+ All Categories
Home > Documents > Call-In Information

Call-In Information

Date post: 05-Jan-2016
Category:
Upload: chipo
View: 35 times
Download: 6 times
Share this document with a friend
Description:
Call-In Information. Web meeting address: https://www.webmeeting.att.com Meeting number 8773361828 Access Code: 1666469 Join the audio portion at: 877-336-1828 Access code 1666469 (yes, they are the same) (the audio will not be able to be streamed through the computer). - PowerPoint PPT Presentation
57
Call-In Information Web meeting address: https://www.webmeeting.att.com Meeting number 8773361828 Access Code: 1666469 Join the audio portion at: – 877-336-1828 Access code 1666469 (yes, they are the same) (the audio will not be able to be streamed through the computer)
Transcript
Page 1: Call-In Information

Call-In Information

• Web meeting address: https://www.webmeeting.att.com – Meeting number 8773361828– Access Code: 1666469

• Join the audio portion at: – 877-336-1828– Access code 1666469 (yes, they are the same)

(the audio will not be able to be streamed through the computer)

Page 2: Call-In Information

Michigan Regional Laboratory System

Laboratory In-service:Laboratory Director and Site

Coordinator Training

Jeffrey P. Massey, Dr.P.H., HCLD(ABB)Michigan Department of Community Health

Quality Manager517-335-8074

[email protected]

Page 3: Call-In Information

Agenda

• Laboratory Quality Commandments• Guidelines for review of QC documentation• Good Laboratory Practice & Laboratory

Documentation• Responsibilities

– Lab Director– Clinical Consultant– Site Coordinator

Page 4: Call-In Information
Page 5: Call-In Information

Laboratory Quality Commandments

Yes, they are carved in stone!

Page 6: Call-In Information

I.I. Thou shall not perform

any clinical test for which thou

are not trained and checked

for competency.

Page 7: Call-In Information

• . II. Thou shall not use any expired reagents or supplies to perform any clinical test, including quality control evaluations

Page 8: Call-In Information

III.III. Thou shall not perform any patient test for which any of the Quality Control checks have failed.

Page 9: Call-In Information

IV. Thou shall not perform any test in any manner except as specified in the written standard operating procedure or manufacturers product insert.

Page 10: Call-In Information

Laboratory Quality Commandments

• There are no exceptions!

Page 11: Call-In Information

Quality Assessmentanswers these questions

Does the procedure work the way it is supposed to work?

Does the data produced reflect a real condition in the client?

Page 12: Call-In Information

Quality Control:Quality Assessment Component

• Intended to detect, reduce, and correct errors prior to the release of patient results.

• QC measures of precision or how well the measurement system reproduces the same result over time and accuracy or if that measurement is what is really present in the patient.

Page 13: Call-In Information

Quality ControlIn practice, we have three parameters to evaluate:

1. The Instrument:• Does the instrument perform as expected*?

2. The Reagents:• Do the reagents perform as expected?

3. The Analyst:• Is the person performing the test correctly?

*within the parameters in the manufacturer’s insert as approved by the FDA

Page 14: Call-In Information

Quality Control: Equipment

• What equipment preventative maintenance is indicated by the manufacturer?

• Has the indicated maintenance been performed:– on schedule?– and documented?

Page 15: Call-In Information

Quality Control: Equipment

• Instrument Checks– Display shows what it is supposed to when

power is turned on.• Self check is ok?

– Instrument check device• Optics Cuvette• Calibration Cassette• Calibration strip, tag, stick

– Calibration checks are within expected range

Page 16: Call-In Information

Quality Control: Reagents

• Guidelines apply to Test reagents and QC materials!

• Determine closed vial (or other container) expiration date specified by manufacturer.– Specific date (March 13, 2011 = March 13, 2011)– End of the month (June 2011 = June 30, 2011)

• Determine open vial expiration date– 30, 60 or 90 days after opening– Open vial expiration date CANNOT exceed the

closed vial expiration date

Page 17: Call-In Information

Quality Control: Reagents

• Store according to manufacturer’s instructions– Protect from moisture and light– Bring to room temperature before opening

packet unless manufacturer says otherwise

• Open only what you need– Monitor/document room temperature if reagents

stored at room temp or test performed at room temp

Page 18: Call-In Information

Quality Control: Reagents

• If any of your reagents are expired, then you can not perform the test for clinical purposes.

• Options:– Find some reagents that are not expired– Use your backup plan, or – Cancel your clinic!

Page 19: Call-In Information

Quality ControlA Reagent Cardinal Rule

• NEVER, EVERNEVER, EVER use expired reagents to test clinical specimens.

