Date post: | 07-Oct-2015 |
Category: |
Documents |
Upload: | ahmed-tawfig-gamal |
View: | 212 times |
Download: | 0 times |
Calculus-detection technologiesand their clinical application
GRIT MEISSNER & THOMAS KOCHER
Subgingival calculus surfaces are usually covered
with a layer of unmineralized and metabolically
active bacteria. The essential component of conven-
tional periodontal therapy is the effective removal of
these bacterial deposits from the root surface, along
with calculus deposits, in order to create a biologi-
cally compatible root surface (10, 45, 63).
While numerous clinical studies have documented
the beneficial effects of complete removal of sub-
gingival calculus on the resolution of inflammation
(11, 45, 63), others have found that gingival tissues
adjacent to root surfaces covered with small polished
calculus spots might have a tendency to heal that is
similar to tissues adjacent to thoroughly cleaned,
calculus-free root surfaces (26, 44). Nevertheless,
periodontal destruction is clearly related to the very
presence of calculus, which may extend the range of
damage associated with plaque microorganisms (36,
61, 64).
Calculus is a porous substance that can adsorb a
variety of toxic products and retain significant levels
of endotoxin, which itself can damage tissue (64).
These toxins are located on, not within, periodontally
diseased root surfaces (7, 22, 43). It was therefore
deduced that extensive removal of cementum is not
necessary, and root surfaces should be treated
carefully during periodontal therapy in order to
selectively remove subgingival calculus and biofilm
without removing the underlying cementum.
Subgingival root debridement currently comprises
the systematic treatment of all diseased root surfaces
using hand-sonic and or ultrasonic instruments,followed by tactile control with a periodontal probe,
explorer or curette, until the root surface feels
smooth and clean. However, traditional tactile per-
ception of the subgingival environment without vis-
ible access before and after treatment frequently
lacks sensitivity, specificity and reproducibility, and
thus may lead to the unwanted removal of cemen-
tum, residual calculus, or both (6, 25, 27, 47, 57).
Clinicians are often uncertain about the nature of a
subgingival root surface while performing periodon-
tal instrumentation. The correct evaluation of a
cleaned surface is key to enable thorough and sub-
stance-sparing debridement. To support the clini-
cians decision to either stop or continue therapy, the
past few years have witnessed the development of
several calculus-detection techniques based on dif-
ferent technologies. Current technologies for calculus
identification include detection-only systems (a
miniaturized endoscope, a device based on light
reflection and a laser that activates the tooth surface to
fluoresce) as well as combined calculus-detection and
calculus-removal systems [an ultrasonic oscillation-
based system that analyzes impulses reflected from
the tooth surface, and a system combining erbium-
doped yttrium aluminium garnet (Er:YAG) and diode
lasers] (Tables 1 and 2). The aim of this article was to
provide a critical review of these devices based on
currently available clinical and experimental data.
Detection-only systems
Fiberoptic endoscopy-based technology
The idea to modify a medical endoscope for peri-
odontal use has, to date, been realized in only one
device (Perioscopy; Perioscopy Inc., Oakland, CA,
USA), which was introduced in the year 2000. Peri-
oscopy is a minimally invasive miniature periodontal
endoscope which is inserted into the periodontal
pocket and permits visualization of the root surface
within the subgingival environment at magnifications
of 2448 (Fig. 1). The system consists of a 1 mm,10,000-pixel fiberoptic bundle surrounded by multiple
189
Periodontology 2000, Vol. 55, 2011, 189204
Printed in Singapore. All rights reserved
2011 John Wiley & Sons A/S
PERIODONTOLOGY 2000
illumination fibers, a light source, an irrigation sys-
tem and a liquid crystal display monitor. Clinicians
can observe the subgingival root surface, tooth
structure and residual calculus in real time. The
magnified images can be viewed on the monitor in
real time, and images and videos can be captured and
saved in computer files. The endoscope may help to
identify, locate and treat calculus spots during
instrumentation of residual calculus at the time of, or
after, scaling. To be proficient in the endoscopic
technique a training period of at least 8 h is necessary
to learn the procedure and practical experience is
required for up to 4 weeks subsequently (59, 60).
In the first clinical study, nonresponding peri-
odontal sites (n = 44; probing depth 58 mm) were
treated by subgingival root debridement with or
without use of the dental endoscope (5). No signifi-
cant changes regarding pocket depth reduction were
reported in either group, 1 and 3 months after
treatment, compared with baseline. Moreover, the
gingival crevicular fluid flow rate, prostaglandin E2and interleukin-1beta levels decreased without
showing significant differences between the groups.
Additionally, a rather long treatment time, of 45 min
per experimental site, was noted for the Perioscopy
procedure.
In a study evaluating the histologic response to the
removal of calculus and biofilm with the aid of the
dental endoscope (65), a total of 12 teeth from six
patients were extracted 6 months after endoscope-
aided scaling and root planing. Histological evidence
showed formation of a long junctional epithelium,
bone repair and no signs of chronic inflammation.
However, a control group that received scaling and
root planing alone was not included and therefore
the incremental effect attributable to the use of the
endoscope was not determined.
A randomized, controlled, clinical study evaluated
the percentage of residual calculus after tooth
extraction (20) in 100 single-rooted teeth of 15
patients. The teeth were treated by hand- and ultra-
sonic instruments until the root surface was found to
be clean, as assessed by either an explorer or the
periodontal endoscope. After extraction, a higher
percentage of residual calculus covering the root
surface was detected microscopically in the explorer
group than in the endoscope group (D = 2.1%). Thedifference was statistically significant only in deeper
pockets and in interproximal sites (pocket depth
> 6 mm; D = 2.9%) compared with buccal sites(pocket depth > 4 mm; D = 1.3%). A correlation wasfound between shorter treatment time and increasing
experience of the operator for treatment with the
endoscope, a finding confirmed by a companion
study (41). However, the treatment results of the
latter study showed some discrepancies. Out of 24
patients, a total of 70 molars were treated in vivo
either by scaling and root planing only or by scaling
and root planing plus dental endoscopy, followed by
extraction. Overall,1.2% less residual calculus cover-
ing the root surface was found in the endoscopy
group (12.3%) compared with the scaling and root
planing group (13,5%). No differences in residual
calculus were found in deep pockets, furcation areas
or on buccal lingual surfaces. Only interproximalpockets with a depth of < 6 mm had significantly less
residual calculus in the endoscope group compared
with the scaling and root planing group. Thus, at least
for multi-rooted teeth, the beneficial effect of the
endoscope-aided scaling and root planing remains
questionable.
