Certificate of Analysis CBD CRYSTALLINE 99.6% CBD CRYSTALLINE 99.6% Matrix: N/A Page 1 of 6 BMH Ventures, Inc. 2900 Gateway Drive, Suite A Pompano Beach FL, USA 33069 (954) 802-8826 [email protected] SAMPLE:DA90228009-001 METRC/Biotrack#185691 Harvest/Lot ID: 185691 Batch#: 18569, Sample Size: N/A -grams Ordered: 02/28/19 Sampled:02/28/19 Completed: 03/04/19 Expires: 03/04/20 Sampling Method: SOP Client Method Image Safety Pesticides - Tested Microbials - NOT Tested Mycotoxins - Tested Heavy Metals - Tested Terpenes - NOT Tested Residual-Solvents - Tested Filth - NOT Tested Water Activity - NOT Tested Moisture - NOT Tested Cannabinoids 0.00% Total THC 98.6% Total CBD Cannabinoids Analyte Weight(%) mg/g D9-THC ND ND THCa ND ND TOTAL THC ND ND CBD 98.6 986 CBDa ND ND TOTAL CBD 98.6 986 CBN ND ND CBDV 0.28 2.88 D8-THC ND ND THCV ND ND CBG ND ND CBGa ND ND CBC ND ND TOTAL CANNABINOIDS 98.88 988.88 ND ND 98.6 ND ND 0.28 ND ND ND ND ND D9-THC THCa CBD CBDa CBN CBDV D8-THC THCV CBG CBGa CBC 4131 SW 47th AVENUE SUITE 1408 DAVIE, FL 33314 1-954-368-7664 info@eviolabsfl.com ___________________ Jorge Segredo Lab Director State License # n/a ISO Accreditation # 97164 This report shall not be reproduced, unless in its entirety, without written approval from EVIO Labs. This report is an EVIO Labs certification. The results relate only to the material or product analyzed. Test results are confidential unless explicitly waived otherwise. Void after 1 year from test end date. Cannabinoid content of batch material may vary depending on sampling error. IC=In-control QC parameter, NC=Non-controlled QC parameter, ND=Not Detected, NA=Not Analyzed, ppm=Parts Per Million, ppb=Parts Per Billion. Limit of Detection (LoD) and Limit Of Quantitation (LoQ) are terms used to describe the smallest concentration that can be reliably measured by an analytical procedure. RPD=Reproducibility of two measurements. Action Levels are State determined thresholds for human safety for consumption and/or inhalation.
Transcript

Certificate of AnalysisCBD CRYSTALLINE 99.6%CBD CRYSTALLINE 99.6%Matrix: N/A

Page 1 of 6

BMH Ventures, Inc.2900 Gateway Drive, Suite APompano BeachFL, USA 33069(954) [email protected]

SAMPLE:DA90228009-001METRC/Biotrack#185691 Harvest/Lot ID: 185691

Batch#: 18569, Sample Size: N/A -gramsOrdered: 02/28/19 Sampled:02/28/19

Completed: 03/04/19 Expires: 03/04/20Sampling Method: SOP Client Method

Image Safety

Pesticides - TestedMicrobials - NOT TestedMycotoxins - TestedHeavy Metals - TestedTerpenes - NOT TestedResidual-Solvents - TestedFilth - NOT TestedWater Activity - NOT TestedMoisture - NOT Tested

Cannabinoids

0.00%Total THC

98.6%Total CBD

CannabinoidsAnalyte Weight(%) mg/gD9-THC ND NDTHCa ND NDTOTAL THC ND NDCBD 98.6 986CBDa ND NDTOTAL CBD 98.6 986CBN ND NDCBDV 0.28 2.88D8-THC ND NDTHCV ND NDCBG ND NDCBGa ND NDCBC ND NDTOTAL CANNABINOIDS 98.88 988.88

ND ND

98.6

ND ND 0.28 ND ND ND ND NDD9-THC THCa CBD CBDa CBN CBDV D8-THC THCV CBG CBGa CBC

4131 SW 47th AVENUE SUITE1408DAVIE, FL [email protected] ___________________

Jorge SegredoLab Director

State License # n/aISO Accreditation #97164

This report shall not be reproduced, unless in its entirety, without written approval from EVIO Labs. This report is an EVIO Labs certification. The results relate only to the material or productanalyzed. Test results are confidential unless explicitly waived otherwise. Void after 1 year from test end date. Cannabinoid content of batch material may vary depending on sampling error.IC=In-control QC parameter, NC=Non-controlled QC parameter, ND=Not Detected, NA=Not Analyzed, ppm=Parts Per Million, ppb=Parts Per Billion. Limit of Detection (LoD) and Limit OfQuantitation (LoQ) are terms used to describe the smallest concentration that can be reliably measured by an analytical procedure. RPD=Reproducibility of two measurements. Action Levels areState determined thresholds for human safety for consumption and/or inhalation.

