Date post: | 27-Oct-2015 |
Category: |
Documents |
Upload: | emir-goekberk-eken |
View: | 37 times |
Download: | 1 times |
T.C.BAHÇEŞEHİR UNIVERSITY
CARDIOVASCULAR SYSTEM MOCK CIRCUIT
Capstone Project
Emir Gökberk Eken
İSTANBUL, 2012
T.C.BAHÇEŞEHİR UNIVERSITY
FACULTY OF ENGINEERING
DEPARTMENT OF MECATRONICS ENGINEERING
CARDIOVASCULAR SYSTEM MOCK CIRCUIT
Capstone Project
Emir Gökberk Eken
Advisor: Dr. Kamuran KADIPAŞAOĞLU
İSTANBUL, 2012
T.C.BAHÇEŞEHİR UNIVERSITY
FACULTY OF ENGINEERINGDEPARTMENT OF MECATRONICS ENGINEERING
Name of the project: Cardiovascular System Mock CircuitName/Last Name of the Student: Emir Gökberk EkenDate of Thesis Defense: 06/07/2012
I hereby state that the graduation project prepared by Emir Gökberk Eken has been completed under my supervision. I accept this work as a “Graduation Project”.
06/07/2012 Dr. Kamuran Kadıpaşaoğlu
I hereby state that I have examined this graduation project by Emir Gökberk Eken which is accepted by his supervisor. This work is acceptable as a graduation project and the student is eligible to take the graduation project examination.
06/07/2012 Prof. Dr. Oktay ÖZCAN
Head of the Department of Mechatronics Engineering
We hereby state that we have held the graduation examination of Your Name and agree that the student has satisfied all requirements.
THE EXAMINATION COMMITTEE
Committee Member Signature
1. Dr. Kamuran Kadıpaşaoğlu ………………………..
2. Prof. Erol Sezer ………………………..
3. Dr. Khalid Abidi ………………………..
ACADEMIC HONESTY PLEDGE
In keeping with Bahçeşehir University Student Code of Conduct, I pledge that this work is
my own and that I have not received inappropriate assistance in its preparation.
I further declare that all resources in print or on the web are explicitly cited.
NAME DATE SIGNATURE
Emir Gökberk Eken 06/07/2012
5
ABSTRACT
CARDIOVASCULAR SYSTEM MOCK CIRCUIT
Emir Gökberk Eken
Faculty of EngineeringDepartment Mechatronics Engineering
Advisor: Dr. Kamuran Kadıpaşaoğlu
JULY, 2012, 69 pages
Heart disease is founded as the highest cause of death in the world. These loses can
reduce by improving heart valves, ventricular assist devices (VADs), total artificial hearts,
heart and lung machines etc. All these medical devices have to be tested on cardiovascular
mock circuits (CVMCs) before starting clinical testing.
At 1975 Donovan et al. designed and built a complete mock circulation for testing
artificial hearts but, atria or ventricles wasn’t include to system. This CVMC has been the
basis for many. Timms et al. (2010) designed and constructed a CVMC which includes
both systemic and pulmonary circulatory systems. They accomplished to produce left
ventricular pressure but, they couldn’t prevent lagging in aortic pressure. Aortic valves
opening and closing times or wrong resistance values may cause this problem. For prevent
this problem more accurate heart valves may use and resistance may optimize with feed-
back control of proportional control valves.
Objective in this study was to design and build CVMC which is drivable with an
elastance based feed-back control for simulate heart functions in various conditions.
CVMC virtually designed by using Solidworks based on results which are obtained in the
Simulink model. Pressure and flow inputs and check valve outputs are implemented into
system by using dSPACE control card. CVMC operated successfully with compressed air
and check valves. However, results are indicated a leaking problem of heart valves.
Key Words: Cardiovascular Mock Circuit, VAD and Heart valve test
6
ÖZET
CARDIOVASCULAR SYSTEM MOCK CIRCUIT
Emir Gökberk Eken
Mühendislik FakültesiMekatronik Mühendisliği Bölümü
Tez Danışmanı: Dr. Kamuran Kadıpaşaoğlu
TEMMUZ, 2012, 69 pages
Kalp hastalıklarından kaynaklı ölümler hala dünyadaki en yüksek sayıdaki ölüm
sebebidir. Bu kayıplar kalp kapakçıkları, ventrikül destek üniteleri (VADs), yapay kalpler,
kalp-akciğer makinaları vs. Geliştirilerek azaltılabilir. Ancak bütün bu araçlar kılinik
testler öncesinde kardiovasculer mock düzenekleride (CVMC) test edilmelidir.
Donovan ekibi 1975 yılında kalp kapakçıklarının denenmesi amacıyla CVMC
dizayn edip kurdular fakat, kulakçık ve karıncıklar sisteme dahil edilmemişti. Bu sistem
bir çok yenisi için temel niteliği taşımaktadır. 2010 yılında Timms ve ekibi sistemic ve
pulmoner tarafların bulunduğu bir CVMC dizayn edip kurdular. Bu sistemle sol karıncık
basıncını başarılı bir şekilkde elde etmelirne karşın aort basıncının gecikmeli olarak elde
edilmesini engelliyemediler. Bu probleme aort kapakçığının zamanında açılıp
kapanamaması yada direnç değerlerinin optimum olmaması sebep olmuş olabilir. Bu
problemi engellemek için daha iyi kalp kapakçıkları seçilerek, dirençlerde geri-besleme
konrollü, doğrusal akış vanalarıyla optimize edilebilinir.
Kalp fonksiyonlarını çeşitli durumlara yerine getirebilicek, geri-beslemeli
elastansla kontrol edilebilicek bir CVMC dizayn edip kurmak bu projenin amacıdır.
Bunun için Matlabdeki simülasyonlardan elde edilen verilerle, SolidWorks de sanal
dizaynı yapıldı. Basınç, akış bilgileri ve çek valflerini kontrolleri dSpace kontol borduna
yerleştirldi. CVMC hava basıncı kullanılarak başarlı birşekilde işletildi. Ancak soruçlar
kalp kaçakçıklarının kaçırdığına işaret etti.
