manual

    TABLE OF CONTENTS

    Appendix A - Menu Operating Instructions A-1

    Calorimeter Operation Page 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1

    Operating Controls Page 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2Program Information and Control Page 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4

    Calibration Data and Controls Page 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-5

    Thermochemical Calculations Page 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-7

    Data Entry Controls Page 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9

    Reporting Controls Page 7 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-11

    Communication Controls Page 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-13

    Diagnostics Page 9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-16

    Appendix B - Calculations B-1

    Calculating the Heat of Combustion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1

    General Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1Thermochemical Corrections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1

    ASTM and ISO Methods Differ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2

    Fuse Correction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3

    ACID and SULFUR Corrections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-3

    ASTM Treatment for Acid and Sulfur . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5

    ISO Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5

    Calculate HNO3 from the Energy Release . . . . . . . . . . . . . . . . . . . . . . . . . . . B-6

    Correction Equation Set 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-7

    Correction Equation Set 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-7

    Correction Equation Set 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-7

    Spiking Samples. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-8Conversion to Other Moisture Bases. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-8

    Conversion to Net Heat of Combustion (Need New) . . . . . . . . . . . . . . . . . . B-8

    Magnitude of Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-9

    Precision Statements and the Confidence Factor . . . . . . . . . . . . . . . . . . . . . B-9

    Calorimeter Control Limits when Benzoic Acid is Used as a Test Sample . . . . B-10

    Appendix C- Standardization C-1

    Standardizing the Calorimeter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1

    Standard Materials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-1

    Automatic Statistical Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2

  • 8/12/2019 Bomba Cal 1266

    5/113

    1266 Calorimeter Operating Instruction Manual

    i-iv

    TABLE OF CONTENTS

    Appendix D - Communication Interfaces D-1

    Required Software Versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-1

    Required Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-2

    Smart Link Configurations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-2

    Smart Link Configuration Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-2

    Smart Link Configuration Cabling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-3

    RS232C Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-3

    Terminal Port Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-4

    Balance Input Driver Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-4

    Mettler 011/012 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-5

    Sartorius Balance Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-5

    Ohaus Balance Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-6

    Generic Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-6

    Computer Interface Description for the 1266/1356 Calorimeters . . . . . . . . . D-6

    Appendix E - Technical Service E-1

    Appendix F - 1266 Calorimeter Parts List F-1

    Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-1

    Hinge Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-2

    5 Year Recomended Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F-3

    Recommended Spare Parts Per 5000 Tests . . . . . . . . . . . . . . . . . . . . . . . . . . F-4

    Appendix G - 1266 Drawings G-1

  • 8/12/2019 Bomba Cal 1266

    6/113

    Figures

    Fig. 2-1 1552 Water Cooler Electrical Circuits . . . . . . . . . . . . . . . . . . . . . . . . 2-2

    Fig. 2-2 1552 Water Cooler Front View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3

    Fig. 2-3 1552 Water Cooler Top View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3

    Fig. 2-4 Swagelok 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3

    Fig. 2-5 Swagelok 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3

    Fig. 2-6 Swagelok 3. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3

    Fig. 2-7 1266 Calorimeter External Plumbing. . . . . . . . . . . . . . . . . . . . . . . . . 2-4

    Fig. 2-8 1266 Calorimeter Electrical Hookup . . . . . . . . . . . . . . . . . . . . . . . . . 2-4

    Fig. 2-9 1266 Calorimeter Back View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6

    Fig. 3-1 1266 Calorimeter Keyboard. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2

    Fig. 4-1 Factory Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2

    Fig. 5-7 Combustion Capsule with Adhesive Tape Seal. . . . . . . . . . . . . . . . . 5-7

    Fig. 9-1 1108 Oxygen Bomb Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-1

    Tables

    Table 4-1 Calorimeter Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6

    Table 9-2 1266 / 1356 Calorimeter Controller Fuses. . . . . . . . . . . . . . . . . . . . 9-5

    Table 9-3 Power Connections - Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6

    Table 9-4 Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6

    Table 9-5 Power Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6

    Table B-1 Settings for ISO and BSI Methods. . . . . . . . . . . . . . . . . . . . . . . . . B-4

    Table B-2 Entered and Reported Acid Value is Taken as Nitric Acid Only. . . B-6

    Table B-3 Entered and Reported Acid Value is Considered Total Acid . . . . . B-6

    Table B-4 Calculate HNO3 if Fixed and Acid is HNO3 Only . . . . . . . . . . . . . . B-6

    Table B-5 Calculate HNO3 if Fixed and Acid is Total Acid . . . . . . . . . . . . . . . B-6

    Table B-6 Calorimeter Control Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-10

    Table B-7 Calorimeter Control Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-11

    Table D-1 Software Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-3

    Table D-2 Smart Link Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-3

    Table D-3 Calorimeter Test Report Field Definitions . . . . . . . . . . . . . . . . . . . D-7

    Table D-4 Test Information Field #1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-8

    Table D-5 Test Information Field #2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-8

    Table D-6 1266 Terminal Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-9

    Table D-7 1266 Balance Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-9Table D-8 1266 Printer Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . D-9

    i-v

    1266 Calorimeter Operating Instruction Manual

    TABLE OF CONTENTS

  • 8/12/2019 Bomba Cal 1266

    7/113

    This manual contains instructions

    for installing and operating the Parr

    1266 Calorimeter. For ease of use, the

    manual is divided into ten chapters.

    Concept of Operation

    Installation

    Instrument Description

    Program Installation and Control

    Operating Instructions

    Corrections & Final Reports

    Reporting Instructions

    Memory Management

    Maintenance Instructions

    Appendices

    Subsections of these chapters are

    identified in the Table of Contents.

    To assure successful installation

    and operation, the user must study all

    instructions carefully before starting to

    use the calorimeter to obtain an

    understanding of the capabilities of the

    equipment and the safety precautions

    to be observed in the operation.

    Additional instructions concerning

    the installation and operation of

    various component parts and

    peripheral items used with the 1266

    Calorimeter have been included and

    made a part of these instructions.

    No. Description

    201M Limited Warranty

    202M Introduction to Bomb

    Calorimetry

    205M 1108 Oxygen Combustion

    Bomb

    207M Analytical Methods for

    Oxygen Bombs

    245M 1552 Water Cooler

    246M 1563 Water Handling System

    Additional instructions for the

    printer are found in the respective

    package and should be made a part

    of this book.

    264M Dot Matrix Printer DP8340

    Series Users Manual

    1266 Calorimeter Operating Instruction Manual

    I

    P R E F A C E

    Scope

    Customer Service:

    Questions concerningthe installation oroperation of thisinstrument can beanswered by theParr Customer Service

    Department:309-762-7716800-872-7720Fax: [email protected]

  • 8/12/2019 Bomba Cal 1266

    8/113

    1266 Calorimeter Operating Instruction Manual

    II

    These steps are offered to help

    the user become familiar with, install,

    operate and develop the full

    capabilities of the Parr 1266

    Calorimeter.

    1. Review the Concept of Operations,

    Chapter 1, to get an understanding

    of the overall capabilities of the

    calorimeter and microprocessor

    control.

    2. Unpack and install the

    calorimeter in accordance with the

    Installation Instructions, Chapter 2.

    This simple, step-wise procedure

    will acquaint the user with the

    various parts of the calorimeter

    and make it easier to understand

    the operating instructions which

    follow.

    3. Turn to the Instrument Description,

    Chapter 3, to review the keyboard

    controls.

    4. Review the Program Installation

    and Control, Chapter 4, to match

    the factory settings to the intended

    mode of operation. Any required

    changes can be made to the

    program parameters located in

    the Main Menu.

    5. Review the Reporting Instructions,

    Chapter 7, to become familiar with

    the manner in which calorimetry

    corrections are entered. Also

    discussed are generating finalreports, editing and clearing

    memory.

    6. Turn to the Menu Operating

    Instructions, Appendix A, to review

    the menu functions used to modify

    the program contained in the 1266

    Calorimeter. A review of the

    menus will provide a good idea

    of the capabilities and flexibility

    designed into this instrument.

    7. Review the Calculations, Appendix

    B. This provides information about

    calculations performed by the 1266

    Calorimeter.

