+ All Categories
Home > Documents > · by the ARTISTE firmware – if more dose must be applied, the gantry rotation is slowed down or...

· by the ARTISTE firmware – if more dose must be applied, the gantry rotation is slowed down or...

Date post: 21-May-2020
Category:
Upload: others
View: 3 times
Download: 0 times
Share this document with a friend
8
www.siemens.com/healthcare Global Siemens Headquarters Siemens AG Wittelsbacherplatz 2 80333 Muenchen Germany Legal Manufacturer Siemens AG Wittelsbacherplatz 2 DE-80333 Muenchen Germany Global Siemens Healthcare Headquarters Siemens AG Healthcare Sector Henkestrasse 127 91052 Erlangen Germany Phone: +49 9131 84-0 www.siemens.com/healthcare Global Business Unit Siemens AG Medical Solutions Computed Tomography & Radiation Oncology Siemensstr. 1 DE-91301 Forchheim Germany Phone: +49 9191 18 0 Fax: +49 9191 18 9998 In the event that upgrades require FDA clearance, Siemens cannot predict whether or when the FDA will issue its clearance. Therefore, if regulatory clearance is obtained and is applicable to this package, it will be made available according to the terms of this offer. On account of certain regional limitations of sales rights and service availability, we cannot guarantee that all products included in this brochure are available through the Siemens sales organization worldwide. Availability and packaging may vary by country and are subject to change without prior notice. Some/All of the features and products described herein may not be available in the United States. The information in this document contains general technical descriptions of specifications and options as well as standard and optional features which do not always have to be present in individual cases. Siemens reserves the right to modify the design, packaging, specifications, and options described herein without prior notice. Please contact your local Siemens sales representative for the most current information. Note: Any technical data contained in this document may vary within defined tolerances. Original images always lose a certain amount of detail when reproduced. Please find fitting accessories: www.siemens.com/medical-accessories Order No. A91RO-01000-08C1-7600 | Printed in Germany | CC RO 1386 05130.5 | © 05.2013, Siemens AG
Transcript

www.siemens.com/healthcare

Global Siemens HeadquartersSiemens AGWittelsbacherplatz 280333 MuenchenGermany

Legal ManufacturerSiemens AGWittelsbacherplatz 2DE-80333 MuenchenGermany

Global Siemens Healthcare Headquarters Siemens AGHealthcare SectorHenkestrasse 12791052 ErlangenGermanyPhone: +49 9131 84-0www.siemens.com/healthcare

Global Business UnitSiemens AGMedical SolutionsComputed Tomography & Radiation OncologySiemensstr. 1DE-91301 ForchheimGermanyPhone: +49 9191 18 0Fax: +49 9191 18 9998

In the event that upgrades require FDA clearance, Siemens cannot predict whether or when the FDA will issue its clearance. Therefore, if regulatory clearance is obtained and is applicable to this package, it will be made available according to the terms of this offer.

On account of certain regional limitations of sales rights and service availability, we cannot guarantee that all products included in this brochure are available through the Siemens sales organization worldwide. Availability and packaging may vary by country and are subject to change without prior notice. Some/All of the features and products described herein may not be available in the United States.

The information in this document contains general technical descriptions of specifications and options as well as standard and optional features which do not always have to be present in individual cases.

Siemens reserves the right to modify the design, packaging, specifications, and options described herein without prior notice. Please contact your local Siemens sales representative for the most current information.

Note: Any technical data contained in this document may vary within defined tolerances. Original images always lose a certain amount of detail when reproduced.

Please find fitting accessories: www.siemens.com/medical-accessories

Order No. A91RO-01000-08C1-7600 | Printed in Germany | CC RO 1386 05130.5 | © 05.2013, Siemens AG

White Paper

Answers for life.

Modulated arc (mARC): principle and clinical implementationY. Dzierma, F. Nuesken, N. Licht, S. Kremp, Ch. Ruebe

Modulated arc (mARC): principle and clinical implementationY. Dzierma, F. Nuesken, N. Licht, S. Kremp, Ch. Ruebe Department of Radiation Oncology, Saarland University Medical Centre, Homburg/Saar, Germany

