FABRICATION OF A REUSEABLE PLASTIC BASED X-RAY DOSIMETER FOR RADIOTHERAPY By IQBAL TARIQ Thesis submitted in fulfilment of the requirements for the degree of Doctor of Philosophy JANUARY 2016
Transcript
Page 1: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

FABRICATION OF A REUSEABLE PLASTIC

BASED X-RAY DOSIMETER FOR

RADIOTHERAPY

By

IQBAL TARIQ

Thesis submitted in fulfilment of the requirements

for the degree of

Doctor of Philosophy

JANUARY 2016

Page 2: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

FABRICATION OF A REUSEABLE PLASTIC BASED X-RAY DOSIMETER FOR RADIOTHERAPY IQBAL TARIQ PhD, Medical Physics. Department of Nuclear Medicine and radiation therapy, School of Medical Sciences, Universiti Sains Malaysia, Health campus,16150 Kelantan , Malaysia. Introduction

Deposition of radiation energy within the interacting material is termed as dose, whose

unit is Gray (Gy), defined as joule of deposited energy per unit mass(kg) (Khan,

2010).Dosimetry is the measurement or estimation of dose, ideally at a defined point

within the interacting material, but it could also be performed for a plane of 2D or for

a volume of 3D.The goal of radiation therapy is to deliver uniform and homogeneous

prescribed dose to treatment volume but at the same time, save the adjacent healthy

tissues from radiation exposure or keep the doses to un-avoidable healthy tissues, well

below their radiation tolerance (Khan, 2010). The treatment plans are made with the patients

radiographic images X-rays, CT scan and MRI. Due to current treatment complexity

raising the risk for target miss or unnecessary irradiation of healthy tissues, pre-treatment

dosimetric verification of each treatment plan is highly recommended.

Now a days a variety of radiation dosimeters are used serving different purposes. Examples

of the commonl used dosimeters are,Ionization chambers, Silver halide Radiographic films,

Radiochromic films, Thermoluminescence dosimeters, Optically stimulated dosimeters,

Semi conductor detectors. Frick dosimeter, Gel dosimetry, Plastic dosimeters and some

Natural and synthetic materials.

Objectives The aim of the study is to fabricate a reusable tissue equivalent plastic composite dosimeter

with acceptable accuracies, that could serve the routine dosimetric requirements of a

Page 3: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

radiation therapy centre.First, this work seeked to identify the most suitable material to be

used as the ingredient in the plastic composite mix. Secondly, this work seeked to determine

whether the sensitivity of any of these materials can be improved by synthetic introduction

of crystal defects. Thirdly, methods of sample preparations were to be evaluated. Finally,

these samples would be evaluated for their radiation dosimetric properties using X-ray

photon beam from a linear accelerator. They would be evaluated mainly for radiation

sensitivity, repeatability, fading, and dose linearity.

Methods and Materials To prepare the samples in which radiation sensitive material grains were uniformly

embedded within the body of the plastic, different techniques were tried and finally N1(pure

sodium chloride) samples were made with Haake polydrive internal mixing and hot press

technique. Samples prepared by, spraying plastic solution with grains of material, with

solution casting of plastic and material grains and b using hot plate with solution casting

and to melt plastic grains was found unsuitable for further experiments.

Results The final sample of N1 and PMMA composite in 40/60 ratio (wt/wt) has shown

good sensitivity for 3 Gy of 6 MV photons with very linear dose response up to 10 Gy.

From 10 Gy to 15 Gy the sensitivity of the plastic dosimeter was found to be

decreased. For the next observed dose of 20 Gy, the dosimeter appeared to

regain its sensitivity. The doses were delivered for alternate 4 days post optical bleaching of

18 hours. The area of dip in the percentage transmission curve, corresponding the absorbed

dose was found to have an average value of 96.4 arbitrary area units with standard

deviation of 4.85 %. The peak values of the dips were found to be between 460 nm to 472

nm. The fading in response for 5 Gy of 6 MV photons was found to be 16.93% in one week

and 48.07% in one month. Dosimetric evaluation of the materials were conducted with the

Page 4: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

study of their optical properties pre and post irradiation of 2 Gy of 6 MV photons, using

Shimadzu UV 1800 Spectrophotometer.

Conclusions Zeff of the fabricated N1 PMMA composite dosimeter is 10.51.Linear response from 3Gy to

10Gy,and then from 15Gy till the observed dose of 20Gy was observed. The fading in

response for 5 Gy of 6 MV photons was found to be 16.93% in one week and 48.07% in one

month. Dose repeatibility for 5Gy of 6 MV photons was found to have standard deviation of

4.85 %.

Dr. Ahmad Lutfi Yuosoff Supervisor

Professor Dr. Ahmad Zakaria Co Suprvisor

Professor IR.Dr Mariatti Binti Jaafar Co Supervisor

Page 5: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

ii

Dedication

I would like to dedicate my thesis to my beloved parents; Noor Jehan Begum and

Muhammad Ashab uddin (may Allah rest their souls in peace) whom deep love for

me and encouragements was always my source of inspiration and strength, to my

loving and extremely caring wife; Nausheen Fatima, who always stand with me in

facing any kind of trouble and with her utmost efforts either physical or emotional,

made situations favourable for me.

I also want to dedicate my work to my elder brother; Ajaz Tariq to me who is the

sincerest friend and reflection of my father, my loving sisters; Seema Tabassum and

Shabana Anjum; their families, and my sincere in laws, who always remained

concerned and pray for me and my family.

Page 6: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

iii

Acknowledgements

All praises is for Allah subhan wa tala and slawat wassalam to His prophet

Mohammad (salalah wa sallam). Beyond any doubt, what I have achieved is only

due to the blessings of Allah sbhan wa talah whose unmatched kindness made my

dreams true.I would like to express my utmost gratitude to my main

supervisor, Dr. Ahmad Lutfi Yusoff , whose sincere concern, Knowledgeable

advises and kind guidance, from the very first step of my research till the end of my

thesis writing, made my work complete and presentable . I am also thank full to my

co supervisor Professor Ahmad Zakaria and Professor Mariatti Jaafer for their

advises, help and kind concern .

My deepest thanks go to Mr. Nik Ruzman, Mr. Redwan and all the radiation

therapy staff of the Department of Nuclear medicine and Radiation therapy, U.S.M,

for their kind concern and help in irradiation of samples and related works. I also

want to thankful to the staff of CRL (central research lab) of U.S.M, my friend Zaid

and other friends of USM Engineering Campus, the staff of USM Engineering

campus (staff of plastics and rubber section), for their guidance, kind help and

cooperation, regarding the use of equipments and related matters. I acknowledge my

mother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may

Allah rest their souls in peace), deep love and their encouragements to achieve my

goals and their advises and guidance which give me insight in the philosophy of life.

Their sweet memories are my most worthy treasure.

I must mentioned the support and help of my beloved wife Nausheen Fatima who

always stands with me against all odds and making things and situations favourable

for me, with her utmost sincerity, devotion and love. The cooperation of my kids,

Page 7: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

iv

(Umar, Ali, Amna and Ahmed) is something which I would like to acknowledge

with love. I am proud to be father of such enduring, and loving kids. I am deeply

thank full to my brother Ajaz Tariq, my sisters SeemaTabassum, Shabana Anjum

and their families, my brother in law Zia Siddiqui and my in laws, for encouraging

me and my family, and to pray for all of us.

.

Page 8: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

v

List of Contents DEDICATION ....................................................................................................................................... II

ACKNOWLEDGEMENTS ................................................................................................................. III

............................................................................................................................................................ ..IV

LIST OF CONTENTS ........................................................................................................................... V

LIST OF FIGURES ............................................................................................................................ VIII

LIST OF TABLES ............................................................................................................................. XV

LIST OF ABBREVIATIONS ........................................................................................................... XVI

ABSTRAK ....................................................................................................................................... XVII

ABSTRACT ........................................................................................................................................ XX

CHAPTER 1 .......................................................................................................................................... 1

INTRODUCTION.................................................................................................................................. 1

1.1 BACKGROUND ....................................................................................................................... 1 1.2 STATEMENT OF THE PROBLEM ............................................................................................... 2 1.3 PURPOSE OF THE STUDY ........................................................................................................ 3 1.5 SCOPE OF THE STUDY............................................................................................................. 5

CHAPTER 2 .......................................................................................................................................... 6

LITERATURE REVIEW ...................................................................................................................... 6

2.1 RADIATION DOSE AND DOSIMETRY ........................................................................................ 6 2.2 MEDICAL DOSIMETRY ............................................................................................................ 7 2.3 IMPORTANCE OF DOSIMETRY FOR RADIATION THERAPY ........................................................ 7

2.3.1 Treatment planning, treatment planning systems and dosimetry ..................................... 8 2.3.2 Requirement of validating treatment plan ........................................................................ 9

2.4 OVER VIEW OF THE DOSIMETERS ......................................................................................... 10 2.4.1 Ionization chambers ....................................................................................................... 10 2.4.2 Film dosimetry ................................................................................................................ 11

2.4.2.1. Silver halide Radiographic films .......................................................................................... 11 2.4.2.2. Radiochromic films.............................................................................................................. 12

2.4.3. Luminescence dosimetry ............................................................................................ 13 2.4.3.1. Thermoluminescence dosimetry (TLD) .............................................................................. 13 2.4.3.2. Optically stimulated dosimeters .......................................................................................... 15

2.4.3.2.a. CW- OSL ........................................................................................................................ 16 2.4.3.2.b. LM-OSL ........................................................................................................................ 16 2.4.3.2.c. Pulsed OSL .................................................................................................................... 17 2.4.3.3. OSD VERSUS TLD ...................................................................................................... 17

2.4.4. Semi conductor detectors ........................................................................................... 18 2.4.4.a. Diodes ................................................................................................................................. 18 2.4.4.b. PN diodes ............................................................................................................................. 18 2.4.4.c. MOSFET ............................................................................................................................ 20 2.4.4.d. Diamond detector ................................................................................................................ 20

2.4.5. Chemical dosimetry .................................................................................................. 21 2.4.5.a. Frick dosimeter ................................................................................................................... 21

2.4.6. Gel dosimetry ............................................................................................................. 22 2.4.6.a. Fricke gel dosimetry ............................................................................................................ 22 2.4.6.b. Polymer gel dosimetry ....................................................................................................... 23

Page 9: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

vi

2.4.7 Plastic dosimeters.......................................................................................................... 25 2.4.8. Natural and synthetic materials ................................................................................ 27

CHAPTER 3 ........................................................................................................................................ 29

MATERIALS AND METHODS ......................................................................................................... 29