    8. Review Standardization,

    Appendix C. This will serve two

    important functions. First, it

    provides instructions on

    generating the energy equivalent

    factor required to calculate the

    heat of combustion (HOC) of

    unknown samples. Secondly, it

    will give the user the opportunity

    to run tests on a material with a

    known heat of combustion to

    become familiar with the

    instrument and confirm that

    the instrument and operating

    procedures are producing results

    with acceptable precision. Most

    1266 Calorimeters with 1108

    Oxygen Combustion Bombs and

    2000 grams of water will have an

    energy equivalent of

    approximately 2400 +/-15

    calories per C. The steps for

    standardization and

    determinations are identical,

    except for the setting of the

    instrument to the standardization

    or determination mode.

    9. Review the CommunicationInterfacing, Appendix D, for the

    correct installation of any Smart

    Link peripherals to the 1266

    Calorimeter.

    10. After successful standardization,

    the 1266 Calorimeter should be

    ready for testing samples.

    GettingStarted

  • 8/12/2019 Bomba Cal 1266

    9/113

    T

    he Parr 1266 Oxygen Combustion

    Bomb Calorimeter has been

    designed to provide rapid and reliable

    heat of combustion values for solid

    and liquid fuels. This instrument

    combines fourth generation

    microprocessor hardware and software

    with the calorimeter designs, technical

    knowledge and customer support

    developed by Parr in the century that

    calorimetry has been a principal thrust

    of our company. This chapter explains

    the important design and operating

    concepts incorporated into the 1266

    Calorimeter.

    1266 Calorimeter Operating Instruction Manual

    1-1

    CONCEPT OF OPERATION 1

    IsoperibolOperation

    DynamicOperation

    In isoperibol operation, the

    calorimeter jacket is held at a

    constant temperature while heat from

    the burning sample causes the bomb

    and bucket temperature to rise. The

    small heat flow between the bucket

    and its surroundings during a test is

    monitored by a microprocessor in the

    calorimeter, which continuously

    determines the effect of any heat leak

    and applies the necessary correction

    automatically. This system differs from

    adiabatic operation in which the jacket

    temperature must be adjusted

    continuously to match the bucket

    temperature in an attempt to maintain

    a zero temperature differential with no

    heat leaks between the bucket and its

    surroundings. Calorimetrists have

    long recognized the advantages of

    simplification and better precision

    obtainable with a well designed

    and executed isoperibol system as

    opposed to the rapidly changing

    jacket temperature required in an

    adiabatic calorimeter.

    In its Dynamic Operating Mode,

    the calorimeter uses a sophisticated

    curve matching technique to compare

    the temperature rise with a known

    thermal curve to extrapolate the final

    temperature rise without actually

    waiting for it to develop. Repeated

    testing, and over 15 years of routine

    use in fuel laboratories, has

    demonstrated that this technique

    can cut the time required for a test by

    one-half without significantly affecting

    the precision of the calorimeter.

  • 8/12/2019 Bomba Cal 1266

    10/1131-2

    The microprocessor controller

    in this calorimeter has been

    preprogrammed to automatically

    prompt the user for all required data

    and control input and to:

    A. Generate all temperature readings

    in the calorimeter.

    B. Monitor jacket as well as bucket

    temperature.

    C. Confirm equilibrium conditions.

    D. Fire the bomb.

    E. Confirm that ignition has occurred.

    F. Determine and apply all necessary

    heat leak corrections.

    G. Perform all curve matching and

    extrapolations required for

    dynamic operation.

    H. Terminate the test when it is

    complete.

    I. Monitor the conditions within the

    calorimeter and report to the user

    whenever a sensor or operating

    condition is out of normal ranges.

    1266 Calorimeter Operating Instruction Manual

    1

    FullMicro-

    processorBasedProcessControl

    In addition to its process control

    functions, the microprocessor in

    the calorimeter has been

    preprogrammed to:

    A. Collect and store all required test data.

    B. Apply all required corrections for

    combustion characteristics.

    C. Compute and report the heat of

    combustion for the sample.

    The fourth generation software

    built into this calorimeter and accessed

    through the screen menus permit the

    user to customize the operation of the

    calorimeter to meet a wide variety of

    operating conditions including:

    A. A large selection of printing

    options.

    B. Choice of accessories and

    peripheral equipment.

    C. Multiple options in regard to

    handling thermochemicalcorrections.

    D. Choice of ASTM or ISO correction

    procedures.

    E. A variety of memory management

    and reporting procedures.

    F. Complete freedom for reagent

    concentrations and calculations.

    G. Unlimited choice of reporting units.

    H. Automatic bomb usage monitoring

    and reporting.

    I. A choice of Equilibrium or

    Dynamic test methods.

    J. Automatic statistical treatment of

    calibration runs.

    K. Enhanced testing and trouble

    shooting procedure

    The 1266 Calorimeter is equipped

    with three RS232C connections fordirect communication with its printer,

    an attached balance and a computer.

    It is also equipped with the Parr

    Smart Link network connection for

    networking with other calorimeters,

    fuel testing instruments, balances

    and computers.

    Full Micro-processorBased DataAcquisitionand Handling

    FlexibleProgramming

  • 8/12/2019 Bomba Cal 1266

    11/113

    T

    he 1266 Calorimeter is completely

    assembled and given a thorough

    test before it is shipped from the

    factory. If the user follows these

    instructions, installation of the

    calorimeter should be completed

    with little or no difficulty. If the factory

    settings are not disturbed, only minor

    adjustments will be needed to adapt

    the calorimeter to operating conditions

    in the users laboratory.

    1266 Calorimeter Operating Instruction Manual

    2-1

    INSTALLATION 2

    Unpack 1. Unpack the calorimeter carefully.If shipping damage is discovered,

    report it immediately to the

    delivering carrier. The calorimeter

    and all of its component parts are

    packed in one carton. The unit has

    had all internal connections made,

    and other than connections to

    cooling water and oxygen

    supplies, no additional assembly

    should be required. Set the

    calorimeter on a sturdy bench

    or table in a location that is

    reasonably free from drafts and

    protected from sources of radiant

    heat; preferably in an air-

    conditioned room. There should

    be convenient access to running

    water, to a drain and appropriate

    grounded electrical outlet. Running

    water and a drain are not

    required when the unit is used in

    conjunction with a Parr 1563

    Water Handling System.

    2. Raise the cover of the calorimeter

    and swing it vertically to the back.

    Remove the calorimeter bucket

    and discard the packing material.

    Check the calorimeter bucket and

    note the three dimples in the

    bottom which rest on supporting

    pins when the bucket is placed in

    the jacket. The single dimple must

    always be placed to the back of the

    jacket. There will also be two

    ignition wires extending into the

    oval chamber through the opening

    in the left side of the jacket.

    Installingthe WaterHandlingSystem

    3. If operation of the calorimeter will

    be done in conjunction with a 1563

    Water Handling System, make the

    water connections at this time.

    Connect the line supplying waterfrom the filling system to the

    connection at the lower left hand

    corner at the water jacket (rear

    view). The metering valve at this

    connection should be fully open

    when used with a 1563 Water

    Handling System. Adjustment of

    the systems temperature to

    between 30 and 32 C will provide

    water at the correct temperature

    for the 1266 Calorimeter. Step 4

    may be skipped if the water

    connection has been made to

    the 1563 Water Handling System.

    4. Connect a cold water line to theinlet valve in the lower left hand

    corner of the water jacket on the

    back of the calorimeter. This is the

    water connection to the cooling

    coil within the jacket. It will not fill

    the jacket with water. The inlet

    water connection is made with

    1/4 copper tubing furnished for

    this purpose. Keep all water lines

    as short as possible to avoid

  • 8/12/2019 Bomba Cal 1266

    12/1132-2

    unwanted temperature changes.

    If the supply pressure exceeds 60

    psig, it will be necessary to install

    a reducing valve to keep the

    pressure below the 60 psig

    maximum. Ideally, the product

    of the cooling water flow rate

    in liters per minute, and the

    difference between the jacket

    set point and the inlet water

    temperature in degrees C should

    be 3.6 degree-liters per minute.

    1266 Calorimeter Operating Instruction Manual

    2

    5. The 1552 Water Cooler is intended

    to provide a uniform water supply

    in the temperature range from

    7 to 10 C for maintaining the

    desired bucket and jacket

    temperatures for each run. The

    amount of cold water needed for

    the water handling system is

    determined by the thermostat in

    the water handling system.

