Case Report COMMON ARTIFACTS OF PULMONARY ARTERY AND PULMONARY ARTERYWEDGEPRESSURES: RECOGNITIONAND INTERPRETATION Edward A. Schrnitt, MD, and C. 0. Brantigan, AID, FACS, FCCP Schmitt EA, Brantigan CO. Common artifacts of pulmonary artery and pulmonary artery wedge pressure: Recognition and interpreta- tion. J Clin Monit 1986;2:44-52 ABSTRACT. Bedside measurement of pulmonary artery pres- sure and pulmonary artery wedge pressure has an important role in the management of critically ill patients. Unfortu- nately, waveform abnormalities and artifacts commonly dis- tort numeric values and lead to incorrect therapeutic decisions. The clinical impact of these artifacts is magnified by the digital pressure displays used in most intensive care units. We present here an atlas and an analysis of the artifacts that commonly occur. Use of analog rather than digital pulmonary artery wedge pressure data, when combined with an understanding of the physiological characteristics of patients, can prevent critical errors in patient management. KEY WORDS. Lung: pulmonary artery pressure; pulmonary ar- tery wedge pressure. Monitoring: blood pressure, artifact Since the introduction of Swan-Ganz cardiac catheters [1], cardiac catheterization has become a frequently per- formed bedside procedure in intensive care units. The resulting hemodynamic data provide a picture of the patient's status that may be invaluable. Data are dis- played on highly technological digital bedside moni- tors. Unfortunately, digital readouts may detect sim- ply the highs and lows during some standard interval and produce data that are entirely artifactual. Hence, measurements designed to reassure the physician or lead to accurate decisions instead frequently become counterproductive. Analysis of an analog strip-chart re- cording of the pulmonary artery wedge pressure tracing and observation of the status of each individual patient can eliminate artifact from these vital measurements. We present here an atlas of the common artifacts, an analysis of how they occur, and a simple way to handle them using a calibrated analog recorder. From the Department of Surgery, Presbyterian/St. Luke's Medical Center, and St. Joseph Hospital, Denver, CO. Received Dec 24, 1984, and in revised form Aug 14, 1985. Accepted for publication Sep 5, 1985. Address correspondence to Dr Brantigan, 1746 Lafayette St, Denver, CO, 80218. MATERIALS AND METHODS Patients for this study were selected from the intensive care unit at Denver Presbyterian Hospital of Presbyte- rian/St Luke's Medical Center when the bedside moni- tor indicated a high probability of important artifact in pulmonary artery or pulmonary artery wedge pres- sures. Digital readouts made by a model #3474CS4JK digital readout computer (General Electric) were com- pared with an analog reading of pulmonary artery and pulmonary artery wedge pressures obtained with a model RI-5DC recorder (General Scanning, Inc). This device interfaced well with the General Electric system. We wished to obtain characteristic examples, therefore, no attempt was made to assess the frequency of 44
Transcript
Page 1: Common artifacts of pulmonary artery and pulmonary artery ... · tive artifact and a 6-mm positive artifact. The mean pressure recorded by the digital equipment was an aver- age of

Case Report

COMMON ARTIFACTS OF PULMONARY ARTERY AND PULMONARY ARTERY WEDGE PRESSURES: RECOGNITION AND INTERPRETATION Edward A. Schrnitt, MD, and C. 0. Brantigan, AID, FACS, FCCP

Schmitt EA, Brantigan CO. Common artifacts of pulmonary artery and pulmonary artery wedge pressure: Recognition and interpreta- tion.

J Clin Monit 1986;2:44-52

ABSTRACT. Bedside measurement of pulmonary artery pres- sure and pulmonary artery wedge pressure has an important role in the management of critically ill patients. Unfortu- nately, waveform abnormalities and artifacts commonly dis- tort numeric values and lead to incorrect therapeutic decisions. The clinical impact of these artifacts is magnified by the digital pressure displays used in most intensive care units. We present here an atlas and an analysis of the artifacts that commonly occur. Use of analog rather than digital pulmonary artery wedge pressure data, when combined with an understanding of the physiological characteristics of patients, can prevent critical errors in patient management.

