GX400
4
GEISTER The better way to operate TM Powered Surgery - Electrosurgical Units
Transcript
Page 1: Brosur Geister Esu X-400ht

GEI

STER

The better way to operate TM

Powered Surgery -Electrosurgical Units

Page 2: Brosur Geister Esu X-400ht

High Technology Electrosurgical Unit

Up to 2 Twin Foot Pedals

One for monpolar and bipolar use(REF 34-9330) and one for indepen-dentbipolar only use (REF 34-9331 -optional). Water proof (IPX7) with lowvoltage, safe medical grade design..

2 Active Monopolar Outputs

for push-button handswitches with USA 3-pinconnectors or monopolar electrode cablesand handles. Activation even simultaneouslypossible by two surgeons by using handswitch or twin foot pedal. A special universaladapter (REF 34-9590) is optionally availablefor connecting multiple accessories with 2 to8mm plugs.

Ideal for Cardio-Surgery

“Floating type” insulated at high/ lowfrequencies and protected against theuse of defibrillator and 5 preset cardioprograms (#75-79) make this unit anideal choice for cardio-surgery.

100 Memory Programs

Stored programs can be selected sim-ply by the push of one button. Presetprograms are available for cardio-surgery, neuro-surgery, general andlaparoscopic surgery, flexible endos-copy or rigid monopolar endoscopyunder liquid (TUR) and for bipolar onlyuse. The unit always remembers the

last settings it was switched off with.

10 Monopolar Currents

perfectly suitable for all surgical needs including TUR(monopolar resection under liquid) and vaporisation. Also available is a special A. ENDO mode for blend cut withalternate phases of pure cut and coagulation in flexibleendoscopy and A. PURE and A. BLEND modes combinedwith a constant voltage system, that automatically adjusts

the powers by keeping the electric arc constant. This provi-

des for a very delicate way of cutting and minimal superficial

necrosis.

Neutral Electrode Alarm

When the contact surface of a dualsplit-type neutral electrode plate decreases to 50%, output powers willbe limited to max. 200W. The powerwill be totally cut off and alarm will beactivated in case contact decreasesfurther to an insufficient level or incase that the cable gets disconnectedor broken, for maximum safety.

HF Leakage Monitoring

automatically reducing the outputpowers in order to meet, even in criti-cal situations (i.e. contact between pa-tient and metallic parts of OT- table),safe values, reducing potential risksinvolved (i.e. burns).

ESU-X 400HT

Page 3: Brosur Geister Esu X-400ht

Powerful Performance

Output power 400W max. in

monpolar pure cut and 160W max.

in bipolar pure cut. Constant output

power levels through the outputpower control system make this unitparticularly suitable for cutting underwater (e.g. during TUR). Power canbe progressively adjusted to thedesired level under visual control.

Neutral Electrode Connection

with safety control circuit to avoid that burnscan be caused on the patients’ skin. Whendual split-type electrodes are being used, notonly the cable and its connection are beingchecked, but also the proper application of theplate will be monitored continuously.

2 Bipolar Outputs

for cables with 2x4mm USA-typebanana plug connectors (e.g.REF 34-9408) for bipolar cut andcoagulation with maximum safetyand precision.

7 Bipolar Currents

for all surgical needs and with special PRECISEmode for micro precise bipolar coagulation and theSEALING mode for bipolar coagulation and confi-dently ligating and sealing large vessels. AUTOmode with automatic Start/ Stop System and im-pendance sensing for manual operation activatesthe HF current automatically once the tissue isbeing touched.

Total Safety Concept

with complete self-test, hardware/software diagnosis and totalfunction and safety monitoring by two 32 bit microprocessors. Dynamic auto-adjustment of output powers for best performancewhile using the lowest possible powers and reducing damagingside effects on target points (i.e. carbonization or tissue sticking).

Order REF 34-9302

(230V) suggested accessories are

REF 34-9320, 34-9331, 34-9408,

34-9590, 34-9526, 10-7030