24
21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220 Manufacturer: EU Representative: Byron Medical, Inc. Mentor Medical Systems, BV 602 West Rillito Street Zernikedreef 2 Tucson, Arizona 85705 2333 CL 800-777-3434 The Netherlands (31) 715 249 600
Transcript
Page 1: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 Rev D

Byron Medical’s

PSI-TEC III Aspiration Platform (with White Cap)

Instruction Manual

PT-PLT-III-110

PT-PLT-III-220

Manufacturer: EU Representative: Byron Medical, Inc. Mentor Medical Systems, BV 602 West Rillito Street Zernikedreef 2

Tucson, Arizona 85705 2333 CL 800-777-3434 The Netherlands

(31) 715 249 600

Page 2: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

Table of

Contents

I. Introduction Symbol Definitions 1

Use of this Documentation 2

Warnings and Cautions 2

Operational Safety 2

Modifications 2

Damage or Loss in Shipment 2

Incorrect Items Shipped by Byron Medical 3

Policy on Returned Goods 3

Repair Program 4

Warranty Policy 4

II. Operation General Description 5

Operative Controls and Components 5

System Set-up 5 Canisters Set-Up Illustration (See Page 7)

Operating Instructions 6

III. Maintenance Cleaning Instructions 8

IV. Troubleshooting Problems, Causes, and Solutions 9

Appendices A. PSI-TEC III Aspiration Platform Specifications (115V/230V) 10 B. PSI-TEC III Replacement Parts List (115V/230V) 11 C. PSI-TEC III Hand Rail Parts List 19 D. PSI-TEC III Aspiration Platform Schematics (115V/230V) 20 2004 by Byron Medical Byron Medical reserves all rights to this document. No part of this manual may be reproduced, stored, or transmitted in any form or by any means (electronic, mechanical, photocopying, recording, etc.) without prior written permission of Byron Medical. Requests for permission to reproduce or distribute this document should be addressed to:

Byron Medical, Inc., 602 West Rillito Street, Tucson, Arizona 85705

Page 3: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 1 of 21 Rev D

Symbol Definitions Compressed Air Vacuum Air Out Air Return Type “B” Equipment Danger: DO NOT use in the presence of flammable anesthetics Attention, Consult users Manual Risk of Electric Shock. DO NOT remove equipment cover Footswitch Fuse and Fuse Rating

AP

Page 4: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 2 of 21 Rev D

I. Introduction

Use of This Documentation This manual provides instruction for operation, maintenance, and troubleshooting procedures. Users should be thoroughly trained in using this product and applicable medical procedures. Instruction manuals should be made available to the user(s) during the procedure. Follow all instructions contained in this manual pertaining to the devices used in the procedure, with particular attention given to the WARNINGS and CAUTIONS.

Warnings and Cautions WARNINGS and CAUTIONS statements are placed throughout the text referring to procedures, practices, or conditions considered essential to the safety of personnel and patients and/or equipment and property. All WARNINGS and CAUTIONS should be reviewed prior to the use of this equipment. If you have any questions, please contact Byron Medical, (520) 573-0857 or (800)-777-3434.

Operational Safety A thorough review of this entire instruction manual is essential prior to using this product. Byron Medical has taken great care to ensure the safety of the patient and staff. However, features may not be readily discernible without reviewing this documentation. Use of this equipment should therefore not be under taken until the user(s) is fully familiarized with the instructions for assembly and operation. If you have any questions, contact Byron Medical, (520) 573-0857, (800) 777-3434, or FAX (520) 746-1757.

Modifications Modifications of any kind NOT recommended and will void all warranties.

Damage or Loss in Shipment Thoroughly inspect shipment immediately upon arrival. If goods are received short or in damaged condition, it is important that you notify the transportation company and insist on a notation of the loss or damage on the freight bill. Otherwise, it may be difficult to make a claim against the Transportation Company. If concealed loss or damage is discovered, retain all packaging materials, notify the transportation company immediately, and request an inspection. The agent will make an inspection and grant a concealed damage notation. A concealed damage report must be made within seven (7) days of shipment delivery. After seven (7) days, the transportation company reserves the right to refuse any claim for loss or damage.

Page 5: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 3 of 21 Rev D

Incorrect Items Shipped by Byron Medical Please check your shipment immediately for any shortage or incorrect items. If any discrepancies exist, notify Byron Medical, (520) 573-0857 or 800-777-3434, at once. Your prompt attention will ensure credit or exchange. ALL DISCREPANCIES BETWEEN THE PACKING LIST AND THE PRODUCTS RECEIVED MUST BE REPORTED WITHIN 48 HOURS OF RECEIPT OF THE PACKAGE TO QUALIFY FOR A CREDIT.