KEY WORDS. Lung: pulmonary artery pressure; pulmonary ar- tery wedge pressure. Monitoring: blood pressure, artifact

Since the introduction of Swan-Ganz cardiac catheters [1], cardiac catheterization has become a frequently per- formed bedside procedure in intensive care units. The resulting hemodynamic data provide a picture of the patient's status that may be invaluable. Data are dis- played on highly technological digital bedside moni- tors. Unfortunately, digital readouts may detect s im- ply the highs and lows during some standard interval and produce data that are entirely artifactual. Hence, measurements designed to reassure the physician or lead to accurate decisions instead frequently become counterproductive. Analysis o f an analog strip-chart re- cording o f the pulmonary artery wedge pressure tracing and observation o f the status o f each individual patient can eliminate artifact f rom these vital measurements. We present here an atlas o f the common artifacts, an analysis o f how they occur, and a simple way to handle them using a calibrated analog recorder.

From the Department of Surgery, Presbyterian/St. Luke's Medical Center, and St. Joseph Hospital, Denver, CO. Received Dec 24, 1984, and in revised form Aug 14, 1985. Accepted for publication Sep 5, 1985.

Address correspondence to Dr Brantigan, 1746 Lafayette St, Denver, CO, 80218.

MATERIALS AND METHODS

Patients for this study were selected f rom the intensive care unit at Denver Presbyterian Hospital o f Presbyte- rian/St Luke's Medical Center when the bedside moni- tor indicated a high probabili ty o f important artifact in pulmonary artery or pulmonary artery wedge pres- sures. Digital readouts made by a model #3474CS4JK digital readout computer (General Electric) were com- pared with an analog reading o f pulmonary artery and pulmonary artery wedge pressures obtained with a model RI -5DC recorder (General Scanning, Inc). This device interfaced well with the General Electric system. We wished to obtain characteristic examples, therefore, no at tempt was made to assess the frequency of

44

Page 2: Common artifacts of pulmonary artery and pulmonary artery ... · tive artifact and a 6-mm positive artifact. The mean pressure recorded by the digital equipment was an aver- age of

Case Report: Schmitt and Brantigan: Pressure Measurement Artifacts 45

significant differences between the methods; however, we did note that artifacts of varying importance oc- curred in the majority of the patients in the intensive care unit. All standard precautions were taken to assure that accurate data would be obtained by both methods. Proper placement of the pulmonary artery pressure catheter was ascertained, and the pressure lines to the patient were examined to make certain that they were free of air bubbles. The transducers were inspected to make certain that they were free of air and that they contained an appropriate amount of water. The analog strip recorder was carefully zeroed to atmospheric pres- sure at precisely the same altitude as the patient's heart. The patients were supine in bed during the recordings. The digital readout computer provided output exactly as it was designed to on all patients, and agreement with the chart recorder was excellent. However, the digital display always interpreted artifacts or abnormal pressure waves that satisfied the programmed criteria as the pulmonary artery systolic, diastolic, or wedge pres- sures. This occurred because the algorithm for digital display looks at a segment of the pressure tracing, identifies the highest value seen during that interval as systolic, the lowest as diastolic, and continuously aver- ages the waveform to produce a mean. The "mean" reading is usually used to indicate the pulmonary artery wedge pressure. While other, more advanced, al- gorithms improve on this simplistic approach [2], our monitors demonstrate the problem intrinsic to all such digital displays; namely, that they are unable to com- pletely eliminate artifactual waveforms from considera- tion in their associated algorithms. We recommend in- terpretation of the calibrated analog wave and reading the pulmonary artery wedge pressures at end ex- piration.