Anahtar Kelimeler: Kardiovasculer Mock düzeneği, Kalp Kapakçığı ve VAD test
7
TABLE OF CONTENTS
ABSTRACT..........................................................................................................................v
ÖZET....................................................................................................................................vi
TABLE OF CONTENTS....................................................................................................vii
LIST OF FIGURES.............................................................................................................ix
LIST OF TABLES..............................................................................................................xii
LIST OF ABBREVIATIONS............................................................................................xiii
LIST OF SYMBOLS..........................................................................................................xv
I. INTRODUCTION........................................................................................................1
1. Heart.....................................................................................................................1
1.1. Anatomy and Pathology of the Heart as a Pump.........................................1
1.2. The Cardiac Cycle.......................................................................................2
1.3. Frank – Starling Mechanism.......................................................................4
1.4. Resistance....................................................................................................4
1.5. Compliance..................................................................................................5
1.6. Elastance......................................................................................................5
1.8. Heart Disease...............................................................................................9
2. Mechanical Circulatory Assistance (MCA).........................................................11
2.1. Left Ventricular Assist Devices (LVADs)................................................11
2.2. Cardiovascular Mock Circuits (CVMCs)..................................................13
II. OBJECTIVES............................................................................................................18
1. Desired Functions.................................................................................................18
2. Design Criteria.....................................................................................................18
III. MATERIALS and METHODS..................................................................................19
1. Cardiovascular Mock Circuit (CVMC) Simulation...........................................19
1.1. Aim.........................................................................................................19
1.2. Method....................................................................................................19
8
1.3. Simulink Model......................................................................................19
1.4. Design.....................................................................................................20
2. Cardiovascular Mock Circuit Virtual Design....................................................26
2.1. Aim.........................................................................................................26
2.2. Method....................................................................................................26
3. Cardiovascular Mock Circuit Mechanical Building..........................................28
3.1. Heart Valves...........................................................................................29
3.1.1. Heart Valve Selection..........................................................................29
3.1.2. Assembly of Heart Valves...................................................................29
3.2. Check Valves..........................................................................................30
3.3. Pressure Sensors.....................................................................................32
4. Cardiovascular Mock Circuit Control................................................................34
4.1. Aim.........................................................................................................34
4.2. Method....................................................................................................34
4.3. Hardware Installation of dSPACE DS1104............................................34
4.4. Software Installation of Control Desk....................................................35
4.5. Software Installation of Matlab/Simulink..............................................36
IV. RESULTS...................................................................................................................41
1. LV and Ao Pressures..........................................................................................42
2. Left Ventricular Volume....................................................................................44
3. P-V Loop............................................................................................................45
4. Elastance............................................................................................................46
V. DISSCUSION & CONCLUSION.............................................................................47
VI. FUTURE WORK.......................................................................................................48
VII. APPENDIX A............................................................................................................48
REFERENCES....................................................................................................................51
9
LIST OF FIGURES
Figure 1- Structure of the heart, and course of blood flow through the heart chambers and
heart valves............................................................................................................................1
Figure 2- Events of the cardiac cycle for left ventricular function, showing changes in left
atrial pressure, left ventricular pressure, aortic pressure and ventricular volume. ...............3
Figure 3- Effect on the cardiac output (CO) curve of left atrial pressure (LAP)..................4
Figure 4- Pressure – Flow relationship value references on the pipe....................................5
Figure 5- Elastance and Compliance in RV, LV and Ao at systole and diastole..................6
Figure 6- Volume-pressure diagram, demonstrating changes in volume and pressure
during the normal cardiac cycle............................................................................................7
Figure 7- Pressure-Volume Loop and Elastance-Time Curve..............................................7
Figure 8- Ventricular Potential Enegry and Stroke Work Diagram......................................8
Figure 9- Effect on the Demand Oxygen DO2 of pressure volume area (PVA)...................9
Figure 10- HeartMate II LVAD as Bridge to Transplant System in Patients Body............11
Figure 11- Parameters of open air reservoirs in a single subsystem of CVMC model.......16
Figure 12- Parameters of closed air reservoirs in a single subsystem of CVMC model.....16
Figure 13- CVMC chambers and connections to each other..............................................19
Figure 14- Parameters of pneumatic cylinder in a single subsystem of CVMC model......20
Figure 15 - A schematic of open air reservoirs in a single subsystem of CVMC model....21
Figure 16- Matlab/Simulink result of open-air reservoir....................................................21
Figure 17- Figure 18 - A schematic of close air reservoirs in a single subsystem of CVMC
model...................................................................................................................................22
Figure 19- Simulation Result for Compliance of Aorta Chamber......................................22
Figure 20- Pneumatic Check Valves Placement in Left Ventricual System.......................23
Figure 21- Sine function values with respect to time..........................................................23
Figure 22- A Simulink schematic and sine wave function and saturation values of RV of
CVMC model......................................................................................................................24
Figure 23- Result of RV pressure function designed on Simulink.....................................24
Figure 24- A Simulink schematic and sine wave function and saturation values of RV of
CVMC model......................................................................................................................25
10
Figure 25- Result of LV pressure function designed on Simulink......................................25
Figure 26- A schematic of CVMC model...........................................................................25
Figure 27- Left Ventricle Pneumatic System Design..........................................................26
Figure 28- Solidworks Drawing of Designed Chamber......................................................27
Figure 29- Solidworks Design of CVMC...........................................................................28
Figure 30- St. Jude Mechanical Heart Valve......................................................................29
Figure 31- Assembly of Heart Valves.................................................................................29
Figure 32- Pneumatic Check Valve....................................................................................30
Figure 33- Technical Drawing of Pneumatic Check Valves...............................................30
Figure 34- Check Valves Relay Circuit..............................................................................31
Figure 35- a) Calmed® Pressure Transducers for Aort and Left Ventricle b) Edwards™
Pressure Transducers for Pressure Tank.............................................................................32
Figure 36- PBV connector of Pressure Transducers...........................................................32
Figure 37- A Non-inverting Amplifier Circuit....................................................................32
Figure 38- Amplifier Circuit...............................................................................................33
Figure 39- DS1104 R&D Controller Board........................................................................34
Figure 40- Single Board Hardware of DS1104...................................................................35
Figure 41- Control Desk Example Screen...........................................................................35
Figure 42- Simulink Schematics of Pressure Transducers..................................................36
Figure 43- Transonic Systems HT110 Bypass Flow Meter Panel......................................37
Figure 44- Transonic Systems Inc. Ultrasonic Flow Meter Sensor....................................37
Figure 45- Flow meter placement on Y-connector.............................................................38
Figure 46- Simulink Schematics of Flow Sensor................................................................38
Figure 47- Simulink Schematics of Check Valves Controller............................................39
Figure 48- Simulink Schematic of General Connection.....................................................39
Figure 49- dSPACE connections of pressure transducers, flow meter and check valves...40
Figure 50- Workspace, heart monitor, dSPACE board, Flow-meter panel, amplifiers and
power-supply.......................................................................................................................40
Figure 51- Built four chambers Mock circuit......................................................................41
Figure 52- ControlDesk GUI...............................................................................................42
Figure 53- Ideal PLV, PAo with Time Graph.........................................................................42
Figure 54- Recorded LV and Ao Pressures.........................................................................43
11
Figure 55- Pressure Relationships in Aortic Insufficiency vii..............................................44
Figure 56- Ideal Left Ventricular Volume Change.............................................................44
Figure 57- LV Volume Change produced by CVMC.........................................................44
Figure 58- Ideal P-V Loop for Healty Human....................................................................45
Figure 59- Recorded P-V Loop of CVMC..........................................................................45
Figure 60- Mitral and Aortic Valve Regurgitation .............................................................45
Figure 61- Ideal Elastance Graph........................................................................................46
Figure 62- Created Elastance Signal for P2.........................................................................46
12
LIST OF TABLES
Table 1- VAD devices, generations, manufacturers and types...............................12
13
LIST OF ABBREVIATIONS
ADC Analog to Digital Converter
AHF Acute Heart Failure
Ao Aorta
AV Aortic Valve
C Compliance
CHF Chronic Heart Failure
CO Cardiac Output
CVMC Cardiovascular Mock Circuit
DAC Digital to Analog Converter
GUI Graphical User Interface
HF Heart Failure
HR Hearth Rate
I/O Input/output
IVC Isovolumetric or Isovolumic contraction
IVR Isovolumetric or Isovolumic relaxation
LA Left Atrium
LV Left Ventricle
MAP Mean Aortic Pressure
MCA Mechanical Circulatory Assistance
MV Mitral Valve
PA Pulmonary Artery
PA Pulmonary Artery
PC Pulmonary Capillary
PE Potential Energy
PV Pulmonary Venous
PVA Pressure Volume Area
PVR Pulmonary Vascular Resistance
RA Right Atrium
RTI Real-time Interface
RV Right Ventricle
14
SAP Systolic Aortic Pressure
SC Systemic Capillary
SV Stroke Volume
SVR Systemic Vascular Resistance
SW Stroke Work
T Cardiac Cycle Length
TV Tricuspid Valve
VAD Ventricular Assist Device
15
LIST OF SYMBOLS
Symbol ExplanationUnits
SI ConventionalP Pressure Pa mmHgV Volume m³ m³L Length cm mr Radius cm md Diameter cm m
π Mathematical constant[dimension-
less][dimensionless]
∆ Difference[dimension-
less][dimensionless]
µ Dynamic viscosity Pa.s N.s/m²
Q Volumetric flow rate m3 /s mL/min
R Resistance (Hydraulic) Pa.s/m³ mmHg.min/mLC Compliance m³/Pa m³/mmHgE Elastance Pa/m³ mmHg/mL
I. INTRODUCTION
1. Heart
The heart can be considered as a pulsatile pump the duty of which is to receive blood
from a low pressure reservoir and deliver it to high pressure reservoir at a given flow rate.