    If a water cooler will be used,

    make the connections at this time.

    Refer to Fig. 2-4. Install the 1552

    Water Cooler so that heat from the

    compressor will not be directed

    toward the calorimeter. In any

    installation, the cooler should be

    placed at or below the working

    surface of the calorimeter. Allow a

    minimum clearance of 6 inches

    in front of the unit, 3 inches in the

    rear and at least 1 inch on the sides

    for air circulation. Install the two

    232VB Union Male Elbows

    on the water inlet and outlet

    connections (see Fig. 2-3).

    Connections between the

    calorimeter and the cooler should

    be made with the furnished 1/4

    tubing. A temperature adjustment

    screw is located in an opening on

    the front of the cooler. Set this

    adjustment screw at the coldest

    setting (fully clockwisesee Fig. 2-2).

    Note:

    Do not plug in water cooler at this

    time.

    Installing theWater Cooler

    Figure 2-1 1552 Water Cooler Electrical Circuits

  • 8/12/2019 Bomba Cal 1266

    13/113

    1266 Calorimeter Operating Instruction Manual

    2-3

    2

    InstallationContinued

    Figure 2-3 1552 Water Cooler Top View

    Figure 2-4 Swagelok 1

    Figure 2-5 Swagelok 2

    Figure 2-6 Swagelok 3

    When Swagelok Tube Fittings areused, the instructions for installationare:

    1. Simply insert the tubing into the

    Swagelok Tube Fitting. Make sure

    that the tubing rests firmly on the

    shoulder of the fitting and the nut

    is tight.

    2. Before tightening the Swagelok

    nut, scribe the nut at the 6:00 clock

    position.

    3. While holding the fitting body with

    a back-up wrench, tighten the nut

    1-1/4 turns. Watch the scribe mark,

    make one complete revolution and

    continue to the 9:00 clock position.

    4. For 3/16 and 4mm or smaller tube

    fittings, tighten the Swagelok nut

    3/4 turn from finger tight.

    Figure 2-2 1552 Water Cooler Front View

    Note:

    12 1

    2

    3

    4

    5

    67

    8

    9

    10

    11

    12 12

    3

    4

    5

    67

    8

    9

    10

    11

  • 8/12/2019 Bomba Cal 1266

    14/1132-4

    6. Attach a drain line to the discharge

    connection of the solenoid valve

    attached to the back of the

    calorimeter jacket. Flexible plastic

    tubing is normally used for the

    cooling water outlet or drain.

    A small black collar is installed in

    the outlet fitting on the

    calorimeter. Insert the tubing into

    the connector. To remove the

    tubing, press in on the black collar

    and pull out on the tube.

    1266 Calorimeter Operating Instruction Manual

    2

    Fig. 2-7 1266 Calorimeter External Plumbing

    Fig 2-8 1266 Calorimeter Electrical Hookup

    Caution!

    DO NOT TURN ON THECALORIMETER ATTHIS TIME. THECALORIMETER MUSTBE FILLED WITHWATER BEFORETURNING ON HEATERAND PUMP. THIS WILLPREVENT IMMEDIATEHEATER BURNOUT.

  • 8/12/2019 Bomba Cal 1266

    15/113

    2

    1266 Calorimeter Operating Instruction Manual

    2-5

    Caution!

    THE CALORIMETERMUST BE FILLEDWITH WATER BEFORETURNING ON HEATERAND PUMP. THIS WILLPREVENT IMMEDIATEHEATER BURNOUT.

    Install thePrinter

    7. Connect the printer to the

    calorimeter at this time. The Parr

    1755 Printer is configured and

    furnished with a cord to connect

    directly to the RS232C port of the

    calorimeter.

    8. Plug the power line into any

    grounded outlet providing proper

    voltage which matches the

    specification on the nameplate of

    the calorimeter. The calorimeter

    will draw approximately 1000

    watts of power. Grounding is very

    important not only as a safety

    measure, but also to ensure

    satisfactory controller

    performance. If there is any

    question about the reliability of

    the ground connection through

    the power cord, run a separate

    earth ground wire to the

    controller chassis.

    The BalancePortConnection

    9. The balance port connection, if

    used, should be made at this time.

    Refer to figure 2-8 for the correct

    cable to connect the balance to the

    calorimeter.

    Smart LinkConnection

    10. Connection to the Parr Smart Link

    Network, if needed, can be made

    at this time. A detailed discussion

    of the Parr Smart Link can be

    found in Appendix D -

    Communication Interfaces. The

    termination Settings are done in

    software, See Menu page 8.1 line

    5. If one Smart Link cable is

    plugged into the rear of the

    calorimeter (end of Smart Link

    position), no change is required

    as the default value is ON. If two

    cables are used (middle of the

    Smart Link), then the setting must

    be set to OFF.

    11. Connect the printer power cable

    and the printer communications

    cable between the 1755 printer and

    the connection at the rear of the

    calorimeter.

    Fill theWater Jacket

    12. The water jacket of the calorimeter

    must be filled manually with

    approximately 9 liters of water. Tofill the jacket, raise the cover and

    remove the filling plug which is

    the 1/2 diameter plug located on

    top of the jacket between the

    bucket chamber and the right-hand

    cover hinge. Add water through

    this opening until the water level

    is just below the top plate of the

    jacket.

    13. Go to the Calorimeter Main Menu

    and press 1, Calorimeter

    Operation. Press the down arrowkey twice and press the YES key

    to activate the calorimeter water

    pump and heater. As water is

    circulated to the cover of the

    calorimeter, the water level will

    fall. Top off the jacket until it

    remains within approximately 1/8

    below the top plate and replace

    the filling plug. To turn off the

    pump, push the F1 key again.

  • 8/12/2019 Bomba Cal 1266

    16/1132-6

    1266 Calorimeter Operating Instruction Manual

    2

    Fig. 2-9 1266 Calorimeter Back View

    SolenoidValve

    High Temp.Cut Out

    Heater

    CoolingInlet Valve Pump Motor

    Oxygento Bomb

    OxygenInletSolenoid

    Valve Assy.

    Controller

    14. The 1266 Calorimeter is equipped

    with an automatic bomb oxygen

    filling system. Thus system

    consists of an oxygen pressure

    regulator with a relief valve that

    mounts on an oxygen tank. A

    control and measurement system

    is mounted in the calorimeter case.

    The connection tubing to connect

    the regulator to the solenoid

    assembly and the solenoid

    assembly to the bomb itself are

    included.

    To install the regulator, unscrew

    the protecting cap from the oxygen

    tank and inspect the threads on the

    tank outlet to be sure they are

    clean and in good condition. Place

    the ball end of the regulator in the

    outlet and draw up the union nut

    Installing theOxygen FillingSystem

  • 8/12/2019 Bomba Cal 1266

    17/113

    1266 Calorimeter Operating Instruction Manual

    2-7

    2

    15. The calorimeter must be accurately

    standardized prior to actually

    performing calorimetric tests on

    sample materials. Review

    Appendix C - Standardization, in

    order to become familiar with the

    general procedure and

    calculations. The user should

    configure the calorimeter at this

    time to accommodate the desired

    sample weight entry mode. The

    calorimeter can be placed into thestandardization mode on the

    Calorimeter Operation Page, Item

    1, Operating Mode. The YES/NO

    keys are used to select the desired

    operating mode. If two bomb head

    and buckets are being used with

    the calorimeter to maximize

    sample throughput, the

    calorimeter can be configured to

    prompt for a Bomb ID at the start

    of each test. The Bomb ID can

    also be selected on the Calorimeter

    Operations Page, Line 2, using the

    YES/NO keys. Both bomb and

    bucket combinations will need to

    be standardized separately. The

    end result of a standardization test

    is an energy equivalent value, or

    the amount of energy required to

    raise the temperature of the

    calorimeter one degree. Repeated

    standardization with any givenbomb head should yield an energy

    equivalent value with a range of 3

    to 4 calories per degree, centered

    around the mean value for all tests

    using that bomb bucket

    combination. The calorimeter is

    ready for testing samples after a

    suitably constant energy

    equivalent value has been

    obtained.

    tightly, keeping the gages tilted

    slightly back from an upright

    position. Connect the regulator to

    the inlet fitting on the back of the

    calorimeter case. Flexible high

    pressure, 1/8 diameter nylon

    tubing is provided.