The "wedge" state is defined in this paper as the state in which the inflated catheter balloon appears to just occlude the pulmonary artery. We attempted to identify each pressure wave seen on a pulmonary artery wedge pressure tracing and selected as the "true wedge" the portion o f the tracing that most closely approximated left ventricular filling pressure. Occasionally, tracings are uninterpretable and must be recognized as such. Un- interpretable tracings most commonly occur when heart rate and respiratory rate are similar. More often there is a relatively stable portion of the trace in the midst of artifacts and abnormal pressure waves that appears to approximate left ventricular filling pressure. When significant V waves are present we read the wedge at the valley of these tracings; we believe that the regurgitant wave does not represent ventricular filling pressure. In the case of respiratory artifacts we attempt to read the values at end expiration as this is the point at which

atmospheric and alveolar pressure values are most nearly equal. Although most respiratory artifacts are easily identified from examining the tracing or by watching the patient while data are collected, direct air- way pressure recordings are helfpul. Most intensive care units do not have the capability for measuring airway pressures continuously, however, and in our experience such recordings are usually not necessary. The examples given in Figures 1-12, all recorded at a chart speed of 5 mm/s, demonstrate how digital displays lead to errors and illustrate how to interpret an analog waveform.

RESULTS

Figure 1 shows a negative respiratory artifact. The pa- tient had labored breathing secondary to pneumonia, sepsis, and pulmonary edema. The wedge pressure trac- ing (PAW) demonstrates large negative excursions (A); hence, the mean wedge pressure (an average of high and low when determined digitally) was less than 0. The actual wedge pressure was approximately 10 mm Hg and could have been accurately measured by setting the digital readout to "systolic." A similar distortion was produced in the pulmonary artery diastolic (PAD) trac- ing. The true pulmonary artery systolic pressure (PAS) is marked by small arrows. The true diastolic pressure, identified by the large arrows, was approximately 30 mm Hg. The digital equipment interpreted the pulmo- nary artery diastolic pressure at 12 mm Hg. This artifact is seen commonly in patients who generate a large nega- tive intrathoracic pressure to breathe, and can occur when patients are either on or off the ventilator.

Figure 2 shows a positive respiratory artifact. This is the pulmonary artery pressure tracing of a patient re- ceiving controlled ventilation. The patient was making no attempt to breathe on his own, but each time the respirator cycled it created positive intrapleural pressure that produced an artifact (A). The artifact was averaged by the digital equipment into the wedge pressure trac- ing, producing a falsely high reading. The pulmonary artery systolic pressure was distorted in the same fash- ion. Whereas the digital equipment interpreted the sys- tohc pressure as 50 m m Hg, it was actually only 40 mm Hg.

Placing the patient whose tracing is shown in Figure 1 on a mechanical ventilator, such as was done in the second example, could change a falsely low pulmonary artery wedge pressure to a falsely high one. Thus the act of placing a patient on a ventilator could easily cause an increase of 20 m m Hg in the wedge pressure displayed by digital equipment, when in fact there is no marked change in the patient's hemodynamic status.

Page 3: Common artifacts of pulmonary artery and pulmonary artery ... · tive artifact and a 6-mm positive artifact. The mean pressure recorded by the digital equipment was an aver- age of

46 Journal of Clinical Monitoring Vol 2 No 1 January 1986

' j i I ....... I I l l ~ -T-T~ I T o~ ,_~_ p,A,s, i. []

: . . . . . ] . . . . . . ' "-? ......... P A W m --V - - - ,

: o ' r 0 0 ~ . . . . N ~A ~A -o m m H g

Fig I. Negative inspiratory artifact (A), in a patient with labored breathing from pneumonia, sepsis, and pulmonary edema. The catheter balloon was deflated in the tracing on the right, and in- flated in the tracing on the l@. PAW = pulmonary artery wedge; PAS = pulmonary artery systolic; PAD = pulmonary artery di- astolic.

" A . •

A A

, . ram H g .I , ' "

Fig 2. Positive respiratory artifact (A) due to controlled ventila- tion. In the l@ tracing the catheter balloon was iT~ated; in the right tracing, it was deflated.

