+ All Categories
Home > Documents > ContentServer_4.pdf

ContentServer_4.pdf

Date post: 07-Oct-2015
Category:
Upload: ahmed-tawfig-gamal
View: 212 times
Download: 0 times
Share this document with a friend
17
 Calculus-detection technologies and their clinical application G R I T  M EISSNER  & T HOMAS  K OCHER Sub gingiv al calculus surf ace s are usually covered  with a layer of unmineralized and metabolically active bacteria. The essential component of conven- tional periodontal therapy is the effective removal of these bacterial deposits from the root surface, along  with calculus deposits, in order to create a biologi- cally compatible root surface (10, 45, 63).  While numerous clinical studies have documented the ben ecial eff ect s of comp let e removal of sub- gingival calculus on the resolution of inammation (11, 45, 63), others have found that gingival tissues adjacent to root surfaces covered with small polished calculus spots might have a tendency to heal that is similar to tis sue s adj acent to thoroughly cle ane d, calcul us-f ree roo t surfaces (26, 44). Nev ert hel ess, periodontal destruction is clearly related to the very presence of calculus, which may extend the range of damage associated with plaque microorganisms (36, 61, 64). Calculus is a porous substance that can adsorb a variety of toxic products and retain signicant levels of end otoxin, whic h its elf can damage tis sue (64) . These toxins are located on, not within, periodontally dis eas ed root sur fac es (7, 22, 43). It was the ref ore deduced that extensive removal of cementum is not necessar y, and root sur fac es sho ul d be tr eated care fully duri ng periodontal ther apy in order to selectively remove subgingival calculus and biolm  without removing the underlying cementum. Subgingival root debridement currently comprises the systematic treatment of all diseased root surfaces usi ng hand -sonic and ⁄   o r ult ras oni c instruments, followed by tactile control with a periodontal probe, ex plo re r or curette, until the root surface feels smooth and clean. However, traditional tactile per- ception of the subgingival environment without vis- ible access before and aft er treatment fre que ntl y lacks sensitivity, specicity and reproducibility, and thus may lead to the unwanted removal of cemen- tum, residual calculus, or both (6, 25, 27, 47, 57). Clinicians are often uncertain about the nature of a subgingival root surface while performing periodon- tal inst rumentation. The correct ev aluation of a cleaned surface is key to enable thorough and sub- stance- spar ing deb rid eme nt. To support the cli ni- cian‘s decision to either stop or continue therapy, the past few yea rs hav e wit nessed the dev elopme nt of sever al calculu s-det ection techni ques based on dif- ferent technologies. Current technologies for calculus ide nti cation include detection-onl y sys tems (a miniaturiz ed end osc ope , a dev ice bas ed on light reect ion and a lase r that act ivates the tooth sur faceto uoresce) as well as combin ed calculus-detection and calculus-removal systems [an ultrasonic oscillation- based system that analyzes impulses reected from the tooth surface, and a system combining erbium- doped yttrium aluminium garnet (Er:YAG) and diode lasers] (Tables 1 and 2). The aim of this article was to provide a critical review of these devices based on currently available clinical and experimental data. Detection-only systems Fiberoptic endoscopy-based technology The idea to modify a medical end osco pe for peri- odontal use has, to date, been realized in only one device (Per iosc opy ; Per ios copy Inc ., Oakland, CA, USA), which was introduced in the year 2000. Peri- oscopy is a minimally invasive miniature periodontal endoscope whi ch is inse rte d int o the per iodont al pocket and permits visualization of the root surface  within the subgingival environment at magnications of 24–48·  (Fig. 1). The system consists of a 1 mm, 10,000-pixel beroptic bundle surrounded by multiple 189 Periodontology 2000, Vol. 55, 2011, 189–204 Printed in Singapore. All rights reserved  2011 John Wiley & Sons A/S PERIODONTOLOGY 2000
Transcript
  • Calculus-detection technologiesand their clinical application

    GRIT MEISSNER & THOMAS KOCHER

    Subgingival calculus surfaces are usually covered

    with a layer of unmineralized and metabolically

    active bacteria. The essential component of conven-

    tional periodontal therapy is the effective removal of

    these bacterial deposits from the root surface, along

    with calculus deposits, in order to create a biologi-

    cally compatible root surface (10, 45, 63).

    While numerous clinical studies have documented

    the beneficial effects of complete removal of sub-

    gingival calculus on the resolution of inflammation

    (11, 45, 63), others have found that gingival tissues

    adjacent to root surfaces covered with small polished

    calculus spots might have a tendency to heal that is

    similar to tissues adjacent to thoroughly cleaned,

    calculus-free root surfaces (26, 44). Nevertheless,

    periodontal destruction is clearly related to the very

    presence of calculus, which may extend the range of

    damage associated with plaque microorganisms (36,

    61, 64).

    Calculus is a porous substance that can adsorb a

    variety of toxic products and retain significant levels

    of endotoxin, which itself can damage tissue (64).

    These toxins are located on, not within, periodontally

    diseased root surfaces (7, 22, 43). It was therefore

    deduced that extensive removal of cementum is not

    necessary, and root surfaces should be treated

    carefully during periodontal therapy in order to

    selectively remove subgingival calculus and biofilm

    without removing the underlying cementum.

    Subgingival root debridement currently comprises

    the systematic treatment of all diseased root surfaces

    using hand-sonic and or ultrasonic instruments,followed by tactile control with a periodontal probe,

    explorer or curette, until the root surface feels

    smooth and clean. However, traditional tactile per-

    ception of the subgingival environment without vis-

    ible access before and after treatment frequently

    lacks sensitivity, specificity and reproducibility, and

    thus may lead to the unwanted removal of cemen-

    tum, residual calculus, or both (6, 25, 27, 47, 57).

    Clinicians are often uncertain about the nature of a

    subgingival root surface while performing periodon-

    tal instrumentation. The correct evaluation of a

    cleaned surface is key to enable thorough and sub-

    stance-sparing debridement. To support the clini-

    cians decision to either stop or continue therapy, the

    past few years have witnessed the development of

    several calculus-detection techniques based on dif-

    ferent technologies. Current technologies for calculus

    identification include detection-only systems (a

    miniaturized endoscope, a device based on light

    reflection and a laser that activates the tooth surface to

    fluoresce) as well as combined calculus-detection and

    calculus-removal systems [an ultrasonic oscillation-

    based system that analyzes impulses reflected from

    the tooth surface, and a system combining erbium-

    doped yttrium aluminium garnet (Er:YAG) and diode

    lasers] (Tables 1 and 2). The aim of this article was to

    provide a critical review of these devices based on

    currently available clinical and experimental data.

    Detection-only systems

    Fiberoptic endoscopy-based technology

    The idea to modify a medical endoscope for peri-

    odontal use has, to date, been realized in only one

    device (Perioscopy; Perioscopy Inc., Oakland, CA,

    USA), which was introduced in the year 2000. Peri-

    oscopy is a minimally invasive miniature periodontal

    endoscope which is inserted into the periodontal

    pocket and permits visualization of the root surface

    within the subgingival environment at magnifications

    of 2448 (Fig. 1). The system consists of a 1 mm,10,000-pixel fiberoptic bundle surrounded by multiple

    189

    Periodontology 2000, Vol. 55, 2011, 189204

    Printed in Singapore. All rights reserved

    2011 John Wiley & Sons A/S

    PERIODONTOLOGY 2000

  • illumination fibers, a light source, an irrigation sys-

    tem and a liquid crystal display monitor. Clinicians

    can observe the subgingival root surface, tooth

    structure and residual calculus in real time. The

    magnified images can be viewed on the monitor in

    real time, and images and videos can be captured and

    saved in computer files. The endoscope may help to

    identify, locate and treat calculus spots during

    instrumentation of residual calculus at the time of, or

    after, scaling. To be proficient in the endoscopic

    technique a training period of at least 8 h is necessary

    to learn the procedure and practical experience is

    required for up to 4 weeks subsequently (59, 60).

    In the first clinical study, nonresponding peri-

    odontal sites (n = 44; probing depth 58 mm) were

    treated by subgingival root debridement with or

    without use of the dental endoscope (5). No signifi-

    cant changes regarding pocket depth reduction were

    reported in either group, 1 and 3 months after

    treatment, compared with baseline. Moreover, the

    gingival crevicular fluid flow rate, prostaglandin E2and interleukin-1beta levels decreased without

    showing significant differences between the groups.

    Additionally, a rather long treatment time, of 45 min

    per experimental site, was noted for the Perioscopy

    procedure.

    In a study evaluating the histologic response to the

    removal of calculus and biofilm with the aid of the

    dental endoscope (65), a total of 12 teeth from six

    patients were extracted 6 months after endoscope-

    aided scaling and root planing. Histological evidence

    showed formation of a long junctional epithelium,

    bone repair and no signs of chronic inflammation.