Unidirectional characteristic of the fluid flow inside the pump is achieved with check valves.
1.1. Anatomy and Pathology of the Heart as a Pump
Right atrium (RA) receives deoxygenated blood returning from the body via superior
and inferior venae cava (See Figure 1). Blood flows into the right ventricle (RV) through,
the tricuspid valve (TV). It is then pumped through the pulmonary valve into the
pulmonary artery (PA). The PA splits into two to supply blood to both the left and right
lungs, where the blood is oxygenated and flows into the left atrium (LA). Mitral valve
(MV) opens and blood flows into the left ventricle (LV), where it is pumped through, the
aortic valve (AV) into the aorta (Ao) at high pressure. Blood then flows through the high
resistance systemic arteries and the low resistance systemic veins back to the RA. From
this point, the circulatory cycle starts over to continually supply oxygenated blood to the
whole body.
Figure 1- Structure of the heart, and course of blood flow through the heart chambers and heart valves1
1
The circulation through the pulmonary arteries, capillaries and veins is realized by the
RV and the circulation though the systemic arteries, capillaries and veins is realized by
the LV. Since the total pulmonary resistance is almost five times lower than that of the
total systemic resistance, the pressures are proportionally higher in the LV compared to
RV in order to realize the same flow rate across each pump. This relationship can be
expressed by the formula,
Pressure Diffence (∆ P )=Flow (Q)x Resistance (R) (Eq. 1)
1.2. The Cardiac Cycle
The cardiac cycle starts from the beginning of one heartbeat and ends at the
beginning of the next heartbeat; and consists of four phases. Only the cycle in the LV will
be explained below for simplicity, the cycle in the right heart being exactly the same,
except for the differences in the names of the anatomic structures involved.
The first phase is called diastole. Diastole includes rapid inflow of blood into the LV
from the LA through the MV, diastasis and atrial kick. Rapid inflow is caused when the
ventricular pressure falls below the atrial pressure. As a result, the AV valve opens and
blood flows into the ventricle rapidly as ventricular pressure keeps falling due to the
continued relaxation of the ventricular wall. Diastasis is the period during which the heart
is completely relaxed and the inflow rate is slowed down. . Towards the end of diastole,
the atrium contracts, increasing the pressure inside and forcing the remaining blood into
the ventricle. This is called the atrial kick.
The second phase of the cardiac cycle is isovolumetric (or isovolumic) contraction
(IVC). As the ventricle contracts, the pressure inside the left ventricle (left ventricular
pressure, PLV) rapidly increases and forces the closure of the mitral valve. LVP continues
to rise until it reaches the level of the pressure inside the aorta (Aortic pressure, P Ao).
During this short period of rapid pressure rise, left ventricular volume (VLV) doesn’t
change because both AV and MV are closed. As soon as PLV surpasses PAo, AV opens,
blood starts to flow into the aorta, VLV begins to drop and IVC ends.
2
The third phase is called systole. Systole consists of two parts. These are rapid
ejection and deceleration. Rapid ejection occurs when pressure inside the ventricle
becomes higher than the pressure of the aorta then aortic valve open and blood flows into
the aorta. Ventricular pressure remains slightly higher than Aortic pressure in order to
create flow (QLV) through the aortic valve resistance shown in (Eq. 1). However, volume
of ventricle decreases rapidly. Reduced ejection occurs when aortic pressure just above
that the ventricular pressure, but blood flows into Ao because of inertia of blood itself.
The atrial pressure rises slowly during this phase.
The fourth and last phase is Isovolumetric relaxation (IVR), occurs when the
ventricular pressure falls rapidly and aortic pressure become lower than below ventricular
pressure and aortic valve closes (See Figure 22).
All phases listed for LV are also valid for RV.
Figure 2- Events of the cardiac cycle for left ventricular function, showing changes in left atrial
pressure, left ventricular pressure, aortic pressure and ventricular volume. 1
3
1.3. Frank – Starling Mechanism
The Frank Starling law states that the heart has ability of increasing contraction force
and as a result its stroke volume. The Frank Starling law states that as more blood is
flowed into the ventricle, the heart works harder to pump more blood. The Frank Starling
law also states that heart is a loop that doesn’t leak blood and what goes into ventricle
must come out.1 Cardiac Output formula is shown below;
Cardiac Output (CO )=StrokeVolume (SV ) x Heart Rate (HR ) (Eq. 2)
1.4. Resistance
Hydraulic resistance in vessels is defined as the slope of the pressure-flow
relationship obtained by measuring vascular pressure at several fluid flow rates 2.
It is determined by the length (L), cross-section area of the pipe and fluid viscosity
(μ). Cross-section area calculates by using diameter (d) or radius (r). In standard fluid
dynamics notation 3 (See Figure 4 ).
