Continuous wave nuclear magnetic resonance:estimation of spin-system properties from steady-state trajectories
James Christopher Korte
ORCID: 0000-0001-9152-1319
Ph.D Engineering (351AA)
August, 2017
Department of Biomedical Engineering, Melbourne School of Engineering
The University of Melbourne
Submitted in total fulfilment of the degree of
Doctor of Philosophy
2
Abstract
Magnetic resonance imaging (MRI) is a powerful imaging modality, widely used in routine clinical
practice and as an investigational tool in basic science. The contrast in MRI is related to both
the underlying tissue properties, which undergo disease or injury related changes, and to the MRI
method and sequence parameters used. It is the latter with which this thesis is concerned: the design
and implementation of novel MRI acquisition paradigms and associated reconstruction methods.
The majority of MRI methods excite the object of interest with a series of short RF pulses, varying
the weaker spatial magnetic field using the gradients, and ensuring the RF transmitter is inactive
while acquiring a series of decaying MR signals. This regime linearises the inherently nonlinear
behaviour of a magnetic resonance spin-system, allowing the acquired signals to be considered in a
spatial frequency space and an image to be reconstructed using the well known Fourier transform.
It is our assertion that nonlinear behaviour of the magnetic spin signal will lead to advantageous
attributes in future MR methods, just as moving beyond conventional linear spatial gradients to
nonlinear encoding fields led to methods for accelerated imaging and variable spatial resolution.
Reconstruction of spin-system properties from nonlinear MR signals requires algorithms beyond
the Fourier transform. In this thesis we propose spectroscopy, radial projection imaging and re-
laxometry methods as optimisation problems which minimise the mismatch between experimental
measurements and predictions from Bloch equation based signal models. The use of continuous
wave (CW) excitation patterns allows the development of signal models which are computationally
efficient as they rely on analytical solutions of the Bloch equations or matrix inversion via harmonic
balancing, rather than numerical integration.
Ultra-short relaxation methods have been applied to a range of applications and demonstrate that
MRI is finding use in areas far beyond traditional soft-tissue imaging. Soft tissues have an easily
observable long duration MR signal, whereas the signal decays rapidly for harder tissues such as
bone, or in regions that distort the magnetic field due to magnetic susceptibility gradients, such as
the lungs. Rabi modulated CW techniques operating in a fully continuous mode have the potential
to measure ultra-short relaxation signals in a similar range to ‘true’ zero echo time techniques.
Work inspired by quantum optics has shown that exciting a spin-system with a long duration Rabi
modulated RF field leads to a significant steady-state MR signal. The steady-state trajectory is
highly nonlinear and can be expressed as a series of harmonics of the amplitude modulation fre-
quency of the RF field. This harmonic response provides a natural decoupling of the excitation
and measurement bandwidth, and the ability to maintain a steady-state response under low power
excitation reduces the isolation requirements between hypothetical transmit and receive chains.
Our experimental investigation of steady-state trajectories makes use of two pseudo-simultaneous
excitation and measurement protocols. Whilst these methods were adequate to explore the proof-of-
concept applications, hardware modifications are suggested to unlock the full potential of continuous
wave excitation patterns.
This thesis demonstrates that CW excitation patterns allow the construction of efficient prediction
models and elicit an information-rich steady-state response from which underlying spin-system prop-
erties can be reconstructed. It is anticipated that further development of these concepts and related
hardware modifications will lead to new continuous wave imaging paradigms.
3
4
Declaration
This is to certify that
(i) the thesis comprises only their original work towards the Ph.D except where indicated in the
preface;
(ii) due acknowledgement has been made in the text to all other material used; and
(iii) the thesis is fewer than 100,000 words in length, exclusive of tables, maps, bibliographies and
appendices.
James Korte
Date
5
6
Publications
The work presented in this thesis has produced the following publications and conference presenta-
tions.
Journal papers
• Korte, J.C., Layton, K.J., Tahayori, B., Farrell, P.M., Moore, S.M. and Johnston, L.A. “NMR
spectroscopy using Rabi modulated continuous wave excitation”, Biomedical Signal Processing
and Control, 2017, 33, p.422-428
Conference abstracts
• Korte, J.C., Tahayori, B., Farrell, P.M., Moore, S.M. and Johnston, L.A. “Relaxometry via
steady-state ring-locked trajectories”, Proceedings of the 25th Annual Meeting of ISMRM,
Honolulu, USA, 2017
• Korte, J.C., Tahayori, B., Farrell, P.M., Moore, S.M. and Johnston, L.A. “Rabi Modulated
Continuous Wave Imaging”, Proceedings of the 24th Annual Meeting of ISMRM, Singapore,
2016 (Power Pitch Presentation)
• Korte, J.C., Tahayori, B., Farrell, P.M., Moore, S.M. and Johnston, L.A. “Gapped measure-
ment of spin system response to periodic continuous wave excitation”, The Australian and New
Zealand Magnetic Resonance Society, Bay of Islands, New Zealand, 2015 (Oral Presentation)
• Korte, J.C., Layton, K.J., Tahayori, B., Farrell, P.M., Moore, S.M. and Johnston, L.A. “En-
coding chemical shift with Rabi modulated continuous wave excitation”, Proceedings of the
22nd Annual Meeting of ISMRM, Milan, Italy, 2014 (Oral Presentation)
7
8
Acknowledgments
Firstly, I would like to express sincere gratitude to my supervisors for their patient guidance through-
out my PhD candidature. Leigh Johnston for the constant enthusiasm and lively debates, Peter
Farrell for the confusing but ultimately insightful comments, Bahman Tahayori for his mathemat-
ical genius and Stephen Moore for imparting his wide knowledge of parallel computation. I have
learnt a great deal from each of you and have no doubt your individual styles of thinking will shape
my research in the years to come. Additional thanks to the members of my advisory committee,
David Grayden and John Wagner, for keeping my candidature on track and offering perspective and
direction.
I am very grateful to my friends and colleagues from the Melbourne School of Engineering, the
Melbourne Brain Centre Imaging Unit and the Florey Neuroscience Institute: Roger Ordidge, Brad
Moffat, Scott Kolbe, Jon Cleary, Sonal Josan, Camille Shanahan, Amanda Ng, Kelvin Layton, David
Wright, Yasmin Blunck, Warda Syeda, Paul Bloembergen, Dhafer Alahmari, Eric Wang, Muhammad
Hanif, Muhammad Usman Khalid, Rosa Shishegar, Edward Green, Myrte Strik, Julia Neugebauer,
Peter Yoo, Annie Shelton, Sanuji Gajamange, Frederique Boonstra, Errol Lloyd and Yamni Mohan.
In particular, a big thank you to Kelvin and David for teaching me how to drive the MRI. Yasmin,
Warda, Rosa and Ed for all the interesting discussions in our weekly “MRI for idiots” meeting.
I have been lucky enough to visit a few labs during my candidature. I learnt a great deal during a
one month lab visit with the MRI group in Freiburg, my thanks to Jurgen Hennig, Maxim Zaitsev,
Frederik Testud and Sebastian Littin. Thanks to Samuel Patz and Mirko Hrovat for your valuable
comments on my experiments and showing me around your laboratory in Boston. I would also like
to thank Steffen Bollmann and the Centre for Advance Imaging in Queensland for your feedback on
my work.
To my friends for reminding me there is more to life than research. Thanks for all the Friday night
sessions at E55 to vent some frustration when things are not going to plan, for helping me stay
active with great company on hiking and climbing trips and keeping me sharp with the occasional
acoustic rap battle.
My family is a great source of inspiration, I would like to thank my parents Chris and Katherine for
instilling in me the importance of knowledge and supporting my scientific curiosity. To my sisters for
teaching and challenging me; Laura for introducing me to scepticism and debate, Anna for reminding
me intelligence has many forms.
Lastly, I’d like to thank all the medical professionals who have patched me up over the years and
continue to inspire me to pursue research in a health related field.
9
10
Contents
1 Introduction 17
2 Theory 23
2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
2.2 Spin Physics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.2.1 Single atomic particle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.2.2 Ensemble of atomic particles . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
2.3 Bloch equations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
2.3.1 Steady-state solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
2.3.2 Pulse excitation solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
2.4 Measurement Concepts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
2.4.1 Signal detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
2.4.2 Signal formation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
2.4.3 NMR spectroscopy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
2.4.4 NMR relaxometry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
2.5 Imaging Concepts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
2.5.1 Slice selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
2.5.2 Spatial encoding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Appendices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
2.A Bloch equations in the rotating frame . . . . . . . . . . . . . . . . . . . . . . . . . . 59
2.B Bloch equations steady-state solution to constant excitation . . . . . . . . . . . . . . 60
3 Methods 63
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
3.2 Proton Density Imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
3.3 Relaxation Mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
3.3.1 RARE-VTR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
3.3.2 MSME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
3.4 Magnetic Field Mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
3.4.1 B0 mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
3.4.2 B1 mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
4 Rabi continuous wave spectroscopy 73
4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
4.2 Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
4.2.1 Rabi modulated excitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
4.2.2 Observed NMR signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
4.2.3 Spectroscopy as an inverse problem . . . . . . . . . . . . . . . . . . . . . . . . 76
4.3 Methods and Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4.3.1 The response of off-resonance spins . . . . . . . . . . . . . . . . . . . . . . . . 79
4.3.2 Rabi modulated spectroscopy . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
4.4 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
11
12 CONTENTS
4.4.1 The response of off-resonance spins . . . . . . . . . . . . . . . . . . . . . . . . 83
4.4.2 Rabi modulated spectroscopy . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
4.5 Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
4.6 Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
5 Rabi continuous wave imaging 89
5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
5.2 Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
5.2.1 Observed NMR signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
5.2.2 Radial projection imaging as an inverse problem . . . . . . . . . . . . . . . . 93
5.3 Methods and Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
5.3.1 Gapped excitation and measurement protocol . . . . . . . . . . . . . . . . . . 94
5.3.2 Gapped measurement of off-resonance response . . . . . . . . . . . . . . . . . 94
5.3.3 Rabi modulated imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
5.3.4 Rabi modulated imaging contrast . . . . . . . . . . . . . . . . . . . . . . . . . 100
5.4 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
5.4.1 Gapped measurement of off-resonance response . . . . . . . . . . . . . . . . . 100
5.4.2 Rabi modulated imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
5.4.3 Rabi modulated imaging contrast . . . . . . . . . . . . . . . . . . . . . . . . . 102
5.5 Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
5.6 Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
6 Ring-lock excitation 111
6.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
6.2 Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
6.2.1 Steady-state ring-lock response . . . . . . . . . . . . . . . . . . . . . . . . . . 114
6.2.2 Estimation of relaxation constants . . . . . . . . . . . . . . . . . . . . . . . . 114
6.3 Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
6.3.1 Verification of excitation envelope . . . . . . . . . . . . . . . . . . . . . . . . 115
6.3.2 Estimation of relaxation constants . . . . . . . . . . . . . . . . . . . . . . . . 118
6.4 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
6.5 Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
6.6 Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Appendices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
6.A Ring-lock excitation envelope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
6.A.1 Amplitude and phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
6.B Spin-system response under ring-lock excitation . . . . . . . . . . . . . . . . . . . . . 125
7 Conclusion 129
7.1 Summary of original contributions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129
7.2 Future work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130
Bibliography 131
List of Figures
2.1 Magnetic dipole precession . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
2.2 Laboratory and rotating frames . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
2.3 Behaviour of a magnetic dipole in a static magnetic field and an RF field . . . . . . 27
2.4 Static magnetic field leads to bulk magnetisation . . . . . . . . . . . . . . . . . . . . 29
2.5 Random orientation of a precessing nuclear magnetic dipole. . . . . . . . . . . . . . . 29
2.6 Bulk magnetisation response to RF excitation . . . . . . . . . . . . . . . . . . . . . . 30
2.7 Steady-state magnetisation under constant excitation over a range of off-resonance
(low power) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
2.8 Steady-state magnetisation under constant excitation over a range of off-resonance . 35
2.9 Measured and predicted periodic steady-state magnetisation in the time and frequency
domain. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
2.10 Free induction decay signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
2.11 Spin-echo sequence diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
2.12 Gradient-echo sequence diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
2.13 FID relation to spectral width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
2.14 Inversion recovery sequence diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
2.15 Progressive saturation sequence diagram . . . . . . . . . . . . . . . . . . . . . . . . . 50
2.16 Spin-echo CPMG sequence diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
2.17 RF envelope relationship with slice profile . . . . . . . . . . . . . . . . . . . . . . . . 53
2.18 Refocusing a slice selective pulse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
2.19 Spin-echo based k-space trajectory diagram . . . . . . . . . . . . . . . . . . . . . . . 57
2.20 Gradient-echo based k-space trajectory diagram . . . . . . . . . . . . . . . . . . . . . 58
3.1 Fast low angle shot imaging (FLASH) sequence diagram. . . . . . . . . . . . . . . . 64
3.2 RARE sequence diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
3.3 RARE sequence diagram for T2 relaxometry . . . . . . . . . . . . . . . . . . . . . . . 66
3.4 MSME sequence diagram for efficient multi-slice T2 relaxometry . . . . . . . . . . . . 68
3.5 B0 mapping sequence diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
3.6 B1 mapping sequence diagram. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
4.1 Experimental protocol for the incremental measurement of the steady-state magneti-
sation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
4.2 Measured and predicted periodic steady-state magnetisation in the time and frequency
domain. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
4.3 Feasible and selected excitation parameter pair sets. . . . . . . . . . . . . . . . . . . 80
4.4 Trace of covariance matrix of reconstructed spectra. . . . . . . . . . . . . . . . . . . 82
4.5 Measured distributions of off-resonance and RF power. . . . . . . . . . . . . . . . . . 83
4.6 Measured and predicted frequency coefficients of the steady-state magnetisation. . . 84
4.7 Reconstructed NMR spectra of ethanol. . . . . . . . . . . . . . . . . . . . . . . . . . 85
5.1 Radial coordinate system and the Radon transform. . . . . . . . . . . . . . . . . . . 91
5.2 Comparison of spin-system response under gapped Rabi modulated excitation . . . . 95
13
14 LIST OF FIGURES
5.3 Gapped excitation protocol for near-simultaneous transmit and receive . . . . . . . . 96
5.4 Diagram of three test-tube imaging phantom. . . . . . . . . . . . . . . . . . . . . . . 97
5.5 Gapped imaging sequence diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
5.6 Numerical phantom and simulated contrast of Rabi modulated imaging . . . . . . . 101
5.7 Measured distributions of off-resonance and RF power. . . . . . . . . . . . . . . . . . 102
5.8 Measured and predicted frequency coefficients of the steady-state magnetisation. . . 103
5.9 Reference FLASH image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
5.10 Rabi modulated radial imaging results . . . . . . . . . . . . . . . . . . . . . . . . . . 105
5.11 Line profile of imaging results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
5.12 Line profiles of simulated contrast of Rabi modulated imaging. . . . . . . . . . . . . 107
6.1 Simulated spin-system response to ring-lock excitations. . . . . . . . . . . . . . . . . 112
6.2 Steady-state ellipsoid for different relaxation parameters. . . . . . . . . . . . . . . . . 113
6.3 Simulated spin-system response and the iterative excitation protocol. . . . . . . . . . 116
6.4 Numerical validation of excitation envelope. . . . . . . . . . . . . . . . . . . . . . . . 117
6.5 Experimental validation of excitation envelope. . . . . . . . . . . . . . . . . . . . . . 119
6.6 The measured and predicted transverse steady-state magnetisation surface. . . . . . 120
List of Tables
4.1 Results from the analysis of the reference and reconstructed ethanol spectra . . . . . 86
6.1 Measured and estimated relaxation constants under ring-lock excitation . . . . . . . 120
15
16 LIST OF TABLES
Chapter 1
Introduction
Magnetic resonance imaging (MRI) is a powerful imaging modality, widely used in routine clinical
practice and as an investigational tool in basic science. The contrast in MRI is related to both the
underlying tissue properties, which undergo disease or injury related changes, and to the MRI method
and sequence parameters used. A major advantage of this technology is its ability to non-invasively
image the human body using non-ionising electromagnetic radiation. Comparable medical imaging
techniques, such as X-ray CT, use ionising radiation which under certain conditions can increase
the risk of cancer due to DNA damage (National Toxicology Program, 2016; Leuraud et al., 2015;
Behjati et al., 2016). The use of non-ionising radiation makes MRI an ideal tool for long-term clinical
applications, such as the monitoring of disease progression and longitudinal scientific studies. MRI is
a versatile tool which has the ability to acquire images with structural, functional, diffusion, perfusion
and mechanical information using the same scanner. It has been strongly adopted by neurological
disciplines due to the available contrast in brain images, the dynamic brain activity inferred from
functional MRI data and the micro-structural information provided by diffusion imaging.
A conventional MRI scanner generates a strong static magnetic field, superimposed with weaker spa-
tially varying magnetic fields and an excitatory magnetic field which oscillates in the radio frequency
(RF) range. These electromagnetic fields provide the conditions to form a nuclear magnetic reso-
nance (NMR) signal which is measured by an induced voltage in a coil (Bloch, 1946). Early NMR
experiments used continuous wave RF excitation (Purcell et al., 1946; Bloch et al., 1946) but were
largely replaced by pulsed Fourier transform techniques due to improved efficiency and sensitivity
(Ernst and Anderson, 1966). The spatially varying magnetic field, commonly referred to as the gra-
dient field, creates a spatial frequency dependant NMR signal used to perform imaging (Carr, 1953;
Lauterbur et al., 1973). The majority of MRI methods excite the object of interest with a series of
short RF pulses, varying the weaker spatial magnetic field using the gradients, and ensuring the RF
transmitter is inactive while acquiring a series of decaying MR signals. This regime linearises the
inherently nonlinear behaviour of a magnetic resonance spin-system, allowing the acquired signals
to be considered in a spatial frequency space (k-space) (Brown et al., 1982; Ljunggren, 1983; Twieg,
1983) and an image to be reconstructed using the well known Fourier transform. The efficiency of
any method of this class is related to its efficiency in sampling k-space, and is constrained by the
Nyquist criteria. Alternative methods perform image reconstruction via optimisation of generalised
signal models and reduce the sampling constraints of a full k-space acquisition by incorporating
additional information, such as multiple signals from receiver coil arrays (Pruessmann et al., 1999;
Griswold et al., 2002), or impose sparsity requirements on the sampling pattern (Lustig et al., 2007).
17
18 CHAPTER 1. INTRODUCTION
It is often the case that spatial and temporal maps of parameter estimates from the reconstructed
images are desired. Parameter mapping methods often require the measurement of multiple images,
such as images at multiple echo times for relaxation mapping. Accelerated parameter mapping using
model based optimisation methods that utilise Bloch equation simulations, have been demonstrated
to jointly estimate multiple tissue parameter maps from reduced data sets (Block et al., 2009; Sumpf
et al., 2011; Ma et al., 2013).
Nonlinearity of the magnetic resonance spin-system is not typically the focus of any well adopted
spectroscopy or imaging technique. On the contrary, all efforts are made to linearise the observed
spin-system behaviour by use of short RF pulses and acquisition of free-induction decays or echo
signals in the absence of an applied RF field. It is our assertion that nonlinear behaviour of the
magnetic spin signal will lead to advantageous attributes in future MR methods, just as moving be-
yond conventional linear spatial gradients to nonlinear encoding fields led to methods for accelerated
imaging and variable spatial resolution (Hennig et al., 2008; Stockmann et al., 2010).
Work inspired by quantum optics (Cappeller and Muller, 1985) has shown that exciting a spin-
system with a long duration amplitude modulated RF field leads to a significant steady-state MR
signal (Tahayori et al., 2008). The amplitude of the steady-state response is maximised when the
amplitude modulation is set to the average power of the excitation envelope (Layton et al., 2014). We
refer to the excitation pattern necessary to achieve this steady-state behaviour as Rabi modulated
excitation or Rabi excitation. The steady-state trajectory is highly nonlinear and can be expressed
as a series of harmonics of the amplitude modulation frequency of the RF field (Tahayori et al.,
2008). This harmonic response provides a natural decoupling of the excitation and measurement
bandwidth, and the ability to maintain a steady-state response under low power excitation (Tahayori
et al., 2009) reduces the isolation requirements between hypothetical transmit and receive chains.
Rabi modulated excitation is therefore well suited to applications that benefit from the ability
to observe the spin-system during excitation, such as closed-loop control of the magnetisation or
measurement of signals beyond the limits of conventional free induction decay (FID) methods.
In majority of human studies, the magnetic resonance signal comes from water in tissue, or more
specifically the hydrogen proton. The abundance of hydrogen and the influence of the surrounding
chemical environment impact the strength and the duration of the MR signal. Soft tissues have an
easily observable long duration MR signal, whereas the signal decays rapidly for harder tissues such
as bone, or in regions that distort the magnetic field due to magnetic susceptibility gradients, such
as the lungs. An additional fast relaxation application is the imaging of quadrupolar nuclei, such
as sodium and phosphorus, which is becoming more feasible due to stronger signal at higher static
magnetic field strengths (Wang et al., 2009). These rapidly decaying signals lead to poor image
contrast and reduced spatial resolution due to blurring image artefacts. Due to these limitations,
and to relative cost and efficiency, X-ray and X-ray CT are still the dominant clinical tool for many
imaging applications.
The first efforts to image fast relaxing spins was by optimisation of existing techniques based on
gradient or spin-echoes, to reduce the finite echo time, TE, to hardware and safety limits (Robson
and Bydder, 2006). Whilst these optimisations were able to image many tissue types, some with
ultra-short relaxation rates were still unobservable. Free induction decay projection methods such
as ultrashort echo time imaging (Pauly et al., 1989a; Bergin et al., 1991) (UTE) reduced the effective
echo time down to hardware switching limits. This was followed by zero-echo time imaging (Hafner,
1994; Madio and Lowe, 1995) (ZTE), allowing an even shorter effective echo time but at the cost of
a hole in central k-space to be filled by algebraic techniques (Kuethe et al., 1999) or additional data
acquisition (Grodzki et al., 2012). Alternative techniques include single point FID imaging (Nauerth
and Gewiese, 1993), water-and fat-suppressed proton projection imaging (Wu et al., 2003) and
sweep imaging with Fourier transformation (Idiyatullin et al., 2006). As mentioned previously,
19
Rabi modulated excitation drives the magnetisation into a steady-state with favourable properties
for simultaneous excitation and measurement. Rabi modulated techniques operating in a fully
continuous mode have the potential to measure ultra-short relaxation signals in a similar range
to ‘true’ zero echo time techniques such as continuous sweep imaging with Fourier transformation
(Idiyatullin et al., 2012) (cSWIFT).
Sweep imaging with Fourier transformation (Idiyatullin et al., 2006) (SWIFT) is similar to ZTE
imaging in that the gradients are applied during RF excitation. The excitation for SWIFT is a low
power, frequency swept RF pulse rather than a high power hard pulse used in ZTE. SWIFT is based
on early CW NMR techniques such as RapidScan NMR (Gupta et al., 1974) and Correlation NMR
(Dadok and Sprecher, 1974) for fast acquisition of spectra. SWIFT is commonly implemented as a
time-shared excitation and acquisition but is inherently continuous wave, and due to it’s low power
excitation has been demonstrated as a fully continuous method (Idiyatullin et al., 2012) (cSWIFT) .
Continuous SWIFT has zero dead-time allowing the measurement of very short T2 relaxation signals
and broad distributions of chemical shifts. In comparison, the shortest reported dead-time for clinical
ZTE (Weiger et al., 2013a) is 5µs. The effective echo-time of UTE is dominated by ramp-time of
the gradient system and peripheral nerve stimulation constraints. Specific absorption rate concerns
for SWIFT, due to its high duty cycle, have been addressed by gradient modulation (Zhang et al.,
2016) (gSWIFT).
Ultra-short relaxation methods have been demonstrated in a range of applications such as dental
imaging, musculoskeletal imaging, lung imaging, x-nuclei imaging and superparamagnetic iron oxide
nanoparticles (SPION) imaging. SWIFT has been used in studies of bone (Kendi et al., 2011; Luhach
et al., 2014), cartilage (Rautiainen et al., 2013), teeth (Idiyatullin et al., 2011), lungs (Kobayashi
et al., 2015), breast (Nelson et al., 2012), brain calcification (Lehto et al., 2012) and SPIONs (Zhang
et al., 2014). UTE has been used in studies of bone (Robson et al., 2004; Robson and Bydder,
2006), cartilage (Brossmann et al., 1997; Gold et al., 1998), knee (Gatehouse et al., 2004; Rahmer
et al., 2006), achilles tendon (Filho et al., 2009), lung (Bergin et al., 1991), liver (Chappell et al.,
2003), USPIOs (Crowe et al., 2005), sodium (Nielles-Vallespin et al., 2007), phosphorus (Anumula
et al., 2006) and brain calcification (Waldman et al., 2003). ZTE has been used in studies of
teeth (Weiger et al., 2012, 2013b), joints (Weiger et al., 2013a) and the skull (Delso et al., 2015). All
these applications of ultra-short relaxation methods demonstrate that MRI is finding use in areas
far beyond traditional soft-tissue imaging.
Rabi modulated excitation of the hydrogen proton has many exciting properties and potential ap-
plications, but at the commencement of this thesis the existing research was limited to a theoretical
framework and a single experimental study. The theoretical work of Tahayori et al. (2009) chal-
lenged the optimality of existing excitation methods and later developed an efficient framework to
find steady-state trajectories to any periodic RF excitation (Tahayori et al., 2015), using Rabi mod-
ulated excitation as a case study. The work of Layton et al. (2014) derived an analytical solution
of the Bloch equations under on-resonance Rabi modulated excitation and developed experimental
techniques to verify this response in a phantom. This manuscript builds on these existing works and
explores the use of the off-resonance response to Rabi modulated excitation to develop novel mag-
netic resonance methods; analogous to historical nuclear magnetic resonance development, first a
spectroscopy method (Korte et al., 2014, 2017a) followed by an imaging method (Korte et al., 2016).
During the investigation of steady-state trajectories, a known steady-state response to constant RF
excitation was revisited, and with the addition of off-resonance, was used to pose relaxometry as a
Bloch model based optimisation problem (Korte et al., 2017b). The novel contributions in this work
estimate tissue properties via optimisation of Bloch equation models against experimental measure-
ments. The Bloch equation based signal models developed in this manuscript are computationally
efficient as they use analytical solutions or matrix inversion via harmonic balancing (Tahayori et al.,
20 CHAPTER 1. INTRODUCTION
2015) as detailed in Section 2.3.1.2, rather than numerical integration.
Following this introduction, the thesis is organised into five major chapters, and a final concluding
chapter. Of the five major chapters, the first two provide theory of nuclear magnetic resonance
physics and imaging methods relevant to the following three novel chapters. The final chapter
summarises the major contributions of this work and makes suggestions for future research directions.
The theory chapter, Chapter 2, introduces the requisite knowledge upon which the remainder of
the thesis is constructed, containing a summary of NMR physics followed by fundamental MRI
concepts. The physics section explains the behaviour of a spin-system under a static magnetic field,
an oscillating magnetic field and observed relaxation mechanisms. The combined dynamics of the
spin-system are then expressed with the Bloch equations in a laboratory and rotating frame of
reference. Bloch equation solutions for a steady-state response under constant excitation, and in
response to a short RF pulse are discussed in relation to the shift from continuous wave (CW) to
pulsed free induction decay Fourier transform methods. A method used throughout the thesis for
efficient calculation of steady-state response to any periodic excitation is described. Volume signal
methods for the measurement of NMR spectra and estimation of relaxation constants are discussed.
We then move into basic magnetic resonance imaging concepts: k-space, slice selection and image
reconstruction.
The methods chapter, Chapter 3, details the standard imaging methods used to make reference mea-
surements for the experiments in the novel chapters. In particular, we present the fast proton density
imaging method used to acquire reference images in Chapter 5. Relaxation mapping methods are
described for the measurement of spin-lattice and spin-spin relaxation maps, which are used as prior
knowledge in the model based reconstructions in Chapters 4-5 and as the ground truth in Chapter 6.