3.1. DOSIMETRIC MATERIALS .................................................................................................... 30 3.1.1. Beryllium Oxide (BeO).............................................................................................. 30 3.1.2. Sodium Chloride (NaCl) ............................................................................................ 31 3.1.3. Potash Alum (K2SO4.24H2O) .................................................................................... 32 3.1.4. Bamboo salt .............................................................................................................. 33 3.1.5. Aluminium Oxide (Al2O3) .......................................................................................... 34

3.2. HOST MATERIALS ............................................................................................................... 34 3.2.1. UVette vial's plastic .................................................................................................. 35 3.2.2. Poly methyl metaacrylate (PMMA) ........................................................................... 35 3.2.3. Shredded branded plastic .......................................................................................... 36 3.2.4. Schematic representation of sample preparation ...................................................... 37

3.3. METHODS ........................................................................................................................... 38 3.4. SAMPLES ............................................................................................................................. 40

3.4.1. Physical form of the sample material ........................................................................ 40 3.4.2. Methods used for the fabrication of plastic sample .................................................. 41

3.4.2.1. Spray technique ................................................................................................................. 41 3.4.2.1.a. Quantification of spray gun parameters ......................................................................... 43 3.4.2.1.b. Characterisation of the spray gun ................................................................................. 45 3.4.2.1.c. Preparation of plastic solution of desired viscosity ........................................................ 45

3.4.2.2. Solution cast technique ........................................................................................................ 46 3.4.2.2.a. Preparation of branded shredded plastic solution ........................................................... 47 3.4.2.2.b. Cast for branded plastic solution and formation of thin films ......................................... 47

3.4.2.3. HOT PLATE TECHNIQUE ............................................................................................. 48 3.4.2.3.a. Branded plastic solution on hot plate .............................................................................. 48 3.4.2.3.b. Melting PMMA grains on hot plate ............................................................................... 49 3.4.2.3.c. Moulded plastic film/thin cut out on hot plate ............................................................... 49

3.4.2.4. Plastic container ................................................................................................................... 51 3.4.2.4.a. Bored branded plastic sheet .......................................................................................... 52 3.4.2.4.b. Uvette vial used for spectrophotometry ......................................................................... 52 3.4.2.4.c. Transparent plastic vials................................................................................................ 53

3.4.2.5 Preparation of plastic cut outs with melt casting ..................................................................... 53 3.4.2.5.1. Preparation of PMMA grains and materials................................................................... 54 3.4.2.5.2. Preparation of samples with melt blending with Haake poly drive ................................ 55 internal mixer and hot press machines ............................................................................................... 55

3.5 EVALUATIONS FOR THE DOSIMETRIC PROPERTIES OF THE SAMPLES .................................... 61 3.5.1. Plastic Containers ..................................................................................................... 61

3.5.1.a. Materials in Raw form ........................................................................................................ 61 3.5.1.b. Anhydrous materials ............................................................................................................ 62 3.5.1.c. Branded plastic sheets bored container ............................................................................... 64 3.5.1.d. Uvette vial ........................................................................................................................... 64 3.5.1.e. Plastic sheet embedding material grains ............................................................................. 65

CHAPTER 4 ........................................................................................................................................ 75

RESULTS AND DISCUSSION .......................................................................................................... 75

4.1. STUDY OF THE OPTICAL PROPERTIES OF PLASTICS USED AS CONTAINERS AND HOST ............ 75 4.2. PLASTIC CONTAINERS .......................................................................................................... 75

4.2.a. Bored branded plastic branded plastic sheet containers .......................................... 75

Page 10: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

vii

4.2.b. Branded plastic vials and host material ..................................................................... 75 4.3. THIN PLASTIC FILM FABRICATION ........................................................................................ 79

4.3.a. Spray technique .......................................................................................................... 79 4.3.b. Solution cast technique .............................................................................................. 86

4.3.b.1. Branded plastic solution casts .............................................................................................. 86 4.3.b.2. Sandwiched layered film .................................................................................................... 89 4.3.b.3. Techniques with hot plate ................................................................................................... 90 4.3.b.4. Branded plastic solution on hot plate .................................................................................. 90

4.3.c. Melting PMMA grains on hot plate ............................................................................... 91 4.3.c.1. Moulded plastic film/thin cut out on hot plate .................................................................... 92

4.3.d. Preparation of plastic cut outs with melt casting ...................................................... 93 4.4. STUDY OF THE FACTORS OF A SAMPLE AFFECTING ITS OPTICAL PROPERTIES PRE AND ........ 96

post irradiation.................................................................................................................. 96

4.4.a. Optical transparency of irradiated plastics of vial and PMMA ............................... 96 4.4.B. EFFECT OF GRAIN SIZE ..................................................................................................... 98 4.5. BASIC DOSIMETRIC SENSITIVITY EVALUATION OF MATERIALS FOR 6 MV PHOTONS .......... 102

4.5.a. Materials in plastic vial with oil .............................................................................. 102 4.5.a.1. Aluminium oxide (Al2O3 ) ................................................................................................ 102 4.5.a.2. Beryllium oxide (BeO) ..................................................................................................... 103 4.5.a.3. Potash alum ....................................................................................................................... 104 4.5.a.4. Ammonium chloride (Noshadar) ....................................................................................... 105 4.5.a.5 Bamboo salt ......................................................................................................................... 105 4.5.a.6. N1 ..................................................................................................................................... 107 4.5.a.7. N2 ..................................................................................................................................... 108 4.5.a.8 N3 ..................................................................................................................................... 108 4.5.a.9. N4 ..................................................................................................................................... 109 4.5.a.10. N5 ..................................................................................................................................... 109 4.5.a.11. N6 ...................................................................................................................................... 110 4.5.a.12. N7 ...................................................................................................................................... 111 4.5.a.13. N8 ..................................................................................................................................... 111

4.5.b. Material embedded in PMMA ................................................................................. 112 4.5.b.1. Aluminium oxide (Al2O3) ................................................................................................ 112 4.5.b.2. Beryllium oxide (BeO) ..................................................................................................... 113 4.5.b.3 N1 .......................................................................................................................................... 113 4.5.b.4. N8 .......................................................................................................................................... 114

4.6 STUDY OF ENHANCEMENT IN 6 MV PHOTONS SENSITIVITY POST TREATMENT WITH .......... 115 9 MEV ELECTRON OF DIFFERENT MATERIALS ................................................................................ 115

4.6.a. UVette vial .............................................................................................................. 115 4.6.b. Transparent vial with optically clear oil ................................................................. 115

4.6.b.1. Aluminium oxide (Al2O3) ................................................................................................ 116 4.6.b.2. Beryllium oxide (BeO) ...................................................................................................... 117

4.6.b.3. Ammonium chloride (Noshadar) ............................................................................. 119 4.6.b.4. BAMBOO SALT .............................................................................................................. 120 4.6.b.5. Potash Alum ...................................................................................................................... 121 4.6.b.6. N8 ..................................................................................................................................... 122

4.7. N1 (STUDY OF ENHANCEMENT IN X6 SENSITIVITY DUE TO DIFFERENT E9 DOSES) ................. 123 4.7.a. Study of enhancement in X6 sensitivity of N1 due to 1600 Gy of 9 MeV ....................... 124 electrons .................................................................................................................................... 124 4.7.b. Study of enhancement in X6 sensitivity of N1 due to different doses of .................. 125 9 MeV electrons ......................................................................................................................... 125 4.7.c. Study of enhancement in X6 sensitivity due to 400 Gy /Day E9 doses ....................... 127

Page 11: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

viii

( plastic vial & oil) .................................................................................................................... 127 4.7.d. Estimating suitable high dose E9 for optimum 6 MV photon .................................. 131 sensitivity ................................................................................................................................... 131 4.7.e. Single dose of 800 Gy of E9 ....................................................................................... 131 4.7.g. Dosimetric evaluation of N1 sample by fractioned dose of 400 Gy/day to ............. 135 cumulative dose of 800 Gy of 9 MeV electrons ........................................................................ 135 4.7.h. Comparison of graphs of N1 sample (final) for different doses of 6 MV ................. 141 photons post bleaching from 800 Gy 9 MeV electrons fractionated ....................................... 141 treatment .................................................................................................................................... 141

4.8 N1 FRESH PLASTIC CUT OUTS (UN IRRADIATED TO E9) .............................................. 145

4.9 RESPONSE LINEARITY VERSUS DOSE ............................................................................... 150

4.10 DOSE REPEATABILITY ........................................................................................................ 154 4.11 FADING .............................................................................................................................. 155 4.12 COMPARISON OF FABRICATED PMMA N1COMPOSITE DOSIMETER WITH........................... 155 PRESAGE...................................................................................................................................... 155

CHAPTER 5 ...................................................................................................................................... 157

CONCLUSIONS AND RECOMMENDATIONS ............................................................................ 157

5.1 CONCLUSIONS ................................................................................................................... 157 5.2 RECOMMENDATIONS FOR FUTURE RESEARCH .................................................................... 160

REFFERENCES ................................................................................................................................ 161