∆ P=8 μL
π r4Q∨∆ P=128 μL
π d4Q (Eq. 3)
4
Figure 3- Effect on the cardiac output (CO) curve of left atrial pressure (LAP)
Figure 4- Pressure – Flow relationship value references on the pipe
1.5. Compliance
Compliance (C) is the unit change in volume (∆V) of a deformable reservoir under
unit change in pressure (∆P) applied into the reservoir. Also compliance is a passive
property of the reservoir. It is given by the formula,
C=∆ V∆ P
(Eq. 4)
1.6. Elastance
The inverse of the compliance, the elastance is an active property. Elastance (E)
describes the dynamics of the active return to the original shape of the variable volume
reservoir after being passively filled. E is given by the formula,
E=∆ P∆ V
(Eq. 5)
In the context of CV hemodynamics, C is used to describe the elastic properties of
the atria, ventricles, arteries and veins under internal filling pressures.
E, on the other hand, is used to define the active internal contractile power of the
ventricle during systole; or the strength with which major elastic-walled arteries,
(especially Ao and PA) after being distended with systolic blood, squeeze passively and
create internal pressure to propel blood forward during diastole.
5
Figure 5- Elastance and Compliance in RV, LV and Ao at systole and diastole
1.7. Pressure Volume Loop
The physiologic meaning of cardiac phases, compliance and elastance can be better
observed in the content of the pressure volume loops. Pressure-volume diagram shows
simultaneous changes in volume and pressure during the cardiac cycle for the left
ventricle. It consists of four phases, which are the same as those explained previously in
section 1.2.
Phase I: Filling period (diastole). This phase starts the mitral valve opens, a
ventricular pressure value around 0 and volume is about 45-50 milliliters and it called the
end-systolic volume. Blood flows into left ventricle from the left atrium. Left ventricle
pressure (PLV) and volume (VLV) increase to about 115-120 milliliters and 5 mmHg,
respectively end-diastolic volume and pressure.
Phase II: Isovolumetric contraction period. This phase begins when the mitral valve
closes upon the rise of PLV, while the aortic valve stays closed. During contraction the
volume of the left ventricle doesn’t change as both AV and MV are closed, so this phase
called Isovolumetric contraction. On the other hand, pressure in the left ventricle
increases rapidly until pressure of the left ventricle equals the Aortic pressure (PAo) at
about 80 mmHg.
Phase III: Ejection period (systole). This phase begins when the aortic valve opened.
Blood flows into the aorta from left ventricle. During the ejection, pressure continues
rising because of ventricle still contracts. At the same time the left ventricle volume
decreases because of contraction.
6
Phase IV: Isovolumetric relaxation period. This phase begins when the aortic valve
is closed and the mitral valve is closed. The left ventricle pressure falls back to the
diastolic pressure level. During relaxation the volume of the left ventricle doesn’t change,
so this phase called Isovolumetric relaxation.
Elastance based control uses curve shown at Figure 7 . It is only experimentally derived. No method to accurately describe how it is performed.
Figure 7- Pressure-Volume Loop and Elastance-Time Curve
7
Figure 6- Volume-pressure diagram, demonstrating changes in volume and pressure during
the normal cardiac cycle.
StrokeVolume ( SV )=End Diastolic Volume ( V ed )−End Systolic Volume (V es ) (Eq. 6)
Potentatial Energy ( PE )=(Pes−Pbd)(V es−V 0)
2(Eq. 7)
StrokeWork (SW )=∫V ed
V es
PLV . d V LV −∫V es
V ed
PLV . d V LV(Eq. 8)
Figure 8- Ventricular Potential Enegry and Stroke Work Diagram
Pressure Volume Area ( PVA )=Potentatial Energy(PE)+StrokeWork (SW ) (Eq. 9)
Myocardial oxygen demand (DO2 ,myo) is a function of the pressure volume area (PVA).
As shown in the formula.
DO2 ,myo=2.46 PVA+D0 (Eq. 10)
In formula, 2.46 is a dimensionless variable find from given formula; 4
PVA=1,64 .10−5 .mLO2
mmHg. mL∧O2 [ ml ]=20 J
PVA=1,64 .10−5 .mLO2
mmHg. mL.106 mL
m3 .mmHg
133,32 Pa.
20 JmL O2
=2,46.J
Pa. m3 =2,46.JJ
D0 means energy needed for heart’s basal metabolism and D0 doesn’t dependent on
PVA. (see Figure 9)
8
1.8. Heart Disease
When the pumping function of the heart is not enough to provide the necessary flow
to the body at the required pressure, the pump is said to fail.
National Institute for Clinical Excellence “defines heart failure as a complex
syndrome that can result from any structural or functional cardiac disorder that impairs
the ability of the heart to function as a pump to support a physiological circulation.”i
Acute heart failure (AHF) is defined as the rapid or progressive beginning of
symptoms of heart failure AHF can be often cured with treatment.5
Chronic heart failure (CHF) is a long-term condition, heart muscle can’t eject
required amount of blood out of the heart over an extended period, e.g. months to years.
This is called systolic chronic heart failure. If heart muscles are stiff and doesn’t fill up
with blood (diastole), this is called diastolic chronic heart failure.6
End Stage heart disease is the final outcome of many forms of heart disease that
progressed to an end stage or an advanced form of that disease. The heart can no longer
pump enough blood to fulfill the body's metabolic requirements. 6
i National Institute for Clinical Excellence “Chronic heart failure – Management of chronic heart failure in adults in primary and secondary 2003 care.” London
9
Figure 9- Effect on the Demand Oxygen DO2 of pressure volume area (PVA)
Coronary heart disease causes the death of more than 2 million people in European
Union and 5.3 million people in Europe each year by being most important effect in all
death reasons. %50 of coronary heart disease death occurs before age 75 and %15 occurs
before age 65 in Europe.ii
Moreover, 5.2 million adults between ages 40 to 70 year of age in USA have heart
failure and every year 550.000 new cases occur. 1 million of these patients have periodic
heart attacks and half of these patients die in one year. According to the 2002 World
Health Organization Report, 11.8% of males and 10.5% of females in the world
population had troubles from heart disease. High death risk of heart disease, in 2002,
16,655,000 people died in the world.iii National Vital Statistic Report in the USA shows
that 831,000 people died from various forms of heart diseases in 2006 and, in 56,565 the
cause of death was heart failure.iv
ii Annual Report 2009 by European Hearth Networkiii The World Health Report 2003: Shaping the Future. World Health Organization.iv National Vital Statistics Reports, Deaths: Final Data for 2006. U.S. Department Of Health And Hu-man Services
10
2. Mechanical Circulatory Assistance (MCA)
Heart pumps were invented in order to keep alive CHF patients until heart
transplantation. However, donated organs are limited, but the number of patients waiting for a
heart transplant waiting is increasing with the population. Because of that, heart pumps were
designed that can implanted by providing options for patients on whom transplantation
operation can’t be performed (Bridge to Transplant). So, patients started to live with portable
mechanical cardiovascular circulation assistances (MCA) until transplantation. LVAD is a
heart pump that is Bridge to Transplant system by pumping blood from Left Ventricle to
Aorta (See Figure 10).
Figure 10- HeartMate II LVAD as Bridge to Transplant System in Patients Body7
2.1. Left Ventricular Assist Devices (LVADs)
The left ventricular assist device (LVAD) is a mechanical pump which is
implanted to help the heart’s left ventricle pump blood into the aorta. LVADs are used
in patients with end-stage heart failure.