Inhomogeneity in the static magnetic field and the RF magnetic field are also incorporated as prior
knowledge in the estimation models of Chapters 4-6. The details of measurement of field maps are
provided.
The first novel chapter, Chapter 4 entitled Rabi continuous wave spectroscopy, presents the first
proof-of-concept application of Rabi modulated excitation (Korte et al., 2017a). The observed
signal from a spin-system under Rabi modulated excitation in the presence of off-resonance and field
inhomogeneities is described. This expression is then used to pose NMR spectroscopy as an inverse
problem, where the spin-system is interrogated with a series of different Rabi modulated excitation
envelopes. An iterative measurement protocol (Layton et al., 2014) is used to experimentally measure
the steady-state trajectory of the spin-system in two experiments: the first experiment to validate
the off-resonance response model under Rabi modulated excitation, the second to demonstrate the
measurement of NMR spectra under a series of Rabi modulated excitations.
The second novel chapter, Chapter 5 entitled Rabi continuous wave imaging, is the progression
of the spectroscopy experiment in Chapter 4. To allow a wider bandwidth reconstruction, the
harmonic estimation model used in the spectroscopy experiment is modified to include an offset to
the RF carrier. Proton projection imaging is then posed as an inverse problem, where a weak radial
gradient is applied and the spin-system is interrogated with a series of different Rabi modulated
excitation envelopes. A series of estimated proton projections are then used to reconstruct a two
dimensional image using filtered back projection. In order to make this method experimentally
feasible, a gapped excitation protocol, inspired by time shared SWIFT (Idiyatullin et al., 2006),
was implemented. The gapped excitation protocol was verified by experimental measurement of the
off-resonance response. A two dimensional imaging experiment was then performed to demonstrate
proton density projections can be reconstructed from the harmonic response to Rabi modulated
excitation. Numerical simulations were performed to explore the relaxation contrast of this new
imaging method.
21
The third and final novel chapter, Chapter 6 entitled Ring-lock excitation, is the exploration of a
different (non-Rabi modulated) steady-state trajectory. The known response to a constant amplitude
excitation is revisited, with the addition of off-resonance effects. The magnetisation spirals down to a
steady-state point in response to an on-resonance constant excitation. The same constant excitation,
applied off-resonance, drives the magnetisation into a steady-state ring trajectory. We derive an
analytical expression for the ring-lock steady-state response in terms of the excitation power and
excitation off-resonance. Two experiments are conducted using the iterative measurement protocol
describe in Chapter 4. The first experiment verifies the derived excitation envelope by demonstrating
control over the ring-lock response, when relaxation constants are known. The second experiment
utilises ring-lock excitation to estimate relaxation constants, through probing a sample with a series
of different ring-lock excitations which are optimised against a theoretical ring-lock model.
This thesis demonstrates that information-rich steady-state trajectories of the spin-system can be
used to reconstruct chemical shift spectra, proton density projections and estimate relaxation rates.
The described psuedo-simultaneous measurement protocols allow the investigation of continuous
wave phenomena on a modern scanner. The novel contributions establish the benefit of contin-
uous wave excitation patterns in conjunction with model based estimation and suggest hardware
modification for a new class of continuous wave steady-state methods.
22 CHAPTER 1. INTRODUCTION
Chapter 2
Theory
Contents2.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
2.2 Spin Physics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.2.1 Single atomic particle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
2.2.2 Ensemble of atomic particles . . . . . . . . . . . . . . . . . . . . . . . . . 27
2.3 Bloch equations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
2.3.1 Steady-state solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
2.3.2 Pulse excitation solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
2.4 Measurement Concepts . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
2.4.1 Signal detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
2.4.2 Signal formation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
2.4.3 NMR spectroscopy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
2.4.4 NMR relaxometry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
2.5 Imaging Concepts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
2.5.1 Slice selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
2.5.2 Spatial encoding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Appendices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
2.A Bloch equations in the rotating frame . . . . . . . . . . . . . . . . . . . 59
2.B Bloch equations steady-state solution to constant excitation . . . . . 60
2.1 Introduction
This chapter describes the theory which provides the foundation for the methods (Chapter 3) and
the experimental chapters (Chapters 4-6). The physics of nuclear magnetic resonance is presented
from a classical point of view, for a single spin and then for an ensemble of spins to form the bulk
magnetisation. The Bloch equations are defined in the laboratory and rotating frames and some
relevant solutions are discussed. Fundamental measurement concepts such as nuclear induction,
echo formation and techniques for acquiring NMR spectra and estimating relaxation constants are
described. Imaging concepts such as slice selection, the k-space formalism, phase encoding and
23
24 CHAPTER 2. THEORY
frequency encoding are presented with some imaging sequence examples. The contents of this
chapter are derived from a range of sources (Levitt, 2001; Liang et al., 2000; Haacke et al., 1999;
Abragam, 1961; Ernst et al., 1990; Slichter, 1978) where each topic can be found in greater detail.
2.2 Spin Physics
This section describes how a subatomic particle with half-integer spin is observed as a magnetic
dipole, its behaviour in the presence of additional magnetic fields and how the behaviour of a group
of these particles form a macroscopic dipole known as the bulk magnetisation vector.
2.2.1 Single atomic particle
Atoms are composed of particles with known attributes such as mass, electric charge, magnetism
and angular momentum. Angular momentum is a combination of orbital angular momentum, such
as an electron orbiting a nucleus, and spin angular momentum, such as a proton rotating on it’s axis
which is often referred to as spin.
2.2.1.1 Nuclear magnetic dipole
A particle’s magnetic moment, µ, is proportional to it’s spin angular momentum, as stated by a
fundamental symmetry theorem (Levitt, 2001),
µ = γS, (2.1)
where γ is the gyromagnetic ratio in units of rad · s−1 ·T−1. The magnitude of the magnetic dipole
can be expressed as
|µ| = γ~√s(s+ 1), (2.2)
where ~ = h/2π is the reduced Plank’s constant and the spin quantum number, s, can take zero,
positive integer and half integer values (0, 12 , 1,
32 , . . . ,
n2 ). The gyromagnetic ratio is often expressed
in units of Hz ·T−1 and has a positive value for a hydrogen proton but is negative for certain other
particles such as an electron.
2.2.1.2 Behaviour in a static magnetic field
A magnetic dipole in an external magnetic field will rotate to align with that external magnetic
field, such as the bar magnet of a compass aligning with the earth’s magnetic field. A nuclear
magnetic dipole also possesses spin angular momentum and exhibits slightly different dynamics to
a bar magnet as shown in Figure 2.1.
The nuclear magnetic dipole will experience a torque,
τ = µ× b, (2.3)
when in an external magnetic field, b. The bold symbols are vectors in a Cartesian coordinate space
with the orthonormal vectors, i, j,k. An arbitrary vector, v, is expressed in a stationary frame of
reference known as the laboratory frame,
v = vxi+ vyj + vyk. (2.4)
2.2. SPIN PHYSICS 25
Figure 2.1: A single magnetic dipole, µ, under a static magnetic field, B0, precessing at the Larmor frequency, ω0.
We can relate the change of total angular momentum, L, to external torque, τ , using Newton’s
second law of motion,
dL
dt= τ . (2.5)
Here we consider a half-integer spin proton such as a hydrogen proton with a total angular momentum
equal to the spin angular momentum, L = S. Substituting (2.1) into (2.5) gives a differential
equation for a nuclear magnetic dipole in an external magnetic field,
dµ
dt= γµ× b. (2.6)
If we consider a static external magnetic field in the z direction,
b = B0k, (2.7)
the solution of (2.6) gives the dynamics of a nuclear magnetic dipole as,
µxy (t) = µxy (0) e−i γB0 t (2.8a)
µz (t) = µz (0) (2.8b)
where the longitudinal component, µz, is constant and the transverse component is expressed in the
complex plane, µxy = µx + iµy, and precesses around the external field at the Larmor frequency,
ω0 = −γB0. (2.9)
2.2.1.3 Response to an oscillating magnetic field
A nuclear magnetic dipole was shown to precess around a static magnetic field in Section 2.2.1.2.
If an oscillating magnetic field is applied in the xy plane at a similar frequency to the Larmor
frequency (2.9) a secondary precession is observed.
26 CHAPTER 2. THEORY
(a) (b)
Figure 2.2: A single magnetic dipole, under a static magnetic field, B0, shown in two frames of reference. In thelaboratory frame (a) the dipole precesses at the Larmor frequency, ω0. In a rotating frame (b) with the rotationalfrequency set to the Larmor frequency, the dipole appears stationary.
Rotating frame
It is useful to consider the spin-system in a rotating frame of reference where transverse rotations,
such as spin precession, appear static if the frequency of the rotating frame and the spin precession
are matched. This is depicted in Figure 2.2. The transformation from the laboratory frame to the
rotating frame is, i′j′k′
=
cos(ωrott) − sin(ωrott) 0
sin(ωrott) cos(ωrott) 0
0 0 1
ijk
(2.10)
= Rz,ccw (ωrot, t)
ijk
(2.11)
where ωrot is the angular frequency of the rotating frame. An arbitrary vector, v′, is expressed in
the rotating frame as,
v′ = v′xi′ + v′yj
′ + v′yk′. (2.12)
Spin nutation
We examine the behavior of a nuclear magnetic dipole under a static magnetic field, b0, and a weaker
oscillating magnetic field, b1, with an oscillation frequency, ωRF, at the nominal Larmor frequency,
ω0. The weaker oscillating field is often referred to as the RF excitation because at common static
field strengths the Larmor frequency (2.9) resides in the radio-frequency range. We define a circularly
polarised RF excitation in the laboratory frame as
b1 (t) = Be1 (t)
[cos(ωRFt+ θRF (t)
)i+ sin
(ωRFt+ θRF (t)
)j], (2.13)
2.2. SPIN PHYSICS 27
Figure 2.3: A single magnetic dipole, µ, in a Larmor rotating frame, under a static magnetic field, B0, and anoscillating magnetic field, B1, applied in the i′ direction. The oscillating magnetic field causes the magnetic dipole toprecess around i′ at the Rabi frequency, ω1.
where Be1 (t) is the excitation envelope, ωRF is the excitaton carrier frequency and θRF (t) is the
excitation phase offset. The RF excitation appears stationary in the RF rotating frame (2.10) where
ωrot = ωRF and the excitation phase offset is constant θRF (t) = θRF,
b′1 (t) = Be1 (t)
(cos(θRF
)i′ + sin
(θRF
)j′), (2.14)
and for zero excitation phase offset is,
b′1 (t) = Be1 (t) i′. (2.15)
The effect of the weaker oscillating field is to rotate the nuclear magnetic dipole around the excitation
vector, b′1 (t), as shown in Figure 2.3. This secondary rotation has a nutation or Rabi frequency
proportional to the strength of the applied RF excitation,
ω1 = ωe1 (t) = −γBe
1 (t) . (2.16)
See (Levitt, 2001, Chapter 10) for a quantum mechanical derivation of this result.
2.2.2 Ensemble of atomic particles
In the previous section we described the behaviour of a single nuclear magnetic dipole in a static
field and a weaker oscillating field. Experimentally we normally observe the behaviour of a group of
nuclear magnetic dipoles. In this section we introduce the behaviour of an ensemble of spins under
similar condition as considered in Section 2.2.1 where we also observe relaxation phenomena.
To discuss the behaviour of a group of nuclear magnetic dipoles, we define the bulk magnetisation
vector as the vector sum of nuclear magnetic dipoles in a sample,
m =
Ns∑i=1
µi (2.17)
28 CHAPTER 2. THEORY
where Ns is the number of spins in a volume of interest. An ensemble of spins in the absence
of a magnetic field has a random distribution of orientations leading to zero bulk magnetisation
(Figure 2.4a).
2.2.2.1 Spin-Lattice relaxation
An ensemble of spins in a static magnetic field is shown in Figure 2.4b. Each spin of the ensem-
ble will precess around the static field at it’s Larmor frequency, but due to a random transverse
phase distribution there is zero transverse bulk magnetisation. The individual spins experience a
fluctuating magnetic field due to thermal energy and their molecular environment which cause a
precessional wandering (Levitt, 2001, Chapter 2) as shown in Figure 2.5. Lower energy orientations,
those with components aligned with the static magnetic field, are slightly favoured by the wandering
spins which leads to a longitudinal bulk magnetisation. The ratio of low and high energy spins is
described by the Boltzmann relationship,
N↑N↓
= exp
(γB0
KTs
), (2.18)
where N↑ is the number of low energy spins, N↓ is the number of high energy spins, B0 is the mag-
nitude of the static magnetic field, K is the Boltzmann constant and Ts is the absolute temperature
of the spin-system. This can also be expressed as a ratio of the difference of spin orientations and
total number of spins,
N↑ −N↓Ns
≈ γB0
2KTs. (2.19)
If we consider a hydrogen sample, γ = 42.58 MHz · T−1, at room temperature, Ts = 300 K, and
a field strength, B0 = 1 T, using (2.19) approximately three in every million spins contribute to
the longitudinal bulk magnetisation at thermal equilibrium. The bulk magnetisation at thermal
equilibrium can be expressed as (Haacke et al., 1999)
M0z =
1
4ρ0γ22
KTsB0, (2.20)
where ρ0 is the spin density.
When the static field, B0, is turned on the bulk magnetisation is not formed instantaneously. The
relaxation rate of the spin-system towards the thermal equilibrium bulk magnetisation, M0z , is depen-
dant on molecular environment, static field strength and temperature. The longitudinal relaxation
of the bulk magnetisation is described by the phenomenological model,
mz (t) =(mz (0)−M0
z
)e−
tT1 +M0
z , (2.21)
where T1 is the longitudinal relaxation constant, also referred to as the spin-latice relaxation constant.
The relaxation model (2.21) can also be expressed as the first order system,
dmz (t)
dt=
1
T1
(M0
z −mz (t)). (2.22)
2.2.2.2 Spin-Spin relaxation
An ensemble of spins in a static magnetic field will relax to a non-zero longitudinal bulk magneti-
sation as discussed in Section 2.2.2.1. The bulk magnetisation has the same response to an RF
2.2. SPIN PHYSICS 29
(a) (b)
Figure 2.4: An ensemble of randomly oriented magnetic dipoles (a) have zero bulk magnetization. When placed in astatic magnetic field (b) the distribution of spin orientation is slightly skewed, creating a bulk magnetisation, m, inthe direction of the static magnetic field.
Figure 2.5: A nuclear magnetic dipole ‘wandering’ through precessional orientations due to the fluctuating magneticfields caused by thermal energy and the molecular environment.
30 CHAPTER 2. THEORY
(a) (b)
Figure 2.6: (a) Bulk magnetisation at thermal equilibrium and (b) the bulk magnetisation after an RF excitation.
excitation as a single spin, rotating around the applied excitation envelope, causing the longitudinal
bulk magnetisation to be rotated into the transverse plane (Figure 2.6). Additionally, the bulk
magnetisation precesses at the Larmor frequency in the same manner as an individual spin.
As the individual spins experience slightly fluctuating magnetic fields, their individual precession
frequencies also vary over time, leading to a dephasing of the transverse bulk magnetisation. We
define the transverse bulk magnetisation in the rotating frame as a complex expression,
m′xy = m′x + im′y. (2.23)
The dephasing of the spins which form the bulk magnetisation leads to a decay of the transverse
signal which is described by a phenomenological model in the rotating frame,
m′xy (t) = m′xy (0) e−tT2 (2.24)
where T2 is the transverse relaxation constant, also referred to as the spin-spin relaxation constant.
When this relationship is expressed in the laboratory frame,
mxy (t) = mxy (0) ei ω0 te−tT2 (2.25)
it is analogous to the precession of a nuclear magnetic dipole (2.8a) with an additional decay term.
The transverse decay (2.24) of the bulk magnetisation can also be expressed as the first order system,
dm′xy (t)
dt= − 1
T2m′xy (t) (2.26)
2.3 Bloch equations
The spin physics in Section 2.2 describe a range of behaviours that can be expressed by a system
of ordinary differential equations. The Bloch equations (Bloch, 1946) are defined in the laboratory
frame as,
d
dtm (t) = γm (t)× b (t)− mxi+myj
T2− (mz −M0
z )
T1k, (2.27)
2.3. BLOCH EQUATIONS 31
where m = [mxmymz]T
is the bulk magnetisation vector, b = [bx by bz]T
is the applied magnetic
field, M0z is the bulk magnetisation at thermal equilibrium, T1 is the longitudinal relaxation constant
and T2 is the transverse relaxation constant.
The Bloch equations (2.27) can be expressed in the rotating frame as,
d
dtm′ (t) = γm′ (t)× b′eff (t)−
m′xi′ +m′yj
′
T2− (m′z −M0
z )
T1k′ (2.28)
in which,
b′eff (t) = b′ (t) +ωrot
γk′ (2.29)
and wherem′ =[m′xm
′ym′z
]Tis the bulk magnetisation vector in the rotating frame, b′ =
[b′x b′y b′z
]Tis the applied magnetic field in the rotating frame. Detail of the transformation into the rotating
frame can be found in Appendix 2.A. The external magnetic field in the rotating frame of reference
is the vector sum of the static and oscillating fields,
b′ (t) = b′0 + b′1 (t) . (2.30)
If we consider the bulk magnetisation in the RF rotating frame, ωrot = ωRF, and substitute the
static field (2.7) and the RF excitation field (2.13) into (2.30), the effective magnetic field (2.29) is,
b′eff (t) = B0k′ +Be
1 (t)(
cos(θRF (t)
)i′ + sin
(θRF (t)
)j′)
+ωRF
γk′
=1
γ
(γBe
1 (t) cos(θRF (t)
)i′ + γBe
1 (t) sin(θRF (t)
)j′ + (ωRF + γB0)k′
)=
1
γ
(− ωe
1 (t) cos(θRF (t)
)i′ − ωe
1 (t) sin(θRF (t)
)j′ + (ωRF − ω0)k′
)=
1
γ
(− ωe
1,x (t) i′ − ωe1,y (t) j′ −∆k′
), (2.31)
where the RF excitation field has been split into orthogonal components,
ωe1,x (t) = ωe
1 (t) cos(θRF (t)
)(2.31a)
ωe1,y (t) = ωe
1 (t) sin(θRF (t)
), (2.31b)
and deviation between an isochromat’s Larmor frequency and the rotating RF frame, is expressed
as off-resonance,
∆ = ω0 − ωRF, (2.32)
which is broken down into individual sources of off-resonance,
∆ = ω0 (1 + δω0)− ωRF (1 + δωrf) , (2.33)
where δω0encompasses chemical shift effects, field inhomogeneities and gradients and δrf is a fre-
quency off-set to the RF transmitter carrier frequency. We can express the Bloch equations in the
RF rotating frame (2.28) in a matrix form by substitution of the effective magnetic field (2.31),
d
dtm′ (t) = Ω (t)m′ (t) +R1m0 (2.34)
32 CHAPTER 2. THEORY
where
Ω(t) =
−R2 −∆ ωe1,y (t)
∆ −R2 −ωe1,x (t)
−ωe1,y (t) ωe
1,x (t) −R1
, (2.34a)
m′ (t) =[m′x m′y m′z
]T, (2.34b)
m0 =[
0 0 M0z
]T, (2.34c)
R1 =1
T1, R2 =
1
T2. (2.34d,e)
The Bloch equations can be solved via numerical integration, but by applying certain assumptions
we can derive analytical solutions to gain more insight into the spin-system response. A secondary
advantage of these solutions is computational efficiency, making simulation of more complex spin-
systems tractable.
2.3.1 Steady-state solutions
Here we consider steady-state solutions of the Bloch equations under continuous wave (CW) exci-
tation. First we discuss an analytical expression of spin response under a constant CW excitation.
This is followed by the description of a method for efficient calculation of spin response under any
periodic excitation where we introduce the Rabi modulated CW excitation envelope and response
as a case study particularly relevant to this thesis.
2.3.1.1 Constant excitation
The spin-system response to constant excitation (Bloch, 1946) was important for historic CW NMR
experiments. We define the constant amplitude RF excitation,
ωe1,x (t) = ω1,x (2.35a)
ωe1,y (t) = 0. (2.35b)
which is applied for an adequate duration, > 5T1, driving the bulk magnetisation into a steady-state
which relates to a rate of change of the bulk magnetisation,
d
dtm′ (t) = 0. (2.36)
Substitution of (2.35) and (2.36) into the Bloch equations (2.34) gives the steady-state solution,
m′x =ω1,x ∆ (T2)
2
1 + (∆T2)2
+ (ω1,x)2T1T2
M0z (2.37a)
m′y =−ω1,x T2
1 + (∆T2)2
+ (ω1,x)2T1T2
M0z (2.37b)
m′z =1 + (∆T2)
2
1 + (∆T2)2
+ (ω1,x)2T1T2
M0z , (2.37c)
the full derivation of which is detailed in Appendix 2.B. The set of possible steady-state magnetisa-
tions vectors lie on an elliptical manifold, as noted by (Abragam, 1961, Chapter 3),
1 =
(m′xrx
)2
+
(m′yry
)2
+
(m′z − czrz
)2
(2.38)
2.3. BLOCH EQUATIONS 33
where
rx = ry =M0
z
2
√T2
T1(2.38a)
rz =M0
z
2(2.38b)
cz =M0
z
2. (2.38c)
The steady-state ellipsoid (2.38) is located in the upper hemisphere of the Bloch sphere as shown in
Figure 2.7b.
Low power excitation case
If we consider a low power RF limiting case,
(ω1,x)2T1T2 1 (2.39)
then the transverse magnetisations (2.37a) and (2.37b) reduce to,
m′x = M0z ω1,x
∆
∆2 +(
1T2
)2
(2.40a)
m′y = −M0z ω1,x
(
1T2
)∆2 +
(1T2
)2
(2.40b)
where (2.40a) and (2.40b) are commonly referred to as the dispersion and absorption line shapes,
shown in Figure 2.7a. The absorption (2.40b) is a negative Lorentzian peak,
L (δ, µ,Γ, a) =a
π
[Γ
(δ − µ)2 + Γ2
](2.41)
where in this case δ is off-resonance, µ is the peak centre, Γ is the half width at half maximum
(HWHM) and a is the peak area. The dispersion and absorption line shapes are also observed when
applying a higher power excitation but have a broader response (Figure 2.8).
2.3.1.2 Periodic excitation
In the majority of excitation cases an analytical solution to the Bloch equations is not available. A
general approach to calculate the spin-system response to periodic excitation waveforms has been
proposed by Tahayori et al. (2015). This technique is referred to as harmonic balancing.
As in Tahayori et al. (2015) we consider the Bloch equations (2.34) under a periodic excitation with
the same period, Tm, for both ωe1,x and ωe
1,y components. We can re-write the transformation matrix,
Ω(t), as
Ω (t) = B +Rx
∞∑n=−∞
β(n)x einωmt +Ry
∞∑n=−∞
β(n)y einωmt (2.42)
34 CHAPTER 2. THEORY
−6 −4 −2 0 2 4 6−0.5
0
0.5
m′ x
∆ (Hz)
−6 −4 −2 0 2 4 6−0.4
−0.2
0m
′ y
∆ (Hz)
−6 −4 −2 0 2 4 6
0.6
0.8
1
m′ z
∆ (Hz)
(a)
(b)
Figure 2.7: Analytical (black line) and numericaly integrated (colored circles) solutions for steady-state magnetisationunder a constant excitation. The excitation was set as, ω1,x = 1.0/
√T1T2, to maximise the absorption peak. (a)
Orthogonal components of the steady-state magnetisation across a range of off-resonance, ∆. (b) Steady-state mag-netisation vector (black line) over a range of off-resonance, ∆. All possible magnetisation vectors are contained by theBloch sphere (light grey) and all steady-state solutions under constant excitation lie on the surface of a steady-stateellipsoid (dark grey).
2.3. BLOCH EQUATIONS 35
−6 −4 −2 0 2 4 6−0.5
0
0.5
m′ x
∆ (Hz)
−6 −4 −2 0 2 4 6−0.1
−0.05
0m
′ y
∆ (Hz)
−6 −4 −2 0 2 4 60
0.5
1
m′ z
∆ (Hz)
(a)
(b)
Figure 2.8: Analytical (black line) and numericaly integrated (colored circles) solutions for steady-state magnetisationunder a constant excitation. The excitation was set as, ω1,x = 10.0/
√T1T2, to demonstrate that the Lorentzian
lineshape also exist under higher power excitation. (a) Orthogonal components of the steady-state magnetisationacross a range of off-resonance, ∆. (b) Steady-state magnetisation vector (black line) over a range of off-resonance,∆. All possible magnetisation vectors are contained by the Bloch sphere (light grey) and all steady-state solutionsunder constant excitation lie on the surface of a steady-state ellipsoid (dark grey).
36 CHAPTER 2. THEORY
where
B =
−R2 −∆ 0
∆ −R2 0
0 0 −R1
, Rx =
0 0 0
0 0 −1
0 1 0
, Ry =
0 0 1
0 0 0
−1 0 0
, (2.42a-c)
and ωm = 2π/Tm is the fundamental frequency, β(n)x , β
(−n)y are the Fourier series coefficients of the
excitation envelopes, ωe1,x, ω
e1,y. As the excitation envelopes are real signals, β(n) = β∗(n), where
β∗(n) is the complex conjugate.
The magnetisation and its rate of change can be expressed as a Fourier expansion,
m =
∞∑l=−∞
c(l)eilωmt (2.43a)
m =
∞∑l=−∞
ilωmc(l)eilωmt (2.43b)
where the Fourier coefficient vector is,
c(l) =[c(l)x c(l)y c(l)z
]T. (2.43c)
Substitution of (2.43) into the Bloch equations (2.34) gives,
∞∑l=−∞
c(l)eilωmt = Ω (t)
∞∑l=−∞
ilωmc(l)eilωmt +R1m0
= B
∞∑l=−∞
c(l)eilωmt +R1m0 · · ·
+Rx
∞∑l=−∞
∞∑n=−∞
c(l)β(n)x ei(l+n)ωmt · · ·
+Ry
∞∑l=−∞
∞∑n=−∞
c(l)β(n)y ei(l+n)ωmt. (2.44)
We apply the Fourier shift theorem to (2.44) and shuffle the order of summation
∞∑l=−∞
(Bc(l) +Rx
∞∑n=−∞
β(n)x c(l−n) +Ry
∞∑n=−∞
β(n)y c(l−n) − ilωmc
(l)
)eilωmt = −R1m0 (2.45)
Equation (2.45) can then be expanded into an infinite matrix form,
PC = Q, (2.46)
2.3. BLOCH EQUATIONS 37
where
P =
. . .
A+ i2ωmI R(β
(1)x , β
(1)y
)R(β
(2)x , β
(2)y
)R(β
(3)x , β
(3)y
)R(β
(4)x , β
(4)y
)R(β
(−1)x , β
(−1)y
)A+ iωmI R
(β
(1)x , β
(1)y
)R(β
(2)x , β
(2)y
)R(β
(3)x , β
(3)y
)R(β
(−2)x , β
(−2)y
)R(β
(−1)x , β
(−1)y
)A R
(β
(1)x , β
(1)y
)R(β
(2)x , β
(2)y
)R(β
(−3)x , β
(−3)y
)R(β
(−2)x , β
(−2)y
)R(β
(−1)x , β
(−1)y
)A− iωmI R
(β
(1)x , β
(1)y
)R(β
(−4)x , β
(−4)y
)R(β
(−3)x , β
(−3)y
)R(β
(−2)x , β
(−2)y
)R(β
(−1)x , β
(−1)y
)A− i2ωmI
. . .