APPENDICES

Page 12: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

ix

List of Figures

Figure 2.1. Model for thermo luminescence ...................................................................................... 14 Figure 3.1: Raw Potash Alum ............................................................................................................ 33 Figure 3.2: Shredded UVette vial plastic ........................................................................................ 35 Figure 3.3: PMMA grains of 2 to 3 mm in dimension ...................................................................... 36 Figure 3.4: Shredded branded plastic .............................................................................................. 36 Figure: 3.5 Materials for the proposed plastic composite radiation dosimeter ................................. 37 Figure 3.6: Shimadzu UV 1800 Spectrophotometer .......................................................................... 38 Figure 3.7: Cary eclipse florescence spectrophotometer for florescence study ................................... 39 Figure 3.8: Optical bleaching of samples with two bulbs of 100 watts................................................ 39 Figure 3.9: Spray gun ........................................................................................................................ 42 Figure 3.10: External mix .............................................................................................................. .....42 Figure 3.11: Internal mix.....................................................................................................................42 Figure 3.12 Mini Spray gun K-3A .................................................................................................... 43 Figure 3.13 Casts made by ordinary glass slides for plastic solution ................................................. 47 Figure 3.14 : Local brand plastic solution heated on hot plate .......................................................... 48 Figure: 3.15: PMMA grains melted on hot plate at 190 oC ................................................................ 49 Figure 3.16: Mould fabrication with glass slides for plastic solutions ............................................... 50 Figure 3.17: Mini 3D oven fabricated with glass slides and heated over hot plate .................... 50 Figure 3.18:Uvette special plastic vials for spectroscopic .................................................................. 51 studies for wavelength ranges 220 nm to 1600 nm .............................................................................. 51 Figure 3.19: Plastic sheet board containers ...................................................................................... 52 Figure 3.20: Transparent plastic vial ................................................................................................ 53 Figure 3.21: Planetary Mono Mill grinder used for grinding samples .............................................. 55 Figure 3.22: Haake poly drive internal mixer .................................................................................... 56 Figure 3.23: Hot press machine for moulding sample sheets by using hot and cold press .............. 57 Figure 3.24: Mechanical sieve for having different grain sizes of material ...................................... 59 Figure 3.25: Summarized steps of sample preparation ....................................................................... 60 Figure 3.26: Materials in transparent plastic vial kept in light sleeves ...................................... 62 Figure 3.27 : Plastic samples of N1 in light tight sleeves \ ............................................................. 65 Figure 3.28: Labelled plastic samples .............................................................................................. 65 Figure 3.29: Labelled plastic samples of N1 ...................................................................................... 66 Figure 4.1: Optical transparencies of, transparent empty vial and with oil, material free ................. 76 PMMA cut out empty UVette vial, nothing in spectrophotometer ........................................................ 76 Figure 4.2 : Ratio of Optical Transparencies of empty Uvette vial to Empty ..................................... 77 plastic vial (Un-normalised curves) ..................................................................................................... 77 Figure 4.3: Ratio of Optical transparencies of Plastic vial with clear oil to ..................................... 78 same empty plastic vial (unnormalised curves) .................................................................................. 78 Figure 4.4: Ratio of Optical transparencies of material free PMMA ................................................. 79 clear plastic cut out (2 mm) to empty plastic vial (un normalised curves) .......................................... 79 Figure 4.5: Dried sprayed plastic solution with solute to solvent ....................................................... 82 proportion 1.967 % .............................................................................................................................. 82 Figure 4.6: Dried sprayed plastic solution with solute to solvent proportion 3.622 % ....................... 83 Figure 4.7: Dried sprayed plastic solution with solute to solvent proportion 1.967 %.......................83 Figure 4.8: Dried sprayed plastic solution with solute to solvent proportion 3.622 % .................... .83 Figure 4.9: Dried sprayed plastic solution with h solute to solvent proportion 1.967 %......................83 Figure 4.10: Dried sprayed plastic solution with solute to solvent proportion 3.622 %.................... .83 Figure 4.11: Dried sprayed plastic solution with solute .................................................................... 84 to solvent proportion 4.89 % .............................................................................................................. 84 Figure 4.12: Dried sprayed plastic solution ....................................................................................... 85 with solute to solvent proportion 2.849 %........................................................................................... 85 Figure 4.13: Dried sprayed plastic solution ........................................................................................ 85 with solute to solvent proportion 2.849 %........................................................................................... 85 Figure 4.14: Branded plastic solution containing N1 grains, dried over glass .................................. 86 slide at room temperature ................................................................................................................... 86

Page 13: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

x

Figure 4:15: Branded plastic solution containing N1 grains ............................................................. 87 dried in petty dish at room temperature ............................................................................................... 87 Figure 4.16: Branded plastic solution containing N1 grains dried in hot air drier ............................. 88 chamber at 50 0C for 24 hours ............................................................................................................. 88 Figure 4.17: Branded plastic solution containing N1 grains dried in hot air ..................................... 88 drier chamber at 700C for 24 hour ...................................................................................................... 88 Figure 4.18: Layered plastic film made by solution cast method ......................................................... 89 Figure 4.19 : Branded plastic solution dried over hot plate ............................................................... 90 Figure 4.20 : Melted PMMA grains over glass slide using hot plate ................................................... 91 Figure 4.21: Solid plastic cut out embedding N1 grains moulded on hot plate ................................. 92 Figure 4.22: PMMA N1(composite)sample..... .................................................................................. 93 Figure 4.23 :Material free PMMA sample.............................................................................................93 Figure 4.24 : PMMA N8 (composite) sample ...................................................................................... 93 Figure 4.25 :PMMA Al2O3 (composite) sample .................................................................................. 94 Figure 4.26 : PMMA BeO (composite) sample......................................................................................94 Figure 4.27 : The final plastic sample of N1125 µm with 40 to 60 ratio with PMMA (wt/wt),collected from melt mixing machine......................................................................................................................95 Figure 4.28 :The final plastic sample of N1125 µm with 40 to 60 ratio with PMMA (wt/wt),as plastic sheet (2 mm, obtained after hot press of sample...................................................................................95 Figure 4.29 SEM pictures of the final PMMA N1 (grounded,125 μm),40/60 (wt/wt) sample ............ 96 Figure 4.30: Ratio of normalised curves of optical transparencies of vial with oil, irradiated with .. 97 400 G y/day dose of 9 MeV electrons, with respect to un irradiated. .............................................. 97 Figure 4.31 Ratio of normalised curves of optical transparencies of material free PMMA ............... 97 (2 mm) cut out irradiated with different 9 MeV electrons with respect to un irradiated..................... 97 Figure 4.32: Ratio of optical transparencies of un irradiated N1raw to ............................................. 99 N1(125 μm) grounded .......................................................................................................................... 99 Figure 4.33: Ratio of optical transparencies of un irradiated raw N8 to N8 .................................... 100 Grounded (63 µm) .............................................................................................................................. 100 Figure 4.34: Optical transparency curves (normalised) of N8 and N1 (raw, unexposed),samples .. 101 in plastic vial immersed in clear oil .................................................................................................. 101 Figure 4.35: Ratio of the relative optical transparencies of un irradiated, N1(raw .......................... 101 to grounded N1, to N8 (raw to grounded N8 ) ................................................................................. 101 Figure 4.36: Ratio of normalised curves of optical transparencies Al2O3 in vial with oil, ............... 103 irradiated with 2 Gy dose of 6 MV photons, with respect to un irradiated sample ........................... 103 Figure 4.37: Ratio of normalised curves of optical transparencies BeO in vial with ........................ 104 oil, irradiated with 2 Gy dose of 6 MVphotons, with respect to un irradiated sample ...................... 104 Figure 4.38: Ratio of normalised curves of optical transparencies Potash alum in .......................... 104 vial with oil, irradiated with 2 Gy dose of 6 MV photons, with respect to un .................................. 104 irradiated sample ............................................................................................................................... 104 Figure 4.39 : Ratio of normalised curves of optical transparencies of ammonium ........................... 105 Chloride in vial with oil irradiated with 2 Gy dose of X6, with respect to ....................................... 105 un irradiated sample ......................................................................................................................... 105 Figure 4.40: Ratio of normalised curves of optical transparencies Bamboo salt in .......................... 106 vial with oil, irradiated with 2 Gy dose of 6 MV photons, with respect to un ................................... 106 irradiated sample .............................................................................................................................. 106 Figure 4.41: Ratio of normalised curves of optical transparencies N1 salt ....................................... 107 in vial with oil ,irradiated with 2 Gy dose of 6 MV photons, with respect ........................................ 107 to un irradiated sample ...................................................................................................................... 107 Figure 4.42: Ratio of normalised curves of optical transparencies N2 salt in................................... 108 vial with oil, irradiated with 2 Gy dose of 6 MV photons, with respect to un .................................. 108 irradiated sample ............................................................................................................................... 108 Figure 4.43: Ratio of normalised curves of optical transparencies N3 salt in vial with .................... 108 oil, irradiated with 2 Gy dose of 6 MV photons, with respect to un irradiated sample ..................... 108 Figure 4.44: Ratio of normalised curves of optical transparencies N4 salt in vial with oil, .............. 109 irradiated with 2 Gy dose of 6 MV photons, with respect to un irradiated sample ........................... 109 Figure 4.45: Ratio of normalised curves of optical transparencies N5 salt in vial with .................... 109 oil, irradiated with 2 Gy dose of 6 MV photons, with respect to un irradiated sample ..................... 109

Page 14: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

xi

Figure 4.46: Ratio of normalised curves of optical transparencies N6 salt in vial with oil, .............. 110 irradiated with 2 Gy dose of 6 MV photons, with respect to un irradiated sample ........................... 110 Figure 4.47: Ratio of normalised curves of optical transparencies N7 salt in vial with oil, .............. 111 irradiated with 2 Gy dose of 6 MV photons, with respect to un irradiated sample ........................... 111 Figure 4.48 : Ratio of normalised curves of optical transparencies N8 salt in vial with oil, ............. 111 irradiated with 2 Gy dose of 6 MV photons, with respect to un irradiated sample ........................... 111 Figure 4.49: Ratio of optical transparencies of plastic cut outs of PMMA (0.5mm thick)................ 112 embedding Al2O3 grains irradiated to 2 Gy of 6 MV photons to its un irradiated state .................. 112 Figure 4.50: Ratio of optical transparencies of plastic cut outs of PMMA (0.5mm thick)................ 113 embedding BeO grains irradiated to 2 Gy of 6 MV photons to its un irradiated state .................... 113 Figure 4.51: Ratio of optical transparencies of plastic cut outs of PMMA (0.5mm thick)................ 114 embedding N1 grains irradiated to 2 Gy of 6 MV photons to its un irradiated state ....................... 114 Figure 4.52 :Ratio of optical transparencies of plastic cut outs of PMMA (2 mm) .......................... 114 embedding N8 grains irradiated to 5 Gy of 6 MV photons to its un irradiated state ........................ 114 Figure 4.53 :Ratio of normalised optical transparencies curves of Al2O3 (in vial with oil), ........... 116 irradiated with 2 Gy dose of 6 MV photons with respect to un irradiated sample ............................ 116 Figure 4.54: Ratio of normalised optical transparencies curves of Al2O3 (in vial with .................... 116 oil), irradiated with 2 Gy dose of 6 MV photons (post irradiation of 2 k Gy of E9 ), ...................... 116 with respect to un irradiated sample .................................................................................................. 116 Figure 4.55 : Ratio of normalised optical transparency curves of Al2O3 (in vial with oil), irradiated with 2 Gy of 6 MV photons (post irradiation of 2 k Gy of 9 MeV 2 Gy dose of 6 MV photons unexposed to electrons ....................................................................................................................... 117 Figure 4.56 : Ratio of normalised curves of optical transparencies of BeO (in vial with oil), ......... 118 irradiated with 2 Gy dose of 6 MV photons, with respect to un irradiated sample ............................ 118 Figure 4.57: Ratio of normalised optical transparencies curves of BeO (in vial.............................. 118 oil), irradiated with 2 Gy dose of 6 MV photons post irradiation of 2 k Gy of 9 MeV ...................... 118 electrons, with respect to un irradiated sample ................................................................................. 118 Figure 4.58: Ratio of normalised optical transparencies curves of BeO (in vial with oil), .............. 119 irradiated with 2 Gy dose of 6 MV photons post irradiation of 2 k Gy of 9 MeV electrons ............. 119 with respect to 2 Gy dose of 6 MV photons unexposed to electrons ................................................. 119 Figure 4.59 :Ratio of normalised optical transparencies curves of ammonium chloride ................. 119 ( Noshadar) in(vial with oil), irradiated with 5 Gy of 6 MV photons, to unexposed sample ............. 119 Figure 4.60: Ratio of normalised optical transparencies curves of Bamboo salt (vial with .............. 120 oil), irradiated with 2 Gy of 6 MV photons, with respect to unexposed sample ............................... 120 Figure 4.61: Ratio of optical transparencies of potash alum, 2 Gy of X6 exposed to post .............. 121 bleaching of 80 Gy of E9 ................................................................................................................... 121 Figure 4.62: Ratio of optical transparencies of potash alum 2 Gy of X6 exposed to post ................ 122 bleaching of 200 Gy of E9 ............................................................................................................... 122 Figure 4.63: Ratio of normalised optical transparencies curves of N8 (vial with oil, average of ..... 123 three samples),irradiated with 2 Gy of 6 MV photons, post irradiation of first E9 treatment of ..... 123 80 Gy of 9 MeV electrons and optically bleached, with respect to unexposed sample ................... 123 Figure 4.64: Ratio of normalised curves of N1( vial in oil ) optical transparencies, post 2 Gy dose of 6 MV photons and unexposed sample ............................................................................................ 124 Figure 4.65: Ratio of normalised curves of N1( vial in oil )optical transparencies, after 2 Gy dose 124 Figure 4.66 : Ratio of normalised curves of N1( vial in oil ) photons, post optical bleaching of .... 125 1600 Gy of 9 MeV electrons with respect to 2 Gy of 6 MV photons delivered to fresh sample ...... 125 Figure 4.67:Average of the Ratio of optical transparencies of N1( vial in oil ) after 2 Gy of X6 .... 125 photons (post 200 Gy E9 treatment )with respect to 2 Gy of X6 un irradiated sample ................... 125 Figure 4.68: Average of the Ratio of optical transparencies of N1( vial in oil ) after 2 Gy ............. 126 of 6 MV photons ( post 200 Gy of E9 treatment) respect to 2 G yX6 to fresh sample ...................... 126 Figure 4.69: Average of the Ratio of optical transparencies of N1(oil in vial ) after 2 Gy ............. 126 of 6 MV photons ( post 200 Gy of E9 treatment) respect to 2 Gy of X6 to fresh sample ................ 126 Figure 4.70: Average of the Ratio of optical transparencies of N1oil in vial after 2 Gy ................. 127 6 MV photons post 100 Gy of E9 treatment with respect to 2 Gy of X6 to fresh ............................. 127 sample ................................................................................................................................................ 127 Figure 4.71: Ratio of optical transparencies post 5 GyX6 to un exposed sample ............................ 127

