  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    1/40

    Instruction Manual

    ChargeSwitchgDNABuccal Cell

    Kits

    Forpurification of genomic DNA fromhuman

    Catalog nos. CS11020, CS11021, CS11020-10,andCS11021-10

    buccal swabs

    Version A

    7Janua ry 2005

    25-0819

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    2/40

    ii

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    3/40

    Table of Contents

    Table of Contents..................................................................................................iii

    Kit Contents and Storage ..................................................................................... v

    Accessory Products..............................................................................................vi

    Introduction ........................................................................................1

    Overview .................................................................................................................1

    Experimental Outline ............................................................................................4

    Methods............................................................................................... 5

    General Information Individual Samples .......................................................5

    Isolating Genomic DNA from Individual Samples ..........................................8

    General Information Automated Sample Processing .................................14

    Automated Genomic DNA Isolation ................................................................18

    Automated Genomic DNA Isolation Normalized Kit ................................23

    Troubleshooting ...................................................................................................27

    Appendix...........................................................................................30

    Technical Service ..................................................................................................30

    Purchaser Notification and Product Qualification .........................................32

    iii

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    4/40

    iv

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    5/40

    Kit Contents and Storage

    Types of Kits This manual is supplied with the following products.

    Product Number ofPurifications

    Catalog no.

    ChargeSwitchgDNA Normalized Buccal CellKit

    50

    960

    CS11020

    CS11020-10

    ChargeSwitchgDNA Buccal Cell Kit 50

    960

    CS11021

    CS11021-10

    Shipping andStorage

    All components of the ChargeSwitchgDNA Buccal CellKits are shipped at room temperature. Upon receipt, store

    the Proteinase K at 4C. Store all other components at roomtemperature.

    All components are guaranteed stable for 6 months, ifstored properly.

    Contents The components supplied in the ChargeSwitchgDNABuccal Cell Kits are listed below.

    Note:Some reagents in the kit may be provided in excess of theamount needed.

    Catalog no.

    Components CS11020 CS11021 CS11020-10 CS11021-10

    ChargeSwitch

    Lysis Buffer(L11) 50 ml 50 ml 960 ml 960 ml

    ChargeSwitchMagneticBeads

    2 x 1 ml 2 x 1 ml 40 ml 40 ml

    Proteinase K (20 mg/ml in 50mM Tris-HCl, pH 8.5, 5 mMCaCl2, 50% glycerol)

    500 l 500 l 9.6 ml 9.6 ml

    ChargeSwitchPurification

    Buffer (N6)

    5 ml 5 ml 100 ml 125 ml

    ChargeSwitchWash Buffer(W12)

    100 ml 100 ml 2 x 960 ml 2 x 960 ml

    ChargeSwitchElution Buffer(E5; 10 mM Tris-HCl, pH 8.5)

    15 ml 15 ml 2 x 100 ml 2 x 100 ml

    v

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    6/40

    Accessory Products

    Additional

    Products

    The table below lists additional products available from

    Invitrogen that may be used with the ChargeSwitchgDNABuccal Cell Kits. In addition, the table lists a selection ofChargeSwitchgDNA Kits that are available for purificationof genomic DNA from other sources. For more informationabout these and other ChargeSwitchgDNA Kits, refer toour Web site at www.invitrogen.comor call TechnicalService (see page 30).

    Product Amount Catalog no.

    MagnaRack 1 rack CS15000

    96-Well Magnetic Separator 1 rack CS15096

    ChargeSwitchgDNA Micro Tissue Kit 50 purifications CS11203

    ChargeSwitchgDNA Mini Tissue Kit 25 purifications CS11204

    ChargeSwitchgDNA 20 l Blood Kit 96 purifications CS11010

    ChargeSwitchgDNA 100 l Blood Kit 50 purifications CS11000

    ChargeSwitchgDNA 1 ml Blood Kit 20 purifications CS11001

    ChargeSwitchgDNA 1 ml Serum Kit 50 purifications CS11040

    ChargeSwitchgDNA 50 l Sheep BloodKit

    50 purifications CS11300

    ChargeSwitchgDNA Mini Bacteria Kit 50 purifications CS11301

    ChargeSwitchForensic DNAPurification Kit

    100 purifications CS11200

    Quant-iTDNA Assay Kit, HighSensitivity

    1000 assays Q33120

    vi

    http://www.invitrogen.com/http://www.invitrogen.com/
  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    7/40

    Introduction

    Overview

    Introduction The ChargeSwitchgDNA Buccal Cell Kits allow rapid andefficient purification of genomic DNA from human buccalswabs. After preparing the lysates, you may purify DNA inless than 15 minutes using the ChargeSwitchTechnology.Depending on the kit used, samples may be handledindividually or in an automated system using a liquidhandling robot. For more information about the

    ChargeSwitch

    Technology, see page 3.

    Intended Usefor the Kits

    The ChargeSwitchgDNA Buccal Cell Kits are designed toallow isolation of the following amounts of genomic DNAfrom human buccal cell swabs or pelleted cells from a

    mouthwash. Samples can be stored for up to 2 weeks at 4Cbefore processing without a noticeable loss in DNA yield orquality. The purified genomic DNA is suitable for use in

    downstream applications such as PCR.

    ChargeSwitchgDNA Normalized Buccal Cell Kits:Produce a normalized yield of genomic DNA at a

    concentration of 1-3 ng/l in a total volume of 150 l.

    ChargeSwitchgDNA Buccal Cell Kits:Purify up to

    6 g of genomic DNA.