• Expired reagents should be:– Discarded– Marked: Warning: EXPIRED. Use for

training only. Not for clinical use.

Page 20: Call-In Information

Quality Control: Reagents• QC material has an established value (or range of

values) for the analyte being measured by the test. • QC frequency minimally must follow those

established by the lab director for non-waived tests or is stated in the manufacture’s insert for waived tests.

• Record the results of QC materials at the time of testing.

• Do not proceed with patient testing if QC values are not acceptable

Page 21: Call-In Information

Quality Control ”The job is not done until the paperwork is

complete”

• Quality Control Log Sheets– The frequency or interval for QC testing depends upon

the test: check your procedure manual for appropriate intervals

– Document instrument control checks• Make sure the machine works before you proceed to test the

reagents.

– Document results at the time you obtain the result – don’t rely on your memory by filling in the result later

Page 22: Call-In Information

Quality Control Quality Control Log Sheets

• Document lot numbers and expiration dates (both open vial and closed vial expiration dates)

• Test both high and low level controls. (Positive and negative if pregnancy or urine dipstick).

Page 23: Call-In Information

Reporting Results:Positive

• Be consistent in how positive results are documented

• Positive– P– Pos– Positive

– Do not use the Symbol “+”

Page 24: Call-In Information

Reporting Results:Negative

• Be consistent in how negative results are documented:

• Negative– N– Neg– Negative

– Do not use the symbol “-”

Page 25: Call-In Information

Reporting Results:Quantitative

• Use 1+, 2+, 3+, 4+

• Do not use +, ++, +++, ++++

Page 26: Call-In Information

Quality Control Quality Control Log Sheets

• If the results for high and/or low controls were out of range, consult the Trouble Shooting Guide

• Record any corrective action(s) taken on the QC Log– Your surveyor will say, “If you didn’t record it

then I must assume that you didn’t do it”.

Page 27: Call-In Information

Quality ControlHave a back up plan!

• If your instrument or reagents fail the quality control checks; You may not You may not use that procedure for clinical use that procedure for clinical testing testing until;– Corrective action has been taken and – The instrument and reagent checks pass

all quality control checks.

• There are no exceptions to this ruleThere are no exceptions to this rule.

Page 28: Call-In Information

The Penalty for Reporting Results if there is a QC Failure ……

Page 29: Call-In Information

Quality Control

• QC Logs need to be reviewed:– By the Site Coordinator on a regular basis

(monthly or quarterly)– By the Laboratory Director on a regular basis

• May review a representative sample rather than all records

– Your assigned Technical Consultant will review if requested (Enhanced Services Option)

• Corrective action is required if problems are identified

• QC Records need to be filed for 2 years in a manner that allows ready retrieval.

Page 30: Call-In Information

Quality Control: Testing Personnel

• Qualifications– Moderately complex– PPM– Waived

• Training– Cover all phases of testing– Checklists helpful

• Documented Competency– Direct observation– Trouble shooting scenarios

Page 31: Call-In Information

Quarterly Chart AuditAn Important Quality Assessment activity

• Required for non-waived testing, recommended for waived testing

• Quarterly Chart Audit.– Verify that the QC Logs are completed for each test

performed• Lot # & expiration date, QC within acceptable limits,

corrective action initiated as required

– Randomly select 3-5 clients for each test. Verify that laboratory data is accurately transcribed to their chart; QC was performed and acceptable on the day of testing, testing personnel was trained and competent on the day of testing.

Page 32: Call-In Information

Quality Control

“Quality in never an accident. It is always the result of an intelligent effort. It is the will to produce a superior thing.”

John Ruskin

Page 33: Call-In Information

Good Laboratory Practice

• Quality Records & Documentation: Characteristics of a Good Record– Overall appearance– Permanence (pen not pencil)– Initials vs. Signature on record– Timing– Completeness– Appropriate error correction

Page 34: Call-In Information

Overall Appearance

• Include document control elements

• Keep forms simple

• Allow enough room for data entry

• Entries must be legible

Page 35: Call-In Information

Characteristics of a Good Record:Overall appearance/permanence

• Legibility is essential• Write in permanent

ink– No pencil– No White Out

Is this Jan or Jun? Is this 2000 or 2006?

Is this 1 or 7?

Page 36: Call-In Information

Quality RecordPermanence

• Always use:– Black or blue indelible ink

• DO NOT USE:– Markers, felt tip pens, gel pens, glitter

pens, pencil, liquid paper, White Out

Page 37: Call-In Information

Characteristics of a Good Record:Timing

Room Completed

(Place a √)

Initials Date Date

105 √ JPM 4/16/10 4/15/10

106 √ JPM 4/16/10 4/15/10

Date completed: April 15, 2010

WRONG CORRECT

By waiting to record tasks, you may forget to fill out the documentation. A later date causes speculation.If it is not documented – the task was not done.

