Table of Contents Chapter 1 Overview......................................................................................................... 1
1.1 Safety Instructions ............................................................................................... 1
1.1.1 Symbols ..................................................................................................... 1
1.1.2 Labels ........................................................................................................ 3
1.1.3 Biohazard Considerations.......................................................................... 3
1.1.4 Electrical Safety ......................................................................................... 4
1.1.5 Electromagnetic compatibility .................................................................... 6
1.2 Components......................................................................................................... 9
1.2.1 Appearance................................................................................................ 9
1.2.2 Control Panel ............................................................................................11
1.2.3 Probes ......................................................................................................11
1.3 Equipment Installation and Debugging .............................................................. 12
1.3.1 Power Connection.................................................................................... 12
1.3.2 Probe Connection .................................................................................... 12
1.3.3 Probe Bracket, Cable Hook and Gel Cup Installation .............................. 13
1.3.4 Installation Environment........................................................................... 14
Chapter 2 System Controls and Display Functions ................................................... 15
2.1 System Controls................................................................................................. 15
2.1.1 Routine functional keys............................................................................ 15
2.1.2 System functional keys ............................................................................ 17
2.1.3 Image-mode keys .................................................................................... 17
2.1.4 Image functional keys .............................................................................. 18
2.1.5 Body Markers and Text keys.................................................................... 19
2.1.6 Record functional keys............................................................................. 20
2.1.7 Other functional keys ............................................................................... 21
2.2 Display function.................................................................................................. 22
2.2.1 Screen Interface ...................................................................................... 22
2.2.2 Monitor adjustment .................................................................................. 23
Chapter 3 Instrument Operating .................................................................................. 24
3.1 Precautions ........................................................................................................ 24
3.2 Turning the System on ....................................................................................... 24
3.3 System Settings ................................................................................................. 25
3.3.1 General Settings ...................................................................................... 26
3.3.2 Measurement Settings ............................................................................. 29
3.3.3 Setting Image Processing ........................................................................ 31
II
3.3.4 Version Information.................................................................................. 32
3.4 Selecting Probe and Frequency......................................................................... 32
3.4.1 Selecting Probe ....................................................................................... 32
3.4.2 Selecting Frequency ................................................................................ 32
3.5 Selecting Examination Type............................................................................... 33
3.6 New Patients...................................................................................................... 34
3.7 Switch Image Mode ........................................................................................... 35
3.7.1 B Mode .................................................................................................... 35
3.7.2 2B Mode .................................................................................................. 36
3.7.3 4B Mode .................................................................................................. 38
3.7.4 B/M Mode ................................................................................................ 39
3.8 Adjusting Image Parameter................................................................................ 41
3.8.1 Adjusting image brightness and contrast ................................................. 41
3.8.2 Gain Adjusting.......................................................................................... 42
3.8.3 Changing the Dynamic Range ................................................................. 42
3.8.4 Frame Correlation.................................................................................... 43
3.8.5 Edge Enhancement and Image Smoothing ............................................. 43
3.8.6 Gray Scale ............................................................................................... 44
3.8.7 Changing the Density............................................................................... 44
3.8.8 Adjusting Focus ....................................................................................... 45
3.8.9 Adjusting imaging depth........................................................................... 45
3.8.10 Adjusting image width ............................................................................ 45
3.8.11 Image Flip .............................................................................................. 46
3.8.12 Image Zoom........................................................................................... 47
3.8.13 Changing the M Sweep Speed .............................................................. 47
3.8.14 Harmonic Imaging.................................................................................. 47
3.8.15 Changing Power .................................................................................... 48
3.9 Annotation and Arrows ....................................................................................... 48
3.9.1 Annotation................................................................................................ 48
3.9.2 Arrows...................................................................................................... 49
3.10 Body Markers................................................................................................... 50
3.11 Biopsy Guideline .............................................................................................. 51
3.12 CINE ................................................................................................................ 53
3.13 Image Managing and Printing .......................................................................... 54
3.13.1 Image Managing .................................................................................... 54
3.13.2 Printing................................................................................................... 56
Chapter 4 Measurements and Reports........................................................................ 57
4.1 Overview of measurement ................................................................................. 57
III
4.1.1 Special Measurement Menu .................................................................... 57
4.1.2 Measurement Results .............................................................................. 58
4.1.3 Measurement Report ............................................................................... 58
4.2 General measurement — B-mode ..................................................................... 60
4.2.1 Distance Measurement ............................................................................ 60
4.2.2 Area/CIRC Measurement......................................................................... 63
4.2.3 Volume Measurement .............................................................................. 65
4.2.4 Angle Measurement ................................................................................. 68
4.2.5 Histogram Measurement.......................................................................... 69
4.3 General measurement — M-mode..................................................................... 71
4.3.1 Distance................................................................................................... 71
4.3.2 Time......................................................................................................... 72
4.3.3 Slope........................................................................................................ 74
4.3.4 Heart rate................................................................................................. 74
4.4 Gynecology Measurements ............................................................................... 75
4.4.1 Uterus Measurement ............................................................................... 76
4.4.2 Cervix Measurement................................................................................ 76
4.4.3 Ovary Measurement ................................................................................ 77
4.4.4 Follicle Measurement............................................................................... 77
4.4.5 Gynecologic Report ................................................................................. 78
4.5 Obstetrics Measurements .................................................................................. 79
4.5.1 Obstetrics Input........................................................................................ 79
4.5.2 Fetus Selection ........................................................................................ 80
4.5.3 Obstetrics Measurements ........................................................................ 81
4.5.4 Amniotic fluid index .................................................................................. 84
4.5.5 Fetal heart rate......................................................................................... 85
4.5.6 Neck Transparency .................................................................................. 85
4.5.7 Fetal physiology score ............................................................................. 85
4.5.8 Obstetrics Report ..................................................................................... 87
4.6 Orthopedics Measurements ............................................................................... 88
4.6.1 Orthopedics Report.................................................................................. 89
4.7 Urology Measurements ...................................................................................... 90
4.7.1 PSAD Measurement ................................................................................ 91
4.7.2 Prostate Volume Measurement................................................................ 91
4.7.3 Residual Urine Volume Measurement...................................................... 92
4.7.4 Urology Report......................................................................................... 92
4.8 Small Part Measurements.................................................................................. 93
4.8.1 Small Parts Measurement........................................................................ 94
IV
4.8.2 Small Parts Measurement Report ............................................................ 94
Chapter 5 Cleaning and Maintenance ......................................................................... 96
5.1 Cleaning............................................................................................................. 96
5.1.1 Cleaning the Main Unit............................................................................. 96
5.1.2 Cleaning the Probe .................................................................................. 97
5.2 Maintenance ...................................................................................................... 98
5.2.1 Main Unit Maintenance ............................................................................ 99
5.2.2 Caring Probe.......................................................................................... 101
5.2.3 Caring Coupling Agent ........................................................................... 102
Chapter 6 Technical Description................................................................................ 103
6.1 Technical Data ................................................................................................. 103
6.2 System Requirements...................................................................................... 106
6.2.1 Power supply Requirements .................................................................. 106
6.2.2 Connection with Peripheral Devices ...................................................... 106
6.2.3 Operating Environment Requirements................................................... 106
6.3 Transport and Storage ..................................................................................... 107
6.3.1 Requirement for Transport and Storage Environment............................ 107
6.3.2 System Transport................................................................................... 107
6.3.3 Probe Transport ..................................................................................... 107
6.4 Disposal of Packing Materials .......................................................................... 107
6.5 Statement on Consumables ............................................................................. 108
1
Chapter 1 Overview
1.1 Safety Instructions
Do not operate the ultrasound imaging system until you fully understand the safety
considerations and procedures presented in this manual.
1.1.1 Symbols
Table1-1 is provided for your identification of important symbols located in labels on the
ultrasound imaging system and probes.
Table 1-1 Symbols
Symbol Description
Type B applied part
Protected against the effects of temporary immersion
in water
Caution: Risk of electric shock. Do not open.
Caution: Refer to user’s manual.
Monitor contrast control
Monitor brightness control
Protective Grounding
Portable Ultrasound Diagnostic Scanner User’s Manual
2
Equipotential grounding connection
USB interface
Video output
On (mains power supply)
OFF (mains power supply)
Alternating Current (AC)
Date of manufacture
Manufacturer
Indicates separate collection for electrical and electronic
equipment in compliance with the Waste Electrical and
Electronic Equipment (WEEE) Directive
Trademark of Landwind
Chapter 1 Overview
3
1.1.2 Labels
Figure 1-1 Schematic Diagram of Label Position
Figure 1-2 Identification Label
1.1.3 Biohazard Considerations
When using the system, please pay attention to the following biosafety-related points:
Caution: For neonatal head imaging, Landwind recommends that you exercise
special care during neonatal cephalic scanning to avoid possible damage to the
posterior region of the eye. The ultrasound energy emitted by the probe easily
Portable Ultrasound Diagnostic Scanner User’s Manual
4
penetrates the fontanels of the infant.
Caution: Landwind makes every effort to manufacture safe and effective probes.
You must take all necessary precautions to eliminate the possibility of exposing
patients, operators, or third parties to hazardous or infectious materials. These
precautions should be considered in the use of any application that may indicate the
need for such care, and during endocavity scanning; during biopsy or puncture
procedures; or hen scanning patients with open wounds.
Caution: To eliminate the possibility of exposing patients, operators, or third parties
to hazardous or infectious materials, handle dangerous or infectious materials all
along according to the medical specifications for handling biohazard castoffs.
Caution: The assessment of biological effects of diagnostic ultrasound on humans is
a subject of ongoing scientific research. This system and all diagnostic ultrasound
procedures should be used for valid reasons, for the shortest period of time, and at
the lowest mechanical and thermal indices necessary to produce clinically acceptable
images.
Caution: To ensure the safety of examined patients, do not scan the same part of an
examined patient consecutively for a long period of time.
Caution: When making ultrasonic scanning, the instrument shall be used with
ultrasound coupling agent, and no biologic compatibility problem is found. However,
to ensure safety, only qualified ultrasound coupling agent in compliance with related
standard shall be used.
Caution: Do not use the system if irregular and unusual repeated video refreshes
occur. Discontinuous scanning frequency is a hardware fault, please remove it before
use.
1.1.4 Electrical Safety
The system is designed in line with the safety standard of EN 60601-1, and use Type B
isolation for the patient part. To ensure safety to the greatest extent, please pay attention
Chapter 1 Overview
5
to the following Cautions:
Caution: Improper ground of the instrument may result in the danger of electric
shock. The only means to prevent electric shock is to use three-phase cable and
plug. The power cable shall be inserted into a grounded socket, and the ground
wire should not be moved away or removed.
Caution: Only cable hooks provided by Landwind can be used.
Caution: Do not attempt to modify the instrument. If it is really necessary to do so,
please notify us or our agent.
Caution: Operators of the instrument shall be qualified service-trained
professionals or shall use the instrument under the supervision of qualified
professionals.
Caution: To make electrical or mechanical connection between this instrument and
the equipment of other manufactures, confirm that it complied with the
IEC60601-1-1:1995 standard, and recheck the leakage current and other safety
indices of the entire set of system, to avoid possible hazard arising from leakage
current overlapping.
Caution:Upon the expiration of the service life of the system, please notify your
dealer or Landwind to handle properly.
Caution: To avoid electrical shock and damage to the ultrasound system, turn off
and unplug the equipment from the AC power outlet before cleaning and
disinfecting.
Caution: Do not open the enclosure of this unit, as some components inside it
carry high voltages that may cause hazards. If it is really necessary to do so, please
contact your dealer or Landwind to handle properly.
