Cell Counting Kit-8 with Annexin V, FITC Assay Kit Technical Manual Cytotoxicity Assay and Apoptosis Assay General Information CCK-8 is an one-bottle solution; no premixing of components is required. CCK-8, being nonradioactive allows sensitive colorimetric assays to determine the number of viable cells in cell proliferation and cytotoxicity assays. WST-8 * is reduced by dehydrogenases in the cells giving an orange colored formazan, which is soluble in the tissue culture medium (Figure 1). The amount of formazan dye generated by dehydrogenases in cells is directly proportional to the number of living cells. Especially, since the toxicity of CCK-8 is so low, the same cells can be used for other cell based assays such as apoptosis assay and DNA fluorometric assay after the CCK-8 assay is completed. [ *Patent No. US 6,063,587 EP 0908453 JP 2757348 Canada 2,251,850] Figure 1. Mechanism of the cell viability detection with Cell Counting Kit-8 Kit Contents [AC10-01] - Cell Counting Kit-8 : 100 tests * 1 1) Cell Counting Kit-8: 1 mL bottle x 1 - Annexin V, FITC Apoptosis Detection Assays Kit : 20 Assays * 2 1) Annexin V, FITC Conjugate: 20 Assays x 1 2) PI Solution: 50 Assays x 1 3) Annexin V Binding Buffer: 50 Assays x 1 [AC10-05] - Cell Counting Kit-8 : 500 tests * 1 1) Cell Counting Kit-8: 5 mL bottle x 1 - Annexin V, FITC Apoptosis Detection Assays Kit : 100 Assays * 2 1) Annexin V, FITC Conjugate: 100 Assays x 1 2) PI Solution: 50 Assays x 2 3) Annexin V Binding Buffer: 50 Assays x 2 * 1 One test corresponds to one well on a 96-well plate. * 2 One assay corresponds to assay with cell concentration of 1 x 10 6 cells / ml. CCK-8 is stable for over one year at 0-5°C with protection from light. Annexin V, FITC Kit is stable for 6 months at 0-5°C with protection from light. Cell Counting Kit-8 - plate reader (450 nm filter) - 96-well plate - CO 2 incubator - 10 μl and 100 - 200 μl multi-channel pipettes Annexin V, FITC Apoptosis Detection Assays Kit - Adjustable pipettes - Flow cytometer or fluorescence microscope *Approximate fluorescence maximun excitation/emission: Annexin V, FITC: 494 nm / 518 nm; PI: 535 nm / 617 nm - Phosphate buffered saline (PBS) - Deionized water Storage Required Equipment and Materials 301.987.2667 1.877.987.2667 [email protected] www.dojindo.com 30 West Gude Drive, Suite 260 Rockville, MD 20850, USA Annexin V stained cells are used to indicate cell membrane changes that occur in the early stage of apoptosis. Once apoptosis is initiated, the phosphatidylserine present in the inner cell membrane migrates through the cell membrane of the lipid bilayer. Annexin V specifically binds to phosphatidylserine in the presense of protein-dependent Ca ion . By using fluorescent labeled Annexin V, the apoptotic cells can be identified using flow cytometry or fluorescence microscopy. Due to the different wavelength characteristics of FITC labeled Annexin V and PI nuclei stain, it is possible to perform double staining simultaneously and observe the stages of apoptosis in the cell membrane(Figure 2). Live Cell Apoptotic Cell Late-Apoptotic Cell Cytoplasmic Membrane Phosphatidyl serine Annexin V, FITC PI (Propidium iodide) Figure 2. Mechanism of the apoptosis detection with Annexin V, FITC The kit contains Cell Counting Kit-8 for cell proliferation assay and Annexin V, FITC Kit for apoptosis detection. The advantage of WST-8* dye is a characteristic of low toxicity to the cells, and it allows to combine cell viability assay based on WST-8 (CellCounting Kit-8) and apoptosis assay (Annexin V, FITC Kit). This combinational assay leads two sets of data from the same cell sample after treating the cells. The result gives not only viability change, but also it gives the cause of cell viability change that can be from the dehydrogenase activity loss or presence of apooptotic cells. [ *Patent No. US 6,063,587 EP 0908453 JP 2757348 Canada 2,251,850]
Transcript
Page 1: Cell Toxicity

Cell Counting Kit-8 with Annexin V, FITC Assay Kit

Technical Manual

Cytotoxicity Assay and Apoptosis Assay

General Information

CCK-8 is an one-bottle solution; no premixing of components is required. CCK-8, being nonradioactive allows sensitive colorimetric assays to determine the number of viable cells in cell proliferation and cytotoxicity assays. WST-8* is reduced by dehydrogenases in the cells giving an orange colored formazan, which is soluble in the tissue culture medium (Figure 1). The amount of formazan dye generated by dehydrogenases in cells is directly proportional to the number of living cells. Especially, since the toxicity of CCK-8 is so low, the same cells can be used for other cell based assays such as apoptosis assay and DNA fluorometric assay after the CCK-8 assay is completed.[*Patent No. US 6,063,587 EP 0908453 JP 2757348 Canada 2,251,850]

Figure 1. Mechanism of the cell viability detection with Cell Counting Kit-8

Kit Contents[AC10-01]

- Cell Counting Kit-8 : 100 tests *1

1) Cell Counting Kit-8: 1 mL bottle x 1- Annexin V, FITC Apoptosis Detection Assays Kit : 20 Assays *2

1) Annexin V, FITC Conjugate: 20 Assays x 12) PI Solution: 50 Assays x 13) Annexin V Binding Buffer: 50 Assays x 1

[AC10-05]- Cell Counting Kit-8 : 500 tests *1

1) Cell Counting Kit-8: 5 mL bottle x 1- Annexin V, FITC Apoptosis Detection Assays Kit : 100 Assays *2

1) Annexin V, FITC Conjugate: 100 Assays x 12) PI Solution: 50 Assays x 23) Annexin V Binding Buffer: 50 Assays x 2

*1 One test corresponds to one well on a 96-well plate. *2 One assay corresponds to assay with cell concentration of 1 x 106 cells / ml.