Policy on Returned Goods NO RETURNS OR EXCHANGES WILL BE ACCEPTED UNLESS YOU HAVE RECEIVED A RETURN OF GOODS AUTHORIZATION NUMBER (RGA) FROM BYRON MEDICAL. THIS NUMBER MUST BE ON THE RETURN BOX AND ALL WRITTEN COMMUNICATION. Return numbers will expire if items are not returned within 30 days. Credit will only be issued if the merchandise is in saleable condition. If the product has been used in surgery, the product must be CLEANED WITH A DISINFECTANT and a written statement confirming such must accompany the product(s). Full credit will be issued for any item returned in 30 days that is in undamaged and saleable condition. A restocking charge of 25% will be charged to any product returned between 31-90 days. A credit memo, based on the dollar amount, will be issued to the account. OPENED PACKAGES OR BOXES containing disposable items are not returnable for credit or exchange. Each package or sales unit specifically states “Not Returnable if Package Seal is Broken.” NO CREDIT WILL BE ISSUED ON ANY ITEM RETURNED AFTER 90 DAYS. CREDIT MEMOS EXPIRE TWO YEARS AFTER DATE OF ISSUE.

Page 6: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 4 of 21 Rev D

Repair Program

If repairs are necessary due to damage other than that incurred during initial shipment (see section labeled DAMAGE OR LOSS IN SHIPMENT), contact Byron Medical to return your unit. A Return of Goods Authorization number must be obtained from Byron Medical’s Customer Service Department prior to returning any merchandise. When requesting a Return Goods Authorization number, please follow the return policy as listed under the section labeled POLICY ON RETURNED GOODS. Then, please carefully repack and return it post-paid to:

Byron Medical, Inc. Attn: Repairs Department

RGA #:________________ 602 West Rillito Street Tucson, Arizona 85705

(520) 573-0857 (800) 777-3434

Repairs must be made by Byron Medical or by an approved authorized agent. Attempting repair without prior authorization nullifies all warranties.

Warranty Policy Byron Medical products are manufactured for use only by qualified medical personnel who are trained in their use. This equipment carries a one-year warranty against defects from date of sale, which warrants it to be free from defects in material and workmanship. This warranty is valid only to the original purchaser, and will be voided if transferred to a third party. Any Byron Medical product with such defects returned will be promptly repaired or replaced at no charge to the customer. However, you are responsible for the cost of shipping the product to us. The warranty does not apply to damage caused by misuse, mishandling, improper operation, excessive voltage, and/or abuse of the product. Repairs or modifications performed other than by Byron Medical or an approved authorized repair facility will nullify this warranty. For details and warranty information, call Customer Service Department, (520) 573-0857 or (800) 777-3434.

Page 7: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 5 of 21 Rev D

II. Operation

General Description The PSI-TEC III Aspiration Platform is a third generation aspiration system with a high flow and powerful vacuum pump designed to evacuate fluid, soft tissue and general tissue removal from a surgical site. It also has a compressor used to drive pneumatic components, all on an integrated cart making it compact, mobile, attachments located at working level, and a pneumatically operated footswitch.

Operative Controls and Components

1. Rails (Assembly Required)

2. ON/OFF Switches (Vacuum and Air Pumps)

3. Vacuum Level Read-out (0-30 in of Hg)

4. Pneumatic Footswitch

5. Compressed Air (Return/In and Out)

System Set-Up 1. REMOVE ¾” Button Head Screws (8 each) from unit, and assemble rails to unit. a. Insert Flat Head Screws (8 each), and tighten screws with enclosed Hex b. Attach top rail L-Key (See detailed view on page 19). c. While holding top rail, secure sides with d. Using Male Knobs (2 each), tighten IV pole mount the Female Knobs (2each). to back of railing, and then insert IV bag holder.

1

4

2

35

Page 8: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 6 of 21 Rev D

2. With unit in “OFF” position, attach pneumatic footswitch and power cord to back of unit, and then plug-in power cord to electrical outlet. (Figure 1)

4. Insert IP Canisters

liners, and with 3. Place canister bracket and finger tips push down

dual-canister holder with on liners. (Figure 3) canisters on rail. (Figure 2)

5. Attach disposable filter to connection port

turning clock-wise until hand tight. (Figure 4)

6. Canister Set-Up, see page 5.

7. Remove white cap by turning counter-clockwise . Replace cap by placing it over the (AIR-OUT) port, then turn cap clockwise .

Operating Instructions

1. Switch “ON” aspirator vacuum and compressed air pumps , unit is ready for operation. Depressing the footswitch will activate or deactivate aspirator vacuum (suction) system.

NOTE – See INSTRUCTION MANUAL of applicable unit for proper connection and operation of pneumatic components with the PSI-TEC Systems.