Caution: Do not use the instruments in an environment with combustible gases or
anesthetic to avoid explosions.
Portable Ultrasound Diagnostic Scanner User’s Manual
6
Caution: Do not pour any fluid onto the system surfaces, as fluid seepage into the
electrical circuitry may cause excessive leakage current or system failure.
Caution: This unit must be serviced once every 12 months to ensure its safety and
functionality. Electric safety test must be performed at periodic intervals according
to the local safety regulations or as needed.
Caution: To avoid the possibility of static shock and damange to the instrument,
avoid the use of aerosol spray cleaners on the monitor screens.
Caution: The power supply of the instrument can only be connected to a special
power inteface. Never connect an unspecified device to the port of this instrument.
Pay attention to correct connection of the grounding wire; otherwise, electric shock
may easily occur.
Caution: Do not use the system in damp or wet environments. Keep it away from
inflammables and explosives.
Caution: First check this unit and its parts every time before you use it. Timely
replace any components with sign of damage.
Caution: Use the probe with care. Do not use damaged probes. Do not use probes
with cracks or surface crimps, since such probes might have their insulation
damaged ,which easily causes electric shock to patients.
1.1.5 Electromagnetic compatibility
Portable Ultrasound Diagnostic Scanner has passed the tests that confirmed its
compliance with electromagnetic compatibility (EMC) requirements of IEC
60601-1-2:2001+A1:2004. It is CISPR 11: Group 1 Class А.
Caution: Special EMC requirements are imposed on medical equipment. The
following should be taken into account when installing and operating such
equipment.
Strong electromagnetic fields induced over the air or the wires by portable and mobile
communication devices or powerful sources of radio-frequency emission can cause
Chapter 1 Overview
7
failures in Portable Ultrasound Diagnostic Scanner operation. Such failures are
manifested in impaired quality of ultrasound images produced by the system or in
distortion thereof, in distortion of measurement results, instability of the system operation
and other abnormal manifestations.
In the event that such disturbance is detected when operating the system, its source shall
be localized and the following measures shall be taken.
— To detect the disturbance source, disconnect the nearby devices and turn them on
again.
— Having detected the interference-emitting device, relocate or re-orient it.
— Increase the separation distance between the interference emitting device and
Portable Ultrasound Diagnostic Scanner .
— Do not use the equipment with operating frequencies close to frequencies of Portable
Ultrasound Diagnostic Scanner.
— The personnel in charge of monitoring the medical devices shall be trained to identify
potential problems associated with electromagnetic emission.
— Use of cellular phones and computers shall be restricted in the areas where
interference-sensitive devices are located.
— Electromagnetic emission shall be removed or eliminated technically, for example,
through shielding.
Caution: Use of probes, cables and accessories other than stated in the operational
documentation can entail RF emission enhancement or degradation of Portable
Ultrasound Diagnostic Scanner stability.
Declaration:
All cables, probes and other accessories, whose compliance with requirements of IEC
60601-1-2:2001+A1:2006 36.201 – EMISSION and 36.202 –IMMUNITY is declared by
the manufacturer.
Portable Ultrasound Diagnostic Scanner User’s Manual
8
Preferable operation conditions and electromagnetic compatibility levels of Portable
Ultrasound Diagnostic Scanner are given in Tables 1-2 ,1-3.
The system is designed to be operated in the below electromagnetic environment.
Table 1-2 EMISSION
EMISSION(EN 60601-1-2:2001+A1:2006)
Test Type Basic Standard
Mains terminal disturbance voltage
Radiated disturbance
EN 55011:1998+A1:1999+A2:2002
Harmonic current emissions EN 61000-3-2:2006
Voltage fluctuations & flicker EN 61000-3-3:1995+A1:2001+A2:2005
Table 1-3 IMMUNITY
IMMUNITY(EN 60601-1-2:2001+A1:2006)
Test Type Basic Standard
Electrostatic discharge immunity EN 61000-4-2:1995+A1:1998+A2:2001
Radiated, radio frequency
electromagnetic field immunity EN 61000-4-3:2006
Electrical fast transient/burst EN 61000 4-4:2004
Surge immunity EN 61000-4-5:2006
Chapter 1 Overview
9
Immunity to conducted disturbances
Induced by RF fields EN 61000-4-6:1996+A1:2001
Voltage dips and short interruptions
immunity EN 61000-4-11:2004
Power frequency magnetic field immunity EN61000-4-8:1993+A1:2001
1.2 Components
1.2.1 Appearance
1 Cable hook 2. Coupling agent case
3. Probe bracket 4. Keyboard
5. Contrast/brightness adjustment 6. Monitor
Figure 1-3 Front Left View of the Ultrasonic Diagnosis System
Portable Ultrasound Diagnostic Scanner User’s Manual
10
1. Output Video port 2. USB port 3. Power switch
4. RS-232-C serial port 5. Equipotential grounding connection 6. Fuse case
7. VGA terminal 8. Power input jack 9. LAN port
10. Remote terminal 11. Probe port
Figure 1-4 Rear View of the Ultrasonic Diagnosis system
Chapter 1 Overview
11
1.2.2 Control Panel
This section describes the overview of the control panel of the instrument.
Figure 1-5 Overview of Operation Panel
1.2.3 Probes
Multi-frequency capability is available for all imaging probes. You can connect two probes
to the ultrasound imaging system, with one probe being the active probe. The
configuration of probes depends on your purchase contract.
Model Type Description
35C60H convex probe center frequency: 3.5 MHz, R=60mm
65C13H convex probe center frequency: 6.5 MHz, R=13mm
75L40H linear probe center frequency: 7.5 MHz, L=40mm
Portable Ultrasound Diagnostic Scanner User’s Manual
12
1.3 Equipment Installation and Debugging
1.3.1 Power Connection
This section describes how to connect the 100-240V~ 50/60Hz power supply for the unit.
Please ensure a stable power supply for the equipment. The power interface card is
located on the back of the instrument. Check the power supply of the unit to ensure that it
is within the specified range. Connect one end of the power cable to the power input jack
(8) (see Figure 1-4), and the other to the power supply.
Caution: Do not connect any device to the external power outlet on your own.
Caution: Confirm the power voltage is the same as the nominal voltage of the
instrument before connection.
Caution: This instrument adopts two fuses. Do not change fuses on your own;
otherwise, it may cause the danger of electric shock.
This equipment includes two other fuses, which are only used by professional
equipment maintenance personnel.
1.3.2 Probe Connection
Operating procedure;
1. Turn the probe’s latch to the open position and plug the probe into the slot
(Figure 1-6).
2. Turn the latch clockwise to the locked position.
Figure 1-6 Connecting a probe
Chapter 1 Overview
13
Caution: Before probe connection, check whether latch is in open position. Do not
insert or pull out the connector in switch-on status to avoid damage to the probe.
Once the probe is connected the main unit, do not insert or pull out it at will.
Caution: Never force a probe into place. If you feel any resistance, reposition the
probe connector and try again.
Caution: Before disconnecting a probe, be sure to open the connector by turning the
latch to the open position.
1.3.3 Probe Bracket, Cable Hook and Gel Cup Installation
1. Probe Bracket and Gel Cup Installation.
The installation of Probe bracket an gel cup adopts the method of directly inserting
the position bore and locked, shown as follow:
Figure 1-7 Probe Bracket and Gel Cup Installation.
2. Cable Hook Installation
The installation of Cable Hook adopts the method of revolved to lock , shown as
follow:
Portable Ultrasound Diagnostic Scanner User’s Manual
14
Figure 1-8 Cable Hook Installation
1.3.4 Installation Environment
The instrument shall be installed in a stable place far from electric fields, devices with
strong magnetic fields and high-voltage devices. Avoid direct sunlight on the monitor
screen and keep a dark indoor environment to facilitate in image observation.
Furthermore, keep a ventilated and dustproof environment.
15
Chapter 2 System Controls and Display Functions
2.1 System Controls
2.1.1 Routine functional keys
Trackball
It is the main tool of screen interface and is for selection and location. Trackball can be
used in conjunction with other keys, when the machine is in different states, the trackball
has different functions. Under normal circumstances, trackball controls the cursor position.
Set
It is a Multi-function key, used in conjunction with the trackball. Its function changes in the
different state automatically, such as setting the cursor location, positioning location map,
noting the location or selecting menu item, increasing the parameter value.
Exit
Exit the current operation mode. It also used to switch the endpoints in DiST-LINE and Ellipse measurement.
FREEZE
Freeze or unfreeze image
Portable Ultrasound Diagnostic Scanner User’s Manual
16
Gain
It is a Multi-function key which can adjust the total gain of the system, switch the input
method and measurement items and turn page to view the measurement results.
Adjust the time gain compensation
Menu
Open or close the menu of real-time image.
Meas
Enter the conventional measurement function.
Probe
Select and switch to probes.
Chapter 2 System Controls and Display Functions
17
2.1.2 System functional keys
New Patient
Press this key to open the new patient dialog box to input the information of new patient.
Press this key again to close the dialog box.
Exam
Press this key to open exam dialog box to preset examination subjects. Press this key
again to close the dialog box.
Setup
Enter or exit the main menu system settings dialog box.
2.1.3 Image-mode keys
B Mode
Switch to single B-mode.
2B Mode
Switch to 2B -mode.
Portable Ultrasound Diagnostic Scanner User’s Manual
18
4B Mode
Switch to 4B-mode.
B/M Mode
Switch to B/M-mode or single M-mode.
2.1.4 Image functional keys
Zoom
Start or close B-mode image zoom function
Depth up
Increase scanning depth.
Depth down
Reduce scanning depth.
Focus up
Chapter 2 System Controls and Display Functions
19
Rise up focal zone position.
Focus down
Adjust down focal zone position.
Cine
Start the movie playback function.
2.1.5 Body Markers and Text keys
Mark
Press this key to enter the position-work state, and the body markers and the location of
the scanning probe can be input at the designated location of the screen. Press the key
again to exit this mode.
Text
Press this key to enter the text-work state, and input texts at the designated location of the
screen. Press the key again to exit this mode.
Arrow
Portable Ultrasound Diagnostic Scanner User’s Manual
20
Press this key to enter the arrow-work state, make arrow mark at the designated location
of the screen. Press the key again to exit this mode.
Clear
Press this key to clear the measurement results, body marks, arrows and text messages
on the screen.
2.1.6 Record functional keys
Report
Open or close the report.
Review
Open or close the image reviewing mode.
Save
Save images.
Chapter 2 System Controls and Display Functions
21
Activation the video printing function.
2.1.7 Other functional keys
Calculating key
Press this key to enter the special measurement state.
Biopsy
Enter or exit the puncture biopsy line display mode.
Backspace key
Press this key can delete a character from right to left when in text input state.
Shift
It is used to access characters on the upper right corner of the keyboard.
Caps
It used to access capital or lowercase letters.
Portable Ultrasound Diagnostic Scanner User’s Manual
22
2.2 Display function
2.2.1 Screen Interface
1 Trademark of Landwind;
2 patient information area: Name of hospital, Name of patient, Number of medical record, exam, etc;
3 Image information area: Total gain (G), Dynamic range (D), Frame correlation (FC), Frame rate (FR),
Edge enhancement and image smoothing (EE);
4 Date & Time display area;
5 USB logo: U-disk is inserted to the system;
6 Gray bar;
7 Horizontal scale: For images in B mode, it indicates horizontal distance (the distance between
different point is 10mm), and the scale varies with the magnification ratio of images. For images in M
mode, it indicates scanning time (the scanning time between different points is 1 second), and the
scale varies with the scanning speed of the M mode.