CCK-8 is stable for over one year at 0-5°C with protection from light.Annexin V, FITC Kit is stable for 6 months at 0-5°C with protection from light.

Cell Counting Kit-8- plate reader (450 nm filter) - 96-well plate- CO2 incubator- 10 μl and 100 - 200 μl multi-channel pipettes

Annexin V, FITC Apoptosis Detection Assays Kit- Adjustable pipettes- Flow cytometer or fluorescence microscope *Approximate fluorescence maximun excitation/emission: Annexin V, FITC: 494 nm / 518 nm; PI: 535 nm / 617 nm - Phosphate buffered saline (PBS)- Deionized water

Storage

Required Equipment and Materials

301.987.2667 1.877.987.2667

[email protected] 30 West Gude Drive, Suite 260

Rockville, MD 20850, USA

Annexin V stained cells are used to indicate cell membrane changes that occur in the early stage of apoptosis. Once apoptosis is initiated, the phosphatidylserine present in the inner cell membrane migrates through the cell membrane of the lipid bilayer. Annexin V specifically binds to phosphatidylserine in the presense of protein-dependent Ca ion. By using fluorescent labeled Annexin V, the apoptotic cells can be identified using flow cytometry or fluorescence microscopy. Due to the different wavelength characteristics of FITC labeled Annexin V and PI nuclei stain, it is possible to perform double staining simultaneously and observe the stages of apoptosis in the cell membrane(Figure 2).

Live Cell Apoptotic Cell Late-Apoptotic Cell

CytoplasmicMembrane

Phosphatidyl serineAnnexin V, FITC

PI(Propidium iodide)

Figure 2. Mechanism of the apoptosis detection with Annexin V, FITCThe kit contains Cell Counting Kit-8 for cell proliferation assay and Annexin V, FITC Kit for

apoptosis detection. The advantage of WST-8* dye is a characteristic of low toxicity to the cells, and it allows to combine cell viability assay based on WST-8 (CellCounting Kit-8) and apoptosis assay (Annexin V, FITC Kit). This combinational assay leads two sets of data from the same cell sample after treating the cells. The result gives not only viability change, but also it gives the cause of cell viability change that can be from the dehydrogenase activity loss or presence of apooptotic cells. [*Patent No. US 6,063,587 EP 0908453 JP 2757348 Canada 2,251,850]

Page 2: Cell Toxicity

0

2,000

4,000

6,000

8,000

10,000

0 0.008 0.04 0.2 1Num

ber o

f Apo

ptos

is Ce

lls

Paclitaxel (μmol/l)

Annexin V Assay after CCK-8 AssayAnnexin V Assay (witout CCK-8 Assay)

0

1,000

2,000

3,000

4,000

5,000

6,000

0 0.1 0.2 0.3 0.4 0.5Num

ber o

f Apo

ptos

is Ce

lls

Staurosporine (μg/ml)

Annexin V Assay after CCK-8 AssayAnnexin V Assay (without CCK-8 Assay)

30 West Gude Drive, Suite 260 Rockville, MD 20850, USA

301.987.2667 1.877.987.2667

[email protected]

If you require assistance, please contact Dojindo technical service.

For Research Use Only.Not for Use in Diagnostic Purpose

Cell Counting Kit-8 with Annexin V, FITC AssayRevised, Nov. 5 2012

General Protocol

Cell Counting Kit-8 Assay and Annexin V, FITC Assay are independent assay from each other.When performing Cell Counting Kit-8 Assay and Annexin V, FITC Assay, please follow the following general protocol.

1. Please use own technical manual for each assay.2. Cell Counting Kit-8 assay should be done before performing Annexin V, FITC assay.3. For Cell Counting Kit-8 Assay, do not use SDS or HCl which are used to stop the

color development reaction.4. Please run Annexin V, FITC assay immediately after Cell Counting Kit-8 assay.

Precautions

Data

Continuous culture is possible without killing cells.

Figure 3. Observation of HeLa cells at 24 hours after the addition of WST-8(CCK-8).

Cells after the cell proliferation assay were used for induction of apoptosis (dark line) and cells without performing the cell proliferation assay were used for induction of apoptosis (light line). As a result, similar readings were observed, which confirms that the same cells can be reused for induction of apoptosis after the cell proliferation assay.

Figure 4. The number of viable cells after induction of apoptosis.- Cells: Jurkat cell (1x105 cells/well)- Induction of Apoptosis: Staurosporine 3.5 hours (5%CO2, 37 °C)- After the induction of apoptosis, cell proliferation was measured using Cell Counting Kit-8.

Figure 5. The number of viable cells after induction of apoptosis.- Cells: MDA-MB-231 cell (2x105 cells/well)- Induction of Apoptosis: Paclitaxel (μmol/l) 4 days (5%CO2, 37 °C)- After the induction of apoptosis, cell proliferation was measured using Cell Counting Kit-8.

Cell Counting Kit-8 Assay for Cell Proliferation*Please see the technical manual of "Cell Counting Kit-8".

After measuring the absorbance with the microplate reader,use the same plate(cells) to perform Annexin V, FITC Assay immediately.The medium with Cell Counting Kit-8 should have orange coloring.

Annexin V, FITC Assay for Apoptosis Detection*Please start from step 1. of Annexin V Staining General Protocol

on the technical manual of "Annexin V, FITC Apoptosis Detection Kit".


Recommended