Certificate of AnalysisCBD CRYSTALLINE 99.6%CBD CRYSTALLINE 99.6%Matrix: N/A

Page 2 of 6

BMH Ventures, Inc.2900 Gateway Drive, Suite APompano BeachFL, USA 33069(954) [email protected]

SAMPLE:DA90228009-001METRC/Biotrack#185691 Harvest/Lot ID: 185691

Batch#: 18569, Sample Size: N/A -gramsOrdered: 02/28/19 Sampled:02/28/19

Completed: 03/04/19 Expires: 03/04/20Sampling Method: SOP Client Method

Pesticides LOQ Action Level Result Units TypeDaminozide 0.02 1 ND ppm Plant growth regulatorAcephate 0.01 0.4 ND ppm InsecticideFlonicamid 0.01 1 ND ppm Pyridine Insecticide, AphicideOxamyl 0.01 1 ND ppm Carbamate Insecticide, Acaricide,

NematicideMethomyl 0.01 0.4 ND ppm Carbamate Insecticide, Acaricide,

MetaboliteThiamethoxam 0.01 0.2 ND ppm Neonicotinoid InsecticideImidacloprid 0.01 0.4 ND ppm Neonicotinoid InsecticideDimethoate 0.01 0.2 ND ppm Organophosphate Insecticide,

Acaricide, MetaboliteAcetamiprid 0.01 0.2 ND ppm InsecticideThiacloprid 0.01 0.2 ND ppm Neonicotinoid Insecticide,

MolluscicideAldicarb 0.02 0.4 ND ppm Insecticide, NematicideDichlorvos 0.05 0.1 ND ppm Organophosphate Insecticide,

Acaricide, MetabolitePropoxur 0.01 0.2 ND ppm Carbamate Insecticide, AcaricideCarbofuran 0.01 0.2 ND ppm Insecticide, NematicideCarbaryl 0.01 0.2 ND ppm InsecticideImazalil 0.01 0.2 ND ppm Imidazole FungicideMetalaxyl 0.01 0.2 ND ppm Phenylamide FungicideChlorantraniliprole 0.01 0.2 ND ppm InsecticidePhosmet 0.01 0.2 ND ppm Organophosphate Insecticide,

AcaricideSpiroxamine 0.01 0.4 ND ppm Morpholine FungicideNaled 0.01 0.5 ND ppm Organophosphate Insecticide,

AcaricideMethiocarb 0.01 0.2 ND ppm Carbamate Insecticide,

Molluscicide, Bird repellentAzoxystrobin 0.01 0.2 ND ppm FungicidePaclobutrazol 0.01 0.4 ND ppm Triazole Plant growth regulator;

FungicideMalathion 0.01 0.2 ND ppm Organophosphate Insecticide,

AcaricideMyclobutanil 0.01 0.2 ND ppm Triazole FungicideBifenazate 0.01 0.2 ND ppm InsecticideSpirotetramat 0.02 0.2 ND ppm Tetramic acid InsecticideEthoprophos 0.01 0.2 ND ppm Insecticide, NematicideFenoxycarb 0.01 0.2 ND ppm Carbamate InsecticideKresoxim-methyl 0.01 0.4 ND ppm Strobilurin Fungicide, BactericideTebuconazole 0.01 0.4 ND ppm Triazole FungicideDiazanon 0.01 0.2 ND ppm Organophosphate Insecticide,

Acaricide, Repellent

4131 SW 47th AVENUE SUITE1408DAVIE, FL [email protected] ___________________

Jorge SegredoLab Director

State License # n/aISO Accreditation #97164

This report shall not be reproduced, unless in its entirety, without written approval from EVIO Labs. This report is an EVIO Labs certification. The results relate only to the material or productanalyzed. Test results are confidential unless explicitly waived otherwise. Void after 1 year from test end date. Cannabinoid content of batch material may vary depending on sampling error.IC=In-control QC parameter, NC=Non-controlled QC parameter, ND=Not Detected, NA=Not Analyzed, ppm=Parts Per Million, ppb=Parts Per Billion. Limit of Detection (LoD) and Limit OfQuantitation (LoQ) are terms used to describe the smallest concentration that can be reliably measured by an analytical procedure. RPD=Reproducibility of two measurements. Action Levels areState determined thresholds for human safety for consumption and/or inhalation.