8 Focus position indicator;
9 Vertical scale: Indicates the vertical distance of an image (the distance between different points is
10mm), and the scale varies with the magnification ration of the image;
10 TGC curve display area;
11 Current active image display depth;
12 Quick button; 13 Cine display indicator;
14 Image display area;
15 Direction Logo: L indicates probe direction, corresponding to the small protrusion on the probe
handle;
16 System working mode;
18 17 16
14
13
12
11
10
9
8
7 6
5 4 3 1 2
15
Chapter 2 System Controls and Display Functions
23
17 Imaging mode;
18 Probe information display area;
Figure 2-1 Layout of Screen Operation Interface
Note on EMC: Using this instrument in strong electromagnetic areas near radio stations or
similar units may cause visual interference upon the monitor screen. However, after
proper design and test, the instrument can withstand similar interferences without any
permanent damage.
2.2.2 Monitor adjustment
After starting the instrument, adjust the brightness and contrast of the monitor or you may
make such adjustment during check at any time. For detailed adjustment, refer to 3.8.1
section.
24
Chapter 3 Instrument Operating
3.1 Precautions
Make sure that the AC power connector plug for the ultrasound system is a three-prong
grounded plug, and ensure that the AC power in the plug is 100-240V~ 50/60Hz.
Check all power suppliers and connection cables to make sure no such damage as
scratch and breakage.
Check the probes and connecting parts to make sure no such damage as scratch and
desquamate.
Illustrations are added in appropriate places of this chapter, and please note that the
actual product shall apply if the details of illustrations differ from the actual product.
3.2 Turning the System on
Make sure that one end of the power cable is inserted into the power input jack on the
power interface card (refer to Figure 1-4) of the instrument, and the other end is inserted
into the AC power jack. Turn on the main power switch (for its position, refer to Figure 1-4.
Press to turn on, and press to turn off), the system startup screen appears. The
system shows the main screen, displaying B-mode images as default (as shown in Figure
3-12).
Please check whether the instrument is started normally, and check the images and
operation panel to make sure there is no abnormality. If no fault occurs, apply coupling
agent on the transmission surface of the probe, and make preparation for patient
examination.
Caution: Please stop operation if any fault is found, and contact your supplier for
troubleshooting.
Caution: Black boot time is about 30 seconds to respond, which is a normal
phenomenon, please be patient.
Chapter 3 Instrument Operations
25
Caution: To ensure long-term stable and reliable work of the instrument, do not turn
on the instrument immediately after power-off. Please wait for at least one minute.
3.3 System Settings
System Settings are used to set the power-on operating environment, status, and
configuration parameters of different examination modes. Preset values are saved in the
memory of the machine so that it will not get lost upon power failure and will ensure the
system runs at the expected state each time when powered on. Reset default setting can
restore the parameters in factory.
Press the 【Setup】 key on the keyboard to access the system settings. The system
settings dialog is displayed on the screen, and the General Settings menu appears by
default, as shown in Figure 3-1.
Setting items in the dialog include General Settings, Measurement Settings, Image
Processing Settings and Version information.
Figure 3-1 main menu of System Setup
Portable Ultrasound Diagnostic Scanner User’s Manual
26
3.3.1 General Settings
General Settings include Date And Time, Time Format, TGC Autohide, Default Port, Cine
Playback Speed, Hospital Name, System Language, System Environment, Network
Status, Exam Settings, Report mode and Biopsy Settings.
3.3.1.1 Setting Date And Time
Move cursor to Date And Time item, and then press the 【Set】 key to click it, then the
time setting screen appears. After adjusting system time, press to save setting
and press (or press【Setup】key) to exit the system setting.
Figure 3-2 Date and Time Settings
3.3.1.2 Setting Time Format
Move cursor to the Time Format item, and then press the 【Set】key to click it, then move
cursor to your favorite date format and press the 【Set】key to click it. Press to
save setting and press (or press【Setup】key) to exit the system setting.
Figure 3-3 Setting Time Format
3.3.1.3 Setting TGC Autohide
The system supports two modes of displaying TGC curve: AutoHide and NoneAutoHide.
Make a choice according to yourself. The detail is the same as 3.3.1.2.
Chapter 3 Instrument Operations
27
3.3.1.4 Setting Default Port
Choose probe port I or II as the startup probe by default by yourself. The detail is the same
as 3.3.1.2.
3.3.1.5 Setting Cine Playback Speed
The system supports 1/1, 1/2, 1/4, 1/8 four kinds of speeds to choose, the detail is the
same as 3.3.1.2.
3.3.1.6 Setting Hospital Name
Move cursor to the hospital name setting item, and then press the【Set】key to click it.
Input a name in the hospital name text box. English is the default language. After inputting
hospital name, press to save setting and press (or press【Setup】key)
to exit the system setting.
Figure 3-4 Setting Hospital Name
3.3.1.7 Choosing System Language
The system supports two language modes: English and Chinese. The detail is the same
as 3.3.1.2 section.
3.3.1.8 Modifying Network Status
Move cursor to the network setting item, and then press the 【Set】key to click it. The
screen will appear shown as Figure 3-5. Setting network according to your actual need,
and then press to save setting and press (or press【Setup】key) to
exit the system setting.
Portable Ultrasound Diagnostic Scanner User’s Manual
28
Figure 3-5 Network Status
3.3.1.9 Exam Settings
The system supports the exams including Abdomen, Gynecology, Obstetrical, Small Part,
Orthopaedy, Urology and Emergency. After choosing, the system will initialize the current
selected examination type which is the default exam when you restart the system.
3.3.1.10 Setting Report Mode
The system supports two report modes: Conventional Report and Specialist Report. The
detail is the same as 3.3.1.2.
3.3.1.11 Biopsy Settings
In this menu, you can set the biopsy direction, start point, biopsy angle and depth prompt.
The detail is the same as 3.3.1.2.
Figure 3-6 Biopsy Settings
Chapter 3 Instrument Operations
29
3.3.2 Measurement Settings
Measurement Settings includes Meas Setup, EFW Formula, HR Cycles and GA
Referance (see Figure 3-7).
Operation procedures:
1. Move the cursor to the (▼) and press the 【Set】 key to choose the default
measurement method.
2. After choosing, move the cursor to the button to save setting.
3. Repeat steps 1-2 if you want to amend other items. Press the button or
【Setup】key if you want to exit measurement menu.
Table 3-1 gives the conventional measurement items and its measurement methods.
Table 3-2 lists the formula for calculating fetal weight and gestational age reference.
Figure 3-7 the Menu of Measurement Settings
Portable Ultrasound Diagnostic Scanner User’s Manual
30
Table 3-1 General measurement setting
Measurement item Measurement method
Default Area/Circ Method Ellipse, Trace
Default Distance Method Line, Trace
Default Distance Unit cm, mm
Default Weight Unit Gram, Ounce, Pound
Table 3-2 EFW Formula and GA Referance
EFW and gestational age
measurement item Measurement formula
EFW Formula Tokyo, Osaka, Hadlock1, Hadlock2, Hadlock3,
Hadlock4, Shepard, Merz1, Hansmann, Campbell
GS Tokyo, Hellman, Rempen, Hansmann, China
CRL Tokyo, Hadlock, Nelson, Robinson, Rempen,
Hansmann, Osaka, China
BPD Tokyo, Hadlock, Jeanty, Kurtz, Merz, Hansmann,
Rempen, Chitty(0-I), Osaka, China
FL Tokyo, Hadlock, Jeanty, Hohler, Merz, Hansmann,
Chitty, Osaka, China
HUM Jeanty, Merz
HC Hadlock, Merz, Hansmann, Chitty(pltd)
AC Hadlock, Merz
FTA Osaka
Chapter 3 Instrument Operations
31
FT Mercer
OFD Hansmann
CER Hill
3.3.3 Setting Image Processing
Image Processing Settings can preset IP values. The items include Dynamic Range
(30-150dB, with a step of 6dB), Frame correlation (0-7 adjustable), M Scan Speed (0-3
adjustable), Edge Enhance (0-6 adjustable), Line Density (0 -2), witch is shown in Figure
3-8.
Figure 3-8 Image Processing Settings
The detail is as follows:
Press button, then you can input IP name, dynamic range, frame correlation,
M scan speed, edge enhance and line density. Press button to save a new
set of IP.
Portable Ultrasound Diagnostic Scanner User’s Manual
32
Select any existing IP in the combo box, and press button, then input the
new name, and press button to save the setting.
Select any existing IP in the combo box, and press button to delete a set of
existing IP value.
Select any existing IP in the combo box, and press button to apply a set of
existing IP value. At this point, both the image area of information and real-time
images will change.
Press button or【Setup】key to exit System Setup.
3.3.4 Version Information
To see the current software and Firmware version.
3.4 Selecting Probe and Frequency
3.4.1 Selecting Probe
Two ports and names of corresponding probes info is shown in the probe info area. The
current port is highlight, and its work frequency is shown.
Press 【Probe】to switch probes connected to the system. When first pressing it, switch
the current port to another port. Every time pressing it again, switch the current port to
another port. If another port doesn’t have a probe, it will not switch.
When the system is connected to two probes, select probe port activated when start up by
system setting. Please refer to 3.3.1.4 section.
3.4.2 Selecting Frequency
Operating procedures is as follows:
1. Under real-time status, press the 【Menu】key and the main menu will be displayed in
nether left of the screen.
2. Move the cursor or rotate【Gain】knob to select the PROBE FREQ menu item.
Chapter 3 Instrument Operations
33
3. Press 【Set】key or 【Gain】knob to choose this menu.
4. Rotate【Gain】 knob to change probe frequency, and press 【Set】 key or 【Gain】
knob to finish it.
5. Press Exit item to exit the main menu.
Figure 3-9 the main menu of B/M mode
3.5 Selecting Examination Type
Press 【Exam】key to enter exam menu and select examination type. The examination
types of the system include Abdomen, Gynecology, Obstetrical, Small Part, Emergency,
Orthopaedy and Urology. After choosing, the name of the current selected examination
type will be displayed on the main screen.
Portable Ultrasound Diagnostic Scanner User’s Manual
34
Figure 3-10 the menu of the exam
3.6 New Patients
Press 【New Patient】key to enter new patients menu. Input name, sex, age and other
personal information and automatically generate ID. Patient’s name is the required
information.
Chapter 3 Instrument Operations
35
Figure 3-11 New Patient
3.7 Switch Image Mode
The system provides five scanning modes: B, 2B, 4B, B/M and M. Press the image-mode
keys to switch over under real-time status.
In the following, the convex probe will be taken as an example to describe the individual
mode.
3.7.1 B Mode
Press 【B】key to activate B Mode, and the image area will display B Mode image, default
mode of the system.
Portable Ultrasound Diagnostic Scanner User’s Manual
36
Figure 3-12 B Mode
3.7.2 2B Mode
Press 【2B】key to switch to 2B image mode.
The operating procedure is as follows:
1. Press the 【2B】key, and a B-mode image appears on the left of the screen (See
Figure 3-13(a)).
2. Press the【2B】key again to activate the right of the screen, and the screen displays
two B-mode images in parallel. Where, the left one is a frozen image, and the right
one is a real-time image. “L” symbol appears on the upper part of a real-time image
(See Figure 3-13(b)).
3. You can continuously press the【2B】key to switch between these two images.
4. Under real-time status, press【Depth Up】or【Depth Down】key to adjust the depth
of real-time images;
Chapter 3 Instrument Operations
37
5. Press the【FREEZE】key to freeze an image, and press the【FREEZE】key again to
unfreeze the image;
6. In real-time mode, press the mode switching key in a mode instead of the 2B mode
can switch to another mode.
