NATIONAL HUB FOR HEALTHCARE INSTRUMENTATION DEVELOPMENT
CENTRE FOR CALIBRATION ANDTESTING OF MEDICAL EQUIPMENT
Department of Science & TechnologyGovt. of India
It may not be the disease... It may not be the doctor's fault... It may not be the medication...
Have you calibrated your medical equipment recently???IT’S WORTH SOMEONE’S LIFE!!!
BUT THE FAULTY MEDICAL EQUIPMENT THAT KILLS!!!
Calibration Team (For Technical Queries)Contact : Project Scientist
Mob : 98408 62782PH : 044 - 2235 7942 Email ID : [email protected] / yahoo.com
Business Liaison Team (For Business Queries)Contact : Business Liason Officer /
Business Liason ExecutivePH : 044 - 2235 7941 / 7938 Email : [email protected] / [email protected]
THE COORDINATORNHHID
2nd Floor, Kalanjiyam Building, Opp. to Mining Engineering, CEG Campus,
Anna University, Chennai - 600025 Phone: 044-2235 7938 / 7941 / 7942
Email : [email protected] / yahoo.com Website : www.nhhid.org
[email protected] GO INDIGENOUS SERVE MANKINDwww.nhhid.org
Dr. S. PoonguzhaliTechnical Manager, CCTME
Associate ProfessorDept. of ECE, Anna University
About the CentreNational Hub for Healthcare Instrumentation Development (NHHID) was established in Anna University with DST funding in 2011 for promoting and accelerating indigenous development of healthcare instrumentation. Centre for Calibration and Testing of Medical Equipment (CCTME) is a sophisticated laboratory of NHHID equipped with state-of-the-art calibrating equipment, IEC standards, operating procedures and trained professionals guided by faculty experts to perform calibration of medical devices to ensure reliable performance for quality healthcare. We provide credible traceability certificates along with calibration reports.
What is Calibration?Calibration is the act of checking or comparing accuracy of measuring equipment with a Standard under specified conditions.
Why Calibrate Medical equipment?Calibrating equipment periodically ensures;
• Accuracy• Efficiency• Safety• Reliability• Longer life of equipment• Uniform Performance of multiple units
Benefits of calibrating medical devices Superior quality of healthcare Enhance patients confidence Mandate for accreditation process (like NABH) Provides patients safety against electrical hazards Reduces risk due to equipment malfunction and increases
patients safety
OUR TECHNICAL MANAGEMENT TEAM
Dr. S. Muttan HOD
Dept of ECE, Anna University
Dr. M. SasikalaDeputy Coordinator, NHHID
Quality Manager CCTMEDirector of Medical Electronics
Professor, Centre for Medical Electronics, Anna University
Dr. Prof. S. MeenakshisundaramCo-Coordinator, NHHIDDean, ACTECH Campus
Dr. Bharanidharan GaneshanAssociate Professor
Radiation Safety Officer, Medical Physics, Anna University
Our expert group is trained & certified in ISO 17025 LQMS & Measurement Uncertainty.
1GO INDIGENOUS SERVE MANKIND
Our services On-site and In-house testing and calibration of medical devices Diagnosis of problems and recommendation for service Quality Assurance check for radiological equipment as per
AERB protocols Free recalibration (within Chennai)
Advantages of our services Brand value Authenticity Constant updates Training
Standards followedISO/IEC 17025 - General requirements for the competence of testing and calibration laboratories.IEC 60601-1:2005 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.IEC 62353:2014 - Medical electrical equipment - Recurrent test and test after repair of medical electrical equipment.IEC 60601-2-12 - Medical electrical equipment - Part 2: Particular requirements for the safety of lung ventilators for medical use.IEC 60601-2-4 - Medical electrical equipment -Part 2: Particular requirements for the safety of cardiac defibrillators and cardiac defibrillators monitors.IEC 60601-2-49 - Medical electrical equipment - Part 2-49: Particular requirements for the safety of multifunction patient monitoring equipment.IEC 60601-2-2 - Medical electrical equipment - Part 2-2: Particular requirements for the safety of high frequency surgical equipment.IEC 60601-2-24 - Medical electrical equipment - Part 2-24: Particular requirements for the safety of infusion pumps and controllers.And other standards.
