Chandra Calibration Status Chandra Calibration Status Highlights from the Chandra Calibration Review New results on the HRMA effective area Updates to the ACIS contamination model Flux Cross-Calibration with Other X- ray Telescopes
Transcript
Page 1: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

Chandra Calibration StatusChandra Calibration Status

Highlights from the Chandra Calibration Review

New results on the HRMA effective area

Updates to the ACIS contamination model

Flux Cross-Calibration with Other X-ray Telescopes

Page 2: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

Proper Stellar Motions with Chandra

Page 3: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

Temperature-dependent ACIS gain corrections

Page 4: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.
Page 5: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.
Page 6: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

Recent HRMA Calibration Results

CALDB Version

Page 7: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.
Page 8: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.
Page 9: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.
Page 10: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.
Page 11: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.
Page 12: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

Optical depth of the contaminant at the Mn-L and F-K complex on Optical depth of the contaminant at the Mn-L and F-K complex on ACIS-SACIS-S

Updates to the ACIS contamination Model

Page 13: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

External Calibration Source (ECS)

Courtesy of Dan McCammon

Page 14: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

External Calibration Source (ECS)

Page 15: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

External Calibration Source (ECS)

Page 16: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

External Calibration Source (ECS)

Page 17: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

Surface Roughness of ACIS Filter ContaminantSurface Roughness of ACIS Filter Contaminant

External Calibration Source (ECS)

Page 18: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

Effect of Surface Roughness on Measured Optical DepthsEffect of Surface Roughness on Measured Optical Depths

External Calibration Source (ECS)

Page 19: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

Revised ACIS Contamination ModelRevised ACIS Contamination Model

CALDB New

Page 20: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

The Coma Cluster

1999

2009 CALDB

2009

New

Page 21: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

LETG/ACIS-S Observation of Mkn 421LETG/ACIS-S Observation of Mkn 421

CALDB New

Page 22: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

Flux Cross-Calibration with Other X-Ray TelescopesFlux Cross-Calibration with Other X-Ray Telescopes

OVII

OVIII

Ne IX

Ne X

Page 23: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

Simultaneous Chandra/XMM-Newton AGN ObservationsSimultaneous Chandra/XMM-Newton AGN Observations

Page 24: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.
Page 25: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.
Page 26: Chandra Calibration Status  Highlights from the Chandra Calibration Review  New results on the HRMA effective area  Updates to the ACIS contamination.

Additional AO-11 Calibration ObservationsAdditional AO-11 Calibration Observations

A set of 5 LETG/ACIS-S observations of Mkn421

A set of 4 observations of Abell 1795 on ACIS-I and ACIS-S

Measure the C-K edge optical depth at the aim-point on S3 with a LETG/ACIS-S observation of Mkn421