(a)
(b)
Figure 3-13 2B Mode
Portable Ultrasound Diagnostic Scanner User’s Manual
38
3.7.3 4B Mode
Under real-time status, press the 【4B】key to switch to 4B-mode images.
Figure 3-14 4B Mode
Operating procedure:
1. Press the 【4B】mode key continuously to display four B-mode images (as shown in
Figure 3-14). Only one image is in real time at the same time.
2. Press the 【4B】 key for four times, the screen will display four B-mode images in
parallel, including one real-time image, and other three frozen images. “L” symbol
appears on the upper part of the real-time image.
3. In the 4B mode, press the【4B】key repeatedly to switch over between the four images.
4. Press the 【FREEZE】 key to freeze an image, and press the【FREEZE】key again
to unfreeze the image.
Chapter 3 Instrument Operations
39
5. In real-time mode, press the mode switching key in a mode instead of the 4B mode
can switch to another mode.
3.7.4 B/M Mode
Under real-time status, press the 【B/M】key to switch to BM-mode images.
Operating procedure:
1. Pressing the 【B/M】key continually, up B down M, left B right M and single M modes
will appear(shown as figure 3-15). A dotted line (i.e., sampling line) appears in the
B-mode imaging area. You may use the trackball to move the sampling line; the
M-mode displays the movement of the tissue on the sampling line in this position.
2 Press the 【FREEZE】 key to freeze the B-mode and M-mode images at the same
time, and press the 【FREEZE】key once again to unfreeze the B-mode and M-mode
images.
3 The system provides four scanning speeds (0-3). Press 【Menu】key to enter the main
menu and setting it in the M Mode scanning speed menu (Refer to 3.8.13 section).
4 In the real-time status, press the mode switching key instead of the B/M mode can
switch to another mode.
Portable Ultrasound Diagnostic Scanner User’s Manual
40
(a) Up B Down M Mode
(b) Left B Right M Mode
Chapter 3 Instrument Operations
41
Rotate the brightness dial to appropriate brightness. Too bright result in weak
contrast, and too dark result in misty image.
Rotate the contrast dial to appropriate contrast. For clear profile, the contrast
shall be bigger; but for features of objects ,it shall be small.
(c) Full M Mode
Figure 3-15 B/M Mode
3.8 Adjusting Image Parameter
3.8.1 Adjusting image brightness and contrast
3.8.1.1 Adjusting image brightness
3.8.1.2 Adjusting image contrast
Portable Ultrasound Diagnostic Scanner User’s Manual
42
3.8.2 Gain Adjusting
3.8.2.1 Adjusting the Total Gain
Adjust the 【Gain】 knob to control the total gain of the entire image to obtain satisfying
images. Clockwise rotation may increase the gain, and anticlockwise rotation may
decrease the gain. You cannot adjust gain in freezing status.
The total gain ranges from 0 dB to 100 dB, with the minimum adjustable step of 5. The
gain value directly appears on the upper part of the screen. When checking 2B or 4B
Mode with many images, it will only display the value of activated image.
3.8.2.2 Adjusting the TGC
To compensate for weak signals or over-bright signals at various depths, you can adjust
Time Gain Compensation (TGC) using the TGC slide controls. The TGC provides six slide
controls for adjusting image quality under real-time status.
Use the system presets to turn on the display of the TGC curve. The curve provides an
on-screen representation of the TGC settings.(Refer to 3.3.1.3 section)
In B/M-mode, TGC adjustments affect both the B-mode image and the M-mode sweep.
The TGC slide controls have no effect during freezing.
To adjust the TGC slide controls: Move a slide control to the right to increase a setting or
to the left to decrease a setting.
3.8.3 Changing the Dynamic Range
You can adjust the overall contrast resolution of a B-mode image and the overall contrast
resolution of an M-mode sweep. For B-mode and M-mode, the dynamic range values are
30 dB to 150 dB with a step of 6dB. The dynamic range value displays in the imaging
parameters area at the top of the screen. In 2B mode or 4B mode with more than one
image, the value displays only for the active image.
Operating procedure:
Chapter 3 Instrument Operations
43
1. Under real-time status, press the 【Menu】key, and the main menu appears on the
lower part of the screen.
2. Move the track ball to place the cursor on the option DYNAMIC RANGE; Press the
【Set】 key to set parameter values or use the 【Gain】 knob to adjust parameters,
and press 【Gain】 knob to choose it.
3. Adjust parameter by 【Gain】 knob.
4. Press 【Menu】key and it will disappear. Directly exit the adjustment status or move
the cursor to the EXIT option and then press 【Set】 to exit.
3.8.4 Frame Correlation
The frame correlation adjustment involves superposing and averaging B-mode images in
neighboring frames to eliminate noise from the images and produce clearer image details.
The frame correlation range is from 0 to 7, totaling 8 levels. 0 means no frame correlation,
while 7 means the largest frame correlation. The current frame correlation value is also
displayed on the menu item.
Operating procedure:
1. Under real-time status, press the 【Menu】key, and the main menu appears on the
lower part of the screen.
2 Move the track ball to place the cursor on the option FRAME CORR. Press the 【Set】
key or press 【Gain】 knob to choose it.
3 Adjust parameter by rotating【Gain】 knob.
4 Press 【Menu】key or move the cursor to the EXIT option and then press 【Set】to
close menu.
3.8.5 Edge Enhancement and Image Smoothing
Edge enhancement highlights image profile so that the border of tissue structure of an
image can be displayed more clearly, while image smoothing has the reverse effect.
Portable Ultrasound Diagnostic Scanner User’s Manual
44
When adjusting, the number displayed on the side of the menu item is the edge
enhancement strength, which is shown as Table 3-3.
Table 3-3 Edge Enhancement and Image Smoothing
EE6 strong
EE5 medium
EE4 light Edge Enhancement
EE3 close
EE3 close
EE2 light
EE1 medium Image Smoothing
EE0 strong
The detail refers to 3.8.3 section.
3.8.6 Gray Scale
3.8.6.1 Setting Gray Map
The system supports 16 kinds of Gray Map (0-15 adjustable). It does not work when
choose OFF option.
The detail refers to 3.8.3 section.
3.8.6.2 Setting Gray Reject
The system supports 16 kinds of Gray Reject (1-16 adjustable). It does not work when
choose OFF option). The detail refers to 3.8.3 section.
3.8.7 Changing the Density
When in B Mode real-time status, press number key【3】can adjust the density, that is
consequently displaying high density, middle density and low density image. Increasing
density could improve image detail, but decrease frame rate.
Density sign differentiates high, middle and low density. H means high density, M means
Chapter 3 Instrument Operations
45
middle density, and L means low density.
3.8.8 Adjusting Focus
The system includes 64 groups of focus combinations. Each focus combination comprises
different positions and number of focuses. The focus control keys(【Focus Up】or 【Focus
Down】) and numeric keys(【5】or 【6】)can be used to adjust image focus. Under B mode
real-time status, focus adjustment brings clearer images in the target area. In B/M mode
and single M mode, the system does not support multi-focus.
Increasing the number of focus can gain higher resolution in different depth, at the same
time; the number of focus can affect frame rate of the image: the more focus ,the lower
frame rate.
For visual display, the focus display mode on the screen is displayed on the left or right of
the vertical scale line by the or symbol.
Operating procedure:
Under real-time status, press numeric keys【5】or【6】to decrease or increase the
number of focus. The range is from 1 to 4. When overstep this step, pressing 【5】
or【6】 will not change the number. B/M Mode has only one focus.
Under real-time status, press【Focus up】and【Focus down】can adjust the position
of the focus. Every kind of focus number can have 16 kinds of focus combination.
When the focus is in the highest or lowest position, press this key again, the position
of the focus will keep the same.
3.8.9 Adjusting imaging depth
Increase or decrease imaging depth by【Depth Up】and【Depth Down】.
3.8.10 Adjusting image width
The system provides images with four steps width.
Operating procedure:
Portable Ultrasound Diagnostic Scanner User’s Manual
46
1. Under real-time status, press【Menu】to display the main menu.
2. Move the trackball or turn【Gain】to place the cursor in the IMAGE WIDTH item.
3. Press【Set】key or【Gain】key to choose this menu item.
4. Image narrowing, broaden, restore default and exit buttons will be displayed in the
bottom of the screen.
5. Press numeric keys 【3】or 【4】to make images narrowing or broadening. Press
numeric key【5】to restore image width.
6. Press numeric key 【6】to exit this adjusting mode.
3.8.11 Image Flip
Figure 3-16 Image Flip
Image flip can be adjusted on the menu, including U/D flip, L/R flip, and B/W flip.
Chapter 3 Instrument Operations
47
Operating procedure:
Under real-time status (except single M mode ), press numeric key 【1】, then the
image will flip 180° along vertical direction.
In B Mode real-time status, press numeric key 【2】, then the image will flip 180° along
level direction.
Under real-time status, press 【Menu】key to choose B/W FLIP item to achieve
black-white inversion of the image. The detail refers to 3.8.3 section.
3.8.12 Image Zoom
Single B mode images under real-time status have zoom function. Measurement,
annotation, body-mark are available in the zoomed image.
Operating procedure:
1. Pressing【Zoom】key or pressing【4】key, the zoom frame will appear.
2. Rotate 【Gain】knob to adjust the size of the zoom frame (two sizes available) so as
to magnify the image twice or three times.
3. Move the trackball to place zoom frame in the concerned area of the image.
4. Press 【Zoom】key or press【4】key to zoom selected area.
5. Press 【Zoom】key or press【4】key to exit zoom mode.
3.8.13 Changing the M Sweep Speed
The system provides four kinds of scanning speed: 0-3. 0 expresses the lowest speed,
while 3 expresses the fastest speed. The detail refers to 3.8.3 section.
3.8.14 Harmonic Imaging
The function of Harmonic Imaging is available for low-frequency probe (The system
supports the probe of 35C60H.). The detail refers to 3.8.3 section.
Portable Ultrasound Diagnostic Scanner User’s Manual
48
3.8.15 Changing Power
The power ranges from 0 to 100, with step of 5. The detail refers to 3.8.3 section.
3.9 Annotation and Arrows
3.9.1 Annotation
Press【Text】key to active annotation function.
Operating procedure:
1. Move the cursor to the wanted position, and then press 【Set】to confirm input
position.
2. After inputting needed text, press 【Enter】key to confirm it.
3. Repeat step 1-2 if you want to input other characters continuously.
4. Press 【Text】key again to exit the annotation function, and the annotations are still
displayed on the images.
5. Press 【Clear】key to clear all annotations and exit the text function.
Chapter 3 Instrument Operations
49
Figure 3-17 Annotation Function
3.9.2 Arrows
Press【Arrow】key to activate arrows. A solid arrow is displayed in the image area ,which
can be rotated 360° along eight directions.
Operating procedure:
1. Turn 【Gain】knob to change arrow direction.
2. Move the trackball to adjust the position of the arrow.
3. Press 【Set】key to confirm entering an arrow.
4. Repeat step 1-3 if you want to continue.
5. Press 【Arrow】again to close arrow function, and the arrows are still displayed on the
images.
6. Press 【Clear】key to clear all the arrows and close the arrow function.
Figure 3-18 Arrow Function
Portable Ultrasound Diagnostic Scanner User’s Manual
50
3.10 Body Markers
Markers are graphics that display on-screen to indicate the anatomical structure under
evaluation and to indicate the orientation of the probe to the structure. Every examination
subject will have fixed body markers.
Operating procedure:
1. Press 【Mark】key to activate body markers function. Corresponding body markers
will be displayed on the bottom of the screen.