Figure 3 shows positive and negative artifacts occur- ring together. The patient had stiff lungs and labored respiration and was receiving assisted ventilation. The wedge pressure tracing was distorted by a 24-mm nega- tive artifact and a 6-mm positive artifact. The mean pressure recorded by the digital equipment was an aver- age of two artifacts. Similarly, the systolic pulmonary artery pressure contained a 24-mm positive artifact, and the diastolic pressure contained a negative 22-mm ar- tifact. This mean pressure was also an average of two artifacts. Thus, for this patient, all o f the information provided by the digital equipment was incorrect.

Figure 4 shows catheter whip. The pulmonary artery tracings demonstrate a spike-like deformity best seen in diastole, which is attributable to catheter whip. This spike covers very little actual area in the trace, but would be detected by the digital monitoring equipment and labeled as diastolic pressure. The spike deformity is presumed to be caused by motion of the catheter during the cardiac cycle, and is dampened when the catheter is

relatively fixed in the wedge position. The sharply pointed systolic peak may represent catheter whip as well. The pulmonary artery wedge pressure tracing on the right shows less distortion.

Figure 5 shows combined catheter whip and inspira- tory artifact. The patient had prosthetic mitral and aortic valves. In the systolic and diastolic tracings from the pulmonary artery, one can see the prominent posi- tive and negative spikes that cover essentially no area on the tracing, but which caused marked negative and posi- tive artifacts. On the wedge tracing one can see a 14- mm deflection caused by the negative intrapleural pres- sure resulting from the patient's attempts to breathe.

Figure 6 shows automatic positive end-expiratory pressure (auto-PEEP). The patient had tachypnea, labored breathing and laryngeal edema. The pulmonary artery pressure tracing shows a typical positive and negative respiratory artifact (A), as well as the very nar- row negative spike that indicates catheter whip. Note that in this case, because the patient was breathing spon- taneously, the positive artifact was not caused by a res- pirator but rather by the patient's attempts to exhale against an obstructed larynx (auto-PEEP). This situa- tion produced a prolonged expiratory phase best seen in

Page 4: Common artifacts of pulmonary artery and pulmonary artery ... · tive artifact and a 6-mm positive artifact. The mean pressure recorded by the digital equipment was an aver- age of

Case Report: Schmitt and Brantigan: Pressure Measurement Artifacts 47

Fig 3. A recording from a patient with labored respiration, stiff lungs, and assisted ventilation, producing both positive and nega- tive artifacts. The catheter balloon is inflated in the left tracing and deflated in the right tracing.

~ ~ I I ~ t ' , ~ , ] ~ +80

mm Hg Fig 4. Negative artifact from catheter whip. The catheter balloon is inflated in the right tracing and deflated in the left.

t v l I : ' T : ~

! :

I 8°nsplr|tory

mmH|

Fig 5. Recording showing catheter whip and inspiratory artifact. The balloon is deflated in the left tracing and inflated in the right.

the wedge pressure tracing. We would prefer to read the wedge pressure (PAWP) at the end o f expiration, but the point where intrapleural pressure approximates at- mospheric pressure occurs during such a brief time that it is difficult to find that exact spot. The arrow indicates where we think it is located in this tracing.

Figure 7 shows V waves. This tracing was obtained from a patient in congestive heart failure. When the

catheter balloon was properly inflated, the pressure peaks never entirely disappeard. This tracing was taken from a patient's chart, and shows how an inexperienced observer, who did not recognize the presence of abnor- mal V waves, tried to obliterate them by overinflating the balloon. This maneuver is dangerous to the patient, first because it may lead to pulmonary artery rupture and second because the data obtained are not related to the patient's hemodynamic state. Note that the pulmo- nary artery wedge pressure recorded by the observer (40 mrn Hg) was higher than the pulmonary artery diastolic

Page 5: Common artifacts of pulmonary artery and pulmonary artery ... · tive artifact and a 6-mm positive artifact. The mean pressure recorded by the digital equipment was an aver- age of

48 Journal of Clinical Monitoring Vol 2 No 1 January i986

Fig 6. Positive and negative artifacts resulting from labored breath- ing, automatic positive end-expiratory pressure, tachypnea, and catheter whip. The balloon is deflated in the left tracing and in- flated in the right. The arrowhead indicates our best estimate of true pulmonary artery wedge pressure (PAWP). A indicates posi- tive and negative artifacts.