Page 15: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

xii

Figure 4.72: Ratio of optical transparencies of 5 Gy X6 ( post bleaching of 400 Gy of E9) to 5 Gy 6 MV photons exposed, fresh sample..........................................................................................128 Figure 4.73:Ratio of optical transparencies of samples ( post treatment of 800 Gy of E9 ) to ......... 128 fresh samples, for 5 Gy dose of 6 MV photons .................................................................................. 128 Figure 4.74: Ratio of optical transparencies of, 5 Gy X6 ( post bleaching cumulative 1600 Gy ..... 129 E9) to 5 Gy 6 MV photons exposed, fresh sample ........................................................................... 129 Figure 4.75: Ratio of optical transparencies of, samples ( post treatment with 2000 Gy of ............ 129 E9 to fresh samples, for 5 Gy dose of 6 MV photons. ..................................................................... 129 Figure 4.76:Ratio of optical transparencies of, samples ( post treatment with 2400 Gy of .............. 130 E9) to fresh samples, for 5 Gy dose of X6 photons. ....................................................................... 130 Figure 4.77: Ratio of optical transparencies of samples ( post treatment with 2800 Gy .................. 130 of E9) to fresh samples for 5 Gy dose of X6 photons .................................................................... 130 Figure 4.78: Optical transparency of N1 (final plastic sample 2 mm exposed with 0.5 Gy of X6 post treatment with E9 single dose of 800 Gy ......................................................................................... 132 Figure 4.79: Optical transparency of N1 final plastic sample 2 mm exposed with 1 Gy X6 post .... 132 treatment with E9 single dose of 800 Gy .......................................................................................... 132 Figure 4.80: Optical transparency of N1 final plastic sample (2 mm) exposed with 1.5 Gy of X6 . 133 post treatment with E9 single dose of 800 Gy .................................................................................. 133 Figure 4.81: Comparison of N1 (2 mm) samples ratios of optical transparencies for different X6 doses (after optical bleaching from 800 Gy single dose of E9) to its freshly bleached state ..................... 133 Figure 4.82: Optical transparency of N1 (final plastic sample 2 mm) exposed with 0.5 Gy X6 post treatment with E9 single dose of 1200 Gy ........................................................................................ 134 Figure 4.83 : Optical transparency of N1 (final plastic sample(2 mm)) exposed with 1.5 Gy X6 post treatment with E9 single dose of 1200 Gy ........................................................................................ 134 Figure 4.84 : SEM images of 800 Gy (single)dose exposed N1 final) plastic sample ..................... 135 Figure 4.85: Ratio of the average optical transparencies, of N1(40/60 2 mm) for 0.5 Gy of ........... 136 6 MV photons post treatment with 800 Gy of E9, to its bleached state ............................................ 136 Figure 4.86: Ratio of the average optical transparencies, of N1 (40/60 2 mm) for 1 Gy .................. 136 of 6 MV photons post treatment with 800 Gy of E9, to its bleached state .................................... 136 Figure 4.87: Ratio of the average optical transparencies, of N1(40/60 2 mm) for 1.5 Gy of ........... 137 X6 photons post treatment with 800 Gyof E9, to its bleached state .............................................. 137 Figure 4.88: Ratio of the average optical transparencies, of N1(40/60 2 mm) for 2 Gy of ........... 137 6 MV photons post treatment with 800 Gy of E9, to its bleached state ......................................... 137 Figure 4.89: Ratio of the average optical transparencies, of N1(40/60 2 mm) for 3 Gy of .............. 138 X6 post treatment with 800 Gy of E9, to its bleached state .......................................................... 138 Figure 4.90: Ratio of the average optical transparencies, of N1(40/60 2 mm) for 5 Gy ................. 138 X6 photons post treatment with 800 Gy of E9, to its bleached state ............................................ 138 Figure 4.91: Ratio of the average optical transparencies, of N1(40/60 2 mm) for 7.5 Gy of ........... 139 X6 photons post treatment with 800 Gy of E9, to its bleached state ............................................... 139 Figure 4.92: Ratio of the average optical transparencies of N1(40/60 2 mm) for10 Gy of .............. 139 6 MV photons post treatment with 800 Gy of E9, to its bleached state ........................................... 139 Figure 4.93: Ratio of the average optical transparencies of N1(40/60 2 mm) for 15 Gy ................ 140 of X6 photons post treatment with 800 Gy of E9, to its bleached state .......................................... 140 Figure 4.94: Ratio of the average optical transparencies, of N1(40/60 2 mm) for 20 Gy ............... 140 of 6 MV photons post treatment with 800 Gy of E9, to its bleached state. ..................................... 140 Figure 4.95: Ratio of optical transparencies of 0.5 Gy X6 (post bleaching of fractioned 800 Gy E9) to 0.5 Gy X6 given to unexposed sample .......................................................................................... 141 Figure 4.96: Ratio of optical transparencies of 1 Gy X6 (post bleaching of fractioned 800 Gy E9) to 1 Gy X6 given to unexposed sample ................................................................................................. 142 Figure 4.97: Ratio of optical transparencies of N1 plastic sample for 2 Gy of 6 MV photons (post 142 bleaching from fractionated dose of 800 Gy of 9 MeV electron) to the same dosed fresh sample .... 142 Figure 4.98: Ratio of optical transparencies of N1 plastic sample for 5 Gy of 6 MV photons (post 143 bleaching from fractionated dose of 800 Gy of E9) to the same dosed fresh sample ........................ 143 Figure 4.99: Ratio of optical transparencies of N1 plastic sample for 7.5 Gy of 6 MV photons (post bleaching from fractionated dose of 800 Gy of E9) to the same dosed fresh sample ....................... 143 Figure 4.100 :Ratio of optical transparencies of N1 plastic sample for 15 Gy of 6 MV photons (post . bleaching from fractionated dose of 800 Gy of E9) to the same dosed fresh sample ...................... 144

Page 16: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

xiii

Figure 4.101: Ratio of optical transparencies of N1 plastic sample for 20 Gy of 6 MV photons (post bleaching from fractionated dose of 800 Gy of E9) to the same dosed fresh sample ....................... 144 Figure 4.102: Optical transparency of fresh N1 (final sample) for 0.5 Gy of X6 with respect to their un irradiated fresh sample transparency ........................................................................................... 145 Figure 4.103 : Optical transparency of fresh N1 (final sample) for 1 Gy of X6 with respect to their un irradiated fresh sample transparency ........................................................................................... 145 Figure 4.104 : Optical transparency of fresh N1 (final sample) for 2 Gy of X6 with respect to their un irradiated fresh sample transparency ........................................................................................... 146 Figure 4.105 : Optical transparency of fresh N1 (final sample) for 3 Gy of X6 with respect to .. 146 their un irradiated fresh sample transparency ................................................................................... 146 Figure 4.106 : Optical transparency of fresh N1 (final sample) for 5 Gy of X6 with respect to their un irradiated fresh sample transparency .......................................................................................... 147 Figure 4.107 : Optical transparency of fresh N1 (final sample) for 7.5 Gy of X6 with respect to their un irradiated fresh sample transparency ................................................................................... 147 Figure 4.108 : Optical transparency of fresh N1 (final sample) for 10 Gy of X6 with respect to .. 148 their un irradiated fresh sample transparency ................................................................................... 148 Figure 4.109 : Optical transparency of fresh N1 (final sample) for 15 Gy of X6 with respect to their un irradiated fresh sample transparency ........................................................................................... 148 Figure 4.110 : Optical transparency of fresh N1 (final sample) for 20 Gy of X6 with respect to .. 149 their un irradiated fresh sample transparency ................................................................................... 149 Figure 4.111: Normalized optical transparency curve of fresh N1 plastic sample (2 mm) post irradiation of 20 Gy of X6 .............................................................................................................. 150 Figure 4.112 : Normalized optical transparency curve of fresh N1 plastic sample (2 mm) post irradiation of 15 Gy of X6 ............................................................................................................... 150 Figure 4.113 : Normalized optical transparency curve of fresh N1 plastic sample (2 mm) post irradiation of 10 Gy of X6 ....................................................................................................... 151 Figure 4.114 : Normalized optical transparency curve of fresh N1 plastic sample (2 mm) post irradiation of 7.5 Gy of X6 ..................................................................................................... 151 Figure 4.115 : Normalized optical transparency curve of fresh N1 plastic sample (2 mm) post irradiation of 5 Gy of X6 ......................................................................................................... 151 Figure 4.116 : Normalized optical transparency curve of fresh N1 plastic sample (2 mm) post ....... 152 irradiation of 3 Gy of X6 ................................................................................................................ 152 Figure 4.117 : Dose versus response curve of fresh N1 plastic samples ......................................... 153