Certificate of AnalysisCBD CRYSTALLINE 99.6%CBD CRYSTALLINE 99.6%Matrix: N/A

Page 3 of 6

BMH Ventures, Inc.2900 Gateway Drive, Suite APompano BeachFL, USA 33069(954) [email protected]

SAMPLE:DA90228009-001METRC/Biotrack#185691 Harvest/Lot ID: 185691

Batch#: 18569, Sample Size: N/A -gramsOrdered: 02/28/19 Sampled:02/28/19

Completed: 03/04/19 Expires: 03/04/20Sampling Method: SOP Client Method

Pesticides LOQ Action Level Result Units TypePropiconazole 0.01 0.4 ND ppm Triazole FungicideClofentezine 0.01 0.2 ND ppm Tetrazine AcaricideSpinosad (Spinosyn A) 0.01 0.2 ND ppm InsecticidePrallethrin 0.05 0.2 ND ppm Synthetic pyrethroid InsecticideTrifloxystrobin 0.01 0.2 ND ppm Strobilurin FungicidePiperonyl butoxide 0.01 3 ND ppm Cyclic aromatic; Performance

enhancer, SynergistChlorpyrifos 0.01 0.2 ND ppm Organophosphate InsecticideHexythiazox 0.01 1 ND ppm Carboxamide AcaricideEtoxazole 0.01 0.2 ND ppm Diphenyl oxazoline AcaricideSpiromesifen 0.01 0.2 ND ppm Tetronic acid InsecticidePyrethrins (Pyrethrin I) 0.01 1 ND ppm InsecticideFenpyroximate 0.01 0.4 ND ppm Pyrazolium Acaricide, InsecticidePyridaben 0.01 0.2 ND ppm Pyridazinone Insecticide, AcaricidePermethrins 0.05 0.2 ND ppm Pyrethroid InsecticideAbamectin B1a 0.02 0.5 ND ppm InsecticideEtofenprox 0.01 0.4 ND ppm Pyrethroid InsecticideBifenthrin 0.01 0.2 ND ppm Acaricide, InsecticideFludioxonil 0.01 0.4 ND ppm Phenylpyrrole FungicideFipronil 0.02 0.4 ND ppm Phenylpyrazole InsecticideCypermethrin 0.02 1 ND ppm Pyrethroid Insecticide, Veterinary

substanceMevinphos 0.01 0.1 ND ppm Organophosphate Insecticide,

AcaricideDimethomorph 0.01 0.1 ND ppm Morpholine FungicideFenhexamid 0.01 0.1 ND ppm Hydroxyanilide FungicideCoumaphos 0.01 0.2 ND ppm InsecticideSpinosad (Spinosyn D) 0.01 0.2 ND ppm Insecticide

4131 SW 47th AVENUE SUITE1408DAVIE, FL [email protected] ___________________

Jorge SegredoLab Director

State License # n/aISO Accreditation #97164

This report shall not be reproduced, unless in its entirety, without written approval from EVIO Labs. This report is an EVIO Labs certification. The results relate only to the material or productanalyzed. Test results are confidential unless explicitly waived otherwise. Void after 1 year from test end date. Cannabinoid content of batch material may vary depending on sampling error.IC=In-control QC parameter, NC=Non-controlled QC parameter, ND=Not Detected, NA=Not Analyzed, ppm=Parts Per Million, ppb=Parts Per Billion. Limit of Detection (LoD) and Limit OfQuantitation (LoQ) are terms used to describe the smallest concentration that can be reliably measured by an analytical procedure. RPD=Reproducibility of two measurements. Action Levels areState determined thresholds for human safety for consumption and/or inhalation.

