This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory quality management system (refer to joint ISO-ILAC-IAF Communiqué dated January 2009). CERTIFICATE OF ACCREDITATION ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044 This is to certify that Micro Precision Calibration Korea 775 Gyeongin-ro Yeongdeungpo-gu Seoul, Korea 150-972 has been assessed by ANAB and meets the requirements of international standard ISO/IEC 17025:2005 and national standards ANSI/NCSL Z540-1-1994 AND ANSI/NCSL Z540.3-2006 while demonstrating technical competence in the field of CALIBRATION Refer to the accompanying Scope of Accreditation for information regarding the types of calibrations to which this accreditation applies. AC-1969.13 Certificate Number Certificate Valid: 10/21/2016-10/31/2018 Version No. 001 Issued: 10/21/2016
Transcript
Page 1: CERTIFICATE OF ACCREDITATION · 2016. 12. 1. · This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates

This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005.

This accreditation demonstrates technical competence for a defined scope and the operation of a laboratory

quality management system (refer to joint ISO-ILAC-IAF Communiqué dated January 2009).

CERTIFICATE OF ACCREDITATION

ANSI-ASQ National Accreditation Board 500 Montgomery Street, Suite 625, Alexandria, VA 22314, 877-344-3044

This is to certify that

Micro Precision Calibration Korea 775 Gyeongin-ro Yeongdeungpo-gu

Seoul, Korea 150-972

has been assessed by ANAB and meets the requirements of international standard

ISO/IEC 17025:2005 and national standards

ANSI/NCSL Z540-1-1994 AND

ANSI/NCSL Z540.3-2006

while demonstrating technical competence in the field of

CALIBRATION

Refer to the accompanying Scope of Accreditation for information regarding the types of

calibrations to which this accreditation applies.

AC-1969.13

Certificate Number

Certificate Valid: 10/21/2016-10/31/2018 Version No. 001 Issued: 10/21/2016

Page 2: CERTIFICATE OF ACCREDITATION · 2016. 12. 1. · This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates

Version 001 Issued: 10/21/2016 Page 1 of 12

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005,

ANSI/NCSL Z540-1-1994, & ANSI/NCSL Z540.3-2006

Micro Precision Calibration Korea 775 Gyeongin-ro Yeongdeungpo-gu,

Seoul, Korea 150-972 Jeong-Hyun Park Phone: +82 70 7826 4450

[email protected] www.microprecision.com

CALIBRATION

Valid to: October 31, 2018 Certificate Number: AC-1969.13

I. Electromagnetic - DC/Low Frequency

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty (±)]

Reference Standard

or Equipment

DC Voltage - Source3

Up to 220 mV

220 mV to 2.2 V

(2.2 to 11) V

(11 to 22) V

(22 to 220) V

220 V to 1.1 kV

10 µV/V + 0.9 µV

9 µV/V + 1.4 µV

9 µV/V + 4.6 µV

9 µV/V + 9.1 µV

10 µV/V + 0.12 mV

12 µV/V + 0.68 mV

Fluke 5700A Opt 03

DC Voltage – Source3

Fixed Values

1 V

1.018V

10 V

1.6 µV/V

1.6 µV/V

1.2 µV/V

Fluke 732A

DC Voltage3 – Measure

Up to 100 mV

100 mV to 1V

(1 to 10) V

(10 to 100) V

100 V to 1 kV

(1 to 10) kV

(10 to 20) kV

15 µV/V

21 µV/V

10 µV/V

12 µV/V

12 µV/V

1 mV/V + 4 V

4 mV/V

HP 3458A Opt 002

Vitrek 4700

Page 3: CERTIFICATE OF ACCREDITATION · 2016. 12. 1. · This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates

Version 001 Issued: 10/21/2016 Page 2 of 12

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty (±)]