,
(2.46a)
C =[· · · c(−2) c(−1) c(0) c(1) c(2) · · ·
], (2.46b)
Q = [· · · 0 0 −R1m0 0 0 · · · ] , (2.46c)
and we define,
A , B +R(β(0)
x , β(0)y
)(2.46d)
R (βx, βy) , βxRx + βyRy. (2.46e)
If we truncate the coefficients in the infinite matrix, P , we can invert the linear system (2.46) and
solve for the Fourier coefficients of the steady-state response of the spin-system,
C = P−1Q. (2.47)
Rabi modulated continuous wave excitation
Here we introduce the Rabi modulated continuous wave excitation which is a major topic in two ex-
perimental chapters (Chapter 4 and Chapter 5), and a good candidate to demonstrate the harmonic
balancing technique. The Rabi modulated RF envelope is defined as
ωe1 (t) = ω1 (1 + α cosωmt) (2.48)
where ω1 is the average Rabi frequency, ωm is the modulation frequency and α is the modulation
level. When the excitation pattern (2.48) is applied on the i′ axis we have the excitation envelopes,
ωe1,x (t) = ω1 (1 + α cosωmt) (2.49a)
ωe1,y (t) = 0, (2.49b)
which have the related Fourier coefficients,
β(0)x = ω1; β(−1)
x = β(1)x =
αω1
2; β(n)
x = 0, n 6= −1, 0, 1. (2.50)
Substitution of (2.49) into the harmonic balance finite matrix (2.46) gives a tridiagonal matrix,
P rabi =
. . .
A+ i2ωmI D 0 0 0
D A+ iωmI D 0 0
0 D A D 0
0 0 D A− iωmI D
0 0 0 D A− i2ωmI. . .
(2.51)
38 CHAPTER 2. THEORY
0 2 4 6 8 10−0.2
0
0.2
0.4
mobs
x
(norm
alised)
time(ms)
−1000 −500 0 500 1000−0.1
−0.05
0
0.05
0.1
frequency(Hz)
Re(
cobs
x
)
(a)0 2 4 6 8 10−0.2
0
0.2
0.4
mobs
x
(norm
alised)
time(ms)
−1000 −500 0 500 1000−0.1
−0.05
0
0.05
0.1
frequency(Hz)
Re(
cobs
x
)
(b)
Figure 2.9: Spin system response of the water phantom under Rabi modulated excitation. (a) Measured (red circles)and theoretical (black line) periodic steady-state magnetisation waveform for excitation parameters α = 1, ω1 =ωm = 100 Hz and δrf = 180 Hz. (b) Measured (coloured circles) and theoretical (black crosses) harmonics of steady-state magnetisation. DC component (blue), first (green), second (purple), third (red) fourth (orange) and fifth (grey)harmonics.
where,
A =
−R2 −∆ 0
∆ −R2 −ω1
0 ω1 −R1
, D =αω1
2Rx. (2.51a,b)
We can solve for the harmonics of the spin-system response using (2.47) and reconstruct the spin-
system time response via (2.43a) with an experimental example shown in Figure 2.9.
It has been shown that the series coefficients can be approximated by CN ≈(α2
)N ( ω1
ωm
)N/N !,
where N is the number of harmonics considered in the solution (Tahayori et al., 2015). Therefore,
the series solution coefficients always converge to zero. When αω1 < 2ωm the convergence rate is
very fast and a few terms are sufficient to obtain accurate results.
2.3.2 Pulse excitation solutions
The focus of this manuscript is CW excitation but the majority of modern NMR methods use RF
pulses, short relatively high power envelopes to tip the bulk magnetisation into the transverse plane.
In this section we explore the on-resonance response to a pulse excitation envelope and compare the
Fourier transform of a free induction decay signal to the spin-system response under constant CW
excitation.
2.3. BLOCH EQUATIONS 39
Figure 2.10: Free induction decay (FID) signal diagram. An RF pulse tips the bulk magnetisation into the transverseplane where it can be observed by an induced voltage in a coil. The exponential transverse decay is characterised bythe spin-spin relaxation rate, T2, but experimentally we observe a faster decay, T ∗2 , due to field inhomogeneity.
2.3.2.1 Response to RF Pulse
The response of the bulk magnetisation is often expressed as a tip angle away from thermal equilib-
rium,
α =
∫ωe
1 (t) dt. (2.52)
A time efficient excitation pulse shape is a rectangular function,
ωe1 (t) = Π (t, τ) =
−γB1 0 ≤ t ≤ τ0 otherwise
(2.53)
which has a corresponding flip angle (2.52) for an on-resonance spin,
α = −γB1τ. (2.54)
This flip angle expression (2.52) is useful for considering the effect of excitation field strength on
the bulk magnetisation. The flip angle response for off-resonance isochromats is considered in Sec-
tion 2.5.1.
2.3.2.2 Free induction decay
If we consider a case when the bulk-magnetisation has been flipped into the transverse plane, α =
π/2, under no further excitation we observe a free induction decay (FID) signal (Figure 2.10). The
40 CHAPTER 2. THEORY
Fourier transform of a FID signal from a single on-resonance isochromat (2.25) is,
ρxy (ω) = Fmxy (0) ei ω0 te−
tT2
=
∫ ∞0
mxy (0) ei ω0 te−tT2 e−iωt dt
= mxy (0)
∫ ∞0
e
[i(ω0−ω)− 1
T2
]tdt
=mxy (0)
i (ω0 − ω)− 1T2
[e
[i(ω0−ω)− 1
T2
]t
]∞0
=−mxy (0)
i (ω0 − ω)− 1T2
−i (ω0 − ω)− 1T2
−i (ω0 − ω)− 1T2
= mxy (0)
1T2
(ω0 − ω)2
+(
1T2
)2 − i(ω0 − ω)
(ω0 − ω)2
+(
1T2
)2
(2.55)
where ω is the Fourier pair of time. If the thermal equilibrium bulk magnetisation has been rotated
into the transverse plane, mxy (0) = 0− iM0z , then (2.55) becomes,
ρxy (ω) = M0z
(ω0 − ω)
(ω0 − ω)2
+(
1T2
)2 − i1T2
(ω0 − ω)2
+(
1T2
)2
(2.56)
or as transverse components,
ρx (ω) = M0z
(ω0 − ω)
(ω0 − ω)2
+(
1T2
)2
(2.57a)
ρy (ω) = −M0z
1T2
(ω0 − ω)2
+(
1T2
)2
. (2.57b)
The Fourier transform of a free induction decay signal (2.57) gives Lorentzian absorption and disper-
sion lineshapes that are comparable to the constant excitation steady-state response (2.40), shown
again here for clarity,
m′x = M0z ω1,x
(ω0 − ωRF)
(ω0 − ωRF)2
+(
1T2
)2
(2.40a)
m′y = −M0z ω1,x
1T2
(ω0 − ωRF)2
+(
1T2
)2
. (2.40b)
This result is important as it demonstrates that the Fourier transform of a single FID is equivalent
to observing the response to a series of constant power CW excitations with different frequency
offsets.
2.4. MEASUREMENT CONCEPTS 41
2.4 Measurement Concepts
This section covers the experimental measurement of the bulk magnetisation via an induced voltage
in a coil. The effects of field inhomogeneity on the free induction decay signal are considered in
the context of NMR spectroscopy. Echo formation using refocusing pulses and field gradients is
described and applied to volumetric NMR relaxometry.
2.4.1 Signal detection
After excitation by an RF field, the bulk magnetisation generates an oscillating magnetic field which
can be detected via induction with a conductive coil (Bloch et al., 1946). The flux in a coil is
expressed with the reciprocity principle,
Φ (t) =
∫object
br (r) ·m (r, t) dr (2.58)
where r is a spatial vector, br (r) is the magnetic field generated at a point in space by a unit current
in the receive coil and m (r, t) is the bulk magnetisation with spatial dependance in the laboratory
frame.
The induced voltage is proportional to the rate of change of the flux, as per Faraday’s law of
induction,
V (t) = −∂Φ (t)
∂t
= − ∂
∂t
∫object
br (r) ·m (r, t) dr
= − ∂
∂t
∫object
[br,x (r)mx (r, t) + br,y (r)my (r, t) + br,z (r)mz (r, t)
]dr. (2.59)
The transverse components of the bulk magnetisation, mxy, are the primary cause of induced volt-
age as they fluctuate at a much faster rate than the longitudinal components, mz, leading to the
simplification of (2.59),
V (t) = −∫
object
[br,x (r)
∂mx (r, t)
∂t+ br,y (r)
∂my (r, t)
∂t
]dr. (2.60)
Assuming phase sensitive quadrature detection of a FID and ignoring relaxation effects, equation
(2.60) becomes the signal equation,
s (t) =
∫object
b∗r,xy (r) mxy (r, 0) e−i∆(r)t dr (2.61)
where b∗r,xy (r) is the complex conjugate of br,xy (r). If the main static field is spatially and temporally
varying then so is the off-resonance,
∆ (r, t) = ω0 (r, t)− ωRF
= −γB0 (r, t)− ωRF (2.62)
and equation (2.61) is modified to,
s (t) =
∫object
b∗r,xy (r) mxy (r, 0) e−i∫ t0
∆(r,τ)dτ dr. (2.63)
Full derivation of (2.61) and (2.63) is in (Liang et al., 2000, Chapter 3). Here we have defined the
resonance offset with respect to the radio frequency carrier.
42 CHAPTER 2. THEORY
2.4.2 Signal formation
Here we introduce the NMR signal types that are observed via nuclear induction on a modern MRI
scanner. Free induction decays, spin echoes and gradient echoes are discussed. Other signals such
as stimulated echoes and rotary echoes are excluded as they are not used in any of the experimental
chapters in this thesis.
2.4.2.1 Free-induction decay
The FID signal (2.25) is the precessing and exponentially decaying transverse bulk magnetisation
(Figure 2.10). The exponential decay of the bulk magnetisation is characterised by a spin-spin
relaxation rate, T2. Inhomogeneity in the B0 field causes a range of off-resonance frequencies which
causes a dephasing of the signal and an observed FID with faster decay than predicted by the
spin-spin relaxation constant. The two decays are shown in Figure 2.13, where the faster decay is
expressed as,
1
T *2
=1
T2+
1
T ′2(2.64)
where T ′2 is the decay attributed to field inhomogeneity.
2.4.2.2 Spin-echo
In the presence of B0 field inhomogeneity the FID signal can become dominated by T ′2 and decay
much faster than the intrinsic T2 decay. The spin-echo or Hahn echo (Hahn, 1950) is described in
this section as per Haacke et al. (1999) and allows us to recover a signal with T2 weighting in the
presence of B0 field inhomogeneity.
Spin echoes are the basis of many standard imaging methods used in this manuscript to take ref-
erence measurements. The T2 weighted signal of a spin echo is useful for measuring relaxation
maps (Section 3.3.1 and Section 3.3.2). The insensitivity to B0 inhomogeneity is beneficial for the
measurement of B1 maps (Section 3.4.2).
Spin-echo formation
A spin-echo is formed via RF excitation with an excitation pulse and a refocusing pulse (Figure 2.11).
The first pulse is an excitation pulse with angle, α = αxi′, the second is a refocusing pulse with
angle, β = βyj′, with an inter-pulse duration, τ . The phase accumulation of an isochromat with
off-resonance, ∆, is
θ (r, t) = ∆ (r) t. (2.65)
In Figure 2.11, we consider a case where the excitation pulse is applied along the i′ axis causing
the magnetisation to rotate by αx = π2 , aligning it with the j′ axis. During the period, τ , between
excitation and refocusing pulses an isochromat will accumulate a phase of
θ (r, τ -) = ∆ (r) τ (2.66)
where τ - is the timepoint directly before the refocusing pulse. We assume the transverse signal
to be completely dephased before the refocusing pulse is applied along the j′ axis, rotating the
2.4. MEASUREMENT CONCEPTS 43
Figure 2.11: Spin-echo sequence diagram for excitation angle, αx = π2
, and refocusing angle, βy = π. The responsefor three isochromats at on-resonance (black), positive off-resonance (red) and negative off-resonance (blue) is shown.Isochromats (a) at equilibrium (b) are excited into the transverse plane (c) where they dephase and (d) have theirphase flipped and (e) rephase to form an echo.
dephased transverse magnetisation by βy = π. The refocusing pulse effectively flips the phase of
each isochromat
θ(r, τ+
)= −θ (r, τ -) (2.67)
where τ+ is the timepoint directly after the refocusing point. In the period after the refocusing
pulse, the isochromat will accumulate phase
θ (r, t− τ) = θ(r, τ+
)+ ∆ (r) (t− τ)
= −∆ (r) τ + ∆ (r) (t− τ)
= ∆ (r) (t− 2τ) . (2.68)
It is apparent from (2.68) that all isochromats will refocus to zero phase at time t = 2τ which is
often referred to as the spin echo time,
TE = 2τ, (2.69)
and corresponds to the peak magnitude of the echo.
Spin-echo envelope
The envelope of the transverse magnetisation during a spin-echo can be expressed as,
d
dt|mxy| = −RSE
2 (t) |mxy| , (2.70)
44 CHAPTER 2. THEORY
where the effective decay rate is,
RSE2 (t) =
R′2 +R2 0 < t < τ
−R′2 +R2 τ < t < 2τ = TE
R′2 +R2 t > 2τ = TE
. (2.70a)
The solution of (2.70) is,
|mxy| (t) = |mxy| (0)
e− tT∗2 0 < t < τ
e−tT2 e− (TE−t)
T ′2 τ < t < 2τ = TE
e−tT2 e− (t−TE)
T ′2 = e− tT∗2 e
TET ′2 t > 2τ = TE
, (2.71)
and shows that a spin-echo signal is T2 weighted at the echo time, t = TE,
|mxy| (TE) = |mxy| (0) e−TET2 . (2.72)
Spin-echo diffusion effects
The preceding analysis of a spin-echo assumes a spatially varying and time invariant off-resonance,
∆ (r). If we consider diffusion, when an isochromat may wander through space during the echo
period, 2τ , the off-resonance becomes time dependant, ∆ (r, t). This leads to imperfect rephasing
and a reduction of the expected echo magnitude (2.72). Diffusion theory, methods and applications
are covered in detail in Jones (2010).
2.4.2.3 Gradient Echo
A gradient recalled echo is formed via switching the polarity of the linear gradient fields (Fig-
ure 2.12). Gradient echoes are used in fast imaging methods which can be used for proton density
mapping (Section 3.2) and are sensitive to off-resonance effects allowing the acquisition of B0 maps
(Section 3.4.1).
Consider the phase accumulation of off-resonance isochromats (2.65) under a linear gradient in the
x direction,
θ (r, t) = ∆ (r) t
=(ω0 [1 + δω0
(r)] + γ Gx− ωRF
)t
=(
∆B0(r) + γ Gx
)t, (2.73)
where G is the gradient field strength and ∆B0 (r) encompasses remaining sources of off-resonance.
In Figure 2.12, we consider a case where an excitation pulse has caused the magnetisation to rotate
by α = π2 into the transverse plane. For simplicity we ignore any dephasing in the period between
the excitation pulse and the gradient and consider t = 0 to be at the start of the positive gradient
lobe. At the end of the positive gradient lobe an isochromat will have accumulated a phase of
θ (r, τ -) =(
∆B0(r) + γ Gx
)τ. (2.74)
2.4. MEASUREMENT CONCEPTS 45
The phase accumulation during the negative gradient is expressed as,
θ (r, t− τ) = θ (r, τ -) +(
∆B0 (r)− γ Gx)
(t− τ)
=(
∆B0 (r) + γ Gx)τ +
(∆B0 (r)− γ Gx
)(t− τ)
= ∆B0 (r) t+ γ Gx (2τ − t) . (2.75)
It is apparent from (2.75) that isochromats will rephase at time, t = 2τ , to form a gradient echo
peak. At the gradient echo time, TE = 2τ , there is a remaining loss of phase coherence,
θ (r, TE) = TE ∆B0(r) (2.76)
causing a loss of intensity and giving the gradient echo a T ∗2 weighting.
46
CHAPTER
2.THEORY
Figure 2.12: Gradient-echo sequence diagram for excitation angle for three groups of isochromats at spatial locations (circles) r, (arrows) 0 and (diamonds) −r. Eachof the three spatial location has the same main field distribution represented by (blue) a slightly slower off-resonance (black) an on-resonance and (red) a slightly fasteroff-resonance. All isochromat groups start (a) at equilibrium (b) are excited into the transverse plane (c) where they dephase due to main field inhomogeneities then (d)dephase due to main field inhomogeneities and the applied gradient (e) partially rephase due to a reversed gradient to form an echo. The loss of phase coherence due tomain field inhomogeneity remains at echo time (e) giving the gradient echo a T ∗2 weighting
2.4. MEASUREMENT CONCEPTS 47
2.4.3 NMR spectroscopy
We have seen in previous sections that the precession frequency of a spin is proportional to the
strength of the magnetic field. A circulating electron cloud generates a local magnetic field contri-
bution which can shield or de-shield a nucleus from the B0 field. This perturbation of the magnetic
field alters the precession frequency of spin, creating an offset to the Larmor frequency known as
chemical shift. This relationship between local chemical environment and frequency allows us to
probe the chemical composition of a sample via NMR spectroscopy.
To allow spectral comparison of data across magnetic field strengths, it is common to express chem-
ical shift in parts-per-million (ppm) with respect to a reference compound,
δcs =ωiso − ωref
ωref, (2.77)
where ωiso is the isochromat frequency and ωref is the reference frequency. Chemical shift (2.77) can
be expressed in the Bloch equations as a source of off-resonance (2.33),
∆ = ω0 (1 + δω0 + δcs)− ωRF (1 + δωrf) , (2.78)
where here we set the reference frequency as the Larmor frequency, ωref = ω0.
Historically, NMR spectra were acquired under continuous wave excitation (Arnold et al., 1951),
slowly sweeping the frequency through the spectral range to collect peaks of the spectrum. Modern
techniques use a broadband excitation pulse and the Fourier transform to obtain a spectrum (Ernst
and Anderson, 1966).
2.4.3.1 Continuous Wave NMR spectroscopy
The first CW-NMR spectrum of ethanol (Arnold et al., 1951) demonstrated the ability of NMR to
distinguish chemical groups of methyl, methylene and hydroxyl by their chemical shift and relative
areas. In this type of spectroscopy experiment a CW excitation is swept across the frequency range,
slow enough that the transverse response of the spin-system can be modelled by the steady-state
solution to the Bloch equations (2.40). A limitation of these early methods was a slow spectral
sweep rate to avoid artifacts from previously excited spins. These efficiency issues were addressed
by later improvements to CW spectroscopy (Gupta et al., 1974; Dadok and Sprecher, 1974).
2.4.3.2 Pulsed Fourier Transform NMR spectroscopy
Fourier transform NMR (FT-NMR) using pulsed excitation (Ernst and Anderson, 1966) provided
a large efficiency and sensitivity improvement on the CW-NMR methods. The spectral range is
excited with a single short duration hard RF pulse and the following FID is Fourier transformed
to reconstruct the spectrum as shown in Figure 2.13. This method of acquiring NMR spectra is
equivalent to a slow passage CW-NMR spectra as discussed in Section 2.3.2.2.
Static field inhomogeneity
Inhomogeneity of the B0 field and spin-spin interactions lead to spectral line broadening in both
CW-NMR and FT-NMR techniques. If we assume the field inhomogeneity is Lorentzian (2.41) it
can be related to the FID relaxation rate,
1
T *2
=1
T2+ γ ΓB0
(2.79)
48 CHAPTER 2. THEORY
(a) (b)
Figure 2.13: Free induction decay relationship to spectral width. (a) A basic FID experiment showing the envelopefor T2 decay in a homogeneous B0 field (dashed line) and the envelope for T ∗2 decay in an inhomogeneous B0 field(solid line). (b) The spectral width of a Fourier transform of an FID in a homogeneous B0 field (dashed line) and inan inhomogeneous B0 field (solid line).
where ΓB0is the HWHM of the Lorentzian distribution (Abragam, 1961). This relationship (2.79)
describes the importance of shimming the B0 field to achieve a narrow line width when performing
NMR spectroscopy (Figure 2.13b).
2.4.4 NMR relaxometry
The NMR signal is effected by spin-lattice and spin-spin relaxation effects (Section 2.2.2.1 and
Section 2.2.2.2). Here we present common methods for acquiring an appropriately weighted volume
signal and estimating the mean T1 and T2 relaxation constants. The techniques described here are
the basis for the relaxation mapping methods employed in Section 3.3.
2.4.4.1 Spin-lattice relaxometry
T1 relaxation is the exponential decay of the bulk-magnetisation towards thermal equilibrium (Sec-
tion 2.2.2.1). Spin-lattice relaxation rates can be measured using inversion recovery, saturation
recovery and progressive saturation techniques.
Inversion recovery
Vold et al. (1968) measured the spin-lattice relaxation of NMR spectra using an inversion preparation
pulse before a standard FID sequence. Inversion recovery applies a π excitation pulse to invert the
equilibrium magnetisation (Figure 2.14) which then recovers over a period, TIR, before a π2 pulse
excites the magnetisation into the transverse plane for acquisition of an FID. The experiment is
repeated for a range of inversion periods, TIR, and the relaxation rate, T1, is estimated by fitting
the measured decay curve to the expected signal from (2.21),
|mxy| (t) =∣∣∣M0
z
(1− 2 e−
TIRT1
)∣∣∣ e− (t−TIR)T∗2 . (2.80)
2.4. MEASUREMENT CONCEPTS 49
Figure 2.14: Inversion recovery sequence diagram. Spin-lattice relaxation rates can be estimated from a series of FIDmeasurement over a range of inversion recovery periods, TIR.
As the inverted magnetisation recovers toward equilibrium is crosses the origin, causing a signal null
with a related inversion time, T nullIR , which can also be used to estimate the relaxation rate,
T1 =T null
IR
ln 2. (2.81)
Additionally, this signal null can be used to suppress signal from a spin population with a known
relaxation rate.
Saturation recovery
Saturation recovery measurements are similar to inversion recovery but with a magnetisation prepa-
ration to saturate the spin-system. The magnetisation can be reliably nulled by a series of π pulses
(Markley et al., 1971) or more efficiently by rotating the magnetisation into the transverse plane
and dephasing with a gradient field (McDonald and Leigh, 1973). Similar to inversion recovery, the
experiment is repeated with a range of saturation recovery times, TSR, and the relaxation rate, T1,
is estimated by fitting the measured decay curve to the expected signal from (2.21),
|mxy| (t) = M0z
(1− e−
TSRT1
)e− (t−TSR)
T∗2 . (2.82)
Progressive saturation
Freeman and Hill (1971) measured T1 relaxation by placing the spin-system in a spin-lattice depen-
dant steady-state using a series of π2 pulses. When the excitation pulses are applied faster than the
longitudinal magnetisation can recover (Figure 2.15) the steady-state longitudinal magnetisation is
dependant on the relaxation rate and the repetition rate,
mz (TR-) = M0
z
(1− e−
TRT1
)(2.83)
50 CHAPTER 2. THEORY
Figure 2.15: Progressive saturation sequence diagram. Spin-lattice relaxation rates can be estimated from a series ofpartially saturated FID measurements over a range of repetition times, TR.
giving a measured steady-state transverse signal
|mxy| (t) = M0z
(1− e−
TRT1
)e− tT∗2 . (2.84)
Similar to inversion recovery, the experiment is repeated with a range of repetition times, TR, and the
relaxation rate, T1, is estimated by fitting the measured decay curve to the expected signal (2.84). A
series of excitation pulses will generate spin and stimulated echoes (Hahn, 1950), which in this case
are undesirable and can be suppressed by using a changing field gradient to dephase the transverse
magnetisation on each repetition.
2.4.4.2 Spin-spin relaxometry
T2 relaxation results in the decay of the transverse magnetisation and is often characterised using a
spin-echo (SE) or a multi-spin-echo (MSE) experiments.
Relaxation measurement
Hahn (1950) measured spin-spin relaxation rates by acquiring data from a series of SE experiments
with varied echo times to generate an exponentialy decaying curve. The measurements were then
matched with a decay curve (2.72). This method has limited efficiency due to a long repetition time,
TR ≥ 5T1, and has estimation error due to diffusion effects. Carr and Purcell (1954) demonstrated
that a T2 weighted echo-train is formed by multiple refocusing pulses (Figure 2.16). The MSE
method can measure a T2 decay curve in a single repetition and is less susceptible to diffusion effects
due to shorter echo times. Using coherent pulses and a π/2 phase shift between the excitation and
refocusing pulses compensates for cumulative error from imperfect refocusing pulses (Meiboom and
Gill, 1958).
2.5. IMAGING CONCEPTS 51
Figure 2.16: Carr-Purcell-Meiboom-Gill (CPMG) spin-echo sequence diagram. The spin-echo train is established withan excitation pulse, αx = π
2, and a series of coherent refocusing pulses, βy = π.
2.5 Imaging Concepts
This brief history of early imaging techniques is summarised from a general history of MR imag-
ing (Edelman, 2014). The first MR images were acquired under a gradient field and were formed by
back-projecting projections of proton density (Lauterbur et al., 1973). The first human MR image
was of a finger (Mansfield and Maudsley, 1977) and was taken using a line scanning method (Mans-
field et al., 1976) where a line volume is excited then frequency encoded during readout. The first
body MR image was acquired using the field focused nuclear magnetic resonance (FONAR) (Dama-
dian et al., 1977) method which measured an image by moving a small signal volume, created with
a small homogeneous field region, through space. Two dimensional imaging in reasonable time was
possible with the Fourier transform based spin warp imaging (Edelstein et al., 1980) which forms
the basis of the majority of modern MR imaging methods. The efficiency of imaging experiments
was greatly improved by the development of echo-planar imaging (EPI) (Mansfield, 1977), fast low
angle shot imaging (FLASH) (Haase et al., 1986) and rapid acquisition with relaxation enhancement
(RARE) (Hennig et al., 1986).
This section describes image formation from NMR signals with spatial frequency dependence due to
the application of magnetic field gradients. We introduce slice selective pulses, the k-space formalism,
phase encoding and frequency encoding. A spin-echo imaging sequence and a gradient echo imaging
sequence are used as examples to discuss fundamental Cartesian k-space sampling strategies.
2.5.1 Slice selection
Imaging an entire three dimensional volume is time consuming and often a reduced set of two
dimensional slices provide adequate information for clinical applications. A slice of space can be
selectively excited by applying a gradient during RF excitation. The bandwidth and frequency
response of the RF excitation are related to the thickness and spatial profile of the excited slice.
We explore slice selection using the Bloch equations (2.34) and apply an RF excitation, ωe1,x =
−γBe1 (t), a slice selection gradient in the z direction, ∆ = −γGssz, and assume a short pulse
duration to minimise relaxation effects,
d
dtm′ (t) = Ωss (t)m′ (t) (2.85)
52 CHAPTER 2. THEORY
where
Ωss(t) =
0 γGssz 0
−γGssz 0 γBe1 (t)
0 −γBe1 (t) 0
. (2.85a)
Assuming a small tip angle m′z = M0z and d
dtm′z = 0 simplifies (2.85) to,
dm′xdt
= γGsszm′y (2.86a)
dm′ydt
= −γGsszm′x + γBe
1 (t)M0z (2.86b)
dm′zdt
= 0. (2.86c)
The transverse components of (2.86) are expressed as a dynamic complex signal
dm′xy (z, t)
dt= −iγGsszm
′xy (z, t) + iγBe
1 (t)M0z (z) , (2.87)
which can be solved using a Laplace transform (Liang et al., 2000) assuming there is no initial trans-
verse magnetisation and the excitation envelope, Be1 (t), is symetric about half the pulse duration
t = τp/2, yielding an expression for the slice excitation profile in the z direction
im′xy (z, τp)
γM0z (z)
eiγGsszτp/2 =
∫ τp/2
−τp/2Be
1
(t+
τp2
)ei2πksst dt (2.88)
where the slice selection frequency is
kss =γ
2πGssz. (2.88a)
2.5.1.1 Slice profile
From the small tip angle approximation result we can state the slice profile (2.88) is proportional to
the inverse Fourier transform of the excitation envelope,∣∣m′xy (z, τp)∣∣ ∝ F -1
Be
1
(t+
τp2
)(2.89)
when relaxation effects are negligible. This relationship has been shown in experiment to be a
reasonable approximation under larger tip angles (Pauly et al., 1989b,c). To design a RF pulse for
a target slice profile we invert the relationship (2.89),
Be1
(t+
τp2
)∝ F
∣∣m′xy (z (kss) , τp)∣∣ . (2.90)
We define an ideal slice profile as a rectangular function,
∣∣m′xy (z (kss) , τp)∣∣ = Π (z,∆z) =
1 |z| > ∆z/2
0 otherwise(2.91)
where ∆z is the slice width in the z direction. Using relationship (2.90) and the rectangular slice
profile (2.91) the slice selective excitation envelope is,
Be1 (t) ∝ sinc
[π∆kss
(t− τp
2
)]. (2.92)
2.5. IMAGING CONCEPTS 53
Figure 2.17: The (top right) slice profile is related to shape and duration of the (top left) excitation pulse shape. Thestrength of the gradient field, Gss, affects the spatial width of the slice, ∆z.