~" Y ' F - ~-Vr,wl, ~ "TCV~LTH~ ~ E ~ ; 1 7 1 : . i i : ' 1 ~ ' i . ~ ~ ~

\]2)! ~ ; : t ~ J ! ! [ ~ i , ! '. I I 1 , 1 1

. . . . . . . J~" t i 7 ; , v | t ' 3 7 ~ ; ' # ! : - • [

....... ~:~ ~ !ii$1!:/[',.v W a v e s i !

H i . . . . . . . . . . . . . . .

Fig 7. Tracings show V waves and results of a dangerous attempt to compensate for them by overinflation of the catheter balloon. The balloon is deflated in the upper left half of the tracing and inflated in the upper right half. In the lower tracing the balloon is deflated in the left third of the tracing, inflated during the middle third, and overinflated in the right third.

pressure. In interpreting left ventricular filling pressures such as this one, the observer should ignore the V wave and not try to obliterate it.

Figure 8 shows positive and negative respiratory ar- tifacts associated with V waves. In this example, the respiratory artifacts in the pulmonary artery wedge pressure tracing are the same as those seen in the previ- ous examples. The tracing has a large V wave that is seen in systole. When the balloon was inflated, the sys- tolic pressure variation was not dampened out, and the

digital electronics averaged the peak of the V wave with the negative respiratory artifact to obtain an artifactually low value for the pulmonary artery wedge pressure.

Figure 9 shows inspiratory artifact and an overinflated catheter balloon. In this example the typical appearance of the rapidly rising pulmonary artery wedge pressure tracing seen in the right side o f the diagram is a charac- teristic example of overinflation of the pulmonary artery balloon. The mechanism for this characteristic pattern is unclear; either the end of the catheter was occluded by being forced into the pulmonary artery wall or the bal- loon tip herniated over the end of the catheter. In either case, pressure in the flow device or in the balloon was measured, rather than pressure in the patient. On the left side of the tracing, however, there is a saw-toothed pattern that was produced by an inexperienced observer who knowingly attempted to overinflate the balloon

Page 6: Common artifacts of pulmonary artery and pulmonary artery ... · tive artifact and a 6-mm positive artifact. The mean pressure recorded by the digital equipment was an aver- age of

Case Report: Schmitt and Brantigan: Pressure Measurement Artifacts 49

I I I . L L I ~ . . . . . . . . . . . . : ; .. : ', n ', :, := --; :: : ~i i I+ t+ t - I t+ t l+b~ / ~ ~ _ ~ f l ~

l f l : a ~ ~ r ~ m + ~ ' + ' ' a : : : a i ~ '="' m m I'l~.~d~tM:gt~i~HiliHiiii-iiii giagl ~ + u - e , + m - ~ m : : ] : : ' ] m - 1 4 ~ ! i :~ i : - : ! ! ! i i i i ] i] ] i } i i i i l i ~ : q ~ ] ~ / k : : : : : : : : : : : : : : : : : : H + ~ m + ~

Fig 8. Positive and negative respiratory artifacts (A) associated with V waves. The catheter balloon is deflated in the left side of the tracing and inflated in the right side.

| t l i i I 8o] ...... i I - ] ] , : - {

. _ _ I ]. - . ; - . . 1 . . . . . . . . i _

] ] i B I L L ] t~

-1 - ' 7 - [ -! m m ] i t _ _ _ b

,! I T T T i , : i i ' I

I ' ' I

1 1 2 " i i ! ~ 1

* I , i } !