Page 38: Call-In Information

Quality Records and Documentation: Completeness of Record

• Decimals

• Not applicable (N/A)

• Not in Use / Test Not Performed

• Expression of dates

• Arrows/ditto marks

• Initials vs. Signature

Page 39: Call-In Information

Completeness of RecordNot Applicable (N/A)

• When completing a form, indicate the spaces or blanks that do not require a response.

• Record “NA” in spaces or blanks that do not apply to the situation. Do not leave spaces blank

Page 40: Call-In Information

Completeness of RecordNot Applicable (N/A)

Room Completed

(Place a √)

Date Initials

105 √ 4/16/10 JPM

106

Blanks look as though operator forgot to complete the entry

Page 41: Call-In Information

Completeness of RecordNot Applicable (N/A)

Room Completed

(Place a √)

Date Initials

105 √ 4/16/10 JPM

106 N/A 4/16/10 JPM

Fill in any entry spaces that are not to be completed with “N/A”

Page 42: Call-In Information

Completeness of RecordNot in Use or Test Not Performed

• When an instrument is not used that day, enter “Not in Use” in the appropriate space. Include your initials and the date where needed.

• When a Test was not performed, enter “TNP” in the appropriate space, and include your initials and the date where needed.

Page 43: Call-In Information

Completeness of RecordsExpression of Dates

• All dates should be entered as month, day, year– Examples

• April 16, 2010• Apr 16, 2010• 4/16/10• 04/16/2010

• The year must be recorded on all forms

Page 44: Call-In Information

Correction of Erroneous Entries

• DO Not:– Use “White Out” (e.g., Liquid Paper)– Use erasors– Obliterate – Write over– Tape over

Page 45: Call-In Information

Appropriate Error Correction:Obliteration

• Do Not Obliterate an Entry

– What would a reviewer think?– Site Coordinator– Laboratory Director– Surveyor

Page 46: Call-In Information

Appropriate Error CorrectionWrite Over

• Do Not write over incorrect entries

• Again – what would a reviewer think??

Page 47: Call-In Information

Procedure for correcting erroneous entries

• 1. Incorrect Result

• 2. Draw a single line through incorrect entry

• 3. Write correct result next to incorrect entry

• 4. Initial and date the correction

Page 48: Call-In Information

Laboratory Director: Primary Responsibilities

(both Waived & Non-Waived Labs)

• Has ultimate responsibility for ALL testing

• Must ensure quality laboratory services

• Assign in writing the duties/responsibilities for each person involved in all phases of the testing process

Page 49: Call-In Information

Responsibilities of Lab Director:Waived Laboratory

• Ensure testing is performed EXACTLY as stated by the manufacturer (including QC)

• Ensure that laboratory staff follow good laboratory practice

• Permit announced or unannounced inspections

• Maintain current CLIA certification

Page 50: Call-In Information

Laboratory Director Responsibilities (for moderate complexity testing)

• Ensure safe working environment• Ensure tests chosen provide quality of

results required for client care• Ensure enrollment in PT program• Ensure Quality Control & Quality

Assessment programs are established & maintained

• Ensure corrective action is taken & documented

Page 51: Call-In Information

Laboratory Director Responsibilities (cont.)

• Ensure test results only reported when test is functioning properly

• Ensure reports of test results include information for interpretation

• Ensure consultation of test results

• Ensure the lab employs sufficient number of testing staff

Page 52: Call-In Information

Laboratory Director Responsibilities (cont.)

• Ensure testing staff are trained and competent

• Ensure procedure manual is available (review and sign on annual basis)

• Specify in writing the responsibilities of each consultant and testing staff

Page 53: Call-In Information

Responsibilities That CAN Be Delegated

• Selection of test menu

• Enrollment in PT program

• Maintenance of QC/QA program

• Corrective action program

• Training and Competency of staff

• Make procedure manual available to staff

Page 54: Call-In Information

Responsibilities that CAN be Delegated to Clinical Consultant

• Interpretation of test results

• Consultation available to clients concerning quality of test results

Page 55: Call-In Information

Responsibilities that CANNOT be delegated by Lab Director

• Ensure safe working environment

• Ensure proper size of testing staff

• Ensure overall quality of test results

• Ensure annual review and signature of test procedures

Page 56: Call-In Information

Questions?

Page 57: Call-In Information

Questions???


Recommended