    Attach the bomb filling hose

    (1/8 tubing) to the tubing

    connector nearest the back of the

    calorimeter case. These hoses

    should be routed so that they will

    not kink or come in contact with

    any hot surface. All connections

    should be checked for leaks. Any

    leaks detected must be stopped

    before proceeding. Instructions for

    operating the filling connection are

    in the Operating Instructions

    chapter. The pressure regulator

    was set at the factory to deliver

    oxygen at 450 psig, the

    recommended charging pressure,

    and should be checked before

    starting to use the system by

    observing the pressure attained

    during an actual filling operation.

    To do this, assemble the oxygen

    bomb without a charge and attach

    the filling hose to the bomb inlet

    valve. Then push the 02 button on

    the calorimeter control panel and

    observe the delivery pressure as

    shown on the 0-600 psi gage while

    oxygen is flowing into the bomb.

    Adjust the regulator, if needed, to

    bring the pressure to 450 psig. If

    there is any doubt about the

    setting, release the gas from the

    bomb and run a second check.

    Standardization

    Installing theOxygen Filling

    SystemContinued

  • 8/12/2019 Bomba Cal 1266

    18/1133-1

    A

    ll calorimeter configurations and

    operations are handled by a menu-

    driven system operated from the

    easy-to-use keyboard with a large,

    bright display. The settings and

    controls are organized into nine

    main sections or pages. This is

    the MAIN MENU.

    Main Menu

    1 Calorimeter Operation

    2 Operating Controls

    3 Program Information and

    Control

    4 Calibration Data and Control

    5 Thermochemical Calculations

    6 Data Entry Controls

    7 Reporting Control

    8 Communication Control

    9 Diagnostics

    1266 Calorimeter Operating Instruction Manual

    INSTRUMENT DESCRIPTION3

    Any page number or line number

    on a page can be selected by:

    Cursor Control - The up and

    down arrow keys can be used to move

    the reverse video to the desired

    page/line number. Once the desired

    page/line number is highlighted, press

    the ENTER key.

    Note:

    When the cursor highlights the line

    number and title, as opposed to the

    data field, a page or sub-menu can be

    accessed by pressing the ENTER key.

    (See Command Codes in this chapter.)

    Number Control - Page/Line

    numbers can also be selected by

    simply pressing the desired page/line

    number on the keyboard.

    Page andLine NumberSelection

    Types ofControls

    Types of Controls The controls that

    change the data field information in

    the menus will be one of the following:

    1. Toggles. These data fields

    contain ON/OFF or YES/NO

    choices. The right and left arrowkeys are used to change the

    setting to the desired choice.

    Once the desired setting is

    displayed, press the ENTER key

    to record the change.

    2. Option Selection. These data

    fields contain a list of options. The

    right and left arrow keys are used

    to step through the options. Once

    the desired option is displayed,

    press the ENTER key to record

    the change.

    3. Value Entry Fields. These data

    fields are used to enter data into

    the calorimeter. Most data fields

    will accept any value entered by

    the user through the keyboard.Certain data fields may only

    accept certain values from a table

    built into the menu similar to the

    option selection described above.

    Once the correct value is dis-

    played in the field, press the

    ENTER key to record the value.

    4. Data Displays. Most of these

    data fields display values that

    have been calculated by the

  • 8/12/2019 Bomba Cal 1266

    19/113

    calorimeter and are informational

    only. Certain display values can be

    overridden by the user entering a

    desired value through the

    keyboard.

    5. Command Codes. Line

    numbers that access sub-menus

    or statements initiating an action

    (the controller will ask, Do you

    wish to continue (Yes or No)? to

    prevent accidental disruptions to

    the program and/or stored data).

    Note:

    To return to previous pages from

    a sub-menu, press ESCAPE. To

    return to the Main Menu, press

    SHIFT, ESCAPE. Remember, no

    value or setting chosen is stored

    in the program until the ENTER

    key is pressed.

    1266 Calorimeter Operating Instruction Manual

    3-2

    3

    Types of ControlsContinued

    Keyboard Before using the 1266 Calorimeter,the operator should understand the

    various keys on the keyboard (see

    Figure 3-1). The brief descriptions

    given here will identify these keys and

    their functions. Instructions in later

    chapters provide detailed operating

    instructions.

    7 8 9

    4 5 6

    1 2 3

    . 0 E

    ESCAPEREPORTRESETSHIFT

    SKIP

    HELP

    PRINTSCREEN

    CLEARMEM.

    CLEARENTRY

    NO YES

    F1 F2

    F3FILL

    O2

    ENTER

    START

    Fig. 3-1 1266 Calorimeter Keyboard

  • 8/12/2019 Bomba Cal 1266

    20/1133-3

    Key Description

    Moves the cursor UP on anyof the menu pages.

    NO Changes menu functions

    that are either an ON/OFF or

    YES/NO toggle to OFF, NO,

    or to select an option from

    the list provided.

    YES Changes menu functions

    that are either an ON/OFF or

    YES/NO toggle to ON, YES,

    or to select an option from

    the list provided.

    Moves the cursor DOWN onany of the menu pages.

    START This key is used to start a

    test. After pressing, the

    controller will prompt the

    user to enter all preliminary

    data required for the test.

    F1 Used to enter sample IDs

    and weights when a series of

    samples will be weighed

    before they are tested. This

    preweighing sequence will

    prompt the user for sample

    IDs and weights.

    F2 not used

    F3 This key is used to initiate

    edit capabilities for sample

    weight, bomb ID, spike

    weight, fuse, acid, and sulfur

    or hydrogen values.

    02 Fill This key is used to activate

    the oxygen filling system

    used to fill the bombs. The

    reset key can be used to

    abort the filling procedure.

    CLEAR MEM. Clears tests from the

    memory. When it is pressed,

    the controller will prompt for

    the beginning and ending

    Sample IDs to be deleted

    from memory.

    PRINTSCREEN Prints information

    (except graphics) which is

    currently shown on the

    instrument display to the

    printer.

    HELP Used to access a set of help

    menus that are available to

    assist the user in setting up

    and/or operating this

    calorimeter.

    SHIFT This key is used to change

    the function of a key in much

    the same way that the shift

    key operates on a typewriter.

    For example, the numeric

    keys (1-9) can be used with

    the SHIFT key to provide

    access up to nine

    programmed functions.

    0-9 & (.) The DECIMAL and 0-9 keysare used for all manual data

    entries. The decimal key can

    also be used for other

    functions described later

    in this manual.

    E This key is used to enter a

    negative number when

    1266 Calorimeter Operating Instruction Manual

    3

    KeyboardContinued

  • 8/12/2019 Bomba Cal 1266

    21/113

    required. It is entered before

    the digits and also used to

    indicate that the numbers

    are being entered in

    scientific notation.

    CLEAR ENTRY Erases incorrect

    values and error messages

    from the display. If the user

    is not satisfied with the value

    shown on the display, press

    the CLEAR ENTRY key and

    re-enter the value.

    ENTER The ENTER key is used in

    two ways: 1.To access

    sub-menus. 2.To store data

    field information into the

    controller.

    If there is a sub-menu, the

    line will be highlighted when

    the cursor is on that line.

    Press the ENTER key to

    access the sub-menu.

    If there is no sub-menu,

    the data field will be

    highlighted. If new data

    field information is selected

    or entered through the

    keyboard by the user, the

    ENTER key must be pressed

    to store the information. (See

    Types of Controls in this

    chapter for a complete

    description.)

    RESET Aborts a test or to escape

    from a special sub-routine.

    (i.e. reporting or memory

    management.)

    REPORT Initiates the reporting

    process. Reports can be

    sent to the display, to the

    attached printer, or over

    the Smart Link to another

    controller. Before any

    preliminary reports can be

    made final, the controller

    will prompt the user to

    provide any missing data;

    such as fuse, acid and/or

    sulfur corrections.

    SKIP Allows the user to omit the

    entry of data requested by

    the calorimeter, to pass over

    a prompted request for

    information already held in

    memory, and to select a data

    field when a double column

    appears (see Page 5 -

    Thermochemical

    Corrections, Lines 1-6).

