Considerations in Particle Sizing Part 2: Specifying a Particle Size Analyzer Introduction In Part 1 we stated that the aim is to provide a pathway through the decision-making process of choosing a particle sizing analyzer by means of asking and answering three general questions: 1. How do I classify the various techniques? 2. How do I set specifications (either quantitative or qualitative)? 3. Which technique(s) have the best chance of solving my problems? We started by classifying the different particle sizing techniques in four ways: (i) size range, (ii) degree of separation (i.e., fractionation), (iii) imaging vs. non-imaging methods and (iv) weighting: intensity, volume, surface and number. Information Content A fifth way to classify a particle sizer is by information content. This final major classification revolves around the amount of information required to solve a problem. There are two key questions to ask that determine which techniques are useful. 1. What do you want? Averages, widths, tables & graphs, etc. 2. How will you use it? Process control, QC or R&D applications If all that is needed is an average particle size, then a single-moment instrument is sufficient. For average length and width, an ensemble averaging instrument will suffice. However, the more information needed, the more resolution that is required. But caveat emptor regarding the “zero-to-infinity” trap set by over-hyped marketing claims made for many instruments.
Transcript
Page 1: Considerations in Particle Sizing · 2019-11-27 · Considerations in Particle Sizing Part 2: Specifying a Particle Size Analyzer Introduction In Part 1 we stated that the aim is

Considerations in Particle Sizing Part 2: Specifying a

Particle Size Analyzer

Introduction

In Part 1 we stated that the aim is to provide a pathway through the

decision-making process of choosing a particle sizing analyzer by

means of asking and answering three general questions:

1. How do I classify the various techniques?

2. How do I set specifications (either quantitative or qualitative)?

3. Which technique(s) have the best chance of solving

my problems?

We started by classifying the different particle sizing techniques in four ways:

(i) size range, (ii) degree of separation (i.e., fractionation), (iii) imaging vs. non-imaging methods and (iv) weighting: intensity, volume, surface and

number.

Information Content A fifth way to classify a particle sizer is by information content. This

final major classification revolves around the amount of information

required to solve a problem. There are two key questions to ask

that determine which techniques are useful.1. What do you want? Averages, widths, tables &

graphs, etc.

2. How will you use it? Process control, QC or R&D

applications

If all that is needed is an average particle size, then a

single-moment instrument is sufficient. For average length

and width, an ensemble averaging instrument will suffice. However, the more information needed, the more

resolution that is required. But caveat emptor regarding

the “zero-to-infinity” trap set by over-hyped marketing

claims made for many instruments.

Page 2: Considerations in Particle Sizing · 2019-11-27 · Considerations in Particle Sizing Part 2: Specifying a Particle Size Analyzer Introduction In Part 1 we stated that the aim is

distribution has several, closely-spaced features, a true high-resolution technique is an imperative.

SpecificationsSpecifications are of two types: quantitative and

qualitative.

QuantitativeSpecifications of this type comprise size range, throughput and definitions: accuracy, precision, reproducibility and resolution.

Size RangeThis was discussed in Part 1 in the section on

Classification of Techniques (Part 1 - Figure 1).

Answering the first question, “What do you want?”, may not be easy but often follows from the answer to the

second question. For example, in most process control environments, varying a single parameter is reasonable, but varying multiple parameters is difficult. In this case, opt for one piece of information, which might be 90%

of particles less than a stated size. For QC, an average and a measure of distribution width is often sufficient, though sometimes the second piece of information is

nothing more than a spec such as d90 < 2 µm. Generally, only in an R&D environment is it useful to consider asking for more information. Additional size distribution information, often hard to come by reliably, might be the skewness of a single, broad distribution. It could also be the size and relative amounts of several

peaks in a multi-modal distribution or the existence of a few particles at one extreme of a distribution. Where the

10,000 nm

Figure 1

PARTICLE SIZING TECHNIQUES AND THEIR COVERED RANGES

NFFF: Normal Field-Flow Fractionation

CHDF: Capillary Hydrodynamic Fractionation

SED: Gravitational Sedimentation

XDC: X-ray Disc Centrifugation

DCP: Disc Centrifuge Photosedimentometry

DLS: Dynamic Light Scattering

TEM/SEM: Transmission/Scanning Electron Microscopy

NTA: Nanoparticle Tracking Analysis

IG: Induced Grating

TOT: Time of Transition

TOF: Time of Flight

E&OZC: Electro- & Optical-Zone Counters

OM/IA: Optical Microscopy/Image Analysis

FD: Fraunhofer Diffraction

SLS: Static Light Scattering

S&HFFF: Steric & Hyperlayer Field-Flow Fractionation

AAS: Acoustic Attenuation Spectroscopy

TOF

TOT

Sieves

E&OZC

OM/IA

FD

SLS

S&HFFF

NFFF

CHDF

SED

XDC

DCP

DLS

TEM/SEM

NTA

IG

AAS

10 nm 100 nm 1000 nm

Lubrizol Life Science

Figure 1

Page 3: Considerations in Particle Sizing · 2019-11-27 · Considerations in Particle Sizing Part 2: Specifying a Particle Size Analyzer Introduction In Part 1 we stated that the aim is