The RF pulse envelope has a finite duration, and is commonly truncated to a fixed number of lobes,
nl, which gives a relationship between pulse duration, gradient strength and slice thickness,
τp =2nl
∆kss=
4πnl
γGss∆z, (2.93)
shown in Figure 2.17.
2.5.1.2 Slice rephasing
The transverse magnetisation excited by a slice selective excitation (2.88) suffers from a phase
distortion,
θss (τp, z) =γGsszτp
2. (2.94)
As the spatial phase variation is linear it is possible to rephase the slice using a rephasing gradi-
ent. The area of the rephasing gradient lobe must equal half the slice selection lobe as shown in
Figure 2.18.
2.5.2 Spatial encoding
Spatial encoding is achieved with magnetic gradient fields which vary over space, generally an MRI
scanner will have hardware to generate linear gradients in three orthogonal direction. The gradient
54 CHAPTER 2. THEORY
Figure 2.18: A slice selective excitation causes a phase distortion across the slice which can be corrected with arefocusing gradient lobe.
fields are commonly generated by water cooled electro-magnets at a field strength much smaller in
magnitude than the main magnetic field. The gradient fields introduce an audio frequency fluctuation
which is superimposed with the radio-frequency precession due to the B0 field.
2.5.2.1 Spatial frequency k-space
Here we introduce k-space (Brown et al., 1982; Ljunggren, 1983; Twieg, 1983) and for simplicity
consider a one dimensional case. The signal equation (2.63) directly following a π/2 pulse can be
re-written,
s (t) =
∫object
ρ (r) e−i∫ t0
∆(r,τ)dτ dr, (2.95)
where the effictive spin density is,
ρ (r) = b∗r, xy (r)1
4ρ0 (r)
γ22
KTsB0. (2.95a)
When considering a one dimensional case in the z direction then signal (2.95) is,
s (t) =
∫object
ρ (z) e−i∫ t0
∆(z,τ)dτ dz (2.96)
where the projection of spin density in the z direction is,
ρ (z) =
∫ρ (r) dx dy, (2.96a)
and assuming no B0 field inhomogeneity, the off-resonance (2.33) from a linear gradient is,
∆ (z, t) = γGz (t) z (2.96b)
where Gz is the rate of change of the z gradient and is often expressed in units mT/m. The linear
spatial dependence of the gradient allows us to express the signal equation in k-space
s (kz) =
∫object
ρ (z) e−i2πkzz dz (2.97)
2.5. IMAGING CONCEPTS 55
where spatial frequency is defined as
kz =γ
2π
∫ t
0
Gz (τ) dτ. (2.97a)
Signal equation (2.97) makes it clear that our observed NMR signal under a linear gradient is a
natural Fourier transform of the pseudo spin density,
s (kz) = F ρ (z) . (2.98)
Furthermore, it is possible to reconstruct the pseudo spin density from the observed signal using an
inverse Fourier transform,
ρ (z) =
∫object
s (kz) ei2πkzz dk (2.99)
= F -1 s (kz) . (2.100)
General k-space
The one dimensional result can be extended to two and three dimensional cases (Liang et al., 2000,
Chapter 5),
s (k) =
∫object
ρ (r) e−i2πk·r dr (2.101)
where spatial frequency is defined as
k =γ
2π
∫ t
0
G (τ) dτ (2.101a)
and
k = [kx ky kz]T
(2.101b)
G = [GxGyGz]T. (2.101c)
A proton density volume is reconstructed by inverse Fourier transforming an adequately sampled
k-space,
ρ (r) =
∫object
s (k) ei2πk·r dk
= F -1 s (k) . (2.102)
2.5.2.2 Phase encoding
A gradient is applied for a duration Tpe to phase encode the NMR signal (2.101),
k =γ
2π
∫ Tpe
0
Gpe (τ) dτ. (2.103)
If the gradients are constant during the phase encode duration then,
k =γ
2πGpeTpe. (2.104)
Phase encoding moves the signal to a new k-space location as shown in Figure 2.19 and Figure 2.20.
Only the net area under the phase encoding gradients is important which gives some flexibility when
designing gradient envelopes.
56 CHAPTER 2. THEORY
2.5.2.3 Frequency encoding
A gradient is applied during acquisition to frequency encode the NMR signal (2.101),
k (t) =γ
2π
∫ t
0
Gfe (τ) dτ. (2.105)
If the gradients are constant during signal acquisition then,
k (t) =γ
2πGfe t. (2.106)
Frequency encoding moves the signal through a k-space trajectory during the acquisition period as
shown in Figure 2.19 and Figure 2.20. More complex gradient envelopes can be used to sample
k-space with non-Cartesian trajectories.
2.5.
IMAGIN
GCONCEPTS
57
Figure 2.19: Multiple spin-echo based k-space trajectory diagram. A slice selective pulse is followed by (1) a phase encoding gradient during the slice refocusing lobe. (2)The refocusing pulse causes a phase inversion in the complex plane. (3) A phase encoding gradient shift the starting point before (4) a frequency encoding gradient isapplied during the acquisition of a spin-echo and (5) a phase encoding gradient shifts the k-space position in preparation for the next echo formation and acquisition.
58
CHAPTER
2.THEORY
Figure 2.20: A gradient-echo based k-space trajectory diagram. A slice selective pulse is followed by (1) a phase encoding gradient during the slice refocusing lobe and (2)a frequency encoding gradient during the acquisition of a spin-echo. The phase encoding gradient is changed to acquire a carestian k-space lines to fill the two dimensionk-space.
2.A. BLOCH EQUATIONS IN THE ROTATING FRAME 59
Appendix
2.A Bloch equations in the rotating frame
We derive an expression for the dynamics of the bulk magnetisation as observed in the rotating
frame. The bulk magnetisation in the laboratory frame is previously defined as
m = mxi+myj +mzk (2.A.1)
and in the rotating frame as
m′ = m′xi′ +m′yj
′ +m′zk′. (2.A.2)
The dynamics of a general rotating frame of reference basis, i′, j′,k′, is
di′
dt= ω × i′ (2.A.3a)
dj′
dt= ω × j′ (2.A.3b)
dk′
dt= ω × k′. (2.A.3c)
If we consider a transverse rotation, ω = ωrotk, then (2.A.3) can be expressed as a transform from
the the laboratory basis to the rotating basis,i′j′k′
=
cos(ωrott) − sin(ωrott) 0
sin(ωrott) cos(ωrott) 0
0 0 1
ijk
(2.A.4)
= Rz,ccw (ωrot, t)
ijk
, (2.A.5)
where Rz, ccw is a counter-clockwise rotation in the transverse plane.
The time derivative of the bulk magnetisation observed from the laboratory frame is,
dm
dt=dmx
dti+
dmy
dtj +
dmz
dtk. (2.A.6)
The time derivative of the rotating bulk magnetisation as observed in the rotating frame is,(dm′
dt
)R
=d
dt
(m′x (t) i′ +m′y (t) j′ +m′z (t)k′
)=dm′xdti′ +
dm′ydtj′ +
dm′zdtk′ (2.A.7)
and observed from the laboratory frame,
dm′
dt=
d
dt
(m′x (t) i′ (t) +m′y (t) j′ (t) +m′z (t)k′ (t)
)=dm′xdti′ +
dm′ydtj′ +
dm′zdtk′ +m′x
di′
dt+m′y
dj′
dt+m′z
dk′
dt. (2.A.8)
60 CHAPTER 2. THEORY
Substitution of (2.A.3) and (2.A.7) into (2.A.8) gives
dm′
dt=
(dm′
dt
)R
+m′xω × i′ +m′yω × j
′ +m′zω × k′
=
(dm′
dt
)R
+ ω ×(m′xi
′ +m′yj′ +m′zk
′)
=
(dm′
dt
)R
+ ω ×m′. (2.A.9)
Considering these time derivatives we can state,
dm
dt=dm′
dt6=(dm′
dt
)R
(2.A.10)
Substitution of (2.27) into (2.A.9) gives us an expression for the dynamics of the bulk magnetisation
as observed in the rotating frame,(dm′
dt
)R
=dm′
dt− ω ×m′
= γm′ × b′ − ω ×m′ −m′xi
′ +m′yj′
T2− (m′z −M0
z )
T1k′
= γ
(m′ × b′ +m′ × ω
γ
)−m′xi
′ +m′yj′
T2− (m′z −M0
z )
T1k′
= γm′ ×(b′ +
ω
γ
)−m′xi
′ +m′yj′
T2− (m′z −M0
z )
T1k′
= γm′ × b′eff −m′xi
′ +m′yj′
T2− (m′z −M0
z )
T1k′ (2.A.11)
where
b′eff = b′ +ω
γ
= b′ +ωrot
γk′. (2.A.11a)
2.B Bloch equations steady-state solution to constant exci-
tation
We derive the steady-state spin-system response to a constant amplitude and constant frequency
RF excitation,
ωe1,x (t) = ω1,x (2.B.1a)
ωe1,y (t) = 0 (2.B.1b)
which is applied for a long duration. When the magnetisation reaches a steady-state, the rate of
change of the bulk magnetisation is,
d
dtm′ (t) = 0. (2.B.2)
2.B. BLOCH EQUATIONS STEADY-STATE SOLUTION TO CONSTANT EXCITATION 61
Applying these conditions (2.B.1) and (2.B.2) to the Bloch equations in the rotating frame (2.34),0
0
0
=
−R2 −∆ 0
∆ −R2 −ω1,x
0 ω1,x −R1
m′xm′ym′z
+R1
0
0
M0z
(2.B.3)
or,
0 = −R2m′x −∆m′y (2.B.4a)
0 = ∆m′x −R2m′y − ω1,xm
′z (2.B.4b)
0 = ω1,xm′y −R1m
′z +R1M
0z . (2.B.4c)
Arrange (2.B.4a) and (2.B.4c) to be functions of m′y,
m′x = − ∆
R2m′y (2.B.5a)
m′z =ω1,x
R1m′y +M0
z (2.B.5c)
Substitute (2.B.5a) and (2.B.5c) into (2.B.4b)
0 = −∆2
R2m′y −R2m
′y −
ω21,x
R1m′y − ω1,xM
0z (2.B.6)
m′y = −ω1,xM0z
(ω2
1,x
R1+
∆2
R2+R2
)−1
= −ω1,xM0z
R1R2
R1R22 +R1∆2 +R2ω2
1,x
=−ω1,xT2
1 + (T2∆)2
+ T1T2ω21,x
M0z (2.B.7)
gives an expression for the rotating y component as a function of excitation amplitude, off-resonance,
relaxation constants and the magnetisation at thermal equalibrium. Substitution of (2.B.7) into
(2.B.5a) yeilds the rotating x component, giving us transverse components
m′x =ω1,x∆T 2
2
1 + (T2∆)2
+ T1T2ω21,x
M0z (2.B.8a)
m′y =−ω1,xT2
1 + (T2∆)2
+ T1T2ω21,x
M0z (2.B.8b)
m′z =1 + (∆T2)
2
1 + (T2∆)2
+ T1T2ω21,x
M0z (2.B.8c)
62 CHAPTER 2. THEORY
Chapter 3
Methods
Contents3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
3.2 Proton Density Imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
3.3 Relaxation Mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
3.3.1 RARE-VTR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
3.3.2 MSME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
3.4 Magnetic Field Mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
3.4.1 B0 mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
3.4.2 B1 mapping . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
3.1 Introduction
This chapter builds on the imaging concepts introduced in Chapter 2 and describes the imaging
methods used to support the developement of novel methods detailed in Chapters 4-6. The pre-
sented imaging methods are used to acquire reference data or to measure parameters for use in
reconstruction models. A fast low shot imaging (FLASH) sequence (Haase et al., 1986) is used to
acquire proton density maps. Relaxation mapping is performed using rapid acquisition with relax-
ation enhancement and variable repetition time (RARE-VTR) and multi-slice multi-echo (MSME)
methods. B0 field maps are measured with a multiple gradient-echo method (Kanayamay et al.,
1996) using the Bruker FieldMap sequence. B1 field maps are acquired using a standard spin-echo
sequence, using a double angle method (Wang et al., 2005). The imaging parameters specific to each
experiment can be found in the Methods section of each chapter.
3.2 Proton Density Imaging
Fast low angle shot imaging (FLASH) (Haase et al., 1986) is an efficient imaging protocol. A basic
FLASH sequence diagram is shown in Figure 3.1. The method uses a low flip angle slice-selective
excitation and measures a phase and frequency encoded gradient echo. These properties allow for a
63
64 CHAPTER 3. METHODS
Figure 3.1: Fast low angle shot imaging (FLASH) sequence diagram.
short repetition time, TR, and the efficiency can be further improved by the use of spoiler or rewinder
gradients. The dynamic steady-state gradient echo intensity can be expressed as (Liang et al., 2000)
|mxy (TE)| = M0z e−TE/T
∗2 sinα
1− e−TR/T1
1− e−TR/T1 cosα. (3.1)
Image contrast can be optimised for a given application by selection of flip angle, α, echo time,
TE and repetition time, TR. A proton density image is measured using FLASH in Chapter 5 as a
reference image to compare the performance to an image reconstructed the steady-state response
under Rabi modulated excitation. The proton density image is acquired by using a short echo time,
TE, to reduce the T ∗2 weighting and a small flip angle, α, to reduce the T1 weighting.
3.3 Relaxation Mapping
Relaxation measurements were acquired in all experiments, with both relaxation constants acquired
using spin-echo based methods. In the Rabi modulated spectroscopy (Chapter 4) and Rabi mod-
ulated imaging experiments (Chapter 5) the measured relaxation rates were used to calculate the
expected off-resonance response and to construct the forward models used to reconstruct spectra and
radial projections of proton density. In Chapter 6, measured relaxation rates were used to calculate
a ring-lock excitation envelope to control the bulk magnetisation on a steady-state manifold, and in
a second experiment to assess the accuracy of a ring-lock relaxometry technique.
The signal magnitude in a multiple spin-echo experiment is weighted by T1 and T2 which allows the
estimation of relaxation constants from measurements with appropriately selected repetition times,
TR, and effective echo times, kTE. A multiple echo spin-echo experiment with an initial excitation
angle, α, followed by Nse refocusing pulses with a fixed refocusing angle, π, has a signal intensity at
3.3. RELAXATION MAPPING 65
Figure 3.2: RARE sequence diagram. Multiple spin-echoes are acquired to fill multiple lines of k-space in a singlerepetition time. RARE has a higher efficiency in comparison to a standard spin-echo sequence which acquires a singleline of k-space per repetition.
the kth echo (Mills et al., 1986),
|mxy (kTE)| = M0z sinα
[1 + (−1)
Nse+1e−TR/T1 − ase
] e−kTE/T2
bse, (3.2)
where
ase =
Nse∑n=1
(−1)Nse+n
2e−(TR−(2n−1)τ)/T1 (3.2a)
bse = 1 + (−1)Nse+1
e−TR/T1 cosα. (3.2b)
If the excitation angle is α = π/2, the echo intensity becomes dependant on the sequence parameters,
TR and TE ,
|mxy (kTE)| = M0z
[1 + (−1)
Nse+1e−TR/T1 − ase
]e−kTE/T2 , (3.3)
where,
ase =
Nse∑n=1
(−1)Nse+n
2e−(TR−(2n−1)τ)/T1 (3.3a)
This expression (3.3) is used to select sequence parameters and describe the signals acquired by the
RARE-VTR and MSME methods.
3.3.1 RARE-VTR
Rapid acquisition with relaxation enhancement (RARE) (Hennig et al., 1986) improves the efficiency
of a basic spin-echo sequence by acquiring multiple spin-echoes to fill multiple lines of k-space in a
66 CHAPTER 3. METHODS
Figure 3.3: RARE sequence diagram for T2 relaxometry. To estimate T2, multiple k-spaces are acquired over a rangeof effective echo times. This diagram shows a RARE factor = 2, where during each repetition two k-space lines areacquired for each effective echo time.
3.3. RELAXATION MAPPING 67
single repetition time (Figure 3.2). The number of k-space lines acquired per repetition, which is
also the scan acceleration factor, is referred to as the RARE factor. This acquisition strategy gives
a non-uniform T2 weighting across k-space which is related to the point spread function (PSF) in
the image domain. Lower RARE factors should be used measuring samples with fast T2 decay to
shorten the effective echo time and avoid image artifacts.
Spin-lattice relaxation, T1, maps can be measured by repeating a RARE experiment with a variable
repetition time (RARE-VTR). The spin-spin weighting of the echoes is minimised by using a short
echo-time, TE, reducing equation (3.3) to,
|mxy (TR)| = M0z
(1− e−
TRT1
), (3.4)
which is a similar result to progressive saturation in Section 2.4.4.1.
Spin-spin relaxation, T2, maps can be measured from a RARE-VTR experiment by using the spin-
echoes to fill multiple k-spaces at different effective echo times rather than multiple lines of the same
k-space (Figure 3.3). Details of the T2 fitting are the same as for MSME discussed in Section 3.3.2.
3.3.2 MSME
The multi-slice multi-echo (MSME) method can be used for efficient T2 mapping. Spin-spin relax-
ation, T2, maps can be measured with a multiple spin-echo experiment. The spin-lattice weighting
of the echoes is minimised by using a long repetition time, TR, reducing equation (3.3) to,
|mxy (kTE)| = M0z e− kTET2 (3.5)
which is similar to 2.72 and can be used to estimate T2 as discussed in Section 2.4.4.2.
The necessity for a long repetition time, TR, causes a long experimental time, where the majority of
time is waiting for the magnetisation to return to thermal equilibrium. To improve efficiency, multiple
slices can be interleaved (Figure 3.4), acquiring data from a single slice while the magnetisation of
other slices return to equilibrium.
68
CHAPTER
3.METHODS
Figure 3.4: Multi-slice multi-echo (MSME) sequence diagram for efficient multi-slice T2 relaxometry. A long repetition is required to reduce T1 weighting of the spin-echoeswhich causes a long experiment time in a single slice experiment. MSME reduces this inefficiency but acquiring data from other spatial locations as previously excitedslices return to equilibrium.
3.4. MAGNETIC FIELD MAPPING 69
Figure 3.5: B0 mapping sequence diagram.
3.4 Magnetic Field Mapping
Field maps were measured in the novel experiments (Chapters 4-6) to inform the magnetisation
prediction models. In the Rabi modulated spectroscopy experiments (Chapter 4) a B0 and a B1
map were used to predict the off-resonance harmonic response, and a B1 map was incorporated in
the forward model used to reconstruct NMR spectra. In the Rabi modulated imaging experiments
(Chapter 5) a B0 and a B1 map were used to predict the off-resonance harmonic response, and a B0
map was incorporated in the forward model used to reconstruct projections of proton density. In
Chapter 6 both B0 and B1 maps were acquired for two ring-lock experiments; in the first experiment
to predict the steady-state trajectory, in the second to inform a model used to estimate relaxation
constants.
3.4.1 B0 mapping
The B0 field near the isocentre of a MR spectrometer is designed to provide a homogeneous field
over a specified spatial volume. When a phantom or patient is placed in the scanner, the field is
distorted. Field mapping methods allow the measurement of the spatial variation of the B0 field
and allow correction using the system’s shim coils or during reconstruction.
Early field mapping measurements (Maudsley et al., 1979, 1984) were taken with chemical shift
imaging which was time consuming, but much faster than manual movement and measurement with
a small NMR probe. A spin-echo based method (Sekihara et al., 1985) derived a field map from the
phase evolution between two spin-echo experiments with different echo times. The efficiency of this
method was improved by Prammer et al. (1988) who used the first and third echoes in a Carr-Purcell
echo train. A gradient recalled echo method (Schneider and Glover, 1991) calculates B0 strength
from the phase difference between two gradient echo experiments with different echo times. The
efficiency was improved by Kanayamay et al. (1996) who proposed acquiring both gradient echoes
in a single repetition.
In our experiments the B0 mapping sequence (Kanayamay et al., 1996) is used to excite a slice and
generate three gradient echo images (Figure 3.5). The phase difference between the first and third
70 CHAPTER 3. METHODS
echo images,
δφ (r) = φ3 (r)− φ1 (r)
= ∆ (r) δT (3.6)
is proportional to the off-resonance, ∆ (r), and the duration between the first and third echo, δT .
The field map is the expressed as a deviation from the Larmor frequency and is related to the phase
difference (3.6) by
δω0(r) = ∆ (r) =
δφ (r)
δT. (3.7)
The echo spacing δT can be selected to ensure any large spin populations with a chemical shift, such
as fat, undergo a 2π phase shift and don’t corrupt the B0 map.
3.4.2 B1 mapping
Inhomogeneity in the B1 field is caused by coil geometry and dielectric effects resulting in spatially
non-uniform flip angles and image intensity artifacts. Early mapping of the excitation field involved
moving a small pickup coil and sample to different spatial locations and making multiple measure-
ments. These methods were replaced by imaging methods which are more efficient and reflect the
actual experimental conditions.
B1 field maps can be calculated from the intensity of a single gradient-echo or spin-echo experiment
(Hornak et al., 1988) but suffer from intensity changes due coil sensitivity, proton density and
relaxation effects. Theses confounding factors are cancelled out by calculating the ratio of two
intensity images acquired from two experiments (Stollberger et al., 1988) where the excitation angle
is often doubled in the second experiment. The sensitivity of these double angle methods (DAM) can
be improved when using a spin-echo sequence by setting the refocusing excitation angle to twice the
excitation angle (Insko and Bolinger, 1993). Efficiency improvements to double angle methods for
application in-vivo are the use of compensation pulses (Stollberger and Wach, 1996) and saturation
pulses (Cunningham et al., 2006). Alternative intensity based methods are actual flip angle imaging
(Yarnykh, 2007) and a stimulated echo approach (Akoka et al., 1993). An recent phase based method
makes use of the Bloch-Seigert shift (Sacolick et al., 2010).
In our experiments we use the spin-echo double angle method as described by Stollberger et al.
(1988) and shown in Figure 3.6. The signal intensity for a spin-echo with arbitrary excitation angle,
α, and refocusing angle, β, is (Wang et al., 2005)
|mxy (r)| = M0z e−TE/T2S (r)
sinα[1− e−TR/T1 cosβ − (1− cosβ) e−TR/T1eTE/2T1
]1− e−TR/T1 cosα cosβ
, (3.8)
where S (r) is the coil sensitivity. If we ensure, TR T1, TE T1, and, β = 2α, then (3.8) simplifies
to,
|mxy (r)| = C (r)S (r) sin3 α (r) , (3.9)
where C (r) is the signal weighting due to relaxation constants and proton density. The intensity
ratio of two spin-echo experiments, with flip angles, α1, α2, can be expressed as
λ (r) =|mxy,2 (r)||mxy,1 (r)|
=sin3 α2 (r)
sin3 α1 (r), (3.10)
3.4. MAGNETIC FIELD MAPPING 71
Figure 3.6: B1 mapping sequence diagram. Two spin-echo experiments are conducted where the refocusing angle isset to twice the excitation angle, that is β = 2α. The excitation and refocusing angles are doubled in the secondexperiment, α2 = 2α1 and β2 = 2β1. The B1 map is calculated from the ratio of the two reconstructed intensityimages using equation (3.12).
which removes the dependance on tissue and coil parameters. If the excitation angle in the second
experiment is set as α2 = 2α1, then the flip angle can be calculated from the intensity ratio (3.10),
α1 (r) = arccos
(λ (r)
8
)1/3
. (3.11)
The B1 map can be calculated from (3.11) and (2.52) assuming a rectangular pulse of duration τ1,
B1 (r) =α1 (r)
γτ1=
1
γτ1arccos
(λ (r)
8
)1/3
. (3.12)
It is often more useful to express the excitation field inhomogeneity as a ratio of measured and
calibrated field strengths,
B1 (r)
B1=α1 (r)
α1, (3.13)
as this removes the excitation field strength dependence.
This chapter has outlined the standard MRI sequences employed in this thesis to support the devel-
opment of novel methods. Chapters 4-6 now proceed to detail these contributions.
72 CHAPTER 3. METHODS
Chapter 4
Rabi continuous wave spectroscopy
Contents4.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
4.2 Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
4.2.1 Rabi modulated excitation . . . . . . . . . . . . . . . . . . . . . . . . . . 74
4.2.2 Observed NMR signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
4.2.3 Spectroscopy as an inverse problem . . . . . . . . . . . . . . . . . . . . . . 76
4.3 Methods and Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
4.3.1 The response of off-resonance spins . . . . . . . . . . . . . . . . . . . . . . 79
4.3.2 Rabi modulated spectroscopy . . . . . . . . . . . . . . . . . . . . . . . . . 79
4.4 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
4.4.1 The response of off-resonance spins . . . . . . . . . . . . . . . . . . . . . . 83
4.4.2 Rabi modulated spectroscopy . . . . . . . . . . . . . . . . . . . . . . . . . 83
4.5 Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 86
4.6 Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
4.1 Introduction
This chapter explores the magnetic resonance response to an amplitude modulated RF excitation
envelope, which we refer to as Rabi modulated excitation. The experiments presented in this chapter
have been published in (Korte et al., 2014, 2017a). They were conducted to demonstrate that
the steady-state spin system response to Rabi modulated CW excitation contains off-resonance
information, in this case chemical shift information. We extend the original experiment (Layton
et al., 2014), which was limited to on-resonance excitation, to investigate the response of the spin
system to off-resonance Rabi modulated CW excitation. We propose a CW method by which to
perform NMR spectroscopy, posed as an inverse problem. Chemical shift information is encoded in
a series of Rabi modulated CW excitations. A forward model is constructed from a periodic solution
of the Bloch equation, and is used to reconstruct a simple NMR spectrum of ethanol. An algorithm
based on the A-optimality criteria (Chaloner and Verdinelli, 1995) is used to select a theoretically
optimal set of excitation parameter pairs.
73
74 CHAPTER 4. RABI CONTINUOUS WAVE SPECTROSCOPY
Historically, NMR spectra were obtained under continuous wave RF excitation by sweeping either
the strength of the main magnetic field or the carrier frequency of the excitation magnetic field, with
the spectrum generated by plotting the magnitude of the resulting NMR signal against the sweep
range (Bloch et al., 1946; Purcell et al., 1946). These methods were overshadowed by the more
efficient technique in which a sample is excited by a powerful, short duration, RF pulse (Ernst and
Anderson, 1966). The spectrum is obtained directly from a Fourier transform of the free induction
decay signal which is a broadband combination of all precessing isochromats.
A resurgence of interest in CW NMR spectroscopy methods (Blumich, 1984; Dadok and Sprecher,
1974) has influenced the development of new magnetic resonance imaging techniques capable of
imaging samples with ultra-fast spin-spin relaxation (Idiyatullin et al., 2006, 2012). The optimality
of pulse excitation sequences has also been challenged (Tahayori et al., 2008, 2009) with the sug-
gestion that CW excitation may improve signal intensity for lower energy excitation. Recent work
(Layton et al., 2014), inspired by quantum optics (Cappeller and Muller, 1985), has experimentally
demonstrated that a spin system excited by a Rabi modulated CW achieves substantial periodic
steady-state magnetisation. The frequency components of this steady-state magnetisation are re-
stricted to harmonics of the excitation modulation frequency (Tahayori et al., 2009) and a maximum
harmonic magnitude is achieved when a secondary resonance condition is met (Layton et al., 2014).