    Important:The DNA yield varies and is dependent on several factorsincluding the technique of the person taking the swab, whether the

    donor is a high or low shedder, and the type of swab used.

    continued on next page

    1

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    8/40

    Overview, continued

    Advantages Use of the ChargeSwitchgDNA Buccal Cell Kits to isolate

    genomic DNA provides the following advantages:

    Uses a magnetic bead-based technology to isolategenomic DNA without the need for hazardouschemicals, centrifugation, or vacuum manifolds

    Rapid and efficient purification of genomic DNA fromhuman buccal swabs in less than 15 minutes followingsample preparation and lysis

    Simple lysis with Proteinase K without the need for anymechanical lysis

    Minimal contamination with RNA

    The purified genomic DNA demonstrates improveddownstream performance in applications including PCR

    Includes a kit designed for automated processing of largenumbers of samples in 96-well plates using a liquidhandling robot

    SystemSpecifications

    Starting Material: Human buccal swabs

    Elution Volume: 150 l

    DNA Yield: 1-3 ng/l in 150 l (Normalized Buccal

    Cell Kit) or up to 6 g (Buccal Cell Kit)

    DNA Size: Varies (depends on quality of startingmaterial

    continued on next page

    2

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    9/40

    Overview, continued

    The

    ChargeSwitchTechnology

    The ChargeSwitchTechnology (CST) is a novel magnetic

    bead-based technology that provides a switchable surfacecharge dependent on the pH of the surrounding buffer tofacilitate nucleic acid purification. In low pH conditions, theCSTbeads have a positive charge that binds the negativelycharged nucleic acid backbone (see figure below). Proteinsand other contaminants are not bound and are simplywashed away in an aqueous wash buffer. To elute nucleicacids, the charge on the surface of the bead is neutralized byraising the pH to 8.5 using a low salt elution buffer (see

    figure below). Purified DNA elutes instantly into this elutionbuffer, and is ready for use in downstream applications.

    ChargeSwitch

    Magnetic BeadSpecifications

    Bead Binding Capacity: 5-10 g genomic DNA per mg

    Bead Size: < 1 m

    Bead Concentration: 25 mg/ml (Buccal Cell Kitsonly)

    3.125 mg/ml (NormalizedBuccal Cell Kits only)

    Storage Buffer: 10 mM MES, pH 5.0, 10 mMNaCl, 0.1% Tween 20

    AutomatedLiquidHandling

    Use of the ChargeSwitchgDNA Buccal Cell Kits has beendemonstrated on the Tecan Genesisrobotic workstation topurify DNA in a fully automated system from largenumbers of buccal cell swabs in a 96-well format. Otherliquid handling robots are suitable provided that each isequipped with a gripper arm, a 96-well magnetic separator,and other additional hardware as described on page 14. Thismanual provides general guidelines and a protocol that may

    be used to develop a script for your robot. For moreinformation, see www.invitrogen.comor call TechnicalService (page 30).

    Genesisis a registered trademark of Tecan AG Group

    3

    http://www.invitrogen.com/http://www.invitrogen.com/
  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    10/40

    Experimental Outline

    Introduction The figure below illustrates the basic steps necessary to

    purify genomic DNA from your buccal cell swab using oneof the ChargeSwitchgDNA Buccal Cell Kits.

    4

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    11/40

    Methods

    General Information Individual Samples

    Introduction This section provides general information needed to use theChargeSwitchgDNA Buccal Cell Kits (Catalog nos.CS11020 or CS11021) to process individual samples. If youare using a liquid handling robot to process large numbersof samples, see General Information Automated SampleProcessing, page 14.

    User SuppliedMaterials

    In addition to the reagents supplied with the kit, you need tohave the following materials on hand before beginning:

    A magnetic separation rack suitable for use with 1.5 mlmicrocentrifuge tubes (see below)

    Sterile, 1.5 ml microcentrifuge tubes

    Vortex mixer

    20 l, 200 l, and 1 ml sterile, pipette tips

    Water bath at 37C

    MagnaRack

    The MagnaRackavailable from Invitrogen (Catalog no.CS15000) is a two-piece magnetic separation rack for use inprotocols with magnetic beads, and consists of a magnetic

    base station and a removable tube rack. The tube rack canhold up to 24 microcentrifuge tubes. The tube rack fits ontothe magnetic base station in two different positions,

    associating the row of 12 neodymium magnets with a singlerow of 12 tubes for simple on the magnet and off themagnet sample processing (see figure below). For moreinformation, see www.invitrogen.comor call TechnicalService (page 30).

    continued on next page

    5

    http://www.invitrogen.com/http://www.invitrogen.com/
  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    12/40

    General Information Individual Samples,continued

    SafetyInformation

    Follow the safety guidelines below when using theChargeSwitchgDNA Buccal Cell Kits.

    Treat all reagents supplied in the kit as potentialirritants.

    Always wear a suitable lab coat, disposable gloves, andprotective goggles.

    If a spill of the buffers occurs, clean with a suitable

    laboratory detergent and water. If the liquid spillcontains potentially infectious agents, clean the affectedarea first with laboratory detergent and water, thenwith 1% (v/v) sodium hypochlorite or a suitablelaboratory disinfectant.

    Dispose of biological samples and all liquid wastegenerated during the purification procedure as

    biohazardous waste.

    Handling theChargeSwitch

    MagneticBeads

    Follow the guidelines below when handling theChargeSwitchmagnetic beads.

    Do not freeze the beads as this irreparably damagesthem. Store the beads at room temperature.

    Always keep the beads in solution. Do not allow themto dry out as this renders them non-functional.

    When using the beads, resuspend thoroughly in thestorage buffer by vortexing before removal.

    Discard beads after use. Do not reuse.

    continued on next page

    6

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    13/40

    General Information Individual Samples,continued

    Elution Buffer ChargeSwitchElution Buffer (E5; 10 mM Tris-HCl, pH 8.5)is supplied with the kit for eluting the DNA from theChargeSwitchMagnetic Beads. For best results, use ElutionBuffer (E5) to elute the DNA. Alternatively, TE Buffer, pH8.5-9.0 is acceptable. Note that the pH must be between 8.5-9.0 otherwise the DNA will not elute. Do not use water forelution.