' I I [ I i ' I . i *

H. I .......... i ! ,

------ i" -4--

# Fig 9. Inspiratory artifacts (small arrows) and attempts to compen- sate for them by overinflation of the balloon (large arrowheads). The balloon catheter is overinflated briefly in the left tracing, then deflated slightly to obtain a true pulmonary artery wedge pressure tracing that was not recognized; thus, the balloon was again over- inflate& When the problem was recognized the balloon was deflated to an appropriate level.

enough to eliminate the unrecognized inspiratory ar- tifact. At the extreme right side of the diagram is the true pulmonary artery wedge pressure tracing, with the prominently displayed typical negative respiratory ar- tifact. We present this example so that this waveform may be instantly recognized and the balloon im- mediately deflated when this dangerous situation is rec- ognized.

Figure 10 shows hiccoughs and the pulmonary artery wedge pressure. This example demonstrates the charac- teristic distortion of the pulmonary artery wedge pres- sure tracing produced by hiccoughs. A negative respira- tory artifact can be seen as well. Although distortions produced in this way would seem self-explanatory, in this patient these artifacts produced erroneous digital data that was accepted as valid by the staff.

Figure 11 shows pulsus paradoxus. This example il- lustrates the moment-to-moment variation in the stroke volume of a patient in congestive heart failure. One must be arbitrary in selecting the point on the tracing that represents the pulmonary artery wedge pressure, and then be consistent in interpreting subsequent trac- ings. In this example, the point selected by the digital equipment is probably as consistent and accurate as that selected using analog data.

Figure 12: This example shows a chaotic respiratory pattern that occurred while the patient was receiving mechanical ventilation. Observation of the patient was required to correlate the various positive and negative pressure spikes with respiration. However, our final condusion was that the pulmonary artery systolic and diastolic data were uninterpretable. The wedge pressure tracing, on the other hand, was interpretable. The posi- tive spikes seen on the tracing are artifacts caused by the cycling of the respirator and the resulting positive in- trapleurat pressure. The negative spikes represent the patient's attempts at inspiration. When these attempts were not vigorous enough to initiate a respirator cycle, they were followed by a plateau rather than by a posi- tive artifact. Thus the true pulmonary artery wedge

Page 7: Common artifacts of pulmonary artery and pulmonary artery ... · tive artifact and a 6-mm positive artifact. The mean pressure recorded by the digital equipment was an aver- age of

50 Journal of Clinical Monitoring Vot 2 No I January 1986

: i I i ! i }~

: ill{_!. :, i I I

! ; ! i i ; i ] ;

-7 :---|-:-I ,, n ~ - m m H , ,

: ' i t - ,

Fig 10. Tracing shows hiccoughs (A) and pulmonary artery wedge pressure. The catheter balloon is inflated throughout the tracing.

,~lllfllt; ., llilil,ltilltl N I t I ' ~ .,.I lillltttttlttlllNiili ItlitL!l!~! mmlitliiillTtl tl l l l . l lt l l l i lhlllft~ lt~,tl~ltiitlliillillmt iittllll|flltl Ilk!lt|tlitll:l 4~t~o ! tt~ll!,lLIIm.,,,,,,, l i~ t t !H~Hi l ! i l l ~Itl~IHUN fllllllllllli | .... ~ii i~tl!l!!ltUllll!!t!HJtHlNtlilliilltii!Utttt

. . . . . . . ';"';'";'":: iT ,i fiI Itill ,,',}t!it ii, ilIF.'IHIttlii!tttltlII!IltUltN!,,,,,, .......... ,,,,,,m;immiztltd,' J. "~1til [ ~ .......................

' t I t i , , t i ' , " llhfl,li.ttit.ll,tllmlRhtNttl iIIIttllIitltlItlIIItt,I!HUlIIII . $ !~.giii~ImlmmllllH,tt~tfii]tllill!tltt ~ i ~ ~ ~ ~ ~ ' i " ~ i ~ lliltl!i-tlilitltttllltl,ll,tllt,ltill-t,t~liltr

Fig 1I. Tracing shows pulsus paradoxus. The lines indicate an approximation of pulmonary artery systolic, pulmonary artery di- astolic, and pulmonary artery wedge pressures. The catheter bal- loon is deflated in the left side of the tracing and inflated in the right.

pressure in this patient was approximately 6 mm Hg, represented by the plateau seen in the tracing. Continu- ous display of airway pressure would have made inter- pretation easier, as the plateau could have been iden- tified quickly as end expiration.