Page 17: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

xiv

Page 18: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

xv

List of tables

Tables Page no

3.1 Summarized dosimetric experiments .....................................................................69 - 74

Page 19: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

xvi

List of abbreviations BANG BIS Acrylamide Nitrogen Gelatin BeO Beryllium Oxide CT Computed Technology CW-OSL Continuous waves optically stimulated dosimetery CCD Charge coupled device EBT External Beam Therapy E9 9 MeV Electrons HDR High Dose Rate HVLP High volume low pressure IMRT Intensity modulated Radiation Therapy LM-OSL Linearly modulated optically stimulated luminescence LVLP Low volume low pressure MOSFET Metal Oxide semi-conductor Field Effect Transistor MRI Magnetic Resonance imaging MV Million volt OCT Optical CT scan OD Optical Density OSL Optically stimulated luminescence PC Poly carbonates PE Poly ethylene PMMA Poly methyl meta acrylate PMT Photo multiplier tube PVA Poly vinyl alcohol RIA Radiation induced attenuation RTC Relative transparency curve SRS Stereo radio surgery TL Thermoluminescence TLD Thermo Luminescence dosimeter TPS Treatment planning system UV Ultra violet XO Xylenol Orange X6 6 MV photons

Page 20: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

xvii

FABRIKASI DOSIMETER GUNASEMULA SINAR-X

BERASASKAN PLASTIK UNTUK RADIOTERAPI

ABSTRAK

Tujuan kajian ini adalah untuk menghasilkan dosimeter komposit plastik untuk

mengukur dos sinaran foton sinar-X yang digunakan dalam radioterapi. Kriteria

untuk dosimeter yang dicadangkan adalah kos yang rendah, penggunaan yang

mudah, bahan bukan toksik, dan boleh diguna semula. Sebagai dosimeter, ia harus

mempunyai sensitiviti dan kebolehpercayaan yang cukup dalam lingkungan ukuran

dos kepentingan dalam radioterapi.

Dalam membuat dosimeter berasaskan plastik, zarah bahan aktif yang sensitif kepada

sinaran tertanam secara seragam, di dalam posisi yang tetap, dalam plastik lut sinar.

Perubahan dalam spektrum transmisi optik bagi bahan telah digunakan sebagai

kaedah bacaan, menggunakan Shimadzu UV 1800 spektrofotometer. Perubahan nilai

ini dikalibrasi untuk mendapatkan nilai dos serapan. Pelunturan optik sampel yang

telah menerima dedahan sinaran dilakukan dengan menggunakan dua buah mentol

biasa berkuasa 100 watt dengan dedahan selama 18 jam.

Antara tiga belas bahan yang telah dipilih untuk kajian kesesuaian mereka sebagai

bahan aktif terhadap sinaran, natrium klorida tulen gred penyelidikan didapati

merupakan bahan yang paling sesuai. Bahan-bahan lain yang turut dikaji ialah

berilium oksida, tujuh jenis garam yang boleh dimakan dari di pasaran tempatan,

aluminium oksida tulen, kalium alum (tawas), ammonium klorida (Noshadar) dan

garam buluh. PMMA (polymethyl metaacrylate) telah dipilih sebagai bahan hos.

Sampel plastik yang baik telah dapat dihasilkan menggunakan Haake poly drive

pencampur dalaman dan mesin tekan panas, berbentuk lembaran plastik setebal

Page 21: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

xviii

2 mm, dengan zarah natrium klorida (125 μm) tertanam secara seragam dalam nisbah

berat 40/60 natrium klorida kepada PMMA. Kesemua penilaian bahan aktif telah

dijalankan dengan sampel diletakkan di dalam botol plastik lutsinar yang

mengandungi minyak yang lut sinar. Botol plastik ini didapati mempunyai ciri-ciri

o p t i k y a n g s a n g a t d e k a t s a m p e l P M M A t a n p a b a h a n a k t i f .

Pemerhartian mendapati bahawa rawatan sampel menggunakan dos elektron 9 MeV,

sehingga 2800 Gy, tidak mengubah sensitiviti sampel terhadap sinaran. Sampel

plastik akhir yang dinyatakan di atas, mengandungi natrium klorida yang dilabelkan

sebagai N1, telah menunjukkan tindak balas kepada foton 6 MV, dengan tindak balas

dos didapati linear dalam julat 3 Gy hingga 10 Gy. Sensitiviti kelihatan menurun

antara 10 Gy hingga15 Gy, selepas itu sensitiviti kelihatan kembali untuk dos

terakhir yang dinilai, iaitu 20 Gy. Penurunan sensitiviti pada 15 Gy berkemungkinan

besar berpunca daripada pengurangan elektron terperangkap, sepatutnya

menghasilkan penyerapan optik dalam julat 395 kepada 550nm, disebabkan

penukaran-ke atas pendarfluor hasil daripada penyerapan semasa ukuran, dalam julat

570-625 nm. Nilai Zeff bagi dosimeter berasaskan plastik PMMA yang mengandungi

40% N1 mengikut berat telah dikira dan dipastikan sebagai 10.51 amu. Sampel N1

yang mendapat dos sinaran 5 Gy 6 MV foton telah menunjukkan nilai kepudaran

16.92% untuk satu minggu dan 48.06% untuk satu bulan. Kebolehulangan dos

dianggarkan daripada tiga ujian menggunakan tiga sampel yang berasingan dengan

nilai dos serapan 5 Gy 6 MV foton. Dos sinaran telah diberikan selang 4 hari selepas

pelunturan optik selama 18 jam. Kawasan lembah pada spektra transmisi optik yang

sepadan dengan dos serapan didapati mempunyai nilai purata 96.4 unit-arbitrari

dengan sisihan piawai 4.85%. Kawasan lembah di dalam lekuk transmisi tersebut

terletak dalam julat 460 nm hingga 472nm. Dosimeter berasaskan plastik yang

Page 22: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

xix

terhasil didapati mantap, murah, mudah untuk digunakan dan boleh diguna semula.

Ia didapati sesuai untuk mengukur dos X-ray foton, dalam julat respons dos linear,

dari 3 Gy hingga 10 Gy, dan untuk julat dos 15 Gy dan 20 Gy.

Page 23: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

xx

FABRICATION OF A REUSEABLE PLASTIC BASED X-RAY

DOSIMETER FOR RADIOTHERAPY

ABSTRACT

The aim of this study was to fabricate a plastic composite dosimeter for the

measurement of radiation doses of X -ray photons used in radiotherapy. The criteria

for the proposed dosimeter were low cost, ease of use, non toxic, and reusable.

As a dosimeter, it should have acceptable sensitivity and reliability in the dose

measurements range of interest in radiotherapy.

In making plastics based dosimeters, radiation sensitive material grains were

uniformly embedded in transparent plastic host where they are kept spatially

fixed. Changes in optical transmission spectra of the materials was used as readout

method, using Shimadzu UV 1800 spectrophotometer. These changes were than

calibrated to get the absorbed doses. Optical bleaching of the irradiated samples were

performed with two 100 watts ordinary bulbs for 18 hours. Among the thirteen

materials that were selected for the study of their suitability as the embedding

material, research grade pure sodium chloride was found to be the most suitable

material. Other materials studied were beryllium oxide, seven different types of

edible salts available in local markets, un doped aluminium oxide, potash alum

(tawas), ammonium chloride (Noshadar) and bamboo salt. PMMA (polymethyl

metaacrylate) was chosen as the host material. The final plastic sample was produced

using Haake poly drive internal mixer and hot press machine, as 2 mm thick plastic

sheet, with uniformly embedded grounded (125µm) pure sodium chloride grains in

40/60 sodium chloride to PMMA wt/wt ratio. All the evaluations of embedding

materials were carried out with their samples placed inside a transparent plastic vial

Page 24: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

xxi

containing optically clear oil. This plastic vial was found to have optical properties

very close to the final, material free, PMMA sample. It was observed that treatment

of samples using 9 MeV electron doses of up to 2800 Gy does not changed their

sensitivity. The aforementioned final plastic sample, containing sodium chloride

labelled as N1, have shown response to 6 MV photons with a linear dose response

from 3 Gy to10 Gy. The sensitivity appeared to be decreased between 10 Gy to15

Gy, after which appeared to recover for the last observed dose of 20 Gy. The

decrease in 15 Gy response is postulated to be due to the depletion of the trapped

electrons of interest, supposed to produce optical absorption at 395 to 550nm, due to

fluorescence up-conversion resulting from 570 to 625 nm absorption during

measurements. The Zeff of the PMMA plastic based dosimeter containing 40% N1

by weight was calculated to be 10.51 a.m.u. The N1 samples irradiated with 5 Gy

6MV photons has shown fading of 16.92% and 48.06% for one week and one month

respectively. The dose reproducibility was estimated with the average response of

three fresh samples to 5 Gy 6 MV photons. The doses were delivered for alternate 4

days post optical bleaching of 18 hours. The area of dip in the percentage

transmission curve corresponding the absorbed dose was found to have an

average value of 96.4 arbitrary area units with standard deviation of 4.85%. The

peak values of the dips were found to be between 460 nm to 472 nm. The fabricated

plastic based dosimeter was found to be robust, cheap, easy to use and reusable and

suitable for measuring doses of X-ray photons, in the dose ranges of linear responses,

3 G y to10 G y and fo r obse rved h igher doses , 15 G y and 20 G y.

Page 25: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

1

CHAPTER 1

INTRODUCTION

1.1 Background A variety of dosimeters are in use for different purposes these days. They vary in

their sensitivities for different radiations, range of measurable doses, physical form

and suitability for a particular usage. Although achievements have been made in

developing materials that are very sensitive and can work at very high doses but

they fall in the advanced materials category and have very costly.

Some introduced dosimeters have high dose threshold and they are difficult to

develop due to, their toxicity (plastic dosimeters), their sensitivity to impurities and

atmospheric conditions (Fricke chemical dosimeters and Fricke gel dosimeter,

polymer gel dosimeter).

To use chemical and gel dosimeters the introduction of the container not only

made its routine usage difficult but required corrections in dose estimation

made its usage difficult . Further the extraction of dose information in most of the

dosimeters required diagnostic machines such as MRI, CT scan and Ultra sound

which may be difficult due to availability of these machines at for immediate

dosimetric use. Regarding reusability only ionization chambers, OSLD's and TLD's

are reusable.