The Rabi resonance condition has also been investigated in CW electron paramagnetic resonance
(EPR) (Saiko et al., 2015).
We anticipate that the results we present here will find application in the acquisition of MR informa-
tion from ultra-fast relaxation samples. The proposed excitation method can maintain an observ-
able steady-state magnetisation for such samples and provides an alternative to existing ultra-fast
relaxation methods such as Sweep Imaging with Fourier Transformation (Idiyatullin et al., 2006),
Ultra-short echo time imaging (Bergin et al., 1991) and Zero-echo time imaging (Weiger et al., 2011).
Such methods have beneficial clinical applications such as the assessment of bone and connective
tissue (Robson and Bydder, 2006), measurement of sodium concentration in brain tissue (Ouwerkerk
et al., 2003) and the detection of iron oxide nano-particles delivered to tumours (Wang et al., 2014).
4.2 Theory
4.2.1 Rabi modulated excitation
As introduced in Chapter 2, Rabi modulated excitation is an amplitude modulated RF field with
the previously defined envelope function,
ωe1 (t) = ω1 (1 + α cosωmt) , (2.48)
where ω1 = −γB1 is the average of the excitation envelope, B1 is the average excitation field
strength, α is the modulation level and ωm is the modulation frequency. It is known that under this
excitation the spin system achieves a significant periodic steady-state magnetisation and that the
magnitude of this steady-state is maximised when the Rabi resonance condition, ωm = ω1, is met
(Layton et al., 2014).
4.2. THEORY 75
4.2.2 Observed NMR signal
The observed noise-free NMR signal under Rabi modulated excitation is,
mobs (t, α, ωm, δrf, T1, T2) = . . . (4.1)∫∫m (t, α, ωm, ω1, δω0
, δrf, T1, T2) p (ω1) p (δω0) dω1 dδω0
where δω0is any deviation from the Larmor frequency, ω0, due to main field inhomogeneities,
chemical shift effects or applied gradients. The p (δω0) distribution represents deviations from the
Larmor frequency, p(ω1) represents inhomogeneities in the strength of the excitation field and δrf is
an offset of the excitation carrier frequency from the Larmor frequency.
The signal equation can be approximated by numerical integration over a regular grid of J excitation
amplitudes and K off-resonances
mobs (t, α, ωm, δrf, T1, T2) ≈ . . . (4.2)
J∑j=1
K∑k=1
m(t, α, ωm, ω
(j)1 , δ(k)
ω0, δrf, T1, T2
)p(ω
(j)1
)p(δ(k)ω0
).
The steady-state magnetisation can be described by a complex Fourier series, restricted to harmonics
of the modulation frequency, ωm,
m =
∞∑l=−∞
c(l)eilωmt, (4.3)
m =
∞∑l=−∞
ilωmc(l)eilωmt, (4.4)
where c(l) =[c(l)x c
(l)y c
(l)z
]Tare the Fourier coefficients. In matrix notation, the relationship
becomes
m = eTC (4.5)
where
e =[. . . e−2iωmt e−iωmt 1 eiωmt e2iωmt . . .
]T(4.5a)
C =[. . . c(−2) c(−1) c(0) c(1) c(2) . . .
]T. (4.5b)
The bulk magnetisation in the frequency domain, C, is predicted by a harmonic balance solution of
the Bloch equations (2.47) as described in Section 2.3.1.2. The observed bulk magnetisation (4.2) is
transformed into the frequency domain by substitution of (4.5) and (2.47) into (4.2)
Cobs (α, ωm, δrf, T1, T2) ≈ . . . (4.6)
J∑j=1
K∑k=1
C(α, ωm, ω
(j)1 , δ(k)
ω0, δrf, T1, T2
)p(ω
(j)1
)p(δ(k)ω0
)
=
K∑k=1
S(α, ωm, δ
(k)ω0, δrf, T1, T2
)p(δ(k)ω0
)where
S (α, ωm, δω0 , δrf, T1, T2) =
J∑j=1
C(α, ωm, ω
(j)1 , δω0 , δrf, T1, T2
)p(ω
(j)1
). (4.6a)
76 CHAPTER 4. RABI CONTINUOUS WAVE SPECTROSCOPY
4.2.3 Spectroscopy as an inverse problem
It is possible to reconstruct a NMR spectrum under Rabi modulated excitation, as chemical shift
effects are encoded in the observed NMR signal. A linear system can be constructed from (4.6) by
exciting the spin system with a set of N excitation parameter pairs
(α(1), ω(1)m ) . . . (α(N), ω
(N)m )
and subsequent measurement of the transverse magnetisation,
Hx = z (4.7)
where
H =
S(α(1), ω
(1)m , δ
(1)ω0
)· · · S
(α(1), ω
(1)m , δ
(K)ω0
)S(α(2), ω
(2)m , δ
(1)ω0
)· · · S
(α(2), ω
(2)m , δ
(K)ω0
)...
. . ....
S(α(N), ω
(N)m , δ
(1)ω0
)· · · S
(α(N), ω
(N)m , δ
(K)ω0
)
(4.7a)
x =[p(δ(1)ω0
). . . p
(δ(K)ω0
) ]T(4.7b)
z =[Cobs
(α(1), ω(1)
m
)Cobs
(α(2), ω(2)
m
). . . Cobs
(α(N), ω(N)
m
) ]T. (4.7c)
The discrete NMR spectrum, x, is reconstructed from the known forward model, H, and an observa-
tion vector, z, by solving the linear system (4.7). In this formulation (4.7), the relaxation constants,
T1 and T2, and the RF offset, δrf, are assumed to be known constants and are omitted for notational
simplicity.
4.3 Methods and Materials
Two experiments were performed on a 4.7T Bruker BioSpec small bore MRI scanner, using the
experimental protocol developed in (Layton et al., 2014) and summarised in Figure 4.1. Rabi mod-
ulated excitation (2.48) was applied for an initial duration of T=1000 ms to allow the magnetisation
to reach a steady-state, after which the free induction decay (FID) was measured. A steady-state
magnetisation waveform, mobs, was incrementally acquired by selecting the first FID point, and
repeating the process with an increase, ε, to the excitation duration on each repetition. Whilst this
method is inefficient, it allows the measurement of the steady-state magnetisation waveform, such
as Figure 4.2a, without hardware modification.
The incremental increase to the excitation duration, ε, is used to set the sampling frequency of
the measured steady-state magnetisation. The experimental sampling frequency is determined by
the modulation frequency of the Rabi excitation envelope and the desired harmonic content of
the measured steady-state response. The finite switching time between RF excitation and signal
acquisition is fast enough (∼5µs) that relaxation effects on the first points of the FID are considered
negligible.
In the first experiment, the steady-state harmonics, Cobs, are measured over a range of off-resonance,
δrf, and compared to the predicted steady-state harmonics by integrating the periodic solution of the
Bloch equations over the measured p(δω0) and p(ω1) field distributions. In the second experiment,
ethanol spectra, p(δω0), are reconstructed from a series of steady-state harmonics measured under a
range of Rabi modulated excitation envelopes.
4.3. METHODS AND MATERIALS 77
Figure 4.1: Experimental protocol for the incremental measurement of the steady-state transverse magnetisationwaveform. The spin system is perturbed by an excitation, ωe
1 (t), causing a transverse response mxy. By repeatedlyperturbing the spin system with an increasing excitation duration it is possible to construct an observed transversemagnetisation mobs
xy .
78 CHAPTER 4. RABI CONTINUOUS WAVE SPECTROSCOPY
0 2 4 6 8 10−0.2
0
0.2
0.4
mobs
x
(norm
alised)
time(ms)
−1000 −500 0 500 1000−0.1
−0.05
0
0.05
0.1
frequency(Hz)
Re(
cobs
x
)
(a)0 2 4 6 8 10−0.2
0
0.2
0.4
mobs
x
(norm
alised)
time(ms)
−1000 −500 0 500 1000−0.1
−0.05
0
0.05
0.1
frequency(Hz)
Re(
cobs
x
)
(b)
Figure 4.2: Spin system response of the water phantom (T1 = 287 ms, T2 = 150 ms) under Rabi modulated excitation.(a) Measured (red circles) and theoretical (black line) periodic steady-state magnetisation waveform for excitationparameters α = 1, ω1 = ωm = 100 Hz and δrf = 180 Hz. (b) Measured (coloured circles) and theoretical (blackcrosses) harmonics of steady-state magnetisation. DC component (blue), first (green), second (purple), third (red)fourth (orange) and fifth (grey) harmonics.
4.3. METHODS AND MATERIALS 79
Two spherical phantoms were used in this study, one filled with tap water and the other with ethanol.
Both phantoms were doped with Magnevist R© to reduce the experimental time.
4.3.1 The response of off-resonance spins
We investigated the effects of off-resonance on the observed steady-state magnetisation waveform by
offsetting the excitation carrier frequency from the Larmor frequency. A spherical phantom of Gd-
doped water (T1 = 287 ms, T2 = 150 ms) was selected for its narrow off-resonance distribution. Off-
resonance measurements were taken from δrf = −400 Hz to δrf = 400 Hz with 20 Hz increments. The
excitation envelope had parameters α = 1 and ω1 = ωm = 100 Hz. A discrete Fourier transform was
applied to the measured steady-state magnetisation waveform to extract its harmonic components.
A theoretical curve was generated from the periodic solution of the Bloch equations, numerically
integrated over the measured p (δω0) and p(ω1) distributions using (4.6).
The distribution of off-resonances, p(δω0), was measured using a B0 field mapping sequence (Kanaya-
may et al., 1996) described in Section 3.4.1. Data was acquired using a multiple gradient echo se-
quence with effective echo times (TE = 1.95, 6.22 ms), FA = 20, TR = 20 ms, FOV = 4.5 cm and
128×128×128 matrix. The distribution of excitation field strengths, p(ω1), was measured with a B1
mapping sequence (Stollberger et al., 1988) described in Section 3.4.2. Data was acquired with a
MSME sequence over 9 slices, 1 mm thickness, FOV = 4 cm, 64×64 matrix, TE = 12 ms, TR = 2.5
s. The MSME scans were taken for two excitation angle, α, and refocusing angle, β, configurations;
α/β = 45/90 in the first scan, α/β = 90/180 in the second scan.
The average T1 relaxation rate was calculated from the spatial average of T1 maps, which were
measured as described in Section 3.3.1. Data was acquired with a RARE-VTR sequence over 9
slices, 1 mm thickness, FOV = 4 cm, 32×32 matrix, TE = 12 ms, TR = 4500, 1500, 800, 500, 400,
300, 250, 215 ms, RARE factor 2. The average T2 relaxation rate was calculated from the spatial
average of T2 maps, which were measured as described in Section 3.3.2. Data was acquired with a
MSME sequence using 32 echoes with a 12 ms echo spacing, 9 slices, 1 mm thickness, FOV = 4 cm,
64×64 matrix, TR = 2500 ms.
4.3.2 Rabi modulated spectroscopy
A proof of concept experiment was undertaken to demonstrate the ability to encode chemical shift
information in a series of steady-state magnetisation trajectories. Our objective was to reconstruct
the spectrum of a spherical phantom of gadolinium doped ethanol (T1 = 120 ms, T2 = 43 ms) from
the response to a series of CW excitations. A reference spectrum from a single, 2048-point, FID was
acquired with a dwell time of 100µs. Spectra and variance are plotted on a ppm scale,
δ =δω0× 106
ω0+ δtms, (4.8)
where δtms = 1.25 ppm is an offset commonly observed when using Tetramethylsilane (TMS) as a
reference compound. No reference compound was added to the doped ethanol sample used in our
experiments.
4.3.2.1 Excitation parameter selection
Experimental parameters were selected from a set of candidate parameter pairs, Ω = (α, ωm),defined by a dense grid adhering to the power limitation of the excitation coil, as illustrated in
80 CHAPTER 4. RABI CONTINUOUS WAVE SPECTROSCOPY
0 5 10 150
100
200
300
400
500
600
α
ωm(H
z)
Figure 4.3: Feasible (light grey) excitation parameter pairs Ω, set of gridded (dark grey) excitation parameter pairsΩgrid, set of optimal (black) excitation parameter pairs Ωopti for a 101 point reconstruction from -600 Hz to 600 Hzand a homogeneous B1 field.
Figure 4.3. Two reduced sets of 500 parameter pairs were then selected for experiments. The first
set, Ωgrid, was selected as a coarse grid of the candidate parameter pairs. The second set, Ωopti,
was constructed iteratively to minimise the theoretical variance of the reconstructed spectrum, as
detailed in Algorithm 1.
At each measurement selection step, k, in Algorithm 1, the covariance, Xk, for every candidate
parameter pair in the candidate set was calculated using,
Xk = Xk−1 −Xk−1HTk
(HkXk−1H
Tk +W
)−1
HkXk−1 , (4.9)
where W is the covariance of observation noise. The optimal parameter pair was selected from the
candidate set using the A-optimality criteria (Dette, 1997),
minimise(α,ωm)∈Ω
Tr (Xk) , (4.10)
where Tr denotes the matrix trace.
The selected excitation parameter pair was then added to Ωopti and used to update Xk. The
selection cycle was repeated until the required number of excitation parameter pairs were selected.
The variance of the optimal set is shown in Figure 4.4.
The initial covariance was set to X0 = σ2I where the variance, σ2 = 10,000, and I is an identity
matrix, providing an initial condition with negligible regularisation. The observation noise covariance
matrix was set to W = σ2obsI, with a standard deviation σobs = 7.2× 10−3, which is approximately
5% of the mean signal.
4.3.2.2 Measurement
Steady-state magnetisation waveforms were acquired for every CW excitation parameter pair in
Ωgrid and Ωopti. The excitation carrier frequency was offset from the system frequency by δrf = 400
4.3. METHODS AND MATERIALS 81
Algorithm 1 Select the optimal set, Ωopti, of N excitation parameter pairs from a candidate set,Ω.
function OptimalSet(Ω, N)X0 ← σ2IΩopti ← . an empty setk ← 1 . measurement counterwhile k ≤ N do
(α′, ω′m)← argmin(α,ωm)∈Ω
Tr (Xk(α, ωm))
add (α′, ω′m) to Ωopti
remove (α′, ω′m) from ΩXk ←Xk(α′, ω′m) . update covariancek ← k + 1
end whilereturn Ωopti
end function
Hz to centre the spectrum and reduce the required bandwidth. The DC component and first five
harmonics were extracted from each steady-state magnetisation and recorded in a measurement
vector z.
Relaxation constants T1 and T2 were measured before each experiment, using a rapid acquisition
with relaxation enhancement with variable repetition time (RARE-VTR) scan as described in Sec-
tion 3.3.1, to ensure accuracy of the encoding matrix H. Scans were taken with RARE factor 2
over a single 1 mm slice, FOV = 4 cm, 64×64 matrix, TR = 200, 400, 800, 1500, 3000, 4500 ms, TE
= 11, 33, 55, 77, 99 ms. The measured relaxation maps were spatially averaged to produce a single
T1 and T2 constant value per phantom.
The distribution of excitation field strengths, p(ω1), was measured with a B1 mapping sequence
(Stollberger et al., 1988) described in Section 3.4.2. Data was acquired with a MSME sequence over
9 slices, 1 mm thickness, FOV = 4 cm, 64×64 matrix, TE = 12 ms, TR = 2.5 s. The MSME scans
were taken for two excitation angle, α, and refocusing angle, β, configurations; α/β = 45/90 in
the first scan, α/β = 90/180 in the second scan.
Reference spectra were measured with a single FID (FA = 90, 2048 complex points, spectral width
5.0 kHz). The Fourier transform of the FID was manually phase corrected and the real component
used as a reference spectrum. All measurements were duplicated to verify the reproducibility of
results.
4.3.2.3 Reconstruction
The spectrum, x, was reconstructed by least squares optimisation with a non-negative constraint:
minimisex∈[0,∞)
‖Hx− z‖2. (4.11)
The forward model, H, was constructed from the Fourier series approximation of the Bloch equation,
numerically integrated over the measured p(ω1) distribution using (4.7). The optimisation algorithm
was initialised with the spectrum, x, set as a vector of zeros.
82 CHAPTER 4. RABI CONTINUOUS WAVE SPECTROSCOPY
1234560
1
2
3
x 10−4
δ (ppm)
Tr(X)
Figure 4.4: Trace of covariance matrix of reconstructed spectra (4.10) for the gridded (grey triangles) set of excitation
parameter pairs Ωgrid and the optimal (black circles) set of excitation parameter pairs Ωopti. Mean variance for the
optimal set is 1.4× 10−4 and 2.7× 10−4 for the gridded set.
4.3.2.4 Spectrum analysis
Let x be the area normalised version of the reconstructed spectra, x. Lorentzian peaks were fitted
to x, to assess the accuracy of peak locations and the ratio of peak areas. A Lorentzian peak is
defined as
L (δ, µ,Γ, a) =a
π
[Γ
(δ − µ)2 + Γ2
](4.12)
where δ is a discrete vector of off-resonance, µ is the peak centre, Γ is the half width at half maximum
(HWHM) and a is the peak area.
As ethanol has three distinct peaks, the objective function is
minimiseµ,Γ,a
∥∥∥∥∥x−3∑i=1
L (δ, µi,Γi, ai)
∥∥∥∥∥2
+ g
∥∥∥∥∥1−3∑i=1
ai
∥∥∥∥∥2
(4.13)
where
µ =
µ1
µ2
µ3
, Γ =
Γ1
Γ2
Γ3
, a =
a1
a2
a3
(4.13a-c)
and g is a weighting factor for the unity area term. The constrained minimisation function, fmincon,
from MATLAB R© was used to solve optimisation problems (4.11) and (4.13).
4.4. RESULTS 83
−100 −50 0 50 1000
0.01
0.02
0.03
0.04
0.05
off-resonance, δω0(Hz)
p(δ
ω0)
(a)
0.7 0.8 0.9 1 1.1 1.2 1.30
0.02
0.04
0.06
0.08
0.1
0.12
0.14
RF amplitude scaling (×ω1)
p(ω
1)
(b)
Figure 4.5: Measured distributions of the water phantom. (a) Off-resonance distribution p (δω0 ) extracted from the
histogram of a B0 field map. (b) RF power distribution p (ω1) extracted from the histogram of a B1 field map.
4.4 Results
4.4.1 The response of off-resonance spins
The measured and theoretical steady-state magnetisation waveform, for a single measurement, δrf= 180 Hz, is shown in Figure 4.2. The spectrum of this steady-state magnetisation (Figure 4.2b)
shows that information is restricted to harmonics of the Rabi frequency, ω1 = 100 Hz. The measured
p (δω0) and p (ω1) distributions, used in the generation of the predicted harmonic curves, are shown
in Figure 4.5a and Figure 4.5b, respectively. The variation in the harmonics is shown over a range of
off-resonances (Figure 4.6). The experimental measurements agree with the theoretical curves and
show that the relative strengths of the harmonic components are influenced by off-resonance effects.
4.4.2 Rabi modulated spectroscopy
Reconstructed ethanol spectra from the reference FID, gridded parameter set Ωgrid and optimal
parameter set Ωopti are shown in Figure 4.7. These reconstructions demonstrate that under Rabi
modulated CW excitation, chemical shift information can be encoded in the steady-state magneti-
sation.
84 CHAPTER 4. RABI CONTINUOUS WAVE SPECTROSCOPY
−400 −300 −200 −100 0 100 200 300 400
−0.1
−0.05
0
0.05
0.1
0.15
δrf (Hz)
Re(
cobs
x
)
(a)
−400 −300 −200 −100 0 100 200 300 400
−0.1
−0.05
0
0.05
0.1
0.15
δrf (Hz)
Im(
cobs
y
)
(b)
Figure 4.6: Harmonic curves of the water phantom under Rabi modulated CW excitation with parameters α = 1 and
ω1 = ωm = 100 Hz. Measured (circles) and theoretical (solid line) frequency coefficients of steady state magnetisation.
DC component (blue), first (green), second (purple), third (red) fourth (orange) and fifth (grey) harmonics.
4.4.
RESULTS
85
1234560.00
0.20
0.40
δ (ppm)
Ωgrid
spectru
m(areanorm
alised)
1234560.00
0.20
0.40
δ (ppm)
Ωopti
spectru
m(areanorm
alised)
(a)
1234560.00
0.20
0.40
δ (ppm)
Ωgrid
spectru
m(areanorm
alised)
1234560.00
0.20
0.40
δ (ppm)
Ωopti
spectru
m(areanorm
alised)
(b)
Figure 4.7: NMR spectrum of doped ethanol phantom from reference FID (grey) and reconstruction from Rabi modulated CW excitation (black). (a) Reconstruction from
the Ωgrid measurement set. (b) Reconstruction from the Ωopti measurement set.
86 CHAPTER 4. RABI CONTINUOUS WAVE SPECTROSCOPY
Table 4.1: Results from the spectrum analysis, with a g = 0.01, of the reference and reconstructed ethanol spectra.
The parameters of three fitted Lorentzian peaks are centre µ, half width half max Γ and area a. The chemical shift δ1
is between CH3 and CH2 and the chemical shift δ2 is between CH2 and OH. Peak centre, peak width and chemical
shifts are listed in parts per million (ppm).
CH3 Peak CH2 Peak OH Peak Summary
µ Γ a µ Γ a µ Γ a δ1 δ2 aratio
REFERENCE
Literature (Arnold, 1956) 1.250 0.500 3.709 0.333 5.365 0.167 2.459 1.656 3.0 : 2.0 : 1.0
FID 1.305 0.07 0.524 3.737 0.07 0.341 5.525 0.08 0.156 2.432 1.788 3.1 : 2.0 : 0.9
MEASUREMENT 1
Optimal 1.272 0.05 0.533 3.751 0.04 0.332 5.589 0.07 0.135 2.479 1.838 3.2 : 2.0 : 0.8
Grid 1.270 0.05 0.527 3.753 0.04 0.334 5.567 0.06 0.138 2.483 1.815 3.2 : 2.0 : 0.8
MEASUREMENT 2
Optimal 1.266 0.05 0.534 3.746 0.05 0.336 5.575 0.07 0.130 2.480 1.829 3.2 : 2.0 : 0.8
Grid 1.268 0.05 0.521 3.744 0.04 0.333 5.552 0.04 0.146 2.476 1.808 3.1 : 2.0 : 0.9
The results of the Lorentzian fitting are shown in Table 4.1 where the reference entries for relative
peak shifts (δ1, δ2) and area ratios (aratio) are experimental results from (Arnold, 1956). The
results for both parameter sets, for both measurements, and the result from the reference FID
are comparable. The measured and reconstructed spectra differ slightly from those observed in
(Arnold, 1956) with larger relative peak shifts, enlarged CH3 peak areas and reduced OH peak
areas. Spectra reconstructed from the optimal parameter set, Ωopti, are slightly more accurate than
those reconstructed from the gridded parameter set, Ωgrid, which suffer from an artifact near 0.25
and 6.25 ppm.
4.5 Discussion
The results from the investigation of off-resonance spins verify that our periodic solution of the Bloch
equations (4.6) can predict off-resonance behaviour of the spin system under Rabi continuous wave
excitation. Furthermore, the hypothesis that spectral information is restricted to harmonics of the
Rabi frequency (Layton et al., 2014) for the resonant case, ∆ = 0 Hz, has now been extended to
off-resonance cases.
Variations between prediction and measurement can be attributed to a number of factors such as
measurement error of relaxation time constants, measurement error of field distributions or phase
error between the prediction model and measurements. Whilst the error introduced by each of these
factors should be minimal, the combined effect of these errors on the prediction model, H, is the
subject of ongoing investigation.
Results from the Rabi modulated spectroscopy experiment demonstrate that it is possible to encode
chemical shift in a series of CW excitations and reconstruct a spectrum. This method requires the
measurement of relaxation constants to ensure an accurate forward model, H, this requirement may
be removed in future experiments by joint estimation (Bretthorst et al., 2005).
The reconstructions from the optimal and gridded parameter sets are comparable and the lack of
an artifact around 0.25 and 6.25 ppm in the optimal reconstruction may be attributed to the lower
theoretical variance in these regions as shown in Figure 4.4. The reconstruction routine (4.11) also
converges in less iterations when reconstructing with measurements from the optimal parameter set.
4.6. CONCLUSION 87
Imperfections in the spectra reconstructed from measurements under Rabi modulated excitation can
be attributed to error in the forward model, H. The receiver attenuation was fixed during each
experiment which may have introduced error when measuring magnetisation waveforms with a lower
relative power, where experimental imperfections such as low SNR begin to dominate.
In this chapter spectroscopy is posed as an inverse problem thus it is important to consider the
stability of the reconstruction. The low theoretical variance of the estimated spectra, shown in
Figure 4.4, provides an exact quantification of the predicted accuracy and demonstrates that the
inverse problem is well conditioned. The reconstruction algorithm was tested with multiple starting
points and converged to the same solution as presented. Currently, this method has the potential
to produce non-physical spectra, such as the small spurious peaks around 4.5 ppm in Figure 4.7.
This is likely due to minor imperfections in experimental setup and with an improved acquisition
strategy this should not be an issue.
The proposed method is not time efficient and in Chapter 5 we improve efficiency with a gapped
excitation and measurement protocol, similar to (Idiyatullin et al., 2006). Future work will con-
sider hardware modification (Brunner et al., 2011, 2012) to allow simultaneous transmit and receive.
In (Brunner et al., 2011, 2012), sideband modulation was used to spectrally isolate the excitation
and induced NMR signal. Similar filtering could be used here but no sideband modulation would
be required as the Rabi modulated response contains harmonic information outside the excitation
bandwidth. Alternative strategies for decoupling the transmit and NMR signal are analog cancela-
tion (Ozen et al., 2017b) and signal isolation in the digital domain under low power excitation (Ozen
et al., 2017a).
With a gradient applied during the Rabi spectroscopy experiment, the reconstruction vector, x, is a
projection of proton density along the gradient direction rather than a spectrum. Imaging will thus
result from the acquisition of a sufficient number of projections, to be reconstructed using a filtered
back-projection algorithm or an iterative algorithm, such as the conjugate gradient method. This
idea is explored in Chapter 5.
The ability to maintain an observable steady-state magnetisation under Rabi modulated excitation
may provide better SNR than methods that measure a single fast decaying FID signal. Rabi modu-
lated steady-state methods are true zero echo time techniques, if hardware modifications are made
to allow simultaneous excitation and observation. Measurement of ultra-short relaxation samples
such as sodium, bone or connective tissue will be explored in future work.
4.6 Conclusion
Rabi modulated CW excitation causes an observable periodic steady-state magnetisation with spec-
tral information at harmonics of the Rabi frequency. This steady-state magnetisation is influenced
by off-resonance effects, in a manner that can be accurately predicted by a periodic solution of the
Bloch equations. In this chapter, we have verified the existence of the off-resonance response, by
experimental measurement of the spin-system under Rabi modulated excitation. We have further
demonstrated the utility of the off-resonance harmonic response by encoding chemical shift infor-
mation in a series of measurements and the reconstruction of doped ethanol spectra. In Chapter 5,
using a more efficient measurement protocol, we extend this idea to encode radial projections under
Rabi modulated excitation and reconstruct a proton density image.
88 CHAPTER 4. RABI CONTINUOUS WAVE SPECTROSCOPY
Chapter 5
Rabi continuous wave imaging
Contents5.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
5.2 Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
5.2.1 Observed NMR signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
5.2.2 Radial projection imaging as an inverse problem . . . . . . . . . . . . . . 93
5.3 Methods and Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
5.3.1 Gapped excitation and measurement protocol . . . . . . . . . . . . . . . . 94
5.3.2 Gapped measurement of off-resonance response . . . . . . . . . . . . . . . 94
5.3.3 Rabi modulated imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
5.3.4 Rabi modulated imaging contrast . . . . . . . . . . . . . . . . . . . . . . . 100
5.4 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
5.4.1 Gapped measurement of off-resonance response . . . . . . . . . . . . . . . 100
5.4.2 Rabi modulated imaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
5.4.3 Rabi modulated imaging contrast . . . . . . . . . . . . . . . . . . . . . . . 102
5.5 Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
5.6 Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
5.1 Introduction
In Chapter 4 the off-resonance response to a series of Rabi modulated RF excitations was used to
encode chemical shift information and reconstruct NMR spectra. In this chapter we extend the idea
to the imaging problem, and implement a proof of concept continuous wave (CW) imaging technique
(Korte et al., 2016). Again, we exploit the response to Rabi modulated excitation, where the spin-
system achieves an observable periodic magnetisation (Layton et al., 2014) and the steady-state
magnetisation can be fully described as harmonics of the excitation envelope modulation frequency
(Tahayori et al., 2015).