Reference Standard

or Equipment

DC Current3 – Source

Up to 220 µA

220 µA to 2.2 mA

(2.2 to 22) mA

(22 to 220) mA

220 mA to 2.2 A

Up to 330 A

330 A to 3.3 mA

(3.3 to 33) mA

(33 to 330) mA

330 mA to 3 A

(3 to 10) A

(1 to 1 000) A

68 µA/A + 11 nA

35 µ A/A + 11 nA

35 µA/A + 0.12 µA

40 µA/A + 1.1 µA

55 µA/A + 34 µA

0.15 mA/A + 20 nA

0.1 mA/A + 50 nA

0.1 mA/A + 0.25 µA

0.1 mA/A + 2.5 µA

0.38 mA/A + 40 µA

1.7 mA/A

2.6 mA/A

Fluke 5700A Opt 03

Fluke 5500A

Fluke 5500A

with 50 turn Coil

DC Current3 - Measure

Up to 100 nA

100 nA to 1 A

(1 to 10) A

(10 to 100) A

100 A to 10 mA

(10 to 100) mA

100 mA to 1 A

0.49 mA/A

68 mA/A

34 mA/A

36 mA/A

30 mA/A

46 mA/A

1.4 mA/A

HP 3458A Opt 002

DC High Current3 –

Measure

(1 to 2) A

(2 to 10) A

(10 to 100) A

0.1 mA/A + 3 µA

0.3 mA/A + 30 µA

0.3 mA/A + 3 µA

1625 Shunt monitored

with Multimeter HP

3458A

Inductance3 - Source

Fixed Points

at 1 kHz

100 µH

1 mH

10 mH

100 mH

1 H

10 H

0.25 µH/H

0.18 µH/H

0.6 µH/H

0.18 µH/H

0.75 µH/H

0.32 µH/H

GR Standard Inductors

Model 1482 Series

Inductance - Measure3

12 Hz to 10 kHz 100 H to 10 H

12 Hz to 10 kHz 1.2 mH/H

GenRad 1689

Impedance Bridge

Resistance3 - Source

Up to 11 Ω

(11 to 33) Ω

(33 to 110) Ω

(110 to 330) Ω

330 Ω to 1.1 kΩ

(1.1 to 3.3) kΩ

(3.3 to 11) kΩ

(11 to 33) kΩ

(33 to 110) kΩ

(110 to 330) kΩ

330 kΩ to 1.1 MΩ

0.1 mΩ/Ω + 8 mΩ

0.1 mΩ/Ω + 15 mΩ

0.1 mΩ/Ω + 15 mΩ

0.1 mΩ/Ω + 15 mΩ

0.1 mΩ/Ω + 15 mΩ

0.1 mΩ/Ω + 60 mΩ

0.1 mΩ/Ω + 60 mΩ

0.1 mΩ/Ω + 0.6 Ω

0.1 mΩ/Ω + 0.6 Ω

0.1 mΩ/Ω + 6 Ω

0.2 mΩ/Ω + 55 Ω

Fluke 5500A

Page 4: CERTIFICATE OF ACCREDITATION · 2016. 12. 1. · This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates

Version 001 Issued: 10/21/2016 Page 3 of 12

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty(±)]

Reference Standard

or Equipment

Resistance3 – Source

(Cont.)