Page 26: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

2

1.2 Statement of the problem

Although the radiation dosimeters used these days have covered specific dosimetric

requirements of a particular usage, none of them fit in the criteria of a tissue

equivalent, reusable, easy to use, low cost and reliable dosimeter for routine

dosimetry of therapeutic ranges of doses.

For example Ionization chambers with all of their merits for dosimetry, have

drawbacks in terms of size, requirements of correction factors and stable

electrical. On the other hand semiconductor dosimeters such as diodes, and

MOSFETS, are non tissue equivalent and their performances are, affected by

temperature, dose rates and cumulative doses.

Among the tissue equivalent dosimeters, the diamond detectors are affected by dose

rates and the cumulative doses, Silver halide and radio chromic films on the other

hand are sensitive to dose rates and require energy corrections. Furthermore silver

halide films showed non linear behaviour that needs to be corrected.

Tissue equivalent synthetically prepared dosimeters such as, LiF:Mg,Ti,

LiF:Mg,Cu,P, Li 2 B 4 O 7 :Mn although are good for the therapeutic range of

dosimetry, but are expensive and require dedicated readout systems for their dose

data acquisitions. Other tissue equivalent dosimeters, such as chemical and plastic

dosimeters have shown high dose thresholds. The gel dosimeters, although

provides good results in the therapeutic range of doses, suffered with the corrections

required for its containers.

Recently introduced, tissue equivalent plastic dosimeter, PRASAGE (Zeff varies from

6.4 to 16.3 depending on the compositions), has established as a good dosimeter

qualifying for dosimetry of therapeutic range of doses.

Page 27: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

3

The methods of producing all these tissue equivalent dosimetrs are not simple and in

most of the cases, handling of toxic materials and steps of preparation has potential

hazards for health. The simple and non toxic preparation and the reusability of a

dosimeter results in its low cost.Except ionization chambers, doped materials, diodes

and MOSFETS, all the tissue equivalent materials are one time useable. Furthermore

their readout methods require either expensive dedicated equipments or

sophisticated diagnostic machines, which cannot be used on routinely in most

medical centres.

Hence, fabrication of a tissue equivalent radiation dosimeter, which could be used

on routine basis in radiation therapy centres with acceptable accuracy and reliability

in its dose measurements of therapeutic ranges, which is reusable, non toxic, using

easily available low cost ingredients , easy fabrication method and have ease in use,

is still a challenge for the researchers.

1.3 Purpose of the study It was planned to fabricate a tissue equivalent, nontoxic, easy to develop, low cost

and robust dosimeter, which should be capable of handling therapeutic dose ranges

with acceptable accuracy and confidence of reproducibility in its routine

use. The purpose of the study is to develop a tissue equivalent robust plastic

composite dosimeter suitable for validating dose ranges used in radiotherapy.

Its non-toxicity, ease in its, preparation, handling, simple readout method and its

reusability was of prime concern.Search for the most suitable low cost materials and

simple methods for the preparation and dose data acquisitions were also considered.

Page 28: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

4

1.4 Objectives of the study

The aim of the study is to fabricate a reusable tissue equivalent plastic composite

dosimeter with acceptable accuracies, that could serve the routine dosimetric

requirements of a radiation therapy centre. The simple preparation technique easy

read out methods were of prime concern. The cost of the dosimeter was also

considered. The sample was supposed to be of PMMA, uniformly embed sieved

grains of the selected radiation sensitive material. To achieve this goal, smaller goals

for this work have been identified.

First, this work seeked to identify the most suitable material to be used as the

ingredient in the plastic composite mix. This could be achieved through independent

evaluations of the respective materials for their responses to radiation in term of

changes in their optical properties.

Secondly, this work seeked to determine whether the sensitivity of any of these

materials can be improved by synthetic introduction of crystal defects. This can be

done by using high dose of particulate radiation such as neutron or electron.

Thirdly, methods of sample preparations were to be evaluated. The aim of this

evaluation is to determine the most suitable technique that can produce plastic

composite samples having favourable properties such as uniform physical

dimensions and uniform distribution of the embedding materials. Once the sample

preparation technique has been identified, the fourth minor objective is to actually

fabricate the samples for the rest of evaluation needed in this work.

Finally, these samples would be evaluated for their radiation dosimetric properties

using X-ray photon beam from a linear accelerator. They would be evaluated mainly

for radiation sensitivity, repeatability, fading, and dose linearity. Further dosimetric

studies, such as energy and dose rate dependency, would be studied if time allowed.

Page 29: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

5

1.5 Scope of the study

The selected radiation sensitive material grains should be embedded uniformly in the

body of the plastic as the optical response depends on it. Non-uniformities in either

the number of particles per unit mass of the plastic composite or the thickness of the

sample may cause error in observations.

Since the selected materials are supposedly light sensitive, any exposure to light even

for a small duration, from the time of irradiation until readouts may result in loss of

response signal. Performing the whole procedure in light tight condition helps in

getting the true dose versus response relationship.

The UV Spectrophotometer, which uses intensity based CCD detectors, is a less

sensitive instrument compared to fluorescence spectrophotometer, which uses PMT's

as detectors. Hence the radiation sensitivity of the final plastic composite dosimeter

may appear to be less with UV spectrophotometer than with fluorescence

spectrophotometer.

Page 30: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

6

CHAPTER 2

LITERATURE REVIEW

2.1 Radiation dose and dosimetry Ionizing radiations interacts with matter in different ways. The possibility of a

mode of interaction depends upon the type and energy of radiation and the physical

properties of the interacting matter. Ionizing radiations may lose its energy in

matter either by ionizing, exciting or by making the atoms or molecules

vibrate. Deposition of radiation energy within the interacting material is termed as

dose, whose unit is Gray (Gy), defined as joule of deposited energy per unit

mass(kg) (Khan, 2010).

Dosimetry is the measurement or estimation of dose, ideally at a defined point

within the interacting material, but it could also be performed for a plane of 2D

or for a volume of 3D. The properties of ionizing radiations, their interacting

mechanisms, and the relationship between doses versus different material's

responses, has been successfully exploited in health services, different industries,

scientific research and technologies. The use of ionizing radiation in plastic

industry, sterilizing of medical stuff and food items, quality assurance of different

goods such as metal sheets, environmental monitoring, archaeological and

geological dating, forensic applications, development of new materials for space

crafts and in medical sciences are examples of its usage.

Page 31: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

7

2.2 Medical dosimetry

The dose measurements in medical sciences was rapidly advancing and becoming

ever complexing. The needs of radiation dosimetry for, personal monitoring or

radiation protection, diagnostic radiology or for therapeutic radiology were different

leading to different methods, and related sophistications. Although the basic

concern of medical dosimetry is to estimate the biological damages resulting from

ionizing radiation exposures, the radiation therapy differs in that, the exposures were

pre planned.

The goal of radiation therapy is to deliver uniform and homogeneous prescribed

dose to treatment volume but at the same time, save the adjacent healthy tissues

from radiation exposure or keep the doses to un-avoidable healthy tissues, well

below their radiation tolerance (Khan, 2010).

2.3 Importance of dosimetry for radiation therapy

The challenges in radiation dosimetry is due to the variety in cancer tumours, types,

its anatomical site, its shape, and adjacent healthy tissues. The diagnostic images

used for marking treatment volumes may not be used for a particular radiation

treatment plan as the geometries of organs may change during the planned posture.

The cure or control of cancerous tumour, depends upon the coverage of the

treatment volume in the prescribed dose. Hence radiation treatments were needed

to be planed and must be validated prior to the actual treatments to optimize the

goals of radiation therapy.

Ideally, the 3D dose distribution of a treatment plan, within a tissue equivalent

phantom, similar to patients treated anatomy, is needed to be validated. New types

Page 32: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

8

of radiation therapy machines and techniques generally made the dose delivery faster

and its distribution over a treatment volume more precise, however due to the

sophistications involved, it is increasingly desirable that the actual dose delivered to

the patient be checked possibly for each treatment plan.

In more complicated treatment techniques such as stereotactic radio surgery,

conformal radiotherapy, Intensity modulated radiation therapy (IMRT),

Brachy therapy, in which are situations may arise where junctions between two fields

or regions of steep dose gradient exists, or the un-avoidable inclusion of some vital

organ or healthy tissues in the prescribed radiation fields. In such conditions 3D

dosimetry will be very useful, with spatially detailed point to point accurate dose

measurements, for the quality assurance of a treatment plan and hence to get the

ultimate benefits of the radiation therapy.

2.3.1 Treatment planning, treatment planning systems and dosimetry The treatment planning systems (TPS) used the dose distribution data in water

phantom. These doses were delivered with standard protocols and field sizes, used

in therapeutics. The TPS then interpolate or extrapolate the dose values for

different depths to calculate three dimensional doses distributions for a particular

treatment plan.

The beam data of every machine for a given field size is generally different,

due to different scatter factors, may be due to different machine design and

variations during production. Since TPS used interpolated or extrapolated

values of parameters, it might show different dose distributions, from the actual

treatment machines even for standard field sizes.

Page 33: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

9

Serious dosimetric errors such as under dosing, over dosing, missing the treatment

volume or inclusion of healthy tissues or vital organs, which were considered in the

treatment plan but may wrongly appeared in TPS. Such errors especially may

occurred at the junction of different density tissues or at the field junctions where

dose gradient appeared. Such errors ultimately cause unsuccessful treatment.

2.3.2 Requirement of validating treatment plan The treatment plans are made with the patients radiographic images X-rays, CT scan

and MRI. The requirement for validating a treatment plan is also to match the

patient's planed and the actual treatment setup. This matching of the patient's

planned treatment positioning with the actual treatment set-up is very important

as the treatment volumes are related to the surface marking of the beam entrance

points. This set up matching has to be done with a tissue equivalent phantoms

similar to the patient's treatment anatomy, whose 3D dose measurement will then

validate a treatment plan. Further, in the cases where the treatment volume or

the surface anatomy changed post irradiations, any modification in the plan needed

to be verified the same way. Due to current treatment complexity raising the risk for

target miss or unnecessary irradiation of healthy tissues, pre-treatment dosimetric

verification of each treatment plan is highly recommended.

Ideally, validations of the treatment plans are performed on real treatment machines

with real treatment parameters using tissue equivalent phantoms that faithfully

represents the anatomy of the patients treatment volume .The 3D dose distribution in

these phantoms should accurately indicates the actual dose distribution.

Page 34: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

10

2.4 Over view of the dosimeters Generally, dosimetry is performed using radiation sensitive volumes, whose

responses should be dose dependent and measurable using suitable read out system.

The response of any new dosimeter should be able to be confirmed and calibrated

with a standard dosimetry system, i.e. using ionization chambers and electrometers.

The performance of a dosimetry system is the net performances of its two

components, that is the radiation sensitive material and its response read out system.