It is known that off-resonance effects influence the steady-state harmonics, as shown in the previous
chapter to encode chemical shift information and reconstruct ethanol spectra. Here, we focus a
different source of off-resonance, linear gradient fields, which are applied during a Rabi modulated
89
90 CHAPTER 5. RABI CONTINUOUS WAVE IMAGING
excitation envelope to encode spatial projections of proton density. Harmonics of the steady-state
response are acquired under a series of Rabi excitation parameters and a range of radial gradients.
For each gradient direction a model based optimisation is performed to reconstruct radial projections
of proton density, followed by filtered back projection (FBP) of all the reconstructed projections to
form a proton density image.
Rabi modulated excitation has potential to image ultra-short T2 tissues, which existing techniques
such as UTE (Bergin et al., 1991), ZTE (Weiger et al., 2011) and SWIFT (Idiyatullin et al., 2006)
have shown is clinically valuable in the diagnosis of muscular skeletal injury and disease (Robson
and Bydder, 2006; Weiger et al., 2013a; Luhach et al., 2014). A time-shared measurement protocol,
which rapidly switches between excitation and data acquisition, is used to observe the response to
Rabi modulated excitation in this chapter. This gapped protocol is an order of magnitude faster
than the iterative protocol used in Chapter 4, but still suffers from the finite switching time between
the transmitter and receiver. Rabi modulated excitation and the spin-system response exhibit
properties which may allow simultaneous excitation and measurement, but require non-standard
hardware. Observations made with the gapped excitation protocol should allow measurement of
ultra-fast relaxation signal in the same order as UTE, ZTE and time shared SWIFT methods;
a simultaneous excitation and measurement implementation would allow measurement of signals
similar to continuous SWIFT (Idiyatullin et al., 2012).
5.2 Theory
As introduced in Chapter 2, Rabi modulated excitation is an amplitude modulated RF field with
the previously defined envelope function
ωe1 (t) = ω1 (1 + α cosωmt) (2.48)
where ω1 = −γB1 is the average of the excitation envelope, B1 is the average excitation field
strength, α is the modulation level and ωm is the modulation frequency. It is known that under this
excitation the spin system achieves a significant periodic steady-state magnetisation and that the
magnitude of this steady-state is maximised when the Rabi resonance condition, ωm = ω1, is met
(Layton et al., 2014).
5.2.1 Observed NMR signal
The observed noise-free NMR signal under Rabi excitation, here with radial 2D spatial dependance,
is,
mobs (α, ω1, ωm,∆, T1, T2) = . . . (5.1)∫∫m (α, ωm, ω1 (r, s) ,∆ (r, s) , T1 (r, s) , T2 (r, s)) ρ (r, s) dr ds,
5.2. THEORY 91
Figure 5.1: Radial coordinate system and the Radon transform.
where there are projections in the z direction for spin density, ρ, excitation field strength, ω1, and
relaxation constants, T1 and T2,
ρ (r, s) =
∫object
ρ (r) dz (5.1a)
ω1 (r, s) =
∫object
ω1 (r) dz (5.1b)
T1 (r, s) =
∫object
T1 (r) dz (5.1c)
T2 (r, s) =
∫object
T2 (r) dz. (5.1d)
The projection of off-resonance (2.33) in the z direction is,
∆ (r, s) =
∫object
[ω0
(1 + δω0
(r) + g (r))− ωrf
(1 + δrf (r)
)]dz, (5.1e)
where δrf is an RF carrier frequency offset δω0encompasses B0 field inhomogeneity and chemical
shift and g is the gradient field contribution. The radial coordinate system is defined for an abitrary
projection angle, θ, [r
s
]=
[cos θ sin θ
− sin θ cos θ
] [x
y
], (5.2)
and the Radon transform,
ρθ (r) =
∫ρθ (r, s) ds, (5.3)
is shown graphically in Figure 5.1.
To simplify (5.1), we assume a homogenous RF excitation field, ω1 = ωm, and approximate the
spatially dependant relaxation constants as two average relaxation constants, T av1 and T av
2 . We
92 CHAPTER 5. RABI CONTINUOUS WAVE IMAGING
assume the B0 field inhomogeneity and chemical shift is adequately represented by the distribution,
p (δω0), for all voxels. These assumptions reduce equation (5.1) to,
mobs
(α, ωm, δrf, T
av1 , T av
2
)= . . . (5.4)∫∫∫m(α, ωm, δrf, δω0
, g (r, s) , T av1 , T av
2
)p (δω0
) ρ (r, s) dδω0dr ds.
We re-order the integration in the simplified observed NMR signal (5.4),
mobs
(α, ωm, δrf, T
av1 , T av
2
)= . . . (5.5)∫∫m(α, ωm, δrf, δω0
, g (r, s) , T av1 , T av
2
)p (δω0
) dδω0
∫ρ (r, s) ds dr.
A radial gradient is defined as,
g (r) = Gr r, (5.6)
where Gr is a constant gradient strength. The Radon transform (5.3) and radial gradient (5.6) are
applied to (5.5) which gives,
mobs
(α, ωm, δrf, T
av1 , T av
2
)= . . . (5.7)∫∫m(α, ωm, δrf, δω0 , g (r) , T av
1 , T av2
)p (δω0) dδω0 ρθ (r) dr.
The integral (5.7) can be numerically approximated over a discrete grid,
mobs (α, ωm, δrf, Tav1 , T av
2 ) ≈ . . . (5.8)
J∑j=1
K∑k=1
m(α, ωm, δrf, δ
(j)ω0, g(r(k)
), T av
1 , T av2
)p(δ(j)ω0
)ρθ
(r(k)
)and expressed in the frequency domain by substitution of (4.5),
Cobs (α, ωm, δrf, Tav1 , T av
2 , Gr) ≈ . . . (5.9)
J∑j=1
K∑k=1
C(α, ωm, δrf, δ
(j)ω0, g(r(k)
), T av
1 , T av2
)p(δ(j)ω0
)ρθ
(r(k)
)
=
K∑k=1
S(α, ωm, δrf, r
(k), T av1 , T av
2 , Gr
)ρθ
(r(k)
)where
S (α, ωm, δrf, r, Tav1 , T av
2 , Gr) =
J∑j=1
C(α, ωm, δrf, δ
(j)ω0, Grr, T
av1 , T av
2
)p(δ(j)ω0
). (5.9a)
The expression for predicting the observed Rabi harmonics under a gradient field (5.9) has a sim-
ilar form to the expression (4.6) used in the Rabi spectroscopy inverse formulation. The proton
projection expression (5.9) is different from (4.6) as it integrates out the dependance on main field
inhomogeneity and chemical shift effects, δω0, reducing the off-resonance dependance to gradient
sources, g (r).
5.3. METHODS AND MATERIALS 93
5.2.2 Radial projection imaging as an inverse problem
It is possible to reconstruct a radial projection of proton density under Rabi modulated excitation,
as the off-resonance information from the gradient field is encoded in the observed NMR signal.
A linear system can be constructed from (5.9) by exciting the spin system with a set of N exci-
tation parameter tuples
(α(1), ω(1)m , δ
(1)rf ) . . . (α(N), ω
(N)m , δ
(N)rf )
and subsequent measurement of
the transverse magnetisation,
Hxθ = zθ (5.10)
where:
H =
S(α(1), ω
(1)m , δ
(1)rf , r
(1))
· · · S(α(1), ω
(1)m , δ
(1)rf , r
(K))
S(α(2), ω
(2)m , δ
(2)rf , r
(1))
· · · S(α(2), ω
(2)m , δ
(2)rf , r
(K))
.... . .
...
S(α(N), ω
(N)m , δ
(N)rf , r(1)
)· · · S
(α(N), ω
(N)m , δ
(N)rf , r(K)
)
(5.10a)
xθ =[ρθ
(r(1))
. . . ρθ
(r(K)
) ]T(5.10b)
zθ =[Cobs,θ
(α(1), ω(1)
m , δ(1)rf
)Cobs,θ
(α(2), ω(2)
m , δ(2)rf
). . . Cobs,θ
(α(N), ω(N)
m , δ(N)rf
) ]T(5.10c)
A discrete projection of proton density, xθ, with K points is reconstructed from the known forward
model, H, and an observation vector, zθ, by solving the linear system (5.10). If multiple projec-
tions are acquired with the same excitation envelope parameters and gradient strength, the forward
model (5.10a) only needs to be computed once, and can be used to reconstruct all proton density
projections. In this formulation (5.10), the average relaxation constants, T av1 and T av
2 , and gradient
amplitude, Gr, are assumed to be known constants and are omitted as functional arguments for
notational simplicity.
5.2.2.1 Image reconstruction
If we encode and reconstruct the radial proton density, xθ, for a range of projections, θ = [0, π), we
can construct a sinogram. For a given projection angle, θ, the radial gradient (5.6) is related to the
linear gradients by,
Gx = Gr cos θ (5.11a)
Gy = Gr sin θ. (5.11b)
A proton density image, ρ (r, s), can then be reconstructed from the sinogram using a standard
filtered back-projection algorithm, such as ‘iradon’ in MATLAB R©.
5.3 Methods and Materials
Two experiments were conducted on a 4.7T Bruker Biospec small bore scanner with an AVANCE III
console. In the first experiment the off-resonance response of a spin-system under Rabi modulated
94 CHAPTER 5. RABI CONTINUOUS WAVE IMAGING
excitation is measured using a gapped measurement protocol, developed to improve experimental
efficiency from the Rabi modulated experiments in Chapter 4. In the second experiment a series
of proton density projections are encoded under Rabi modulated excitation and used to recon-
struct a two dimensional proton density image. A Rabi modulated projection imaging simulation
was conducted on a numerical phantom with spatially varying relaxation constants to investigate
reconstructed image contrast.
5.3.1 Gapped excitation and measurement protocol
A gapped excitation measurement protocol (Korte et al., 2015) (Figure 5.3), similar to that used in
SWIFT (Idiyatullin et al., 2006), was used to achieve near-simultaneous transmit and receive. To
maintain a similar steady-state magnetisation trajectory as that observed under CW excitation, the
gapped Rabi modulated excitation envelope must be power corrected,
ωe1,gap (α, ω1, ωm, t, dcycle, Tgap) =
100dcycle
ωe1(α, ω1, ωm, t) tloc (t, Tgap) ≤ dcycleTgap
0 otherwise, (5.12)
where
tloc (t, Tgap) = t− floor
(t
Tgap
)Tgap (5.12a)
where dcycle is duty cycle and Tgap is the gapped sampling period. A comparison of the spin-system
steady-state response under continuous wave, ωe1, and gapped excitation envelope, ωe
1,gap, is shown
in Figure 5.2. The gapped excitation response oscillates around the continuous wave response, due
to the increased power during the excitation period, and relaxation during the acquisition period.
The gapped response intercepts the continuous wave response, approximately in the centre of the
acquisition period.
5.3.2 Gapped measurement of off-resonance response
To verify the gapped excitation protocol, the harmonic off-resonance response was experimentally
measured using a spherical phantom of Gd-doped water (T1 = 44 ms, T2 = 38 ms), selected for
its narrow off-resonance distribution. We investigated the effects of off-resonance on the observed
steady-state magnetisation waveform by offsetting the excitation carrier frequency from the Larmor
frequency. Off-resonance measurements were taken from δrf = −400 Hz to δrf = 400 Hz with an 8 Hz
increment. The gapped excitation envelope had parameters, α = 1, ω1 = ωm = 50 Hz, duty cycle,
dcycle = 90% and sampling period, Tgap = 1.67 ms. A discrete Fourier transform was applied to the
measured steady-state magnetisation waveform to extract its harmonic components. A theoretical
curve was generated from the periodic solution of the Bloch equations, numerically integrated over
the measured p (δω0) and p(ω1) distributions using (4.6).
The distribution of off-resonances, p(δω0), was measured using a B0 field mapping sequence (Kanaya-
may et al., 1996) described in Section 3.4.1. Data was acquired using a multiple gradient echo se-
quence with effective echo times (TE = 1.58, 4.43 ms), FA = 30, TR = 35 ms, FOV = 6 cm and
64×64×64 matrix. The distribution of excitation field strengths, p(ω1), was measured with a B1
mapping sequence (Stollberger et al., 1988) described in Section 3.4.2. Data was acquired with a
MSME sequence over 10 slices, 2 mm thickness, FOV = 6 cm, 128×128 matrix, TE = 20 ms, TR = 1
s. The MSME scans were taken for two excitation angle, α, and refocusing angle, β, configurations;
α/β = 45/90 in the first scan, α/β = 90/180 in the second scan.
5.3. METHODS AND MATERIALS 95
2 4 6 8 10
−0.2
0
0.2
time (ms)
my
2 4 6 8 100
159
time (ms)
ωe 1(H
z)
(a)
2 4 6 8 10
−0.2
0
0.2
time (ms)
my
2 4 6 8 100
159
time (ms)
ωe 1(H
z)
(b)
Figure 5.2: Comparison of spin-system response under Rabi modulated excitation, numerical simulation of Blochequation. (a) Rabi excitation envelope in continuous wave (grey line) and gapped (black area) mode. (b) Steady-statetrajectory in response to continuous wave excitation (grey line) and gapped excitation (black lines), relaxation duringthe measurement period (red lines) to acquire a single sample (red dots).
96 CHAPTER 5. RABI CONTINUOUS WAVE IMAGING
Figure 5.3: Gapped excitation protocol for near-simultaneous transmit and receive. The sample is excited by a Rabimodulated CW envelope, ωe
1 (t), causing the magnetisation to reach a periodic trajectory, mxy . The Fourier transformof this magnetisation, cxy , has information restricted to harmonics of the excitation envelope modulation frequency,ω1.
5.3. METHODS AND MATERIALS 97
Figure 5.4: Diagram of three test-tube phantom.
The average T1 relaxation rate was calculated from the spatial average of T1 maps, which were
measured as described in Section 3.3.1. Data was acquired with a RARE-VTR sequence over 1
slice, 1 mm thickness, FOV = 6 cm, 128×128 matrix, TE = 80 ms, TR = 200, 400, 800, 1500, 3000,
4500 ms, RARE factor 8. The average T2 relaxation rate was calculated from the spatial average
of T2 maps, which were measured as described in Section 3.3.2. Data was acquired with a MSME
sequence using 9 echoes with a 22 ms echo spacing, 1 slice, 1 mm thickness, FOV = 6 cm, 128×128
matrix, TR = 1 s.
5.3.3 Rabi modulated imaging
To demonstrate Rabi modulated imaging, an imaging phantom of three test tubes of Gadolinium
doped water (T1 = 41 ms, T2 = 33 ms) was aligned in the longitudinal axis (Figure 5.4). A coil insert
was 3D printed to hold the three test tubes in position, and ensure their longitudinal alignment.
This relatively accurate positioning allowed the three dimensional phantom to be considered as a
two dimensional object. An alternative strategy is to use a phantom which is thin in one dimension,
our method provides a larger proton signal.
5.3.3.1 Reference measurements
A reference image was acquired with the FLASH protocol as described in as described in Section 3.2
for 10 axial slices with a 2 mm thickness, FA = 30, TE = 4 ms, TR = 100 ms, FOV = 6 cm,
Matrix=128x128. The axial slices were averaged to construct a single two dimensional proton density
image. A Radon transform of the reference proton density image was used to generate projections
98 CHAPTER 5. RABI CONTINUOUS WAVE IMAGING
along the same projection angles, θ, as used in the Rabi CW experiment. From the reference image
projections, an under-projected reference image was reconstructed using a standard FBP algorithm.
The distribution of off-resonances, p(δω0), was measured using a B0 field mapping sequence (Kanaya-
may et al., 1996) described in Section 3.4.1. Data was acquired using a multiple gradient echo se-
quence with effective echo times (TE = 1.58, 4.43 ms), FA = 30, TR = 35 ms, FOV = 6 cm and
64×64×64 matrix.
The average T1 relaxation rate was calculated from the spatial average of T1 maps, which were
measured as described in Section 3.3.1. Data was acquired with a RARE-VTR sequence over 1
slice, 1 mm thickness, FOV = 6 cm, 128×128 matrix, TE = 80 ms, TR = 200, 400, 800, 1500, 3000,
4500 ms, RARE factor 8. The average T2 relaxation rate was calculated from the spatial average
of T2 maps, which were measured as described in Section 3.3.2. Data was acquired with a MSME
sequence using 9 echoes with a 22 ms echo spacing, 1 slice, 1 mm thickness, FOV = 6 cm, 128×128
matrix, TR = 1 s.
5.3.3.2 Rabi modulated measurements
Using the gapped excitation measurement protocol, the phantom was excited by a set of N = 4030
Rabi modulated CW excitations(α(1), ω
(1)1 , δ
(1)rf
), · · · ,
(α(N), ω
(N)1 , δ
(N)rf
)where δrf is an offset to
the RF carrier frequency. The modulation level, α, ranged from 0.5 to 5.0, the modulation frequency,
ωm = ω1, ranged from 30 Hz to 90 Hz and the offset to RF carrier, δrf, ranged from -4.5 kHz to 4.5
kHz. For each CW excitation the phantom was measured over 18 projections angles, θ, as shown
in the sequence diagram (Figure 5.5). A low gradient strength of 102.1 Hz/mm was used to reduce
experimental time, and is 0.5% of the maximum available.
5.3.
METHODSAND
MATERIA
LS
99
Figure 5.5: Sequence diagram for the pseudo-continuous wave imaging experiment. The phantom is excited by a series of gapped Rabi modulated excitations. The lineargradients in the x and y directions are applied to take measurements over a range of 2D radial projections.
100 CHAPTER 5. RABI CONTINUOUS WAVE IMAGING
5.3.3.3 Reconstruction
The DC component and first five harmonics of the envelope modulation frequency, ω1, were extracted
from the measured steady-state magnetisation, mxy, and used to construct a measurement vector,
zθ, for each projection. The forward model matrix, H, was constructed from a Fourier series
approximation of the Bloch equation, numerically integrated over a voxel distribution, ρ (δω0). Each
proton density projection, xθ, was solved by least squares optimisation with a nonnegative and
smoothness constraint,
minimisexθ∈[0,∞)
(1− g) ‖Hxθ − zθ ‖2 + g ‖ ∆fd xθ ‖2, (5.13)
where g is a smoothing factor and ∆fd is a finite difference matrix. A two dimensional proton density
image was formed from the radial projections using a standard FBP algorithm.
5.3.4 Rabi modulated imaging contrast
The imaging experiment (Section 5.3.3) used a phantom with homogeneous relaxation, to match the
assumption of average relaxation constants, T av1 and T av
2 , used in the formulation of imaging as an
inverse problem (5.10). An imaging simulation was conducted to test if proton density projections
would be accurately reconstructed in a numerical phantom with multiple relaxation constants, and
to explore the imaging contrast under Rabi modulated excitation.
A two-dimensional numerical phantom of sixteen test tubes (Figure 5.6) has constant proton density
(Figure 5.6c) and a range of relaxation times; T1 relaxation rates of 1000, 2000, 3000, 4000 ms
(Figure 5.6a), T2 relaxation rates of 250, 500, 750, 1000 ms (Figure 5.6b). Imaging signals were
simulated over a 256x256 voxel grid via harmonic balancing of the Bloch equations (2.47) integrated
over a Gaussian voxel distribution, p (δω0), using (4.6). Rabi modulated excitation was simulated
for a grid of N = 500 excitation parameter pairs
(α(1), ω(1)m ) . . . (α(N), ω
(N)m )
as used in the
Rabi spectroscopy experiment (Figure 4.3). The spin system response was generated for 36 radial
projections with five degree increments of the projection angle, θ. Image reconstruction was identical
to the imaging experiment (Section 5.3.3.3) with a discrete proton density reconstruction vector,
xθ, of K = 64 points. The forward model was constructed with the average relaxation constants,
T av1 = 2500 ms and T av
2 = 600 ms. Projections of proton density were reconstructed via optimisation
of objective function (5.13) with no regularisation, g = 0.
5.4 Results
5.4.1 Gapped measurement of off-resonance response
The measured p (δω0) and p (ω1) distributions, used in the generation of the predicted harmonic
curves, are shown in Figure 5.7a and Figure 5.7b, respectively. The variation in the harmonics of
the response to the Rabi modulation frequency are shown over a range of off-resonances (Figure 5.8).
The experimental measurements are very similar to the theoretical curves, with some minor phase
error visible in the measurements at δrf = −320 Hz and δrf = 152 Hz.
5.4. RESULTS 101
x (mm)
y (
mm
)
0 5 10 15 20
0
5
10
15
20 0
1
2
3
4
(a)
x (mm)y
(m
m)
0 5 10 15 20
0
5
10
15
20 0
0.2
0.4
0.6
0.8
1
(b)
x (mm)
y (
mm
)
0 5 10 15 20
0
5
10
15
20 0
0.2
0.4
0.6
0.8
1
(c)
x (mm)
y(m
m)
0 5 10 15 20
0
5
10
15
20 0
0.2
0.4
0.6
0.8
1
(d)
Figure 5.6: Numerical phantom and simulated contrast of Rabi modulated imaging. Numerical phantom of 16 testtubes with spin system properties defined by (a) T1 map (b) T2 map and (c) proton density map. (d) is the FBP ofa sinogram reconstructed from a series of simulated Rabi modulated responses.
102 CHAPTER 5. RABI CONTINUOUS WAVE IMAGING
−400 −200 0 200 4000
1
2
3
4
5
6x 10
−3
p(δ
ω0)
off-resonance, δω 0(Hz)
(a)
0.8 0.9 1 1.1 1.20
0.05
0.1
0.15
0.2
0.25
p(ω
1)
RF amplitude scaling, (×ω1)
(b)
Figure 5.7: Measured distributions of the water phantom used in the gapped measurement of off-resonance experiment.(a) Off-resonance distribution p (δω0 ) extracted from the histogram of a B0 field map. (b) RF power distributionp (ω1) extracted from the histogram of a B1 field map.
5.4.2 Rabi modulated imaging
The reference image (Figure 5.9) is a z projection of proton density and displays three defined test
tubes, aligned along the longitudinal axis. The reference sinogram (Figure 5.10a) and FBP image
(Figure 5.10b) show an expected decrease in image quality due artifacts from the reduced number
of projection angles. Two reconstruction cases are considered for the measurements taken under
Rabi modulated CW excitation; the first assuming B0 homogeneity and the second incorporating
a measured voxel distribution in the forward model. The results assuming B0 homogeneity show
a sinogram (Figure 5.10c) with a consistent artifact through all projections, and a FBP image
(Figure 5.10d) with three clearly defined test tubes but have some error in shape and size. Including
knowledge of the voxel distribution in the forward model removes the artifact from the sinogram
(Figure 5.10e) and leads to a blurred FBP image (Figure 5.10f).
A line profile over one test tube (Figure 5.11), is used to assess the quality of the FBP images
against the reference FLASH image. The line profile for the reference FLASH image (Figure 5.11a)
has excellent SNR, the small signals at the edge of the test-tube are due to water trapped in the
thread of the test-tube cap. The line profile for the FBP of an under-projected reference image
(Figure 5.11b) has lower SNR than the reference image but maintains a strong transition between
the background and the test-tube. The line profile of the Rabi modulated FBP image (Figure 5.11c)
has the lowest SNR and a blurred transition between the background and the test-tube.
5.4.3 Rabi modulated imaging contrast
The imaging results from the simulated Rabi modulated response from a numerical phantom of 16
test tubes are shown in Figure 5.6d and as line profiles in Figure 5.12. The FBP image (Figure 5.6d)
shows 16 test tubes with a range of intensities, there is an artifact in the centre of the image and a
circular streaking artifact. The vertical line profiles (Figure 5.12a) over test tubes with a decreasing
T2 in the positive y direction (Figures 5.12c,5.12e,5.12g and 5.12i) show a related decrease in image
intensity. The horizontal line profiles (Figure 5.12b) over test tubes with a increasing T1 in the
positive x direction (Figures 5.12d,5.12f,5.12h and 5.12j) show a related increase in image intensity.
5.4. RESULTS 103
−400 −200 0 200 400−15
−10
−5
0
5
10
15
δrf(Hz)
Re(cobs
x)
(a)
−400 −200 0 200 400−15
−10
−5
0
5
10
15
δrf(Hz)
Im(c
obs
y)
(b)
Figure 5.8: Harmonic curves of the water phantom under Rabi modulated CW excitation with parameters α = 1and ω1 = ωm = 50 Hz, acquired with a gapped excitation protocol of duty cycle, dcycle = 90% and sampling period,Tgap = 1.67 ms. Measured (circles) and theoretical (solid line) frequency coefficients of steady state magnetisation.DC component (blue), first (green), second (purple), third (red) fourth (orange) and fifth (grey) harmonics.
104 CHAPTER 5. RABI CONTINUOUS WAVE IMAGING
x (mm)
y(m
m)
−20 0 20
−30
−20
−10
0
10
20
30 0
0.2
0.4
0.6
0.8
1
Figure 5.9: Reference image is the average of 10 axial slices acquired with a FLASH protocol. The ring around eachtest tube is due to water trapped in the test tube cap thread.
All line profiles (Figure 5.12) have a non-zero background signal, in the fastest T2 case (Figure 5.12j)
the magnitude of test tube signal is similar to the background signal.
5.5 Discussion
The measurement of off-resonance spins under power corrected gapped Rabi modulated excitation
(Figure 5.8) verifies that the CW spin system response can be observed under a gapped excitation
and measurement protocol. Variations between prediction and measurement can be attributed to
cumulative error in the prediction model, H, from relaxation and field distribution measurements.
There is also a noticeable phase jump in the measurements at δrf = −320, 152 Hz which may indicate
some mild instability in phase coherence between measurements. This discrepancy between model
and measurement may also be an artifact related to the gapped measurement protocol (Idiyatullin
et al., 2008).
The gapped protocol offers a significant efficiency improvement over the iterative protocol used in
Chapter 4 and Chapter 6 which makes observation of the transient response feasible for future
studies. For an experiment where steady-state is assumed after 5T1 and the same time is allowed
for relaxation to equilibrium in each repetition, the approximate experimental duration for iterative
and gapped protocols is,
T iterN = 20N T1Nharm (5.14)
T gapN = 10N T1 (5.15)
where N is the number of steady-states to be measured and Nharm is the number of measured steady-
state harmonics. In our steady-state experiments, the efficiency is improved by a factor of twelve
when acquiring enough gapped data for six harmonics. The theoretical limit, assuming steady-state
after 5T1 and smooth transitions between different steady-state trajectories can be achieved, gives
5.5. DISCUSSION 105
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
(a)
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
(b)θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1(c)
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1(d)
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
(e)
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
θ()
r(m
m)
0 50 100 150
−40
−20
0
20
400
0.2
0.4
0.6
0.8
1
0
0.2
0.4
0.6
0.8
1
(f)
Figure 5.10: (a) Reference FLASH sinogram (b) FBP of reference FLASH sinogram (c) Rabi CW sinogram (d) FBPof Rabi CW sinogram (e) Rabi CW sinogram with voxel distribution in forward model (f) FBP of Rabi CW sinogramwith voxel distribution in forward model.
106 CHAPTER 5. RABI CONTINUOUS WAVE IMAGING
(a) (b) (c)
−30 −20 −10 0 10 20 300
0.5
1
x (mm)
ρ(δ
ω0)
(d)
Figure 5.11: Line profile (red line) marked on (a) Reference FLASH image (b) FBP of under-projected referenceFLASH sinogram (c) FBP of Rabi CW sinogram. (d) Line profile of proton density from reference FLASH image(thick line), FBP of under-projected reference FLASH sinogram (thin line) and FBP of Rabi CW sinogram (dashedline).