[email protected] www.nhhid.org
CALIBRATION FACILITY AND CAPABILITY
FOR
FOR
SUCTION PUMP
BOYLES APPARATUS
ANESTHESIA WORKSTATION
PULSE OXIMETERVITAL SIGNS SIMULATOR
GAS FLOW ANALYZER
ECG MACHINE
BP APPARATUS
PATIENT MONITORING
SYSTEM
VENTILATOR
BIPAP
FLOW METER
3GO INDIGENOUS SERVE MANKIND
FOR
FOR
DEFIBRILLATOR
PACEMAKER
DIATHERMY / CAUTERY
DEFIBRILLATOR ANALYZER
ELECTROSURGICAL ANALYZER
[email protected] www.nhhid.org
FOR
FOR
INFUSION PUMP ANALYZER
ELECTRICAL SAFETY ANALYZER
SYRINGE PUMP
INFUSION PUMP
APPLICABLE FOR ALL MEDICAL DEVICES
CALIBRATION FACILITY AND CAPABILITY
5GO INDIGENOUS SERVE MANKIND
FOR
FOR
INCUBATOR ANALYZER
DIGITAL LUX METER
INFANT INCUBATOR
BABY WARMER
OT LIGHT
[email protected] www.nhhid.org
FOR
FOR
FETAL SIMULATOR
DIALYSIS ANALYZER
FETAL DOPPLER
FETAL MONITOR
DIALYSIS MACHINE
CALIBRATION FACILITY AND CAPABILITY
7GO INDIGENOUS SERVE MANKIND
FOR
FOR
PHOTOTHERAPY RADIOMETER
DIGITAL TACHOMETER
PHOTOTHERAPY UNIT
TMT MACHINE
CENTRIFUGE
[email protected] www.nhhid.org
Analytical Balance
FOR
FOR
PIPETTE
AUTOCLAVE
HOT AIR OVEN / INCUBATOR
9GO INDIGENOUS SERVE MANKIND
INDOOR AIR QUALITY TESTING
PRESSURE CALIBRATOR AND GAUGE
OPERATION THEATRE /OTHER COOLING UNITS
SUCTION APP & ALL GAUGES (+ & - VE)
FOR
FOR
[email protected] www.nhhid.org
F1 CLASS WEIGHTS
PH MEASURING INSTRUMENT
PH METER
BABY AND ADULT WEIGHING MACHINE
FOR
FOR
11GO INDIGENOUS SERVE MANKIND
SOUND LEVEL METER
SMART DIFFERENTIAL PRESSURE GAUGE
ICU NOISE TESTING
AIR VENT PRESSURE
FOR
FOR
[email protected] www.nhhid.org
FOR
X-RAY DOSIMETER
CTIQ PHANTOM
MAMMOGRAPHY PHANTOM
HEAD AND BODY
PHANTOM
TOR IQ II PHANTOM
RADIOLOGICAL ANALYZERS
13GO INDIGENOUS SERVE MANKIND
SURVEY METER
RADIATION SURVEY
FOR
X-ray (Mobile, Fixed, Portable) CT Mammography Fluoroscopy C-arm Interventional Radiography Linear Accelerator Dental X-ray (OPG & IOP) DEXA Scan (BMD)
[email protected] www.nhhid.org
NATIONAL HUB FOR HEALTHCARE INSTRUMENTATION DEVELOPMENT CENTRE FOR CALIBRATION
Anna University, Chennai
Test Report Number : NHHID-AU/2015-Customer Name and Address
Description Make / Model/ Type Serial Number Calibrated at Condition of the equipment Date of calibration Recommended due date for next calibrationPartial / Full Calibration
ENVIRONMENTAL CONDITIONS OF MEASUREMENT Temperature Relative Humidity
DETAILS OF MASTER INSTRUMENTS USED FOR CALIBRATION
Description Make
1. Vital Signs Simulator Fluke/ProSim8
1. The Standards used are traceable to National / International Standards.2. This report refers only to the item calibrated and shall not be reproduced in fu
the issuing laboratory. 3. This report is not valid unless it is signed by the authorized signatory.4. Any error in this report should be brought to 5. The result reported is valid at the time of and under the stated conditions of measurement.6. This is a test report not a certificate.