2.4.1 Ionization chambers Among all types of dosimeters, ionization chambers are accepted to standard

dosimeters in clinical applications. Generally, ionization chambers have basic

construction of a cylindrical shell of air equivalent material that contain a suitable

gas of known volume and density. There is a fine electrode at the centre of this

cylindrical construction, maintained at high voltage, normally around

300-400 volts (Khan, 2010).

The thickness of the solid shell was made such that all the charges produced by the

ionization due to photons within a range of specified energies, reached the

electrodes and an electronic equilibrium establishes between the outgoing and

incoming electrons in measuring volume. The charge collected was proportional

to absorbed dose.Using relation between average energy required for producing

an ion pair in air as 33.85 eV / ion pair (Khan, 2010), amount of charge

collected can be converted to absorbed dose.

The doses of different energy photons could by measured by using a build up cap

whose thickness corresponds with the range of electronic equilibrium to the

Page 35: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

11

measuring volume of the chamber. Ionization chambers with build-up caps were

near to tissue equivalent and considered as standard point dose detectors .For dose

measurement temperature pressure and humidity corrections are required.

2.4.2 Film dosimetry Radiographic film was originally used for diagnostic radiography, later it found its

uses in personal monitoring. Now the use of radiographic films were also common

for portal imaging in therapeutic radiology.

2.4.2.1. Silver halide Radiographic films Radiographic films consists of a transparent thin base cellulose acetate or

polyester resin sandwiched between emulsion layers put together using thin layers

of adhesive. The emulsion is a mixture of gelatine and photosensitive grains of

silver halide micro crystals. The finished surfaces of film was protected with

layers known as super coating. The silver iodo bromide crystals embedded in

layers are sensitive to light and ionizing radiations. Unfortunately, the sensitivity

to ionizing radiations is generally much more lower to that of light.

The crystal defects enable silver ions Ag+ which moves freely with in the crystal.

Upon radiation exposure, bromide ion of the crystal absorbs light and gives off

electron, and left in gelatine as neutral bromide atom.

The electrons are then captured by the wandering silver ions converting them to

silver atoms. Depending on the amount of radiation received ,the silver atoms form

clumps together, forming latent image, i.e. a map of the radiation intensity.

Upon irradiation the exposed crystals were reduced to grains of metallic silver.

During the fixation of film processing, the unexposed grains are washed away,

Page 36: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

12

leaving behind a clear film. The silver grain within the film appeared black hence,

areas with higher radiation exposures have higher accumulation of silver grains,

leading to appearance.

Densitometers with suitable calibration are used to estimate the doses of the

different areas of exposed film, using the varying optical transparencies in a unit

known as optical density (Akselrod et al.) (Thomas III et al., 1984).

Although the radiographic film's response to radiation is nonlinear and depend on

radiation energy and processing parameters, their ease of use, inexpensiveness, high

spatial resolution and long period record keeping capability, keeps them in use for

general dsoimetric field checks of megavoltage therapy machines, both for photons

and electrons (Khan, 2010). With suitable choice of filters, radiographic films served

as personal monitoring device.

2.4.2.2. Radiochromic films

Radio chromic films are commonly used in therapeutic dosimetry. These films

are transparent and almost tissue equivalent. They contain grain less dye, which

upon irradiation gets polymerized and turn to blue colour. The change in optical

transparency was then determined with densitometer with proper calibration, the

results can be used to estimate absorbed dose. These films are self-developing,

thus they do not require chemical processing.

The response is dose rate independent. In standard to ambient conditions ,i.e. except

excessive humidity, they are comparatively less energy dependent than silver halide

films. Although radio chromic films are generally less sensitive than silver halide

films, however they provide higher resolution dose measurement, attributed due

to grain less feature of the dye, therefore they are found to be valuable in validating

Page 37: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

13

multi field radiotherapy plan with high dose gradient regions. Gafchromic HS (high

sensitivity), EBT (External beam therapy), films were found to be useful in the

dose ranges 0.1 Gy to 8 Gy and 0.5 Gy to 40 Gy and ranges respectively

(Jordan, 2006). These and many other radio chromic films with little differences

in, sensitivity, energy dependency, linearity and read out methods were

successfully used in therapeutic radiology.

2.4.3. Luminescence dosimetry

Luminescence is the process of emission of visible and near visible light from

certain materials, upon stimulation . A more detailed classification of luminescence

according to the with emission delay includes fluorescence and phosphorescence

with emission times ~ 10 - 8 s and >> 10-8 s respectively. It was found t hat some

solids, after irradiating with ionizing radiations upon heating or illuminating with

lights of suitable frequencies, emits light. In most cases, the light intensity was found

to be proportional to the absorbed doses.

This property of stimulated luminescence was exploited for dosimetry and named

according to the used mode of stimulation, as thermally simulated and optically

stimulated, luminescence dosimetry. The stimulated luminance used for dosimetry

was explained with the band theory of solids (Podgorsak, 2003).

2.4.3.1. Thermoluminescence dosimetry (TLD) In solids insulators, defects in the crystal lattice modifies the forbidden energy

bands between valence and conduction band by creating discontinuous energy levels

within the forbidden band, and create cites (L,T) where a hole or electron could be

tapped. Figure 2.1 gives a simplified model of these defects in term of the energy

Page 38: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

14

band. These defects may be intrinsic, extrinsic or result of some ionizing

radiation exposures, causing displacement of negative ion and hence generating a

vacancy for an electron. When such an insulator is irradiated with ionizing

radiations, the generated electrons and holes from the valence band may trapped

in these sites.

Figure 2.1. Model for thermo luminescence

If the natural de-trapping is negligible and the energy depth 'E' of the electron

trapping site below the conduction band is large, the life time of the trapped electrons

would be significantly large, establishing a meta stable state. When the irradiated

material is heated with constant rate of elevating temperature trapped electrons get

energized after receiving their energy from lattice vibrations leading to electron

ejection. The radioactive recombination of these liberated electrons with the

trapped holes, known as luminescence centres, gives out luminescence that

corresponds with some temperature is known as glow peaks. These glow peaks

could give information about the traps and the structural defects of the material

(Mandavia, 2011).

After heating the TLD to a specific high temperature, all the traps are emptied.

The TLD was then termed as zeroed as it was prepared for its re usage .

Thermo luminescence was originally used for measuring nuclear radiations by

Page 39: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

15

Daniel in early 1950's. Successful use of TLD for archaeological dating was done in

1960's by Aitken and Mejdahl and for geological dating in 1980's by Wintle

and Huntley, ( Botter-Jensin, 1997).

TLD's having favourable properties such as high sensitivity, reusability, reliability

and ease of use are now established as a popular tool for dosimetry. Nowadays

commercially TLD'S are available that varies in the physical forms(rods, tablets ,

ribbons, small size grain, powder), tissue equivalency, sensitivity, operational

conditions making them suitable for a wide range of dosimetric application . In

medical dosimetry, near tissue equivalent TLD's such as LiF: Mg, Ti, LiF:

Mg, Cu, P, Li2B4O7: Mn and non tissue equivalents such as CaSO4: Dy,

Al2O3: C, CaF2:Mn , as most sensitive dosimeters, are in common use

(Mandavia, 2011).

2.4.3.2. Optically stimulated dosimeters

The phenomenon of optically stimulated luminescence is the same as thermally

stimulated luminescence except that the mode of de-excitation which is optical

rather than thermal. This technique was originally used for dating by Huntley

in 1985, and further studied by, Hutt in 1988 , Aitken and Smith in 1988,

Spooner and Questiaux in 1990, Poolton and Bailif in 1989, Bùtter-Jensen in 1991.

Bùtter-Jensen and Duller in 1992, established it as a tool for dating, and

Godfrey-Smith and Haskell in 1993, Haskell in 1993, Bailiff in 1995, Bøtter- Jensen

in 1995 used for dosimetry for Hiroshima and Chernobyl nuclear events.

An insulator with defects in its lattice, either by some impurity or dislocation of

an ion, naturally present or synthetically produced, when irradiated with

ionizing radiation, generates electrons by leaving behind holes in the valence band.

Page 40: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

16

These electrons and holes may get trapped at the cites of defects, below the

conduction band and above the valence band respectively. Upon Illumination with

suitable frequency of light, luminescence may be observed. The emitted light may be

longer or of shorter wave length than stimulation wave length. The emitting light is

a characteristic of the material but the intensity was found to be proportional to the

absorbed dose. Different modes of optical excitations used were namely,

Continuous wave optically stimulated dosimetry (CW-OSL), Linear modulated

(LM-OSL) and Pulse OSL (P-OSL).

2.4.3.2.a. CW- OSL

In this mode of stimulation, samples were stimulated with constant intensity light

source which may be either laser or a broad band source. The luminescence was

measured simultaneously with the stimulating illumination. The luminescence

wavelength are distinguished from excitation wave lengths, using mono chromoters.

The decay curve of luminescence may or may not be exponential, revealing

different de trapping modes of trapped electrons and information about the

lattice defects. The total luminescence i.e. the total area under luminescence

curve, was proportional to the absorbed dose (McKeever, 2001).

2.4.3.2.b. LM-OSL

In this technique, instead of constant intensity of excitation wave lengths,

continuously increased intensity was used. The continuous increase in the intensity

of light, emptied trapped electrons, at their different intensity values and hence can

easily distinguished. This intensity with verses emptied electrons gives information

about the depth of the traps. The luminescence peaks at different points in time

Page 41: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

17

of a linearly increased intensity ramp, gives information about the contribution

of different electron traps of different photo ionization cross section ,to the total

luminescence (McKeever, 2001).

2.4.3.2.c. Pulsed OSL In Pulsed OSL, the excitation wavelengths are used to illuminate the samples for

such short intervals of time that suits the life time of the luminescence

(Bøtter-Jensen, 1997).

Since luminescence does not depleted much in a short pulse, multiple observations

could be made for single dose (McKeever, 2001). Further, since measurement

was done after illumination, the stimulation and emitted wavelengths are easily

distinguishable without any aid of filters, which simplifies the method and

increased the measuring efficiency of the luminescence.

2.4.3.3. OSD VERSUS TLD The general features of the two modes of luminescence dosimeters i.e, TLD and

OSLD can be summarized as follows :-

1. Thermo luminescence dosimetry, as a single time dose measuring technique, with

comparatively complex method and thermal quenching, is a less sensitive

technique than non-destructive optically stimulated luminescence dosimetry, which

offers multiple observations for single dose .

2. Thermoluminescence output is a whole sample's response, where as in OSL, with

thin laser beams as optical exciting tool, even a grain could be selected from the

whole sample to get an optical response. This ease of site selection from a sample

for dose estimation could be exploited in re assessment of absorbed or in dose

mapping .

Page 42: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

18

3. It was observed that TL observations could be taken after taking OSL

observations from the same sample and that in many cases separate samples with

same dose gave same response, either use TL or OSL. This suggests that the

electron traps for TL and OSL may not be the same.

