STILL, the Most Innovative and Versatile Coaxial System that... Protects both inspiratory and expiratory machine ports with a single, dual-chamber, high efficiency filter Helps maintain patient body temperature as reported in independent studies 1,2,5 Reduces the torque on the airway device, because the lightweight flex tubing can be: > Custom shaped near the airway > Expanded to the ideal length between patient and machine to minimize draping Reduce the number of circuit lengths stocked, due to flex tubing that expands to multiple lengths Minimizes circuit types stocked, because the Flex2 can be used wherever an adult or pediatric two-limb is used > Whenever an adult or pediatric two limb is used the effective inner and outer diameter is equivalent to a standard two limb pediatric circuit. > Elbow of the Flex2 has no more dead space than a pediatric 2-limb wye Requires significantly less storage space, because flex tubing is packaged at its most compressed length KingFlex2 TM Coaxial breathing circuit Produces less overall disposable waste for the hospital when using Flex2 for surgery, transport, and the PACU Can provide lower cost options when using the Flex2 for surgery, transport, and the PACU Protects Both Patient and Machine Simplifies Inventory Reduces Disposable Waste and Cost
Transcript
Page 1: Coaxial breathing circuit · 1. The Universal F Breathing System Reduces the Number of Hypothermic Patients Entering the PACU. Thompson, Rast, Carson. 2. Evaluation of the King Systems

STILL, the Most Innovative and Versatile Coaxial System that...

• Protects both inspiratory and expiratory machine ports with a single, dual-chamber, high efficiency filter• Helps maintain patient body temperature as reported in independent studies1,2,5

• Reduces the torque on the airway device, because the lightweight flex tubing can be: > Custom shaped near the airway > Expanded to the ideal length between patient and machine to minimize draping

• Reduce the number of circuit lengths stocked, due to flex tubing that expands to multiple lengths

• Minimizes circuit types stocked, because the Flex2 can be used wherever an adult or pediatric two-limb

is used > Whenever an adult or pediatric two limb is used the effective inner and outer diameter is

equivalent to a standard two limb pediatric circuit.

> Elbow of the Flex2 has no more dead space than a pediatric 2-limb wye

• Requires significantly less storage space, because flex tubing is packaged at its most compressed length

KingFlex2TM

Coaxial breathing circuit

• Produces less overall disposable waste for the hospital when using Flex2 for surgery, transport, and the PACU• Can provide lower cost options when using the Flex2 for surgery, transport, and the PACU

Protects Both Patient and Machine

Simplifies Inventory

Reduces Disposable Waste and Cost

Page 2: Coaxial breathing circuit · 1. The Universal F Breathing System Reduces the Number of Hypothermic Patients Entering the PACU. Thompson, Rast, Carson. 2. Evaluation of the King Systems

KingFlex2 Filter and HME Performance Data5,6

* Bacterial efficiency determined per test method ASTM F2101, which simulates the actual airborne transmission of Staphylococcus aureus in water droplets.6

** Viral efficiency determined per test method ASTM F2101, which simulates the actual airborne transmission of bacteriophage phi x174 in water droplets.6

*** Particulate efficiency determined per Nelson Laboratories’ test protocol, which transmits dry and inert 0.3um airborne particulates. An efficiency of >99.97% is a requirement for a filter to be rated as HEPA.4

Ambu Inc.6230 Old Dobbin Lane Columbia, MD 21045 Tel. 800 262-8462 Fax 800 262-8673ambuUSA.com

Circuits

Thermal Efficiency 2,5 (oC @ 4l/m)Absolute

Humidity 5

@500ml Vt

Insp

Temp

Initial

Insp

Temp

@ 15 min

Insp

Temp

King single-limb 23.2 29.2 + 6.0 26.3 mg/l

Std two-limb 23.1 23.9 + 0.8 18.3 mg/l

KingFlex2 Circuit Performance Data

1. The Universal F Breathing System Reduces the Number of Hypothermic Patients Entering the PACU. Thompson, Rast, Carson.2. Evaluation of the King Systems Universal F Anesthesia Breathing Circuit. Campbell, Branson, Johnson, et. al.3. Transmission by water droplets per ASTM F2101 test method, CDC Principles of Epidemiology, and Boston University ”Transmission of Infectious Disease”.

4. Examples of diseases spread by airborne respiratory water droplets. Boston University ”Transmission of Infectious Disease” and Wikipedia-Airborne-Disease.5. Data on file.6. Testing and reported results by Nelson Laboratories

Part # Type Size DesignCO2Port

Volume

(ml)

Weight

(g)

Resistance

(cm H2O)

Absolute

Humidity

Bacterial

Efficiency*

Viral

Efficiency**

0.3um

Particulate

Efficiency ***

F2 Dual Filter FILTER DUAL Straight No 149 60 1.0 @ 0.5 I/s n/a 99.9999% 99.9996% >99.97%

GVS-4244HEPA

HMEFSTND Straight Yes 52 40 1.4 @ 0.5 l/s 30 @ 500ml Vt 99.9999% 99.9999% >99.97%

MS-111 HMEF STND Straight Yes 55 34 1.1 @ 0.5 I/s 38 @ 1000ml Vt 99.99% 99.99% ---

KN- HMEF HMEF SML Straight Yes 38 25 2.5 @ 0.5 I/s 31 @ 500ml Vt >99.99% >99.99% ---

MS-110 HME STND Straight Yes 55 33 0.5 @ 0.5 I/s 38 @ 1000ml Vt --- --- ---

MS-102 HME SML Straight No 22 13 0.7 @ 0.5 I/s 35 @ 500ml Vt --- --- ---

Data on Common Bacteria and Viruses- Including Those Used for Filter Efficiency Tests

Flex2

Inspiratory

Resistance 5

(cm H2O)

Expiratory

Resistance 5

(cm H2O)

Compliance 5

(ml/cm H2O)

72” 1.23 @ 0.5 l/s 1.41 @ 0.5 l/s 2.20 @ 60cm H2O

108” 1.67 @ 0.5 l/s 2.04 @ 0.5 l/s 2.83 @ 60cm H2O

Key/Options:• 72” (1.8m) Flex2 circuit, expandable from 21” to 72” (0.5 to 1.8m)• 108” (2.7m) Flex2 circuit, expandable from 38” to 108” (1.0 to 2.7m)• Circuit extensions expandable up to 72” (1.8m)

493

5041

31 -

V0

1 -

03/

2019

- A

mbu Inc.

Tech

nic

al d

ata

may

be

mo

dif

ied w

itho

ut

furt

her

noti

ce.

References/Studies:

US: Rx only

NameBacteria, Virus, or

Particulate

Flex2 Filter

Efficiency6

Bacteria/Virus/

Particulate

Average Size (um)5

Transmission of Diseases

(actual or in tests)3,4,5

Staphylococcus aureus TEST bacteria 99.9999% 0.90 diameter airborne water droplets

Neisseria meningitidis bacteria --- 0.80 diameter airborne water droplets

Bacillus anthracis bacteria --- 6.50 length1.15 diameter airborne water droplets

Myobacterium tuberculosis bacteria --- 2.75 length0.35 diameter airborne water droplets

Phi X174 TEST virus 99.9996% 0.027 diameter airborne water droplets

Flu type A- H1N1 & H3N2 virus --- 0.100 diameter airborne water droplets

Flu type B- Victoria & Yamagata virus --- 0.100 diameter airborne water droplets

Norovirus virus --- 0.030 diameter airborne water droplets

Inert particulate for HEPA testing TEST particulate >99.97% 0.300 diameter dry particulate only


Recommended