Brewer Ozone Spectrophotometer Internal Stray Light Issues Vladimir Savastiouk Tom McElroy Some points for discussion
Transcript

Brewer Ozone SpectrophotometerInternal Stray Light Issues

Vladimir SavastioukTom McElroy

Some points for discussion

September 12-16, 2011 13th Brewer Workshop Beijing, China 2

Topics for Today

• Follow on from the cross-section talk

• Differential cross-sections

• Instrumental stray light

• Conclusions

September 12-16, 2011 13th Brewer Workshop Beijing, China 3

Calibrating Single Brewers

• It has been demonstrated that Brewers can be independently characterized and absolutely calibrated and be in excellent agreement

• This makes the absolute reference robust since it does not rely on one instrument remaining ‘stable’

• This is at odds with doing a linear 2-parameter fit for the calibration of single Brewers

September 12-16, 2011 13th Brewer Workshop Beijing, China 4

Instrument Stray Light

• Both single Brewers and Dobsons have a stray light problem at large ozone paths

• Some quantification of the problem has been done – both analytically and by observation

• A correction for the effect is needed

• The use of a 2-point calibration is NOT an appropriate response to the issue

September 12-16, 2011 13th Brewer Workshop Beijing, China 5

Comparison of Brewers and Dobsons

Dobson AD

Brewer #007

September 12-16, 2011 13th Brewer Workshop Beijing, China 6

Stray Light

From SAUNA campaign data

Work byAlexander Cede

September 12-16, 2011 13th Brewer Workshop Beijing, China 7

Problems with a 2-parameter Fit• The two parameter fit has no physical basis

• 2-parameter linear fit is a poor representation of the observed performance of the single Brewers (SAUNA)

• Poor performance at low ozone values

• Poor performance at large slant amounts

• Inability to independently calibrate single instruments

Not recommended!

September 12-16, 2011 13th Brewer Workshop Beijing, China 8

Non-linear Retrievalx = (Fo - B m - F) / (A mu) x = X - k ( mu * X )s

equating…

(A mu) [ X - k ( mu * X )s ] - (Fo - B m - F) = 0.0

X – ‘true’ ozone Fo – extraterrestrialk – nonlinearity constant s – power lawmu – ozone airmass m – Rayleigh airmass

F – absorption function x – Single Brewer O3

A – differential absorption coefficient B – Rayleigh scattering coefficient

} Fromdispersion test

September 12-16, 2011 13th Brewer Workshop Beijing, China 9

Solutions

• SAUNA suggested that:– s = 2.16– k = 0.016 (with ozone in cm)

• But…– can fix s and / or k or retrieve all 4 parameters– probably best to retrieve s from a comparison

between singles and doubles (changing ozone)

• A non-linear, least squares solution works

September 12-16, 2011 13th Brewer Workshop Beijing, China 10

Conclusion

• At this point it is suggested that the s and k determined in the SAUNA campaign be used with single Brewer MkII and MkV

• Then do a retrieval for the Fo using calculated cross-sections and a comparison to a calibrated instrument or direct Langley plot