(1.1 to 3.3) MΩ

(3.3 to 11) MΩ

(11 to 33) MΩ

(33 to 110) MΩ

(110 to 330) MΩ

0.2 mΩ/Ω + 55 Ω

0.6 mΩ/Ω + 0.55 kΩ

1 mΩ/Ω + 0.55 kΩ

5.1 mΩ/Ω + 5.5 kΩ

5.2 mΩ/Ω + 17 kΩ

Fluke 5500A

Resistance3 - Source

Fixed Points

1.0 Ω

1.9 Ω

10 Ω

19 Ω

100 Ω

190 Ω

1 kΩ

1.9 kΩ

10 kΩ

19 kΩ

100 kΩ

190 kΩ

1 MΩ

1.9 MΩ

10 MΩ

19 MΩ

100 MΩ

0.13 mΩ

0.13 mΩ

37 µΩ

35 µΩ

23 µΩ

23 µΩ

17 mΩ

17 mΩ

16 mΩ

16 mΩ

18 mΩ

18 mΩ

26 Ω

27 Ω

52 Ω

62 Ω

0.15 kΩ

Fluke 5700A

w/option 03

Resistance3 - Measure

(0 to 10) Ω

(10 to 100) Ω

100 to 100 kΩ

100 k to 1 M

(1 to 10) MΩ

(10 to 100) MΩ

100 MΩ to 1 GΩ

23 µΩ/Ω

20 µΩ/Ω

12 µΩ/Ω

21 µΩ/Ω

71 µΩ/Ω

0.58 mΩ/Ω

5.7 mΩ/Ω

HP 3458A Opt 002

AC Resistance -

Source and Measure3

100 Hz to 10 kHz

Up to 200 mΩ

200 mΩ to 25 Ω

(25 to 400) Ω

400 Ω to 6 kΩ

(6 to 100) kΩ

100 kΩ to 10 MΩ

0.87 mΩ

5.6 mΩ

90 mΩ

1.4 Ω

23 Ω

30 kΩ

Genrad 1689 with

AC Resistors

Capacitance3 - Source

at 1 kHz

(330 to 490) pF

490 pF to 1.1 nF

(1.1 to 3.3) nF

(3.3 to 11) nF

(11 to 33) nF

(33 to 110) nF

(110 to 330) nF

330 nF to 1.1 µF

(1.1 to 3.3) µF

(3.3 to 11) µF

5.9 pF/F + 11 pF

6.2 pF/F + 11 pF

16 pF/F + 15 pF

16 pF/F + 11 pF

0.15 nF/F + 0.11 nF

0.15 nF/F + 0.11 nF

7.3 pF/F + 0.34 nF

3.2 pF/F + 1.1 nF

5.2 pF/F + 3.4 nF

4.3 pF/F + 11 nF

Fluke 5500A

Page 5: CERTIFICATE OF ACCREDITATION · 2016. 12. 1. · This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates

Version 001 Issued: 10/21/2016 Page 4 of 12

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty(±)]

Reference Standard

or Equipment

Capacitance3 - Source

at 1 kHz

(Cont.)

(11 to 33) µF

(33 to 110) µF

(110 to 330) µF

330 µF to 1.1 mF

4.8 pF/F + 34 nF

8.2 pF/F + 0.12 µF

0.54 µF/F + 0.34 µF

11 µF/F + 0.34 µF

Fluke 5500A

Capacitance -

Measure 3

12 Hz to 100 kHz

Up to 10 pF

(10 to 100) pF

(100 to 400) pF

400 pF to 1.6 nF

(1.6 to 6.4) nF

6.4 pF to 25 nF

(25 to 100) nF

(100 to 400) nF

400 nF to 1.6 µF

(1.6 to 6) µF

(6 to 25) µF

(25 to 100) µF

100 µF to 1 mF

46 nF

57 nF

91 nF

0.36 pF

1.4 pF

5.7 pF

23 pF

91 pF

0.36 nF

1.4 nF

5.7 nF

34 nF

2.4 µF

GenRad 1689

Electrical Stimulation

of Thermocouples3

Type E

Type J

Type K

Type S

Type T

(-250 to -100) ºC

(-100 to 650) ºC

(650 to 1 000) ºC

(-210 to -100) ºC

(-100 to 760) ºC

(760 to 1 200) ºC

(-200 to -100) ºC

(-100 to 120) ºC

(120 to 1 000) ºC

(1 000 to 1 372) ºC

(0 to 250) ºC

(250 to 1 400) ºC

(1 400 to 1 767) ºC

(-250 to -150) ºC

(-150 to 0) ºC

(0 to 400) ºC

0.50 ºC

0.17 ºC

0.21 ºC

0.27 ºC

0.17 ºC

0.23 ºC

0.50 ºC

0.43 ºC

0.46 ºC

0.46 ºC

0.45 ºC

0.47 ºC

0.55 ºC

0.63 ºC

0.58 ºC

0.53 ºC

Fluke 5500A

Page 6: CERTIFICATE OF ACCREDITATION · 2016. 12. 1. · This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates

Version 001 Issued: 10/21/2016 Page 5 of 12

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty(±)]

Reference Standard

or Equipment

Electrical Calibration of

RTDs3

Pt 385, 100 Ω

Pt 3926, 100 Ω

Pt 3916, 100 Ω

Pt 385, 200 Ω

Pt 385, 500 Ω

Pt 385, 1 kΩ

PtNi 385, 100 Ω

Cu 427, 10 Ω

(-200 to 0) °C

(0 to 100) °C

(100 to 400) °C

(400 to 630) °C

(630 to 800) °C

(-200 to 0) °C

(0 to 100) °C

(100 to 400) °C

(400 to 630) °C

(-200 to -190) °C

(-190 to 0) °C

(0 to 300) °C

(300 to 600) °C

(600 to 630) °C

(-200 to 100) °C

(100 to 260) °C

(260 to 600) °C

(600 to 630) °C

(-200 to 100) °C

(100 to 260) °C

(260 to 600) °C

(600 to 630) °C

(-200 to 0) °C

(0 to 260) °C

(260 to 600) °C

(600 to 630) °C

(-80 to 100) °C

(100 to 260) °C

(-100 to 260) °C

0.08 °C

0.08 °C

0.14 °C

0.18 °C

0.25 °C

0.05 °C

0.07 °C

0.1 °C

0.12 °C

0.25 °C

0.05 °C

0.08 °C

0.1 °C

0.23 °C

0.04 °C

0.05 °C

0.14 °C

0.16 °C

0.05 °C

0.06 °C

0.09 °C

0.11 °C

0.03 °C

0.05 °C

0.07 °C

0.23 °C

0.08 °C

0.14 °C

0.3 °C

Fluke 5500A

Page 7: CERTIFICATE OF ACCREDITATION · 2016. 12. 1. · This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates

Version 001 Issued: 10/21/2016 Page 6 of 12

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty(±)]

Reference Standard

or Equipment

AC Voltage3 - Source

Up to 220 mV

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

220 mV to 2.2 V

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

(2.2 to 22) V

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100 kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

(22 to 220) V

(10 to 20) Hz

(20 to 40) Hz

40 Hz to 20 kHz

(20 to 50) kHz

(50 to 100 kHz

(100 to 300) kHz

(300 to 500) kHz

500 kHz to 1 MHz

220 V to 1.1 kV (15 to 50) Hz

50 Hz to 1 kHz

0.68 mV/V + 18 µV

0.27 mV/V + 11 µV

0.12 mV/V + 11 µV

0.41 mV/V + 11 µV

1 mV/V + 34 µV

1.2 mV/V + 34 µV

2 mV/V + 45 µV

4.1 mV/V + 0.12 mV

0.68 mV/V + 0.12 mV

0.2 mV/V + 34 µV

0.1 mV/V + 8 µV

0.16 mV/V + 23 µV

0.32 mV/V + 90 µV

0.54 mV/V + 0.17 mV

1.4 mV/V + 0.46 mV

2.7 mV/V + 1.1 mV

0.68 mV/V + 1.2 μV

0.2 mV/V + 0.34 mV

0.1 mV/V + 0.08 mV

0.16 mV/V + 0.23 mV

0.32 mV/V + 0.45 mV

0.68mV/V + 1.9 mV

1.6 mV/V + 5.6 mV

3.4 mV/V + 11 mV

0.7 mV/V + 11 mV

0.2 mV/V + 3.4 mV

0.1 mV/V + 1.1 mV

0.28 mV/V + 4.5 mV

0.7 mV/V + 11 mV

1.8 mV/V + 0.13 V

6.1 mV/V + 0.13 V

15 mV/V + 0.25 V

2.3 mV/V + 23 mV

0.1 mV/V + 4.5 mV

Fluke 5700A Opt 03

Page 8: CERTIFICATE OF ACCREDITATION · 2016. 12. 1. · This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates

Version 001 Issued: 10/21/2016 Page 7 of 12

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty(±)]

