Date post: | 20-Jan-2016 |
Category: |
Documents |
Upload: | walter-elliott |
View: | 236 times |
Download: | 0 times |
Chapter 22
GC & LC
22-1 Gas Chromatography1. Schematic diagram
22-1 Gas Chromatography2. Columns : open tubular columns
22-1 Gas Chromatography
A) m.p.(gas) - s.p. 1) s.p.: solid ( using adsorption ) ex: SiO2
column ages: Si-O-H cause tailing peak.
2) s.p.: liquid ( GLC, using partition) a range of polarities (Table 22-1), “like dissolves like”
Decrease thickness of stationary phase leads to a) Resolution (H)b) tr
c) Sample capacity
22-1 Gas Chromatography B) The effects of column polarity on separation
Like dissolves like (a) S.P: nonpolar, b.p. dependent (b) S.P: polar
Figure 22-4 Resolution of trans fatty acids in hydrogenated food oil improves when the stationary phase is changed from DB-23 to HP-88 (aryl group)
P.484
How changing the S.P. can affect separation
22-1 Gas ChromatographyC) Common solid s.p. :
a) Porous carbon : larger molecules bind more tightly than small ones, flexible molecules bind more than rigid ones
b) Molecular sieves : inorganic materials with nanometer-size cavities that retain & separate small molecules : H2, O2, N2, CO2, CH4. (Fig. 22-5)
c) Guard column
Collect nonvolatile components that would otherwise be injected into a column and never be eluted.
22-1 Gas Chromatographypacked column vs. open tubular column
higher resolutionlower sample capacity
22-1 Gas Chromatography
3. Temperature programming
temp of column v.p. solute,
tr
sharpens peaksisothermal : constant temp.temp. programming (gradient) : raise the column temp. during the
separation.
22-1 Gas Chromatography -9
Figure 22-6 (a) Isothermal and (b) programmed temperature chromatography of linear alkanes through a packed column with a nonpolar stationary phase.
4. Carrier Gas
22-1 Gas Chromatography
22-1 Gas Chromatography
5. Sample Injection
1) gasses, liquids, or solids
vaporized, not decomposition
2) injection time bands broader
3) injected by syringe (manual or automatic injection)
22-1 Gas Chromatography
Figure 22-7 Injection port operation for (a) split, (b) splitless, and (c) on-column injection into an open tubular column.
22-1 Gas Chromatographysplit injection (350℃) (only 0.1-10% sample)
Routine method
concentrated sample
high resolution
dirty samples
could cause thermal decomposition
splitless injection (220℃) (80%)For quantitative analysis and for analysis of trace components of
mixture
high resolution
solvent trapping (Tsolvent < 40℃) for dilute sample
cold trapping (Tsolute < 150℃) for high-boiling solutes
on-column injection (50℃) (100%)
best for thermally unstable solutes.
22-1 Gas Chromatography
5. Detectors
Qualitative analysis :mass spectrometer, IR
Quantitative analysis :area of a chromatographic peak.
22-1 Gas Chromatography
a) Thermal conductivity detector:
-most general way
-responds to everything
-not sensitive enough for high resolution.
b) Flame ionization detector :
-most popular
-mainly responds hydrocarbons (C-H)
c) Electron capture detector : -for compounds containing atoms with high electron affinities.
-sensitive for halogen, C=O, NOx, & orgaometallic compounds.
22-1 Gas Chromatography
d) Mass Spectrometric Detection and Selected Reaction Monitoring :- A mass spectrometer is the single most versatile detector.- Total Ion Chromatogram (TIC)- selected ion monitoring (SIM) at on value of m/z- selected reaction monitoring (SRM) = tandem mass =
MS/MS- Multiple reaction monitoring (MRM)
QQQ Mass Spectrometer
Precursor ion (parent ion) vs. Product ions (daughter ion)
Solid phase extraction (SPE)
Caffeine as example
Caffeine (13C) as an internal standard
22-2 Liquid Chromatography
1. open, gravity-feed column 2. closed column (under high pressure)
packed with micron-size particles. (HPLC)
3. stationary phase : a. adsorption : silica (SiO2xH2O), alumina
(Al2O3xH2O),b. molecular exclusion,
c. ion-exchange, affinity
22-2 Liquid Chromatographycompete with ▲ for binding on s.p.
the more strongly bind to s.p.eluent strength
22-2 Liquid Chromatography
4. Eluent strength : Table 22.2
The more polar solvent
eluent strength
tr
5. Gradient elution : increased the eluent strength during the separation in liquid chromatography.
22-3 High-Performance Liquid Chromatography (HPLC)
1. Through a closed column, and needs high pressure.
2. s.p. particles size microporous particles of silica
with diameters of 1.5-10 um
s.p. m.p. faster,
i.e. C in van Deemter eqn.
resolution
22-3 High-Performance Liquid Chromatography (HPLC)
22-3 HPLC
22-3 HPLC
3. Stationary phasea) Normal-phase chromatography : polar s.p.
and less polar solvent. Eluent strength is increased by adding a more polar solvent.
b) Reversed-phase chromatography : low-polarity s.p. and polar solvent. Eluent strength is increased by adding a less polar solvent.
22-3 HPLCc) Bonded stationary phase.
polar vs. nonpolar
d) Optical isomersD- & L-amino acidsfor drug industry
see p.494 for R = polar or nonpolar
22-3 HPLCd) Optical isomers separation
ex: for ant-inflammatory drug Naproxen
4. Columna) Guard columnb) Injection valve
22-3 HPLC
22-3 HPLC
5. Solvents a) Isocratic elution :
elution with single solvent or a constant solvent mixture
b) Gradient elution : solvent is changed continuously from a weak eluent strength to a strong eluent strength by mixing more and more of a strong solvent to a weak solvent during the chromatography.
22-3 HPLC
A : KH2PO4(aq)
B: CH3CN(l)
Figure 22-20 Isocratic HPLC separation of a mixture of aromaticcompounds at 1.0 mL/min on a 0.46×25 cm Hypersil ODS column (C18 on 5-μm silica) at ambient temperature (~22 )℃ :(1) benzyl alcohol; (2) phenol; (3) 3’, 4’-dimethoxyacetopheneone; (4) benzoin; (5) ethyl benzoate;(6) toluene; (7) 2,6-dimethoxytoluene; (8) o-methoxybiphenyl.
22-3 HPLC
The gradient can be used to resolve all peaks by reducing the time from 2 h to 38 min.
Detectors- Ultraviolet detector
- Electrochemical detector
redox reaction
- Fluorescence detector
LC-MS
- ESI (Electrospray ionization)
- APCI (atmospheric pressure chemical ionization)
P.500
Figure 22-23 Atmospheric pressure chemical ionization interface between liquid chromatography column and mass spectrometer.