    The protocol suggests eluting the genomic DNA in 150 l of

    ChargeSwitchElution Buffer (E5). When using theChargeSwitchgDNA Buccal Cell Kit, you may vary theamount of ChargeSwitchElution Buffer (E5) used to obtaingenomic DNA in the desired final concentration. For bestresults, always use a volume of ChargeSwitchElutionBuffer (E5) that is equal to or greater than the volume ofChargeSwitchMagnetic Beads used in the protocol.If thevolume of ChargeSwitchElution Buffer (E5) is lower thanthe volume of beads used, DNA elution is incomplete. You

    may need to perform a second elution to recover all DNA.Important:When using the ChargeSwitchgDNA Normalized

    Buccal Cell Kit, elute the genomic DNA in 150 l of ChargeSwitchElution Buffer (E5). Do notvary the amount of ChargeSwitchElution Buffer (E5) used otherwise the yield will no longer be

    normalized at 1-3 ng/l.

    7

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    14/40

    Isolating Genomic DNA from IndividualSamples

    Introduction This section provides guidelines and instructions to isolategenomic DNA from human buccal swabs or pelleted cellsfrom a mouthwash using the reagents supplied in the kit.Note that the protocol is optimized for efficient purificationof DNA from small sample volumes. Depending on thevolume of your sample, some further optimization of theprotocol may be required.

    StartingMaterial

    Use this procedure to isolate genomic DNA from humanbuccal cell swabs or pelleted cells from a mouthwash.

    Process samples immediately after collection or store at 4Cfor up to 2 weeks. Do notstore unprocessed samples atroom temperature as buccal swabs contain bacteria andnucleases that will degrade DNA.

    The ChargeSwitchMagnetic Beads supplied in theChargeSwitchgDNA Buccal Cell Kit differ inconcentration from those supplied in the ChargeSwitchgDNA Normalized Buccal Cell Kit.Although thepurification protocol is identical for both kits, you mustusethe ChargeSwitchMagnetic Beads supplied with each kit toobtain the DNA yields specified on page 1. Do notsubstitute ChargeSwitchMagnetic Beads provided in theChargeSwitchgDNA Buccal Cell Kit for those in the

    ChargeSwitch

    gDNA Normalized Buccal Cell Kit.

    continued on next page

    8

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    15/40

    Isolating Genomic DNA from IndividualSamples, continued

    MaterialsNeeded

    Have the following materials on hand before beginning:

    Buccal swab(s)

    MagnaRack(Catalog no. CS15000)

    Sterile 1.5 ml microcentrifuge tubes

    Vortex mixer

    Sterile pipette tips (20 l, 200 l, and 1 ml)

    Water bath

    Components Supplied with the Kit

    ChargeSwitchLysis Buffer (L11)

    Proteinase K

    ChargeSwitchMagnetic Beads

    ChargeSwitchPurification Buffer (N6)

    ChargeSwitch

    Wash Buffer (W12) ChargeSwitchElution Buffer (E5) or TE Buffer (not

    supplied; 10 mM Tris-HCl, 1 mM EDTA, pH 8.5)

    continued on next page

    9

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    16/40

    Isolating Genomic DNA from IndividualSamples, continued

    BeforeStarting

    Perform the following before beginning:

    1. Set a water bath at 37C.

    2. Prepare a Lysis Mix: For each sample, mix 1 ml of

    ChargeSwitchLysis Buffer (L11) and 10 l of ProteinaseK to prepare the Lysis Mix. If you are isolating DNAfrom multiple samples, you may scale up the volume ofreagents used and prepare a master Lysis Mix.

    3. Vortex the tube containing the ChargeSwitchMagneticBeads to fully resuspend and evenly distribute the beadsin the storage buffer.

    4. Prepare a Purification Mix: For each sample, mix 40 lof ChargeSwitchMagnetic Beads (fully resuspended;

    see above) and 100 l of ChargeSwitchPurificationBuffer (N6) to prepare the Purification Mix. If you areisolating DNA from multiple samples, you may scale upthe volume of reagents used and prepare a masterPurification Mix.

    Preparing theLysate

    Follow the procedure below to prepare a lysate from thehuman buccal cell swab.

    1. Transfer the human buccal cell sample to a sterilemicrocentrifuge tube.

    2. Add 1 ml of Lysis Mix (see above) to the tube, making

    sure that the sample is completely immersed in the LysisMix.

    3. Incubate the sample at 37C for 20 minutes.

    4. Proceed to Binding DNA, next page.

    continued on next page

    10

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    17/40

    Isolating Genomic DNA from IndividualSamples, continued

    Binding DNA Follow the procedure below to bind the DNA to theChargeSwitchMagnetic Beads.

    1. Transfer the digested supernatant (from Step 3, previouspage) into a new, sterile microcentrifuge tube.

    2. Gently pipet up and down the Purification Mixcontaining the ChargeSwitchMagnetic Beads (seeprevious page) to fully resuspend the beads.

    3. Add 140 l of Purification Mix to the sample and pipetup and down gently 5 times to mix.

    Important:Use a 1 ml pipette tip set to 900 l to mix the sample.Make sure that the tip is submerged, and pipet up and downgently to avoid forming bubbles.

    4. Incubate at room temperature for 1 minute to allow theDNA to bind to the ChargeSwitchMagnetic Beads.

    5. Place the sample in the MagnaRackfor 1 minute or until

    the beads have formed a tight pellet.6. Without removing the tube from the MagnaRack,

    carefully remove the supernatant and discard. Take carenot to disturb the pellet of beads by angling the pipettesuch that the tip is pointed away from the pellet (seefigure below).

    7. Proceed immediately to Washing DNA, next page.

    continued on next page

    11

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    18/40

    Isolating Genomic DNA from IndividualSamples, continued

    Washing DNA 1. Without removing the tube from the MagnaRack, add1 ml of ChargeSwitchWash Buffer (W12) to the sample.Direct the ChargeSwitchWash Buffer over the pellet ofmagnetic beads in such a way that the beads are brieflyresuspended in solution.