Taken together, only one clinical study to date has
investigated the clinical effects after the application of
fiberoptic technology. No differences were found
regarding pocket depth reduction between scaling and
root planing alone and endoscope-aided scaling and
root planing. Histologic healing, which was assessed
on extracted teeth 6 months after endoscope-aided
scaling and root planing, was not compared with
scaling and root planing alone in a randomized clinical
study. Microscopic analysis of root surfaces after
endoscopy-aided scaling and root planing showed a
Table 1. Automated calculus-detection technologies
Treatment goal Technology Device name
Calculus detection only Fiberoptic endoscopy Perioscopy
Spectro-optical technology Detectar
Autofluorescence Diagnodent
Combined calculus detection and
removal
Ultrasound Perioscan
Laser and autofluorescence Keylaser3
190
Meissner & Kocher
Table 2. Studies reviewed in this article
Instrument Reference Design Sample size Method Results
Diagnodent (31) In vitro study 10 teeth, 271 sites Fluorescence was mea-
sured at five teeth and
reproducibility was
tested (at all five teeth)
Effect of root
debridement on
fluorescence was tested
A clean root surface was
indicated with a median
value of 6.2, in contrast to a
median value of 57.7 on
the root where calculus was
found
Not influenced by the fluid
High reproducibility
Fluorescence values after root
debridement were similar to
those for a clean root surface
Diagnodent (17) In vitro study A total of 30 teeth,
For each medium,
10 teeth were
included
Fluorescence was
measured in medium,
air, saline solution
and blood
Significant differences in
fluorescence between calculus
and cementum in all fluids
Air: cementum, 0.4; calculus,
54.1
Saline solution: cementum,
0.4; calculus, 60.7
Blood: cementum, 2.1;
calculus, 39.6
Diagnodent (16) In vitro study A total of 40 teeth;
20 teeth were
included for each
treatment
Hand instrumentation
with and without
Diagnodent. In total,
120 surfaces were
evaluated
Surface area of residual calculus
Multirooted teeth:
hand instrumentation:
0.5 0.48 107 lm2
Diagnodent: 0.27 0.43 107
lm2 (P = 0.02)Single-rooted teeth:
hand instrumentation:
0.19 0.37 107 lm2
Diagnodent:
0.11 0.26 107 lm2
(P = 0.19)
Keylaser 3 (30) In vitro study Twenty teeth
covered with
subgingival calculus
were treated with
an ERL
ERL (140 mJ per pulse,
10 Hz), with a chisel-
shaped glass-fiber tip
(0.4 1.65 mm); waterirrigation (1 ml min)Fluorescence threshold
level of 5 [U] was reduced
at intervals of 1 [U] for
every laser treatment
Threshold 5 [U]; the median
residual calculus was
11 (078)%
Threshold 1 [U]; the median
residual calculus was 0 (026)%
Laser-treated cementum thick-
ness [median, 80 (0250) mm]
Untreated opposite side [med-
ian, 90 (30250) mm] (P < 0.05)
Keylaser 3 (53) Randomized,
single-masked
study
Twelve patients,
each with six
periodont
ally diseased
single-rooted teeth
Three teeth per patient
were treated with an ERL
[ERL1, 100 mJ per pulse;
ERL2, 120 mJ per pulse;
ERL3, 140 mJ per pulse;
10 Hz; water irrigation;
chisel-shaped glass-fiber
tip (0.4 1.65 mm);transmission factor 0.85]
and three teeth per
patient were treated
with the Vector system
or hand instrumentation,
or were untreated
(control)
Histologically, ERL produced
homogeneous and smooth root
surfaces
Calculus was almost selectively
removed, no thermal damage,
no cementum loss, mean
treatment time needed with
the ERL was comparable to that
for hand instrumentation
Hand instrumentation resulted
in significantly higher values for
residual calculus and in more
root surface damage than laser
treatment
191
Calculus-detection technologies
Table 2. (Continued)
Instrument Reference Design Sample size Method Results
Keylaser 3 (55) Randomized
clinical study
Twenty-four peri-
odontally diseased
single-rooted teeth
ERL, water irrigation
[160 mJ per pulse and
chisel-shaped tip
(1.65 0.5 mm);calculated energy density
19.4 J cm2 per pulse;10 Hz] vs. hand
instrumentation
Histologically, calculus was
selectively removed No thermal
damage
Results obtained following
treatment with the ERL were
comparable to those obtained
by hand instruments
Keylaser 3 (56) Randomized,
controlled,
split-mouth
study
Twenty patients,
single-rooted teeth
[n = 407 for laser
treatment (ERL),
n = 383 for UI]
multirooted teeth
(n = 269 for laser
treatment, n = 247
for UI
ERL, water irrigation
[160 mJ per pulse; 10 Hz;
chisel-shaped tip
(1.65 0.5 mm); calcu-lated energy density
136 mJ per pulse; or
chisel-shaped tip
(1.1 0.5 mm); calculatedenergy density 114 mJ
per pulse]
Average treatment time in both
groups was 5 min for
single-rooted teeth and 9 min
for multirooted teeth
All clinical parameters
investigated showed improve-
ment in both groups, which was
significant between baseline
and 6 months post-treatment
Bleeding on probing:
ERL: baseline, 40%; 6 months,
17%
UI: baseline, 46%; 6 months,
15%
Clinical attachment level gain:
ERL: after 3 months, 1.48 0.73;
after 6 months, 1.11 0.59
UI: after 3 months, 1.53 0.67;
after 6 months, 1.11 0.46
There were no statistically
significant differences between
the groups
Keylaser 3 (62) Single masked,
randomized,
controlled,
split-mouth
design study
Twenty patients at
recall visit with at
least two residual
pocket depths of
> 5 mm in each jaw
Treatment either by ERL
[160 mJ per pulse;10 Hz;
water irrigation;
chisel-shaped tips
(0.5 1.1 mm)] or by apiezoelectric ultrasonic
scaler (UI) (Piezon Master
400; EMS, Nyon,
Switzerland)
Clinical and microbiologic
effects at 1 and 4 months
post-treatment were
evaluated
Baseline:
Mean pocket depth: ERL, 6 mm;
UI, 5.8 mm
After 1 month significant differ-
ences:
Mean pocket depth reduction:
ERL, 0.9 mm; UI, 0.5 mm
(P < 0.05)
Mean clinical attachment
level gain: ERL, 0.5 mm; UI,
0.06 mm (P < 0.01)
After 4 months no significant
differences:
Mean pocket depth reduction:
ERL, 1.1 mm; UI, 1.0 mm
Mean clinical attachment level
gain: ERL, 0.6 mm; UI, 0.4 mm
Both treatment modalities
resulted in reduction of subgin-
gival microflora, with no
differences between the groups
The patients preference waslaser instrumentation
192
Meissner & Kocher
Table 2. (Continued)
Instrument Reference Design Sample size Method Results
Keyaser 3 (13) Single-blinded,
randomized,
controlled,
specific quad-
rant design
study
Seventy-two
patients with
periodontal
disease
Treatment per quadrant:
hand instruments
(Gracey curettes
(Hu Friedy), feedback-
controlled ERL (160 mJ
per pulse;10 Hz; water
irrigation; chisel-shaped
tips of 0.5 1.65 and0.5 1.1 mm), sonic
scaler (SONICflexs system
LUX 2003 L; KaVo)
and a piezoelectric
ultrasonic scaler (Piezon
Master 400, EMS)
Bacterial samples were
investigated at baseline,
and at 3 and 6 months
post-treatment
All four treatment modalities
resulted in a significant
reduction of Porphyromonas
gingivalis, Prevotella intermedia,
Tannerella forsythia and
Treponema denticola after
3 months. Laser and sonic
instrumentation failed to
reduce Aggregatibacter
actinomycetemcomitans
significantly
The patients preferencewas UI
Perioscopy (5) Randomized
patient
matched-site
design study
Six patients on
maintenance
therapy, 44 sites
with pocket depth
58 mm
Group A: scaling and root
planing plus explorer
Group B: scaling and root
planing plus Perioscopy
Treatment until root
surface was considered
to be clean
Evaluation of plaque
index, bleeding on prob-
ing, clinical attachment
level after 3 months
Plaque index, bleeding on
probing, clinical attachment
level gain: no significant
differences after 3 months
between the groups
Pocket depth: decrease of
2 mm in both groups, nosignificant differences
Treatment duration unrealistic
for clinical use
Perioscopy (20) Randomized
clinical and
in vitro study
Fifteen patients, a
total of 100 sites,
Single-rooted teeth
Group A: scaling and
root planing plus
explorer
Group B: scaling and
root planing plus
Perioscopy
Treatment until root
surface was considered to
be clean
Tooth extraction
immediately after therapy
Microscopic evaluation of
residual calculus
2.1% more residual calculus in
the explorer group
Statistical significance only in
interproximal sites (pocket
depth > 6 mm; 2.9%)
Treatment duration: endoscope
group showed a significant
decrease of time with
increasing experience of the
operator
Perioscopy (41) Randomized
clinical and
in vitro study
Twenty-four
patients, a total of
70 molars
Group A: scaling and root