Power Cord

Pneumatic Footswitch

Figure 1

Connection Port

Figure 2

Figure 3

Figure 4

NOTE – Vacuum will not occur in the IP Canister system without BOTH liners installed.

Page 9: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 7 of 21 Rev D

Canister Set-Up Illustration Step 1 Attach disposable filter to connection port turning clock-wise until hand tight. Step 2 Connect filter tubing from disposable filter to port A on CANISTER 2. Step 3 Connect aspirator tubing from port B on CANISTER 2 to port C on CANISTER 1. Step 4 Connect canister tubing E on CANISTER 2 to port D on CANISTER 1.

Step 5 Connect other canister tubing G on CANISTER 1 to port F on CANISTER 2.

Step 6 Using sterile technique, open tubing package (PT-5558 PSI-TEC); connect one end to port H on CANISTER 1, and other end to the desired cannula instrumentation.

A

B

Canister 2Canister 1

C

D

E

G H

Disposable Filter

F

Page 10: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 8 of 21 Rev D

CAUTION – STOP surgical aspiration PRIOR to removing liners from canisters.

WARNING – Single canister usage NOT recommended, as there is a greater probability of filter/ pump contamination.

2. (To CHANGE Filled Liner):

a. Switch aspirator vacuum [ ] “OFF” when liner in CANISTER 1 is full (1500 - 1700cc).

b. Disconnect aspiration tubing from port A on CANISTER 1.

c. Disconnect canister tubing from port B on CANISTER 2, and then connect it to port A on CANISTER 1. Capping the port prevents waste overflow of liner.

d. Remove liner, dispose liner and tubing according to

your guidelines for HAZARDOUS WASTE DISPOSAL.

6. (To REINSTALL New Liner):

a. Place new liner in CANISTER 1, and with finger tips

push down on lid of liner.

b. Connect loose end of canister tubing from CANISTER 1 to port B on CANISTER 2.

c. Reconnect aspiration tubing to port A on CANISTER 1, switch “ON” aspirator vacuum.

III. Maintenance Cleaning Instructions The unit exterior should be wiped down with a damp soft cloth and germicide. DO NOT use cleaning solutions containing ALCOHOL or SOLVENTS. WARNING – DO NOT IMMERSE IN LIQUID. STORE COVERED AND IN A CLEAN DRY PLACE. CONTACT PROPER AUTHORIZED PERSONNEL FOR SERVICING.

DO NOT autoclave or ETO sterilize aspirator, canisters, liners, or single-use items.

Canister 2Canister 1

B A

Canister LINER Removal

Page 11: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 9 of 21 Rev D

IV. Troubleshooting

Problems, Causes, and Solutions

Problem Possible Cause Solution Footswitch not activated. Depress Footswitch.

No vacuum

System not receiving power. Plug Power Cord into electrical source.

Vacuum Switch [ ] “OFF”.

Push Vacuum Switch [ ] “ON”.

Footswitch will not control aspirator.

Footswitch not connected to unit.

Connect footswitch to unit (rear).

Power cord not plugged-in to electrical outlet.

Plug-in power cord to electrical outlet.

System will not turn “ON”.

Fuses blown on back of unit tripped located just above power cord (female connector)

Unplug power cord, replace blown fuse, reconnect power cord, switch on vacuum, and depress footswitch.

IP Canister Liner(s) not properly sealed.

With vacuum [ ] “ON’ gently push down on lid.

Low vacuum.

Tubing improperly connected. Check all tubing at connection points for proper seal.

Aspirator stalls after quick restart.

Aspiration pumps still under load (vacuum).

Release vacuum by venting with air (remove cannula from patient and restart).

No Compression System not receiving power. Plug Power Cord into electrical source

Air Switch [ ] “OFF”. Push Air Switch [ ] “ON”.

Pneumatic component is inoperative.

Pneumatic component is improperly connected to aspiration compressed air [ ] fittings.

Ensure pneumatic component is properly connected. Check for visible damage to connectors and tubing.

If still inoperative, Contact Byron Medical, (800) 777-3434 or (520) 573-0857 for technical support.