2. Select body marker using numeric keys 【1~8】. If there are more than eight body
markers, pressing numeric key 【9】can turn to next page, while 【0】 to the previews
page. A small pictograph like probe will be displayed on the screen.
(a)
Chapter 3 Instrument Operations
51
(b)
Figure 3-19 Body Markers
3. Rotate 【Gain】knob to adjust probe pictograph’s direction, and press 【Set】 key to
confirm it.
4. Press【Clear】key to clear body markers. Repeat 1-3 to reselect body marker.
3.11 Biopsy Guideline
Caution: Needle biopsy is highly dangerous for both patients and operating
doctors. The use of puncture rack and its accessories shall strictly follow the correct
needling sequence to avoid unnecessary risks to patients.
Caution: The biopsy guideline displayed on the equipment monitor is for reference
only. During biopsy, aspiration, puncture and related operations, the user shall
confirm the correct needling position.
Caution: Biopsy, aspiration and puncture are intervention operations with high
risks. Operators shall strictly stick to the principle of operation of sterilization.
The biopsy guideline is used to help the operator to move the probe to an appropriate
Portable Ultrasound Diagnostic Scanner User’s Manual
52
position during puncture operation so that the end of the probe with the puncture rack is
located in the ideal needling position. The biopsy guideline can only be used under
real-time status.
Operating procedure:
1. In single B mode real-time status, press the【Biopsy】key. The screen displays a
biopsy guideline.
2. Press numeric keys【1】and【2】to adjust offset of the biopsy guideline.
3. Press numeric keys【3】and【4】to adjust angle of the biopsy guideline.
4. Press numeric key【5】to restore the default settings.
5. Press numeric key【6】to display the depth of the biopsy guideline.
6. Press the【Biopsy】key again to exit biopsy function.
Figure 3-20 Biopsy Guideline
Chapter 3 Instrument Operations
53
3.12 CINE
During B mode real-time imaging, the system places the most recently acquired images
as well as image currently on screen into a CINE memory buffer. This memory buffer can
store 512 frames. Film indicator will be displayed on right bottom of the image, which tells
us current frame and saved frame number.
Operating procedure:
1. Press 【Cine】key or 【FREEZE】key to enter CINE mode under real-time status.
2. Rotate 【Gain】knob to play forward or backward manually.
3. Press 【Gain】key to play automatically, and press 【Gain】key again to stop.
4. Press 【FREEZE】key again to exit CINE mode, and enter real-time image status.
Figure 3-21 CINE
Portable Ultrasound Diagnostic Scanner User’s Manual
54
3.13 Image Managing and Printing
3.13.1 Image Managing
3.13.1.1 Image Storage
In the status of image freezing, press 【Save】key to store images.
3.13.1.2 Image Browsing and Exporting
Press 【Review】key to enter image browsing window.
Figure 3-22 Image Reviewing
Picture layout, 1×1,2×2 two formats available
Zoom the selected image ( Double-clicking the selected image can
also achieve full-screen zoom function .). Press【Exit】key to exit this
mode.
Chapter 3 Instrument Operations
55
View the previous image
View the next image
Delete all images
Delete the selected image
Print the selected image
Export all images
Export the selected image
Caution: You’d better select optional 512M Netac U265 or 1G Netac U208 provided
by Landwind. The system only supports FAT32 file-system. Before using your U-Disk,
please format it in your computer. After inserting an U-Disk, mark will be
displayed on the upper right of the screen, then you can store images in the U-Disk.
After storing, move the cursor to this USB mark, and press 【Set】key, then the U-Disk
information and operating dialog box will appear, click button and after a
tip box appeared, extract the U-Disk. System update is only used by the engineer
authorized by Landwind. Do not operate it casually, or accept the consequences.
Portable Ultrasound Diagnostic Scanner User’s Manual
56
3.13.2 Printing
3.13.2.1 Printing Using a Laser Printer
Portable Ultrasound Diagnostic Scanner supports HP LaserJet P2015 printer.
Laser printer connection
Connect the data line of laser printer to the USB port of the system.
Printing Images and Reports
Open【Review】window, choose an image an click icon to print the
selected image.
Open【Report】menu, and click button to print a report.
3.13.2.2 Printing Using a Video Printer
Video printer connection
Connect one end of the video printer to the video output port of the system, and
the other end to the remote port of the system.
Printing an Image
We recommend only printing an image when freezing.
To print an image, please press the【Print】key.
57
Chapter 4 Measurements and Reports
4.1 Overview of measurement
The system provides the operation methods with general measurement (B-mode general
measurement, M-mode general measurement), and special measurement (abdominal
measurement, obstetrical measurement, orthopedics measurement, urology
measurement and small parts measurement).
4.1.1 Special Measurement Menu
The system’s special measurement is equipped with a special measuring menu. Take the
obstetrics measurement menu as an example to introduce the use of the special
measurement menu.
1.Title 2.Measurements(which have sub-menu)
3. Measurements 4.Exit the menu
5.Return the main menu
Figure 4-1 Special Measurement Menu-Obstetrical
In special measurement mode, press【2】key to enter obstetrics measurement menu, as
shown in Figure 4-1. Move cursorl and move the cursor to a measurement item, then
press【Set】key. If the there is logo on the item, you can enter the sub-menu; if there’s
not, you can enter the appropriate measurement mode. Choose RETURN to return the
Portable Ultrasound Diagnostic Scanner User’s Manual
58
main menu of obstetrics. Choose EXIT to exit the main menu of obstetrics.
4.1.2 Measurement Results
Measuring results are displayed on the left side of the screen. If there are more measuring
information, the arrow will appear to carry out instructions. In measurement condition, you
can use 【Gain】knob to turn the page up and down to view the measuring results, as
shown in Figure 4-2.
Figure 4-2 the displayed screen of the measuring results
4.1.3 Measurement Report
The measurement report is used to record the measurement results. A measurement
report contains the report name, patient information, measurement and calculation results,
doctor diagnosis and more. The user can edit and print measurement reports. This system
provides general measurement report and exam-specific measurement report.
The exam-specific measurement reports include gynecological report, obstetrical report,
orthopedic report, urological report and small report.
Chapter 4 Measurements and Reports
59
General measurement report:
Figure 4-3 General measurement report
Operating procedure:
Click and buttons to select the appropriate
diagnostic image.
Input the diagnostic results in the column of DIAGNOSE.
Click button to print report.
Click button or press【Report】key to exit the report mode.
Portable Ultrasound Diagnostic Scanner User’s Manual
60
4.2 General measurement — B-mode
The B-mode includes distance measurement, area/perimeter measurement, volume
measurement, angle measurement and histogram function.
Under real-time status, press 【Meas】key into the general measurement, or press
【FREEZE】key, and then press the numeric key 【5】 to enter the general measurement,
as shown in Figure 4-4.
Figure 4-4 B-mode General Measurement
4.2.1 Distance Measurement
Distance measurement is representative of the measurement on B-mode.
Press the numeric key【1】to enter the distance measurement mode. repeatedly pressing
【1】key can switch over between two measurement modes: line and trace.
4.2.1.1 DIST-LINE Measurement
Function: Measure the distance between two points in Ultrasound image.
Chapter 4 Measurements and Reports
61
Figure 4-5 DIST-LINE Measurement
Measurement:
1. Repeatedly press 【1】key to switch over the DIST-LINE measurement, at this
time a “+” cursor appears on the screen;
2. Use the trackball to move the “+” cursor to the starting point of the
measurement. press 【Set】key to determine the starting point;
3. Use the trackball to move the cursor, at this time between the starting point
and the cursor will always be a dotted line connection, the screen shows
real-time measurement result on the left;
4. Use the trackball to move the cursor to the end of the measurement.(Press
【Exit】key to switch the endpoints of the segments ) Press 【Set】 key, and
the measurement result will be finally confirmed, which means the
measurement finishes;
5. Repeat steps 2 to 4 to continue a next beeline measurement. The screen can
display a maximum of four measurement results. If it does beyond four,
Portable Ultrasound Diagnostic Scanner User’s Manual
62
subsequent measurement results will automatically replace the previous ones.
6. Press【Clear】key to clear the measurement data and exit this mode.
Caution: An image can be measured four times at maximum, with different cursor
forms for each measurement.
4.2.1.2 DIST-TRACE Measurement
Function: Calculate the distance manually with the trackball.
Figure 4-6 DIST-TRACE Measurement
Measurement:
1. Repeatedly press 【1】key to switch over the DIST-TRACE Measurement, at this
time the “+” cursor appears on the screen;
2. Use the trackball to move the cursor to the starting point of the measurement,
and press【Set】key;
3. Use the trackball to make the cursor move along the edge of the target mobile,
outline the trace to determine the end point and then press【Set】key, at this
time real-time results will display on the left of the screen;
Chapter 4 Measurements and Reports
63
4. Repeat steps 2 to 3 to continue the next track measurement;
5. Press 【Clear】key to clear the measurement data and exit this mode.
4.2.2 Area/CIRC Measurement
4.2.2.1 Ellipse method
Function: Measure the circumference and area of a closed area by using ellipse approach
Operating procedures:
Figure 4-7 Ellipse Measurement
Measurement:
1. Repeatedly press【2】key to switch over the Ellipse method measurement
mode, at this time a “+” cursor appears on the screen;
2. Move cursorl and choose the starting point of the long axis. Press 【Set】key
to confirm the starting point;
3. Move cursorl to move the cursor to the end of the long axis (Press【Exit】key
to switch the endpoints of the elliptical axes), and then press 【Set】key to
Portable Ultrasound Diagnostic Scanner User’s Manual
64
confirm the end;
4. Move cursorl and choose the start-stop point of the short axis. Press 【Set】
key to confirm the starting and end point, at this time, the result shows on the
left of the screen ;
5. Repeat steps 3 to 4 to continue the Ellipse measurement, at most 4 groups;
6. Press 【Clear】key to clear the measurement data and exit this mode.
4.2.2.2 Trace method
Function: The trace measurement is used to measure the area and circumference
manually with the trackball.
Figure 4-8 Trace Measurement
Measurement:
1. Repeatedly press【2】key to switch over the trace measurement mode, at this
time a “+”cursor appears on the screen;
Chapter 4 Measurements and Reports
65
2. Move the trackball to choose the starting point of the track, and press【Set】
key;
3. Move the trackball to choose the end point of the track, and press【Set】key, at
this time, the system will automatically connect the start and the end of the
track in a circle, at the same time, on the left of the screen will display the
results;
4. Repeat steps 2 to 3 to continue the trace measurement, at most 4 groups;
5. Press【Clear】key to clear measurement data and exit this mode.
4.2.3 Volume Measurement
4.2.3.1 3-LINE
Function: Calculate the volume with three distance measurements, volume measurement
need two images: longitudinal and transverse.
Figure 4-9 3-LINE Measurement
Measurement:
Portable Ultrasound Diagnostic Scanner User’s Manual
66
1. Repeatedly press 【3】key to switch over 3-line measurement mode, at this
time a “+”cursor appears on the screen;
2. Use Section 4.2.1.1 DIST-LINE measurement, in a different plane to measure
the length of the segments, at this time, the actual measurement value of the
volume will be displayed on the left of the screen ;
3. Repeat step 2 to continue the 3-line measurement, at most 4 groups;
4. Press【Clear】key to clear measurement data and exit this mode.
4.2.3.2 1-LINE
Function: Calculate volume with one distance measurement.
