+ All Categories
Home > Documents > Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer...

Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer...

Date post: 19-Jan-2020
Category:
Upload: others
View: 25 times
Download: 0 times
Share this document with a friend
30
Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi , Alexander Blacky, Walter Koller Clinical Institute of Hospital Hygiene Medical University of Vienna
Transcript
Page 1: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions

Magda Diab-Elschahawi, Alexander Blacky, Walter Koller

Clinical Institute of Hospital HygieneMedical University of Vienna

Page 2: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Medical equipment

A new very expensive hightech medical instrument for diagostic or therapeutic purposes (wonderfull for the clinician – a nightmare for hospital hygienist) does

it end up being a single use product because we cannot reprocess it?!

Page 3: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

STERRAD100NX™ Sterilizer

Page 4: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Why the STERRAD 100NX Sterilizer?

ONLY NEEDSa power pointH2O2 cartridges provided by the manufacturer to run

DOES NOTleave any toxic residues norgenerates harmful waste

Cycle times are shortWorking temperatures are low >> “gentle” processing of thermo-labile instruments

Page 5: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Aim of the study

• Evaluate the efficacy of the Sterrad 100 NX under challenging conditions

• Challenges:

1. Carrier materials (TIT, PU and PE)2. Wrappings (1 or 3 times)*3. Organic and inorganic burdens **not according to manufacturers instructions

Page 6: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Working hypothesis

To verify that the Sterrad 100NX sterilizer in its present setting can

consistently provide a minimum sterility assurance level (SAL) of 10-6 in

presence of above mentioned challenges.

Page 7: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Carrier materials

Certain materials, particularly polymers,might not be compatible with H2O2.Depending on the type of polymer,different degrees of polymer surfacemodifications induced by H2O2 and otherplasma-based sterilization techniqueshave been observed.

Page 8: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Carrier materials

Sterilant type and concentration as wellas parameters such as:• temperature, • pressure and of course • cycle time of a sterilization process will determine which materials can besafely processed.

Page 9: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Carrier materials

• Titanium (TIT)• Polyethylene (PE)• Polyurethane (PU)

Single versus threefold wrapping of inoculated carriers (sized 20 x 5 mm) with Tyvek® sterilization pouches.

Page 10: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Test organism

• Spores of Geobacillus stearothermophilus (ATCC 7953) were used as indicator organism (at least one million spores per carrier).

• Spore preparation was done following the method described by Pflug IJ. After harvesting and cleaning spore pellets were resuspended in either 5% FBS or in hard water.

Page 11: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Test load

• Two perforated stainless steel baskets with a standardized load of surgical instruments without lumens such as forceps, scissors, clamps and retractors

• >> Total weight of the test load was 5 kG.

Page 12: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

BI-Distribution

• In testing the efficacy of a sterilization process, the BIs should be placed in several places considered to be the most difficult sites in the sterilizer load to sterilize. We therefore distributed our BIs trying to reflect probable key positions.

Page 13: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Test load + BIs

Test load upper shelf Test load lower shelf

Page 14: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Untreated 300 ppm 600 ppm 1200 ppm 5% FBS

2 3 41 2 3 41 2 3 41 2 3 41 2 3 41

Carrier type (PU, PE, TIT)

Standard half cycle #1: Quantitative evaluationStandard half cycle #2-4: Qualitative evaluation

Standard half Cycle #

Experimental setting

Page 15: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Cycles (full cycles)

• Standard cyclefor the sterilization of most surgical instruments >> about 47 min. cycle time

AND

• Flex Scope cyclefor the sterilization of flexible endoscopes >> about 42 min. cycle time.

Page 16: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Why use half cycles?„overkill method“

• The principle of the half-cycle approach is to challenge a sterilization process with BIs (usually containing at least 106 spores per carrier) at sterilization times equal to half of the full cycle. Inactivating a BI with an initial population of 106 in the half-cycle means at least a 6-log reduction has been attained. Extrapolating the inactivation kinetics of a half-cycle which inactivated 106 spores will provide a 10-6 SAL for the full cycle. Reference: EN ISO 14937

Page 17: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Worst-case conditions