    2. Leave the sample in the MagnaRackfor 1 minute oruntil the beads have formed a tight pellet.

    3. Without removing the tube from the MagnaRack,carefully remove the supernatant and discard. Take carenot to disturb the pellet of beads by angling the pipettesuch that the tip is pointed away from the pellet (seefigure on page 11).

    4. Repeat Steps 1-3.

    5. Proceed to Eluting DNA, below.

    Eluting DNA 1. Remove the tube containing the pelleted magnetic beadsfrom the MagnaRack(Step 4, above). There should beno supernatant in the tube.

    2. Add 150 l of ChargeSwitchElution Buffer (E5) (or TEBuffer, pH 8.5) to the tube and pipet up and down gently10 times to resuspend the magnetic beads.

    Important: Do not use water for elution. The DNA will not elutedue to the poor buffering capacity of water.

    3. Incubate at room temperature for 1 minute.

    4. Place the sample in the MagnaRackfor 1 minute or untilthe beads have formed a tight pellet.

    5. Without removing the tube from the MagnaRack,carefully remove the supernatantcontaining the DNAtoa sterile microcentrifuge tube. Take care not to disturbthe pellet of beads by angling the pipette such that the tipis pointed away from the pellet (see figure on page 11).

    Note:If the eluate containing the DNA is discolored, repeatSteps 4-5.

    6. Discard the used magnetic beads. Do not reuse the beads.

    continued on next page

    12

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    19/40

    Isolating Genomic DNA from IndividualSamples, continued

    Storing DNA Store the purified DNA at -20C or use immediately for PCRor other appropriate downstream application. Avoidrepeatedly freezing and thawing DNA.

    QuantitatingDNA Yield

    To quantitate yield of your DNA, we recommend using theQuant-iTDNA Assay Kit, High Sensitivity (Catalog no.Q33120) available from Invitrogen. This kit contains a state-

    of-the-art quantitation reagent, pre-diluted standards, and apre-made buffer to allow sensitive and accuratefluorescence-based quantitation of dsDNA. For moreinformation about the Quant-iTDNA Assay Kit, seewww.invitrogen.comor call Technical Service (page 30).

    13

    http://www.invitrogen.com/http://www.invitrogen.com/
  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    20/40

    General Information Automated SampleProcessing

    Introduction This section provides general information to use theChargeSwitchgDNA Buccal Cell Kits (Catalog nos.CS11020-10 and CS11021-10) to process large numbers ofsamples in 96-well format using an automated liquidhandling robot. If you wish to process small numbers ofsamples individually, see General Information IndividualSamples, page 5.

    HardwareRequirements

    The ChargeSwitchchemistry is ideal for purification ofDNA using a liquid handling robot, avoiding the need forcentrifugation steps or the use of ethanol or chaotropic salts.You will need to have the following hardware to performautomated processing of buccal swabs using one of theChargeSwitchgDNA Buccal Cell Kits:

    Any liquid handling robotic workstation with a gripperarm

    Appropriate tips for liquid dispensing and aspiration(see below for factors to consider)

    96-Well Magnetic Separator (see page 15)

    Shaker

    Incubator, heat block, or water bath for heating samples

    96 x 2 ml deep well plate(s) (Greiner, Catalog no. 780270or Abgene, Catalog no. AB-0932)

    96 x 300 l U-Bottomed microtiter plate (Greiner,Catalog no. 650201)

    For an example of how to set up the deck, see page 16.

    Tip Selection You may use any tips of choice to dispense and aspirateliquid during the purification procedure. Consider thefollowing factors when choosing an appropriate tip to use.

    Fixed vs. disposable tips

    Tip size vs. head size

    Conductive or non-conductive

    Sterile or non-sterile

    Filtered or non-filtered

    continued on next page

    14

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    21/40

    General Information Automated SampleProcessing, continued

    96-WellMagneticSeparator

    The 96-Well Magnetic Separator available from Invitrogen(Catalog no. CS15096) is a magnetic separation rack that canhold up to 96 samples in a deep well plate. The deep wellplate fits onto the magnetic base station, associating thearray of 24 neodymium magnets with the samples for on themagnet and off the magnet sample processing (see figures

    below). For more information, see www.invitrogen.comorcall Technical Service (page 30).

    continued on next page

    15

    http://www.invitrogen.com/http://www.invitrogen.com/
  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    22/40

    General Information Automated SampleProcessing, continued

    Deck Set Up Once you have the required hardware, you will need toconfigure the deck of your liquid handling robotappropriately to process samples. You may use any suitableconfiguration of your choice. An example is provided below.

    Location Trough Contents Plate

    1 96-well Deep Well plate #1

    2 Lysis Mix (i.e.ChargeSwitchLysisBuffer (L11) + Proteinase K)

    3 Purification Mix (i.e.ChargeSwitchPurification Buffer (N6) +ChargeSwitchMagnetic Beads)

    4 ChargeSwitchWash Buffer (W12)

    5 ChargeSwitchElution Buffer (E5)

    6

    7 96-well Deep Well plate #2

    8 Waste

    9 96-well Magnetic Separator

    10 Shaker

    11 96-well Sample Tray

    12 ChargeSwitchLysis Buffer (L11)

    13 96-well U-bottomed microtiterplate (for final elution)

    continued on next page

    16

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    23/40

    General Information Automated SampleProcessing, continued

    Primary LiquidHandlingParameters

    The table below lists the primary liquid handlingparameters required to isolate DNA using the automatedprotocol. Use the parameters and guidelines provided, aswell as the protocol on pages 21-22 to program your robot.