ThroughputThe novice often mistakenly assumes that the

measurement duration is sufficient to characterize the typical time per sample. Sometimes the measurement duration is only a fraction of the actual time per cycle. Throughput is the total sum over all the following: sample preparation, analysis, data reduction/printout/ interpretation and cleanup.

Throughput is probably most important to a QC

laboratory where, often, large numbers of samples must be run in one day. Speed of analysis is sometimes a major consideration even for one measurement in process

control applications. Sample preparation may be as short as a few minutes or require overnight. Warm-up, calibration or instrument adjustment all add to the overall time. Generally, with most modern instruments, the actual measurement or analysis time can be short. Yet, for broad distributions, sieving and sedimentation techniques

(including field-flow fractionation) are relatively slow

compared to most forms of light scattering. Single particle counting (SPC) is fast for narrow distributions but can be slow for broad distributions. Data reduction and printout are fast given modern computers. The time to interpret the data depends on the analyst and what

criteria have been set. Cleanup time is often seriously underestimated.

Finally, it is wise to consider whether a fast measurement or analysis time is worth it if the sum of all the other times

is considerable. If the total throughput time is not much different a higher resolution but slower technique is a

better choice.

DefinitionsAccuracy is a measure of how close an experimental result is to the “true” value. For irregularly shaped

particles, techniques that cannot be calibrated, or any other set of conditions where a “true” value is unknown or not well defined, accuracy has no meaning. For spheres and other simple shapes, accuracy can be established by comparison between several techniques. Surprisingly, below one µm, absolute accuracy is typically no better than 3%.

Precision is a measure of the variation in repeated

measurements under the same conditions (instrument, sample, and operator). Accuracy (associated with

systematic error) and precision (associated with random error) are related: the results of many measurements may

group tightly together (high precision, low random error) but the mean of the group may be far from the true value

(low accuracy, high systematic error). However, if a

measurement is highly accurate, then repeated

measurements must be grouped around the true value. Still, accurate mean values may consist of either high or low precision. In such cases, precision limits accuracy. Precision limits resolution and reproducibility and is a

useful criterion by which to assess instruments even

when accuracy cannot be determined.

Resolution is a measure of the minimum detectable

differences between distinct features in a size

distribution. For broad, unimodal distributions, resolution is still an important concept. If the measured breadth of a distribution is meaningful, then the instrument that produces it should be able to separate narrow size peaks

closer than or equal to that breadth. Otherwise the

measured breadth is really an instrumental broadening

effect. Generally, SPC and fractionators produce high resolution size distributions and ensemble averaging

devices (light scattering and diffraction instruments) produce medium to low resolution size distributions. Resolution is a function of the signal-to-noise ratio of the instrument. Reporting more than this is like magnifying the noise; more numbers are obtained, but they are meaningless. The particle size of many APIs is typically above one µm and the size distribution is very broad and a common assertion is that resolution would seem to be

unimportant. However, if the fundamental resolution of an instrument is undetermined, then one cannot really know if the broad distribution is hiding practical and

possibly significant information.

Reproducibility is a measure of the variation between

different machines, operators, sample preparations, etc. It becomes most important when comparing the results

produced on two different machines of the same type. Such a situation is quite com- mon where multiple

particle sizers of one make and model have been

purchased for use in different laboratories and/or locations. It is surprising how often the resolution

(expressed as a range of values) exceeds the basic

precision for any one of the machines. In such cases, it is useful to have round-robin tests conducted on the

same sample and, under the same set of prescribed conditions, to isolate any machine-to-machine variations. A classic example is the big differences obtained on FD instruments with high angle light scattering detectors

from the same manufacturer because of evolving

Page 4: Considerations in Particle Sizing · 2019-11-27 · Considerations in Particle Sizing Part 2: Specifying a Particle Size Analyzer Introduction In Part 1 we stated that the aim is

purpose rather than a poor job on a wide variety of amples.

Life-Cycle Cost: The basic instrument cost is only one

factor to consider. The total price is best judged in terms of the life-cycle cost. This includes purchase price, operating cost, maintenance, and repair costs. Every instrument needs some type of maintenance. It may be as simple as cleaning air filters once a month; it may be as difficult as replacing mechanical parts or aligning an optical system. And every instrument will, sooner or later, require repairs. If labor is intensive, the life-cycle cost can be quite high. If special solvents or expensive

environmental costs are involved, the life-cycle cost may be high enough to consider alternate choices.