• Cleaning process not validated >> sub-standard washing or rinsing

• Human error >> Pushing the wrong button: Standard instead of Flex Scope cycle

Page 18: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Percentage of BIs with no growth BIs wrapped once BIs wrapped three

timesCondition/ challenge

PU PE TIT PU PE TIT

Untreated 100% 96.6% 100% 100% 100% 100%300ppm 23.3% 56.6% 63.5% 16.6% 40% 93.3%600ppm 76.6% 90% 86.6% 26.6% 80% 86.6%1200ppm 30% 40% 90% 16.6% 56.6% 76.6%5% FBS 93.3% 96.6% 80% 80% 90% 93.3%

Qualitative Results

Page 19: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Qualitative Results

BIs wrapped once

0

20

40

60

80

100

120

untreated 300ppm 600ppm 1200ppm 5%FBS

condition/challenge

%of

pro

cess

ed B

Is w

ith n

G

PUPETIT

Page 20: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Qualitative Results

BIs wrapped three times

0

20

40

60

80

100

120

untreated 300ppm 600ppm 1200ppm 5%FBS

condition/challenge

% o

f pre

oces

sed

Bis

with

nG

PU

PE

TIT

Page 21: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Untreated condition

• Our qualitative results show that irrespective of the number of wrappings (1 or 3) in the untreated condition sterilization by the Sterrad 100NX was equally effective on all three carrier materials, reaching a log10 reduction rate of ≥ 6 under Standard half cycle conditions.

Page 22: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Organic/inorganic challenge

• When an organic or inorganic challenge was added to our spore carriers, with none of the three carrier materials a log10 reduction rate of 6 could consistently be achieved under Standard half cycle conditions.

Page 23: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Carrier materials

• Sterilization by the Sterrad 100NX was least effective on the PU carriers (considering organic and inorganic challenge as well as once and three times wrapping).

Page 24: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Influence of wrapping (single/tripple)

• Threefold wrapping was beneficial for TIT in certain conditions (organic challenge and 300ppm qualitative results), while it impaired the sterilizing ability of H2O2 for PU and PE (wrapped three times with organic and inorganic challenge).

Page 25: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Influence of wrapping (single/tripple): Hypothesis

H2O2 trapped by the three layers of wrapping

>> had therefore longer time to act as sterilant.

Beneficial effect for an inert material (TIT) / Adverse effect formaterial incompatible with H2O2 sterilization:

The surplus of H2O2 might be absorbed by such materials and might thus be prevented from reaching relevant surfaces in sufficient concentrations.

Valid explanation for PU (a material known to absorb H2O2 varyingwith its micro-structure) it does not apply to PE known to be inertregarding H2O2 resorption.

Page 26: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Reduction factor (mean from 10 positions)

BIs wrapped once

BIs wrapped three times

Condition/ challenge

PU PE TIT PU PE TIT

Untreated 5.60 5.26 5.73 5.67 5.81 5.19300ppm 3.65 5.13 5.24 3.89 4.95 4.82600ppm 5.06 5.62 5.09 3.84 5.41 5.531200ppm 5.07 4.74 6.17 5.60 4.56 6.125% FBS 5.57 4.37 5.29 5.01 4.47 4.72

Quantitative Results

Page 27: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

ConditionCondition BIs wrapped onceBIs wrapped once BIs wrapped three timesBIs wrapped three times

untreated PU position 2 PE position 1-3, 6,7

TIT position 1,8-10 PU position 3,7,10

300ppm none none

600ppm PE position 3,6,9,10 none

1200ppm TIT position 1-10 TIT all positions except 8

5% FBS PU position 1 none

TIT position 4 none

Quantitative Results

Positions where a log reduction rate of ≥ 6 was reached

Page 28: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Carrier positions

• Sometimes big variations in log10 reduction rates found for our BIs under one specified condition:

>> uneven repartition of the spore preparation on the carrier material.>> non-uniform distribution of H2O2 vapor in the sterilization chamber, therefore a limited ability of the vapor to reach different positions within the test load equally well.

• Question: Is hydrogen peroxide vapor distributed homogenously within the sterilization chamber? No sensors positioned within the test load.

Page 29: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Recommendations regarding the STERRAD100NX™

• Low temperature hydrogen peroxide plasma offers a very promising sterilization technology.

• Significance of a thorough and validated cleaning of

contaminated items before being exposed to sterilization in the STERRAD100NX™ Sterilizer was clearly demonstrated.

• We also recommend to strictly adhering to the manufacturers recommendations specified in their User’s Guide regarding the correct cycle and permitted materials of medical devices for their processing in the Sterrad 100NX sterilizer.