IntroductionThe modulated arc (mARC) technique (Salter et al., 2011; Kainz et al., 2011) has recently become available for Siemens ARTISTE linear accelerators as an analogue to RapidArc and volume modulated arc therapy (VMAT) available for Varian and Elekta machines (Otto, 2008; for a recent review see Teoh et al., 2011). All techniques offer highly conformal treatment, since inversion is per-formed similarly to intensity-modulated radiotherapy (IMRT) for a large number of beam directions, which may create a complete or partial gantry rotation; a notable decrease in treatment time is accomplished by the continuous gantry rotation and multi-leaf collimator (MLC) movement. How-ever, mARC differs from VMAT in underlying philosophy and practical implementation. The purpose of this work is to give a short overview on the characteristics and tech-nical principles of mARC treatment (compare Salter et al., 2011 and Kainz et al., 2011 for more detail) and the prac-tical application in the clinic. In addition to providing the physics background of planning and dosimetric verification, treatment planning and patient treatment for a number of clinical cases is presented. The work was carried out on the Siemens ARTISTE at the Institute of Radiation Oncology of the Saarland University Medical Centre in Homburg/Saar (Germany). This acceler-ator is equipped with 160 multi-leaf collimator (MLC) and 6 MV flat and 7 MV flattening-filter-free (FFF) beam modalities with dose rates between 50 and 300 MU/min in flat and 500 to 2000 MU/min in FFF mode Treatment planning was performed in the Prowess Panther TPS.

Background – concept of mARC treatmentIn conventional arc treatment as implemented by Varian and Elekta, continuous mode delivery is applied, meaning that dose is delivered continually while the gantry and MLC move. In contrast to this, the mARC beam is only switched on while the MLC remains static, although the gantry moves along an arclet of a few degrees. This is why this mode of delivery has been called “burst mode” (Salter et al., 2011). In fact, this way of delivery approxi-mates more closely the dose calculation as it is generally performed in treatment planning systems, since the TPS usually calculates the dose based on static fields at the control point angles. In contrast to “real” IMRT, where the dose would be delivered statically at these control points, the mARC beam is switched on in a small arclet around each control point (usually 2°-5°), which makes delivery faster since the gantry can continually rotate. The maxi-mum width of the arclet is defined by the user; if the dose can be delivered in a shorter time than needed to rotate through the arclet, the angle will be automatically decreased by the ARTISTE firmware – if more dose must be applied, the gantry rotation is slowed down or stopped so that the maximum arclet angle is never exceeded. In this way, the gantry speed is automatically modulated so that the arclet angle is minimised. At the same time, the firmware linearly reduces the dose rate for small-MU segments to ensure accurate linearity. It is possible to include “hybrid fields”, i.e. fields where more than one segment is irradiated from the same gantry direction; for these fields the gantry will be stopped until

Modulated arc (mARC)

2

Figure 1: Schematic illustration of mARC delivery (from Siemens mARC Planning Guide). Between two control points, radiation is off and the MLC leaves more into the required aperture shape for the upcoming optimisation point. In the vicinity of each optimisation point, over a small arclet (maximum gantry angle range restricted by the user and symmetrically placed around the optimisation point), the MLC remains fixed and radiation is on. Not shown here, hybrid fields can be included to irradiate several segments of different MLC configuration from the same gantry angle, equivalently to normal step-and-shoot beams.

Modulated arc (mARC)

the desired number of segments has been irradiated. Between one arclet and the next, the MLC is moved to the new configuration without dose delivery. This offers a free choice of MLC configuration for the treatment plan, since intermediate configurations are not irradiated, and can therefore allow for better plan optimisation than continuous-mode delivery.

Physics concepts Machine data modelling and commissioning is performed in exactly the same way as usual; no further requirements for mARC facilities exist. At our institution, the mARC upgrade was carried out at an ARTISTE linac equipped with a flat 6 MV and flattening-filter-free 7 MV beam line (Multiple-X); both energies were modelled and com-missioned in Prowess. The dosimetric properties of the linac can be taken from Dzierma et al. (2012).

a. Treatment planning

Treatment planning in the Prowess Panther TPS is per- formed similarly to IMRT treatment planning. For com-paratively simple target volumes (e.g., brain metastases, prostate), a single arc with optimisation points spaced every 10° will give good quality plans and result in short treatment times. After the definition of the arc isocenter, collimator angle and control points, the optimisation is carried out in the same way as in IMRT planning.