DISCUSSION

To understand the physiological characteristics of a dig- ital readout is to understand how the machine arrives at the numbers that it displays. The monitors in use at Denver Presbyterian Hospital are typical o f those used in many hospitals. They examine the output of the transducer for a discrete 4-second period o f time known as the "scanning interval" and then calculate and display values obtained f rom this time period. They update and display the data for each new scanning interval. In this equipment, the scanning interval is almost, but not ex-

actly, the time required for one screen width of analog pressure data. The machine then takes the lowest value observed and designates it as the "diastolic" value. It takes the highest value and designates it as the "systolic" value. The "mean" is an arithmetic average over time. If the lowest value seen is an artifact, then the value dis- played for the "diastolic" value will be an artifact. If the highest value is an artifact, then the "systolic" value will be an artifact. If both are artifacts, then the mean will be an arithmetic average including two artifacts.

More sophisticated computer algorithms have been devised to deal with positive and negative artifacts, but they do not deal with many o f the more complex com- binations o f artifacts shown here, nor do they call atten- tion to changes in artifacts. Hence, even state-of-the-art algorithms are not reliable when compared with analog waveform interpretation by an experienced observer. However, they are probably more accurate than obser- vations made by the usual monitor user.

The normal pulmonary artery wedge pressure tracing is not a straight line, but is made up o f A, C, and V waves, which can be seen in most patients if the moni- toring system has sufficient fidelity. For this reason,

Page 8: Common artifacts of pulmonary artery and pulmonary artery ... · tive artifact and a 6-mm positive artifact. The mean pressure recorded by the digital equipment was an aver- age of

Case Report: Schmitt and Brantigan: Pressure Measurement Artifacts 51

t ! ~ ' mm HI; si Balloon T

Up~ I40

Fig 12. Tracings from a patient receiving mechanical ventilation. Airway pressure measurement would have helped in this example. The catheter balloon is deflated in the top tracing and inflated in the bottom tracing.

therapy is usually based on the mean value. The refer- ence point for pressure values is atmospheric pressure. Any pressure that is present at the catheter tip, be it intrapleural or intravascular pressure, is measured by the transducer, compared to atmospheric pressure, and displayed on the screen. Large negative excursions in pleural pressure may be produced by a patient making a great effort to inspire. Large positive excursions may be produced by the respirator struggling against stiff lungs or by a patient exhaling against an obstructed larynx (auto-PEEP) [3]. (End expiration is considered the op- t imum time for measuring pulmonary artery pressures, because intrathoracic pressure is closest to atmospheric pressure at end expiration.) Other changes, such as pul- sus paradoxus, that cause actual variations in cardiac output are more complicated [4] and require individual attention. When large V waves are superimposed on such a pattern the situation is even more complex.

Other authors have documented the effects reported here [5-12]. Some authors have described one or more artifacts and discussed their physiological characteris- tics. Others have provided normative statistical infor- mation about how incorrect the monitoring informa- tion may be. Our approach is more patient-specific. We have documented the important patterns of artifact and have provided a practical and inexpensive approach to handle them.

Berryhill and colleagues [7], in an elegant discussion, documented the effects of pleural pressure fluctuations on pulmonary artery pressure readings. They recom- mended the use of a chart recorder and measurements at

end expiration. They also recommended continuous monitoring of airway pressure so that end expiration could be identified. In our experience, however, the end of expiration is usually easily identified without such monitoring.