    Parameter Aim Guidelines

    [Magnetic BeadPreparation]

    To resuspend beadsprior to mixing withsolution

    Only required once

    Beads stay in suspension for up to45 minutes

    [Mixing #1] Used to mix beads orbead/DNA pelletwith buffer

    Aspirate/dispense at 400-500 l

    Aspirate/dispense position fixed 1-2 mmabove the well bottom

    Use tips/volume setting at 80 l volume

    [Dispense liquid] Normal liquidparameters for adding

    a reagent to each well

    Aspirate/dispense at 300-400 l

    Use multi-dispense if appropriate to save

    time

    [Transfersupernatant towaste]

    To remove anddiscard supernatant

    Aspirate slowly at 50-100 l/second

    Aspirate off the entire liquid volume usingliquid detect and tracking or setting fixedheight 1 mm above the well bottom

    Do not disturb pellet

    Dispense to waste

    [Transfer

    supernatant toanother plate]

    To transfer

    supernatant toanother plate

    Aspirate slowly at 50-100 l/second

    Aspirate off the entire liquid volume usingliquid detect and tracking or setting fixedheight 1 mm above the well bottom

    Do not disturb pellet

    Dispense slowly at 50-100 l/second

    Avoid splashing

    [Final DNAElution]

    To dispense the eluatecontaining DNA

    Dispense at 10 l/second

    Aspirate from position fixed 1 mm abovethe well bottom

    Avoid bead carry-over

    Dispense into new plate at 2 mm above thewell bottom

    17

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    24/40

    Automated Genomic DNA Isolation

    Introduction This section provides a general protocol for automated

    isolation of genomic DNA from human buccal cell swabs ina 96-well format using the ChargeSwitchgDNA Buccal CellKit (Catalog no. 11021-10). Use this general protocol todevelop the script for your liquid handling robot.

    Note:If you are using the ChargeSwitchgDNA Normalized BuccalCell Kit (Catalog no. 11020-10), follow the protocol on pages 23-26.

    The ChargeSwitchMagnetic Beads supplied in theChargeSwitchgDNA Buccal Cell Kit differ in concentrationfrom those supplied in the ChargeSwitchgDNANormalized Buccal Cell Kit. Use onlythe ChargeSwitchMagnetic Beads supplied in the ChargeSwitchgDNABuccal Cell Kit in this protocol.

    MaterialsNeeded

    Have the following materials on hand before beginning:

    Liquid handling robot configured to process samples in96-well plates

    Buccal swabs

    96 x 2 ml deep well plates

    96 x 300 l U-bottomed microtiter plate

    Optional: 96 x 2 ml glass-filled, polypropylene,UnifilterMicroplate (Whatman, Catalog no. 7720-7235;see the next page)

    Components Supplied with the Kit

    ChargeSwitchLysis Buffer (L11)

    Proteinase K

    ChargeSwitchMagnetic Beads

    ChargeSwitchPurification Buffer (N6)

    ChargeSwitchWash Buffer (W12)

    ChargeSwitchElution Buffer (E5) or TE Buffer (not

    supplied; 10 mM Tris-HCl, 1 mM EDTA, pH 8.5)

    continued on next page

    Unifilteris a registered trademark of Whatman

    18

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    25/40

    Automated Genomic DNA Isolation,continued

    ImportantGuidelines

    To maximize DNA yield, follow these recommendationswhen processing your samples:

    Ensure that the robotic tips enter the wells of the plateswithout interfering with the pellet of beads.

    When removing supernatant, leave samples on the 96-Well Magnetic Separator and aspirate slowly to ensurethat the pellet of beads is not disturbed.

    When resuspending pelleted ChargeSwitchMagneticBeads, make sure that all beads are fully resuspended tomaximize DNA recovery.

    To maximize DNA yield, make sure that all WashBuffer is removed before elution.

    To maximize DNA yield, make sure that the beads arefully resuspended during the elution step.

    LysateVolume

    The first step of the genomic DNA isolation protocolrequires addition of Lysis Mix to the sample. You may addeither 1 ml or 1.4 ml of Lysis Mix to the sample. Note thatsome of the Lysis Mix may be absorbed by the swab,resulting in a lower volume of lysate being available forpurification. To maximize DNA yield, we recommend using1.4 ml of Lysis Mix (Step 2, page 21). To prepare Lysis Mix,see the next page.

    Using a FilterPlate

    To maximize recovery volume and minimize contaminanttransfer during lysate preparation, you may prepare lysatesin a 96 x 2 ml filter plate (see Steps 1-2, page 21), thendirectly filter the supernatant into a 96 x 2 ml deep wellplate to perform the remainder of the purificationprocedure. We recommend using the 96 x 2 ml glass-filledpolypropylene UnifilterMicroplate from Whatman

    (Catalog no. 7720-7235). Other 96 x 2 ml filter plates aresuitable.

    continued on next page

    19

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    26/40

    Automated Genomic DNA Isolation,continued

    BeforeStarting

    Perform the following before beginning:

    Prepare Lysis Mix: For each sample, mix 1 ml of

    ChargeSwitchLysis Buffer (L11) and 10 l ofProteinase K to prepare the Lysis Mix. If lysing in 1.4 mlvolume, mix 1.4 ml of ChargeSwitchLysis Buffer (L11)and 10 l of Proteinase K for each sample. Scale up thevolume of reagents used (based on number of samples)to prepare a master mix.

    Prepare Purification Mix: For each sample, mix 100 l

    of ChargeSwitchPurification Buffer (N5) and 40 l ofChargeSwitchMagnetic Beads to prepare thePurification Mix (make sure that the beads arethoroughly resuspended). If lysing in 1.4 ml volume,mix 140 l of ChargeSwitchPurification Buffer (N5)and 40 l of ChargeSwitchMagnetic Beads for eachsample. Scale up the volume of reagents used (based on

    number of samples) to prepare a master mix.

    continued on next page

    20

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    27/40

    Automated Genomic DNA Isolation,continued

    AutomatedProtocol

    Follow this protocol to isolate genomic DNA from buccalswabs. The volumes given are on a per sample basis.