    However, a control group that received scaling and

    root planing alone was not included and therefore

    the incremental effect attributable to the use of the

    endoscope was not determined.

    A randomized, controlled, clinical study evaluated

    the percentage of residual calculus after tooth

    extraction (20) in 100 single-rooted teeth of 15

    patients. The teeth were treated by hand- and ultra-

    sonic instruments until the root surface was found to

    be clean, as assessed by either an explorer or the

    periodontal endoscope. After extraction, a higher

    percentage of residual calculus covering the root

    surface was detected microscopically in the explorer

    group than in the endoscope group (D = 2.1%). Thedifference was statistically significant only in deeper

    pockets and in interproximal sites (pocket depth

    > 6 mm; D = 2.9%) compared with buccal sites(pocket depth > 4 mm; D = 1.3%). A correlation wasfound between shorter treatment time and increasing

    experience of the operator for treatment with the

    endoscope, a finding confirmed by a companion

    study (41). However, the treatment results of the

    latter study showed some discrepancies. Out of 24

    patients, a total of 70 molars were treated in vivo

    either by scaling and root planing only or by scaling

    and root planing plus dental endoscopy, followed by

    extraction. Overall,1.2% less residual calculus cover-

    ing the root surface was found in the endoscopy

    group (12.3%) compared with the scaling and root

    planing group (13,5%). No differences in residual

    calculus were found in deep pockets, furcation areas

    or on buccal lingual surfaces. Only interproximalpockets with a depth of < 6 mm had significantly less

    residual calculus in the endoscope group compared

    with the scaling and root planing group. Thus, at least

    for multi-rooted teeth, the beneficial effect of the

    endoscope-aided scaling and root planing remains

    questionable.

    Taken together, only one clinical study to date has

    investigated the clinical effects after the application of

    fiberoptic technology. No differences were found

    regarding pocket depth reduction between scaling and

    root planing alone and endoscope-aided scaling and

    root planing. Histologic healing, which was assessed

    on extracted teeth 6 months after endoscope-aided

    scaling and root planing, was not compared with

    scaling and root planing alone in a randomized clinical

    study. Microscopic analysis of root surfaces after

    endoscopy-aided scaling and root planing showed a

    Table 1. Automated calculus-detection technologies

    Treatment goal Technology Device name

    Calculus detection only Fiberoptic endoscopy Perioscopy

    Spectro-optical technology Detectar

    Autofluorescence Diagnodent

    Combined calculus detection and

    removal

    Ultrasound Perioscan

    Laser and autofluorescence Keylaser3

    190

    Meissner & Kocher

  • Table 2. Studies reviewed in this article

    Instrument Reference Design Sample size Method Results

    Diagnodent (31) In vitro study 10 teeth, 271 sites Fluorescence was mea-

    sured at five teeth and

    reproducibility was

    tested (at all five teeth)

    Effect of root

    debridement on

    fluorescence was tested

    A clean root surface was

    indicated with a median

    value of 6.2, in contrast to a

    median value of 57.7 on

    the root where calculus was

    found

    Not influenced by the fluid

    High reproducibility

    Fluorescence values after root

    debridement were similar to

    those for a clean root surface

    Diagnodent (17) In vitro study A total of 30 teeth,

    For each medium,

    10 teeth were

    included

    Fluorescence was

    measured in medium,

    air, saline solution

    and blood

    Significant differences in

    fluorescence between calculus

    and cementum in all fluids

    Air: cementum, 0.4; calculus,

    54.1

    Saline solution: cementum,

    0.4; calculus, 60.7

    Blood: cementum, 2.1;

    calculus, 39.6

    Diagnodent (16) In vitro study A total of 40 teeth;

    20 teeth were

    included for each

    treatment

    Hand instrumentation

    with and without

    Diagnodent. In total,

    120 surfaces were

    evaluated

    Surface area of residual calculus

    Multirooted teeth:

    hand instrumentation:

    0.5 0.48 107 lm2

    Diagnodent: 0.27 0.43 107

    lm2 (P = 0.02)Single-rooted teeth:

    hand instrumentation:

    0.19 0.37 107 lm2

    Diagnodent:

    0.11 0.26 107 lm2

    (P = 0.19)

    Keylaser 3 (30) In vitro study Twenty teeth

    covered with

    subgingival calculus

    were treated with

    an ERL

    ERL (140 mJ per pulse,

    10 Hz), with a chisel-

    shaped glass-fiber tip

    (0.4 1.65 mm); waterirrigation (1 ml min)Fluorescence threshold

    level of 5 [U] was reduced

    at intervals of 1 [U] for

    every laser treatment

    Threshold 5 [U]; the median

    residual calculus was

    11 (078)%

    Threshold 1 [U]; the median

    residual calculus was 0 (026)%

    Laser-treated cementum thick-

    ness [median, 80 (0250) mm]

    Untreated opposite side [med-

    ian, 90 (30250) mm] (P < 0.05)

    Keylaser 3 (53) Randomized,

    single-masked

    study

    Twelve patients,

    each with six

    periodont

    ally diseased

    single-rooted teeth

    Three teeth per patient

    were treated with an ERL

    [ERL1, 100 mJ per pulse;

    ERL2, 120 mJ per pulse;

    ERL3, 140 mJ per pulse;

    10 Hz; water irrigation;

    chisel-shaped glass-fiber

    tip (0.4 1.65 mm);transmission factor 0.85]

    and three teeth per

    patient were treated

    with the Vector system

    or hand instrumentation,

    or were untreated

    (control)

    Histologically, ERL produced

    homogeneous and smooth root

    surfaces

    Calculus was almost selectively

    removed, no thermal damage,

    no cementum loss, mean

    treatment time needed with

    the ERL was comparable to that

    for hand instrumentation

    Hand instrumentation resulted

    in significantly higher values for

    residual calculus and in more

    root surface damage than laser

    treatment

    191

    Calculus-detection technologies

  • Table 2. (Continued)

    Instrument Reference Design Sample size Method Results

    Keylaser 3 (55) Randomized

    clinical study

    Twenty-four peri-

    odontally diseased

    single-rooted teeth

    ERL, water irrigation

    [160 mJ per pulse and

    chisel-shaped tip

    (1.65 0.5 mm);calculated energy density

    19.4 J cm2 per pulse;10 Hz] vs. hand

    instrumentation

    Histologically, calculus was

    selectively removed No thermal

    damage

    Results obtained following

    treatment with the ERL were

    comparable to those obtained

    by hand instruments

    Keylaser 3 (56) Randomized,

    controlled,

    split-mouth

    study

    Twenty patients,

    single-rooted teeth

    [n = 407 for laser

    treatment (ERL),

    n = 383 for UI]

    multirooted teeth

    (n = 269 for laser

    treatment, n = 247

    for UI

    ERL, water irrigation

    [160 mJ per pulse; 10 Hz;

    chisel-shaped tip

    (1.65 0.5 mm); calcu-lated energy density

    136 mJ per pulse; or

    chisel-shaped tip

    (1.1 0.5 mm); calculatedenergy density 114 mJ

    per pulse]

    Average treatment time in both

    groups was 5 min for

    single-rooted teeth and 9 min

    for multirooted teeth

    All clinical parameters

    investigated showed improve-

    ment in both groups, which was

    significant between baseline

    and 6 months post-treatment

    Bleeding on probing:

    ERL: baseline, 40%; 6 months,

    17%

    UI: baseline, 46%; 6 months,

    15%

    Clinical attachment level gain:

    ERL: after 3 months, 1.48 0.73;

    after 6 months, 1.11 0.59

    UI: after 3 months, 1.53 0.67;

    after 6 months, 1.11 0.46

    There were no statistically

    significant differences between

    the groups

    Keylaser 3 (62) Single masked,

    randomized,

    controlled,

    split-mouth

    design study

    Twenty patients at

    recall visit with at

    least two residual

    pocket depths of

    > 5 mm in each jaw

    Treatment either by ERL

    [160 mJ per pulse;10 Hz;

    water irrigation;

    chisel-shaped tips

    (0.5 1.1 mm)] or by apiezoelectric ultrasonic

    scaler (UI) (Piezon Master

    400; EMS, Nyon,

    Switzerland)

    Clinical and microbiologic

    effects at 1 and 4 months

    post-treatment were

    evaluated

    Baseline:

    Mean pocket depth: ERL, 6 mm;

    UI, 5.8 mm

    After 1 month significant differ-

    ences:

    Mean pocket depth reduction:

    ERL, 0.9 mm; UI, 0.5 mm

    (P < 0.05)

    Mean clinical attachment

    level gain: ERL, 0.5 mm; UI,

    0.06 mm (P < 0.01)

    After 4 months no significant

    differences:

    Mean pocket depth reduction:

    ERL, 1.1 mm; UI, 1.0 mm

    Mean clinical attachment level

    gain: ERL, 0.6 mm; UI, 0.4 mm

    Both treatment modalities

    resulted in reduction of subgin-

    gival microflora, with no

    differences between the groups

    The patients preference waslaser instrumentation

    192

    Meissner & Kocher

  • Table 2. (Continued)

    Instrument Reference Design Sample size Method Results

    Keyaser 3 (13) Single-blinded,

    randomized,

    controlled,

    specific quad-

    rant design

    study

    Seventy-two

    patients with

    periodontal

    disease

    Treatment per quadrant:

    hand instruments

    (Gracey curettes

    (Hu Friedy), feedback-

    controlled ERL (160 mJ

    per pulse;10 Hz; water

    irrigation; chisel-shaped

    tips of 0.5 1.65 and0.5 1.1 mm), sonic

    scaler (SONICflexs system

    LUX 2003 L; KaVo)

    and a piezoelectric

    ultrasonic scaler (Piezon

    Master 400, EMS)

    Bacterial samples were

    investigated at baseline,

    and at 3 and 6 months

    post-treatment

    All four treatment modalities

    resulted in a significant

    reduction of Porphyromonas

    gingivalis, Prevotella intermedia,

    Tannerella forsythia and

    Treponema denticola after

    3 months. Laser and sonic

    instrumentation failed to

    reduce Aggregatibacter

    actinomycetemcomitans

    significantly

    The patients preferencewas UI

    Perioscopy (5) Randomized

    patient

    matched-site

    design study

    Six patients on

    maintenance

    therapy, 44 sites

    with pocket depth

    58 mm

    Group A: scaling and root

    planing plus explorer

    Group B: scaling and root

    planing plus Perioscopy

    Treatment until root

    surface was considered

    to be clean

    Evaluation of plaque

    index, bleeding on prob-

    ing, clinical attachment

    level after 3 months

    Plaque index, bleeding on

    probing, clinical attachment

    level gain: no significant

    differences after 3 months

    between the groups

    Pocket depth: decrease of

    2 mm in both groups, nosignificant differences

    Treatment duration unrealistic

    for clinical use

    Perioscopy (20) Randomized

    clinical and

    in vitro study

    Fifteen patients, a

    total of 100 sites,

    Single-rooted teeth

    Group A: scaling and

    root planing plus

    explorer

    Group B: scaling and

    root planing plus

    Perioscopy

    Treatment until root

    surface was considered to

    be clean

    Tooth extraction

    immediately after therapy

    Microscopic evaluation of

    residual calculus

    2.1% more residual calculus in

    the explorer group

    Statistical significance only in

    interproximal sites (pocket

    depth > 6 mm; 2.9%)

    Treatment duration: endoscope

    group showed a significant

    decrease of time with

    increasing experience of the

    operator

    Perioscopy (41) Randomized

    clinical and

    in vitro study

    Twenty-four

    patients, a total of

    70 molars

    Group A: scaling and root

    planing plus explorer

    Group B: scaling and root

    planing plus Perioscopy

    Treatment until root

    surface was considered

    to be clean

    Tooth extraction immedi-

    ately after therapy

    Microscopic evaluation

    of residual calculus

    1.2% more residual calculus in

    the explorer group

    Statistical significance only in

    interproximal sites (pocket

    depth < 6 mm; 2.6%)

    No differences in residual

    calculus in deep pockets, furca-

    tion areas or on buccal lingualsurfaces

    Treatment duration: endoscope

    group showed a significant de-

    crease of time with increasing

    experience of the operator

    193

    Calculus-detection technologies

  • Table 2. (Continued)

    Instrument Reference Design Sample size Method Results

    Perioscopy (65) Clinical and

    histological

    study

    Six patients, a

    total of 12 teeth

    Scaling and root planing

    plus Perioscopy

    Tooth extraction

    6 months after therapy

    Histologic evaluation

    No control group

    Histologically: formation

    of a long junctional epithelium,

    evidence of bone repair,

    no signs of chronic

    inflammation

    DetecTar (23) Randomized,

    single-masked

    study

    Eight patients, a

    total of 44 teeth

    (176 surfaces)

    Teeth extracted

    immediately after

    treatment

    Microscopic

    evaluation

    Group A: no treatment,

    calculus detection by

    DetecTar

    Group B: scaling and

    root planing + DetecTar

    until teeth were

    considered to be clean

    Control of the detection:

    results after extraction

    Group A: n = 96 surfaces;

    79.4% sensitivity and 95.1%

    specificity

    Group B: n = 80 surfaces

    (n = 58 initially positive,

    n = 22 initially negative)

    DetecTar (24) Randomized,

    controlled

    clinical study

    One-hundred

    patients with

    plaque-associated

    gingivitis

    Group A (n = 50):

    supragingival

    debridement + oral

    hygiene instruction

    and motivation

    Group B (n = 50):

    supragingival

    debridement + oral

    hygiene instruction and

    motivation + Detectar

    Detectar group:

    Plaque index (baseline 57.5%,

    after 4 weeks 27.1%)

    Bleeding on probing

    (baseline 19.1%, after

    4 weeks 7.1%)

    Control group:

    Plaque index (baseline 60.5%,

    after 4 weeks 41.9%)

    Bleeding on probing (baseline

    23.1%, after 4 weeks 14.5%)

    DetecTar (32) In vitro ran-

    domized study

    Twenty extracted

    periodontally

    involved, calculus-

    covered teeth

    Teeth were scanned:

    (a) with different

    working tip

    angulations of the

    fibreoptic (0, 10, 45

    or 90)(b) with different ambient

    fluids (blood and saline

    solution)

    Results were compared

    with clinical and

    histological findings

    Specificity:

    100% in blood

    95-100% for all

    angulations in saline

    solution

    Sensitivity:

    Nearly 100% for all

    angulations in saline solution

    In blood:

    100% for 90 angulation89% for 45 angulation

    70% for 10 to 0 angulation

    Perioscan (39) In vitro study Ten teeth, 200

    measurements

    Detection results were

    compared with visual

    findings on calculus

    and cementum

    surfaces

    Calculus and cementum were

    distinguishable with a sensitivity

    of 88% and a specificity of 76%

    Perioscan (38) In vitro study Thirty-four teeth,

    1363 measurements

    Detection results were

    compared with visual

    findings, by moving the

    instrument tip over the

    calculus and cementum

    surfaces

    Calculus and cementum

    were distinguishable with a

    sensitivity of 76% and a

    specificity of 86%

    Perioscan (40) In vitro study Fifty extracted,

    periodontally

    involved, calculus-

    covered teeth

    Calculus was removed

    stepwise, in order to

    determine the discrimi-

    native capability

    The smallest, recognizable

    residual deposits had an average

    diameter of 219 mm, an area of

    21,600 mm2 and a circumfer-

    ence of 748 mm; Sensitivity was

    73% and specicity 80%

    194

    Meissner & Kocher

  • small benefit only in interproximal sites, in particular

    in single-rooted teeth with deep pockets, and in

    multirooted teeth with relatively shallow pockets.

    Spectro-optical technology

    The spectro-optical approach to calculus detection

    uses a light-emitting diode and fiberoptic technology,

    and is currently used by only one device, the Detec-

    Tar (Dentsply Professional, York, PA, USA) (Fig. 2).

    The characteristic spectral signature of subgingival

    calculus, which is caused by absorption, reflection

    and diffraction when irradiated by red light, is sensed

    by an optical fiber and converted into an electrical

    signal that is analyzed by a computer-processed

    algorithm. The DetecTar device comes as a portable

    cordless handpiece with a curved periodontal probe

    that has millimeter markings to measure pocket

    depths. Without any tactile pressure, the subgingival

    root surface can be scanned by the instrument. As

    soon as calculus is detected, the operator receives the

    information on calculus localization by audible and

    luminous signals.

    Only a few investigations have evaluated spectro-

    optical technology as a diagnostic instrument in

    periodontology. The ability to detect subgingival

    calculus in vitro was tested in 20 freshly extracted

    teeth affected by periodontitis, and the results were

    compared with clinical and histological findings (32).

    In addition, the influence of different working-tip

    angulations (0, 10, 45 and 90) of the fiberoptic probeand of different ambient fluids (blood and saline

    solution) were studied. The specificity was only

    slightly influenced by the type of irrigation fluid,

    being 100% in blood and 95-100% in saline solution

    for all angulations. The sensitivity in saline solution

    Table 2. (Continued)

    Instrument Reference Design Sample size Method Results

    Perioscan (37) In vivo

    randomized,

    clinical study

    Sixty-three buccal

    subgingival tooth

    surfaces

    Teeth were scanned

    in situ

    Detection results were

    compared with visual

    findings after extraction

    Calculus and cementum

    were distinguishable with a

    sensitivity of 91% and a

    specificity of 82%

    The positive predictive

    value was 0.59 and the

    negative predictive value

    was 0.97

    ERL, Er:YAG laser; UI, ultrasonic instrumentation.