5.5. DISCUSSION 107
y(m
m)
x (mm)0 5 10 15 20
0
5
10
15
20
(a)
y(m
m)
x (mm)0 5 10 15 20
0
5
10
15
20
(b)
0 2 4 6 8 10 12 14 16 18 200
0.5
1
y (mm)
ρ(δ
ω0)
(c)
0 2 4 6 8 10 12 14 16 18 200
0.5
1
x (mm)
ρ(δ
ω0)
(d)
0 2 4 6 8 10 12 14 16 18 200
0.5
1
y (mm)
ρ(δ
ω0)
(e)
0 2 4 6 8 10 12 14 16 18 200
0.5
1
x (mm)
ρ(δ
ω0)
(f)
0 2 4 6 8 10 12 14 16 18 200
0.5
1
y (mm)
ρ(δ
ω0)
(g)
0 2 4 6 8 10 12 14 16 18 200
0.5
1
x (mm)
ρ(δ
ω0)
(h)
0 2 4 6 8 10 12 14 16 18 200
0.5
1
y (mm)
ρ(δ
ω0)
(i)
0 2 4 6 8 10 12 14 16 18 200
0.5
1
x (mm)
ρ(δ
ω0)
(j)
Figure 5.12: Line profiles of simulated contrast of Rabi modulated imaging. (a) Vertical line profiles (red lines) alongtest-tubes with constant T1 and a decreasing T2 in the positive y direction. The vertical red lines shown in (a) areplotted from left to right as (c),(e),(g) and (i). (b) Horizontal line profiles (red lines) along test-tubes with constantT2 and a increasing T1 in the positive x direction. The horizontal red lines shown in (b) are plotted from top tobottom as (d),(f),(h) and (j).
108 CHAPTER 5. RABI CONTINUOUS WAVE IMAGING
a best case experimental time
T limN = 10T1 +N
2π
ωavm
(5.16)
where ωavm is the average Rabi modulation of the set of N excitation parameters. Using the imaging
experiment parameters, data acquisition for a single projection is, T iterN = 5.5 hours, using the
iterative protocol, T gapN = 28 minutes, using the gapped protocol and T lim
N = 1.1 minutes, using
a theoretical ideal protocol. These experimental times are to encode a chemical shift spectrum or
a single projection of proton density, for a radial imaging experiment these are multiplied by the
number of projections to calculate the total experimental time.
The results of the radial imaging experiment (Figure 5.10 and Figure 5.11) demonstrate that Rabi
modulated excitation can encode off-resonant information, in this case off-resonance induced by
linear field gradients. The gradients used in the imaging experiment lead to a larger off-resonance
bandwidth to be encoded than the spectroscopy experiment of Chapter 4. This was achieved by
shifting the encoding function across the off-resonance range using the RF carrier offset δrf.
The image (Figure 5.10b) reconstructed from the reference sinogram (Figure 5.10a) shows an ex-
pected streak artifact from the reduced number of projections. Rabi modulated proton density
projections reconstructed with a homogeneous voxel form the sinogram (Figure 5.10c) which is
comparable to the reference sinogram (Figure 5.10a). The horizontal lines in the Rabi modulated
sinogram (Figure 5.10c) may be due to uneven coverage of the encoding function, this may be im-
proved by a finer increment of δrf. Another potential source of the horizontal line artifact in the
sinogram (Figure 5.10c) is related to the use of a gapped excitation envelope, which causes a bullseye
artifact in SWIFT (Idiyatullin et al., 2008) and can be removed by a gap cycling strategy during
acquisition (Corum et al., 2015). The FBP image (Figure 5.10d) from homogeneous voxel sinogram
(Figure 5.10c), shows three test tubes with a similar back projection artifact to the reference image
(Figure 5.10b) and error in the diameter of the test tubes. Rabi modulated proton density projec-
tions reconstructed with a measured voxel distribution in the forward model, H, form a smooth
sinogram (Figure 5.10e) without the horizontal artifact seen in reconstructions assuming a homoge-
neous voxel (Figure 5.10c) and mildly improves the size of the test tubes in the reconstructed image
(Figure 5.10f).
The image line profiles (Figure 5.11d) show an expected loss of SNR in the FBP cases (Figure 5.11b
and 5.11c) due to the under projected streak artifact. The line profile of the Rabi modulated FBP
(Figure 5.11c) fails to reconstruct the sharp edge of the test tube as shown in the reference image
(Figure 5.11a). This loss of high frequency information may be an effect of regularisation in the
reconstruction objective function (5.13) or due to uncorrected B0 field inhomogeneity effects.
The contrast simulation results (Figure 5.12) demonstrate that an object with variable relaxation
rates can be reconstructed using a forward model computed with two average relaxation rates, T av1
and T av2 . In addition, the intensity of the FBP image with be weighted by the actual relaxation
rates. The line profiles showed two general image intensity trends; a faster T1 constant gives a higher
image intensity, a slower T2 constant gives a higher image intensity.
Filtered back projection was an adequate method for this experiment, better reconstruction tech-
niques such as the iterative methods used in CT may lead to a better reconstructed image quality.
We anticipate that further improvements to experimental efficiency will allow the acquisition of more
excitation parameters under a higher gradient and lead to more accurate reconstruction, without
prior knowledge of a voxel distribution.
5.6. CONCLUSION 109
5.6 Conclusion
This chapter introduces a gapped excitation and measurement protocol which can be used to study
spin-system response to periodic excitation. The greater efficiency of the gapped protocol made an
imaging experiment feasible and will allow future investigation of the transient spin-system response
under Rabi modulated excitation. The proof of concept imaging experiment demonstrates that
under Rabi modulated CW excitation, gradient localisation information can be encoded in the
steady-state magnetisation and used to reconstruct proton density images. Rabi modulated imaging
was simulated for a numerical phantom with variable relaxation rates and reconstructed images show
T1 and T2 contrast. Future work will focus on improving the efficiency of acquisition, the accuracy
of reconstruction and evaluation of the methods ability to image samples with ultra-short spin-spin
relaxation.
110 CHAPTER 5. RABI CONTINUOUS WAVE IMAGING
Chapter 6
Ring-lock excitation
Contents6.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
6.2 Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 113
6.2.1 Steady-state ring-lock response . . . . . . . . . . . . . . . . . . . . . . . . 114
6.2.2 Estimation of relaxation constants . . . . . . . . . . . . . . . . . . . . . . 114
6.3 Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
6.3.1 Verification of excitation envelope . . . . . . . . . . . . . . . . . . . . . . 115
6.3.2 Estimation of relaxation constants . . . . . . . . . . . . . . . . . . . . . . 118
6.4 Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 118
6.5 Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
6.6 Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Appendices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
6.A Ring-lock excitation envelope . . . . . . . . . . . . . . . . . . . . . . . . 122
6.A.1 Amplitude and phase . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
6.B Spin-system response under ring-lock excitation . . . . . . . . . . . . . 125
6.1 Introduction
In previous chapters 4 and 5, we explored potential applications of the steady-state magnetisation
under continuous wave periodic RF excitation, in particular Rabi modulated excitation. In this
chapter, we revisit a well known steady-state response of the Bloch equations, with the addition
of off-resonance. Under continuous wave, constant amplitude RF excitation, the Bloch equations
predict that the bulk magnetisation will reach a steady-state magnetisation on an ellipsoid (Bloch,
1946). The steady-state ellipsoid is located in the upper hemisphere of the Bloch sphere (Figure 6.1).
The power of the constant amplitude excitation determines the location of the steady-state magneti-
sation on this manifold. A very low power excitation will perturb the magnetisation only slightly
from thermal equilibrium, as power is increased the steady-state magnetisation will move down the
ellipsoid, approaching saturation at the centre of the Bloch sphere. The ability to saturate the
NMR signal can be used to estimate spin-spin relaxation rates with saturation recovery (Markley
111
112 CHAPTER 6. RING-LOCK EXCITATION
(a) (b) (c)
Figure 6.1: Steady-state ring trajectory (black ring) on steady-state manifold (purple ellipsoid) in the Bloch sphere(grey sphere). Trajectories are shown for a (a) low power case (b) slightly higher power case (c) slower ring-lockfrequency with the same power as (b).
et al., 1971) and partial saturation (Freeman and Hill, 1971) methods. Unwanted solvent peaks
are saturated in NMR spectroscopy (Jesson et al., 1973) and similar frequency selective saturation
methods (Bottomley et al., 1984; Haase et al., 1985) are used to saturate water and lipid signals in
chemical shift imaging.
When a constant amplitude RF excitation is applied off-resonance, the magnetisation is driven into
a steady-state ring trajectory (Figure 6.1), on the previously described ellipse. The off-resonance is
achieved by amplitude modulating the excitation envelope, or by frequency offset of the transmitter
RF carrier. An alternative way to consider the response, is viewing a spin-lock (Redfield, 1955;
Hartmann and Hahn, 1962) response from an off-resonance frame of reference, with the magnetisation
decaying to a ring rather than a point. Not a completely fair comparison, as spin-lock applications
focus on the transient dynamics for T1ρ estimation or contrast (Sepponen et al., 1985; Santyr et al.,
1989, 1994), and are generally not applied for a long enough duration to reach steady-state. In this
chapter, we refer to an off-resonance constant amplitude envelope as ring-lock excitation, named
for the ring shaped steady-state trajectory of the spin-system; and in recognition of similarities to
spin-lock excitation. It is most useful though to consider the excitation as an off-resonance partial
saturation pulse, where the saturation level is influenced by both the amplitude and the frequency
offset.
The ring-lock excitation envelope can control the steady-state magnetisation on an ellipsoid, with
adequate knowledge of the spin-system properties, such as relaxation rates. We demonstrate this
control experimentally and develop a model based method for estimating the ratio of relaxation
constants (Korte et al., 2017b). In our method, we apply off-resonance partial saturation pulses and
derive an analytical steady-state signal model, which is dependent on the excitation off-resonance,
in addition to the relaxation constants.
The shape of the steady-state manifold is dependant on relaxation constants, as shown in Fig-
ure 6.2. We exploit the relationship between excitation parameters and unknown relaxation con-
stants to develop a volume relaxometry technique. Existing steady-state free precession (Carr,
1958) (SSFP) methods such as IR-TrueFISP (Scheffler and Hennig, 2001; Schmitt et al., 2004),
DESPOT (Homer and Beevers, 1985; Deoni et al., 2003), DESS (Bruder et al., 1988; Welsch et al.,
2009) and TESS (Heule et al., 2014) generate relaxation maps via optimisation of signal models,
dependent on relaxation constants T1 and T2. We plan to develop our volumetric ring-lock relaxom-
etry technique into a relaxation mapping method, utilising the elliptical nature (Hargreaves, 2012)
of balanced steady-state free precession (Oppelt et al., 1986) (bSSFP).
6.2. THEORY 113
(a) (b) (c)
Figure 6.2: Steady-state ring trajectory (black ring) on steady-state manifold (purple ellipsoid) in the Bloch sphere(grey sphere). Steady-state ellipsoids are shown for three different tissue relaxation parameters (a) white matter (b)lipid (c) cerebrospinal fluid.
6.2 Theory
A spin-system under constant amplitude RF excitation will reach a steady-state magnetisation on
an ellipsoid (Bloch, 1946) whose shape is defined by the relaxation constants and is influenced by
off-resonance effects and imperfect excitation (Abragam, 1961). We introduce a constant frequency
modulation, ωrl, to the constant amplitude excitation, |ωe1|, which induces a ring-locked steady-state
trajectory on the surface of the ellipsoid (Figure 6.3a). The ring-lock excitation envelope is defined
as,
γBe1 (t) = |ωe
1| eiφ(t) (6.1)
where
|ωe1| =
√T2
T1
(1− a1 + a
)[(ωrl + ∆)
2+
(1
T2
)2 ](6.1a)
φ (t) = ωrlt+ θ − tan−1
(1
T2 (ωrl + ∆)
)+π
2. (6.1b)
The amplitude and phase of the envelope are expressing in terms of the estimated spin parameters
and the desired response parameters. The estimated spin parameters are relaxation, T1, T2, and
off-resonance, ∆. The desired ring response parameters are phase offset, θ, and elevation on the
steady-state ellipsoid, a ∈ (−1, 1]. The desired elevation, a, is related to the longitudinal component
of the steady-state magnetisation (6.2c); at thermal equilibrium, a = 1, the magnetisation becomes
saturated as, lima→−1
. When the estimated spin parameters used in the ring-lock envelope are accurate,
the spin-system response is,
mx =M0
z
2
√T2
T1(1− a2) sin (ωrlt+ θ) (6.2a)
my =M0
z
2
√T2
T1(1− a2) cos (ωrlt+ θ) (6.2b)
mz =M0
z
2(1 + a) . (6.2c)
114 CHAPTER 6. RING-LOCK EXCITATION
6.2.1 Steady-state ring-lock response
The estimated tissue parameters may differ from the actual tissue parameters for relaxation T1, T2,
and off-resonance ∆, which leads to a different ring-lock response with parameters for elevation, a,
and phase offset, θ. The difference between the desired steady-state trajectory parameters, (a, θ),
and the observed steady-state trajectory parameters,(a, θ), is also influenced by an excitation
field inhomogeneity factor, Bmod1 . The observed steady-state trajectory, in response to ring-lock
excitation (6.1), is defined as,
mx =M0
z
2
√T2
T1(1− a2) sin
(ωrlt+ θ
)(6.3a)
my =M0
z
2
√T2
T1(1− a2) cos
(ωrlt+ θ
)(6.3b)
mz =M0
z
2(1 + a) (6.3c)
where observed elevation is,
a(a, ωrl, T1, T2,∆, T1, T2, ∆, B
mod1
)=
1− β1 + β
(6.3d)
β = Bmod1
2 T1 T2
T1 T2
(1− a1 + a
) (ωrl + ∆)2 +(
1T2
)2
(ωrl + ∆)2 +(
1T2
)2
(6.3e)
and observed phase is,
θ = θ − tan−1
(1
T2 (ωrl + ∆)
)+ tan−1
(1
T2
(ωrl + ∆
)). (6.3f)
For derivation details of the ring-lock excitation envelope and spin-system steady-state response
see Appendix 6.A and Appendix 6.B, respectively. We have verified this analytical solution using
numerical simulations (Figure 6.4) and harmonic balancing (Tahayori et al., 2015) of the Bloch
equations, an experimental verification is detailed in the methods and results sections.
6.2.2 Estimation of relaxation constants
If we consider a case when parameters for the ring-lock envelope amplitude (6.1a) are not known, we
can treat the amplitude, |ωe1|, as an adjustable power. The elevation on the steady-state ellipsoid,
a, can then be expressed in terms of the power, |ωe1|, and frequency, ωrl, of the ring-lock envelope,
a(|ωe
1|, ωrl, T1, T2, ∆, Bmod1
)=
1− β1 + β
(6.4a)
β =T1
T2
|ωe1|2Bmod
12(
ωrl + ∆)2
+(
1T2
)2
. (6.4b)
This relationship (6.4) gives an observed magnetisation,
mmodelxy
(|ωe
1|, ωrl, T1, T2, t)
= . . . (6.5)∫∫mxy
(|ωe
1|, ωrl, T1, T2, ∆ (δω0) , Bmod
1 , t)p(Bmod
1
)p (δω0
) dBmod1 dδω0
.
6.3. METHODS 115
which is influenced by unknown tissue parameters, T1, T2, and excitation parameters, |ωe1|, ωrl. Re-
laxation constants can therefore be estimated from a series of ring-lock responses, acquired under a
range of excitation parameters, |ωe1|, ωrl, using the objective function,
minimiseT1, T2∈[0,∞)
‖ ∣∣mmodelxy
(|ωe
1|, ωrl, T1, T2
)∣∣− ∣∣mobsxy (|ωe
1|, ωrl)∣∣ ‖
2. (6.6)
6.3 Methods
Two experiments were conducted on a 4.7T Bruker Biospec scanner with an AVANCE III console, to
test the ring-lock excitation envelope. The first to experimentally verify the analytical response of the
spin-system (6.2) under ring-lock excitation; using measured spin-system properties to control the
magnetisation on the surface of the steady-state ellipsoid. A second, proof-of-concept experiment,
demonstrates the estimation of relaxation constants from a series of steady-state measurements,
acquired under a range of ring-lock excitation envelopes.
In both experiments, an iterative measurement protocol (Layton et al., 2014; Korte et al., 2017a)
(Figure 6.3b) was used to incrementally measure the steady-state magnetisation. Rapid acquisition
with relaxation enhancement with variable repetition time (RARE-VTR) scans and multi-slice multi-
echo (MSME) scans were used to measure T1 and T2 maps respectively. The measured relaxation
maps were spatially averaged to produce a single T1 and T2 constant per phantom. The distri-
bution of off-resonances, p (δω0), was measured using a field-mapping sequence (Kanayamay et al.,
1996). The distribution of excitation field strength, p(Bmod
1
), was measured with a B1 mapping
sequence (Stollberger et al., 1988). Distributions were extracted via a histogram of non-background
voxels.
6.3.1 Verification of excitation envelope
The first experiment applied an excitation with the ring-lock envelope (6.1) to a spherical phantom
of water to demonstrate control of the steady-state magnetisation. Ring-locked trajectories were
measured over a range of elevations, a, from -0.95 to 1.0 with a 0.05 increment, using a constant
excitation frequency, ωrl = 50 Hz, measured relaxation constants, T1 = 3.20 s and T2 = 2.02
s, and off-resonance ∆ = 0 Hz. For each measurement point, two incremental acquisitions were
required, the first as shown in Figure 6.3b to acquire the transverse magnetisation, the second with
a hard pulse directly before the FID to measure the longitudinal magnetisation. The predicted
response was generated by numerically integrating the analytical response (6.5) over the measured
field distributions, incorporating the measured relaxation constants.
The distribution of off-resonances, p(δω0), was measured using a B0 field mapping sequence (Kanaya-
may et al., 1996) described in Section 3.4.1. Data was acquired using a multiple gradient echo se-
quence with effective echo times (TE = 1.94, 6.22 ms), FA = 20, TR = 20 ms, FOV = 4.5 cm and
64×64×64 matrix. The distribution of excitation field strengths, p(Bmod
1
), was measured with a B1
mapping sequence (Stollberger et al., 1988) described in Section 3.4.2. Data was acquired with a
MSME sequence over 18 slices, 2 mm thickness, FOV = 4.5 cm, 64×64 matrix, TE = 12 ms, TR = 2.5
s. The MSME scans were taken for two excitation angle, α, and refocusing angle, β, configurations;
α/β = 45/90 in the first scan, α/β = 90/180 in the second scan.
The average T1 relaxation rate was calculated from the spatial average of T1 maps, which were
measured as described in Section 3.3.1. Data was acquired with a RARE-VTR sequence over 1 slice,
116 CHAPTER 6. RING-LOCK EXCITATION
(a)
(b)
Figure 6.3: (a) Simulated spin-system response of the magnetisation (dashed line) approaching a steady-state ring-trajectory on an ellipse (purple) in the upper half of the Bloch sphere (grey). (b) The iterative excitation protocolapplies the excitation envelope, γBe
1 (t), for 5T1 until the spin-system has reached steady-state, after which a freeinduction decay (FID) is measured. A steady-state magnetisation waveform, mobs
xy , can be acquired by selecting thefirst FID point, and repeating the process with an incremental increase to the excitation duration on each repetition.The longitudinal magnetisation is measured in the same manner, with the addition of a hard pulse directly beforeeach FID.
6.3. METHODS 117
0 1 2 3 4 5 6−1
0
1
t(s)
mx
(a)
5.96 5.97 5.98 5.99 6−0.2
0
0.2
t(s)
mx
(b)
0 1 2 3 4 5 6−1
0
1
t(s)
my
(c)
5.96 5.97 5.98 5.99 6−0.2
0
0.2
t(s)
my
(d)
0 1 2 3 4 5 6−1
0
1
t(s)
mz
(e)
5.96 5.97 5.98 5.99 60
0.5
1
t(s)
mz
(f)
Figure 6.4: Numerical validation of excitation envelope with (a),(c),(e) simulated magnetisation from numericalintegration of Bloch equations. (b),(d),(f) steady-state magnetisation under ring-lock excitation with a comparisonof numerically integrated magnetisation (line) and magnetisation predicted by analytical solution of Bloch equations(circles). The excitation parameters used were ring-lock frequency, ωrl = 50Hz, desired elevation, a = 0.2, estimatedrelaxation constants, T1 = 500ms, T2 = 250ms, and estimated off-resonance, ∆ = 20Hz. Spin parameters used wererelaxation constants, T1 = 1200ms, T2 = 300ms, and off-resonance, ∆ = −10Hz.
118 CHAPTER 6. RING-LOCK EXCITATION
4 cm thickness, FOV = 4 cm, 32×32 matrix, TE = 11 ms, TR = 50, 100, 200, 400, 800, 1500, 3000,
6000, 9000, 12000, 15000 ms, RARE factor 1. The average T2 relaxation rate was calculated from the
spatial average of T2 maps, which were measured as described in Section 3.3.2. Data was acquired
with a MSME sequence using 128 echoes with a 50 ms echo spacing, 1 slice, 8 mm thickness, FOV
= 4 cm, 32×32 matrix, TR = 13.2 s.
6.3.2 Estimation of relaxation constants
In a second experiment, we measured the transverse steady-state magnetisation, mobsxy , of a spherical
phantom of Gadolinium doped water (T1 = 360 ms, T2 = 200 ms). Steady-state measurements were
acquired under RF excitation with the ring-lock envelope (6.1), over a grid of excitation parameters;
excitation amplitude, |ωe1|, from 100 Hz to 1800 Hz with an increment of 100 Hz and ring-lock
frequency, ωrl, from 25 Hz to 400 Hz with a 25 Hz increment. Relaxation constants were estimated
from the grid of observations, by solving the objective function (6.6) with a constrained non-linear
optimisation algorithm, fmincon, from MATLAB R©. In addition to the estimation from experimental
measurements, the relaxation constants were estimated from a “perfect” observation, generated from
the observation model (6.5) using the relaxation constants measured with standard methods. The
relaxation constants estimated from experimental measurement were used to generate a predicted
response surface, by numerically integrating the analytical response (6.5) over the measured field
distributions.
The distribution of off-resonances, p(δω0), was measured using a B0 field mapping sequence (Kanaya-
may et al., 1996) described in Section 3.4.1. Data was acquired using a multiple gradient echo
sequence with effective echo times (TE = 1.94, 6.22 ms), FA = 20, TR = 20 ms, FOV = 4.5 cm
and 128×128×128 matrix. The distribution of excitation field strengths, p(Bmod
1
), was measured
with a B1 mapping sequence (Wang et al., 2005) described in Section 3.4.2. Data was acquired
with a MSME sequence over 18 slices, 2 mm thickness, FOV = 4.5 cm, 128×128 matrix, TE = 12
ms, TR = 2.5 s. The MSME scans were taken for two excitation angle, α, and refocusing angle, β,
configurations; α/β = 45/90 in the first scan, α/β = 90/180 in the second scan.
The average T1 relaxation rate was calculated from the spatial average of T1 maps, which were
measured as described in Section 3.3.1. Data was acquired with a RARE-VTR sequence over 1
slice, 2 mm thickness, FOV = 3 cm, 256×256 matrix, TE = 10 ms, TR = 50, 100, 200, 400, 800,
1600, 3200, 6400 ms, RARE factor 2. The average T2 relaxation rate was calculated from the spatial
average of T2 maps, which were measured as described in Section 3.3.2. Data was acquired with a
MSME sequence using 64 echoes with a 10 ms echo spacing, 1 slice, 2 mm thickness, FOV = 3 cm,
256×256 matrix, TR = 5.0 s.
6.4 Results
Verification of the excitation envelope and steady-state response are the shown in Figure 6.5. The
steady-state measurements are purple and the analytically predicted steady-state trajectories are
shown in black. The steady-state trajectories (Figure 6.5a-6.5c) form ellipsoids viewed from a range
of angles. In Figure 6.5b the elevation, or longitudinal magnetisation, mz, of measured steady-state
ring trajectories is consistent with the model prediction. In Figure 6.5c the phase of measured
steady-state ring trajectories is consistent with the model prediction. The average magnitude of the
transverse magnetisation, or each ring-lock response is shown in Figure 6.5d, with minor error (3.1%
M0z ) between measurement and model prediction.
6.4. RESULTS 119
−0.5
0
0.5 −0.5
0
0.50
0.2
0.4
0.6
0.8
1
mymx
mz
(a)
−0.5 0 0.50
0.2
0.4
0.6
0.8
1
mx
mz
(b)
−0.5 0 0.5−0.5
0
0.5
mx
my
(c)
0
0.2
0.4
0.6
0.8
1
0 0.1 0.2 0.3 0.4 0.5
mean(|mxy|)
mean(m
z)
(d)
Figure 6.5: Experimental validation of excitation envelope using predicted (black) and measured (purple) steady-statering-trajectory response, displayed in (a) three-dimensional (b) side-on and (c) top-down views. The mz and mxy
components of each ring trajectory are averaged to generate (d) a mean profile of the ellipse.
120 CHAPTER 6. RING-LOCK EXCITATION
(a) (b)
Figure 6.6: (a) The transverse steady-state magnetisation, predicted (black) and measured (purple) over the excitationparameter space. (b) The error surface of the difference between measured and predicted transverse magnetisation(purple) and a reference plane at zero (black).
Table 6.1: Measured and estimated relaxation constants.
T1 (ms) T2 (ms) T1/T2 ratio
REFERENCE
RARE-VTR 360.23±11.84 -1.8057
MSME - 199.50±2.60
METHOD
Ideal surface 360.20 199.48 1.8057
Measured surface 97.19 53.81 1.8059
The estimation of relaxation constants results are visualised in Figure 6.6. The magnitude of trans-
verse steady-state magnetisation, |mxy|, is shown as a surface across a grid of ring-lock envelope
parameters (Figure 6.6a), where measurements are purple and the black surface is generated using
the relaxation parameters estimated via an optimisation routine. Figure 6.6b is a surface represent-
ing error between the transverse magnetisation and the predicted transverse magnetisation. The flat
black surface can be considered a zero error reference, the purple surface is the error between mea-
sured and predicted magnitude of the transverse magnetisation; maximum error is approximately
4.1% M0z .
The relaxation constants from standard measurements and estimations from ring-lock response,
are shown in Table 6.1. The reference measurements for spin-lattice relaxation, T1 = 360 ms,
and for spin-spin relaxation, T2 = 200 ms, giving a relaxation ratio, T1/T2 = 1.8. Relaxation
constants estimated from the simulated “perfect” observation, T1 = 360 ms, T2 = 200 ms, match
those measured with standard methods. Relaxation constants estimated from the experimental
observations, T1 = 97 ms, T2 = 54 ms, do not match those measured with standard methods, but
the relaxation ratio is accurate, T1/T2 = 1.8.
6.5. DISCUSSION 121
6.5 Discussion
The results of the first experiment (Figure 6.5), to control the magnetisation on the steady-state
ellipsoid are generally in good agreement with the analytic prediction integrated over the measured
field distributions. The linearly increasing phase of the measurements shows the steady-state ring
trajectory is locked to the envelope frequency. The similarity between measured and predicted
elevation, demonstrates that with known spin-system parameters the ring-lock envelope can drive
the steady-state response to a desired elevation. The minor error in the transverse steady-state
remains an open question; the measured transverse response is very similar to what we would expect
from a single on-resonance isochromat, as opposed to our prediction model which is integrated over
measured field distributions. Exploratory measurements, not shown here, also show unexpected
change in the ellipse profile dependant on the ring-lock frequency used, which could possibly indicate
magnetisation transfer effects. This will be investigated in future work, using distilled rather than
tap water.