Manipulations of the patient or the respirator have been proposed to eliminate intrapleural pressure ar- tifacts. Davison and associates [8] measured pulmonary artery wedge pressure at end expiration in ventilated patients and compared this value with that obtained when patients were disconnected from the ventilator. The values were the same. Although their conclusion that pulmonary artery wedge pressure should be mea- sured at end expiration, rather than when patients were disconnected from the respirator, was valid, their deter- mination that pulmonary artery wedge pressure was the same when patients were and were not receiving me- chanical ventilation simply indicated that their patients were not critically ill. Mechanical ventilation may pro- duce important pulmonary artery and pulmonary artery wedge pressure changes that are not artifactual. Thus the physiological characteristics of patients receiving mechanical ventilation should be defined while they are being ventilated. Many patients cannot tolerate even a brief interruption of ventilatory support.

Schuster and Seeman [9] proposed that patients be paralyzed briefly to obtain accurate pressure data. This is another intervention that might change the patient's physiological status. Elimination of muscle tone can be expected to decrease venous return and make the pul- monary artery wedge pressure lower, as it did in their study. Physicians must take care that attempts to re- move measurement artifacts do not in themselves pro- duce misinformation.

Our purpose is to provide examples of most of the common and some of the uncommon patterns of pul- monary artery pressure seen in critically ill patients. It is important that physicians caring for these patients rec- ognize from waveform patterns displayed on the moni- tor screen when these patterns will not be interpreted accurately by the digital display. Our preferred ap- proach when such a pattern is recognized is to obtain a calibrated analog waveform, either by a strip-chart re- corder or, if the monitors are capable of generating a calibrated graphic display of waveforms, by directly in- terpreting the waveform from the monitor. The cali- brated strip-chart recorder is preferable because it allows more leisurely study of the waveforms and provides a permanent record for retrospective review and possible reconsideration. A chart recorder is inexpensive and can be interfaced with most monitoring equipment. In addition, it can be moved from patient to patient as needed.

Page 9: Common artifacts of pulmonary artery and pulmonary artery ... · tive artifact and a 6-mm positive artifact. The mean pressure recorded by the digital equipment was an aver- age of

52 Journal of Clinical Monitoring Vol 2 No 1 January 1986

REFERENCES

1. Swan HJC, Ganz W, ForresterJ, et al: Cardiac catheteriza- tion with a flow directed, balloon tipped catheter. N Engl J Med 1970;283:447-451

2. Ellis DM: Interpretation of beat-to-beat blood pressure values in the presence ofventilatory changes. J Clin Monit 1985;1:65-70

3. Pepe PE, Marini JJ: Occult positive end-expiratory pres- sure in mechanically ventilated patients with airflow ob- struction. Am Rev Respir Dis 1982;126:166-170

4. McGregor M: Pulsus paradoxus. N Engl J Med 1979; 301:480-482

5. Rice DL, Awe RJ, Gaaseh WH, et al: Wedge pressure measurement in obstructive pulmonary disease. Chest 1974;66:628-632

6. Permutt S: Relation between pulmonary arterial pressure and pleural pressure during the acute asthmatic attack. Chest 1973;63:25S-28S

7. Berryhill RE, BenumotJL, Rauscher LA: Pulmonary vas- cular pressure reading at the end of exhalation. Anes- thesiology 1978;49:365-368

8. Davison D, Parker M, Harrison RA: The validity of determination of pulmonary wedge pressure during mechanical ventilation. Chest t978;73:352-355

9. Schuster DP, Seeman MD: Temporary muscle paralysis for accurate measurement of pulmonary artery occlusion pressure. Chest 1983;84:593-597

10. Maran AG: Variables in pulmonary capillary wedge pres- sure: Variation with intrathoracic pressure, graphic and digital recorders. Crit Care Med t980;8:102-105

11. Quinn KQ, Quebbeman EJ: Pulmonary artery pressure monitoring in the surgical intensive care unit. Arch Surg 1981 ;116:872-876

12. Cengiz M, Crapo RO, Gardner RM: The effect of ventila- tion on the accuracy of pulmonary artery and wedge pres- sure measurements. Crit Care Med 1983;11:502-507


Recommended