    Fig. 1. Endoscopy-based technology. The Perioscopy

    (Perioscopy Inc., Oakland, CA, USA) uses a minimally

    invasive miniature periodontal endoscope, which is

    inserted into the periodontal pocket, to detect calculus.

    Fig. 2. Spectro-optical technology. The DetecTar (Dents-

    ply Professional, York, PA, USA) uses a light-emitting

    diode and fiberoptic technology to detect calculus.

    195

    Calculus-detection technologies

  • was nearly 100% for all angulations. In blood, the

    sensitivity decreased with smaller tip angulations

    (100% sensitivity with angulation 90, 89% sensitivitywith angulation 45 and 70% sensitivity with angu-lation 100). The combination of saline solution asthe ambient fluid and a working-tip angulation of 90 which, however, cannot be achieved in the

    periodontal pocket resulted in the most accurate

    measurements.

    A recent clinical study sought to determine the

    utility of the spectro-optical technology for subgingi-

    val calculus removal (23). A total of 44 teeth (176

    surfaces) were included in the study. In an untreated

    control group, a total of 96 untreated surfaces were

    scanned in vivo using the DetecTar. In the treatment

    group, treatment was initiated upon obtaining posi-

    tive signals from the spectro-optical device, and the

    treatment was continued until no signal was elicited.

    Clinical calculus findings were documented by visual

    and microscopic examination after tooth extraction.

    The control group showed a sensitivity of only 79.4%

    and a specificity of 95.1%. Of 58 tooth surfaces that

    initially showed calculus and which were conse-

    quently treated until they tested negative for calculus,

    10 (17%) remained partly covered with calculus,

    whereas 48 (83%) were completely calculus-free.

    Nevertheless, nine (41%) of the 22 surfaces that were

    initially identified as calculus-free (and therefore

    untreated) did, in fact, harbor calculus. However, the

    number of false-negative readings may have been

    caused by incomplete surface scanning as a result of

    limited access of the instrument and problems with

    guiding the instrument. No sensitivity or specificity

    data for the treatment group were calculated from

    the published results. Additionally, the study only

    recorded the clinical presence or absence of subgin-

    gival calculus deposits for each surface (without exact

    localization on the respective surface), and a highly

    heterogeneous group of surfaces, with pocket depths

    ranging from 1 to 10 mm, was evaluated. Therefore,

    false-negative results may have been caused by an

    incomplete scanning process, technological limits of

    the device, or a combination of both. These aspects

    cannot be discriminated in vivo if the exact location of

    the device during scanning is not definitively known.

    Altogether, the utility of the spectro-optical

    technology for calculus detection has not yet been

    thoroughly investigated.

    Autofluorescence-based technology

    The ability of calculus to emit fluorescent light fol-

    lowing irradiation with light of a certain wavelength

    enables the detection of calculus, and several in vitro

    studies have examined the autofluorescence of dental

    root surfaces and calculus (8, 12, 18, 26, 30, 45). Oral

    microorganisms and their metabolites (metal-free

    porphyrins, metalloporphyrins and other chromato-

    phores) are assumed to contain the fluorophores that

    are emitted from dental calculus and from carious

    lesions (14, 21, 29). Several distinct fluorescence

    bands between 570 and 730 nm were identified on

    calculus specimens, which could be elicited with light

    of wavelength 400420 nm, but could not be found

    on clean root surfaces (9). Another study found

    characteristic autofluorescence emission peaks for

    calculus and dentin caries at 700 and 720 nm,

    respectively, which were elicited by light of wave-

    lengths 635 and 655 nm, respectively (33). On

    surfaces covered by bacterial cells or blood, the

    autofluorescence intensity was reduced.

    In order to differentiate calculus from the healthy

    tooth surface, a fluorescence-ratio method based on

    autofluorescence induced by a blue light-emitting

    diode of 405 nm has been developed (48). Calculus

    and healthy tooth surfaces exposed to light wave-

    lengths of 487 and 628685 nm were used to create a

    calculus parameter, R, which was selected to define a

    relationship between the integrated intensities spe-

    cific for calculus and for healthy teeth in the 628 to

    685- and the 477 to 497-nm wavelength regions,

    respectively. A cut-off threshold of R = 0.2 was able to

    distinguish dental calculus from healthy teeth with

    100% sensitivity and 100% specificity under various

    experimental conditions in vitro.

    A diagnostic instrument, based on different auto-

    fluorescence intensities after stimulation with red

    light, claims to distinguish healthy from carious tooth

    substance (Diagnodent; KaVo, Biberach, Germany)

    (Fig. 3). An indium gallium arsenide phosphate

    (InGaAsP)-based red laser diode (< 1 mW) sends light

    with a wavelength of 655 nm through an optical fiber

    onto the root surface, which is then induced to

    fluoresce. The emitted fluorescent light returning

    from the tooth tissue is captured by surrounding

    optical fibers and transmitted to an integrated photo

    diode, which serves as the fluorescence detector.

    Optical effects caused by reflected light and ambient

    light are eliminated by a band-pass filter and mod-

    ulation of the fluorescent light, respectively. The

    device was primarily developed for caries diagnosis

    and launched as a stand-alone device about 10 years

    ago. Based on a multitude of clinical studies, it is

    considered to be a reliable caries detector on occlusal

    and smooth surfaces, showing high levels of sensi-

    tivity (92.1%) and specificity (100%), a high level of

    196

    Meissner & Kocher

  • reproducibility (kappa value: in vitro, 0.9; in vivo,

    0.9) and a good interexaminer and intra-examiner

    agreement (21, 34, 35, 42, 46, 49, 58).

    Later, the device was further refined to enable cal-

    culus detection. The fluorescence intensities are

    measured, transformed and shown on a digital dis-

    play as relative calculus-detection values from 0-99.

    According to themanufacturer, values of 40 indicatemineralized deposits, whereas values of between 5

    and 40 indicate very small calcified plaque sites (not

    further specified) or residual calculus following partial

    cleaning, and values of 5 indicate a clean root sur-face. Values indicating calculus are indicated by a

    beep with an increasing audiotone frequency as the

    display value increases. The manufacturer thus pro-

    vides a small-size device, which is claimed to be able

    to detect both caries and calculus, and which can be

    handled easily with no further training required.

    The autofluorescence-based device for calculus

    detection has been evaluated only in in vitro studies

    so far, with any patient-derived clinical evidence

    lacking. Surfaces of extracted periodontally involved

    teeth, which were partly covered with calculus and

    moistened with saline solution or blood, were scan-

    ned using the device (17, 31). The fluorescence

    signals detected were compared with visual and

    histological findings. The presence of calculus was

    significantly correlated with a higher intensity of

    fluorescence (17, 31). A median value of 6.2 was

    obtained for clean root surfaces and a median value

    of 57.7 was obtained for calculus, which was not

    influenced by the presence of fluid. Additionally, high

    reproducibility for measurements after 6 and 24 h

    could be shown (31). The second study found relative

    fluorescence values in air (cementum, 0.4; calculus,

    54.1), in saline solution (cementum, 0.4; calculus,

    60.7) and in blood (cementum, 2.1; calculus, 39.6).

    With a cut-off value of 5, sensitivity and specificity in

    all three media were 100% (17). Another study sim-

    ulated a clinical situation based on a mannequin

    model and compared the effectiveness of root-

    surface instrumentation when supported by the

    application of two different diagnostic instruments

    (the autofluorescence-based system vs. a conven-

    tional explorer) (16). Forty extracted periodontally

    involved teeth (120 surfaces for each diagnostic

    group) were treated with conventional Gracey

    curettes until this method indicated a clean root

    surface. For multirooted teeth, calculus detection

    using autofluorescence resulted in a significantly

    smaller total area covered with residual calculus than

    if diagnostics was based on a conventional explorer.

    However, in single-rooted teeth, the two study groups

    revealed a comparable amount of residual calculus.

    In summary, when used in vitro, the autofluores-

    cence-based system could differentiate between cal-

    culus and cementum with great reproducibility. In a

    preclinical situation, a superior effect of the system

    compared with manual use of an explorer could be

    shown only on molars. The diagnostic value of the

    autofluorescence-based system needs to be assessed

    in the clinical setting, and its effect on treatment

    outcomes determined.