The early VADs replicated the heart by using a "pulsatile" action where blood is
alternately sucked into the pump from the left ventricle then forced out into the aorta.
Continuous flow pumps have the advantage of simplicity resulting in smaller size
and higher safety. These devices are referred to as second generation LVADs. A side
effect is that the user will not have a pulse or that the pulse intensity will be seriously
reduced. These are second generation LVADs.
11
Third generation LVADs removed bearings in the pump using either
hydrodynamic or electromagnetic suspension or both together, and reducing the
number of moving parts to one.
Device Generation Manufacturer Type
Novacor Generation 1 World Heart Pulsatile.
HeartMate XVE Generation 1 Thoratec Pulsatile.
HeartMate II Generation 2 Thoratec Continuous axial flow
HeartMate III Generation 2 Thoratec Continuous axial flow
Jarvik 2000 Generation 2 Jarvik Heart Continuous axial flow
MicroMed De-
Bakey VADGeneration 2 MicroMed Continuous axial flow
VentrAssist Generation 3 Ventracor Continuous centrifugal flow
HVAD Generation 3 HeartWare Continuous centrifugal flow
DuraHeart Generation 3 Terumo Continuous centrifugal flow
Süha Küçükaksu et al. placed three MicroMed DeBakey VAD to three male
patients, aged 37, 41 and 40, all had end-stage left heart failure. These are the 1st
implantations of a left ventricular assist device in Turkey. 8
12
Table 1- VAD devices, generations, manufacturers and types.
Producing this LVAD in Turkey has some advantages. Because people who will
use LVAD can attend stages for medical opinion of designing and producing of
LVAD, the rate of deaths caused by using imported LVADs in a wrong way will
decrease. For testing LVAD before humans it has to be tested on Cardiovascular
Mock Circuits (CVMCs) and animals.
13
2.2. Cardiovascular Mock Circuits (CVMCs)
CVMCs are used as a mechanical representation of human cardiovascular system
in vitro. Early versions of CVMCs were built for testing artificial heart valves
CVMCs also can be used for testing of total artificial hearts, artificial lungs and many
other artificial organs. In these systems stepping motors used for create an artificial
heart beat.9
Kolff et al. (1959) 10 constructed a mock circuit which consisted of both
pulmonary and systemic sides. Compressed air used to operate ventricles, also to
obtain pressures of aortic and pulmonary arterial chambers. No resistance valves used
in this study.
Donovan et al. (1975) 11 designed and built a complete mock circulation for
testing new artificial heart designs in an artificial environment. However, Donovan et
al didn’t include atria or ventricles which limit the cardiovascular device testing
capabilities. This device used acrylic material to build a chamber, for replicate aortic,
systemic venous, pulmonary arterial, pulmonary venous and ventricular components.
Dimensions of this system are 605mm wide, 403mm high and 202mm deep. This
cardiovascular mock circuit has been the basis for many CVMCs since.
Verdonck et al. (1992) 12 designed and built a complete mock circulation for
testing mitral valves. For this reason aortic and mitral valves in the system could
change easily. The left ventricle verb made by silicone and latex atrium were both
anatomically shaped and mounted in water filled Plexiglas tanks where the pressures
are controlled by an external circuit. This CVMC was later used by Vandenberghe et
al. in 2003, to estimate the hydrodynamic performance of the PUCA II LVAD. 13
Baloa et al. (2001) 14 studied control of cardiovascular mock circuit by using the
elastance-based control. This study used the elastance of the ventricle to change
control output and develop a new strategy. The authors concluded that this design was
successful in using elastance-based control for simulating different cardiac conditions.
14
Koenig et al. (2004) 15 produced a systemic mock circulation to test pulsatile and
continuous ventricular assist devices (VADs). A mock ventricle was created of
ellipsoid shape with mounts for the aortic and mitral valves. This system was used
with a pressure chamber to replicate desired pressure values. A resistance realized by
a latex tube in the system. However, the pulmonary circulation to complement the
systemic side was not included in the system.
Pantalos et al. (2004) 16 constructed a CVMC to replicate the Frank Starling
response for several situations. In this system flexible polyurethane was used for
produce atrium and ventricle. A polyurethane aorta was connected to the ventricle.
Pressure volume loops representing natural situations were produced also this system
can be replicate the Frank Starling response for all situations. The primary limitation
of this system is there is not having a pulmonary vasculature mock circulatory system.
Gregory S.D. (2004) 17 designed a mock circuit that used compliance based
control. Means calculating pressure using compliance. Also Mock Circuit didn’t
accurately replicate the elastic nature of pressure and volume values in Figure 2.
Timms et al. (2005) 18 designed and constructed a complete CVMC for testing of
VADs. The purpose of this study was to design and construct a CVMC that was
capable of replicating several situations. This mock loop was not shown to produce
accurate pressure and flow data for several situations. Limitations in this study
involve the use of slow check valves for use in replacing the natural heart valves.
Timms et al. (2010) 19 designed and constructed a CVMC which includes both
systemic and pulmonary circulatory systems. Also simulation includes calculations of
circuit volumes, resistances, and compliances. However, the system was not
accurately replicating the human cardiovascular system. There is a lagging between
pressures of LV and Ao. Aortic valves opening and closing times may cause this
problem.
15
In CVMCs, Air/water tanks (chambers) are used to model the venous and arterial
compliances and venous, arterial, and other system flow resistances.
Water is used instead of blood. Pressures are generated by water in hydraulic
columns. Pressure is calculated with water height (h) in these columns, density (p) of
water and acceleration of gravity (g) by given the formula,
P=pgh (Eq. 11)
Qout=P−Pb
R(Eq. 12)
h=Q¿−Qout
A(Eq. 13)
Qout=pgh−Pb
R(Eq. 14)
h=Q¿−
pgh−Pb
RA
(Eq. 15)
h=Q¿−
pgh−Pb
RA
(Eq. 16)
h=R .Qi n
A− pgh
A R+
Pb
A R(Eq. 17)
h=−pghA R
+ 1A (R .Q¿+
Pb
R ) (Eq. 18)
pgA
= 1Ch
(Eq. 19)
h= −1Ch R
h+ 1A (R . Q¿+
Pb
R ) (Eq. 20)
1st order kinematics exponential decay charge.
16
Figure 11- Parameters of open air reservoirs in a single subsystem of CVMC model.
A=π r2 (Eq. 21)
γ= r (r-1)(α-1) (Eq. 22)
P=pgh+γpgH
H −h−Patm (Eq. 23)
Qout=P−Pb
R(Eq. 24)
h=Q∈−Q outA
(Eq. 25)
Vt=h . A (Eq. 26)
V=H . A (Eq. 27)
C=(V−Vt )2
K+(V−Vt )2(Eq. 28)
Figure 12- Parameters of closed air reservoirs in a single subsystem of CVMC model.
17
II. OBJECTIVES
1. Desired Functions
Goal of this project is to design and build Cardiovascular Mock Circuit which is drivable
with an elastance based feed-back control that can replicate heart’s pressure-volume relation-
ship accurately. For this goal pressure volume relationship was calculated using elastance.