planing plus explorer
Group B: scaling and root
planing plus Perioscopy
Treatment until root
surface was considered
to be clean
Tooth extraction immedi-
ately after therapy
Microscopic evaluation
of residual calculus
1.2% more residual calculus in
the explorer group
Statistical significance only in
interproximal sites (pocket
depth < 6 mm; 2.6%)
No differences in residual
calculus in deep pockets, furca-
tion areas or on buccal lingualsurfaces
Treatment duration: endoscope
group showed a significant de-
crease of time with increasing
experience of the operator
193
Calculus-detection technologies
Table 2. (Continued)
Instrument Reference Design Sample size Method Results
Perioscopy (65) Clinical and
histological
study
Six patients, a
total of 12 teeth
Scaling and root planing
plus Perioscopy
Tooth extraction
6 months after therapy
Histologic evaluation
No control group
Histologically: formation
of a long junctional epithelium,
evidence of bone repair,
no signs of chronic
inflammation
DetecTar (23) Randomized,
single-masked
study
Eight patients, a
total of 44 teeth
(176 surfaces)
Teeth extracted
immediately after
treatment
Microscopic
evaluation
Group A: no treatment,
calculus detection by
DetecTar
Group B: scaling and
root planing + DetecTar
until teeth were
considered to be clean
Control of the detection:
results after extraction
Group A: n = 96 surfaces;
79.4% sensitivity and 95.1%
specificity
Group B: n = 80 surfaces
(n = 58 initially positive,
n = 22 initially negative)
DetecTar (24) Randomized,
controlled
clinical study
One-hundred
patients with
plaque-associated
gingivitis
Group A (n = 50):
supragingival
debridement + oral
hygiene instruction
and motivation
Group B (n = 50):
supragingival
debridement + oral
hygiene instruction and
motivation + Detectar
Detectar group:
Plaque index (baseline 57.5%,
after 4 weeks 27.1%)
Bleeding on probing
(baseline 19.1%, after
4 weeks 7.1%)
Control group:
Plaque index (baseline 60.5%,
after 4 weeks 41.9%)
Bleeding on probing (baseline
23.1%, after 4 weeks 14.5%)
DetecTar (32) In vitro ran-
domized study
Twenty extracted
periodontally
involved, calculus-
covered teeth
Teeth were scanned:
(a) with different
working tip
angulations of the
fibreoptic (0, 10, 45
or 90)(b) with different ambient
fluids (blood and saline
solution)
Results were compared
with clinical and
histological findings
Specificity:
100% in blood
95-100% for all
angulations in saline
solution
Sensitivity:
Nearly 100% for all
angulations in saline solution
In blood:
100% for 90 angulation89% for 45 angulation
70% for 10 to 0 angulation
Perioscan (39) In vitro study Ten teeth, 200
measurements
Detection results were
compared with visual
findings on calculus
and cementum
surfaces
Calculus and cementum were
distinguishable with a sensitivity
of 88% and a specificity of 76%
Perioscan (38) In vitro study Thirty-four teeth,
1363 measurements
Detection results were
compared with visual
findings, by moving the
instrument tip over the
calculus and cementum
surfaces
Calculus and cementum
were distinguishable with a
sensitivity of 76% and a
specificity of 86%
Perioscan (40) In vitro study Fifty extracted,
periodontally
involved, calculus-
covered teeth
Calculus was removed
stepwise, in order to
determine the discrimi-
native capability
The smallest, recognizable
residual deposits had an average
diameter of 219 mm, an area of
21,600 mm2 and a circumfer-
ence of 748 mm; Sensitivity was
73% and specicity 80%
194
Meissner & Kocher
small benefit only in interproximal sites, in particular
in single-rooted teeth with deep pockets, and in
multirooted teeth with relatively shallow pockets.
Spectro-optical technology
The spectro-optical approach to calculus detection
uses a light-emitting diode and fiberoptic technology,
and is currently used by only one device, the Detec-
Tar (Dentsply Professional, York, PA, USA) (Fig. 2).
The characteristic spectral signature of subgingival
calculus, which is caused by absorption, reflection
and diffraction when irradiated by red light, is sensed
by an optical fiber and converted into an electrical
signal that is analyzed by a computer-processed
algorithm. The DetecTar device comes as a portable
cordless handpiece with a curved periodontal probe
that has millimeter markings to measure pocket
depths. Without any tactile pressure, the subgingival
root surface can be scanned by the instrument. As
soon as calculus is detected, the operator receives the
information on calculus localization by audible and
luminous signals.
Only a few investigations have evaluated spectro-
optical technology as a diagnostic instrument in
periodontology. The ability to detect subgingival
calculus in vitro was tested in 20 freshly extracted
teeth affected by periodontitis, and the results were
compared with clinical and histological findings (32).
In addition, the influence of different working-tip
angulations (0, 10, 45 and 90) of the fiberoptic probeand of different ambient fluids (blood and saline
solution) were studied. The specificity was only
slightly influenced by the type of irrigation fluid,
being 100% in blood and 95-100% in saline solution
for all angulations. The sensitivity in saline solution
Table 2. (Continued)
Instrument Reference Design Sample size Method Results
Perioscan (37) In vivo
randomized,
clinical study
Sixty-three buccal
subgingival tooth
surfaces
Teeth were scanned
in situ
Detection results were
compared with visual
findings after extraction
Calculus and cementum
were distinguishable with a
sensitivity of 91% and a
specificity of 82%
The positive predictive
value was 0.59 and the
negative predictive value
was 0.97
ERL, Er:YAG laser; UI, ultrasonic instrumentation.
Fig. 1. Endoscopy-based technology. The Perioscopy
(Perioscopy Inc., Oakland, CA, USA) uses a minimally
invasive miniature periodontal endoscope, which is
inserted into the periodontal pocket, to detect calculus.
Fig. 2. Spectro-optical technology. The DetecTar (Dents-
ply Professional, York, PA, USA) uses a light-emitting
diode and fiberoptic technology to detect calculus.
195
Calculus-detection technologies
was nearly 100% for all angulations. In blood, the
sensitivity decreased with smaller tip angulations
(100% sensitivity with angulation 90, 89% sensitivitywith angulation 45 and 70% sensitivity with angu-lation 100). The combination of saline solution asthe ambient fluid and a working-tip angulation of 90 which, however, cannot be achieved in the
periodontal pocket resulted in the most accurate
measurements.
A recent clinical study sought to determine the
utility of the spectro-optical technology for subgingi-
val calculus removal (23). A total of 44 teeth (176
surfaces) were included in the study. In an untreated
control group, a total of 96 untreated surfaces were
scanned in vivo using the DetecTar. In the treatment
group, treatment was initiated upon obtaining posi-
tive signals from the spectro-optical device, and the
treatment was continued until no signal was elicited.
Clinical calculus findings were documented by visual
and microscopic examination after tooth extraction.
The control group showed a sensitivity of only 79.4%
and a specificity of 95.1%. Of 58 tooth surfaces that
initially showed calculus and which were conse-
quently treated until they tested negative for calculus,
10 (17%) remained partly covered with calculus,
whereas 48 (83%) were completely calculus-free.
Nevertheless, nine (41%) of the 22 surfaces that were
initially identified as calculus-free (and therefore
untreated) did, in fact, harbor calculus. However, the
number of false-negative readings may have been
caused by incomplete surface scanning as a result of
limited access of the instrument and problems with
guiding the instrument. No sensitivity or specificity
data for the treatment group were calculated from
the published results. Additionally, the study only
recorded the clinical presence or absence of subgin-
gival calculus deposits for each surface (without exact
localization on the respective surface), and a highly
heterogeneous group of surfaces, with pocket depths
ranging from 1 to 10 mm, was evaluated. Therefore,
false-negative results may have been caused by an
incomplete scanning process, technological limits of
the device, or a combination of both. These aspects
cannot be discriminated in vivo if the exact location of
the device during scanning is not definitively known.