Procedure used for Calibration *DUT - Device Under Test N/A-Limits of performance not applicable
CALIBRATED BY
HUB FOR HEALTHCARE INSTRUMENTATION DEVELOPMENT CALIBRATION AND TESTING OF MEDICAL EQUIPMENT
Anna University, Chennai - 25 TEST REPORT
-02/001 DATE: 23/02/15 M/S. Health center CEG Campus, Anna university, Guindy, Chennai-600025
DETAILS OF THE DUT* ECG Machine Schiller / AT 2 /CF 12525452 On-Site Good 23/02/15
Recommended due date for next calibration 22/02/16 Partial
ENVIRONMENTAL CONDITIONS OF MEASUREMENT 25.0°C - 29.0°C 62% RH - 65% RH
DETAILS OF MASTER INSTRUMENTS USED FOR CALIBRATION
Make / Model Sl. No Traceability
Fluke/ProSim8 200624 TEKTRONIX
CONDITIONS OF ISSUE OF REPORT 1. The Standards used are traceable to National / International Standards.
refers only to the item calibrated and shall not be reproduced in full without the written approval
is not valid unless it is signed by the authorized signatory. rror in this report should be brought to our knowledge within 15 days from the date of this report.esult reported is valid at the time of and under the stated conditions of measurement.
NHHID-AU/STP/06
Limits of performance not applicable VERIFIED BY AUTHORISED SIGNATORY
HUB FOR HEALTHCARE INSTRUMENTATION DEVELOPMENT – NHHID MEDICAL EQUIPMENT
Page 01 of 02
DETAILS OF MASTER INSTRUMENTS USED FOR CALIBRATION
Traceability Calibration Validity
TEKTRONIX 13/06/2018
ll without the written approval of
our knowledge within 15 days from the date of this report. esult reported is valid at the time of and under the stated conditions of measurement.
AUTHORISED SIGNATORY
M/S. Health CentreCEG Campus,Anna University,Guindy, Chennai-600025
Calibration EngineeringTechnical ManagerCalibration Engineer
15GO INDIGENOUS SERVE MANKIND
NATIONAL HUB FOR HEALTHCARE INSTRUMENTATION DEVELOPMENT CENTRE FOR CALIBRATION
Anna University, Chennai
Test Report Number : NHHID-AU/2015- 1.0 VISUAL INSPECTION OF DUT 1.1 Device clean & decontaminated 1.2 No physical damage to case, display,1.3 Switches & Controls operable are correctly aligned1.4 Display Intensity adequate for daytime1.5 Control number, labeling and warnings present are legible1.6 Inlets & hoses 1.7 Power Cord, Accessory cables and charges1.8 Filters & vents clean
2.0 PERFORMANCE TEST 2.1 Heart Rate
Sl. No Standard Value - Master (bpm)
1 80 2 100 3 120 4 130 5 140
2.2 Amplitude
Sl. No Standard Value - Master (mV)
1 1 2 2.5
Conclusion / Remarks: 1. The measured values are within the specified limits
CALIBRATED BY
HUB FOR HEALTHCARE INSTRUMENTATION DEVELOPMENT CALIBRATION AND TESTING OF MEDICAL EQUIPMENT
Anna University, Chennai - 25 TEST REPORT
-02/001 DATE: 23/02/15
PASS √
display, mounts, or connectors √ Switches & Controls operable are correctly aligned √
time use √ Control number, labeling and warnings present are legible √
1.7 Power Cord, Accessory cables and charges √
Indicated Value-DUT (bpm)
Error/ Deviation
(±bpm) 80 0
100 0 120 0 130 0 140 0
Indicated Value-DUT (mV)
Error/ Deviation
(±mV) 1 0
2.4 0.1
specified limits of standard.
VERIFIED BY AUTHORISED SIGNATORY
HUB FOR HEALTHCARE INSTRUMENTATION DEVELOPMENT – NHHID MEDICAL EQUIPMENT
23/02/15 Page 02 of 02
FAIL N/A √ √
Allowable Tolerance
(bpm) Remarks
±5%
√ √ √ √ √
Allowable Tolerance
(mV) Remarks
±5% √ √
AUTHORISED SIGNATORY
Calibration Engineering Technical ManagerTechnical ManagerCalibration Engineer
[email protected] www.nhhid.org
Critical care equipmentsCALIBRATION CHARGES
S. No. EQUIPMENT TESTING PARAMETER Calibration
Charges
1. BP Apparatus, Torniquet, Insufflator Pressure Test, Leak Test 200
2. ECG Machine/TMT ECG Waveform Simulation, Amplitude* Electrical Safety Test* 600
3. Pulse Oximeter O2 Saturation, Heart Rate 400
4. Patient Monitor System ECG Simulation, NIBP, SpO2,Electrical Safety Test* 1200
5. Ventilator, Anesthesia Workstation
Full Breath Test, Pressure/Leakage Test, Continuous Flow Test, Electrical Safety Test*
1200
6. Defibrillator Energy, Charge Time, Electrical Safety Test* 800
7. Infusion Pump/Syringe Pump Flow Test, Occlusion Test, Dual Flow Rate Test*, PCA Test* 600