Product of this project will be used for VADs tests for this purpose product must be sturdy.
Product life has to be at least 5 years. However, product has also to be cheap either. On the
other hand system must be accurate for trustable test results. That means system has to repro-
duce values of pressure, volume, flow of cardiovascular system within 5% error of pressure
and flow values. Modularity is another function desired for CVMC. System has to allow for
later improvements. Control of CVMC system has to be elastance based feed-back control for
reducing errors.
2. Design Criteria
Product life depends on material selection and design of parts. In order to choose material,
all of the criteria have to take into account. Body of chambers must be transparent and rigid.
Piston, top and bottom covers must be though, rigid and strong and leak-proof.
Low cost Materials have to be chosen. However, this criterion conflicts to first criteria
about transparency, rigidity and strength.
In order to make an accurate model and getting an accurate aortic pressure and flow
graphs, designing the right mechanical model is important. The mechanical design must be
compatible with both electrical and hydro-dynamical model of our vascular system.
For modularity, design of parts has to be simple, replaceable and standardized as much as
possible.
Elastance based control requires simultaneously and continuously pressure and flow sensor
inputs. Feed-back control changes output according to inputs, elastance and feed-back signal.
19
III. MATERIALS and METHODS
1. Cardiovascular Mock Circuit (CVMC) Simulation
1.1. Aim
Create a simulation model to replicate the function of the heart. This was to be
used to design and improve CVMC by finding dimensions of mock circuit
components such as length and diameters of pipe for suitable resistance or height and
diameters of chambers.
1.2. Method
Computer program called MATLAB/ SIMULINK (The Mathworks, Natick, MA,
U.S.A) was chosen to simulate CVMC in visual environment.
1.3. Simulink Model
A Simulink model was needed to simulate the functions and components of the
CVMC. Components of the CVMC are shown in the (Figure 13).
Figure 13- CVMC chambers and connections to each other.
20
1.4. Design
There are eight chambers for simulating different parts of cardiovascular system.
These are Venus Pool and Right Atrium (RA), Systemic Capillary (SC), Aorta (Ao),
Left Ventricle (LV), Pulmonary Venous (PV), Pulmonary Capillary (PC), Pulmonary
Artery (PA) and Right Ventricle (RV).
In the Left and Right Ventricles, simulated by pneumatic cylinders and
generate pressure and volume. These cylinders have to supply maximum pressure, in
the limitation of volume change.
Figure 14- Parameters of pneumatic cylinder in a single subsystem of CVMC model.
Six chambers of mock circuit are open to air, except Aorta which is the only
chamber close to atmosphere. Because of high pressure of Aorta chamber if it is open
air, it has to be high about 2 meters. Instead of using 2 meters high chamber, water
height can be reduced by using close chamber with pressurized air above the water
column.
To model the CVMC, subsystems for left atrium (LA), left ventricle (LV), aorta
(Ao), systemic capillaries (SC), right atrium (RA), right ventricle (RV) and pulmonary
arterial system (PA) were created. The pressure and flow output of each system was
21
calculated using the flow output from the previous subsystem and the pressure output
from the following subsystem. A schematic of CVMC model can be seen in Figure
19. Parameters of a single subsystem for open air reservoirs which are LA, SC, RA,
PA and Lungs, were shown in the Figure 15.
Figure 15 - A schematic of open air reservoirs in a single subsystem of CVMC model.
22
Figure 16- Matlab/Simulink result of open-air reservoir
Figure 17- Figure 18 - A schematic of close air reservoirs in a single subsystem of CVMC model.
-2 -1.5 -1 -0.5 0 0.5 1 1.5 20
0.2
0.4
0.6
0.8
1
1.2x 10
-5
Figure 19- Simulation Result for Compliance of Aorta Chamber
Resistance has to be replicated by mock circuit. Also Systemic vascular
resistance (SVR) and pulmonary vascular resistance (PVR) can change with a control
valve to desired resistance level.
For create a ventricular pressure, sine wave function used and signal saturated.
23
RV pressure changes between 5 mmHg to 25 mmHg. Also systole and diastole times
are known.
Figure 20- Pneumatic Check Valves Placement in Left Ventricual System
Pneumatic valves α and β controlled LV pressure so that,
P1V 1+P2 V 2=¿Constant
In systole:
˙(P2 V 2)=α ( P1−P2 ) ,(α open)0 ,(α closed )
and, in diastole:
˙(P2 V 2)=β ( Patm−P2 ) ,(β open)
0 ,( β closed )
Figure 21- Sine function values with respect to time.
24
Amplitude=20 (sin( pi
2 )−sin (0 ))sin( pi
2 )−sin ( 1501000
. pi)(Eq. 29)
Bias=−( 20(sin ( pi2 )−sin (0 ))
sin ( pi2 )−sin( 150
1000. pi)
−25) (Eq. 30)
Frequency=2π∗HR
60(Eq. 31)
Phase=2π∗1501000
(Eq. 32)
Figure 22- A Simulink schematic and sine wave function and saturation values of RV of CVMC model.
Figure 23- Result of RV pressure function designed on Simulink.
25
Figure 24- A Simulink schematic and sine wave function and saturation values of RV of CVMC model.
Figure 25- Result of LV pressure function designed on Simulink.
Figure 26- A schematic of CVMC model.
26
2. Cardiovascular Mock Circuit Virtual Design
2.1. Aim
By using the results obtained in the Simulink model of a Cardiovascular Mock Circuit
(CVMC), a new CVMC was virtually designed to produce accurate pressure and flow
diagrams. The new CVMC was required to replicate the human circulatory system.
2.2. Method
SolidWorks (Dassault Systèmes, Vélizy, France) is 3D CAD (Computer Aided
Design) program is used to design the product before production. SolidWorks was used to
design six air/water chambers and chamber for left ventricle (LV) and right ventricle
(RV).
2.2.1. The Ventricle
Figure 27- Left Ventricle Pneumatic System Design
27
2.2.2.The Chamber
The six chambers which are Venus Pool and Right Atrium (RA), Systemic
Capillary (SC), Aorta (Ao), Pulmonary Venous (PV), Pulmonary Capillary (PC) and
Pulmonary Artery (PA) represented by transparent, horizontal chambers with one hole
at the top for air transfer and three holes at the bottom of the chamber for fluid flow.
Connection between chambers was provided by using flexible transparent Tygon®
tubes.
Figure 28- Solidworks Drawing of Designed Chamber
28
3. Cardiovascular Mock Circuit Mechanical Building
By using the results obtained in the Simulink model and Solidworks drawings of a
Cardiovascular Mock Circuit (CVMC), a new CVMC was mechanically built to produce
accurate pressure and flow diagrams. The new CVMC was required to replicate the
human circulatory system.
CVMC is designed with four chambers for simplify system. However, this system can
improve to eight chamber with adding other chambers.
Figure 29- Solidworks Design of CVMC.