    ESCAPE Used to go up one level in

    the menu structure. For

    example, pressing ESCAPE

    one time will shift the menu

    from sub-menu Page 5.6 to

    5. Pressing ESCAPE a

    second time will return to

    the main menu. Press the

    SHIFT, ESCAPE key to return

    directly to the main menu

    from a sub-menu.

    1266 Calorimeter Operating Instruction Manual

    3-4

    3

    KeyboardContinued

  • 8/12/2019 Bomba Cal 1266

    22/1134-1

    1266 Calorimeter Operating Instruction Manual

    PROGRAM INSTALLATION & CONTROL4

    Units are preprogrammed with

    DEFAULT SETTINGS. See Figure 4-1

    on the next page for a listing of the

    factory default settings.

    These default settings remain in

    effect until changed by the user.

    Should the user ever wish to return to

    the factory default settings, go to Page

    3, Line 7 - Re-load Factory Default

    Settings, press ENTER and YES.

    Battery backed-up memory is

    provided to retain any and all operator

    initiated program changes; even if

    power is interrupted or the unit is

    turned off. If the unit experiences an

    intentional or unintentional Cold

    Restart, the controller will return to

    its default settings.

    DefaultSettings

    The default parameters of the

    1266 Calorimeter can be changed to

    guarantee that the 1266 Calorimeter,

    when cold restarted, will always be in

    the desired configuration before

    beginning a series of tests.

    Users who wish to permanentlyrevise their default settings may do

    so using the following procedure:

    Establish the operating parameters

    to be stored as the user default

    settings.

    Go to Page 3, Line 9 - Save User

    Default Settings, and press the ENTER

    key.

    Press YES, the controller will

    prompt the user for a User Rev.. This

    can be any number that is helpful in

    identifying the users setup. Enternumber and press the ENTER key.

    To re-load the user default setting,

    go to Page 3, Line 8 - Re-load User

    Default Settings, press the ENTER key

    and YES.

    RevisedDefaultSettings

    SoftwareInstallation

    The program in the 1266

    Calorimeter can be extensively

    modified to tailor the unit to a wide

    variety of operating conditions,

    reporting units, laboratory techniques,

    available accessories and

    communication modes.

    In addition, the calculations,

    thermochemical corrections and

    reporting modes can be modified to

    conform to a number of standard test

    methods and procedures.

    Numerous provisions are included

    to permit the use of other reagent

    concentrations, techniques,

    combustion aids and short cuts

    appropriate for the users work.

    Note:

    Changes to the program

    are made by use of the menu structure

    described in Appendix A of this

    manual. Any of these items can be

    individually entered at any time to

    revise the operating program.

  • 8/12/2019 Bomba Cal 1266

    23/113

  • 8/12/2019 Bomba Cal 1266

    24/113

  • 8/12/2019 Bomba Cal 1266

    25/113

    1266 Calorimeter Operating Instruction Manual

    4-4

    4

    Fig.4.1Factory DefaultSettingsContinued

    Discussion for menu page 5.7;

    Items 9 and 0 (calculated HNO3)

    The nitric acid formed in the bomb

    during the combustion process is

    derived from the nitrogen in the air

    that occupies the bomb prior to

    pressurizing it with oxygen. The

    quantity of nitric acid formed is a

    function of the interior volume of the

    bomb, the oxygen filling pressure and

    the quantity of energy released in the

    bomb during a test. For a given bomb

    volume and filling pressure, the

    relationship between the amount of

    nitric acid formed and the energy

    released is reasonably constant.

    If the fixed acid flag for a given

    test mode is ON and calculate HNO3is selected (turned ON), then e1 is

    calculated based on the amount of

    energy released during the test. In this

    case, the amount of nitric acid

    calculated (in calories) is the product of

    the calorimeter energy equivalent, the

    temperature rise and the

    HNO3/Energy factor divided by 1000.

    This HNO3/Energy factor is 1.58 for a

    Parr 1108 style bomb at 30 atmos-

    pheres filling pressure. If the acid is

    HNO3 only flag is ON then the report-

    ed acid value is e1. If this flag is OFF,

    then the reported acid value is e1+e

    2.

    See the discussion in Appendix B

    (Thermochemical Corrections) for a

    more thorough discussion related to

    how these corrections are calculated

    and handled.

    Page 6

    Data Entry Controls

    1 Weight Entry Keyboard

    2 Spike Entry Keyboard

    3 Acid Entry Keyboard

    4 Sulfur Entry Keyboard

    5 Hydrogen Entry Keyboard

    6 Prompt for Bomb ID ON

    7 Use Long Sample IDs OFF

    8 AutoSampleID Controls ON

    9 Auto Controls ON

    0 Weight Warning above 2.0

    Page 6.9

    Sample ID Controls

    1 Automatic Preweigh ID ON

    2 Preweigh ID Increment 1.0

    3 Preweigh ID Number 1.0

    Page 7

    Reporting Controls

    1 Printer Controls

    2 Report Dest Printer

    3 Automatic Reporting ON

    4 Individual Reports OFF

    5 Report Format Text

    6 Overwrite Final Reports OFF

    7 Edit Final Reports OFF

    8 Recalc Final Reports OFF

    9 Use New EE Values OFF

    Page 8

    Communication Controls

    1 Smart Link Controls

    2 Printer Port (RS232C)

    3 Balance Port (RS232C)

    4 Terminal Port (RS232C)

  • 8/12/2019 Bomba Cal 1266

    26/1134-5

    1266 Calorimeter Operating Instruction Manual

    4

    Page 8.1

    Smart Link Controls

    1 Smart Link ID (0-15) 0

    2 Baud Rate 9600

    3 Report Type Short Report

    4 Timeout (x 0.01 s) 50

    5 Smart Link Termination OFF

    6 Use a 4 digit year OFF

    Page 8.2

    Printer Port

    Communications CAL I

    1 Data Bits 8

    2 Parity None

    3 Stop Bits 1

    4 Handshaking Xon/Xoff

    5 Baud Rate 9600

    Page 8.3

    Balance Port Communications

    1 Data Bits 7

    2 Parity Even

    3 Stop Bits 1

    4 Handshaking None

    5 Baud Rate 24006 Balance type Mettler 011

    7 Data Char 9

    8 Log to Printer

    Page 8.4

    Terminal Port Communications

    1 Data Bits 8

    2 Parity None

    3 Stop Bits 1

    4 Handshaking Xon/Xoff

    5 Baud Rate 9600

    6 Terminal Type VT100

    7 Remote Terminal ON

    8 Emulate 1745 ON

    Page 9

    Diagnostics

    1 Self-Test

    2 I/O Diagnostics

    3 Thermometry Diagnostics

    4 Communication Diagnostics

    5 Keyboard Diagnostics

    6 User Defaults Up/Down Load

    7 Data Logger Menu

    8 Print Error List

    Page 9.2

    I/O Diagnostics

    1 Activate Ignition Circuit

    2 Test 02 Fill Function

    3 I/O Driver Diagnostic

  • 8/12/2019 Bomba Cal 1266

    27/113

    Parameter Description

    D0 Calorimeter drift rate (10s time base)

    corrected for systematic heat leaks.

    Tsum The accumulated temperature rise after the sample is

    ignited. This value is the temperature rise for an

    Equilibrium Method test.C0 This is a counter that is initialized to zero at the

    start of a test and then incremented every 10s.

    At firing, the counter is again reset to zero.

    T1 This value is the extrapolated portion of the temperature

    rise for a Dynamic Method test.

    Calorimeter Status

    Alternate Menu 1 (Access by Shift 1)

    The alternate 1 (shift 1) screen

    presents key calorimeter parameters

    and updates them on a real time

    basis. These items include the date,

    time, calorimeter state (idle, pre

    period or post period), bucket and

    jacket temperatures, digital I/O

    images and the heater duty cycle.

    The remaining items on this page

    are parameters directly related to

    the course of the current calorimetric

    test. These items are used by Parr

    Instrument Company technicians to

    evaluate calorimeter performance.

    Some of these items are described

    below.