Certificate of AnalysisCBD CRYSTALLINE 99.6%CBD CRYSTALLINE 99.6%Matrix: N/A

Page 4 of 6

BMH Ventures, Inc.2900 Gateway Drive, Suite APompano BeachFL, USA 33069(954) [email protected]

SAMPLE:DA90228009-001METRC/Biotrack#185691 Harvest/Lot ID: 185691

Batch#: 18569, Sample Size: N/A -gramsOrdered: 02/28/19 Sampled:02/28/19

Completed: 03/04/19 Expires: 03/04/20Sampling Method: SOP Client Method

Residual solvent Action Level(ppm) Pass/Fail Results(ppm)Hexanes (2,3-dimethylbutane) 290 Pass ND1,4-Dioxane 380 Pass NDPentanes (iso-pentane) 5000 Pass NDPentanes (neo-pentane) 5000 Pass NDButanes (iso-butane) 5000 Pass ND2-Butanol 5000 Pass ND2-Ethoxyethanol 160 Pass ND2-Propanol 5000 Pass NDAcetone 5000 Pass NDAcetonitrile 410 Pass NDBenzene 2 Pass NDButanes (n-butane) 5000 Pass NDCyclohexane 3880 Pass NDDichloromethane 600 Pass NDHexanes (2,2-dimethylbutane) 290 Pass NDXylenes-O (1,2-dimethylbenzene) 2170 Pass NDXylenes-M (1,3-dimethylbenzene) 2170 Pass NDXylenes-P (1,4-dimethylbenzene) 2170 Pass NDEthanol 5000 Pass NDEthyl acetate 5000 Pass NDEthylbenzene 2170 Pass NDEthyl ether 5000 Pass NDEthylene Oxide 50 Pass NDHeptane 5000 Pass NDn-Hexane 290 Pass NDIsopropyl acetate 5000 Pass NDMethanol 3000 Pass NDHexanes (2-methylpentane) 290 Pass NDHexanes (3-methylpentane) 290 Pass NDPentanes (n-pentane) 5000 Pass NDPropane 5000 Pass NDTetrahydrofuran 720 Pass NDToluene 1068 Pass NDXylenes 2170 Pass ND

4131 SW 47th AVENUE SUITE1408DAVIE, FL [email protected] ___________________

Jorge SegredoLab Director

State License # n/aISO Accreditation #97164

This report shall not be reproduced, unless in its entirety, without written approval from EVIO Labs. This report is an EVIO Labs certification. The results relate only to the material or productanalyzed. Test results are confidential unless explicitly waived otherwise. Void after 1 year from test end date. Cannabinoid content of batch material may vary depending on sampling error.IC=In-control QC parameter, NC=Non-controlled QC parameter, ND=Not Detected, NA=Not Analyzed, ppm=Parts Per Million, ppb=Parts Per Billion. Limit of Detection (LoD) and Limit OfQuantitation (LoQ) are terms used to describe the smallest concentration that can be reliably measured by an analytical procedure. RPD=Reproducibility of two measurements. Action Levels areState determined thresholds for human safety for consumption and/or inhalation.

Certificate of AnalysisCBD CRYSTALLINE 99.6%CBD CRYSTALLINE 99.6%Matrix: N/A

Page 5 of 6

BMH Ventures, Inc.2900 Gateway Drive, Suite APompano BeachFL, USA 33069(954) [email protected]

SAMPLE:DA90228009-001METRC/Biotrack#185691 Harvest/Lot ID: 185691

Batch#: 18569, Sample Size: N/A -gramsOrdered: 02/28/19 Sampled:02/28/19

Completed: 03/04/19 Expires: 03/04/20Sampling Method: SOP Client Method

Cannabinoid Profile Test Result-Analysis Method :SOP.T.40.020, SOP.T.30.050 Analytical Batch:DA001870Reagent LOT ID Dilution Consumables Id022519.R02 00264746022719.R05 A91237102-GEN

849C4-849AK840C6-840H

Full spectrum cannabinoid analysis utilizing High Performance Liquid Chromatography with UV detection (HPLC-UV). (Method: SOP.T.30.050 for sample prep and Shimadzu HighSensitivity Method SOP.T.40.020 for analysis. LOQ for all cannabinoids is 1 mg/L).Mycotoxin Analysis-Analysis Method :SOP.T.30.065, SOP.T.40.065 Analytical Batch:Analyte Results Action LevelAflatoxin G2 ND 0.02Aflatoxin G1 ND 0.02Aflatoxin B2 ND 0.02Aflatoxin B1 ND 0.02Ochratoxin A+ ND 0.02Aflatoxins B1, B2, G1, G2, and Ochratoxins A testing using LC-MS. (Method: SOP.T.30.065 for Sample Preparation and SOP.T40.065 Procedure for Mycotoxins QuantificationUsing LCMS. LOQ 1.0 ppb). Total Aflatoxins (Aflotoxin B1, B2, G1, G2) must be <20µg/Kg. Ochratoxins must be <20µg/Kg.