    Combined detection treatmentdevices

    Ultrasonic technology

    Ultrasonic calculus-detection technology is based on

    a conventional piezo-driven ultrasonic scaler and is

    similar to the way that one might tap on the rim of a

    glass with a spoon to identify cracks acoustically (28,

    60). An insert at a conventional dental ultrasound

    scaler receives short, weak impulses with a frequency

    of about 50 Hz, which make the inserts distal tiposcillate at a frequency that is dependent upon the

    surface characteristics. The oscillations are con-

    ducted into the piezo-ceramic discs, which transform

    the oscillations into voltage. The voltage level repre-

    sents the intensity of the tip oscillation, while the

    frequency stays the same. The overall signal, con-

    sisting of both the impulse stimulus and the impulse

    response, is evaluated using a computerized system,

    thereby generating information about a given surface

    characteristic.

    Fig. 3. Autofluorescence-based technology. The Diagno-

    dentTM Pen (KaVo, Biberach, Germany) is based on the

    detection of different autofluorescence intensities after

    stimulation with red light.

    197

    Calculus-detection technologies

  • The ultrasonic device currently available (Perio-

    scan; Sirona, Bensheim, Germany) (Fig. 4) providesa detection mode to discriminate between calculus

    deposits and clean roots, along with a treatment

    mode that allows conventional ultrasonic treatment

    at different power levels. When the ultrasonic tip

    touches the tooth surface, the detection results are

    indicated by a light signal integrated both in the

    handpiece and in a display of the table unit (green

    indicates cementum and blue indicates calculus).

    When calculus is detected, an additional acoustic

    signal sounds. The detection mode is only activated

    when no scaling treatment is performed. The detec-

    tion and treatment modes can be used successively

    on the surface of the same tooth. If calculus deposits

    are found, the root surface can be treated with a

    higher power setting, whereas in the absence of cal-

    culus (thus requiring the systematic removal only of

    biofilm), instrumentation can be performed at a

    lower power setting. A prototype of the ultrasonic

    device evaluated the calculus-detection capability

    under laboratory conditions both in static tests

    (yielding a sensitivity of 75% and a specificity of

    82%) and during movements of the probing tip

    (yielding a sensitivity of 88% and a specificity of

    76%) (38, 39). The detection limit was further eval-

    uated by gradually removing calculus from 50 ex-

    tracted teeth until the system stopped discriminating

    calculus deposits. Diameter, circumference and area

    of the smallest recognizable deposit, and of the no

    longer recognizable deposit, were measured, and a

    cut-off point was determined. It could be demon-

    strated that calculus deposits with a diameter of

    0.2 mm could still be recognized with a sensitivity of

    73% and a specificity of 80% (40).

    The only available study involving the clinical

    application of this ultrasound tool tested the

    accuracy by which calculus was detected (37).

    In vivo calculus detection was determined on 63

    subgingival surfaces and compared with visual

    findings after tooth extraction. A prevalence of

    calculus of 22.3% was found on the scanned sur-

    faces, and calculus and cementum were discrimi-

    nated with a sensitivity of 91% and a specificity of

    82%. The positive and negative predictive values

    were 0.59 and 0.97, respectively. The combined

    application of the calculus-detection mode and

    the ultrasonic removal of calculus remain to be

    investigated.

    To sum up, the combined detection-and-treatment

    technology using ultrasound is a promising tool for

    minimally invasive debridement (retaining cemen-

    tum) and selective calculus removal, as shown by a

    study employing an in vivo and ex vivo reconstruc-

    tion technique. However, the long-term clinical out-

    come has not yet been investigated.

    Laser-based technology

    The benefit of laser application in nonsurgical peri-

    odontal therapy is still a matter of debate among

    clinicians (4, 12, 51). Lately, out of a variety of other

    types of lasers, the Er:YAG laser has been considered

    to be the most promising for periodontal therapy (2,

    3, 19). Its ability to ablate soft and hard tissue without

    major thermal side effects qualifies the use of this

    laser for periodontal therapy, and Er:YAG lasers at

    different energy levels have been studied in various

    in vitro and clinical trials. Er:YAG lasers are solid-

    state lasers that emit pulsed infrared light with a

    90% root10% calculus

    Fig. 4. Ultrasound-based calculus-

    detection technology: Perioscan

    (Sirona Dental Systems GmbH,

    Bensheim, Germany). The principle

    includes a fuzzy-logic-based detec-

    tion mode employing ultrasound

    feedback analysis and adds a treat-

    ment mode to the automated

    calculus detection, which uses the

    same tip.

    198

    Meissner & Kocher

  • wavelength of 2940 nm, which is strongly absorbed

    by virtually all biological tissues containing water.

    The effect of Er:YAG lasers is based on photoablation.

    The light-induced tissue evaporation results in water

    release and a concomitant cooling effect on the sur-

    rounding tissue. However, when applied to dental

    hard tissue, which contains a lower amount of water,

    increased thermal effects can occur, and therefore

    water irrigation is required (2).

    The treatment effect of Er:YAG lasers (Keylaser 1 or

    2; Kavo, Biberach, Germany) (Fig. 5) with regard to

    calculus removal has been shown to be comparable

    to conventional root debridement. No major thermal

    damage was found if the laser was applied at lower

    energy levels (radiation energy, 50160 mJ) and with

    concomitant water irrigation (2, 15, 18, 19, 54). A

    number of in vivo and in vitro studies have shown the

    potential of Er:YAG lasers to create a biocompatible

    root surface by removing the smear layer and lipo-

    polysaccharides from the tooth surface, by promoting

    the attachment of periodontal ligament fibroblasts

    and by decreasing the bacterial load (1, 52, 66). By

    contrast, studies have also reported increased tissue

    removal, roughened surfaces and a lower yield of

    calculus removal compared with hand instrumenta-

    tion (3, 15, 18, 19). The effectiveness of calculus

    removal seems to be dependent on the irradiation

    energy level. However, the application of high energy

    levels is also associated with increased and undesir-

    able root-substance loss if applied to a healthy tooth

    structure (2, 18, 19).

    The only commercially available device (Keyla-

    ser3TM; KaVo) combines detection and treatment in a

    feedback-controlled manner for selective removal of

    calculus. The integrated calculus-detection device is

    based on a 655-nm InGaAs diode laser for autofluo-

    rescence-based calculus detection (described above

    as a stand-alone diagnostic tool), whereas a 2940-nm

    Er:YAG laser is used for treatment. The Er:YAG laser is

    only activated to emit light if a preselected autoflu-

    orescence threshold value for the diagnostic laser on

    a scale of 099 is exceeded. As soon as the value falls

    below the threshold, the Er:YAG laser turns off. This

    combination of a diagnostic and a therapeutic laser

    was designed to optimize calculus removal while

    minimizing the undesired side effects of the Er:YAG

    laser.

    The feedback-controlled Er:YAG laser was recently

    evaluated in in vitro and clinical studies to determine

    how different fluorescence-classification thresholds

    would influence the extent of calculus and cement

    removal. Twenty teeth partly covered with calculus

    and irrigated with water were treated from coronal to

    apical direction in contact irradiation mode with

    pulsed infrared radiation [wavelength of 2.940 mm,

    a chisel-shaped glass-fiber application tip (size

    0.4 1.65 mm), 140 mJ per pulse, 10 Hz and calcu-lated energy density of 17.2 mJ cm2) (30). The fluo-rescence threshold varied between 5 (recommended

    by the manufacturer as the lowest threshold value)

    and 1 in order to potentially increase sensitivity. Not

    Fig. 5. Laser-based combined detection and treatment

    technology. The Keylaser 3 (KaVo, Biberach, Germany)

    employs the same detection method depicted in Fig. 3, but

    adds a treatment mode to it.

    199

    Calculus-detection technologies

  • surprisingly, the amount of residual calculus de-

    pended on the laser fluorescence threshold levels. At

    a threshold of 5, the median residual amount of cal-

    culus related to the baseline amount of calculus was

    11% (minimum, 0%; maximum, 78%), whereas at a

    threshold of 1, it was reduced to 0% (minimum, 0%;

    maximum, 26%). However, the laser-treated residual

    cementum was signicantly thinner (median, 80 lm)than the untreated residual cementum (median, 90 lm;P < 0.05). Thus, by reducing the threshold level to 1, the

    sensitivity was increased at the expense of a reduced

    specificity, as indicated by the increase of undesired

    substance loss.

    A different study compared the clinical and histo-

    logical effects of conventional hand instrumentation

    with fluorescence-controlled Er:YAG laser irradiation

    at different device settings (55). Twenty-four peri-

    odontally involved single-rooted teeth were treated

    in vivo and extracted after therapy. Laser treatment

    consisted of fluorescence-controlled Er:YAG laser

    irradiation under water irrigation (160 mJ per pulse,

    chisel-shaped tip of 1.65 0.5 mm, calculated energydensity 19.4 J cm2 per pulse, 10 Hz). All mesial rootsurfaces were treated in vivo under local anesthesia

    until they were considered to be clean. After extraction,

    the distal root surfaces were treated in vitro for com-

    parison. Hand-instrumented teeth were treated accord-

    ingly. Clinically, the use of the Er:YAG laser in vivo

    produced homogeneous and nearly smooth root surfaces

    without visible traces of the tip. Histologically, calculus

    had been selectively removed and no thermal damage

    could be observed. The results were comparable to those

    seen after the use of hand instruments. The treatments

    with the Er:Yag laser and with the hand instruments

    were found to be more effective in vitro than in vivo.