Figure 4-10 1-LINE Measurement
Measurement:
1. Repeatedly press 【3】key to switch over 1-line measurement mode, at this
time a “+”cursor appears on the screen;
2. Use Section 4.2.1.1 DIST-LINE measurement to measure the length of a
segment , at this time, the actual measurement value of the volume will be
Chapter 4 Measurements and Reports
67
displayed on the left of the screen ;
3. Repeat step 2 to continue one distance measurement, at most 4 groups;
4. Press【Clear】key to clear measurement data and exit this mode.
4.2.3.3 LINE+ELLIPSE
Measurement:
1. Repeatedly press 【3】key to switch over LINE + ELLIPSE measurement mode,
at this time a “+”cursor appears on the screen;
2. Using section 4.2.1.1 DIST-LINE measurement to draw a line;
3. Then using section 4.2.2.1 Ellipse measurement to determine an oval; at this
time, the actual measurement value of the volume will display on the left of the
screen;
4. Repeat steps 2 to 3 to continue Distance + Ellipse measurement, at most 4
groups;
5. Press【Clear】key to clear measurement data and exit this mode.
Figure 4-11 LINE+ELLIPSE Measurement
Portable Ultrasound Diagnostic Scanner User’s Manual
68
4.2.3.4 2-ELLIIPSE
Measurement:
1. Repeatedly press 【3】key to switch over Ellipse + Ellipse measurement mode,
mode, at this time a “+”cursor appears on the screen;
2. Using section 4.2.2.1 Ellipse measurement to draw two ellipses in different
planes, at this time, the actual measurement value of the volume will display
on the left of the screen;
3. Repeat step 2 to continue Ellipse + Ellipse measurement, at most 4 groups;
4. Press【Clear】key to clear measurement data and exit this mode.
Figure 4-12 2-ELLIPSE Measurement
4.2.4 Angle Measurement
Function: Measure the inclination of two lines in a plane in the ultrasonic image.
Measurement:
1. Press 【1】key to switch over the angle measurement mode, at this time a “+”
Chapter 4 Measurements and Reports
69
cursor appears on the screen;
2. Use the trackball to move the cursor to the starting point of the measurement,
and press【Set】key;
3. Move cursor to move the cursor to the end point. Press【Set】key again to
confirm the end point. By now, a segment will appear;
4. Move cursor to choose another segment. Press【Set】key to confirm the
starting and end point , at this time, the real-time measurement result of the
angle between the two segments displays on the left of the screen;
5. Repeat steps 2 to 4 to continue the next angle measurement;
6. Press 【Clear】key to clear the measurement data and exit this mode.
Figure 4-13 Angle Measurement
4.2.5 Histogram Measurement
Function: Calculate the gray distribution of ultrasonic echo signals in the specified region.
Measure the interest region by using rectangle and show the results by histogram.
Portable Ultrasound Diagnostic Scanner User’s Manual
70
Caution: Histogram Measurement cannot used in B/M mode.
Measurement:
1. Press the number key【5】into the histogram measurement mode, at this time
a “+” cursor appear in the center of the screen;
2. Move the trackball and press 【Set】key to determine an acme of a rectangle;
3. Move the trackball , and the screen will display a rectangular box of samples;
4. Press【Set】to determine the vertex of the rectangle diagonally, and determine
the histogram measurements region. The histogram results show in the
middle of the screen;
5. Repeat steps 2 to 4 to continue histogram measurement;
6. Press【Clear】key to clear measurement data and exit this mode.
Figure 4-14 Histogram Measurement
The significance of the measurement result values as follows:
total: the sum of the data in the measuring region;
Chapter 4 Measurements and Reports
71
mean: average;
most: the maximum value of the measuring area;
ratio: the maximum percentage of the measuring region;
sd: standard deviation.
4.3 General measurement — M-mode
M-mode general measurement includes distance, time, slope and heart rate
measurements in B/M mode or single M mode real-time state. First press 【Meas】key,
then press number key 【4】 into M mode general measurement, or first press【FREEZE】
key, and then press numeric keys 【5】and【4】into the M mode general measurement.
Figure 4-15 M-mode General Measurement
4.3.1 Distance
Function: Measure the vertical distance between the two points in M mode image.
Measurement:
1. Repeatedly press【4】key to switch over distance measurement mode, at this
Portable Ultrasound Diagnostic Scanner User’s Manual
72
time a “-” cursor appears on the screen;
2. Using the trackball to move the “-” cursor to the starting point of the
measurement, and press【Set】key, then the starting point for measuring
shows the baseline and the short horizontal line marker “-”, at this time, the
cursor can only move up and down, and the measurement results in real-time
display on the left of the screen;
3. Move the cursor up and down to the end of measurement, press 【Set】key,
and the final result shows on the left of the screen. The system automatically
enters the next measurement;
4. Repeat steps 2 to 3 to continue distance measurement, at most 4 groups.
5. Press【Clear】key to clear measurement data and exit this mode.
Figure 4-16 Distance Measurement in M mode
4.3.2 Time
Function: Measure the time interval between two points of the image in M mode.
Measurement:
Chapter 4 Measurements and Reports
73
1. Repeatedly press【4】key to switch over the time measurement mode, at this
time a vertical long-dashed line appeared on the screen;
2. Move the long-dashed line with Trackball to the starting point of the
measurement, press【Set】key, and the starting point for measuring displays a
vertical line. The cursor can only move horizontally to determine the starting
measurement, at the same time, the second long-dashed line appears on the
screen;
3. Move the cursor with Trackball, and the real-time measurement result will be
displayed on the bottom of the screen. Move the cursor to the end of the
measurement, and press【Set】key, then the final measurement results show
on the left of the screen. The system automatically enter the next
measurement after this measurement;
4. Repeat steps 2 to 3 to continue the time measurement, at most 4 groups;
5. Press【Clear】key to clear measurement data and exit this mode.
Figure 4-17 Time Measurement in M mode
Portable Ultrasound Diagnostic Scanner User’s Manual
74
4.3.3 Slope
Function: Calculate the slope between two points by measuring their time and distance in
M mode.
Measurement: Please refer to section 4.2.1.1 DIST-LINE measurement.
Figure 4-18 Slope Measurement in M mode
4.3.4 Heart rate
Function: Measure the time interval between N times HR Cycle in M mode, and calculate
the number of heartbeats per minute.
Measurement:
1. Press【5】key into the heart rate measurement mode, at this time a long –dashed
broken line appears on the screen;
2. Use the trackball to move the long-dashed broken line to the starting point of the
measurement, press 【Set】key, where will display measurement base line and another
long-dashed line. At this time, the cursor can only be moved left and right, and the
measurement result will appear on the left of the screen real-time.
Chapter 4 Measurements and Reports
75
3. Move the cursor left and right to the end point of the measurement, press【Set】key,
and the final measurement result will appear on the left of the screen. After the
measurement, the system will enter next measurement automatically.
4. Press 【Clear】key to clear all measurement data.
5. Repeat steps 2-3 to measure heart rate continually.
Figure 4-19 Heart Rate Measurement in M mode
4.4 Gynecology Measurements
This section describes the system of the Gynecological measurement functions, including
Uterus (length, width, depth, endometrium), cervix (length, width, depth), the left ovary
(length, width, depth), the right ovary (length, width, depth), left-follicle (left-follicle1,
left-follicle2, left-follicle3, left-follicle4), right-follicle (right-follicle1, right-follicle2,
right-follicle3, right-follicle4).
Gynecology measurement only works in the B mode; there are two methods to enter the
Gynecology measurement, as follows:
Portable Ultrasound Diagnostic Scanner User’s Manual
76
In real-time state, press【Calc】key to enter the special measurement, and then
press 【1】key to open the menu of gynecology measurement;
In real-time state, press 【FREEZE】 and【 6】 key to enter the special
measurement, and then press【1】key to open the menu of gynecology
measurement.
Figure 4-20 the main menu of Gynecology Measurement
4.4.1 Uterus Measurement
It is to measure uterus length, width and depth(The system automatically calculates the
size of the uterus)and endometrial thickness. Their measurement methods are distance
measurement in B mode (please refer to Section 4.2.1).
Figure 4-21 the menu of uterus
4.4.2 Cervix Measurement
It is to measure cervix length, width and depth, and then the system automatically
calculates the size of the cervix. The measurement methods are distance measurement in
B mode (please refer to the Section 4.2.1).
Chapter 4 Measurements and Reports
77
Figure 4-22 the menu of cervix
4.4.3 Ovary Measurement
It is to measure Ovary length, width and depth, and then the system automatically
calculates the volume of the ovary, The measurement methods are distance
measurement in B mode (please refer to the Section 4.2.1).
Figure 4-23 the menu of ovary
4.4.4 Follicle Measurement
Ovary can be measured within four follicles (follicle1, follicle2, follicle3, follicle4). After
measuring the diameter of the follicle, the system can automatically calculate the volume
of the follicle, The measurement methods are distance measurement in B mode (please
refer to Section 4.2.1).
Figure 4-24 the menu of follicle
Portable Ultrasound Diagnostic Scanner User’s Manual
78
4.4.5 Gynecologic Report
In the gynecological measurement mode, press 【Report】 on the control panel to enter
the GYN report, as shown in Figure 4-25.
Refer to Section 4.1.3 General Measurement Report for details of the gynecological
measurement reports.
Figure 4-25 Gynecology Measurement
Chapter 4 Measurements and Reports
79
4.5 Obstetrics Measurements
This section describes the system of the obstetrics measurement function, including OB
measurements, AFI, FHR, NT, OB input, biophysical profile score and so on.
There are two methods to enter the obstetrics measurement, as follows:
In real-time state, press【Calc】key to enter the special measurement, and then
press 【2】key to open the menu of obstetrics measurement;
In real-time state, press 【FREEZE】 and【 6】 key to enter the special
measurement, and then press【 2】 key to open the menu of obstetrics
measurement.
Figure 4-26 the main menu of Obstetrics Measurement
4.5.1 Obstetrics Input
Obstetrics input items include the last menstrual period(calendar), the date of
ovulation(calendar), the number of pregnancies, the number of partition and the number
of abnormal pregnancies.
The data of delivery have been calculated from the last menstrual period and the date of
ovulation.
Operating procedure:
Click button to open the date choosing box (see Figure 4-28).
Click the triangle icons on the to choose year, and click the triangle icons
Portable Ultrasound Diagnostic Scanner User’s Manual
80
on the to choose month.
Move cursor and press 【Set】to choose day.
Click button to confirm the date and exit the box, and then the
dates of LMP or ovulate and their EDD are displayed automatically, as shown in
Figure 4-27.
Figure 4-27 Obstetrics Input Box
Figure 4-28 choosing date
4.5.2 Fetus Selection
This system can measure up to 6 fetuses. After typing the number of fetus in the obstetrics
input box, the serial numbers of fetus can be switch in the fetus changing box. Choose to
enter the measurement of the fetus, and the serial number of the fetus is displayed on the
left side of the screen. Fetus No.1 is the default measurement .
Chapter 4 Measurements and Reports
81
Figure 4-29 selecting fetus number
4.5.3 Obstetrics Measurements
Figure 4-30 the menu of Obstetrics Measurement
Click anyone of the obstetrics measurements and enter the appropriate measurement
mode. Then the system automatically calculates the gestational age, which displayed in
the left side of the screen. Table 4-1 gives the methods and application scopes of the
measurements.