Page 30: Challenging the STERRAD 100NX Sterilizer under ... · Challenging the STERRAD 100NX Sterilizer under experimental “clean” and “dirty” conditions Magda Diab-Elschahawi, Alexander

Literatur1. Heeg P. Allgemeine Probleme der Infektionsprophylaxe in der Endoskopie. Hyg Med 1994; 19: 554-559.2. Körber J und Trautmann M. Infektionsrisiko und -prophylaxe bei endoskopischen Interventionen im Gastrointestinaltrakt. Hyg Med

1994; 19: 543-552.3. Babb JR, Bradley CR. Endoscope decontamination: where do we go from here? J Hosp Infect. 1995 Jun; 30 Suppl: 543-51.4. Ishino Y, Ido K, Sugano K. Contamination with hepatitis B virus DNA in gastrointestinal endoscope channels: risk of infection on reuse

after on-site cleaning. Endoscopy. 2005 Jun; 37(6):548-51.5. Lerouge S, Tabrizian M, Wertheimer MR, Marchand R, Yahia L. Safety of plasma-based sterilization: surface modifications of

polymeric medical devices induced by Sterrad and Plazlyte processes. Biomed Mater Eng. 2002;12(1):3-13.6. Chu NS, Favero M. The microbial flora of the gastrointestinal tract and the cleaning of flexible endoscopes. Gastrointest Endosc Clin N

Am. 2000 Apr;10(2):233-44.7. Diab-Elschahawi M, Fürnkranz U, Blacky A, Koller W. Re-evaluation of current A0 value recommendations for thermal disinfection of

reusable human waste containers based on new experimental data. J Hosp Infect 2010 May; 75(1):62-5.8. Deutsche Gesellschaft für Krankenhaushygiene: Mitteilung des Vorstandes der DGKH zum aktuellen Erkenntnisstand der Validierung des

Sterrad/E-Plasma-Sterilisationsverfahrens mit den sich ergebenden Schlußfolgerungen für die Praxis. Hyg Med 1995; 20:52-53.9. Kanemitsu K, Imasaka T, Ishikawa S, Kunishima H, Harigae H, Ueno K et al. A comparative study of ethylene oxide gas, hydrogen

peroxide gas plasma, and low-temperature steam formaldehyde sterilization. Infect Control Hosp Epidemiol. 2005 May; 26(5):486-9.10. EN ISO 14937. Sterilization of health care products. General requirements for characterization of sterilizing agent and the

development, validation and routine control of a sterilization process. Berlin: Beuth; 2000.11. Peters J and Borchers U. Vergleichende Untersuchungen zur Resistenz von Mycobakterium terrae, Aspergillus niger und Bacillus-Sporen

bei der Plasma-Sterilisation, Zentr Steril 1995; 3:163-172.12. Koller W and Lessky E. Microbiological test results and observations with an H2O2 plasma sterilizer. Zentr Steril 1996; 4:79-87.13. Pflug IJ. Microbiology and Engineering of Sterilization processes, 10th edition Minneapolis, MN, USA: University of Minnesota

Environmental Sterilization Laboratory; 1999.14. Okpara J. Requirements for the testing of sterilization processes with Bacillus spores. Zentr Steril 1998; 6 (2):96-112.15. Feldman LA and Hui HK. Compatibility of medical devices and materials with low-temperature hydrogen peroxide gas plasma. Med Dev

Diag Ind 19 (1997), pp. 57–62.16. Okpara-Hofmann J, Knoll M, Dürr M, Schmitt B, Borneff-Lipp M. Comparison of low-temperature hydrogen peroxide plasma

sterilization for endoscopes using various Sterrad™ models. J Hosp Infect 2005; 59:280-285.17. Lipscomb IP, Sihota AK, Keevil CW. Comparison between visual analysis and microscope assessment of surgical instrument cleanliness

from sterile service departments. J Hosp Infect. 2008 Jan; 68(1):52-8. 18. Department of Health. Variant Creutzfeldt-Jakob disease (vCJD): minimizing the risk of transmission. Health Service Circular 1999,

HSC 1999/178.19. Department of Health. Controls assurance in infection control: decontamination of medical devices. Health Service Circular 1999, HSC

1999/179.20. Alfa MJ, Olson N, Alfadhaly A. Cleaning efficacy of medical device washers in North American healthcare facilities. J Hosp Infect.

2010 Feb;74(2):168-177. Epub 2009 Aug 27.

Thank you for your attention!


Recommended