Altogether, the utility of the spectro-optical
technology for calculus detection has not yet been
thoroughly investigated.
Autofluorescence-based technology
The ability of calculus to emit fluorescent light fol-
lowing irradiation with light of a certain wavelength
enables the detection of calculus, and several in vitro
studies have examined the autofluorescence of dental
root surfaces and calculus (8, 12, 18, 26, 30, 45). Oral
microorganisms and their metabolites (metal-free
porphyrins, metalloporphyrins and other chromato-
phores) are assumed to contain the fluorophores that
are emitted from dental calculus and from carious
lesions (14, 21, 29). Several distinct fluorescence
bands between 570 and 730 nm were identified on
calculus specimens, which could be elicited with light
of wavelength 400420 nm, but could not be found
on clean root surfaces (9). Another study found
characteristic autofluorescence emission peaks for
calculus and dentin caries at 700 and 720 nm,
respectively, which were elicited by light of wave-
lengths 635 and 655 nm, respectively (33). On
surfaces covered by bacterial cells or blood, the
autofluorescence intensity was reduced.
In order to differentiate calculus from the healthy
tooth surface, a fluorescence-ratio method based on
autofluorescence induced by a blue light-emitting
diode of 405 nm has been developed (48). Calculus
and healthy tooth surfaces exposed to light wave-
lengths of 487 and 628685 nm were used to create a
calculus parameter, R, which was selected to define a
relationship between the integrated intensities spe-
cific for calculus and for healthy teeth in the 628 to
685- and the 477 to 497-nm wavelength regions,
respectively. A cut-off threshold of R = 0.2 was able to
distinguish dental calculus from healthy teeth with
100% sensitivity and 100% specificity under various
experimental conditions in vitro.
A diagnostic instrument, based on different auto-
fluorescence intensities after stimulation with red
light, claims to distinguish healthy from carious tooth
substance (Diagnodent; KaVo, Biberach, Germany)
(Fig. 3). An indium gallium arsenide phosphate
(InGaAsP)-based red laser diode (< 1 mW) sends light
with a wavelength of 655 nm through an optical fiber
onto the root surface, which is then induced to
fluoresce. The emitted fluorescent light returning
from the tooth tissue is captured by surrounding
optical fibers and transmitted to an integrated photo
diode, which serves as the fluorescence detector.
Optical effects caused by reflected light and ambient
light are eliminated by a band-pass filter and mod-
ulation of the fluorescent light, respectively. The
device was primarily developed for caries diagnosis
and launched as a stand-alone device about 10 years
ago. Based on a multitude of clinical studies, it is
considered to be a reliable caries detector on occlusal
and smooth surfaces, showing high levels of sensi-
tivity (92.1%) and specificity (100%), a high level of
196
Meissner & Kocher
reproducibility (kappa value: in vitro, 0.9; in vivo,
0.9) and a good interexaminer and intra-examiner
agreement (21, 34, 35, 42, 46, 49, 58).
Later, the device was further refined to enable cal-
culus detection. The fluorescence intensities are
measured, transformed and shown on a digital dis-
play as relative calculus-detection values from 0-99.
According to themanufacturer, values of 40 indicatemineralized deposits, whereas values of between 5
and 40 indicate very small calcified plaque sites (not
further specified) or residual calculus following partial
cleaning, and values of 5 indicate a clean root sur-face. Values indicating calculus are indicated by a
beep with an increasing audiotone frequency as the
display value increases. The manufacturer thus pro-
vides a small-size device, which is claimed to be able
to detect both caries and calculus, and which can be
handled easily with no further training required.
The autofluorescence-based device for calculus
detection has been evaluated only in in vitro studies
so far, with any patient-derived clinical evidence
lacking. Surfaces of extracted periodontally involved
teeth, which were partly covered with calculus and
moistened with saline solution or blood, were scan-
ned using the device (17, 31). The fluorescence
signals detected were compared with visual and
histological findings. The presence of calculus was
significantly correlated with a higher intensity of
fluorescence (17, 31). A median value of 6.2 was
obtained for clean root surfaces and a median value
of 57.7 was obtained for calculus, which was not
influenced by the presence of fluid. Additionally, high
reproducibility for measurements after 6 and 24 h
could be shown (31). The second study found relative
fluorescence values in air (cementum, 0.4; calculus,
54.1), in saline solution (cementum, 0.4; calculus,
60.7) and in blood (cementum, 2.1; calculus, 39.6).
With a cut-off value of 5, sensitivity and specificity in
all three media were 100% (17). Another study sim-
ulated a clinical situation based on a mannequin
model and compared the effectiveness of root-
surface instrumentation when supported by the
application of two different diagnostic instruments
(the autofluorescence-based system vs. a conven-
tional explorer) (16). Forty extracted periodontally
involved teeth (120 surfaces for each diagnostic
group) were treated with conventional Gracey
curettes until this method indicated a clean root
surface. For multirooted teeth, calculus detection
using autofluorescence resulted in a significantly
smaller total area covered with residual calculus than
if diagnostics was based on a conventional explorer.
However, in single-rooted teeth, the two study groups
revealed a comparable amount of residual calculus.
In summary, when used in vitro, the autofluores-
cence-based system could differentiate between cal-
culus and cementum with great reproducibility. In a
preclinical situation, a superior effect of the system
compared with manual use of an explorer could be
shown only on molars. The diagnostic value of the
autofluorescence-based system needs to be assessed
in the clinical setting, and its effect on treatment
outcomes determined.
Combined detection treatmentdevices
Ultrasonic technology
Ultrasonic calculus-detection technology is based on
a conventional piezo-driven ultrasonic scaler and is
similar to the way that one might tap on the rim of a
glass with a spoon to identify cracks acoustically (28,
60). An insert at a conventional dental ultrasound
scaler receives short, weak impulses with a frequency
of about 50 Hz, which make the inserts distal tiposcillate at a frequency that is dependent upon the
surface characteristics. The oscillations are con-
ducted into the piezo-ceramic discs, which transform
the oscillations into voltage. The voltage level repre-
sents the intensity of the tip oscillation, while the
frequency stays the same. The overall signal, con-
sisting of both the impulse stimulus and the impulse
response, is evaluated using a computerized system,
thereby generating information about a given surface
characteristic.
Fig. 3. Autofluorescence-based technology. The Diagno-
dentTM Pen (KaVo, Biberach, Germany) is based on the
detection of different autofluorescence intensities after
stimulation with red light.
197
Calculus-detection technologies
The ultrasonic device currently available (Perio-
scan; Sirona, Bensheim, Germany) (Fig. 4) providesa detection mode to discriminate between calculus
deposits and clean roots, along with a treatment
mode that allows conventional ultrasonic treatment
at different power levels. When the ultrasonic tip
touches the tooth surface, the detection results are
indicated by a light signal integrated both in the
handpiece and in a display of the table unit (green
indicates cementum and blue indicates calculus).
When calculus is detected, an additional acoustic
signal sounds. The detection mode is only activated
when no scaling treatment is performed. The detec-
tion and treatment modes can be used successively
on the surface of the same tooth. If calculus deposits
are found, the root surface can be treated with a
higher power setting, whereas in the absence of cal-
culus (thus requiring the systematic removal only of
biofilm), instrumentation can be performed at a
lower power setting. A prototype of the ultrasonic
device evaluated the calculus-detection capability
under laboratory conditions both in static tests
(yielding a sensitivity of 75% and a specificity of
82%) and during movements of the probing tip
(yielding a sensitivity of 88% and a specificity of
76%) (38, 39). The detection limit was further eval-
uated by gradually removing calculus from 50 ex-
tracted teeth until the system stopped discriminating
calculus deposits. Diameter, circumference and area
of the smallest recognizable deposit, and of the no
longer recognizable deposit, were measured, and a
cut-off point was determined. It could be demon-
strated that calculus deposits with a diameter of
0.2 mm could still be recognized with a sensitivity of
73% and a specificity of 80% (40).