    Table 4-1 Calorimeter Status

    1266 Calorimeter Operating Instruction Manual

    4-6

    Fig.4.1Factory Default

    SettingsContinued

    Page 9.2.3

    Calorimeter 1 I/O Diagnostics

    Ignition Fire = Off

    Output = 0x0000

    Input = 0x0cff

    Bucket temperature 27.98814

    Jacket temperature 29.86350

    Heater Duty Cycle 0.00%

    Page 9.7

    Data Logger Controls

    1 Data Logger OFF

    2 Interval In Seconds 10

    3 Data Log Items

    4 Computer Format OFF

    5 Destination Printer

    6 Process Log Dest Printer

    7 Process Log Items

    8 Print Ram Buffer

    9 Clear Ram Buffer

  • 8/12/2019 Bomba Cal 1266

    28/1134-7

    1266 Calorimeter Operating Instruction Manual

    4

    Alternate Report

    Destination Setup

    Alternate Menu 2 (Access by Shift 2)

    These calorimeters are set up, by

    default, to issue a printed report at the

    conclusion of a test. Provision has

    been made to have a duplicate report

    issued at either the terminal or the

    balance port. This feature is enabled

    on menu page A2 (alternate 2). To

    access this page, press the ENTER and

    hold the SHIFT key and then 2. This

    menu page has the following entries:

    1 Alt. Report Destination Null Device

    / Terminal / Balance

    2 Alternate Report Format Data /

    Text

    When the destination is set to the

    null device, duplicate reports are not

    issued. The format for the additional

    report can include the text and

    formatting of the original report (text

    option) or just the unique data items

    associated with the report (data). In the

    later case, datum is delimited with a

    comma and the report is terminated

    with carriage return and line

    feed characters.

    Dry Calculation Setup

    Alternate Menu 3 (Access by Shift 3)

    1 Dry Calculation OFF

    2 Fixed Moisture (%) 0.0

    3 Moisture Entry Mode Keyboard

    4 Moisture Multiplier 5.83

    See Appendix B - Calculations for the

    calculation that uses these parameters

    when Dry Calculation is ON

  • 8/12/2019 Bomba Cal 1266

    29/113

    1266 Calorimeter Operating Instruction Manual

    5-1

    OPERATING INSTRUCTIONS 5

    Operatingthe FillingConnection

    To fill the bomb, connect the hose

    to the bomb inlet valve and push the

    02 button on the calorimeter control

    panel. The calorimeter will then fill the

    bomb to the preset pressure and

    release the residual pressure in the

    connecting hose at the end of the

    filling cycle. The main display will

    read 02 FILL while the bomb is being

    charged. It will take approximately 40

    seconds to fill the bomb. (Pushing the

    RESET key will stop the flow of oxygen

    at any time.) Once the display returns

    to its normal reading, the user can

    disconnect the coupling and proceed

    with the combustion test.

    If the charging pressure fails to

    reach approximately 28 atmospheres

    during the filling cycle, an ERROR 78

    message will be reported on the

    display and/or printer. If this occurs,

    check the system carefully for low tank

    pressure, a closed tank valve or leaks

    in the system.

    The charging cycle can be started

    at any time, but once it is started it

    will lock out any other keyboard or

    reporting activity until it is completed.It will not interrupt a test in progress.

    If the charging cycle should be

    started inadvertently, it can be

    stopped immediately by pushing

    the RESET key.

    During extended periods of

    inactivity, overnight or longer, close

    the tank valve to prevent leakage.

    When changing oxygen tanks, close

    the tank valve and push the 02 FILL

    key to exhaust the system. Do not use

    oil or combustible lubricants on this

    filling system or on any devices

    handling oxygen under pressure. Keep

    all threads, fittings, and gaskets clean

    and in good condition. Replace the two

    394HCJE O-rings in the slip connector

    if the connector fails to maintain a

    tight seal on the bomb inlet valve.

    The recommended 450 psig filling

    pressure is slightly higher than the 30

    atm prescribed in Parr bomb and

    calorimeter instruction manuals. This

    difference is insignificant. Higher or

    lower settings can be used, but the

    bomb must never be filled to more

    than 600 psig (40 atm). If pressures

    below 28 atm are used, the low

    pressure warning will always occur.

    Operating the1108 OxygenBomb

    Detailed instructions for preparing

    the sample and charging the 1108

    Oxygen Bomb are given in Operating

    Instructions No. 205M. Follow these

    instructions carefully, giving particular

    attention to the precautions to be

    observed in charging and handling the

    bomb.

  • 8/12/2019 Bomba Cal 1266

    30/1135-2

    1266 Calorimeter Operating Instruction Manual

    5

    All operations required to

    standardize the 1266 Calorimeter, or

    test an unknown sample, should

    proceed step-wise in the following

    manner:

    1. Turn on the calorimeter and

    activate the pump and heater

    using Line 3 of Page 1, Calorimeter

    Operations. Allow at least 20

    minutes for the controller to warm

    up and the jacket temperature to

    stabilize. Once the jacket

    temperature comes within 0.5 of

    35 C, the calorimeter is ready to

    begin testing. The bomb parts

    should be wetted and then dried in

    the manner used at the conclusion

    of a test. This serves to wet all

    sealing parts, as well as leaving

    the bomb with the same amount

    of residual water which will exist

    in all subsequent testing. If the

    calorimeter takes longer than 20

    minutes to heat up 6 C, either the

    flow rate of the cooling water is

    too high or the temperature of the

    incoming water is too low. If the

    calorimeter jacket temperature

    stays higher than the set point

    (normally 35 C), then the opposite

    of the above conditions exist.

    Keep in mind that the product

    of the cooling water flow as read

    in liters/min and the absolutedifference between the jacket set

    point (normally 35 C) and the inlet

    water temperature in C should be

    between 2 and 4 liter

    degrees/minute. These settings

    will maintain heater duty cycles of

    between 30% and 60%.

    2. Prepare the sample and charge the

    oxygen bomb as described in

    Operating Instructions No. 205M,

    the Filling Connection Section.

    The throughput of the 1266

    Calorimeter can be increased by

    using multiple bombs and water

    buckets. With this arrangement,

    the calorimeter can operate almost

    continuously since the operator

    will be able to empty a bomb and

    recharge it while a run is in

    progress. A bomb and bucket for

    the next run will be ready to go

    into the calorimeter as soon as it is

    opened. Each bomb and bucket

    combination will have to be

    standardized separately and the

    proper energy equivalent for each

    set must be used when calculating

    the heat of combustion.

    3. Fill the calorimeter bucket by first

    taring the dry bucket on a solution

    or trip balance; then add 2000 (+/-

    0.5) grams of water. Distilled water

    is preferred, but demineralized or

    tap water containing less than 250

    ppm of dissolved solids is

    satisfactory. The bucket water

    temperature should be

    approximately 3 to 5 C below the

    jacket temperature. It is not

    necessary to use exactly 2000

    grams, but the amount selected

    must be duplicated within +/-0.5gram for each run. Instead of

    weighing the bucket, it can be

    filled from an automatic pipet, or

    from any other volumetric device

    if the repeatability of the filling

    system is within +/-0.5 ml.

    To speed and simplify the

    Operating theCalorimeter

  • 8/12/2019 Bomba Cal 1266

    31/113

    Operating theCalorimeterContinued

    bucket filling process, and to

    conserve water and energy, Parr

    offers a closed-circuit Water

    Handling System (No. 1563). This

    provides a water supply, cooled to

    the starting temperature and held

    in an automatic pipet ready for

    delivery in the exact amount

    needed to fill the bucket. A 1552

    Water Cooler is required when

    using the 1563 Water Handling

    System. Instructions for this

    automatic system are given in

    Operating Instruction No. 246M.

    4. Set the bucket in the calorimeter.

    Attach the lifting handle to the two

    holes in the side of the screw cap

    and partially lower the bomb in the

    water. Handle the bomb carefully

    during this operation so that the

    sample will not be disturbed. Push

    the two ignition lead wires into the

    terminal sockets on the bomb

    head. Orient the wires away from

    the stirrer shaft so they do not

    become tangled in the stirring

    mechanism. Lower the bomb

    completely into water with its feet

    spanning the circular boss in the

    bottom of the bucket. Remove the

    lifting handle and shake any drops

    of water back into the bucket and

    check for gas bubbles.

    5. Close the calorimeter cover. This

    lowers the stirrer and thermistorprobe into the bucket.