It has generally been observed in arc treatment (also RapidArc and VMAT) that for complicated target volumes, one arc may not be sufficient to create an optimal dose distribution. In these cases where a single arc does not yield good quality plans, a second arclet with a different collimator orientation (and reverse gantry rotation direction) can be included; or more complicated combinations of

ArcletDose

Delivery

LeafShape

Change

Optimization Point

Arclet

Gantry Movement

CP0

CP1

CP2

CP3

CP4 CP5 CP6CP7

CP8

CP9

Beam On

Beam Off

Gantry Speed

a1

a2

a3

a4

a5

3

a number of coplanar or non-coplanar arcs can be chosen (Guckenberger et al., 2009; Vanetti et al., 2009). Alter-natively, and often preferable from the point of view of both simplicity and treatment time, a number of gantry directions can be designated as hybrid fields, with a chosen number of segments (Chan et al., 2011).

b. Dosimetric verification

While treatment planning is fairly straightforward and does not prove any more complicated than for static IMRT, dosimetric verification must be more thoroughly checked before patient treatment. In this context, we shall con-sider two different issues: the accuracy of the dosimetric delivery by mARC as performed by the ARTISTE firmware from a purely technical view-point, and the dosimetric verification of patient treatment plans from the TPS soft-ware.

From the technical side, accuracy of mARC is evaluated in comparison with step-and-shoot IMRT plans. For this reason, we programmed a conversion program from IMRT plans exported in RTP file format into mARC plans (Dzierma et al., submitted to Medical Physics). Both the original step-and-shoot and the converted mARC plans were then irradiated and compared; irradiation times were reduced by about half for the mARC converted plans. This dosimetric test is independent of the quality of the TPS calculation and commissioning and only detects any deviation in dose that arise from the mARC treatment controlled by the ARTISTE firmware. Apart from the

“smearing” of the dose delivery across the arclet angle (set to 4°), no difference should exist between both plans if the technical delivery is accurate. It is recommended to limit arclet widths to 5° maximum to keep dose-smearing effects small. Absolute dosimetric verification of both the

Modulated arc (mARC)

Figure 2: Example of a prostate shrinking field treatment plan (screenshot from Prowess) calculated as a single arc with 10° equiangular-spaced control points. Left: trans verse view, upper right: sagittal view, lower right: beams-eye view. The white line denotes the 95% isodose.

Figure 3: (a) Gamma distribution of GafChromic film measurement of the converted mARC plan, compared with the original step-and-shoot plan. Pass criteria are 3% deviation in local dose and 3 mm distance to agree ment, which are satisfied by over 95% of measurement points. (Figure from Dzierma et al., under review)

4

step-and-shoot and mARC plans by point dose measure-ments with a PTW semiflex ionization chamber (type 31010) showed deviations below 3% local dose, which is within the specifications allowed at our institution for IMRT verification. The absolute dosimetric verification of mARC plans deviated from step-and-shoot plans by no more than 1%.

Planar dose distributions of both plan varieties (step-and-shoot vs. mARC) were measured in an acrylic phantom using GafChromic film. The agreement between both treatment modalities is excellent, with over 95% of points within 3% deviation in dose and 3 mm distance to agree-ment; an example is shown in Figures 3-4 for a hybrid prostate PTV plan.

In principle, validation of the treatment planning system need not explicitly be performed from the technical per-

spective, since dosimetric verification for IMRT plans was already checked in the commissioning phase (Dzierma et al., 2012), and since the deviation of the mARC plans from the step-and-shoot plans is negligible. However, we perform these measurements for complete ness using the PTW Octavius phantom with 729 2D-Array. Furthermore, for original mARC plans (not created by conversion from step-and-shoot plans), dosimetric verifi cation before treat-ment should be carried out for each patient, just as it is done for IMRT treatments. For all plans tested at our insti-tution, over 80% of the points passed the criteria of 3% local dose deviation and 3 mm to agree ment (example shown in Figure 5), which provided an independent verifi-cation of the dose distribution of the mARC plan. Deviations mainly arose in the low-dose area (below 20% maximum dose), while the agreement within the high-dose region is much better.

Modulated arc (mARC)

Figure 4: Example left-right profile (left) and target-gantry profile (right) measured in both films (orange line: original step-and-shoot plan, blue line: mARC converted plan). (Figure from Dzierma et al., under review)

Figure 5: Gamma distribution of Octavius 729 2D-Array measurement of a prostate mARC plan, compared with the dose distribution exported from the Prowess TPS. Only one coronal and one transverse slice are shown. Pass criteria are 5% deviation in local dose and 3 mm distance to agreement, which are satis-fied by over 80% of measurement points in all slices (excluding doses below 10% of the maximum dose).