The only available study involving the clinical
application of this ultrasound tool tested the
accuracy by which calculus was detected (37).
In vivo calculus detection was determined on 63
subgingival surfaces and compared with visual
findings after tooth extraction. A prevalence of
calculus of 22.3% was found on the scanned sur-
faces, and calculus and cementum were discrimi-
nated with a sensitivity of 91% and a specificity of
82%. The positive and negative predictive values
were 0.59 and 0.97, respectively. The combined
application of the calculus-detection mode and
the ultrasonic removal of calculus remain to be
investigated.
To sum up, the combined detection-and-treatment
technology using ultrasound is a promising tool for
minimally invasive debridement (retaining cemen-
tum) and selective calculus removal, as shown by a
study employing an in vivo and ex vivo reconstruc-
tion technique. However, the long-term clinical out-
come has not yet been investigated.
Laser-based technology
The benefit of laser application in nonsurgical peri-
odontal therapy is still a matter of debate among
clinicians (4, 12, 51). Lately, out of a variety of other
types of lasers, the Er:YAG laser has been considered
to be the most promising for periodontal therapy (2,
3, 19). Its ability to ablate soft and hard tissue without
major thermal side effects qualifies the use of this
laser for periodontal therapy, and Er:YAG lasers at
different energy levels have been studied in various
in vitro and clinical trials. Er:YAG lasers are solid-
state lasers that emit pulsed infrared light with a
90% root10% calculus
Fig. 4. Ultrasound-based calculus-
detection technology: Perioscan
(Sirona Dental Systems GmbH,
Bensheim, Germany). The principle
includes a fuzzy-logic-based detec-
tion mode employing ultrasound
feedback analysis and adds a treat-
ment mode to the automated
calculus detection, which uses the
same tip.
198
Meissner & Kocher
wavelength of 2940 nm, which is strongly absorbed
by virtually all biological tissues containing water.
The effect of Er:YAG lasers is based on photoablation.
The light-induced tissue evaporation results in water
release and a concomitant cooling effect on the sur-
rounding tissue. However, when applied to dental
hard tissue, which contains a lower amount of water,
increased thermal effects can occur, and therefore
water irrigation is required (2).
The treatment effect of Er:YAG lasers (Keylaser 1 or
2; Kavo, Biberach, Germany) (Fig. 5) with regard to
calculus removal has been shown to be comparable
to conventional root debridement. No major thermal
damage was found if the laser was applied at lower
energy levels (radiation energy, 50160 mJ) and with
concomitant water irrigation (2, 15, 18, 19, 54). A
number of in vivo and in vitro studies have shown the
potential of Er:YAG lasers to create a biocompatible
root surface by removing the smear layer and lipo-
polysaccharides from the tooth surface, by promoting
the attachment of periodontal ligament fibroblasts
and by decreasing the bacterial load (1, 52, 66). By
contrast, studies have also reported increased tissue
removal, roughened surfaces and a lower yield of
calculus removal compared with hand instrumenta-
tion (3, 15, 18, 19). The effectiveness of calculus
removal seems to be dependent on the irradiation
energy level. However, the application of high energy
levels is also associated with increased and undesir-
able root-substance loss if applied to a healthy tooth
structure (2, 18, 19).
The only commercially available device (Keyla-
ser3TM; KaVo) combines detection and treatment in a
feedback-controlled manner for selective removal of
calculus. The integrated calculus-detection device is
based on a 655-nm InGaAs diode laser for autofluo-
rescence-based calculus detection (described above
as a stand-alone diagnostic tool), whereas a 2940-nm
Er:YAG laser is used for treatment. The Er:YAG laser is
only activated to emit light if a preselected autoflu-
orescence threshold value for the diagnostic laser on
a scale of 099 is exceeded. As soon as the value falls
below the threshold, the Er:YAG laser turns off. This
combination of a diagnostic and a therapeutic laser
was designed to optimize calculus removal while
minimizing the undesired side effects of the Er:YAG
laser.
The feedback-controlled Er:YAG laser was recently
evaluated in in vitro and clinical studies to determine
how different fluorescence-classification thresholds
would influence the extent of calculus and cement
removal. Twenty teeth partly covered with calculus
and irrigated with water were treated from coronal to
apical direction in contact irradiation mode with
pulsed infrared radiation [wavelength of 2.940 mm,
a chisel-shaped glass-fiber application tip (size
0.4 1.65 mm), 140 mJ per pulse, 10 Hz and calcu-lated energy density of 17.2 mJ cm2) (30). The fluo-rescence threshold varied between 5 (recommended
by the manufacturer as the lowest threshold value)
and 1 in order to potentially increase sensitivity. Not
Fig. 5. Laser-based combined detection and treatment
technology. The Keylaser 3 (KaVo, Biberach, Germany)
employs the same detection method depicted in Fig. 3, but
adds a treatment mode to it.
199
Calculus-detection technologies
surprisingly, the amount of residual calculus de-
pended on the laser fluorescence threshold levels. At
a threshold of 5, the median residual amount of cal-
culus related to the baseline amount of calculus was
11% (minimum, 0%; maximum, 78%), whereas at a
threshold of 1, it was reduced to 0% (minimum, 0%;
maximum, 26%). However, the laser-treated residual
cementum was signicantly thinner (median, 80 lm)than the untreated residual cementum (median, 90 lm;P < 0.05). Thus, by reducing the threshold level to 1, the
sensitivity was increased at the expense of a reduced
specificity, as indicated by the increase of undesired
substance loss.
A different study compared the clinical and histo-
logical effects of conventional hand instrumentation
with fluorescence-controlled Er:YAG laser irradiation
at different device settings (55). Twenty-four peri-
odontally involved single-rooted teeth were treated
in vivo and extracted after therapy. Laser treatment
consisted of fluorescence-controlled Er:YAG laser
irradiation under water irrigation (160 mJ per pulse,
chisel-shaped tip of 1.65 0.5 mm, calculated energydensity 19.4 J cm2 per pulse, 10 Hz). All mesial rootsurfaces were treated in vivo under local anesthesia
until they were considered to be clean. After extraction,
the distal root surfaces were treated in vitro for com-
parison. Hand-instrumented teeth were treated accord-
ingly. Clinically, the use of the Er:YAG laser in vivo
produced homogeneous and nearly smooth root surfaces
without visible traces of the tip. Histologically, calculus
had been selectively removed and no thermal damage
could be observed. The results were comparable to those
seen after the use of hand instruments. The treatments
with the Er:Yag laser and with the hand instruments
were found to be more effective in vitro than in vivo.
Laser treatment also resulted in the removal of an in-
creased amount of cementum in vitro compared with
in vivo, whereas for hand instrumentation the in vitro
and in vivo results were comparable The reason for less
substance removal in vivo was assumed to be caused by
the restaining of the pocket tissue with blood and sulcus
fluid, which may have influenced the autofluorescence of
the dental hard tissue in vivo. However, by contrast,
different media (including blood and saline solution) did
not influence the autofluorescence intensity in vitro (17).
Another clinical study compared the clinical
benefit of autofluorescence-controlled Er:YAG laser
radiation with that of a special ultrasonic device
with vertical vibrations of the working tip (Vec-
torTM; Durr, Bietigheim-Bissingen, Germany), and
with hand instrumentation (53). Seventy-two single-
rooted teeth that were scheduled for extraction from
12 patients were randomly treated by the laser (at one
of three energy levels: 100, 120 or 140 mJ per pulse,
10 Hz), the Vector ultrasound system, conventional
hand instruments, or remained untreated. Teeth
were instrumented in vivo under local anesthesia until
they were considered to be clean and then immediately
extracted for analysis. The ultrasound system left sig-
nificantly smaller areas of residual calculus than the two
other therapies, but needed a significantly longer
instrumentation time than the laser and the hand
instruments. However, treatment with the feedback-
controlled Er:YAG laser still resulted in significantly less
residual calculus and less root-surface alterations than
hand instrumentation.