4131 SW 47th AVENUE SUITE1408DAVIE, FL [email protected] ___________________

Jorge SegredoLab Director

State License # n/aISO Accreditation #97164

This report shall not be reproduced, unless in its entirety, without written approval from EVIO Labs. This report is an EVIO Labs certification. The results relate only to the material or productanalyzed. Test results are confidential unless explicitly waived otherwise. Void after 1 year from test end date. Cannabinoid content of batch material may vary depending on sampling error.IC=In-control QC parameter, NC=Non-controlled QC parameter, ND=Not Detected, NA=Not Analyzed, ppm=Parts Per Million, ppb=Parts Per Billion. Limit of Detection (LoD) and Limit OfQuantitation (LoQ) are terms used to describe the smallest concentration that can be reliably measured by an analytical procedure. RPD=Reproducibility of two measurements. Action Levels areState determined thresholds for human safety for consumption and/or inhalation.

Certificate of AnalysisCBD CRYSTALLINE 99.6%CBD CRYSTALLINE 99.6%Matrix: N/A

Page 6 of 6

BMH Ventures, Inc.2900 Gateway Drive, Suite APompano BeachFL, USA 33069(954) [email protected]

SAMPLE:DA90228009-001METRC/Biotrack#185691 Harvest/Lot ID: 185691

Batch#: 18569, Sample Size: N/A -gramsOrdered: 02/28/19 Sampled:02/28/19

Completed: 03/04/19 Expires: 03/04/20Sampling Method: SOP Client Method

Pesticide Analysis-Analysis Method:SOP.T.30.065, SOP.T.40.065 Analytical Batch :DA001877Reagent LOT/ID Dilution Consumables ID022619.R03 A91237102-GENPesticide screen is performed using LC-MS which can screen down to below single digit ppb concentrations for regulated Pesticides. Currently we analyze for 57 Pesticides.(Method: SOP.T.30.065 Sample Preparation for Pesticides Analysis via LCMSMS and SOP.T40.065 Procedure for Pesticide Quantification Using LCMS).Heavy Metals Analysis-Analysis-Method:SOP.T.40.050, SOP.T.30.052 Analytical Batch: DA001879Reagent LOT/ID Dilution Consumables ID013019.01 50011519.01021319.R15021819.R02Heavy Metals screening is performed using ICP-MS (Inductively Coupled Plasma – Mass Spectrometer) which can screen down to below single digit ppb concentrations forregulated heavy metals using Method SOP.T.30.052 Sample Preparation for Heavy Metals Analysis via ICP-MS and SOP.T.40.050 Heavy Metals Analysis via ICP-MS.Metal Result Action-LevelArsenic ND 1.500Cadmium ND 0.500Lead ND 0.500Mercury ND 3Abbreviation:ppm=Parts Per MillionResidual SolventsAnalysis Method:SOP.T.40.032 Analytical Batch :DA001872Residual solvents screening is performed using GC-MS which can detect below single digit ppm concentrations. Currently we analyze for 34 Residual solvents. (Method:SOP.T.30.042 Residual Solvents Analysis via GC-MS).

4131 SW 47th AVENUE SUITE1408DAVIE, FL [email protected] ___________________

Jorge SegredoLab Director

State License # n/aISO Accreditation #97164

This report shall not be reproduced, unless in its entirety, without written approval from EVIO Labs. This report is an EVIO Labs certification. The results relate only to the material or productanalyzed. Test results are confidential unless explicitly waived otherwise. Void after 1 year from test end date. Cannabinoid content of batch material may vary depending on sampling error.IC=In-control QC parameter, NC=Non-controlled QC parameter, ND=Not Detected, NA=Not Analyzed, ppm=Parts Per Million, ppb=Parts Per Billion. Limit of Detection (LoD) and Limit OfQuantitation (LoQ) are terms used to describe the smallest concentration that can be reliably measured by an analytical procedure. RPD=Reproducibility of two measurements. Action Levels areState determined thresholds for human safety for consumption and/or inhalation.