Portable Ultrasound Diagnostic Scanner User’s Manual
82
Table 4-1 Obstetrics Measurements
Measurement Gestational age in table Method
GS Tokyo、Hellman、 Rempen、Hansmann、ChinaDistance
Measurement
CRL Tokyo、Hadlock、Nelson、Robinson、Rempen、
Hansmann、Osaka、China
Distance
Measurement
BPD Tokyo、Hadlock、Jeanty、Kurtz、Merz、
Hansmann、Rempen、Chitty(O-I)、 Osaka、
Distance
Measurement
FL Tokyo、Hadlock、Jeanty、Hohler、Merz、
Hansmann、Chitty、Osaka、China
Distance
Measurement
HUM Jeanty,Merz Distance
Measurement
HC Hadlock,Merz ,Hansmann,Chitty(pltd) Circumference
Measurement
AC Hadlock、Merz Circumference
Measurement
FTA Osaka Area
Measurement
FT Mercer Distance
Measurement
OFD Hansmann Distance
Measurement
CER Hill Distance
Measurement
TTD Fetal weight Distance
Measurement
Chapter 4 Measurements and Reports
83
APTD Fetal weight Distance
Measurement
For all the obstetrics measurements, if it is set up a fetus weight formula in measurement
settings and all the measurements corresponding to this formula has been measured, the
system will calculate the gestational age automatically. Table 4-2 gives every fetus weight
formula required for the measurement of parameters. Please set with the information from
Table 4-2.
Table 4-2 Obstetrics Formula
Formula Name Measurements
Tokyo BPD、APTD、TTD、FL
Osaka BPD、FL、FTA
Hadlock1 AC、FL
Hadlock2 AC、FL、BPD
Hadlock3 AC、FL、HC
Hadlock4 AC、BPD、FL、HC
Shepard AC、BPD
Merz1 AC、BPD
Hansmann BPD、TTD
Campbell AC
Taking the Osaka formula as an example, the operation steps are detailed below.
Portable Ultrasound Diagnostic Scanner User’s Manual
84
1. Press 【Setup】key to enter the menu of system settings, and move cursor to open
the menu of measurement settings.
2. Choose the Osaka formula in EFW Formula Box, and confirm the settings.
3. Enter the obstetrics measurement mode and measure the items of BPD, FL and
FTA.
4. Then system calculates the fetus weight automatically, and the results are displayed
on the left side of the screen, as shown in Figure 4-31.
Figure 4-31 the Fetal weight measurement with “Osaka” formula
4.5.4 Amniotic fluid index
In the 4B mode, after four times distance measurement (refer to Section 4.2.1), the
system automatically calculates the amniotic fluid index.
Chapter 4 Measurements and Reports
85
Figure 4-32 AFI
4.5.5 Fetal heart rate
Fetal heart rate can only be measured in the B/M or M mode. For the measurement
method, please refer to Section 4.3.4 heart rate.
4.5.6 Neck Transparency
The measurement method of neck transparency (NT) is distance measurement in B mode,
referring to Section 4.2.1.
4.5.7 Fetal physiology score
Fetal physiology score is on the basis of the fetal ultrasound images, experiment or
measurement methods to get the fetal development indicators, and then speculate the
state of the fetus by a certain standard score.
The system provides criteria, which is based on the “Vintzileos” formula, as shown in
Table 4-3. Table 4-4 gives the description of the total score.
Figure 4-33 Biophysical Profile Score
Portable Ultrasound Diagnostic Scanner User’s Manual
86
Table 4-3 Fetal Scoring Standard(Vintzileos Formula)
Fetal development
indicators 0 Score 1 Score 2 Score
Observation time
Remarks
Response
fetal heart
rate to
acceleration
FHR FHR
Increase ≤1
FHR increase
≥15/min,
duration≥15S,
1~4 times
FHR
increase≥15/min,
duration ≥15S,
≥5 times
20 minutes
Fetal
movement FM
No
movement 1~2 times ≥3 times 30 minutes
Fetal
breathing
movement
FBM No FBM, or
duration≤30S
FBM ≥1,
duration
30-60S
FBM≥1,
duration≥60S 30 minutes
Fetal tone FT
Limbs
Stretched,
without
bending,
hands open
≥1 times,
stretching
and bending
movement of
limbs of
vertebral
column
≥1 times,
stretching and
bending
movement of
limbs of vertebral
column
30 minutes
The scores
obtained
from fetal
response
experiment
can be
entered
into the
system
through
the
interface.
Placental
grading PL
III-level
placenta
Back wall of
placenta, the
placenta
difficult to
assess
classification
≤II-level placenta
According to the
audio-visual placenta to
the placenta mature
situations as 4 levels: level
0, I, II, III
The depth
of amniotic
fluid
AF AF<1cm 1cm≤AF≤2cm AF>2cm Obtained by measuring
B-mode images
Chapter 4 Measurements and Reports
87
Table 4-4 Description of the Total Score
Total Score Development condition
7-12 score Normal
3-6 score Equivocal
0-2 score Abnormal
4.5.8 Obstetrics Report
In the obstetrical measurement mode, press 【Report】 on the control panel to enter the
obstetrics report, as shown in Figure 4-34.
Refer to Section 4.1.3 General Measurement Report for details of the obstetrical
measurement reports.
Portable Ultrasound Diagnostic Scanner User’s Manual
88
Figure 4-34 Orthopedics Report
4.6 Orthopedics Measurements
This section describes the orthopedics measurement function of the system, including BL,
ARL and IL. They can only be measured in the B mode.
Chapter 4 Measurements and Reports
89
Figure 4-35 the main menu of Orthopedics Measurement
There are two methods to enter Orthopedics measurements, as follows:
In real-time state, press【Calc】key to enter the special measurement, and then
press 【3】key to open the menu of orthopedics measurement;
In real-time state, press 【FREEZE】 and【 6】 key to enter the special
measurement, and then press【3】key to open the menu of orthopedics
measurement.
BL, ARL and IL are three lines, by the identification of BL and ARL, their included angle α
can be measured; by the identification of BL and IL, their included angle β can be
measured. The measurements of α and β are displayed on the left of the screen. The
measurements of BL, ARL and IL refer to Section 4.2.4.
4.6.1 Orthopedics Report
In the orthopedic measurement mode, press 【Report】 on the control panel to enter the
ORTH report, as shown in Figure 4-36.
Refer to Section 4.1.3 General Measurement Report for details of the orthopedic
measurement reports.
Portable Ultrasound Diagnostic Scanner User’s Manual
90
Figure 4-36 Orthopedics Report
4.7 Urology Measurements
This section describes the system of the Urology measurement function, including
prostate volume (length, width, and depth), residual urine volume (length, width, depth)
and PSAD measurement. The prostate volume and residual urine volume can only be
measured on the B mode.
Chapter 4 Measurements and Reports
91
Figure 4-37 the main menu of Urology Measurement
There are two methods to enter Urology measurement, as follows:
In real-time state, press【Calc】key enter the special measurement, and then
press 【4】key to open the menu of Urology measurement;
In real-time state, press 【FREEZE】 and【 6】 key to enter the special
measurement, and then press 【 4 】 key to open the menu of Urology
measurement.
4.7.1 PSAD Measurement
The value of PSAD (serum prostate specific antigen density) can be calculated by PSA
( serum prostate specific antigen) and prostate volume. Firstly, Input the value of PSA in
urology PSA input box, then measure the prostate volume with the methods described in
section 4.7.2. Finally, the system automatically calculates the value of PSAD, and the
result is displayed on the left side of the screen.
Figure 4-38 PSA Input Box
4.7.2 Prostate Volume Measurement
It is to measure prostate length, width and depth, and then the system automatically
calculates the volume of the prostate. The method is the distance measurement (refer to
Section 4.2.1).
Portable Ultrasound Diagnostic Scanner User’s Manual
92
Figure 4-39 the menu of Prostate Volume Measurement
4.7.3 Residual Urine Volume Measurement
By measuring the length, width and thickness diameter, the system automatically
calculates the volume of residual urine. The method is distance measurement (refer to
Section 4.2.1).
Figure 4-40 the menu of RUV Measurement
4.7.4 Urology Report
In the urological measurement mode, press 【Report】 on the control panel to enter the
URO report, as shown in Figure 4-41.
Refer to Section 4.1.3 General Measurement Report for details of the urological
measurement reports.
Chapter 4 Measurements and Reports
93
Figure 4-41 Urology Report
4.8 Small Part Measurements
This section describes the system of the small part measurement functions, including the
left side of the small part (length, width, depth) and the right side of the small part (length,
width, depth).
Figure 4-42 the main menu of Small Part Measurement
Portable Ultrasound Diagnostic Scanner User’s Manual
94
Small parts measurement only in the B mode, there are two methods to enter the small
parts measurement, as follows:
In real-time state, press【Calc】key to enter the special measurement, and then
press 【6】key to open the menu of small parts measurement;
In real-time state, press 【FREEZE】 and【 6】 key to enter the special
measurement, and then press【6】key to open the menu of small parts
measurement.
4.8.1 Small Parts Measurement
It is to measure the small parts’ length, width and depth, and then the system
automatically calculates the volume of the small parts. The method is distance
measurement (refer to Section 4.2.1).
Figure 4-43 the menu of both sides of small parts measurement
4.8.2 Small Parts Measurement Report
In the small parts measurement mode, press 【Report】 on the control panel to enter the
SMT report, as shown in Figure 4-44.
Refer to Section 4.1.3 General Measurement Report for details of the small parts
measurement reports.
Chapter 4 Measurements and Reports
95
Figure 4-44 Small Parts Measurement Report
96
Chapter 5 Cleaning and Maintenance
5.1 Cleaning
Caution: Please turn off the power before cleaning the unit; otherwise, it may
cause electric shock and equipment damage.
Caution: Avoid any fluid from entering the keyboard or monitor of the equipment;
otherwise, it may cause electric shock or equipment damage.
5.1.1 Cleaning the Main Unit
To clean the surface of the ultrasonic diagnosis system:
1. Power off the ultrasound system and disconnect it from the power source.
2. Use a clean gauze pad or lint-free cloth, lightly moistened with a mild
detergent, to wide the surface of the ultrasound system. Take particular care
to clean the areas near the trackball and the slide controls. Ensure these
areas are free of gel and any other visible residue.
3. After cleaning, use a clean, lint-free cloth to dry the surface.
4. After cleaning, reconnect the ultrasound system to the power supply.
To clean the surface of probe bracket and coupling agent cup:
Use neutral cleaning agent to clean the surfaces of the probe brackets and
coupling agent cup, and then use a piece of lint-free cloth to dry them.
To clean the cables and connector of the diagnosis system:
Use neutral cleaning agent to clean the surfaces of the probe brackets and
coupling agent cup, and then use a piece of lint-free cloth to dry them.
Chapter 5 Cleaning an Maintenance
97
5.1.2 Cleaning the Probe
Caution: To avoid electrical shock and damage to the system, disconnect the
probe prior to cleaning or disinfecting.
Caution: Ensure surfaces of the probe are always disinfected during procedures
requiring disinfection.
Caution: To minimize the risk of cross contamination and infectious diseases,
clean and disinfect the transducer for entocoele and surgical operations thoroughly
after use each time.
Caution: The outer surfaces of an endocavity or intraoperative probe should be
checked to ensure there are no unintended rough surfaces, sharp edges, or
protrusions that may cause a safety hazard.
Caution: Probes are sensitive devices——irreparable damage may occur if they
are dropped, knocked against other objects, cut, or punctured. Do not attempt to
repair or replace any part of a probe.
Caution: Do not use probe sheaths containing oil-based coatings or petroleum-or
mineral oil-based ultrasonic coupling agents. Use only water-based ultrasonic
coupling agent.
Caution: Take extreme care when handling or storing probes. They must not be
dropped, jarred, or knocked against other objects. Do not allow probes to come into
contact with any sharp-edged or po9inted object.
Caution: Do not sterilize probes using hot steam, cold gas, or Ethylene Oxide
(ETO) methods. Before applying any other methods which might be recommended
by manufacturers of sterilization equipment, please contact your supplier.