A clinical study compared the microbiological
effects of the Er:YAG laser, hand instruments, sonic
scalers and ultrasonic scalers (13). The controlled,
randomized, single-blinded clinical trial included 72
periodontal patients who had at least one site per
quadrant with a pocket depth of > 4 mm, bleeding
on probing and bone loss of at least 33%. The four
quadrants per patient were randomly assigned to one
of the following four debridement modalities: hand
instruments, a feedback-controlled Er:YAG laser
(Keylaser3; 160 mJ per pulse, 10 Hz, water irrigation,
chisel-shaped tips of 0.5 1.65 and 0.5 1.1 mm), asonic scaler (SONICflexs system LUX 2003 L; KaVo)
or a piezoelectric ultrasonic scaler (Piezon Master
400; EMS, Nyon, Switzerland). Subgingival plaque
samples were obtained at baseline and at 3 and
6 months postoperatively. All four treatments re-
sulted in a significant reduction in the amounts of
Porphyromonas gingivalis, Prevotella intermedia,
Tannerella forsythia and Treponema denticola after
3 months. Laser and sonic instrumentation failed to
significantly reduce the amount of Aggregatibacter
actinomycetemcomitans. Six months post-treatment,
the amount of test bacteria had increased in all study
groups.
Another set of clinical trials compared the clinical
outcome of periodontal treatment by a feedback-
controlled Er:YAG laser or ultrasonic instrumenta-
tion (56). Single-rooted and multirooted teeth with
pocket depths of > 4 mm were randomly treated in
a split-mouth design either by a feedback-controlled
Er:YAG laser (160 mJ per pulse, 10 Hz, chisel-shaped
tip of 1.65 0.5 mm, calculated energy density136 mJ per pulse; or chisel-shaped tip of 1.1 0.5 mm, calculated energy density 114 mJ per pulse)
or by an ultrasonic device (Cavitron Select; Dents-
ply, Konstanz, Germany) (56). At baseline, and 3 and
6 months post-treatment, plaque index, bleeding on
probing, pocket depth, gingival recession and clini-
cal attachment level were measured at six sites per
200
Meissner & Kocher
tooth. Deep pockets showed a tendency to experi-
ence more gingival recession, to gain more clinical
attachment level and to retain more residual pocket
depth compared with moderately deep pockets.
Bleeding on probing and clinical attachment level
improved significantly in both treatment groups
after 6 months compared with baseline. However,
statistically significant differences could not be
observed between the two types of treatment, sug-
gesting that treatment with the Er:YAG laser was
comparable with, but probably not superior to,
ultrasonic instrumentation (56). This conclusion is
in agreement with a subsequent clinical study that
compared the microbiological and short-term clini-
cal effects after Er:YAG laser debridement vs. ultra-
sonic treatment (62). Twenty patients with at least
two pockets with a depth of > 5 mm in each jaw
were included in the study. The pockets were ran-
domized to receive either feedback-controlled
Er:YAG laser treatment (160 mJ per pulse, 10 Hz,
chisel-shaped tip of 1.1 0.5 mm, water irrigation)or piezoelectric ultrasonic treatment (Piezon Master
400; EMS). Clinical attachment level gain and pocket
depth reduction after 1 month were significantly
higher in the laser group (mean pocket depth
reduction, 0.9 mm; mean clinical attachment level
gain, 0.5 mm) than in the ultrasonic group [mean
pocket depth reduction, 0.5 mm (P < 0.05); mean
clinical attachment level gain, 0.06 mm (P < 0.01)],
whereas 4 months after retreatment, no significant
differences were detected between the two treat-
ment modalities (mean pocket depth reduction:
laser, 1.1 mm; ultrasonic, 1.0 mm; and mean clinical
attachment level gain: laser, 0.6 mm; ultrasonic,
0.4 mm). Both treatment modalities yielded a simi-
lar reduction of the subgingival microflora after
4 months.
In conclusion, clinical and histological studies have
shown that laser-based detection and treatment of
calculus can effectively remove subgingival calculus
and preserve root substance. However, the results
were comparable with hand and ultrasonic debride-
ment, and controlled long-term clinical studies are
lacking.
Summary
A number of different technologies have been
incorporated into dental devices for the purpose of
identifying and selectively removing dental calculus.
Some of these new approaches for calculus removal
show promising results under optimum in vitro
conditions. Histological and microscopic findings
after in vivo use point to the potential for some of
these technologies to support or replace conventional
subgingival scaling. Published studies evaluating
clinical parameters, however, exist only for the
ultrasound- and laser-based devices, which combine
calculus detection and treatment. Moreover, con-
trolled randomized clinical trials are lacking for all
currently commercially available dental devices that
are used to identify and selectively remove dental
calculus.
All studies starting out with teeth treated in vivo
and then investigated after extraction have the same
problem in common, namely that clinical parameters
such as pocket depth, gingival recession and clinical
attachment level are assumed to be associated with a
comparable prevalence of calculus. This might not
always be the case and therefore a bias of uncertain
magnitude is introduced, especially if different stud-
ies and methods are compared. Moreover, it is
questionable whether the claimed improvement in
calculus detection in fact has resulted in selective
calculus removal and a concomitant preservation of
cementum. Without histologic examination, it is
impossible to decide whether cementum has actually
also been removed (50). In the case of the laser-based
detection and treatment device, for instance, histo-
logical analysis unveiled that the thorough removal of
calculus also resulted in an unwanted increase in the
amount of cementum removed.
A common problem of the stand-alone diagnostic
devices is that the application of these instruments
requires the systematic scanning of the entire sub-
gingival tooth surface, and, in the case of positive
calculus detection, the detected calculus has to be
located using the therapeutic scaling instrument.
Identifying the exact location of the calculus may be
difficult, thus potentially leading to over-treatment or
under-treatment. This problem relates to the skills of
the clinician rather than to features of the instru-
ment. The combined detection and treatment
instruments aim to overcome this problem.
The influence of operator skills on the outcome
variable has been shown previously and should
always be considered when evaluating the utility of a
particular method of scaling (8). Two different sce-
narios are conceivable: an experienced and trained
clinician will manage more easily the application of
advanced diagnostic procedures, such as the endos-
copy-based system, and thus obtain better results
than an inexperienced operator. Alternatively, a cli-
nician who is highly experienced in traditional scal-
ing methods may achieve less additional benefit by
201
Calculus-detection technologies
using supportive detection devices than a beginner or
a modestly skilled clinician, who may overcome a
lack of manual dexterity by using a supportive diag-
nostic system. These aspects have not been ad-
dressed in the published literature.
The fiberoptic detection technology shows poten-
tial to be a helpful tool in periodontal therapy, but
needs to be studied in clinical studies in direct
comparison with established scaling techniques. The
fiberoptic device currently available is somewhat
difficult to handle and requires additional time and
skills of the operator, especially when used simulta-
neously with scaling and root planing.
Data on the clinical utility of a spectro-optical
device for scaling and root planing are scarce.
Promising results were shown regarding the sensi-
tivity and specificity of calculus detection in vitro.
Whether or not a spectro-optical device is useful for
calculus detection needs to be evaluated in a clinical
setting. To our knowledge, a spectro-optical device is
not currently available for dental use.
To date, published data on autofluorescence-based
detection technology are only available from in vitro
and mannequin model studies. The autofluores-
cence-based system was found to be superior to
scaling and root planing alone only for multirooted
teeth, possibly because of their complicated root
configuration, which makes conventional diagnostics
more difficult. Calculus removal in single-rooted
teeth yielded similar results with and without the use
of the autofluorescence-based system. Its effective-
ness in clinical situations and its impact on clinical
parameters remains to be investigated.