    6. Select determination or

    standardization as appropriate on

    Page 1, Calorimeter Operations,

    Line 1, using the left arrow key to

    toggle between the choices. Press

    Start (Deter.) or Shift Start (STD) to

    begin the test. The calorimeter will

    now prompt the operator for Bomb

    ID number, sample ID number,

    sample weight and spike weight

    in accordance with the instructions

    set into the operating modes on

    menu pages 2 and 6.

    7. The calorimeter will now take over

    and conduct the test. During the

    time it is establishing the initial

    equilibrium, it will display

    PREPERIOD on the status bar.

    Just before it fires the bomb, it will

    sound a series of short beeps to

    warn the user to move away from

    the calorimeter.

    Once the bomb has been

    fired, the status bar will display

    POSTPERIOD. The calorimeter will

    check to make certain that a

    temperature rise occurs and will

    then look for the final equilibrium

    conditions to be met. If it fails to

    meet either the initial or final

    equilibrium conditions, or if it fails

    to detect a temperature rise within

    the allotted time, the calorimeter

    will terminate the test and advise

    the user of the error.

    8. At the conclusion of the test, the

    calorimeter will signal the user.

    9. Open the cover and remove the

    bomb and bucket. Remove the

    bomb from the bucket and open

    the knurled valve knob on the

    bomb head to release the residualgas pressure before attempting to

    remove the cap. This release

    should proceed slowly over a

    period of not less than one minute

    to avoid entrainment losses. After

    all pressure has been released,

    unscrew the cap; lift the head out

    of the cylinder and place it on the

    1266 Calorimeter Operating Instruction Manual

    5-3

    5

  • 8/12/2019 Bomba Cal 1266

    32/1135-4

    1266 Calorimeter Operating Instruction Manual

    5

    support stand. Examine the

    interior of the bomb for soot or

    other evidence of incomplete

    combustion. If such evidence is

    found, the test will have to be

    discarded.

    10. Wash all interior surfaces of the

    bomb with a jet of distilled water

    and collect the washings in a

    beaker.

    11. Remove all unburned pieces of

    fuse wire from the bomb

    electrodes; straighten them and

    measure their combined length in

    centimeters. Subtract this length

    from the initial length of 10

    centimeters and multiply this

    burned length by 2.3 calories per

    cm (for Parr 45C10 Fuse Wire) to

    obtain the fuse correction. The

    scale on the fuse wire card can be

    used to obtain this value directly.

    12. Titrate the bomb washings with a

    standard sodium carbonate

    solution using methyl orange, red

    or purple indicator. A 0.0709N

    sodium carbonate solution is

    recommended for this titration to

    simplify the calculation. This is

    prepared by dissolving 3.76 grams

    of Na2CO3 in the water and

    diluting to one liter. NaOH or KOH

    solutions of the same normality

    may be used.

    13. Analyze the bomb washings to

    determine the sulfur content of the

    sample if it exceeds 0.1%.

    Methods for determining sulfur

    are discussed in Operating

    Instructions No. 207M.

    14. At the end of the testing period,

    turn OFF the calorimeter at the power switch.

    Large particles may not burn

    completely and small particles are

    easily swept out of the capsule by

    turbulent gases during rapid

    combustion.

    Note:

    Particle size is important because

    it influences the reaction rate.

    Compression into a pellet is

    recommended because the pressuredeveloped during combustion can be

    reduced as much as 40% when

    compared to the combustion of the

    material in the powder form. In

    addition in giving controlled burn

    rates, the pelletizing of samples keeps

    the sample in the fuel capsule during

    combustion.

    Materials, such as coal, burn well

    in the as-received or air-dry condition,

    but do not burn completely dry

    samples. A certain amount of moisture

    is desirable in order to control the

    burning rate. Moisture content up to

    20% can be tolerated in many cases,

    but the optimum moisture is best

    determined by trial combustion.

    If moisture is to be added to

    retard the combustion rate, drop waterdirectly into a loose sample or onto a

    pellet after the sample has been

    weighed. Then let the sample stand for

    awhile to obtain uniform distribution.

    OperatingSuggestions

  • 8/12/2019 Bomba Cal 1266

    33/113

    1266 Calorimeter Operating Instruction Manual

    5-5

    5

    Some samples may be difficult to

    ignite or they may burn so slowly that

    the particles become chilled below the

    ignition point before complete

    combustion is obtained. In such

    cases powdered benzoic acid, white

    oil or any other combustible material

    of known purity can be mixed with the

    sample. Ethylene glycol, butyl alcohol

    or decalin may also be used for this

    purpose.

    Note:

    Combustion aids add to the total

    energy released in the bomb and the

    amount of sample may have to be

    reduced to compensate for the added

    charge.

    Also, when benzoic acid is

    combusted for standardization runs or

    for combustion aid purposes, it should

    be in the form of a pellet to avoid

    possible damage to the bomb which

    might result from rapid combustion of

    the loose powder.

    CombustionAids

    Non-volatile samples to be tested

    in Parr oxygen bombs are weighed

    and burned in shallow capsules

    measuring approximately 1 diameter

    and 7/16 deep. These are available in

    stainless steel, fused silica and

    platinum alloyed with 3-1/2% rhodium.

    Stainless steel capsules (43AS) are

    furnished with each calorimeter.

    When combusting samples that

    contain metal particles such as

    aluminum or magnesium, the

    non-metallic (fused silica) 43A3

    Capsule is required, or where the

    superior corrosion resistance of a

    Pt-Rh, the 43A5 Capsule is required.

    The stainless steel capsules will

    acquire a dull gray finish after repeated

    use in an oxygen bomb due to the

    formation of a hard, protective oxidefilm. This dull finish not only protects

    the capsule, but it also promotes

    combustion and makes it easier to

    burn the last traces of the sample.

    It is recommended, therefore, that

    new capsules be heated in a muffle

    furnace at 480 C for four hours to

    develop this protective coating

    uniformly on all surfaces. This

    treatment should be repeated after a

    capsule has been polished with an

    abrasive to remove any ash or other

    surface deposits. Heating in a muffle is

    also a good way to destroy any traces

    of carbon or combustible matter which

    might remain in the capsule from a

    previous test.

    Note:

    After heating, place the capsules

    in a clean container and handle them

    only with forceps when they are

    removed to be weighed on an a

    nalytical balance.

    Foodstuffs and Cellulosic Materials.

    Fibrous and fluffy materials generally

    require one of three modes of

    controlling the burn rate. Fibrous

    materials do not pelletize readily and

    generally require either moisturecontent of combustion aid such as

    mineral oil to retard the burn rate and

    avoid development of high pressures.

    Partial drying may be necessary if the

    moisture content is too high to obtain

    ignition, but if the sample is heat

    sensitive and cannot be dried, a water

    soluble combustion aid such as

    ethylene glycol can be added to

    promote ignition.

    CombustionCapsules

  • 8/12/2019 Bomba Cal 1266

    34/1135-6

    1266 Calorimeter Operating Instruction Manual

    5

    In most cases it may be necessary

    to burn coarse samples without size

    reduction since grinding or drying may

    introduce unwanted changes. There is

    no objection to this if the coarse

    sample will ignite and burn

    completely. Whole wheat grains and

    coarse charcoal chunks are typical of

    materials which will burn satisfactorily

    without grinding and without additives

    or a special procedure.

    Corrosive Samples. The 1108

    Oxygen Bomb is made of a corrosion

    resistant alloy designed to withstand

    the corrosive mixture of sulfuric and

    nitric acids produced in normal fuel

    testing operations. Samples containing

    chlorine and particular samples

    containing more than 20 mg of

    chlorine or samples with high sulfur

    contents will greatly accelerate

    corrosion of the bomb. An alternate

    1108CL Bomb is available constructed

    of an alloy selected to specifically

    resist the corrosive effects of samples

    with high chlorine or chloride.

    While no material will offer

    complete corrosion resistance to these

    samples, the 1108CL Bomb offers

    significantly enhanced corrosion

    resistance for this service.

    CoarseSamples

    Volatile Sample can be handled in a

    Parr 43A6 Platinum Capsule with a

    spun rim, or in a Parr 43AS Alloy

    Capsule which has a sturdy wall with a

    flat top rim. These holders can be

    sealed with a disc of plastic adhesive

    tape prepared by stretching tape

    across the top of the cup and trimmingthe excess with a sharp knife. The seal

    obtained after pressing this disc firmly

    against the rim of the cup with a flat

    blade will be adequate for most

    volatile samples.