Page 12: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 10 of 21 Rev D

PSI-TEC III Aspiration Platform Specifications

PSI-TEC III ASPIRATION PLATFORM 115V

PSI-TEC III ASPIRATION PLATFORM 230V

Size 20.18 in (H) x 22.96 in (D) x 18.32 in (W) 51.26 cm (H)x 58.32 cm (D) x 46.53 cm (W)

20.18 in (H) x 22.96 in (D) x 18.32 in (W) 51.26 cm (H) x 58.32 cm (D) x 46.53 cm (W)

Weight 80 lbs. / 36.288 Kilograms 80 lbs. / 36.288 Kilograms Power Cord 10 ft / 3 m Hospital Grade IEC 10 ft / 3 m Hospital Grade IEC AC Power Input 115 VAC, 60 Hz, 7.5 A 230 VAC, 50 Hz, 3.1 A Fuse T 15 A 250 V T 6.3 A 250V Maximum Vacuum 28 in of Hg. / 711 mm of Hg 28 in of Hg. / 711 mm of Hg Maximum Flow Rate 4.75 CFM (Cubic feet per minute) 4.75 CFM (Cubic feet per minute) Sound Level 52 dB Typical 52 dB Typical Pump Type(s) Two cylinder Piston/Single Piston Two cylinder Piston/Single Piston Maximum Compressor Air Pressure

29 inHg 736.6 mmHg

29 inHg 736.6 mmHg

Internal Pressure 100 psi MAX compressor output Regulated to 60 PSI

100 psi MAX compressor output Regulated to 60 PSI

Tank Capacity 250 psi Minimum (250% Max Compressor Capacity)

250 psi Minimum (250% Max Compressor Capacity)

IEC Classification Class 1 Type B Class 1 Type B Environmental

Operating Temperature 10 deg to 40 deg C 54 deg to 104 deg F

10 deg to 40 deg C 54 deg to 104 deg F

Storage/Shipping Temperature

-30 deg to 50 deg C -22 deg to 122 deg F

-30 deg to 50 deg C -22 deg to 122 deg F

Operating Humidity 30% to 75% relative humidity (non-condensing)

30% to 75% relative humidity (non-condensing)

APPENDIX - A

Page 13: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 11 of 21 Rev D

APPENDIX - B

PSI-TEC III 115V Platform MODEL: PT-PLT-III-110

PARTS LIST CMI NBR DESCRIPTION UOM Qty

2196 TIE, CABLE 4" EA 27 2520 SCREW, 4-40 X ¼ “ BUTTON HD MACHINE SLOTTED EA 1 2552 SCREW, 6-32 X ½ " PHILLIPS PAN HEAD MACHINE EA 6 2605 HEX NUT, 6-32 X ¼ " WIDTH EA 2 2620 HEX NUT, 8-32 X 11⁄32 " EA 4 2625 HEX NUT, 10-32 X 3⁄8 " WIDTH EA 10 2665 LOCK WASHER, #6 INTERNAL TOOTH EA 5 2675 LOCK WASHER, #8 EXTERNAL TOOTH EA 9 2690 LOCKWASHER, 10 INTERNAL TOOTH EA 10 2695 LOCK WASHER, #10 SPLIT EA 3 3126 NON-INSULATED TERMINAL ¼ " HOLE EA 3 4772 TIE-DOWN, FASTRAP EA 4 6112 ADHESIVE OZ 0.20 10176 FAN, 4” 110V MUFFIN EA 1 17060 TUBING, 3⁄16 " SHRINK BLACK IN 12 17217 LOCK WASHER, #6 SPLIT EA 2 19771 PIPE JOINT COMPOUND OZ 0.60 19861 POLY BAG, 6” X 9” EA 1 19974 POWER CORD, 125V 10A TO IEC EA 1 30400 BAG, CLEAR Z-LOCK 0.004M 12” X12” EA 1 32029 RECEPTACLE, 3 PIN EA 4 32030 PLUG, 3 PIN EA 4 32031 MALE TERMINAL PIN EA 20 32032 FEMALE TERMINAL PIN EA 20 32345 COMPRESSOR SIDE MTG BRKT(LH/RH) EA 2 32346 COMPRESSOR BASE PLATE PSI-TEC III EA 1 32347 VACUUM PUMP MTG PLATE (RH+LH) EA 2 32348 VACUUM PUMP MOTOR MOUNT EA 1 32351 AIR RECEIVER END CAP EA 2 32352 AIR RECEIVER TIE ROD EA 4 32353 AIR RECEIVER CYLINDER WALL EA 1 32357 RADIATOR BRACKET (RH + LH) EA 2 32361 ½ "-13 HEX NUT ZINC EA 4 32362 ¼ " LOCKWASHER ZINC EA 26 32364 3⁄8 " LOCKWASHER ZINC EA 4 32365 ½ " LOCKWASHER ZINC EA 4 32367 ⅜ “ – 16 X ½ “ BUTTON HEAD SKT CAP EA 4 32379 BRONZE MUFFLER 1⁄8 " EA 1 32388 ¼ " X 1⁄8 " ADAPTOR EA 1 32393 GAST VACUUM PUMP EA 1