    Laser treatment also resulted in the removal of an in-

    creased amount of cementum in vitro compared with

    in vivo, whereas for hand instrumentation the in vitro

    and in vivo results were comparable The reason for less

    substance removal in vivo was assumed to be caused by

    the restaining of the pocket tissue with blood and sulcus

    fluid, which may have influenced the autofluorescence of

    the dental hard tissue in vivo. However, by contrast,

    different media (including blood and saline solution) did

    not influence the autofluorescence intensity in vitro (17).

    Another clinical study compared the clinical

    benefit of autofluorescence-controlled Er:YAG laser

    radiation with that of a special ultrasonic device

    with vertical vibrations of the working tip (Vec-

    torTM; Durr, Bietigheim-Bissingen, Germany), and

    with hand instrumentation (53). Seventy-two single-

    rooted teeth that were scheduled for extraction from

    12 patients were randomly treated by the laser (at one

    of three energy levels: 100, 120 or 140 mJ per pulse,

    10 Hz), the Vector ultrasound system, conventional

    hand instruments, or remained untreated. Teeth

    were instrumented in vivo under local anesthesia until

    they were considered to be clean and then immediately

    extracted for analysis. The ultrasound system left sig-

    nificantly smaller areas of residual calculus than the two

    other therapies, but needed a significantly longer

    instrumentation time than the laser and the hand

    instruments. However, treatment with the feedback-

    controlled Er:YAG laser still resulted in significantly less

    residual calculus and less root-surface alterations than

    hand instrumentation.

    A clinical study compared the microbiological

    effects of the Er:YAG laser, hand instruments, sonic

    scalers and ultrasonic scalers (13). The controlled,

    randomized, single-blinded clinical trial included 72

    periodontal patients who had at least one site per

    quadrant with a pocket depth of > 4 mm, bleeding

    on probing and bone loss of at least 33%. The four

    quadrants per patient were randomly assigned to one

    of the following four debridement modalities: hand

    instruments, a feedback-controlled Er:YAG laser

    (Keylaser3; 160 mJ per pulse, 10 Hz, water irrigation,

    chisel-shaped tips of 0.5 1.65 and 0.5 1.1 mm), asonic scaler (SONICflexs system LUX 2003 L; KaVo)

    or a piezoelectric ultrasonic scaler (Piezon Master

    400; EMS, Nyon, Switzerland). Subgingival plaque

    samples were obtained at baseline and at 3 and

    6 months postoperatively. All four treatments re-

    sulted in a significant reduction in the amounts of

    Porphyromonas gingivalis, Prevotella intermedia,

    Tannerella forsythia and Treponema denticola after

    3 months. Laser and sonic instrumentation failed to

    significantly reduce the amount of Aggregatibacter

    actinomycetemcomitans. Six months post-treatment,

    the amount of test bacteria had increased in all study

    groups.

    Another set of clinical trials compared the clinical

    outcome of periodontal treatment by a feedback-

    controlled Er:YAG laser or ultrasonic instrumenta-

    tion (56). Single-rooted and multirooted teeth with

    pocket depths of > 4 mm were randomly treated in

    a split-mouth design either by a feedback-controlled

    Er:YAG laser (160 mJ per pulse, 10 Hz, chisel-shaped

    tip of 1.65 0.5 mm, calculated energy density136 mJ per pulse; or chisel-shaped tip of 1.1 0.5 mm, calculated energy density 114 mJ per pulse)

    or by an ultrasonic device (Cavitron Select; Dents-

    ply, Konstanz, Germany) (56). At baseline, and 3 and

    6 months post-treatment, plaque index, bleeding on

    probing, pocket depth, gingival recession and clini-

    cal attachment level were measured at six sites per

    200

    Meissner & Kocher

  • tooth. Deep pockets showed a tendency to experi-

    ence more gingival recession, to gain more clinical

    attachment level and to retain more residual pocket

    depth compared with moderately deep pockets.

    Bleeding on probing and clinical attachment level

    improved significantly in both treatment groups

    after 6 months compared with baseline. However,

    statistically significant differences could not be

    observed between the two types of treatment, sug-

    gesting that treatment with the Er:YAG laser was

    comparable with, but probably not superior to,

    ultrasonic instrumentation (56). This conclusion is

    in agreement with a subsequent clinical study that

    compared the microbiological and short-term clini-

    cal effects after Er:YAG laser debridement vs. ultra-

    sonic treatment (62). Twenty patients with at least

    two pockets with a depth of > 5 mm in each jaw

    were included in the study. The pockets were ran-

    domized to receive either feedback-controlled

    Er:YAG laser treatment (160 mJ per pulse, 10 Hz,

    chisel-shaped tip of 1.1 0.5 mm, water irrigation)or piezoelectric ultrasonic treatment (Piezon Master

    400; EMS). Clinical attachment level gain and pocket

    depth reduction after 1 month were significantly

    higher in the laser group (mean pocket depth

    reduction, 0.9 mm; mean clinical attachment level

    gain, 0.5 mm) than in the ultrasonic group [mean

    pocket depth reduction, 0.5 mm (P < 0.05); mean

    clinical attachment level gain, 0.06 mm (P < 0.01)],

    whereas 4 months after retreatment, no significant

    differences were detected between the two treat-

    ment modalities (mean pocket depth reduction:

    laser, 1.1 mm; ultrasonic, 1.0 mm; and mean clinical

    attachment level gain: laser, 0.6 mm; ultrasonic,

    0.4 mm). Both treatment modalities yielded a simi-

    lar reduction of the subgingival microflora after

    4 months.

    In conclusion, clinical and histological studies have

    shown that laser-based detection and treatment of

    calculus can effectively remove subgingival calculus

    and preserve root substance. However, the results

    were comparable with hand and ultrasonic debride-

    ment, and controlled long-term clinical studies are

    lacking.

    Summary

    A number of different technologies have been

    incorporated into dental devices for the purpose of

    identifying and selectively removing dental calculus.

    Some of these new approaches for calculus removal

    show promising results under optimum in vitro

    conditions. Histological and microscopic findings

    after in vivo use point to the potential for some of

    these technologies to support or replace conventional

    subgingival scaling. Published studies evaluating

    clinical parameters, however, exist only for the

    ultrasound- and laser-based devices, which combine

    calculus detection and treatment. Moreover, con-

    trolled randomized clinical trials are lacking for all

    currently commercially available dental devices that

    are used to identify and selectively remove dental

    calculus.

    All studies starting out with teeth treated in vivo

    and then investigated after extraction have the same

    problem in common, namely that clinical parameters

    such as pocket depth, gingival recession and clinical

    attachment level are assumed to be associated with a

    comparable prevalence of calculus. This might not

    always be the case and therefore a bias of uncertain

    magnitude is introduced, especially if different stud-

    ies and methods are compared. Moreover, it is

    questionable whether the claimed improvement in

    calculus detection in fact has resulted in selective

    calculus removal and a concomitant preservation of

    cementum. Without histologic examination, it is

    impossible to decide whether cementum has actually

    also been removed (50). In the case of the laser-based

    detection and treatment device, for instance, histo-

    logical analysis unveiled that the thorough removal of

    calculus also resulted in an unwanted increase in the

    amount of cementum removed.

    A common problem of the stand-alone diagnostic

    devices is that the application of these instruments

    requires the systematic scanning of the entire sub-

    gingival tooth surface, and, in the case of positive

    calculus detection, the detected calculus has to be

    located using the therapeutic scaling instrument.

    Identifying the exact location of the calculus may be

    difficult, thus potentially leading to over-treatment or

    under-treatment. This problem relates to the skills of

    the clinician rather than to features of the instru-

    ment. The combined detection and treatment

    instruments aim to overcome this problem.

    The influence of operator skills on the outcome

    variable has been shown previously and should

    always be considered when evaluating the utility of a

    particular method of scaling (8). Two different sce-

    narios are conceivable: an experienced and trained

    clinician will manage more easily the application of

    advanced diagnostic procedures, such as the endos-

    copy-based system, and thus obtain better results

    than an inexperienced operator. Alternatively, a cli-

    nician who is highly experienced in traditional scal-

    ing methods may achieve less additional benefit by

    201

    Calculus-detection technologies

  • using supportive detection devices than a beginner or

    a modestly skilled clinician, who may overcome a

    lack of manual dexterity by using a supportive diag-

    nostic system. These aspects have not been ad-

    dressed in the published literature.