29
3.1. Heart Valves
3.1.1. Heart Valve Selection
For simulate heart valves St. Jude artificial heart valves was used. St. Jude mechanical
heart valve is shown in the Figure 30. Both Aortic and Mitral valves are standard
cuff-polyester types. Aortic valve (AV) tissue annulus diameter is 29 mm, Mitral valve (MV)
tissue annulus diameter is 25 mm, and material is pyrolytic carbon and woven polyester cuff.v
Figure 30- St. Jude Mechanical Heart Valve
3.1.2. Assembly of Heart Valves
Artificial heart valves are placed into polypropylene tubes. In that way mitral valve is
connected between LA and LV and Aortic valve is connected between LV and Ao. (see
Figure 31 )
Figure 31- Assembly of Heart Valves
v Ch: Hydrolic compliance.30
3.2. Check Valves
Pneumatic check valves are used for control air flow in the LV system. Check valves
needs air pressure and +24V to change status. However, for
Figure 32- Pneumatic Check Valve
Figure 33- Technical Drawing of Pneumatic Check Valves
Diastole valve, systole valve and compression valve which are shown in Figure 20, were
used as three pneumatic check valves in shown positions.
31
3.3. Pressure Sensors
Figure 35- a) Calmed® Pressure Transducers for Aort and Left Ventricle
b) Edwards™ Pressure Transducers for Pressure Tank
Figure 36- PBV connector of Pressure Transducers
Range of data voltages are measured from pressure transducers are between 0
and 5mV. That’s why voltage amplified with analog amplifiers. As a amplifier Texas
Instruments uA747C is used. Calculations of amplifier based on ideal pop-amp
formula in (Eq. 33). Amplified voltage has to be between -10V and +10V for usability
in the system and also protection of dSPACE hardware.
Figure 37- A Non-inverting Amplifier Circuit
33
a) b)
V out=V ¿+V ¿ x R2
R1
(Eq. 33)
V outmax=0.005V + 0.005 Vx 200000 Ω
150 Ω=6.672 V
Amplifier circuit (shown in Figure 38) is constructed for each pressure
transducer and output voltage is measured between 0V and +6.7V.
Figure 38- Amplifier Circuit
34
4. Cardiovascular Mock Circuit Control
4.1. Aim
CVMC requires a control mechanism. It was vital that the parameters which control
the CVMC could be easily changed.
4.2. Method
dSPACE (dSPACE GmbH, Germany) is a hardware and software package that
supports the tools for control of check valves by collected real-time pressure and flow
data from transduces and flow-meter.
4.3. Hardware Installation of dSPACE DS1104
dSPACE DS1104 R&D Controller Board upgrades PC to a powerful development
system for rapid control prototyping (shown in Figure 39). Real-Time Interface (RTI)
provides Simulink® blocks for graphical I/O configuration. Controller board provides to
configure all I/O graphically, insert the blocks into a Simulink block diagram, and gener-
ate the model code via Simulink.
Figure 39- DS1104 R&D Controller Board
dSPACE DS1104 Single-Board-Hardware has 8 analog to digital converters (ADC), 8
digital to analog converts (DAC), Digital I/O sub-d connector and serial interfaces. ADCs
are used as input connections from pressure transducers and flow-meter. DACs are used
as output connections from Matlab/Simulink to check valves.
35
Figure 40- Single Board Hardware of DS1104
4.4. Software Installation of Control Desk
For the graphical user interface (GUI) Control Desk software was used
(shown in the Figure 41). Control Desk allow user to change parameters in real time and
reading input and output values simultaneously. Also Control Desk provides data accusation.
Figure 41- Control Desk Example Screen
36
4.5. Software Installation of Matlab/Simulink
Simulink is a subprogram Matlab which is using of graphical block approach on
computing environment. dSPACE software includes Simulink blocks which are designed for
take inputs from CVMC via dSPACE to Simulink environment and gives outputs from
Simulink to CVMC.
4.5.1. Pressure Input
Pressure input is obtained from using analog amplified pressure signal in
dSPACE. Two different kinds of pressure transducers are used. Calmed® pressure
transducers are used for measure Ao and LV pressures and Edwards™ pressure
transducer is used measure for air tank pressure in CVMC. In dSPACE analog amplified
pressure signal amplified digitally by 10 times because of dSPACE’s ADC inputs
reduces signal by 10 times for circuit protection.
Digital pressure signal is filtered by “Analog Filter Design” box of Simulink for
reduce noise and increase stability. Voltage of digital pressure signal measured and
mapped by dSPACE. Measured data is compared with Drager heart monitor, as taken as
golden standard of measurement, by using MATLAB, curve fitting function and
quadratic polynomial method. After mapping results are rounded and pressure values are
obtained (see Figure 42).
Figure 42- Simulink Schematics of Pressure Transducers
37
4.5.2. Flow and Volume Input
Analog flow data is obtained by using Transonic Systems Inc. HT110 bypass
flow meter panel (see Figure 43) and Transonic Systems Inc. ultrasonic flow meter
sensor (see Figure 44). Analog signal output of flow meter is transferred from analog
output of HT110 bypass flow meter panel to computer environment via dSPACE.
Figure 43- Transonic Systems HT110 Bypass Flow Meter Panel
Figure 44- Transonic Systems Inc. Ultrasonic Flow Meter Sensor
38
Digital data which is measured by dSPACE compared with flow values which
are read from HT110 Bypass Flow Meter Panel. MATLAB, curve fitting function, linear
polynomial method is used for determine gain between flow value and digital flow data.
Flow value is obtained as a L/min unit and converted to ml/sec (SI unit) by conversion
gain. For finding volume value in chamber, flow value is integrated and summed with a
initial volume.
Figure 45- Flow meter placement on Y-connector
Figure 46- Simulink Schematics of Flow Sensor
39
4.5.3. Check Valve Control
Check valves are control by pulse signal generators. Opening time and duration
are adjusted by hand. Changing timing of these time variables is provided to change dias-
tole and systole duration, heart rate, elastance and stroke volume.
Figure 47- Simulink Schematics of Check Valves Controller
4.5.4. General Connection
Flow sensor, pressure transducers and check valves connected to each other as
shown in Figure 48.
Figure 48- Simulink Schematic of General Connection
40
Figure 49- dSPACE connections of pressure transducers, flow meter and check valves
Figure 50- Workspace, heart monitor, dSPACE board, Flow-meter panel, amplifiers and power-supply.
4.5.5. CONTROLDESK Graphic Based User Interface (GUI)
CONTROLDESK provides a graphic based user interface (GUI) for observe
Simulink system in real-time. CONTROLDESK is used for observe LV and Ao pressure,
P-V loop, elastance control signal and other control parameters.
41
IV. RESULTS
All result are obtained by using CVMC built in Bahcesehir University, Bioengineering
Research Laborotory. Four chamber CVMC shown in the Figure 51.
Figure 51- Built four chambers Mock circuit
42
Figure 52- ControlDesk GUI
The CVMC is successfuly replicated heart physiology in vitro. Pressure values of LV
and Ao and LV volume values are verifed the abiliy of CVMC regenerate physical
conditions stably and accurately.