Page 14: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 12 of 21 Rev D

PSI-TEC III 115V Platform (Continued) MODEL: PT-PLT-III-110

PARTS LIST CMI NBR DESCRIPTION UOM Qty

32394 GAST COMPRESSOR PSI-TEC III EA 1 32395 AIR LOGIC PRESSURE SWITCH EA 1 32396 RADIATOR 6” X 6” EA 1 32397 120 VAC - HUMPHREY ¼ " VALVE EA 1 32399 ¼ " NPT FEMALE CHECK VALVE EA 1 32401 KOBY MUFFLER EA 1 32402 SIDE ISOLATOR EA 4 32403 BOTTOM ISOLATOR EA 8 32404 RELAY R14-11A10-120 EA 2 32411 SWITCH, AIR ACTIVATED EA 1 32413 MOLEX 12 PIN FEMALE EA 1 32414 SHELL 12 PIN MALE EA 1 32415 CASTER, 4" PLATE MOUNT EA 2 32416 CASTER,LOCKING 4" PLATE MOUNT EA 2 32417 PSI-TEC III U SHAPED BAR EA 1 32420 IV POLE MOUNT EA 1 32430 MALE 1⁄8 “- ¼ “ MALE ELBOW EA 1 32431 FLEX HOSE FT 1 32432 SCREW, 10-32 X ½ " BUTTON HEAD CAP EA 11 32434 ¼ " BRONZE MUFFLER EA 1 32435 MALE ¼ " - MALE ¼ " ELBOW EA 1 32437 MALE ¼ " - FEMALE 1⁄8 " ADAPTOR EA 1 32438 ¼ " - ½ " MALE ELBOW EA 1 32439 1⁄8 " - 1⁄16 " RIGHT ANGLE ADAPTOR EA 1 32442 2 FOOT. FLEX HOSE WITH ½ " FEMALE EA 1 32446 BASE PLATE, PSI-TEC III EA 1 32448 SWITCH, 16 AMP LIGHTED ROCKER EA 2 32449 CONN MOLEX BLADE CRIMP EA 19 32450 TEE ¼ “ – ⅜ “ COMPRESSION EA 1 32451 3⁄8 " STAINLESS STEEL TUBE 41” X 6 EA 1 32453 3⁄8 " COMP UNION EA 1 32454 ¼ " - ¼ " MALE TO MALE COUPLER EA 2 32455 RELAY MTG BLOCK WITH RETAINER EA 2 32456 TERMINAL,FORKED CRIMP #8 EA 16 32457 GREEN/YELLOW 16 GAUGE WIRE FT 6.25 32458 BLUE 16 GAUGE WIRE FT 19.17 32459 BROWN 16 GAUGE WIRE FT 32.75 32460 FAN BRACKET EA 1 32462 SIDE SUPPORT EA 3 32465 VACUUM GAUGE 2 ½ " DIAMETER EA 1 32466 1⁄8 “ ID TUBING IN 35 32468 AIR OUTLET CONNECTOR EA 1

Page 15: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 13 of 21 Rev D

PSI-TEC III 115V Platform (Continued) MODEL: PT-PLT-III-110

PARTS LIST CMI NBR DESCRIPTION UOM Qty

32471 ½ " COMPRESSION TO ¼ " NPT EA 1 32472 ¼ "- ¼ " BULK HEAD ADAPTOR EA 2 32476 IV POLE EA 1 32477 POWER ENTRY MODULE 110-220V 10A EA 1 32478 SCREW, ¼ "-20 X ¾ " SOCKET HEAD CAP EA 6 32481 FUSE, 15AMP FAST ACTING EA 2 32483 SCREW, ¼ "-20 X ½ " SOCKET HEAD CAP EA 28 32485 SCREW, 8-32 X 3⁄8 " PAN HEAD PHILLIPS EA 2 32486 O-RING EA 2 32487 SCREW, 5⁄16 "-18 X 1" BUTTON HEAD EA 4 32488 LOCKWASHER, 5⁄16 " ZINC EA 8 32489 NUT, 5⁄16 "-18 ZINC EA 4 32490 ADAPTOR, GAUGE VACUUM EA 1 32491 PLUG, 1⁄8 " HEX PIPE EA 2 32493 1⁄8 " QUICK RELEASE TO 3⁄8 " NPT EA 1 32494 1⁄8 " QUICK RELEASE TO 10-32 EA 1 32495 TUBING, 1⁄8 " ID X ¼ " OD SUPER THIN FT 3.25 32497 CLIP,RELAY RETENTION FOR #32404 EA 2 32502 ATO MUFFLER, ¼ “ MALE EA 1 32507 FILTER, DUAL STAGE EA 1 32527 MACHINE SCREW, 8-32 X 5⁄8 " TRUS EA 3 32541 FOOTSWITCH EA 1 32542 1 ¼ " FLUTED PL KNOB EA 2 32543 1 ¼ " KNOB WITH STUD EA 3 32546 SCREW, 8-32 X ¾ “FLATHEAD EA 4 32555 CONNECTOR, 16 AWG 1.187 BLADE CRIMP EA 2 32556 COUPLER, 3⁄8 “ TUBING TO 1⁄8 “ BARB EA 1 32566 LUG, SOLDER #6 FLAT LOCKING EA 1 32567 TUBING, 3⁄8 “ SPLIT LOOM FT 3 32568 SHIPPING MATERIALS,PSI-TEC III EA 1 32588 PSI-TEC III SIDE RAIL EA 2 32589 PSI-TEC III BACK RAIL EA 1 32591 5⁄32 " HEX WRENCH, SHORT EA 1 32595 3” X 4” PLASTIC BAG EA 1 32599 SAFETY RELIEF VALVE, 100 PSI ¼ “ NPT EA 1 32600 AIR OUTLET SAFETY COVER EA 1 32602 CHAIN, #6 BALL STAINLESS FT 0.25 32603 CLIP, #6 BALL CHAIN STAINLESS EA 2 32605 TEE, ¼ “ NPT TO DUAL ¼ “ NPT FEMALE EA 1 32606 SCREW, 10-32 X 5⁄8 “ BUTTON HEAD MACHINE SLOTTED EA 2 32607 FUSE DRAWER, POWER INLET EA 1 32614 LOCKNUT, ¼ “-20 EA 12