Caution: Do not use abrasive cleaning agents, organic solvents such as benzene,
isopropyl alcohol, or phenol-based substances, or cleaning agents containing
organic solvents to clean or disinfect probes. These substances can damage the
probes.
Portable Ultrasound Diagnostic Scanner User’s Manual
98
Caution: Do not use aerosol spray cleaner on the transducer, as the cleaner
liquid may enter the enclosure and damage the probe.
Operating procedure for probe cleaning and sterilization:
1. Disconnect the probe from the system..
2. Moisten a clean gauze pad with water and wipe the probe to remove any gel
or particles remaining on the probe. If water is not effective, you may use a
pre-enzyme cleaner
3. Carefully wipe the entire probe, including the cable and connector.
4. To disinfect the probe, take care to keep the cable strain relief and connector
of the probe dry while immersing the probe in an approved disinfectant.
5. Carefully follow the disinfectant manufacturer's instructions.
6. After cleaning or disinfecting, use a clean cloth to dry the probe.
5.2 Maintenance
Caution: To maintain the safety and functionality of the ultrasound system,
maintenance must be performed every 12 months. Electric safety test must be
performed at periodic intervals according to the local safety regulations or as needed.
Chapter 5 Cleaning an Maintenance
99
5.2.1 Main Unit Maintenance
5.2.1.1 External Parts Maintenance
Complete the following each day before using the system:
Visually inspect all probes. Do not use a probe which has a cracked, punctured,
or discolored casing, or a frayed cable.
Visually inspect all power cables. Do not turn on the power if a cable is frayed or
split, or shows signs of wear. Contact your dealer for replacement of the power
cable if a cable is frayed or split, or shows signs of wear.
Verify that the trackball, TGC sliding controls, and other controls on the control
panel are clean and free from gel or other contaminants.
Once the system is powered on.
Visually check the on-screen displays and lighting.
Make sure the monitor displays the current date and time.
Verify that the probe identification and indicated frequency are correct for the
active probe.
5.2.1.2 Internal parts Maintenance
1. It should be established science institution. The Internal parts maintenance should be
based on the use of the environment on a regular, generally speaking, it should not
be less than twice a year, recommendations of the late spring and the Mid-Autumn
two seasons, as more humid in the spring, summer and early autumn are hot climate,
these equipments are vulnerable to adverse effects on.
2. Internal maintenance should be chaired by engineer authorized by Landwind, never
blind to open the cover to carry out maintenance.
3. In the implementation of internal maintenance, electricity is strictly prohibited, and
pulling the power plug system to ensure that equipment and personal safety.
Portable Ultrasound Diagnostic Scanner User’s Manual
100
4. Maintenance is completed, it is necessary to connect to the equipment and check
connector to prevent incorrect connections etc, electricity can confirmed after the
correct test machine.
5. Safeguard the process, pulled the prohibition of arbitrary changes to the wiring and
circuit parts of the location, do not free to spin the adjustable components.
6. Prohibit the use of washing internal parts, nor can of gasoline, kerosene and lacquer
thinner etc, to scrub the motor windings, commutator, as well as transformers,
plexiglass, cable and other rubber products. Alcohol and carbon tetrachloride on the
rubber, and other organic material has dissolved, and therefore can not be used to
scrub cables, rubber line, gelatin plate and mechanical spray paint, these organisms
can be used to dry the wet cloth to clean.
7. Connector on the gold-plated electrical contacts (or silver), it is strictly prohibited to
use gauze or other metal objects polished, and not in direct contact with their hands
in order to avoid corrosion caused by perspiration.
8. The proper use of tools in order to prevent the screw fixed as a result of damage to
the ultra-moment.
As manufacturers and installers of ultrasound equipment, Landwind cannot assume
responsibility for the safety properties, reliability, and/or performance of the equipment if:
Installations, extensions, readjustments, modifications, additions, or repairs are
carried out by persons not specifically authorized by Landwind.
Components that affect the safe operation of the system are replaced by parts
not authorized by Landwind.
The electrical installation of the room where the equipment is located does not
meet the power and environment requirements specified in this manual.
The equipment is not used in accordance with the operation instructions.
The system is operated by persons who do not receive good education or
Chapter 5 Cleaning an Maintenance
101
training.
Landwind suggests that you request any person that performs maintenance, or repairs, to
provide you with a certificate showing:
The nature and extent of the work performed
Changes in rated performance
Changes in working ranges
Date of service
Name of the person or firm performing the service
Signature of the person performing the service
Landwind refuses any responsibility whatsoever for repairs performed without the express
written consent of the Landwind.
5.2.2 Caring Probe
Caution: Ensure surfaces of the probe are always disinfected during procedures
requiring disinfection.
Caution: To minimize the risk of cross contamination and infectious diseases,
endocavity and intraoperative probes must be cleaned and high-level disinfected
after each use.
Caution: The outer surfaces of an endocavity or intraoperative probe should be
checked to ensure there are no unintended rough surfaces, sharp edges, or
protrusions that may cause a safety hazard.
Caution: Probes are sensitive devices——irreparable damage may occur if they are
dropped, knocked against other objects, cut, or punctured. Do not attempt to repair
or replace any part of a probe.
Caution: To avoid cable damage, do not roll the system over probe cables.
Caution: Do not use probe sheaths containing oil-based coatings or petroleum-or
Portable Ultrasound Diagnostic Scanner User’s Manual
102
mineral oil-based ultrasonic coupling agents. Use only water-based ultrasonic
coupling agent.
Caution: Take extreme care when handling or storing probes. They must not be
dropped, jarred, or knocked against other objects. Do not allow probes to come into
contact with any sharp-edged or pointed object.
5.2.3 Caring Coupling Agent
Caution: Before use, examine the coupling agent for any material flaws.. Thinning,
bulging, or brittleness of the material indicates damage. Any product showing flaws
should not be used.
Do not store coupling agent below 5°C nor above 57°C. Coupling agent have a limited
shelf life. Before use, examine these products for any material flaws. Some packaging
may list an expiration date. Any product showing flaws, or whose expiration date has
passed, should not be used.
103
Chapter 6 Technical Description
6.1 Technical Data
Scanning mode electronic liner array, electronic convex array;
Scanning angle 25~120°;
Display depth 250mm;
Scanning depth 200mm(35C60H, B mode, 3.5MHz);
Frequency range 2.0~10MHz;
Probe configuration
35C60H(with a center frequency of 3.5MHz, R=60mm);
65C13H(with a center frequency of 6.5MHz, R=13mm);
75L40H(with a center frequency of 7.5MHz, L=40mm);
Display mode B, 2B, 4B, B/M, M;
CINE 512 frames, supporting automatic playback and manual
playback;
Image smooth 0, 1, 2, 3;
Edge enhance 3, 4, 5, 6;
Dynamic range 30~150dB, with a step of 6dB;
Frame correlation 0~7: eight adjustable ranges;
Gray reject 1~16 adjustable ranges;
Gray map 0~15 adjustable ranges;
Linear density H, M, L;
Portable Ultrasound Diagnostic Scanner User’s Manual
104
Image width four optional width;
Focus mode Quadruple focus:16 optional focus positions;
M Sweep speed 0, 1, 2, 3 (four adjustable speed);
Image flip Left/right flip, up/down flip, polarity flip;
Zoom adjustable two levels of zoom frames;
Body Markers Each subject has fixed body markers;
Display information
Landwind trademark, Hospital, name, medical record ID,
Date/Time, probe information, IP Settings, Examination
subjects ,depth, focus, gray scale, direction symbol, etc;
Examination subjects Abdomen, Gynecology, Obstetrical, Small Part, Orthopaedy,
Urology, Emergency;
B-mode
measurements Distance, Area/CIRC, Volume, Angle, Histogram;
M-mode
measurements Time, Heart rate, Slope, Distance;
Specific Measurement:
Gynecology Uterus, Cervix, left/right ovary, left/right follicle;
Obstetrical
OB Measurements(GS、CRL、BPD、FL、HUM、HC、
AC、FTA、FT、OFD、CER、TTD、APTD), AFI(Q1、
Q2、Q3、Q4), FHR, NT, SELECT FNO, OB INPUT,
Biophysical profile score;
Orthopaedy BL、ARL、IL;
Urology Prostate VOL, PSAD, RUV;
Small parts Left, Right;
Annotation function Supporting body marker, puncture, arrow and text
Chapter 6 Technical Description
105
annotation;
Storage function Image Storage, Image Review, Report Print, etc.;
Acoustic Output parameters(B mode, 35C60H, 3.5MHz):
pra (Ma) 0.957
P (MW) 18.64
z at max. Ipi,α (cm) 4.70
deq(zb) (cm) 4.21
fawf (MHz) 3.445
X (cm) 2.81
Associated acoustic
parameter
Dim of Aaprt
Y (cm) 1.45
td (μsec) 0.332
prr (Hz) 3210
pr at max. Ipi (MPa) 1.673 Other Information
Ipa.3 at max. MI (W/ cm2) 101.42
Depth 110mm
Focal Position 50mm
Scan Width 1
Operating Control
Conditions
Line Density H
Monitor 10 inch CRT;
System VGA system, Supporting external VGA or video printer;
Cursor A combination of trackball and keyboard;
Keyboard Foldable and portable keyboard;
External interface USB, LAN, RS232, Video, VGA, Remote;
Weight Net weight: 9.5 kg; Gross weight: 12kg;
Portable Ultrasound Diagnostic Scanner User’s Manual
106
Dimensions 522×365×420mm3
6.2 System Requirements
This section describes the power and environmental requirements for the ultrasound
diagnostic system.
6.2.1 Power supply Requirements
Power supply: 100-240V~ 50/60Hz 150VA
6.2.2 Connection with Peripheral Devices
Only the peripheral devices listed in the operation instructions are approved to use with
the ultrasound system. Any use of other devices with the system will be at the user’s risk
and may void the system warranty. On-board peripheral devices must be installed by an
authorized Landwind representative.
Caution: Landwind can only guarantee the performance and safety of the devices in
the Accessories and Options chapter. If in doubt, consult Landwind service department or
your local Landwind representative.
6.2.3 Operating Environment Requirements
Ambient temperature: 5℃~40℃
Relative humidity: ≤80%
Atmospheric pressure: 700hPa~1060hPa
Far away from equipment with strong electric and magnetic field and high
voltage
Avoiding the exposure of the screen in direct sunlight; light shielding indoors for
easy observation; good ventilation and anti-dust measures.
Chapter 6 Technical Description
107
6.3 Transport and Storage
6.3.1 Requirement for Transport and Storage Environment
Ambient temperature: -20℃~+55℃
Relative humidity: ≤93%
Atmospheric pressure: 700hPa~1060hPa
6.3.2 System Transport
When shipping the system, perform the following tasks, as appropriate.
To prepare the system for shipment over long distances or rough road:
Disconnect the probe first, and then disconnect the power cable and some
peripherals;
Repackage the system with the delivery-attached packages;
Please fix the system properly to avoid side movement of the system during
transport;
To prevent sudden jarring of the system during transport, provide shock cushions
beneath the system.
6.3.3 Probe Transport
Disconnect the probe from the system.
Place the probe correctly back into the probe box which is specially designed to
protect the probe. Do not dispose it.
6.4 Disposal of Packing Materials
Please properly keep all internal and external packing materials of the instrument
for future transportation and movement.
Please store all internal and external packing materials of the instrument
Portable Ultrasound Diagnostic Scanner User’s Manual
108
together.
6.5 Statement on Consumables
This instrument has no replaceable or consumable parts.
In the case of failure, please contact your dealer and never attempt to dismantle
the machine on your own.