Of all these qualitative considerations, support is, perhaps, the most important. When choosing between vendors of similar equipment, the one with better support may tip the scale in its favor. Do not assume that the largest vendor, or the one with the fanciest brochure, will provide the best support. Today, many companies use representatives to sell and service instruments. Just as you would choose any professional service, asking for references and getting second opinions should be an

integral part of the purchase process.

Conclusion to Parts 1 and 2Narrow down the possibilities and then make a choice

Start with Figure 1 and find the overlap of your expected size range with the various techniques that purport to

measure that range. Identify techniques whose midrange covers your expected size range. Don’t know your size range? Get some preliminary measurements made but

pay attention to sampling and sample preparation. The biggest mistake at this point is to choose the apparent

zero-to-infinity devices.

Given the list, narrow it further by deciding if you need imaging (irregular particle shapes that correlate with end-product performance) or not, single particle

counting (absolute concentration) or not, and what degree of information you require.

Now carefully consider the quantitative and qualitative

specifications, giving the most weight to those aspects that pertain to your situation. While automated, high throughput instrumentation is convenient, if it sacrifices

software variations on how best to handle the necessary

light scattering (Mie) corrections.

Qualitative

In addition to quantitative specifications, there are

qualitative ones that are important considerations for the

purchase of any analytical instruments. These include the

following:

Support: Is training, service, and applications assistance available during the installation, warranty period, and

for as long as the instrument is still serviceable? An

instrument might be available at a lower price from a

supplier in another country but check that it comes

with the expected type and level of support. Ask for references to verify any claims that are made. Ask also about any continuing program of development to ward

against obsolescence.

Ease-of-Use: This is a very subjective concept. Will the

instrument be used by experts or by inexperienced

users? Although the goal of a “one button” device is

admirable, it is rarely achieved if for no other reason

than sampling and sample preparation are not

one-button amenable. If this concept is important then, initially, be sure to watch measurements being made – the entire process from sample preparation to clean-up.

Versatility: This is defined as the ability to measure

a wide variety of samples under a wide variety of

conditions. Does the instrument handle samples in air, liquids, or both? Does the instrument work with polar as well as nonpolar liquids? Does the instrument work with dilute samples or concentrates or both? Try to estimate a realistic range of sample types and the corresponding

size ranges intended to be measured. Experience has shown that it is usually better to choose dedicated

instruments that do a good job for their intended

Page 5: Considerations in Particle Sizing · 2019-11-27 · Considerations in Particle Sizing Part 2: Specifying a Particle Size Analyzer Introduction In Part 1 we stated that the aim is

The information contained herein is believed to be reliable, but no representations, guarantees or warranties of any kind are made as to its accuracy, suitability for particular applications or the results to be obtained. The information often is based on laboratory work with small-scale equipment and does not necessarily indicate end-product per-formance or reproducibility. Formulations presented may not have been tested for stability and should be used only as a suggested starting point. Because of the variations in methods, conditions and equipment used commercially in processing these materials, no warranties or guarantees are made as to the suitability of the products for the applications disclosed. Full-scale testing and end-product performance are the responsibility of the user. Lubrizol Advanced Materials, Inc., shall not be liable for and the customer assumes all risk and liability for any use or handling of any material beyond Lubrizol Advanced Materials, Inc.’s direct control. The SELLER MAKES NO WARRANTIES, EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. Nothing con-tained herein is to be considered as permission, recommendation nor as an inducement to practice any patented invention without permission of the patent owner. Lubrizol Advanced Materials, Inc., is a wholly owned subsidiary of The Lubrizol Corporation.

©2019 The Lubrizol Corporation, all rights reserved. All marks are the property of The Lubrizol Corporation. The Lubrizol Corporation is a Berkshire Hathaway company.

HEALTH_RX_TB7_PARTICLESIZINGP2NO1620 OCT 2019

9911 Brecksville Road Cleveland, OH 44141-3201 USA

the resolution you need to make good decisions, consider carefully.

Accuracy, precision, resolution and reproducibility are functions of the size range. Errors are always greatest at the extremes. A common mistake is to check an instrument in its midrange and then proceed to use it at one or other of the extremes. Be skeptical of claims if these refer only to the average size. The average of any distribution is least subject to variation. Even instruments with poor resolution and instrument-to-instrument reproducibility can yield results with 2% precision in the

average. Higher moments such as the measure of width, or skewness, are much more sensitive to uncertainties; so pay

particular attention to the variance in these statistics. If it is not clear from the manufacturer’s literature then ask for clarification

Finally, before purchasing ask the vendor for a list of users who have had the instrument for at least one year. Contact them and ask for their experience with maintenance and repairs.