Page 16: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 14 of 21 Rev D

PSI-TEC III 115V Platform (Continued) MODEL: PT-PLT-III-110

PARTS LIST CMI NBR DESCRIPTION UOM Qty

32615 LOCKNUT, 3⁄8 “-16 EA 4 32616 LOCKNUT, ½ “-13 EA 4 32628 TY-RAP, 30 NYLON EA 1 32644 ¼ “-20 X 1” FLATHEAD COUNTER SINK EA 8 32666 PSI-TEC III SHIPPING CONTAINER EA 1 32680 ENCLOSURE, PLATFORM ALUMINUM EA 1 32682 PENETROX ANTI-OXIDANT OZ 0.50 32694 SCREW, 6-32 X 5⁄8 “ FLAT HEAD MACHINE SCREW PHILLIPS EA 2 32695 SCREW, 8-32 X ½ ” PAN HEAD WITH WASHER EA 10 32696 SCREW, 7⁄16 “-14 X 1 ½ ” STAINLESS EA 1 32697 LOCKWASHER, 7⁄16 “ SPLIT EA 1 32698 HOSE, 7⁄16 “ ID RUBBER FT 0.10 32710 5⁄16 “- ¼ “ QUICK CONNECTOR COUPLING EA 1 32714 PSI-TEC III BACK PANEL EA 1

Page 17: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 15 of 21 Rev D

APPENDIX – B (Continued)

PSI-TEC III 230V Platform MODEL: PT-PLT-III-220

PARTS LIST CMI NBR DESCRIPTION UOM Qty

2196 TIE, CABLE 4" EA 27 2520 SCREW, 4-40 X ¼ “ BUTTON HD MACHINE SLOTTED EA 1 2552 SCREW, 6-32 X ½ " PHILLIPS PAN HEAD MACHINE EA 6 2605 HEX NUT, 6-32 X ¼ " WIDTH EA 2 2620 HEX NUT, 8-32 X 11⁄32 " EA 4 2625 HEX NUT, 10-32 X 3⁄8 " WIDTH EA 10 2665 LOCK WASHER, #6 INTERNAL TOOTH EA 5 2675 LOCK WASHER, #8 EXTERNAL TOOTH EA 9 2690 LOCKWASHER, 10 INTERNAL TOOTH EA 10 2695 LOCK WASHER, #10 SPLIT EA 3 3126 NON-INSULATED TERMINAL ¼ " HOLE EA 3 4772 TIE-DOWN, FASTRAP EA 4 6112 ADHESIVE OZ 0.20 17060 TUBING, 3⁄16 " SHRINK BLACK IN 12 17217 LOCK WASHER, #6 SPLIT EA 2 19771 PIPE JOINT COMPOUND OZ 0.60 19861 POLY BAG, 6” X 9” EA 1 30400 BAG, CLEAR Z-LOCK 0.004M 12” X12” EA 1 32029 RECEPTACLE, 3 PIN EA 4 32030 PLUG, 3 PIN EA 4 32031 MALE TERMINAL PIN EA 20 32032 FEMALE TERMINAL PIN EA 20 32345 COMPRESSOR SIDE MTG BRKT(LH/RH) EA 2 32346 COMPRESSOR BASE PLATE PSI-TEC III EA 1 32347 VACUUM PUMP MTG PLATE (RH+LH) EA 2 32348 VACUUM PUMP MOTOR MOUNT EA 1 32351 AIR RECEIVER END CAP EA 2 32352 AIR RECEIVER TIE ROD EA 4 32353 AIR RECEIVER CYLINDER WALL EA 1 32357 RADIATOR BRACKET (RH + LH) EA 2 32361 ½ "-13 HEX NUT ZINC EA 4 32362 ¼ " LOCKWASHER ZINC EA 26 32364 3⁄8 " LOCKWASHER ZINC EA 4 32365 ½ " LOCKWASHER ZINC EA 4 32367 ⅜ “ – 16 X ½ “ BUTTON HEAD SKT CAP EA 4 32379 BRONZE MUFFLER 1⁄8 " EA 1 32388 ¼ " X 1⁄8 " ADAPTOR EA 1 32395 AIR LOGIC PRESSURE SWITCH EA 1 32396 RADIATOR 6” X 6” EA 1 32397 120 VAC - HUMPHREY ¼ " VALVE EA 1