It may be easy for clinicians to learn how to use
and apply the ultrasonic-driven combined detection
and treatment device because it is similar to the
familiar scaling technique. To provide reliable data
on the benefits of the device, clinical studies are
necessary to investigate changes in pocket depth,
clinical attachment level, bleeding on probing and
occurrence of hypersensitivity after treatment com-
pared with conventional methods.
The laser-based calculus-detection and treatment
technology has shown promising results with respect
to histology and certain clinical parameters in one
study, which, however, was limited to single-rooted
teeth. As the total number of cases in the published
literature is still small, additional studies are neces-
sary to evaluate the clinical benefit of this technology.
Taken together, despite promising laboratory
research results, the new technology-assisted
periodontal treatments have yet to show clinical
superiority in comparison with conventional scaling.
Clinical studies are necessary to assess if the use of
these devices can improve long-term treatment out-
come, with consequences of smaller residual probing
depth, a reduced need for periodontal surgery and
less hypersensitivity after treatment.
Acknowledgment
The work on Perioscan was supported by grants from
the Bundesministerium fur Bildung und Forschung
(BMBF 01 EZ 0025, BMBF 01 EZ 0026) and
from Sirona, Bensheim, Germany. T. Kocher and
G. Meissner have served as consultants to Sirona.
References
1. Ando Y, Aoki A, Watanabe H, Ishikawa I. Bactericidal effect
of erbium YAG laser on periodontopathic bacteria. Lasers
Surg Med 1996: 19: 190200.
2. Aoki A, Ando Y, Watanabe H, Ishikawa I. In vitro studies on
laser scaling of subgingival calculus with an erbium:YAG
laser. J Periodontol 1994: 65: 10971106.
3. Aoki A, Miura M, Akiyama F, Nakagawa N, Tanaka J, Oda S,
Watanabe H, Ishikawa I. In vitro evaluation of Er:YAG laser
scaling of subgingival calculus in comparison with ultra-
sonic scaling. J Periodontal Res 2000: 35: 266277.
4. Aoki A, Sasaki KM, Watanabe H, Ishikawa I. Lasers in
nonsurgical periodontal therapy. Periodontol 2000 2004:
36: 5997.
5. Avradopoulos V, Wilder RS, Chichester S, Offenbacher S.
Clinical and inflammatory evaluation of Perioscopy on
patients with chronic periodontitis. J Dent Hyg 2004: 78:
3038.
6. Biller IR, Kerber PE. Reliability of scaling error detection.
J Dent Educ 1980: 44: 206210.
7. Blomlof L, Lindskog S, Appelgren R, Jonsson B, Weintraub
A, Hammarstrom L. New attachment in monkeys with
experimental periodontitis with and without removal of
the cementum. J Clin Periodontol 1987: 14: 136143.
8. Brayer WK, Mellonig JT, Dunlap RM, Marinak KW, Carson
RE. Scaling and root planing effectiveness: the effect of root
surface access and operator experience. J Periodontol 1989:
60: 6772.
9. Buchalla W, Lennon AM, Attin T. Fluorescence spectros-
copy of dental calculus. J Periodontal Res 2004: 39: 327332.
10. Cobb CM. Non-surgical pocket therapy: mechanical. Ann
Periodontol 1996: 1: 443490.
11. Cobb CM. Clinical significance of non-surgical periodontal
therapy: an evidence-based perspective of scaling and root
planing. J Clin Periodontol 2002: 29(Suppl. 2): 616.
12. Cobb CM. Lasers in periodontics: a review of the literature.
J Periodontol 2006: 77: 545564.
13. Derdilopoulou FV, Nonhoff J, Neumann K, Kielbassa AM.
Microbiological findings after periodontal therapy using
curettes, Er:YAG laser, sonic, and ultrasonic scalers. J Clin
Periodontol 2007: 34: 588598.
202
Meissner & Kocher
14. Dolowy WC, Brandes ML, Gouterman M, Parker JD, Lind J.
Fluorescence of dental calculus from cats, dogs, and
humans and of bacteria cultured from dental calculus. J Vet
Dent 1995: 12: 105109.
15. Eberhard J, Ehlers H, Falk W, Acil Y, Albers HK, Jepsen S.
Efficacy of subgingival calculus removal with Er:YAG laser
compared to mechanical debridement: an in situ study.
J Clin Periodontol 2003: 30: 511518.
16. Folwaczny M, Heym R, Mehl A, Hickel R. The effectiveness
of InGaAsP diode laser radiation to detect subgingival
calculus as compared to an explorer. J Periodontol 2004: 75:
744749.
17. Folwaczny M, Heym R, Mehl A, Hickel R. Subgingival cal-
culus detection with fluorescence induced by 655 nm InG-
aAsP diode laser radiation. J Periodontol 2002: 73: 597601.
18. Folwaczny M, Mehl A, Haffner C, Benz C, Hickel R. Root
substance removal with Er:YAG laser radiation at different
parameters using a new delivery system. J Periodontol
2000: 71: 147155.
19. Frentzen M, Braun A, Aniol D. Er:YAG laser scaling of
diseased root surfaces. J Periodontol 2002: 73: 524530.
20. Geisinger ML, Mealey BL, Schoolfield J, Mellonig JT. The
effectiveness of subgingival scaling and root planing: an
evaluation of therapy with and without the use of the
periodontal endoscope. J Periodontol 2007: 78: 2228.
21. Hibst R, Paulus R, Lussi A. Detection of occlusal caries by
laser fluorescence: basic and clinical investigation. Med
Laser Appl 2001: 16: 205213.
22. Hughes FJ, Smales FC. Immunohistochemical investigation
of the presence and distribution of cementum-associated
lipopolysaccharides in periodontal disease. J Periodontal
Res 1986: 21: 660667.
23. Kasaj A, Moschos I, Rohrig B, Willershausen B. The effec-
tiveness of a novel optical probe in subgingival calculus
detection. Int J Dent Hyg 2008: 6: 143147.
24. Kasaj A, Sculean A, Willershausen B. Studie zur uberpru-
fung der wirksamkeit des detectar-systems hinsichtlich
einer verbesserung der compliance und der mundhygiene.
Parodontologie 2004: 15: 343353.
25. Kepic TJ, OLeary TJ, Kafrawy AH. Total calculus removal:an attainable objective? J Periodontol 1990: 61: 1620.
26. Kocher T, Konig J, Hansen P, Ruhling A. Subgingival pol-
ishing compared to scaling with steel curettes: a clinical
pilot study. J Clin Periodontol 2001: 28: 194199.
27. Kocher T, Ruhling A, Momsen H, Plagmann HC. Effec-
tiveness of subgingival instrumentation with power-driven
instruments in the hands of experienced and inexperi-
enced operators. A study on manikins. J Clin Periodontol
1997: 24: 498504.
28. Kocher T, Strackeljan J, Behr D. Feasibility of computer-
assisted recognition of different dental hard tissues. J Dent
Res 2000: 79: 829834.
29. Konig K, Flemming G, Hibst R. Laser-induced autofluo-
rescence spectroscopy of dental caries. Cell Mol Biol
(Noisy-le-grand) 1998: 44: 12931300.
30. Krause F, Braun A, Brede O, Eberhard J, Frentzen M, Jepsen
S. Evaluation of selective calculus removal by a fluores-
cence feedback-controlled Er:YAG laser in vitro. J Clin
Periodontol 2007: 34: 6671.
31. Krause F, Braun A, Frentzen M. The possibility of detecting
subgingival calculus by laser-fluorescence in vitro. Lasers
Med Sci 2003: 18: 3235.
32. Krause F, Braun A, Jepsen S, Frentzen M. Detection of
subgingival calculus with a novel LED-based optical probe.
J Periodontol 2005: 76: 12021206.