    The fiberoptic detection technology shows poten-

    tial to be a helpful tool in periodontal therapy, but

    needs to be studied in clinical studies in direct

    comparison with established scaling techniques. The

    fiberoptic device currently available is somewhat

    difficult to handle and requires additional time and

    skills of the operator, especially when used simulta-

    neously with scaling and root planing.

    Data on the clinical utility of a spectro-optical

    device for scaling and root planing are scarce.

    Promising results were shown regarding the sensi-

    tivity and specificity of calculus detection in vitro.

    Whether or not a spectro-optical device is useful for

    calculus detection needs to be evaluated in a clinical

    setting. To our knowledge, a spectro-optical device is

    not currently available for dental use.

    To date, published data on autofluorescence-based

    detection technology are only available from in vitro

    and mannequin model studies. The autofluores-

    cence-based system was found to be superior to

    scaling and root planing alone only for multirooted

    teeth, possibly because of their complicated root

    configuration, which makes conventional diagnostics

    more difficult. Calculus removal in single-rooted

    teeth yielded similar results with and without the use

    of the autofluorescence-based system. Its effective-

    ness in clinical situations and its impact on clinical

    parameters remains to be investigated.

    It may be easy for clinicians to learn how to use

    and apply the ultrasonic-driven combined detection

    and treatment device because it is similar to the

    familiar scaling technique. To provide reliable data

    on the benefits of the device, clinical studies are

    necessary to investigate changes in pocket depth,

    clinical attachment level, bleeding on probing and

    occurrence of hypersensitivity after treatment com-

    pared with conventional methods.

    The laser-based calculus-detection and treatment

    technology has shown promising results with respect

    to histology and certain clinical parameters in one

    study, which, however, was limited to single-rooted

    teeth. As the total number of cases in the published

    literature is still small, additional studies are neces-

    sary to evaluate the clinical benefit of this technology.

    Taken together, despite promising laboratory

    research results, the new technology-assisted

    periodontal treatments have yet to show clinical

    superiority in comparison with conventional scaling.

    Clinical studies are necessary to assess if the use of

    these devices can improve long-term treatment out-

    come, with consequences of smaller residual probing

    depth, a reduced need for periodontal surgery and

    less hypersensitivity after treatment.

    Acknowledgment

    The work on Perioscan was supported by grants from

    the Bundesministerium fur Bildung und Forschung

    (BMBF 01 EZ 0025, BMBF 01 EZ 0026) and

    from Sirona, Bensheim, Germany. T. Kocher and

    G. Meissner have served as consultants to Sirona.

    References

    1. Ando Y, Aoki A, Watanabe H, Ishikawa I. Bactericidal effect

    of erbium YAG laser on periodontopathic bacteria. Lasers

    Surg Med 1996: 19: 190200.

    2. Aoki A, Ando Y, Watanabe H, Ishikawa I. In vitro studies on

    laser scaling of subgingival calculus with an erbium:YAG

    laser. J Periodontol 1994: 65: 10971106.

    3. Aoki A, Miura M, Akiyama F, Nakagawa N, Tanaka J, Oda S,

    Watanabe H, Ishikawa I. In vitro evaluation of Er:YAG laser

    scaling of subgingival calculus in comparison with ultra-

    sonic scaling. J Periodontal Res 2000: 35: 266277.

    4. Aoki A, Sasaki KM, Watanabe H, Ishikawa I. Lasers in

    nonsurgical periodontal therapy. Periodontol 2000 2004:

    36: 5997.

    5. Avradopoulos V, Wilder RS, Chichester S, Offenbacher S.

    Clinical and inflammatory evaluation of Perioscopy on

    patients with chronic periodontitis. J Dent Hyg 2004: 78:

    3038.

    6. Biller IR, Kerber PE. Reliability of scaling error detection.

    J Dent Educ 1980: 44: 206210.

    7. Blomlof L, Lindskog S, Appelgren R, Jonsson B, Weintraub

    A, Hammarstrom L. New attachment in monkeys with

    experimental periodontitis with and without removal of

    the cementum. J Clin Periodontol 1987: 14: 136143.

    8. Brayer WK, Mellonig JT, Dunlap RM, Marinak KW, Carson

    RE. Scaling and root planing effectiveness: the effect of root

    surface access and operator experience. J Periodontol 1989:

    60: 6772.

    9. Buchalla W, Lennon AM, Attin T. Fluorescence spectros-

    copy of dental calculus. J Periodontal Res 2004: 39: 327332.

    10. Cobb CM. Non-surgical pocket therapy: mechanical. Ann

    Periodontol 1996: 1: 443490.

    11. Cobb CM. Clinical significance of non-surgical periodontal

    therapy: an evidence-based perspective of scaling and root

    planing. J Clin Periodontol 2002: 29(Suppl. 2): 616.

    12. Cobb CM. Lasers in periodontics: a review of the literature.

    J Periodontol 2006: 77: 545564.

    13. Derdilopoulou FV, Nonhoff J, Neumann K, Kielbassa AM.

    Microbiological findings after periodontal therapy using

    curettes, Er:YAG laser, sonic, and ultrasonic scalers. J Clin

    Periodontol 2007: 34: 588598.

    202

    Meissner & Kocher

  • 14. Dolowy WC, Brandes ML, Gouterman M, Parker JD, Lind J.

    Fluorescence of dental calculus from cats, dogs, and

    humans and of bacteria cultured from dental calculus. J Vet

    Dent 1995: 12: 105109.

    15. Eberhard J, Ehlers H, Falk W, Acil Y, Albers HK, Jepsen S.

    Efficacy of subgingival calculus removal with Er:YAG laser

    compared to mechanical debridement: an in situ study.

    J Clin Periodontol 2003: 30: 511518.

    16. Folwaczny M, Heym R, Mehl A, Hickel R. The effectiveness

    of InGaAsP diode laser radiation to detect subgingival

    calculus as compared to an explorer. J Periodontol 2004: 75:

    744749.

    17. Folwaczny M, Heym R, Mehl A, Hickel R. Subgingival cal-

    culus detection with fluorescence induced by 655 nm InG-

    aAsP diode laser radiation. J Periodontol 2002: 73: 597601.

    18. Folwaczny M, Mehl A, Haffner C, Benz C, Hickel R. Root

    substance removal with Er:YAG laser radiation at different

    parameters using a new delivery system. J Periodontol

    2000: 71: 147155.

    19. Frentzen M, Braun A, Aniol D. Er:YAG laser scaling of

    diseased root surfaces. J Periodontol 2002: 73: 524530.

    20. Geisinger ML, Mealey BL, Schoolfield J, Mellonig JT. The

    effectiveness of subgingival scaling and root planing: an

    evaluation of therapy with and without the use of the

    periodontal endoscope. J Periodontol 2007: 78: 2228.

    21. Hibst R, Paulus R, Lussi A. Detection of occlusal caries by

    laser fluorescence: basic and clinical investigation. Med

    Laser Appl 2001: 16: 205213.

    22. Hughes FJ, Smales FC. Immunohistochemical investigation

    of the presence and distribution of cementum-associated

    lipopolysaccharides in periodontal disease. J Periodontal

    Res 1986: 21: 660667.

    23. Kasaj A, Moschos I, Rohrig B, Willershausen B. The effec-

    tiveness of a novel optical probe in subgingival calculus

    detection. Int J Dent Hyg 2008: 6: 143147.

    24. Kasaj A, Sculean A, Willershausen B. Studie zur uberpru-

    fung der wirksamkeit des detectar-systems hinsichtlich

    einer verbesserung der compliance und der mundhygiene.

    Parodontologie 2004: 15: 343353.

    25. Kepic TJ, OLeary TJ, Kafrawy AH. Total calculus removal:an attainable objective? J Periodontol 1990: 61: 1620.

    26. Kocher T, Konig J, Hansen P, Ruhling A. Subgingival pol-

    ishing compared to scaling with steel curettes: a clinical

    pilot study. J Clin Periodontol 2001: 28: 194199.

    27. Kocher T, Ruhling A, Momsen H, Plagmann HC. Effec-

    tiveness of subgingival instrumentation with power-driven

    instruments in the hands of experienced and inexperi-

    enced operators. A study on manikins. J Clin Periodontol

    1997: 24: 498504.

    28. Kocher T, Strackeljan J, Behr D. Feasibility of computer-

    assisted recognition of different dental hard tissues. J Dent

    Res 2000: 79: 829834.

    29. Konig K, Flemming G, Hibst R. Laser-induced autofluo-

    rescence spectroscopy of dental caries. Cell Mol Biol

    (Noisy-le-grand) 1998: 44: 12931300.