Page 18: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 16 of 21 Rev D

PSI-TEC III 230V Platform(Continued) MODEL: PT-PLT-III-220

PARTS LIST CMI NBR DESCRIPTION UOM Qty

32399 ¼ " NPT FEMALE CHECK VALVE EA 1 32401 KOBY MUFFLER EA 1 32402 SIDE ISOLATOR EA 4 32403 BOTTOM ISOLATOR EA 8 32411 SWITCH, AIR ACTIVATED EA 1 32413 MOLEX 12 PIN FEMALE EA 1 32414 SHELL 12 PIN MALE EA 1 32415 CASTER, 4" PLATE MOUNT EA 2 32416 CASTER,LOCKING 4" PLATE MOUNT EA 2 32417 PSI-TEC III U SHAPED BAR EA 1 32420 IV POLE MOUNT EA 1 32430 MALE 1⁄8 “- ¼ “ MALE ELBOW EA 1 32431 FLEX HOSE FT 1 32432 SCREW, 10-32 X ½ " BUTTON HEAD CAP EA 11 32434 ¼ " BRONZE MUFFLER EA 1 32435 MALE ¼ " - MALE ¼ " ELBOW EA 1 32437 MALE ¼ " - FEMALE 1⁄8 " ADAPTOR EA 1 32438 ¼ " - ½ " MALE ELBOW EA 1 32439 1⁄8 " - 1⁄16 " RIGHT ANGLE ADAPTOR EA 1 32442 2 FOOT. FLEX HOSE WITH ½ " FEMALE EA 1 32446 BASE PLATE, PSI-TEC III EA 1 32448 SWITCH, 16 AMP LIGHTED ROCKER EA 2 32449 CONN MOLEX BLADE CRIMP EA 19 32450 TEE ¼ “ – ⅜ “ COMPRESSION EA 1 32451 3⁄8 " STAINLESS STEEL TUBE 41” X 6 EA 1 32453 3⁄8 " COMP UNION EA 1 32454 ¼ " - ¼ " MALE TO MALE COUPLER EA 2 32455 RELAY MTG BLOCK WITH RETAINER EA 2 32456 TERMINAL,FORKED CRIMP #8 EA 16 32457 GREEN/YELLOW 16 GAUGE WIRE FT 6.25 32458 BLUE 16 GAUGE WIRE FT 19.17 32459 BROWN 16 GAUGE WIRE FT 32.75 32460 FAN BRACKET EA 1 32462 SIDE SUPPORT EA 3 32465 VACUUM GAUGE 2 ½ " DIAMETER EA 1 32466 1⁄8 “ ID TUBING IN 35 32468 AIR OUTLET CONNECTOR EA 1 32471 ½ " COMPRESSION TO ¼ " NPT EA 1 32472 ¼ "- ¼ " BULK HEAD ADAPTOR EA 2 32476 IV POLE EA 1 32477 POWER ENTRY MODULE 110-220V 10A EA 1 32478 SCREW, ¼ "-20 X ¾ " SOCKET HEAD CAP EA 6