First part of the ControlDesk/GUI (Figure 52) shows PLV (Green line) and PAo (Red
line) which are obtained by Calmed pressure transducers via dSPACE.
Second part shows P-V loop of heart, x axis is indicated LV volume and y axis is
indicated LV pressure, in real-time.
In third part in an order PLV, PAo, air-tank pressure and LV volume are numerically
showed.
At the bottom of the screen elastance based air-tank pressure and LV volume are
showed by graphically.
43
1. LV and Ao Pressures
Figure 53- Ideal PLV, PAo with Time Graph
PLV and PAo values recorded in real time and grafically displayed in Figure 54. In all
results maximum pressures are slightly elevated from 120 mmHg to 130-140 mmHg.
However, pressures can reduce by decreasing maximum air-tank pressure, means
decreasing elastance.
A)
B)
C)
Figure 54- Recorded LV and Ao Pressures
44
As you can see from Figure 54 PAo increases before AV open and also PAo
decreases rapidly after AV close. These results indicate AV leaking when it is closed.
The presence of backwards flow across a closed cardiac valve is defined as valvular
regurgitation (insufficiency).vi In this project there is Aortic insufficiency and results are
supported this finding. (See Figure 55)
Figure 55- Pressure Relationships in Aortic Insufficiency vii
2. Left Ventricular Volume
Ideal volume is shown in the Figure 56. Stroke volume (ΔV) in human normal
physiology is approximately 70-80 ml. However, in ideal pressures CVMC produces less
stroke volume than ideal (See Figure 57). That mean resistance between LV and Ao is higher
than human physiology.
Figure 56- Ideal Left Ventricular Volume Change
vi St. Jude Medical Heart Valve Division, St Jude Medical® Mechanical Heart Valve, Physician’s Manual45
Figure 57- LV Volume Change produced by CVMC
3. P-V Loop
Figure 58- Ideal P-V Loop for Healty Human
Figure 59- Recorded P-V Loop of CVMC
Ideal P-V loop for healthy human is shown in the Figure 58 and some of the recorded p-v loops shown in the Figure 59. As you can see in the recorded p-v loops IVC and IVR are not completely isovolumetric that results supports the theory about heart regurgitation (see Figure 60). Also stroke volume is not sufficient but in the Figure 59, C stroke volume is in an acceptable range.
46
III
IV
II
I
P-V Loop Stages
I -Diastole
II -IVC
III -Systole
IV-IVR
A) B) C)
Figure 60- Mitral and Aortic Valve Regurgitation vii
4. Elastance
Figure 61- Ideal Elastance Graph
In CVMC, elastance is directly proportional to P2 (LV air pressure, see Figure 20).
That is the reason control algorithm based on controlling P2.
vii Web source: ` Doppler Changes in Valvular Regurgitation` http://www.echoincontext.com/doppler02/doppler02_02.asp
47
V. DISSCUSION & CONCLUSION
CVMC has been designed to replicate pumping functions of heart and characteristics
of circulatory system.
Cardiovascular mock circulation systems provide an easy and desirable way for
testing cardiac devices during development process. For this reason system has to be portable,
adjustable and most importantly trustable.
Recreate heart pumping function with controlling compressed air in the ventricle was
successful. Results are comparable to heart functions but, system has to simulate different
heart states and failure situations.
The shapes of the PV loops which are obtained with CVMC were similar to in vivo
results. However, an increasing in volume was observed at the beginning of diastole and the
end of the systole. The problem is artificial heart valves (AV- MV) are leaking when they are
closed. This problem gives similar results with Aortic insufficiency.
49
VI. FUTURE WORK
In my master thesis I am planning to improve CVMC. Heart Valves will change with
hydraulic check valves because of static leaking during IVR and IVC. Air check valves will
change with proportional control valves for control air pressure increase and decrease
durations. Resistances will change with computer controlled proportional control valves for
simulate different conditions and diseases of heart and circulatory system. RV and other
chambers will add to CVMC. For new chambers and RV, new pressure and flow sensors will
implement to the system. New ventricle design will be build and implement into system. (see
Appendix A)
VII. APPENDIX A
50
1 Textbook of Medical Physiology 11th ed by Arthur C. Guyton, John E. Hall. Copyright at 2006 by
Elsevier Inc.
2 Scott DA, Fox JA, Cnaan A, Philip BK, Lind LJ, Palleiko MA, Stelling JM, Philip
JH. ,Resistance to fluid flow in veins, 1996
3 Kirby, B.J. (2010). Micro and Nano-scale Fluid Mechanics: Transport in Microfluidic Devices.
Cambridge University Press. ISBN 978-0521119030.
4 Suga H. Ventricular energetics. Physiol. Rev. 1990; 70: 247–77.
5Teerlink, John. Acute Heart Failure [Internet]. Version 17. Knol. 2008 Jul 28.
From:http://knol.google.com/k/john-teerlink/acute-heart-failure/3fkxugmiumnks/1.
6 Bonow RO, Mann DL, Zipes DP, Libby P. Braunwald's Heart Disease: A Textbook of
Cardiovascular Medicine. 9th ed. Philadelphia, Pa: Saunders Elsevier; 2011
7 Bartley P. Griffith, et al. Heartmate ii left ventricular assist system: from concept to first clinical
use. The Annals of Thoracic Surgery, 2001.
8 Deniz Suha Kucukaksu, First Turkish Experience with the MicroMed DeBakey VAD, (TexHeart
Inst J 2003;30:114-20)
9 Pantalos, G., et al., Characterization of an adult mock circulation for testing cardiac support
devices. ASAIO, 2004. 50(1): p. 37-46.
10 Kolff, W., Mock circulation to test pumps designed for permanent replacement of damaged
hearts. Cleveland Clinic Quarterly, 1959. 26: p. 223-226.
11 Donovan, F., Design of a hydraulic analog of the circulatory system for evaluating artificial
hearts. Artificial organs, 1975. 3(4): p. 439-449.
12 Verdonck, P., et al., Computer -controlled in vitro model of the human left heart. Med. & Biol.
Eng & Comput., 1992. 30: p. 656-659.
13 Vandenberghe, S., et al., In vitro evaluation of the PUCA II intra-arterial LVAD. The
International Journal of Artificial Organs, 2003. 26(8): p. 743-752.
14 Baloa, L., J. Boston, and J. Antaki, Elastance -based control of a mock circulatory system.
Annals of Biomedical Engineering, 2001. 29: p. 244-251.
15 Koenig, S., et al., Hemodynamic and pressure-volume responses to continuous and pulsatile
ventricular assist in an adult mock circulation. ASAIO, 2004. 50(1): p. 15 -24.
16 Pantalos, G., et al., Characterization of an adult mock circulation for testing cardiac support
devices. ASAIO, 2004. 50(1): p. 37 -46.
17 Gregory SD, Simulation and Development of a Mock Circulation Loop with Varible Compliance,
Master Thesis, Queensland University of Technology, 2009
18 Timms, D., et al., A complete mock circulation loop for the evaluation of left, right, and
biventricular assist devices. Artificial organs, 2005. 29(7): p. 564-572.