    1. Start with 96 buccal cell samples in a 96 x 2 ml deep wellplate or 96 x 2 ml filter plate.

    2. Add 1 ml (or 1.4 ml) of Lysis Mix (see previous page) and

    incubate at 37C for 20 minutes (use a heating block).

    Note:Optimal incubation parameters (i.e.time) vary depending

    on the sample and automation plasticware, and should bedetermined empirically.

    3. After incubation, transfer or filter as appropriate, asmuch of the lysate as possible to a 96 x 2 ml deep wellplate, without interfering with the samples.

    4. Add 140 l (or 180 l if 1.4 ml of Lysis Mix used) ofPurification Mix (see previous page; make sure that the

    beads are thoroughly resuspended).

    5. Shake at medium fast speed (e.g.pulse, 10 seconds) toevenly distribute the magnetic beads within the solution.

    6. Wait for 10 seconds.

    7. Move samples to the 96-Well Magnetic Separator.

    8. Wait for 60-90 seconds.

    9. Slowly aspirate all of the supernatant and discard,leaving behind the pellet of beads.

    10. While samples are still on the 96-Well MagneticSeparator, add 1 ml of ChargeSwitchWash Buffer(W12).

    11. Wait for 60 seconds or until the beads have formed atight pellet.

    12. Slowly aspirate all of the supernatant and discard,leaving behind the pellet of beads.

    13. While samples are still on the 96-Well MagneticSeparator, add 1 ml of ChargeSwitchWash Buffer(W12).

    continued on next page

    21

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    28/40

    Automated Genomic DNA Isolation,continued

    AutomatedProtocol,continued

    14. Wait for 30-60 seconds.

    15. Slowly aspirate all of the supernatant and discard,leaving behind the pellet of beads.

    16. Move samples to the shaker.

    17. Add 150 l of ChargeSwitchElution Buffer (E5).

    Note:You may vary elution volume depending on your needs.

    Do not elute in volumes < 60 l as the DNA may not completely

    elute from the beads.

    18. Shake rapidly for 1-2 minutes to completely disperse thebeads within the solution.

    19. Move samples to the 96-Well Magnetic Separator.

    20. Wait for 1 minute.

    21. Slowly aspirate supernatant containing the DNAto a

    96 x 300 l U-bottomed microtiter plate.

    Storing DNA Store the purified DNA at -20C or use immediately fordownstream applications such as PCR. Avoid repeatedlyfreezing and thawing DNA.

    QuantitatingDNA Yield

    To quantitate yield of your DNA, use the Quant-iTDNAAssay Kit, High Sensitivity (Catalog no. Q33120). For more

    information, see page 13.

    22

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    29/40

    Automated Genomic DNA Isolation Normalized Kit

    Introduction This section provides a general protocol for automatedisolation of genomic DNA from human buccal cell swabs ina 96-well format using the ChargeSwitchgDNANormalized Buccal Cell Kit (Catalog no. 11020-10). Use thisgeneral protocol to develop the script for your liquidhandling robot.

    Note:If you are using the ChargeSwitchgDNA Buccal Cell Kit(Catalog no. 11021-10), follow the guidelines and protocol on

    pages 18-22.

    The ChargeSwitchMagnetic Beads supplied in theChargeSwitchgDNA Normalized Buccal Cell Kit differ inconcentration from those supplied in the ChargeSwitchgDNA Buccal Cell Kit. Use onlythe ChargeSwitchMagnetic Beads supplied in the ChargeSwitchgDNANormalized Buccal Cell Kit in this protocol.

    MaterialsNeeded

    Have the following materials on hand before beginning:

    Liquid handling robot configured to process samples in96-well plates

    Buccal swabs

    96 x 2 ml deep well plates

    96 x 300 l U-bottomed microtiter plate

    Optional: 96 x 2 ml glass-filled, polypropylene,UnifilterMicroplate (Whatman, Catalog no. 7720-7235;see the next page)

    Components Supplied with the Kit

    ChargeSwitchLysis Buffer (L11)

    Proteinase K

    ChargeSwitchMagnetic Beads

    ChargeSwitchPurification Buffer (N6)

    ChargeSwitchWash Buffer (W12)

    ChargeSwitchElution Buffer (E5) or TE Buffer (notsupplied; 10 mM Tris-HCl, 1 mM EDTA, pH 8.5)

    continued on next page

    23

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    30/40

    Automated Genomic DNA Isolation Normalized Kit, continued

    ImportantGuidelines

    To maximize DNA yield, follow these recommendationswhen processing your samples:

    Ensure that the robotic tips enter the wells of the plateswithout interfering with the pellet of beads.

    When removing supernatant, leave samples on the 96-Well Magnetic Separator and aspirate slowly to ensurethat the pellet of beads is not disturbed.

    When resuspending pelleted ChargeSwitchMagneticBeads, make sure that all beads are fully resuspended tomaximize DNA recovery.

    To maximize DNA yield, make sure that all WashBuffer is removed before elution.

    To maximize DNA yield, make sure that the beads arefully resuspended during the elution step.

    Using a FilterPlate

    Some of the Lysis Mix may be absorbed by the sample,resulting in a lower volume of lysate being available forpurification. To maximize recovery volume and minimizecontaminant transfer during lysate preparation, you mayprepare lysates in a 96 x 2 ml filter plate (see Steps 1-2,page 25), then directly filter the supernatant into a 96 x 2 mldeep well plate to perform the remainder of the purificationprocedure. We recommend using the 96 x 2 ml glass-filled

    polypropylene UnifilterMicroplate from Whatman(Catalog no. 7720-7235). Other 96 x 2 ml filter plates aresuitable.

    continued on next page

    24

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    31/40

    Automated Genomic DNA Isolation Normalized Kit, continued

    BeforeStarting

    Perform the following before beginning:

    Prepare Lysis Mix: For each sample, mix 1 ml of

    ChargeSwitchLysis Buffer (L11) and 10 l ofProteinase K to prepare the Lysis Mix. Scale up thevolume of reagents used (based on number of samples)to prepare a master mix.