8. Surgical Diathermy Power Distribution Test, Output Test, ESA* 800
9. Flow Meter, Fumigation Flow Rate 20010. Boyles Apparatus O2 Flow, N2O balance O2 880
11. BIPAP/CPAPFull Breath Test, Pressure/Leakage Test Continuous Flow Test
880
12. Fetal/ Maternal/CTG Monitor Fetal Heart Rate, Electrical Safety Test* 600
13. Baby Warmer/ Infant IncubatorElectrical Safety Test*, Temperature Test, Relative Humidity, Sound & Air Flow
1200
14. Infant Phototherapy Irradiance 40015. Dialysis Conductivity, Flow, Temperature 160016. Pressure Gauge Pressure 600
17. OT Light, Focus light, camera console, spot light Luminance Test 400
18. Suction Pump/ Vacuum Jar Pressure 200
17GO INDIGENOUS SERVE MANKIND
19. Autoclave/Hot Air Oven Temperature, Pressure*, ESA* 800
20. Incubator/B.O.D Incubator/Hot plate Temperature, ESA* 800
21. Interferential Therapy Electrical Safety 32022. Short wave Diathermy Electrical Safety, Scope* 32023. Nerve Stimulator Electrical Safety, Scope* 32024. Traction Unit, OT table Electrical Safety 32025. TENS Electrical Safety, Scope* 32026. Physiotherapy Equipments Electrical Safety, Scope* 32027. Nebulizer Flow, Pressure 200
28. Tube sealer, ELISA Reader, ELISA Washer, EEG Machine Electrical Safety 320
29.
Hydrometer, PH meter, Calorimeter, Photometer, USG Machine, Ultra sonic cleaner, Electrophoresis apparatus
Electrical Safety 320
30. Water bath, Wax Bath, Paraffin bath Temperature, ESA* 800
31. Platelet Agitator with Incubator Temperature, Agitation* 80032. VDRL Rotator Speed, ESA*, Cycles 200
33. Centrifuge, shaker, Power Drill Speed, ESA* 200
34. Refrigerated Centrifuge Temperature, Speed, timer, Imbalance, ESA* 1000
35. Refrigerator, Deep freezer, Plasma Freezer, thermometer Temperature 800
36.Laryngoscope, Bronchoscope, Foetoscope, Colposcope
Luminance, ESA* 400
37. Thermometer(ILC) Temperature 40038. Pipette Volume 500
39. Air quality Temperature, Humidity, Pressure, Co2, O2, Co 1200
40. ILR Temperature 800
* Based on availability, the test will be performed.
Terms and Conditions1. GST @18% and other applicable taxes at the time of payment.
2. Report Delivery: Within 20 days after settling full payment.
3. Execution of work: Within 14 days from your confirmed order.
4. Billing: Variations from the work order will be taken into account and based on that final bill will be prepared
5. The Standards used for calibration are traceable to National/ International Standards.
6. Quality of service will not be compromised for any reasons.
7. 50% Advance payment before start of work to proceed with PO.
8. Accommodation charges will be bared by the customer, when calibration was performed outside Chennai.
9. Onsite Calibration other than Chennai will be performed only when the number of equipments was more than 150.
Radiological Equipments
S.No Equipment Testing Parameters
Unit Charges
(Rs)
1 X ray As per AERB protocols 5500
2 CT As per AERB protocols 9500
3 Mammography As per AERB protocols 5000
4 Radiation Survey As per AERB protocols 2000
19GO INDIGENOUS SERVE MANKIND
OUR REACH
13000+ 250+
Total Number of Equipment
Calibrated so far
Calibration Capability 59 Different types of devices
Total Number of Hospitals Catered
so far
and counting . . .
[email protected] www.nhhid.org
KEY EVENTS
To create awareness as well as to provide this service to IMA-NHB hospitals, a meeting was held at Koodal Hall of Anna University between NHHID and IMA-TNSB office bearers on 16th April 2016. As an outcome of this meeting, 44 IMA-NHB hospitals in Chennai, Erode, Coimbatore, Madurai and Vellore were benefited by the testing and calibration of their medical equipment. NHHID appeals to the Tamil Nadu hospitals to make use this facility fully to ensure that the treatment rendered in their premises is with calibrated instruments and of higher quality and reliability. Together we should make this joint initiative of DST-TDT and Anna University a big success and worthy model to emulate throughout the country.
MoU between DME and Anna University
DIVMED workshop
Meeting with Indian Medical Association Tamilnadu branch
Inauguration of calibration centre