    30. Krause F, Braun A, Brede O, Eberhard J, Frentzen M, Jepsen

    S. Evaluation of selective calculus removal by a fluores-

    cence feedback-controlled Er:YAG laser in vitro. J Clin

    Periodontol 2007: 34: 6671.

    31. Krause F, Braun A, Frentzen M. The possibility of detecting

    subgingival calculus by laser-fluorescence in vitro. Lasers

    Med Sci 2003: 18: 3235.

    32. Krause F, Braun A, Jepsen S, Frentzen M. Detection of

    subgingival calculus with a novel LED-based optical probe.

    J Periodontol 2005: 76: 12021206.

    33. Kurihara E, Koseki T, Gohara K, Nishihara T, Ansai T,

    Takehara T. Detection of subgingival calculus and dentine

    caries by laser fluorescence. J Periodontal Res 2004: 39: 59

    65.

    34. Lussi A, Hibst R, Paulus R. DIAGNOdent: an optical method

    for caries detection. J Dent Res 2004. 83 Spec No C: C8083.

    35. Lussi A, Megert B, Longbottom C, Reich E, Francescut P.

    Clinical performance of a laser fluorescence device for

    detection of occlusal caries lesions. Eur J Oral Sci 2001:

    109: 1419.

    36. Mandel ID, Gaffar A. Calculus revisited. A review. J Clin

    Periodontol 1986: 13: 249257.

    37. Meissner G, Oehme B, Strackeljan J, Kocher T. Clinical

    subgingival calculus detection with a smart ultrasonic de-

    vice: a pilot study. J Clin Periodontol 2008: 35: 126132.

    38. Meissner G, Oehme B, Strackeljan J, Kocher T. In vitro

    calculus detection with a moved smart ultrasonic device.

    J Clin Periodontol 2006: 33: 130134.

    39. Meissner G, Oehme B, Strackeljan J, Kocher T. Influence of

    handling-relevant factors on the behaviour of a novel

    calculus-detection device. J Clin Periodontol 2005: 32: 323

    328.

    40. Meissner G, Oehme B, Strackeljan J, Kocher T. A new sys-

    tem to detect residual subgingival calculus: in vitro detec-

    tion limits. J Clin Periodontol 2006: 33: 195199.

    41. Michaud RM, Schoolfield J, Mellonig JT, Mealey BL. The

    efficacy of subgingival calculus removal with endoscopy-

    aided scaling and root planing: a study on multirooted

    teeth. J Periodontol 2007: 78: 22382245.

    42. Nemes J, Csillag M, Toth Z, Fazekas A. Reproducibility of

    the laser fluorescence method for the diagnosis of occlusal

    caries. Clinical study. Fogorv Sz 2001: 94: 3336.

    43. Nyman S, Sarhed G, Ericsson I, Gottlow J, Karring T. Role of

    diseased root cementum in healing following treatment

    of periodontal disease. An experimental study in the dog.

    J Periodontal Res 1986: 21: 496503.

    44. Nyman S, Westfelt E, Sarhed G, Karring T. Role of dis-

    eased root cementum in healing following treatment of

    periodontal disease. A clinical study. J Clin Periodontol

    1988: 15: 464468.

    45. Pihlstrom BL, Ammons WF. Treatment of gingivitis and

    periodontitis. Research, Science and Therapy Committee

    of the American Academy of Periodontology. J Periodontol

    1997: 68: 12461253.

    46. Pinelli C, Campos Serra M, de Castro Monteiro Loffredo L.

    Validity and reproducibility of a laser fluorescence system

    for detecting the activity of white-spot lesions on free

    smooth surfaces in vivo. Caries Res 2002: 36: 1924.

    47. Pippin DJ, Feil P. Interrater agreement on subgingival

    calculus detection following scaling. J Dent Educ 1992: 56:

    322326.

    48. Qin YL, Luan XL, Bi LJ, Lu Z, Sheng YQ, Somesfalean G,

    Zhou CN, Zhang ZG. Real-time detection of dental calculus

    by blue-LED-induced fluorescence spectroscopy. J Photo-

    chem Photobiol B 2007: 87: 8894.

    49. Ross G. Caries diagnosis with the DIAGNOdent laser: a

    users product evaluation. Ont Dent 1999: 76: 2124.50. Ruhling A, Wulf J, Schwahn C, Kocher T. Surface wear on

    cervical restorations and adjacent enamel and root

    203

    Calculus-detection technologies

  • cementum caused by simulated long-term maintenance

    therapy. J Clin Periodontol 2004: 31: 293298.

    51. Schwarz F, Aoki A, Becker J, Sculean A. Laser application in

    non-surgical periodontal therapy: a systematic review.

    J Clin Periodontol 2008: 35: 2944.

    52. Schwarz F, Aoki A, Sculean A, Georg T, Scherbaum W,

    Becker J. In vivo effects of an Er:YAG laser, an ultrasonic

    system and scaling and root planing on the biocompati-

    bility of periodontally diseased root surfaces in cultures of

    human PDL fibroblasts. Lasers Surg Med 2003: 33: 140

    147.

    53. Schwarz F, Bieling K, Venghaus S, Sculean A, Jepsen S,

    Becker J. Influence of fluorescence-controlled Er:YAG laser

    radiation, the Vector system and hand instruments on

    periodontally diseased root surfaces in vivo. J Clin Period-

    ontol 2006: 33: 200208.

    54. Schwarz F, Sculean A, Berakdar M, Georg T, Reich E, Becker

    J. Clinical evaluation of an Er:YAG laser combined with

    scaling and root planing for non-surgical periodontal

    treatment. A controlled, prospective clinical study. J Clin

    Periodontol 2003: 30: 2634.

    55. Schwarz F, Sculean A, Berakdar M, Szathmari L, Georg T,

    Becker J. In vivo and in vitro effects of an Er:YAG laser, a

    GaAlAs diode laser, and scaling and root planing on peri-

    odontally diseased root surfaces: a comparative histologic

    study. Lasers Surg Med 2003: 32: 359366.

    56. Sculean A, Schwarz F, Berakdar M, Romanos GE, Arweiler

    NB, Becker J. Periodontal treatment with an Er:YAG laser

    compared to ultrasonic instrumentation: a pilot study.

    J Periodontol 2004: 75: 966973.

    57. Sherman PR, Hutchens LH Jr, Jewson LG, Moriarty JM,

    Greco GW, McFall WT Jr. The effectiveness of subgingival

    scaling and root planning. I. Clinical detection of residual

    calculus. J Periodontol 1990: 61: 38.

    58. Shi XQ, Welander U, Angmar-Mansson B. Occlusal caries

    detection with KaVo DIAGNOdent and radiography: an in

    vitro comparison. Caries Res 2000: 34: 151158.

    59. Stambaugh RV, Myers G, Ebling W, Beckman B, Stambaugh

    K. Endoscopic visualization of the submarginal gingiva

    dental sulcus and tooth root surfaces. J Periodontol 2002:

    73: 374382.

    60. Strackeljan J, Behr D, Kocher T. Fuzzy-pattern recognition

    for automatic detection of different teeth substances. Fuzzy

    Sets and Systems 1997: 85: 275286.

    61. Tan B, Gillam DG, Mordan NJ, Galgut PN. A preliminary

    investigation into the ultrastructure of dental calculus and

    associated bacteria. J Clin Periodontol 2004: 31: 364369.

    62. Tomasi C, Schander K, Dahlen G, Wennstrom JL.

    Short-term clinical and microbiologic effects of pocket

    debridement with an Er:YAG laser during periodontal

    maintenance. J Periodontol 2006: 77: 111118.

    63. Van der Weijden GA, Timmerman MF. A systematic review

    on the clinical efficacy of subgingival debridement in the

    treatment of chronic periodontitis. J Clin Periodontol 2002:

    29(Suppl 3): 5571.

    64. White DJ. Dental calculus: recent insights into occurrence,

    formation, prevention, removal and oral health effects of

    supragingival and subgingival deposits. Eur J Oral Sci 1997:

    105: 508522.

    65. Wilson TG Jr, Carnio J, Schenk R, Myers G. Absence of

    histologic signs of chronic inflammation following closed

    subgingival scaling and root planing using the dental

    endoscope: human biopsies - a pilot study. J Periodontol

    2008: 79: 3641.

    66. Yamaguchi H, Kobayashi K, Osada R, Sakuraba E, Nomura

    T, Arai T, Nakamura J. Effects of irradiation of an erbiu-

    m:YAG laser on root surfaces. J Periodontol 1997: 68: 1151

    1155.

    204

    Meissner & Kocher

  • Copyright of Periodontology 2000 is the property of Wiley-Blackwell and its content may not be copied oremailed to multiple sites or posted to a listserv without the copyright holder's express written permission.However, users may print, download, or email articles for individual use.


Recommended