    Prepare Purification Mix: For each sample, mix 100 l

    of ChargeSwitch

    Purification Buffer (N5) and 40 l ofChargeSwitchMagnetic Beads to prepare thePurification Mix (make sure that the beads arethoroughly resuspended). Scale up the volume ofreagents used (based on number of samples) to preparea master mix.

    Automated

    Protocol

    Follow this protocol to isolate genomic DNA from buccal

    swabs. The volumes given are on a per sample basis.1. Start with 96 buccal cell samples in a 96 x 2 ml deep well

    plate or 96 x 2 ml filter plate.

    2. Add 1 ml of Lysis Mix (see above) and incubate at 37Cfor 20 minutes (use a heating block).

    Note:Optimal incubation parameters (i.e.time) vary dependingon the sample and automation plasticware, and should bedetermined empirically.

    3. After incubation, transfer or filter as appropriate, asmuch of the lysate as possible to a 96 x 2 ml deep wellplate, without interfering with the samples.

    4. Add 140 l of Purification Mix (see above; make sure thatthe beads are thoroughly resuspended).

    5. Shake at medium fast speed (e.g.pulse, 10 seconds) toevenly distribute the magnetic beads within the solution.

    6. Wait for 10 seconds.

    7. Move samples to the 96-Well Magnetic Separator.

    8. Wait for 60-90 seconds.

    9. Slowly aspirate all of the supernatant and discard,leaving behind the pellet of beads.

    continued on next page

    25

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    32/40

    Automated Genomic DNA Isolation Normalized Kit, continued

    AutomatedProtocol,continued

    10. Remove samples from the 96-Well Magnetic Separator.

    11. Add 500 l of ChargeSwitchWash Buffer (W12).

    12. Shake at medium fast speed (e.g.pulse, 10 seconds) toevenly distribute the magnetic beads within the solution.

    13. Move samples to the 96-Well Magnetic Separator.

    14. Wait for 60-90 seconds.

    15. Slowly aspirate all of the supernatant and discard,leaving behind the pellet of beads.

    16. While samples are still on the 96-Well Magnetic

    Separator, add 500 l of ChargeSwitchWash Buffer(W12).

    17. Wait for 30-60 seconds.

    18. Slowly aspirate all of the supernatant and discard,leaving behind the pellet of beads.

    19. Move samples to the shaker.

    20. Add 150 l of ChargeSwitchElution Buffer (E5).

    21. Shake rapidly for 1-2 minutes to completely disperse thebeads within the solution.

    22. Move samples to the 96-Well Magnetic Separator.

    23. Wait for 1 minute.

    24. Slowly aspirate supernatant containing the DNAto a

    96 x 300 l U-bottomed microtiter plate.

    Storing DNA Store the purified DNA at -20C or use immediately fordownstream applications such as PCR. Avoid repeatedlyfreezing and thawing DNA.

    QuantitatingDNA Yield

    To quantitate the yield of your DNA, use the Quant-iT

    DNA Assay Kit, High Sensitivity (Catalog no. Q33120). Formore information, see page 13.

    26

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    33/40

    Troubleshooting

    Introduction Refer to the table below to troubleshoot problems that you

    may encounter when purifying genomic DNA with the kit.

    Problem Cause Solution

    Low DNA yield Incomplete lysis Be sure to add Proteinase Kduring lysis.

    Increase the length of

    incubation at 37C.

    Poor quality ofstarting material Process samples immediatelyafter collection or store the

    sample at 4C. Do not storesamples at room temperature as

    bacteria and nucleases presentin the sample will degrade theDNA.

    Insufficient amount ofChargeSwitch

    Magnetic Beads added

    Vortex the tube containing theChargeSwitchMagnetic

    Beads to fully resuspend thebeads in solution beforepreparing the PurificationMix.

    Before adding PurificationMix to your sample, makesure that the beads are fullyresuspended.

    Pellet of beadsdisturbed or lostduring binding orwashing steps

    Keep the sample in theMagnaRackor 96-WellMagnetic Separator whenremoving supernatant duringthe binding or washing steps.

    Remove the supernatantwithout disturbing the pelletof beads by angling thepipette tip away from the

    pellet.

    Bubbles formedduring mixing steps

    Make sure that the pipette tip issubmerged in the solutionduring mixing.

    continued on next page

    27

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    34/40

    Troubleshooting, continued

    Problem Cause Solution

    Low DNA yield,continued

    Incomplete disso-ciation of DNA fromthe ChargeSwitchMagnetic Beads

    Perform additional mixing ofthe suspension of beads (bypipetting up and down).

    Incorrect elutionconditions

    After adding ChargeSwitchElution Buffer (E5) to thesample, pipet up and down tofully resuspend the magnetic

    beads before incubation.

    Do not use water for elution.Use ChargeSwitchElutionBuffer (E5) or TE, pH 8.5.

    Lysate mixed toovigorously or smallpipette tips usedduring mixing

    Use the appropriate pipettetip set to a volume lower thanthe total volume of solution inthe sample.

    Pipet up and down gently tomix.

    No DNA recovered Water used for elution Do not use water for elution.The elution buffer musthave apH = 8.5-9.0 or the DNA willremain bound to theChargeSwitchMagnetic Beads.Use Elution Buffer (E5) or TE,

    pH 8.5.ChargeSwitchMagnetic Beads storedor handledimproperly

    Store beads at roomtemperature. Do not freeze the

    beads as they will becomeirreparably damaged.

    Make sure that the beads arein solution at all times and donot become dried. Dried

    beads are non-functional.