33. Kurihara E, Koseki T, Gohara K, Nishihara T, Ansai T,
Takehara T. Detection of subgingival calculus and dentine
caries by laser fluorescence. J Periodontal Res 2004: 39: 59
65.
34. Lussi A, Hibst R, Paulus R. DIAGNOdent: an optical method
for caries detection. J Dent Res 2004. 83 Spec No C: C8083.
35. Lussi A, Megert B, Longbottom C, Reich E, Francescut P.
Clinical performance of a laser fluorescence device for
detection of occlusal caries lesions. Eur J Oral Sci 2001:
109: 1419.
36. Mandel ID, Gaffar A. Calculus revisited. A review. J Clin
Periodontol 1986: 13: 249257.
37. Meissner G, Oehme B, Strackeljan J, Kocher T. Clinical
subgingival calculus detection with a smart ultrasonic de-
vice: a pilot study. J Clin Periodontol 2008: 35: 126132.
38. Meissner G, Oehme B, Strackeljan J, Kocher T. In vitro
calculus detection with a moved smart ultrasonic device.
J Clin Periodontol 2006: 33: 130134.
39. Meissner G, Oehme B, Strackeljan J, Kocher T. Influence of
handling-relevant factors on the behaviour of a novel
calculus-detection device. J Clin Periodontol 2005: 32: 323
328.
40. Meissner G, Oehme B, Strackeljan J, Kocher T. A new sys-
tem to detect residual subgingival calculus: in vitro detec-
tion limits. J Clin Periodontol 2006: 33: 195199.
41. Michaud RM, Schoolfield J, Mellonig JT, Mealey BL. The
efficacy of subgingival calculus removal with endoscopy-
aided scaling and root planing: a study on multirooted
teeth. J Periodontol 2007: 78: 22382245.
42. Nemes J, Csillag M, Toth Z, Fazekas A. Reproducibility of
the laser fluorescence method for the diagnosis of occlusal
caries. Clinical study. Fogorv Sz 2001: 94: 3336.
43. Nyman S, Sarhed G, Ericsson I, Gottlow J, Karring T. Role of
diseased root cementum in healing following treatment
of periodontal disease. An experimental study in the dog.
J Periodontal Res 1986: 21: 496503.
44. Nyman S, Westfelt E, Sarhed G, Karring T. Role of dis-
eased root cementum in healing following treatment of
periodontal disease. A clinical study. J Clin Periodontol
1988: 15: 464468.
45. Pihlstrom BL, Ammons WF. Treatment of gingivitis and
periodontitis. Research, Science and Therapy Committee
of the American Academy of Periodontology. J Periodontol
1997: 68: 12461253.
46. Pinelli C, Campos Serra M, de Castro Monteiro Loffredo L.
Validity and reproducibility of a laser fluorescence system
for detecting the activity of white-spot lesions on free
smooth surfaces in vivo. Caries Res 2002: 36: 1924.
47. Pippin DJ, Feil P. Interrater agreement on subgingival
calculus detection following scaling. J Dent Educ 1992: 56:
322326.
48. Qin YL, Luan XL, Bi LJ, Lu Z, Sheng YQ, Somesfalean G,
Zhou CN, Zhang ZG. Real-time detection of dental calculus
by blue-LED-induced fluorescence spectroscopy. J Photo-
chem Photobiol B 2007: 87: 8894.
49. Ross G. Caries diagnosis with the DIAGNOdent laser: a
users product evaluation. Ont Dent 1999: 76: 2124.50. Ruhling A, Wulf J, Schwahn C, Kocher T. Surface wear on
cervical restorations and adjacent enamel and root
203
Calculus-detection technologies
cementum caused by simulated long-term maintenance
therapy. J Clin Periodontol 2004: 31: 293298.
51. Schwarz F, Aoki A, Becker J, Sculean A. Laser application in
non-surgical periodontal therapy: a systematic review.
J Clin Periodontol 2008: 35: 2944.
52. Schwarz F, Aoki A, Sculean A, Georg T, Scherbaum W,
Becker J. In vivo effects of an Er:YAG laser, an ultrasonic
system and scaling and root planing on the biocompati-
bility of periodontally diseased root surfaces in cultures of
human PDL fibroblasts. Lasers Surg Med 2003: 33: 140
147.
53. Schwarz F, Bieling K, Venghaus S, Sculean A, Jepsen S,
Becker J. Influence of fluorescence-controlled Er:YAG laser
radiation, the Vector system and hand instruments on
periodontally diseased root surfaces in vivo. J Clin Period-
ontol 2006: 33: 200208.
54. Schwarz F, Sculean A, Berakdar M, Georg T, Reich E, Becker
J. Clinical evaluation of an Er:YAG laser combined with
scaling and root planing for non-surgical periodontal
treatment. A controlled, prospective clinical study. J Clin
Periodontol 2003: 30: 2634.
55. Schwarz F, Sculean A, Berakdar M, Szathmari L, Georg T,
Becker J. In vivo and in vitro effects of an Er:YAG laser, a
GaAlAs diode laser, and scaling and root planing on peri-
odontally diseased root surfaces: a comparative histologic
study. Lasers Surg Med 2003: 32: 359366.
56. Sculean A, Schwarz F, Berakdar M, Romanos GE, Arweiler
NB, Becker J. Periodontal treatment with an Er:YAG laser
compared to ultrasonic instrumentation: a pilot study.
J Periodontol 2004: 75: 966973.
57. Sherman PR, Hutchens LH Jr, Jewson LG, Moriarty JM,
Greco GW, McFall WT Jr. The effectiveness of subgingival
scaling and root planning. I. Clinical detection of residual
calculus. J Periodontol 1990: 61: 38.
58. Shi XQ, Welander U, Angmar-Mansson B. Occlusal caries
detection with KaVo DIAGNOdent and radiography: an in
vitro comparison. Caries Res 2000: 34: 151158.
59. Stambaugh RV, Myers G, Ebling W, Beckman B, Stambaugh
K. Endoscopic visualization of the submarginal gingiva
dental sulcus and tooth root surfaces. J Periodontol 2002:
73: 374382.
60. Strackeljan J, Behr D, Kocher T. Fuzzy-pattern recognition
for automatic detection of different teeth substances. Fuzzy
Sets and Systems 1997: 85: 275286.
61. Tan B, Gillam DG, Mordan NJ, Galgut PN. A preliminary
investigation into the ultrastructure of dental calculus and
associated bacteria. J Clin Periodontol 2004: 31: 364369.
62. Tomasi C, Schander K, Dahlen G, Wennstrom JL.
Short-term clinical and microbiologic effects of pocket
debridement with an Er:YAG laser during periodontal
maintenance. J Periodontol 2006: 77: 111118.
63. Van der Weijden GA, Timmerman MF. A systematic review
on the clinical efficacy of subgingival debridement in the
treatment of chronic periodontitis. J Clin Periodontol 2002:
29(Suppl 3): 5571.
64. White DJ. Dental calculus: recent insights into occurrence,
formation, prevention, removal and oral health effects of
supragingival and subgingival deposits. Eur J Oral Sci 1997:
105: 508522.
65. Wilson TG Jr, Carnio J, Schenk R, Myers G. Absence of
histologic signs of chronic inflammation following closed
subgingival scaling and root planing using the dental
endoscope: human biopsies - a pilot study. J Periodontol
2008: 79: 3641.
66. Yamaguchi H, Kobayashi K, Osada R, Sakuraba E, Nomura
T, Arai T, Nakamura J. Effects of irradiation of an erbiu-
m:YAG laser on root surfaces. J Periodontol 1997: 68: 1151
1155.
204
Meissner & Kocher
Copyright of Periodontology 2000 is the property of Wiley-Blackwell and its content may not be copied oremailed to multiple sites or posted to a listserv without the copyright holder's express written permission.However, users may print, download, or email articles for individual use.