5

Modulated arc (mARC)

Clinical casesWithin the nine months experience at our clinic, most cases treated have been prostate patients. Both PTV and shrinking field can be planned with a single arc and optimisation points spaced every 10° (see Figure 6 for an example plan treated with 6 MV). For brain tumours, a single arc stereotactic treatment can equally well be planned, as shown by the example plan (Figure 7) for an acoustic neuroma patient treated with FFF 7 MV. Complicated target volumes, e.g., head-and-neck cases with integrated boost, require more planning effort. In these cases, a second arc or the inclusion of a number of hybrid fields may be required to achieve a clinically acceptable dose distribution.Clinical application of the mARC has hitherto been straightforward, with no problems of treatment delivery. A large advantage of the mARC as compared with VMAT

and RapidArc is that the treatment can be easily inter-rupted and resumed, with a known delivered dose at each instant – a case we have admittedly not experienced yet since not interruption occurred and all treatments were delivered as planned. Treatment times were measured for all patients and range between 2 and 5 minutes for a single arc, depending on the use of flat vs. FFF beams and whether or not hybrid fields are included.

SummaryThe mARC presents an alternative to VMAT and RapidArc treatments. Due to the burst mode dose delivery, which interrupts the beam for MLC movement, the delivery time is still slightly longer than for VMAT and RapidArc; at the same time, this technique provides a further degree of freedom for the MLC configuration. The burst delivery technique is also of advantage in case of treatment

Figure 6: Example prostate shrinking field plan, treated with a single arc of energy 6 MV (flat). Optimisation points were spaced every 10°, with no hybrid fields.

6

Modulated arc (mARC)

interruption, as the delivered dose is precisely known at each instant. As long as the arclet angles do not exceed 5°, the delivered dose is calculated very accurately, with good agreement between step-and-shoot and mARC delivery. Verification of the measured mARC plan (Octavius with 2D-Array) against the TPS calculation is straight-forward, and of comparable quality as IMRT verification. From the technical point of view, perfor mance is excellent, with no malfunctions or irradiation problems so far.

The mARC offers the possibility to save significant amounts of time, with single-arc treatments of only a few minutes achieving comparable dose distribution to IMRT plans taking up to twice as long. We are currently imple menting mARC treatment into clinical routine step-by-step for different tumour locations, starting with prostate patients.

References

O.S.H. Chan, M.C.H. Lee, A.W.M. Hung, A.T.Y. Chang, R.M.W. Yeung, A.W.M. Lee, “The superiority of hybrid-volumetric arc therapy (VMAT) technique over double arcs VMAT and 3D-conformal technique in the treatment of locally advanced non-small cell lung cancer – A planning study”, Radiother. Oncol. 101, 298-302 (2011).

Y. Dzierma, N. Licht, F. Nuesken, C. Ruebe, “Beam properties and stability of a flattening-filter-free 7 MV beam – An overview”, Med. Phys. 39(5), 2595-2602 (2012).

Y. Dzierma, N. Licht, F. Nuesken, Ch. Ruebe, “A novel implementation of mARC planning with automated gantry angle optimization for hybrid fields”, Rad. Onc., under review

M. Guckenberger, A. Richter, T. Krieger, J. Wilbert, K. Baier, M. Flentje, “Is a single arc sufficient in volumetric- modulated arc therapy (VMAT) for complex-shaped target volumes?”, Radiother. Oncol. 93(2), 259-65 (2009).

K. Kainz, G.-P. Chen, Y.-W. Chang, D. Prah, X.S. Qi, H.P. Shukla, J. Stahl, X.A. Li, “A planning and delivery study of a rotational IMRT technique with burst delivery”, Med. Phys. 38(9), 5104-5118 (2011).

K. Otto, “Volumetric modulated arc therapy: IMRT in a single gantry arc”, Med. Phys. 35(1), 310-317 (2008).

B.J. Salter, V. Sarkar, B. Wang, H. Shukla, M. Szegedi, P. Rassiah-Szegedi, “Rotational IMRT delivery using a digital linear accelerator in a very high dose rate ’burst mode’”, Phys. Med. Biol. 56, 1931-1946 (2011).

M. Teoh, C.H. Clark, K. Wood, S. Whitaker, A. Nisbet, “Volumetric modulated arc therapy: a review of current literature and clinical use in practice”, BJR 84, 967-996 (2011).

E. Vanetti, A. Clivio, G. Nicolini, A. Fogliata, S. Ghosh-Laskar, J.P. Agarwal, R.R. Upreti, A. Budrukkar, V. Murthy, D.D. Deshpande, S.K. Shrivastava, K.A. Dinshaw, L. Cozzi,

“Volumetric modulated arc therapy for carcinomas of the oro-pharynx, hypo-pharynx and larynx: A treatment planning comparison with fixed field IMRT”, Radiother. Oncol. 92, 111-117 (2009).

Figure 7: Example plan of an acoustic neuroma patient, treated with a single arc of energy 7 MV (FFF). Optimisation points were spaced every 10°, with no hybrid fields.