    Purification Mix didnot containChargeSwitchMagnetic Beads

    Purification Mix should containChargeSwitchPurificationBuffer (N6) + ChargeSwitchMagnetic Beads.

    continued on next page

    28

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    35/40

    Troubleshooting, continued

    Problem Cause Solution

    DNA is degraded Buccal swabs stored atroom temperature

    Process buccal swabsimmediately after collection or

    store at 4C.

    DNA yields varywidely betweensamples

    Variability in samplecollection

    DNA yields can vary dependingon swabbing technique, whetherthe donor is a high or lowshedder, and the type of swabused.

    29

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    36/40

    Appendix

    Technical Service

    World WideWeb

    Visit the Invitrogen Web Resource using your World WideWeb browser. At the site, you can:

    Get the scoop on our hot new products and specialproduct offers

    View and download vector maps and sequences

    Download manuals in AdobeAcrobat(PDF) format

    Explore our catalog with full color graphics

    Obtain citations for Invitrogen products

    Request catalog and product literature

    Once connected to the Internet, launch your Web browser(Internet Explorer 5.0 or newer or Netscape 4.0 or newer),then enter the following location (or URL):

    http://www.invitrogen.com

    ...and the program will connect directly. Click on underlinedtext or outlined graphics to explore. Don't forget to put a

    bookmark at our site for easy reference!

    Contact Us For more information or technical assistance, call, write, fax,or email. Additional international offices are listed on ourWeb page (www.invitrogen.com).

    Corporate Headquarters:

    European Headquarters:Invitrogen Corporation Invitrogen Ltd

    1600 Faraday Avenue Inchinnan Business Park

    Carlsbad, CA 92008 USA 3 Fountain Drive

    Tel: 1 760 603 7200 Paisley PA4 9RF, UK

    Tel (Toll Free): 1 800 955 6288 Tel: +44 (0) 141 814 6100

    Fax: 1 760 602 6500 Tech Fax: +44 (0) 141 814 6117

    E-mail:[email protected]

    E-mail:[email protected]

    continued on next page

    30

    http://www.invitrogen.com/http://www.invitrogen.com/mailto:[email protected]:[email protected]:[email protected]:[email protected]://www.invitrogen.com/http://www.invitrogen.com/
  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    37/40

    Technical Service, continued

    MSDS

    Requests

    To request an MSDS, visit our Web site at

    www.invitrogen.com. On the home page, go to TechnicalResources, select MSDS, and follow instructions on thepage.

    LimitedWarranty

    Invitrogen is committed to providing our customers with high-quality goods and services. Our goal is to ensure that everycustomer is 100% satisfied with our products and our service. If youshould have any questions or concerns about an Invitrogen productor service, contact our Technical Service Representatives.

    Invitrogen warrants that all of its products will perform accordingto specifications stated on the certificate of analysis. The companywill replace, free of charge, any product that does not meet thosespecifications. This warranty limits Invitrogen Corporationsliability only to the cost of the product. No warranty is granted forproducts beyond their listed expiration date. No warranty isapplicable unless all product components are stored in accordancewith instructions. Invitrogen reserves the right to select themethod(s) used to analyze a product unless Invitrogen agrees to a

    specified method in writing prior to acceptance of the order.Invitrogen makes every effort to ensure the accuracy of itspublications, but realizes that the occasional typographical or othererror is inevitable. Therefore Invitrogen makes no warranty of anykind regarding the contents of any publications or documentation. Ifyou discover an error in any of our publications, please report it toour Technical Service Representatives.

    Invitrogen assumes no responsibility or liability for any special,incidental, indirect or consequential loss or damage whatsoever.The above limited warranty is sole and exclusive. No other

    warranty is made, whether expressed or implied, including anywarranty of merchantability or fitness for a particular purpose.

    31

    http://www.invitrogen.com/http://www.invitrogen.com/
  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    38/40

    Purchaser Notification and ProductQualification

    PurchaserNotification

    Limited Use Label License No. 265: ChargeSwitchTechnology

    The use of this product may be covered by European PatentNo. EP1036082B1 and foreign equivalents.

    ProductQualification

    Each kit is functionally tested to ensure conformance with themost current approved product specifications. Current

    specifications consist of tests for:

    Bead size, charge, and binding capacity

    Nucleic acid quality and quantity

    Buffer turbidity, volume, and absence of RNases andDNases

    Kit packaging and labeling accuracy

    For individual lot test results and more information, visit

    www.invitrogen.com to download the Certificate of Analysis.

    2005 Invitrogen Corporation. All rights reserved.

    For research use only. Not intended for any animal or humantherapeutic or diagnostic use.

    32

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    39/40

  • 8/11/2019 Chargeswitch GDNA - Buccal - Manual

    40/40

    Corporate Headquarters:Invitrogen Corporation1600 Faraday AvenueCarlsbad, California 92008Tel: 1 760 603 7200Tel (Toll Free): 1 800 955 6288Fax: 1 760 603 7229Email: [email protected]

    European Headquarters:Invitrogen Ltd

    3 Fountain DriveInchinnan Business ParkPaisley PA4 9RF, UKTel (Free Phone Orders): 0800 269 210Tel (General Enquiries): 0800 5345 5345Fax: +44 (0) 141 814 6287Email: [email protected]

    International Offices:Argentina 5411 4556 0844Australia 1 800 331 627Austria 0800 20 1087Belgium 0800 14894Brazil 0800 11 0575

    Canada 800 263 6236China 10 6849 2578Denmark 80 30 17 40

    France 0800 23 20 79Germany 0800 083 0902Hong Kong 2407 8450India 11 577 3282Italy 02 98 22 201

    Japan 03 3663 7974The Netherlands 0800 099 3310New Zealand 0800 600 200Norway 00800 5456 5456

    Spain & Portugal 900 181 461Sweden 020 26 34 52Switzerland 0800 848 800Taiwan 2 2651 6156UK 0800 838 380For other countries see our Web site


Recommended