Reference Standard

or Equipment

AC Voltage3 - Measure

Up to 10 mV

(1 to 40) Hz

40 Hz to 1 kHz

(1 to 20) kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

10 mV to 10 V

(1 to 40) Hz

40 Hz to 1 kHz

(1 to 20) kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

300 kHz to 1 MHz

(1 to 2) MHz

(10 to 100) V

(1 to 40) Hz

40 Hz to 1 kHz

(1 to 20) kHz

(20 to 50) kHz

(50 to 100) kHz

(100 to 300) kHz

300 kHz to 1 MHz

100 V to 1 kV

(1 to 40) Hz

40 Hz to 1 kHz

(1 to 20) kHz

(20 to 50) kHz

(50 to 100) kHz

(1 to 20) kV

(50 to 60) Hz

0.68 mV/V

0.35 mV/V

0.46 mV/V

1.3 mV/V

5.8 mV/V

46 mV/V

0.16 mV/V

0.12 mV/V

0.2 mV/V

0.36 mV/V

0.98 mV/V

4.8 mV/V

12 mV/V

17 mV/V

0.27 mV/V

0.26 mV/V

0.25 mV/V

0.42 mV/V

1.4 mV/V

4.7 mV/V

17 mV/V

0.52 mV/V

0.49 mV/V

0.71 mV/V

1.4 mV/V

3.5 mV/V

9 mV/V

HP 3458A Opt 002

Vitrek 4700

AC Current3 - Source

Up to 220 μA

40 Hz to 1 kHz

220 μA to 22 mA

40 Hz to 1 kHz

(22 to 220) mA

40 Hz to 1 kHz

220 mA to 2.2 A

40 Hz to 1 kHz

(1 to 1 000) A

60 Hz

0.9 mA/A

0.24 mA/A

0.26 mA/A

0.93 mA/A

5.4 mA/A

Fluke 5720A Opt 03

Fluke 5500A

with Coil

AC Current3 – Measure

(5 to 100) µA

(10 to 20) Hz

(20 to 45) Hz

45 Hz to 5 kHz

4.9 mA/A

2.1 mA/A

1.1 mA/A

HP 3458A Opt 002

Page 9: CERTIFICATE OF ACCREDITATION · 2016. 12. 1. · This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates

Version 001 Issued: 10/21/2016 Page 8 of 12

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty(±)]

Reference Standard

or Equipment

AC Current3 – Measure

(Cont.)

100 µA to 1 mA

(10 to 20) Hz

(20 to 45) Hz

(45 to 100) Hz

100 Hz to 5 kHz

(5 to 20) kHz

(20 to 50) kHz

(50 to 100) kHz

100 mA to 1 A

(10 to 20) Hz

(20 to 45) Hz

(45 to 100) Hz

100 Hz to 5 kHz

(5 to 20) kHz

(20 to 50) kHz

(1 to 2) A

45 Hz to 1 kHz

(2 to 20) A

45 Hz to 1 kHz

(20 to 100) A

45 Hz to 1 kHz

4.8 mA/A

1.9 mA/A

0.91 mA/A

0.57 mA/A

0.91 mA/A

5.0 mA/A

7.9 mA/A

4.8 mA/A

2.1 mA/A

1.2 mA/A

1.4 mA/A

3.6 mA/A

12 mA/A

1.3 mA/A + 0.15 mA

1.3 mA/A + 1.5 mA

1.3 mA/A + 15 mA

HP 3458A Opt 002

Ballantine 1625 w

3458a w 50 Turn Coil

Oscilloscopes3

Leveled Sine Wave

50 kHz Reference

Leveled Sine Wave

Flatness

5 mV to 5.5 V

Relative to 50 kHz

Square Wave

1 M

50

10 Hz to 10 kHz

Time Marker Output

Into 50

Pulse Rise Time

(10, 25, 100, 250) mV

5 mV to 5 V (p-p)

50 kHz to 100 MHz

(100 to 300) MHz

(300 to 600) MHz

1.8 mV to 55 V p-p

1.8 mV to 2.5 V p-p

2 ns to 20 ms

50 ms to 5 s

20 ms to 200 ns

(50 Hz to 5 MHz)