Page 19: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 17 of 21 Rev D

PSI-TEC III 230V Platform(Continued) MODEL: PT-PLT-III-220

PARTS LIST CMI NBR DESCRIPTION UOM Qty

32483 SCREW, ¼ "-20 X ½ " SOCKET HEAD CAP EA 28 32485 SCREW, 8-32 X 3⁄8 " PAN HEAD PHILLIPS EA 2 32486 O-RING EA 2 32487 SCREW, 5⁄16 "-18 X 1" BUTTON HEAD EA 4 32488 LOCKWASHER, 5⁄16 " ZINC EA 8 32489 NUT, 5⁄16 "-18 ZINC EA 4 32490 ADAPTOR, GAUGE VACUUM EA 1 32491 PLUG, 1⁄8 " HEX PIPE EA 2 32493 1⁄8 " QUICK RELEASE TO 3⁄8 " NPT EA 1 32494 1⁄8 " QUICK RELEASE TO 10-32 EA 1 32495 TUBING, 1⁄8 " ID X ¼ " OD SUPER THIN FT 3.25 32497 CLIP,RELAY RETENTION FOR #32404 EA 2 32498 VALVE, 320 ¼ “ FOR 240 VAC EA 1 32499 VALVE, 310 1⁄8 “ FOR 220 VAC EA 1 32500 RELAY, R14-11A10-10-240 EA 2 32502 ATO MUFFLER, ¼ “ MALE EA 1 32504 COMPRESSOR, 230 VOLT GAST EA 1 32505 PUMP, 230 VOLT VACUUM GAST EA 1 32506 FAN, 4” 220/230V MUFFIN EA 1 32507 FILTER, DUAL STAGE EA 1 32527 MACHINE SCREW, 8-32 X 5⁄8 " TRUS EA 3 32541 FOOTSWITCH EA 1 32542 1 ¼ " FLUTED PL KNOB EA 2 32543 1 ¼ " KNOB WITH STUD EA 3 32546 SCREW, 8-32 X ¾ “FLATHEAD EA 4 32555 CONNECTOR, 16 AWG 1.187 BLADE CRIMP EA 2 32556 COUPLER, 3⁄8 “ TUBING TO 1⁄8 “ BARB EA 1 32566 LUG, SOLDER #6 FLAT LOCKING EA 1 32567 TUBING, 3⁄8 “ SPLIT LOOM FT 3 32568 SHIPPING MATERIALS,PSI-TEC III EA 1 32588 PSI-TEC III SIDE RAIL EA 2 32589 PSI-TEC III BACK RAIL EA 1 32591 5⁄32 " HEX WRENCH, SHORT EA 1 32595 3” X 4” PLASTIC BAG EA 1 32596 POWER CORD, 250V 10A EUROPEAN EA 1 32599 SAFETY RELIEF VALVE, 100 PSI ¼ “ NPT EA 1 32600 AIR OUTLET SAFETY COVER EA 1 32602 CHAIN, #6 BALL STAINLESS FT 0.25 32603 CLIP, #6 BALL CHAIN STAINLESS EA 2 32605 TEE, ¼ “ NPT TO DUAL ¼ “ NPT FEMALE EA 1 32606 SCREW, 10-32 X 5⁄8 “ BUTTON HEAD MACHINE SLOTTED EA 2 32607 FUSE DRAWER, POWER INLET EA 1

Page 20: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 18 of 21 Rev D

PSI-TEC III 230V Platform(Continued) MODEL: PT-PLT-III-220

PARTS LIST CMI NBR DESCRIPTION UOM Qty

32614 LOCKNUT, ¼ “-20 EA 12 32615 LOCKNUT, 3⁄8 “-16 EA 4 32616 LOCKNUT, ½ “-13 EA 4 32625 FUSE 6.3A 250V SLOW-BLOW GLOW EA 2 32628 TY-RAP, 30 NYLON EA 1 32644 ¼ “-20 X 1” FLATHEAD COUNTER SINK EA 8 32666 PSI-TEC III SHIPPING CONTAINER EA 1 32680 ENCLOSURE, PLATFORM ALUMINUM EA 1 32682 PENETROX ANTI-OXIDANT OZ 0.50 32694 SCREW, 6-32 X 5⁄8 “ FLAT HEAD MACHINE SCREW PHILLIPS EA 2 32695 SCREW, 8-32 X ½ ” PAN HEAD WITH WASHER EA 10 32696 SCREW, 7⁄16 “-14 X 1 ½ ” STAINLESS EA 1 32697 LOCKWASHER, 7⁄16 “ SPLIT EA 1 32698 HOSE, 7⁄16 “ ID RUBBER FT 0.10 32710 5⁄16 “- ¼ “ QUICK CONNECTOR COUPLING EA 1 32714 PSI-TEC III BACK PANEL EA 1

Page 21: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 19 of 21 Rev D

APPENDIX - C

1” Flat Head Screw (8)

When Attaching Rails To PSITEC III Replace The 3/4” Button Head Screws With The 1” Flat Head Screw

Page 22: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 20 of 21 Rev D

APPENDIX - D

Page 23: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

21-OM-0001 21 of 21 Rev D

APPENDIX – D (Continued)

Page 24: Byron Medical’s PSI-TEC III (with White Cap) · 21-OM-0001 Rev D Byron Medical’s PSI-TEC III Aspiration Platform (with White Cap) Instruction Manual PT-PLT-III-110 PT-PLT-III-220

Byron Medical 602 West Rillito Street Tucson, Arizona 85705

800-777-3434 Printed in the USA


Recommended