Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia. COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY. - PowerPoint PPT Presentation
Popular Tags:
22
Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan Children Hospital Srebrnjak Zagreb, Croatia Ruđer Bošković Institute, Zagreb, Croatia COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY
Transcript
Page 1: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

Đ. Milković, M. Ranogajec-Komor, S. Miljanić, Ž. Knežević and K. Krpan

Children Hospital Srebrnjak Zagreb, Croatia

Ruđer Bošković Institute, Zagreb, Croatia

COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN

DIAGNOSTIC RADIOLOGY

Page 2: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

Our wish is that all children are safe and protected in radiology department!

Page 3: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

INTRODUCTION

Pulmonary X-rays are essential in the diagnostics of lung diseases of children and youth.

Chest radiography represents the majority of radiological examinations.

The starting basis for radiation protection is the exact determination of doses.

Page 4: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

FEATURES:

very low doses at low and variable energies have to be measured

there exists a considerable variation in radiation doses delivered to patients (different X-ray equipment, different staff, etc.).

Page 5: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

AIM

to test a new Shimadzu X-ray unit used for thorax examination of children

to compare a thermoluminescence (TL) dosimetry system based on LiF: Mg,Cu,P with the radiophotoluminescent (RPL) glass dosimetry system (FGD-200).

Page 6: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

MATERIALS AND METHODSIrradiations:

137Cs gamma rays in air – for calibrationISOVOLT 420 X Ray Unit (40-300 kV, 1-20 mA) at the SSDL in air and on the water phantom (plastic bottle, Φ=11 cm, V=2.5 l) – energy dependenceEnergies: 33, 48, 65 keVDnom= 2 mGy (air kerma)

150 kV Shimadzu CH-200M unit in air Phantoms:

water phantom (plastic bottle) doll phantom

Page 7: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

BABY PHANTOM

Page 8: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

Voltage: 70 kVQuantity of charge: 1.6 mAsTime of irradiation: 5 msSize of the focus: 0.6 mmDistance: 150 cm

Page 9: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

BABY FIX

Page 10: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY
Page 11: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

Dosimetry systems

Page 12: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

RESULTS

Energy dependence in SSDL

Doses in diagnostic X-ray unit

Page 13: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

Energy dependence in SSDL

Relative dose in air: the mean values of doses measured (Dmeasured) “in air” relative to delivered doses specified as “air-kerma free-in-air” (Ka).On phantom: the mean values of the doses measured on the phantom relative to delivered doses specified as air kerma free-in-air

Page 14: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

The energy dependence of TL and RPL dosimeters in SSDL

0.00.20.40.60.81.01.21.4

0 20 40 60 80

Mean photon energy (keV)

Rel

ativ

e do

se

0

0.5

1

1.5

2

0 20 40 60 80

Mean photon energy (keV)R

elat

ive

dose

In air On phantom

: TLD : RPLD : TLD : RPLD ● : Calculated values of Hp(10)/Ka

Page 15: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

Mean value and standard deviation (SD) of doses measured on “phantoms “ in

diagnostic unitPhantom Doll (unknown plastic) Bottle (water)

Dosimeter RPL TL RPL TL

Place of dosimeter Back Sternum Back Sternum Entr. Exit Entr. Exit

Mean dose (mGy) 0.040 0.019 0.049 0.022 0.041 0.004 0.030 0.002

SD (mGy) 0.002 0.006 0.008 0.007 0.002 0.001 0.006 0.004

SD(%) 4.3 30.6 15.5 30.0 3.9 28.2 19.5 154.9

Dentrance/Dexit 2.5 2.3 11.3 13.0

Entr.: Entrance

Page 16: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

Doses in diagnostic X-ray unitOn the doll: agreement of the dose values of RPL and TL dosimeters in entrance and exit beams On the water phantom: difference between the mean values measured in the ingoing beam with the two dosimeters

Reason: ▪ different materials of phantoms ▪ different energy absorption characteristics of the two dosimeters (below 50 keV) on water phantom

Page 17: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

CONCLUSIONTLD (LiF:Mg, Cu, P )

(termoluminiscent dosimeter)

RPL(radiophotoluminiscent glass dosimeter)

Page 18: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

TLD (LiF: Mg, Cu, P ) (termoluminiscent dosimeter)

High sensitivityIn spite of its anomalous energy dependence nearly tissue- equivalenceAgreement with earlier results

Page 19: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

RPL (radiophotoluminiscent glass dosimeter)Higher sensitivityEnergy dependance “in air” is better than for LiF:Mg, Cu, P (33-65 keV mean energies)Energy dependance curve on the water phantom changes in opposite direction than the calculated Hp(10) valuesThe absolute difference from Hp(10) is not larger than for LiF:Mg, Cu, P

Page 20: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

CONCLUSION

The measured dose values in X-ray diagnostic unit are in accordance with the characteristics found in SSDL for both dosimeters.

The RPL system seems to be suitable for dosimetric measurements in X- ray diagnostics.

Page 21: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

ZAGREB

Page 22: COMPARISON OF DIFFERENT DOSIMETRY SYSTEMS FOR DOSE MEASUREMENTS IN DIAGNOSTIC RADIOLOGY

Thank you for your

attention!

The authors are grateful to Chiyoda Technol Corporation, Japan for the support of this work.