2.3 % + 340 µV

1.8 % + 120 µV

2.6 % + 120 µV

4.7 % + 120 µV

3.4 % + 120 µV

3.4 % + 120 µV

2.8 parts in 106

0.11 %

2.9 x 10-6

Fluke 5500A/SC600

Page 10: CERTIFICATE OF ACCREDITATION · 2016. 12. 1. · This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates

Version 001 Issued: 10/21/2016 Page 9 of 12

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org

II. Electromagnetic – RF/Microwave

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty(±)]

Reference

Standard or

Equipment

RF Tuned Power

Relative3 – Measure

2.5 MHz - 26.5 GHz

(-0.0 to -10) dB

(-10 to -20) dB

(-20 to -30) dB

(-30 to -40) db

(-40 to -50) dB

(-50 to -60) dB

(-60 to -70) dB

(-70 to -80) dB

(- 80 to -90) dB

(-90 to -100) dB

(-100 to -110) dB

(-110 to -120) dB

0.06 dB

0.08 dB

0.09 dB

0.12 dB

0.15 dB

0.17 dB

0.20 dB

0.25 dB

0.25 dB

0.27 dB

0.32 dB

0.37 dB

HP 8902A w/

HP 11793A

RF Power – Absolute3

Measure

100 kHz to 50 GHz

(-70 to -30) dBm

(-30 to -20) dBm

(-20 to 0) dBm

(0 to 20) dBm

0.64 dBm

0.06 dBm

0.068 dBm

1.2 %

HP 438A, HP 8902A,

E4419A, HP 11722A,

HP 8482A, HP 8484A,

HP 8481A, HP 8485A,

HP 8487A

RF Power3 - Source

(-127 to 13) dBm

(-110 to 10) dBm

10 kHz to 1 GHz

10 MHz to 2 GHz

(2 to 50.0) GHz

1.2 dB

1.2 dB

1.3 dB

HP 8657D

HP 83650L

Amplitude Modulation3 –

Measure

Rate:

50 Hz to 10kHz

20 Hz to 100kHz

Depth: (5 to 99) %

Rate:

50 Hz to 50 kHz

20 Hz to100kHz

Depth: (5 to 99) %

Rate:

20 Hz to100kHz

Depth: (5 to 99)%

150 kHz to 10 MHz

10 MHz to 1.3 GHz

(1.3 to 26.5) GHz

2.5 %

3.6 %

1.6 %

3.6 %

3.2%

HP 8902A

HP 8902A w/

11793A

Harmonic Measurements3 30 Hz to 50 GHz 1.5 dB HP 8565E

Harmonic3 - Distortion 20 Hz to 20 kHz

(20 to 100) kHz

1.5 dB

2.5 dB HP 8903B

Page 11: CERTIFICATE OF ACCREDITATION · 2016. 12. 1. · This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates

Version 001 Issued: 10/21/2016 Page 10 of 12

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty(±)]

Reference

Standard or

Equipment

Frequency Modulation3 -

Measure

Rate:

20Hz to 10kHz

Dev: 40 kHz

Rate:

20 Hz to200kHz

50 Hz to100kHz

Dev: 400 kHz

Rate:

20 Hz to100kHz

50 Hz to100kHz

250 kHz to 10 MHz

10 MHz to 1.3 GHz

(1.3 to 26.5) GHz

2.4 %

7.7 %

1.3 %

7.7 %

1.3 %

HP 8902A

HP 8902A w/

11793A

Harmonic3 - Distortion 20 Hz to 20 kHz

(20 to 100) kHz

1.5 dB

2.5 dB HP 8903B

Digital Modulation3-

Source and Measure

EVM

Power Ramp Relative

Accuracy

Phase and Frequency

Error:

Phase Error

Peak Phase Error

Frequency Error

Output RF Spectrum

Relative RF Power

Uncertainty:

Due to Modulation

Offsets≤1.2 MHz

Offsets≥1.8MHz

Due to Switching

Absolute RF Power

Accuracy

- 0.11 ± 0.60 dB

-0.11 ± 0.60 dB

(1 to 15)°

(3 to 25)°

(18 to 30)°

(18 to 30)°

0.26 dB

0.50 °

2.0 °

5.0 Hz + tfa

0.26 dB

0.36 dB

0.27 dB

0.40 dB

Agilent E4436B &

Agilent E4440A

tfa = time frequency

analysis

Page 12: CERTIFICATE OF ACCREDITATION · 2016. 12. 1. · This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates

Version 001 Issued: 10/21/2016 Page 11 of 12

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org

III. Optical Radiation

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty(±)]

Reference Standard

or Equipment

Optical Power3 –

Measure

850 nm

1 310 nm

1 550 nm

(-6 to -60) dB

(10 to -110) dB

(10 to -110) dB

2.5 %

2.5 %

2.5 %

Agilent Optical Test

System

Optical Wavelength3 -

Measure (700 to 1 650) nm 0.007 nm Agilent 86120B

IV. Time & Frequency

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty(±)]

Reference Standard

or Equipment

Frequency3 - Source

0.01 Hz to 225MHz

10 MHz to 26.5 GHz

2.3 x 1011

1.2 x 107

GPS, HP 58503B

Hp 3325B

GPS, HP 58503B

Agilent 83650L

Frequency3 - Measure

Fixed Point

10 MHz

0.1 Hz to 225 MHz

100 MHz to 5 GHz

500 MHz to 26.5 GHz

5 parts in 1012

2.3 x 1011

2.3 x 1011

1.4 x 1011

GPS, HP 58503B

GPS, HP 58503B

Agilent 53132A

GPS, HP 58503B

Agilent 5361B

Photo Tachometer3 (0 to 100 000) rpm 0.8 rpm Generator with LED

V. Thermodynamics

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty(±)]

Reference Standard

or Equipment

Temperature3 (-40 to 420) ºC 0.24 ºC Hart SPRT

Relative Humidity3 (15 to 78) % RH 1.5 % RH Vaisala HMP 141/HM

70 Humidity Standard

Page 13: CERTIFICATE OF ACCREDITATION · 2016. 12. 1. · This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2005. This accreditation demonstrates

Version 001 Issued: 10/21/2016 Page 12 of 12

500 Montgomery St. Suite 625 Alexandria, VA 22314703-836-0025 www.anab.org

VI. Mechanical

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty(±)]

Reference Standard

or Equipment

Pressure 3

(-14 to 200) psi

Up to 1 000 psi

Up to 10 000 psi

0.054 %

0.052 %

0.08 %

Fluke Pressure

Calibrator

Scales and Balances3

(1 to 200) mg

(0.2 to 200) g

Up to 5 kg

(5 to 200) kg

0.11 mg

0.14 mg

23 mg

28 mg

Class 1 and F Weights

Torque Tools3 (0 to 10) N·m

(10 to 675) N·m

0.30 %

0.20 % Torque System

VII. Dimensional

Parameter/

Equipment Range

Calibration and

Measurement

Capability

[Expressed as

Uncertainty(±)]

Reference Standard

or Equipment

Calipers3

(0 to 900) mm

20 µm

Gage Blocks

Height Gages3

(0 to 900) mm

8.6 µm

Indicators3 (0 to 100) mm (0.7 + 0.1L) µm Gage Blocks,

Height Stand

Micrometers3 Up to 300 mm

(1.4 + 0.1L) µm

Gage Blocks

Notes:

1. Calibration and Measurement Capabilities (CMC) (Expanded Uncertainties) are based on approximately a 95% confidence interval, using

a coverage of k=2. 2. This laboratory provides calibration services in its laboratory and on-site at customer-designated locations. Since on-site conditions are

typically more variable than those in the laboratory, larger measurement uncertainties are expected on-site than what is reported on the

scope of accreditation. 3. On-site calibration is offered for this parameter.

4. The use of (L) stands for Length in inches.

6. The use of (t) is an expression of Time in seconds.

8. This scope is formatted as part of a single document including the Certificate of Accreditation No. AC-1969.13