Date post: | 12-Jun-2015 |
Category: |
Documents |
Upload: | boris-cuesta-vasquez |
View: | 248 times |
Download: | 0 times |
Constant Climate Chamber HPP
13.04.2023
HPP1
Unique selling propositions
Applications
References
ServiceEquipment
Techn. Data
Technology
HOME
Unique Selling Propositions
13.04.2023
HPP2
Extremely economical For stable and precise long-term tests Perfect controllability of temperature and humidity Optimally tuned to the requirements for all ongoing stability
tests (ICH-Q1) high precision in temperature and humidity control competitive price / price-performance ratio
ecological requirements - minimum consumption, energy saving - almost vibration free/low noise
service-friendly
HOME
Technical Data
13.04.2023
HPP3
HPP
Volume 108, 749 litres
Temperature range (without light) +5°C to +70°C
Temperature range (with light) +10°C to +40°C
Humidity range (with light) 10% to 85% rh
Humidity range (without light) 10% to 90% rh
Temperature variation in time(at 37°C)
≤ +/- 0,1°C
Temperature uniformity (at 37°C)
≤ +/- 0,3°C
Accuracy humidity sensor 0,5% rh
Humidity variation in time ≤ +/- 1,5% rh
Safety class Class3.3
Alarm Visual and accoustic
Comparison: Energy consumption peltier/compressor
13.04.2023
HPP4
Peltier technology is particularly economical at temperatures close to ambient temperature
Energy is only required if heating or cooling is needed
Closed Peltier cooling system, no outside air is exchanged
No de-humidification of the chamber and heavily reduced risk of contamination
13.04.2023
HPP5
Simulation of climatic zones on earth
I
II
III
0
10
20
30
40
50
60
70
80
90
100
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95
rela
tive
hu
mid
ity
%rH
temperature °C
Climatic zone I: 21 °C / 45 % rH
Moderate climate
Climatic zone II: 25 °C / 60 % rH / Sub-tropical climate / Mediterranien climate
Climatic zone III: 30 °C / 35 % rH
Hot and dry climate
Climatic zone IV: 30 °C / 65 % rH / Hot and
Humid climate
Intermediate stability test :
30°C +/- 2 °C,
65 % +/- 5 % rh
duration: 6 month
Accelerated test:
40°C +/- 2 °C
75% +/- 5% rh
duration: 6 month
Long term tests:
25°C +/- 2 °C
60% +/-5% rh
duration: 12 month
Requirements on ICH Q1A
I V
HOME
13.04.2023
HPP6
Equipment - standard
2 stainless steel grids Works calibration certificate at +10°C and +37°C at 60% rh Software „Celsius“ MEMorycard USB and printer interface Capactive humidity sensor Active microprocessor control for humidifying and dehumidiying
Constant Climate Chamber HPP 108 NOW with light
HPP7
Volume 108 l
Temperature rangewithout light +5°C up to +70°Cwith light +10°C up to +40°C
Humidity rangewithout light 10 – 90% rh with light 10 – 85% rh
Light efficiency 10.000 Lux 2 LED- light modules available:Cold white light 5.500 K Combination of cold white light 5.500 K and warm white 2.700 K
Equipment - options
13.04.2023
HOME
13.04.2023
HPP8
Equipment - accessories
Locking door security Moisture-tight entry ports Stacking version IQ/OQ documentation
Technology
13.04.2023
HPP9
Advantages of Peltier – Technology
Space-saving No cooling liquids necessary(GreenLab) Setting accuracy Cooling or heating Vibration-free
Technology
13.04.2023
Produktpräsentation HPP10
P-material N-material
Kühlen
Heizen
HOME
Service - Maintenance
13.04.2023
HPP11
HOME
Service-Manual
Maintenance
Service / Error-Messages
Calibration / Adjustment
Application – HPP 108 Light
13.04.2023
HPP12
Plant growing
13.04.2023
HPP13
Breeding of insects
Application – HPP 108 Light
13.04.2023
HPP14
Germination
Application – HPP 108 Light
13.04.2023
HPP15
Breeding of protoplasma and cells
Application – HPP 108 Light
HOME
References
RWTH Aachen: Abteilung Pflanzenphysiologie
13.04.2023
HPP16
HOME