2.4.4. Semi conductor detectors

Semi-conductors have offered a good choice for many dosmetric applications.

Commonly used semi-conductor radiation detectors includes, PN diodes, MOSFET

(Metal-Oxide Semiconductor field effect Transistor), and Diamond detector

(Rosenfeld, 2011).

2.4.4.a. Diodes

Semiconductor diodes has been used as radiation detectors since last 30 years.

Due to their merits, of being very small sized and hence offering high resolution,

and ease of use for real time dosimetry, they are popular dosimeters for quality

assurance in radiation therapy (Podgorsak, 2003).

2.4.4.b. PN diodes

Silicon diodes are commercially available as n-type and p-type. They are fabricated

by counter doping the surfaces to produce opposite type material. The pn junction in

semi-conductor diodes, provides an in build, very high electric field (103 V/m)

across the junction.

When diode was exposed to ionizing radiations, electrons and holes pairs were

generated in the diode, which then diffused through the pn junction and collected by

an electrometer. This induced current through the diode was known as radiation

Page 43: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

19

current and was proportional to dose rate. The total current was proportional to

absorbed dose rate in silicon. The silicon diode and MOSFET have near tissue

equivalency in MV energy range as silicon to water mass stopping collision power

ratio is almost independent of energy in MV range for photons and electrons

conventionally used in radiotherapy. There small sensitive volume and low

ionization energy compare to that of air made them 18000 times more sensitive

than same volume ionization chamber (Rosenfeld, 2011).The Diodes are commonly

used in vivo dosimetry as relative dosimeter, for example in bladder and rectum

dose measurements, entrance and exit dose. It is used with beam entrance dose

measurements routine checks of treatment parameters and setup errors such as

correct technique (SSD OR SAD), use of correct radiation and applicator, dose

rate change are able to be performed (Yorke et al., 2005).

As a relative dosimeter, they have proven to be suitable for dose measurements in

phantoms , especially for small field, high dose gradient regions as in radio surgery,

for depth dose measurements and for beam profiles of photons and electrons, in

commercially available 2D and 3D systems serves as quality assurance tools for

IMRT treatment plans. In spite of its merits, commercially available diodes could

not be used for surface dosimetry, because of its packing which behaves as build up

region. Diodes responses are sensitive to dose rate, temperature, accumulated doses

and the material used. Diodes suffers radiation degradation over time, the degree of

which depend on type of radiation and dose. Hence the use of diodes as

radiation detectors needs calibration against many factors. The response of

diodes were found directional dependent and also affected by spectral

composition of beam (Rosenfeld, 2011).

Page 44: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

20

2.4.4.c. MOSFET

MOSFET Metal-Oxide Semiconductor Field Effect Transistor, due to its very small

size 1μm, and sensitivity to full range of photons and electrons, has been

successfully used for measuring relative doses in brachy therapy, radio surgery,

surface dose measurements diagnostic radiology.

These short life detectors have the same drawbacks as the diodes detectors have

Their responses were dependent on temperature, dose rate, direction of detection,

radiation quality, absorbed doses, and bias voltage during irradiation

(Podgorsak, 2003).

2.4.4.d. Diamond detector

Diamonds, either natural or artificial, when exposed to ionizing radiations,

shows change in its resistance. The detector with a suitable biased voltage

100 V and charge integrating amplifier, when exposed to ionizing radiations

gives signal proportional to the charge induced electron hole pair, which in turn

proportional to the absorbed dose.

Diamonds are low Z material and near to tissue equivalent therefore almost no

energy correction is required. These detectors, have small effective volumes order

of few mm3,very small directional dependency, negligible ; dose rate and

temperature 0.1 % per oC dependence and high radiation damage resistance. Prior

irradiation of the detectors were required for stable responses for each usage,

(Podgorsak, 2003).These detectors are specially suitable for small field high dose

gradient regions such as in radio surgery (SRS), routine treatment set up

checks, and relative dosimetry, (Rosenfeld, 2011).

Page 45: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

21

2.4.5. Chemical dosimetry

Chemical dosimeters are the oldest of all kinds of dosimeters. Originally

the chemical dosimeters were in liquid form but now later it was modified as a

gel dosimeters

2.4.5.a. Frick dosimeter

Fricke & Morse 1927 developed a ferrous sulphate solution as radiation

dosimeter with the irreversible oxidation of ferrous ion to ferric ion by ionizing

radiations (Nemtanu et al., 2008). The absorbed doses were measured by the Fe+3 ion

concentration which changes the absorption of UV radiation of wave lengths

224 nm and 303 nm with spectrophotometer. The absorbance of light at those

particular wavelengths and is directly proportional to the absorbed dose. It was used

for the determination of the doses between 40 Gy to 400 Gy (Soares et al., 1987).

The dosimeter was dose rate radiation quality ( photon and electron) independent

and relatively temperature 0.12% per 0C. It was also found to be very precise

0.15%. However it is easily affected by contaminations and does not show very

good stability over the periods of few years.

Being a contamination sensitive solution dosimeter it is difficult to use it on a

routine bases. Furthermore the Fricke dosimeter could not be used for

measurement requiring high spatial resolution. Since it is an irreversible chemical

process Fricke dosimeter could not be recycled. Other chemical dosimeters

includes ferrous–cupric dosimeter measurable range of doses 5×102 to 104 Gy,

with accuracy of ±02% and found to be LET dependent, Ceric sulphate

dosimeter dose range 102 to 2×105 with accuracy ±02% and the ethanol – chloro

benzene dosimeter dose range 4×102 Gy to 5 ×102 6 × 104 Gy, accuracy ±03%.

Page 46: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

22

2.4.6. Gel dosimetry Gel dosimetry is relatively a new technique. Many different kinds of Gels are used

for the purpose . Most of them are modified formulas of the few basic gels originally

introduced. They offer 3D dosimetry with good accuracy.

2.4.6.a. Fricke gel dosimetry

Fricke chemical dosimeters Ferrous sulphate solution could be considered as the

first dosimeter with potential to be extended as 3D dosimeters .Gore in1984, have

showed that Fricke dosimeters could be probed with MRI to have 3D dose

distribution , (Baldock et al., 2010).

Gels such as gelatine, Agarose were used with infused ferrous sulphate ions to have

some stable 3D dose distribution, the concentration of radiation induced Fe+3 ions,

which is proportional to the absorbed doses, changes the colour of gel. The Fe+3

concentration remained fix in gelatine matrix for some time and could be probed

with suitable technique such as MRI, OCT, Ultra sound. (Baldock et al., 2010)

Fricke gels are water equivalent, hence considered as tissue equivalent, for a large

range of therapeutic radiation (Schreiner, 2004a). It is easy to prepare, inexpensive,

non-toxic since it is prepared in liquid form the dosimeter can have complex

phantom shapes and shows very good reproducibility.

However the Fricke gels have high threshold such as 50 Gy to 75 Gy making

it not suitable for low or intermediate radiation doses applications as encountered

in hospitals. Furthermore, the ferrous ions tends to diffuse in the gel after irradiation,

losing the spatial distribution information within a few hours. Attempts to solve the

problem of ferrous ion diffusion were made through the use of different gels such

as ge la t ine , agarose , sephadex and po lyvinyl a lcohol PVA.

Page 47: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

23

Chelating agent agents such as Xylenol Orange XO were tried, which made a small

improvement in the stability ( Baldock el at 2010) but reduces the sensitivity of the

dosimeter (Schreiner, 2004b).When optical CT (OCT) was used noticed for that

readout the dose sensitivity depends on the wave lengths chosen. Optical

CT for dose readout of Fricke-Agarose –Xylenol Orange gel uses a fast

CCD camera to acquire the image. (Luciani et al., 2006).

The results of OCT readout were found to have good linear relationship in the

dose range up to 10 Gy and with spatial resolution approximately 0.5 mm

(Viti et al., 2006). It was observed that the use of MRI for dose data acquisition

requires relatively longer time than OCT readout , hence the optical technique

was better in limiting the effect of diffusion the 3D gel dosimeter.

2.4.6.b. Polymer gel dosimetry

Historically, the polymer gel dosimetry was successfully introduced by Hoecker and

Watkins in 1958. Boni in 1961, used poly acrylamide as a gamma dosimeter.

(Baldock et al., 2010).

The polymer gel dosimetry was based on radiation induced polymerization in

solutions of monomers and polymers (McJury et al., 2000). The polymerization was

proportional to the absorbed dose. The extent of polymerization thus changes the

density of the set materials. This change in density is proportional to the absorbed

dose. The change in density detectable upon viewing with X-ray CT scanners.

The colour or opacity of irradiated gels also changes and hence could be detected

and measured with optical devices such as optical CT OCT. Additionally MRI

was also used as a readout method. Maryanski in 1996 introduced polymer gel

dosimetry which was based on the polymerization of acrylamide AAm and

Page 48: By IQBAL TARIQmother Noor Jehan Begum (late) and father Muhammad Ashabuddin (late) (may Allah rest their souls in peace), deep love and their encouragements to achieve my goals and

24

Bis monomers infused in an aqueous agarose matrix.

He gave it the name BANANA, (Baldock et al., 2010). The dosimeter was found to

be a tissue equivalent and its performance evaluation a 3D dosimeter was done

using stereotactic radio surgery and HDR brachy therapy plans. MRI scans were

used for 3D dose data acquisition and was found in good agreement with the

planned doses (Maryanski et al., 1996). With small change in the formulation of

BANANA, Maryanski in 1994 introduced BANG. This name later changes to PAG

to distinguish it with from the in house manufactured gels, (Baldock et al., 2010).

BangTm3 and PAG gels were investigated for use in 3D dosimetry with XCT

(Oldham et al., 2001) OCT and MRI (Audet et al., 2002), and was found good

for the purpose. Although the polymer gels were tissue equivalent and do not have

the problem of ion diffusion as in Fricke gels, it is very sensitive to atmospheric

oxygen which inhibited the polymerization process. For this reason the polymer

gels has to be prepared in oxygen free environment. This is found to be an in-

convenient procedure for daily use.

Fong in 2001, introduced a normoxic gel ‘MAGIC’ in which atmospheric oxygen

was bounded in metallo organic compound matrix without the gel for its

preparation (Baldock et al., 2010). Thus gel does not require for an Oxygen free

environment.

Different gels with little changes in the formulation were investigated with different

techniques. Mather in 2002 make use of Ultrasound for acquiring 3D dose

distribution data, Rintoulet in 2003 showed how to use Raman spectroscopic

technique to probe depth dose distribution in electron beams for PEG

(Baldock et al., 2010). Bheag gel (Rabaeh et al., 2008) uses MRI to evaluate its

dosimetric performance, Anthromorphic Barex (Duthoy et al., 2004) was