    The tape used for this purpose

    should be free of chlorine and as low

    in sulfur as possible. Borden Mystic

    Tape, No. M-169-C or 3M Transparent

    Tape, No. 610, is recommended for this

    purpose. The 3M Transparent Tape can

    be ordered through Parr, Part No. 517A.

    The weight of the tape disc must be

    determined separately and a correction

    applied for any elements in the tape

    which might interfere with thedetermination. This can be done by

    running a blank test with tape alone

    using a sample weighing 1.0 gram.

    The compensation for heat of tape

    may be done through the spike option;

    see Page 2.3 - Spike Controls, Line 2 -

    Heat of Combustion of Spike.

    VolatileSamples

    Parr offers the 1104 High Pressure

    Oxygen Bomb designed specifically for

    testing materials which detonate with

    explosive force, burn with unusually

    high energy levels or release large

    volumes of gas. This bomb is much

    heavier in construction than the 1108

    and should be used when testing these

    types of materials. The program in the

    calorimeter can be modified to

    accommodate the slower heat transfer

    later of the 1104 bomb with the

    selections on Page 4, Line 2 of the

    Main Menu

    Explosives andHigh EnergyFuels

  • 8/12/2019 Bomba Cal 1266

    35/113

    1266 Calorimeter Operating Instruction Manual

    5-7

    5

    VolatileSamples

    Continued

    Note:

    Tape should always be stored in a

    sealed container to minimize changes

    in its moisture and solvent content.

    Use the following procedure when

    filling and handling any of these

    tape-sealed sample holders:

    Also cut and attach a small flag to

    the disc (see Figure 5-7). Puncture the

    tape at a point below the flag, then

    re-weigh the empty cup with its tape

    cover. Add the sample with a

    hypodermic syringe; close the opening

    with the flag and re-weigh the filled

    cup. Set the cup in the capsule holder

    and arrange the auxiliary fuse so that it

    touches the center of the tape disc.Just before starting the test, prick

    the disc with a sharp needle to make a

    small opening which is needed to

    prevent collapse of the disc when

    pressure is applied.

    Fill the bomb with the usual

    oxygen charging pressure.

    The calorimeter will fire the bomb

    and complete the test in the usual

    manner.

    Volatile samples are defined as one

    with an initial boiling point below

    180 C per ASTM D-2.

    Low volatile samples with a high

    water content, such as urine or blood,

    can be burned in an open capsule by

    absorbing the liquid on filter paper

    pulp or by adding a combustion aid,

    such as ethylene glycol.

    Fig 5-7 Combustion Capsule with Adhesive Tape Seal

  • 8/12/2019 Bomba Cal 1266

    36/1135-8

    1266 Calorimeter Operating Instruction Manual

    5

    The 1266 Calorimeter has been

    designed to operate with an oxygen

    filling pressure of 30 atm. Significant

    changes from this value are not

    recommended.

    OxygenChargingPressure

    Because of the difference in

    combustion characteristics of the many

    different materials which may be

    burned in an oxygen bomb, it is

    difficult to give specific directions

    which will assure complete

    combustion for all samples.

    The following fundamental

    conditions should be considered

    when burning samples:

    Some part of the sample must be

    heated to its ignition temperature to

    start the combustion and, in burning, it

    must liberate sufficient heat to support

    its own combustion regardless of the

    chilling effect of the adjacent metal

    parts.

    The combustion must produce

    sufficient turbulence within the bomb

    to bring oxygen into the fuel cup for

    burning the last traces of the sample.

    Loose or powdery condition of the

    sample which will permit unburned

    particles to be ejected during a violent

    combustion.

    The use of a sample that contains

    coarse particles which may not burn

    readily. Coal particles which are too

    large to pass a 60 mesh screen may

    not burn completely.

    The use of a sample pellet which

    has been made too hard or too soft.

    Either condition can cause spalling and

    the ejection of unburned fragments.

    Insufficient space between the

    combustion cup and the bottom of the

    bomb. The bottom of the cup should

    always be at least one-half inch above

    the bottom of the bomb or above the

    liquid level in the bomb to prevent

    thermal quenching.

    Excessive moisture or non-

    combustible material in the sample.

    If the moisture, ash and other non

    combustible material in the sample

    are approx. 20% or more of the charge,

    it may be difficult to obtain complete

    combustion. This condition can be

    remedied by adding a small amount of

    benzoic acid or other combustion aid.

    PoorCombustion

  • 8/12/2019 Bomba Cal 1266

    37/113

    1266 Calorimeter Operating Instruction Manual

    6-1

    CORRECTIONS & FINAL REPORTS 6

    EnteringCorrections

    and Obtainingthe FinalReport

    Final reports for each test can be

    obtained whenever the operator is

    prepared to enter any required

    corrections for fuse, acid and sulfur.

    When entering corrections, the

    user can choose from three methods.

    These are:

    Manual Entry

    Fixed Corrections

    Automatic Data Transfer

    Program and Installation chapter

    (also see Figure 4-1 in this manual)

    provide the default settings used

    to setup the method preferred by

    the user. Refer to the Reporting

    Instructions, Chapter 7, for the

    steps necessary to initiate a report

    from the controller.

    Calculated Titer

    A new ASTM alternative

    is available.

    The nitric acid correction can

    be calculated based on the energy

    release of the sample. A discussion

    of this procedure is found in

    Chapter Four for menu page 5.7

    where correction is entered.

    Manual Entry During the reporting process, thecontroller will prompt the user to enter

    the following values:

    Fuse Correction

    Key in the Fuse Correction and

    press the ENTER key. The default

    setting for this value is to be

    entered in calories.

    Acid Correction

    Key in the Acid Correction and

    press the ENTER key. The default

    setting for this value is to be

    entered in milliliters of standard

    alkali required to titrate total acidor calories.

    Sulfur Correction

    Key in the Sulfur Correction and

    press the ENTER key. The default

    setting for this value is to be

    entered as percent sulfur in the

    sample.If fixed values for fuse, acid

    and sulfur are turned OFF on Menu

    Page 5, Lines 1-6, then the user

    must manually enter the values at

    the prompt. If Use Spiking

    Correction on Page 2, Line 3 is

    turned ON, a spiking correction

    must be entered before obtaining a

    Final Report. After the last entry

    has been made, the calorimeter

    will automatically produce a Final

    Report. If values for these correc-

    tions are not available, the opera-

    tor can use the SKIP key to bypassany of the corrections, however, a

    Final Report will not be printed

    until an entry is made for fuse,

    acid and sulfur.

  • 8/12/2019 Bomba Cal 1266

    38/1136-2

    1266 Calorimeter Operating Instruction Manual

    6

    In many cases, fixed values for

    fuse and acid can be used without

    introducing a significant error since

    the corrections are both relatively

    small and constant.

    Fixed sulfur corrections can also be

    used whenever a series of samples will

    be tested with a reasonably constant

    sulfur content.

    Details for applying fixed

    corrections are found in Appendix B,

    Thermochemical Calculations.

    Any value set-up as a fixed

    correction will be automatically applied

    and the controller will not prompt the

    user for this value.

    FixedCorrections

    Acid and/or sulfur corrections can

    be automatically transferred into the

    calorimeter from a Parr 1760 Sulfur

    Analyzer. Sulfur corrections can also

    be transferred from a Parr 1742 Sulfur

    Interface.

    Page 6 - Data Entry Controls, Lines

    1-4 can be set for automatic transfer

    from devices connected via the Smart

    Link when default is set to automatic

    entry.

    The calorimeter will poll the

    attached device when the user initiates

    the reporting process. If a value is not

    found, the user will have the option of

    entering it manually.

    Mention the difference for the new

    bomb style.

    AutomaticData Transfers

  • 8/12/2019 Bomba Cal 1266

    39/113

    1266 Calorimeter Operating Instruction Manual

    7-1

    7

    Report OptionSelection

    The 1266 Calorimeter can transmit

    data four ways:

    RS232C port to a 40 or 80 column

    printer to provide a printed report.

    Smart Link which formats and

    prints the data on another calorimeter

    with an attached printer.

    Use Alternate Menu 2 to send a

    duplicate report to the balance or the

    terminal port

    Use Al


Recommended