The results from the relaxometry experiment (Table 6.1) demonstrate that with an ideal mea-
surement that both relaxation constants can be accurately estimated. The optimisation algorithm
estimates an accurate relaxation ratio from experimental measurements, but inaccurately estimates
the individual relaxation rates. If we consider a term used to calculate the elevation response of the
spin-system,
β =T1
T2
|ωe1|2Bmod
12(
ωrl + ∆)2
+(
1T2
)2
, (6.4b)
which is dominated by the relaxation ratio, T1/T2, when sources of off-resonance,(ωrl + ∆
), are
larger than the inverse spin-spin relaxation, 1/T2. The ring-lock frequencies used for this experiment
ranged from 25 Hz to 400 Hz, which were selected to suppress effects of sample off-resonance, ∆, and
are larger than the inverse spin-spin relaxation, 1/T2 = 5 Hz. This leads to an optimisation surface
with multiple, objectively similar local minima, for the optimisation algorithm to converge to. The
error between measured and optimised surfaces (Figure 6.6b), which is similar to that observed in
the first experiment, may also have contributed to the optimisation algorithm not converging on the
global minimum.
Hargreaves (2012) depicts the bSSFP trajectory on an ellipse, where elevation is dependant on the
flip angle and repetition time. If we consider the pulsed bSSFP excitation envelope as the power
corrected gapped equivalent of a continuous wave constant amplitude excitation, then we can relate
the flip angle and repetition time to the analytical response model developed in this chapter; with
the ring-lock frequency, ωrl, set to zero. If we can maintain a bSSFP signal using an off-resonance
gapped ring-lock excitation, then the presented relaxometry method can also be implemented for
relaxation mapping. The off-resonance excitation properties of the ring-lock envelope may make slice
selection challenging, so a three dimensional bSSFP sequence will be aim for a first implementation.
6.6 Conclusion
The proposed excitation envelope is able to control the steady-state ring-locked magnetisation tra-
jectories. A model for steady-state magnetisation response under this excitation was applied as an
alternate technique for estimation of the ratio of relaxation constants. Future work will focus on
reducing measurement error and application of this excitation envelope to a balanced steady-state
free precession relaxation parameter mapping method.
122
CHAPTER
6.RIN
G-LOCK
EXCITATIO
N
Appendix
6.A Ring-lock excitation envelope
A possible trajectory on the surface of the steady-state elipsoid (2.38) is a ring on a transverse plane anywhere on the ellipsoid,
mx =M0
z
2
√R1
R2(1− a2) sin (ωrlt+ θ) (6.A.1a)
my =M0
z
2
√R1
R2(1− a2) cos (ωrlt+ θ) (6.A.1b)
mz =M0
z
2(1 + a) (6.A.1c)
where, a ∈ (−1, 1], represents elevation in the top hemisphere of the Bloch sphere, ωrl, is rotational frequency of the trajectory and θ is a phase
offset. The derivatives of the trajectory (6.A.1) are,
mx = ωrlM0
z
2
√R1
R2(1− a2) cos (ωrlt+ θ) (6.A.2a)
my = −ωrlM0
z
2
√R1
R2(1− a2) sin (ωrlt+ θ) (6.A.2b)
mz = 0. (6.A.2c)
Substitution of the trajectory (6.A.1) and its derivative (6.A.2) into the Bloch equations (2.34) yelids
ωrlM0
z
2
√R1
R2(1− a2) cos (ωrlt+ θ) = −R2
M0z
2
√R1
R2(1− a2) sin (ωrlt+ θ)−∆
M0z
2
√R1
R2(1− a2) cos (ωrlt+ θ) + ωe
1,y
M0z
2(1 + a)
(6.A.3a)
−ωrlM0
z
2
√R1
R2(1− a2) sin (ωrlt+ θ) = ∆
M0z
2
√R1
R2(1− a2) sin (ωrlt+ θ)−R2
M0z
2
√R1
R2(1− a2) cos (ωrlt+ θ)− ωe
1,x
M0z
2(1 + a)
(6.A.3b)
0 = −ωe1,y
M0z
2
√R1
R2(1− a2) sin (ωrlt+ θ) + ωe
1,x
M0z
2
√R1
R2(1− a2) cos (ωrlt+ θ) . . .
−R1M0
z
2(1 + a) +R1M
0z (6.A.3c)
6.A
.RIN
G-LOCK
EXCIT
ATIO
NENVELOPE
123
re-arranging (6.A.3a),(6.A.3b) for the excitation envelope,
ωe1,x =
1
1 + a
√R1
R2(1− a2)
([∆ + ωrl] sin (ωrlt+ θ)−R2 cos (ωrlt+ θ)
)(6.A.4a)
ωe1,y =
1
1 + a
√R1
R2(1− a2)
([∆ + ωrl] cos (ωrlt+ θ) +R2 sin (ωrlt+ θ)
). (6.A.4b)
Substitution of the the excitation envelope (6.A.4) into the longitudinal trajectory (6.A.3c) gives,
0 = − 1
1 + a
√R1
R2(1− a2)
([∆ + ωrl] cos (ωrlt+ θ) +R2 sin (ωrlt+ θ)
)M0
z
2
√R1
R2(1− a2) sin (ωrlt+ θ) . . .
+1
1 + a
√R1
R2(1− a2)
([∆ + ωrl] sin (ωrlt+ θ)−R2 cos (ωrlt+ θ)
)M0
z
2
√R1
R2(1− a2) cos (ωrlt+ θ) . . .
−R1M0
z
2(1 + a) +R1M
0z (6.A.5)
0 = − 1
1 + a
(R1
R2
(1− a2
))M0z
2
([∆ + ωrl] cos (ωrlt+ θ) sin (ωrlt+ θ) +R2 sin2 (ωrlt+ θ)
). . .
+1
1 + a
(R1
R2
(1− a2
))M0z
2
([∆ + ωrl] sin (ωrlt+ θ) cos (ωrlt+ θ)−R2 cos2 (ωrlt+ θ)
). . .
−R1M0z
(1
2(1 + a)− 1
)(6.A.6)
R1M0
z
2(a− 1) =
−R2
1 + a
(R1
R2
(1− a2
))M0z
2
(sin2 (ωrlt+ θ) + cos2 (ωrlt+ θ)
)(6.A.7)
(1− a) =1
1 + a
(1− a2
)(cos2 (ωrlt+ θ) + sin2 (ωrlt+ θ)
)(6.A.8)
(1− a) (1 + a)
(1− a2)= cos2 (ωrlt+ θ) + sin2 (ωrlt+ θ) (6.A.9)
1 = cos2 (ωrlt+ θ) + sin2 (ωrlt+ θ) (6.A.10)
leads to the well known Pythagorean trigonometric identity (6.A.10). This demonstrates that the excitation envelope (6.A.4) derived from the
transverse magnetisation components (6.A.3a), (6.A.3b) also meet the constraint of the longitudinal magnetisation (6.A.3c).
124
CHAPTER
6.RIN
G-LOCK
EXCITATIO
N
6.A.1 Amplitude and phase
We express the excitation envelope in amplitude and phase terms, for simplicity in power calculations and sequence programing. Using iden-
tity (6.A.11),
A cos (ωrlt+ θ) +B sin (ωrlt+ θ) =√A2 +B2 cos
(ωrlt+ θ − tan−1 B
A
)(6.A.11)
we can rewrite equations (6.A.4a) and (6.A.4b) as
ωe1,x = β
√R2
2 + (∆ + ωrl)2 cos
(ωrlt+ θ − tan−1
(∆ + ωrl
−R2
))(6.A.12a)
ωe1,y = β
√(∆ + ωrl)2 +R2
2 cos
(ωrlt+ θ − tan−1
(R2
∆ + ωrl
))(6.A.12b)
where,
β =
√R1
R2
(1− a1 + a
). (6.A.13)
Apply the identities,
tan−1
(1
x
)= − tan−1 (x) +
π
2for: x > 0 (6.A.14a)
tan−1
(1
x
)= − tan−1 (x)− π
2for:x < 0 (6.A.14b)
and,
tan−1 (−x) = − tan−1 (x) (6.A.15)
to (6.A.12) gives,
ωe1,x = β
√(ωrl + ∆)2 +R2
2 cos
(ωrlt+ θ − tan−1
(R2
∆ + ωrl
)+π
2
)(6.A.16a)
ωe1,y = β
√(ωrl + ∆)2 +R2
2 cos
(ωrlt+ θ − tan−1
(R2
∆ + ωrl
)). (6.A.16b)
6.B.SPIN
-SYSTEM
RESPONSE
UNDER
RIN
G-LOCK
EXCIT
ATIO
N125
We can also express (6.A.16) as,
ωe1,x = β
√(ωrl + ∆)2 +R2
2 cos
(ωrlt+ θ − tan−1
(R2
∆ + ωrl
)+π
2
)(6.A.17a)
ωe1,y = β
√(ωrl + ∆)2 +R2
2 sin
(ωrlt+ θ − tan−1
(R2
∆ + ωrl
)+π
2
)(6.A.17b)
allowing us to express the ring-lock excitation envelope as a complex exponential
|ωe1|eiφ = |ωe
1| (cosφ+ i sinφ) (6.A.18)
where,
|ωe1| =
√R1
R2
(1− a1 + a
)[(∆ + ωrl)2 +R2
2
](6.A.18a)
φ = ωrlt+ θ − tan−1
(R2
∆ + ωrl
)+π
2(6.A.18b)
6.B Spin-system response under ring-lock excitation
The ring-lock excitation envelope (6.A.17) will drive a spin into the target trajectory (6.A.1) only when the excitation parameters are matched to
the spin properties. Here we consider the case where there is a mismatch between the excitation parameters, R1, R2,∆, and the spin parameters,
R1, R2, ∆, leading to a deviation from the desired trajectory, a, θ, the spin will respond with a trajectory a, θ. The mismatched spin trajectory
is,
mx =M0
z
2
√R1
R2(1− a2) sin
(ωrlt+ θ
)(6.B.1a)
my =M0
z
2
√R1
R2(1− a2) cos
(ωrlt+ θ
)(6.B.1b)
mz =M0
z
2(1 + a) (6.B.1c)
126
CHAPTER
6.RIN
G-LOCK
EXCITATIO
N
where a ∈ (−1, 1] represents elevation. The derivative of the trajectory (6.B.1) is,
mx = ωrlM0
z
2
√R1
R2(1− a2) cos
(ωrlt+ θ
)(6.B.2a)
my = −ωrlM0
z
2
√R1
R2(1− a2) sin
(ωrlt+ θ
)(6.B.2b)
mz = 0. (6.B.2c)
Substitution of the trajectory (6.B.1) and its derivative (6.B.2) into the Bloch equations (2.34) yelids
ωrlM0
z
2
√R1
R2(1− a2) cos
(ωrlt+ θ
)= −R2
M0z
2
√R1
R2(1− a2) sin
(ωrlt+ θ
)− ∆
M0z
2
√R1
R2(1− a2) cos
(ωrlt+ θ
)+ ωe
1,y
M0z
2(1 + a)
(6.B.3a)
−ωrlM0
z
2
√R1
R2(1− a2) sin
(ωrlt+ θ
)= +∆
M0z
2
√R1
R2(1− a2) sin
(ωrlt+ θ
)− R2
M0z
2
√R1
R2(1− a2) cos
(ωrlt+ θ
)− ωe
1,x
M0z
2(1 + a)
(6.B.3b)
0 = −ωe1,y
M0z
2
√R1
R2(1− a2) sin
(ωrlt+ θ
)+ ωe
1,x
M0z
2
√R1
R2(1− a2) cos
(ωrlt+ θ
). . .
− R1M0
z
2(1 + a) + R1M
0z (6.B.3c)
Consider equation (6.B.3a) and substitute the ring-lock envelope pattern (6.A.17b),
ωrlM0
z
2
√R1
R2(1− a2) cos
(ωrlt+ θ
)= −R2
M0z
2
√R1
R2(1− a2) sin
(ωrlt+ θ
)− ∆
M0z
2
√R1
R2(1− a2) cos
(ωrlt+ θ
). . . (6.B.4)
+
√R1
R2
(1− a1 + a
)[(∆ + ωrl)2 +R2
2
]sin
(ωrlt+ θ − tan−1
(R2
∆ + ωrl
)+π
2
)M0
z
2(1 + a)
6.B.SPIN
-SYSTEM
RESPONSE
UNDER
RIN
G-LOCK
EXCIT
ATIO
N127
Rearrange (6.B.4) to seperate the excitation and spin parameters,
1
(1 + a)
√R1
R2(1− a2)
[ (∆ + ωrl
)cos(ωrlt+ θ
)+ R2 sin
(ωrlt+ θ
)]= . . . (6.B.5)√
R1
R2
(1− a1 + a
)[(∆ + ωrl)2 +R2
2
]sin
(ωrlt+ θ − tan−1
(R2
∆ + ωrl
)+π
2
)√R1
R2
(1− a1 + a
)[ (∆ + ωrl
)cos(ωrlt+ θ
)+ R2 sin
(ωrlt+ θ
)]= . . . (6.B.6)√
R1
R2
(1− a1 + a
)[(∆ + ωrl)2 +R2
2
]cos
(ωrlt+ θ − tan−1
(R2
∆ + ωrl
)).
Apply (6.A.11) to the LHS of (6.B.6)√R1
R2
(1− a1 + a
)[(∆ + ωrl)2 + R2
2]
cos
(ωrlt+ θ − tan−1
(R2
∆ + ωrl
))= . . . (6.B.7)√
R1
R2
(1− a1 + a
)[(∆ + ωrl)2 +R2
2
]cos
(ωrlt+ θ − tan−1
(R2
∆ + ωrl
)).
Equate the amplitude of sinusoids in (6.B.7),√R1
R2
(1− a1 + a
)[(∆ + ωrl)2 + R2
2]
=
√R1
R2
(1− a1 + a
)[(∆ + ωrl)2 +R2
2
]1− a1 + a
=R1 R2
R1R2
(1− a1 + a
)((∆ + ωrl)
2 +R22
(∆ + ωrl)2 + R22
)to give an expression of the spin elevation,
a(a, ωrl, R1, R2,∆, R1, R2, ∆
)=
1− β1 + β
, (6.B.8)
where
β =R1R2
R1R2
(1− a1 + a
)((∆ + ωrl)
2 +R22
(∆ + ωrl)2 + R22
). (6.B.9)
128
CHAPTER
6.RIN
G-LOCK
EXCITATIO
N
The spin elevation (6.B.8) can also be expressed as,
a(a, ωrl, R1, R2,∆, R1, R2, ∆
)=R1R2 (1 + a)
[(∆ + ωrl
)2 − R22]−R1R2 (1− a)
[(∆ + ωrl)
2+R2
2
]R1R2 (1 + a)
[(∆ + ωrl
)2+ R2
2]
+R1R2 (1− a)[(∆ + ωrl)
2+R2
2
] . (6.B.10)
Equate the phase of sinusoids in (6.B.7), to give an expression for the spin phase,
ωrlt+ θ − tan−1
(R2
∆ + ωrl
)= ωrlt+ θ − tan−1
(R2
∆ + ωrl
)θ = θ − tan−1
(R2
∆ + ωrl
)+ tan−1
(R2
∆ + ωrl
). (6.B.11)
Chapter 7
Conclusion
Magnetic resonance imaging encompasses a wide selection of well developed methods for gaining
insight into the properties of a physical object. In this thesis, we have explored interrogating the spin
system with complex continuous wave excitation envelopes as an alternative method for estimating
spin system properties. The use of periodic excitation envelopes that drive the magnetisation into
steady-state trajectories allows for efficient model construction when optimising objective functions
to estimate system parameters. Using a series of Rabi modulated excitations we pose spectroscopy
and proton projection imaging as inverse problems. In a similar manner, using a series of ring-
lock excitations, we estimate relaxation constants using an analytical model derived from the Bloch
equations. Rabi modulated excitation has favourable properties for simultaneous excitation and
measurement, opening possibilities for observation of ultra-short relaxation spin systems and closed-
loop control. A current limitation of the proposed methods is measurement efficiency. A suggested
solution is to make controlled transitions between steady-state trajectories, and online monitoring of
the magnetisation to detect a steady-state condition. Our experimental investigation of steady-state
trajectories makes use of two pseudo-simultaneous excitation and measurement protocols. Whilst
these methods were adequate to explore the proof-of-concept applications, hardware modifications
are suggested to unlock the full potential of continuous wave excitation patterns.
7.1 Summary of original contributions
Chapter 4 investigated the off-resonance response to Rabi modulated continuous wave excitation.
The off-resonance response is fully expressed by harmonics of the excitation modulation frequency
and varies dependant on the excitation envelope parameters. The accuracy of the harmonic pre-
diction model of the off-resonance steady-state response was measured experimentally, then used to
construct a forward model, posing spectroscopy as an inverse problem. Using the proposed method,
ethanol spectra were reconstructed from a series of measured steady-state harmonics, acquired un-
der Rabi modulated excitation. Improvement to the reconstructed spectra was achieved by using
an A-optimality based algorithm to select experimental excitation parameters.
In Chapter 5, we took the idea of encoding off-resonance information via a series of Rabi modulated
excitations and applied it to imaging. Under a radial gradient, spatial information rather than
chemical shift information is encoded in the harmonic response to Rabi modulated excitation. A
129
130 CHAPTER 7. CONCLUSION
more efficient gapped measurement protocol was implemented and verified by measurement of the off-
resonance response. Projections of proton density were reconstructed from multiple measurements
acquired using the gapped protocol, and these were back-projected to form a two dimensional image,
whose structure was in line with reference data. Preliminary simulations of the imaging method,
applied to a numerical phantom with mixed relaxation rates, showed contrast between different spin
regions.
Chapter 6 revisited the known response to a constant amplitude excitation (Bloch, 1946; Abragam,
1961), applying a frequency offset which generates a steady-state ring trajectory. An analytical
solution of the Bloch equations was derived for the spin-system response under this ring-lock excita-
tion. Power and frequency of the ring-lock excitation envelope are expressed in terms of the system
parameters and a target response for the ring-lock trajectory. To verify the excitation and response
model, a basic spin-system was accurately controlled over the steady-state ellipsoid. Relaxometry
was posed as an optimisation problem, where relaxation rates are estimated from the response to
a series of ring-lock excitations. Simulations of this method were able to accurately estimate spin-
lattice and spin-spin relaxation rates, inferences from experimental data were limited to accurate
prediction of the ratio of the two rates.
7.2 Future work
The methods proposed in this thesis exploit information encoded in steady-state trajectories. The
described measurement techniques assume the spin-system is initially at equilibrium and apply the
excitation for a duration adequate to ensure a steady-state has been achieved. The efficiency of
the methods is limited, as only a small portion of the total experimental time is used for data
acquisition. If the magnetisation could be smoothly transitioned from one steady-state to the next,
the experimental efficiency would be greatly improved. This is a non-trivial problem, even in the
unrealistic case of a single spin; knowledge of the instantaneous spin orientation and an estimate
of relaxation constants would likely be required to maintain a controlled transition. The potential
to observe the spin-system during Rabi modulated excitation could find application here; first to
maintain an estimate of the spin position and secondly to monitor the steady-state transition and test
for a steady-state condition such as periodicity. If the data collection were then considered streaming
from an active experiment, model based estimation could be run throughout data acquisition, and
the experiment terminated once the parameter estimates converged.
One advantage of a fully simultaneous excitation and measurement technique is the absence of an
echo time, which allows the observation of ultra-short relaxation signals. Rabi modulated excitation
has two properties, low power and harmonic decoupling, that may be exploited to allow simultaneous
continuous wave excitation and measurement. It is possible to maintain a significant steady-state
trajectory with low power Rabi modulated excitation envelopes. Other groups have demonstrated
simultaneous excitation and measurement with custom hardware, when the excitation power is kept
to a minimum (Idiyatullin et al., 2012; Ozen et al., 2017a). The frequency decoupling between
Rabi modulated excitation and the harmonic response of the spin-system make it ideal for fre-
quency filtering hardware (Brunner et al., 2012; Zheng et al., 2011). There are certain similarities
to Magnetic Particle Imaging where a mixture of frequency filtering and active analog cancellation
has been suggested (Graeser et al., 2013). It was beyond the scope of this thesis to implement a
hardware solution for simultaneous continuous wave transmission and signal reception. All of the
afore-mentioned techniques, however, are potential candidates for future hardware realisation of the
methods developed in this thesis.
This thesis demonstrates the ability to investigate continuous wave steady-state phenomena with
7.2. FUTURE WORK 131
pseudo-simultaneous excitation and measurement protocols. Continuous wave excitation patterns
allow the construction of efficient prediction models and elicit an information-rich steady-state re-
sponse from which underlying spin-system properties can be reconstructed. It is anticipated that
further development of these concepts and related hardware modifications will lead to new continuous
wave imaging paradigms.
132 CHAPTER 7. CONCLUSION
Bibliography
Abragam, A. (1961). The Principles of Nuclear Magnetism. International series of monographs on
physics. Clarendon Press.
Akoka, S., Franconi, F., Seguin, F., and Le Pape, A. (1993). Radiofrequency map of an NMR coil
by imaging. Magnetic Resonance Imaging, 11(3):437–441.
Anumula, S., Magland, J., Wehrli, S., Zhang, H., Ong, H., Song, H., and Wehrli, F. (2006). Mea-
surement of phosphorus content in normal and osteomalacic rabbit bone by solid-state 3D radial
imaging. Magnetic Resonance in Medicine, 56(5):946–952.
Arnold, J., Dharmatti, S., and Packard, M. (1951). Chemical effects on nuclear induction signals
from organic compounds. The Journal of Chemical Physics, 19(4):507–507.
Arnold, J. T. (1956). Magnetic resonances of protons in ethyl alcohol. Physical Review, 102(1):136.
Behjati, S., Gundem, G., Wedge, D. C., Roberts, N. D., Tarpey, P. S., Cooke, S. L., Van Loo, P.,
Alexandrov, L. B., Ramakrishna, M., Davies, H., et al. (2016). Mutational signatures of ionizing
radiation in second malignancies. Nature Communications, 7:12605.
Bergin, C., Pauly, J., and Macovski, A. (1991). Lung parenchyma: projection reconstruction MR
imaging. Radiology, 179(3):777–781.
Bloch, F. (1946). Nuclear induction. Physical Review, 70(7-8):460.
Bloch, F., Hansen, W. W., and Packard, M. (1946). The nuclear induction experiment. Physical
Review, 70:474–485.
Block, K. T., Uecker, M., and Frahm, J. (2009). Model-based iterative reconstruction for radial fast
spin-echo MRI. IEEE Transactions on Medical Imaging, 28(11):1759–1769.
Blumich, B. (1984). Stochastic NMR imaging. Journal of Magnetic Resonance, 60:37–45.
Bottomley, P., Foster, T., and Leue, W. (1984). Chemical imaging of the brain by NMR. The
Lancet, 323(8386):1120.
Bretthorst, G. L., Hutton, W. C., Garbow, J. R., and Ackerman, J. J. (2005). Exponential parameter
estimation (in NMR) using Bayesian probability theory. Concepts in Magnetic Resonance Part
A, 27(2):55–63.
Brossmann, J., Frank, L. R., Pauly, J. M., Boutin, R. D., Pedowitz, R. A., Haghighi, P., and
Resnick, D. (1997). Short echo time projection reconstruction MR imaging of cartilage: com-
parison with fat-suppressed spoiled GRASS and magnetization transfer contrast MR imaging.
Radiology, 203(2):501–507.
Brown, T. R., Kincaid, B., and Ugurbil, K. (1982). NMR chemical shift imaging in three dimensions.
Proceedings of the National Academy of Sciences, 79(11):3523–3526.
Bruder, H., Fischer, H., Graumann, R., and Deimling, M. (1988). A new steady-state imaging
sequence for simultaneous acquisition of two MR images with clearly different contrasts. Magnetic
Resonance in Medicine, 7(1):35–42.
133
134 BIBLIOGRAPHY
Brunner, D., Pavan, M., Dietrich, B., and Pruessmann, K. (2012). MRI with sideband excitation:
Application to continuous SWIFT. In Proceedings of the 20th Annual Meeting of ISMRM, page
150, Melbourne, Australia.
Brunner, D., Pavan, M., Dietrich, B., Rothmund, D., Heller, A., and Pruessmann, K. (2011). Side-
band excitation for concurrent RF transmission and reception. In Proceedings of the 18th Annual
Meeting of ISMRM, page 625, Montreal, Canada.
Cappeller, U. and Muller, H. (1985). Phase-modulated exciation of a spin-system. Annalen der
Physik, 42(3):250–264.
Carr, H. (1958). Steady-State Free Precession in Nuclear Magnetic Resonance. Physical Review,
112(5):1693.
Carr, H. Y. (1953). Free Precession Techniques in Nuclear Magnetic Resonance. PhD thesis, Harvard
University.
Carr, H. Y. and Purcell, E. M. (1954). Effects of Diffusion on Free Precession in Nuclear Magnetic
Resonance Experiments. Physical Review, 94(3):630.
Chaloner, K. and Verdinelli, I. (1995). Bayesian experimental design: A review. Statistical Science,
pages 273–304.
Chappell, K. E., Patel, N., Gatehouse, P. D., Main, J., Puri, B. K., Taylor-Robinson, S. D., and
Bydder, G. M. (2003). Magnetic resonance imaging of the liver with ultrashort TE (UTE) pulse
sequences. Journal of Magnetic Resonance Imaging, 18(6):709–713.
Corum, C. A., Idiyatullin, D., Snyder, C. J., and Garwood, M. (2015). Gap cycling for SWIFT.
Magnetic Resonance in Medicine, 73(2):677–682.
Crowe, L., Wang, Y., Gatehouse, P., Tessier, J., Waterton, J., Robert, P., Bydder, G., and Firmin,
D. (2005). Ex vivo MR imaging of atherosclerotic rabbit aorta labelled with USPIO—enhancement
of iron loaded regions in UTE imaging. In Proc Intl Soc Mag Reson Med, volume 13, page 115.
Cunningham, C. H., Pauly, J. M., and Nayak, K. S. (2006). Saturated double-angle method for
rapid B1+ mapping. Magnetic Resonance in Medicine, 55(6):1326–1333.
Dadok, J. and Sprecher, R. F. (1974). Correlation NMR spectroscopy. Journal of Magnetic Reso-
nance, 13(2):243–248.
Damadian, R., Goldsmith, M., and Minkoff, L. (1977). NMR in cancer: XVI. FONAR image of the
live human body.
Delso, G., Wiesinger, F., Sacolick, L. I., Kaushik, S. S., Shanbhag, D. D., Hullner, M., and Veit-
Haibach, P. (2015). Clinical evaluation of zero-echo-time MR imaging for the segmentation of the
skull. Journal of Nuclear Medicine, 56(3):417–422.
Deoni, S. C., Rutt, B. K., and Peters, T. M. (2003). Rapid combined T1 and T2 mapping using
gradient recalled acquisition in the steady state. Magnetic Resonance in Medicine, 49(3):515–526.
Dette, H. (1997). Designing experiments with respect to ‘standardized’ optimality criteria. Journal
of the Royal Statistical Society: Series B (Statistical Methodology), 59(1):97–110.
Edelman, R. R. (2014). The history of MR imaging as seen through the pages of Radiology. Radiology,
273(2S):S181–S200.
Edelstein, W. A., Hutchison, J. M., Johnson, G., and Redpath, T. (1980). Spin warp NMR imaging
and applications to human whole-body imaging. Physics in Medicine and Biology, 25(4):751.
BIBLIOGRAPHY 135
Ernst, R. and Anderson, W. (1966). Application of Fourier transform spectroscopy to magnetic
resonance. Review of Scientific Instruments, 37(1):93–102.
Ernst, R. R., Bodenhausen, G., and Wokaun, A. (1990). Principles of Nuclear Magnetic Resonance in
one and two dimensions. International series of monographs on chemistry: 14. Oxford : Clarendon
Press, 1990, c1987.
Filho, G. H., Du, J., Pak, B. C., Statum, S., Znamorowski, R., Haghighi, P., Bydder, G., and
Chung, C. B. (2009). Quantitative characterization of the achilles tendon in cadaveric specimens:
T1 and T ∗2 measurements using ultrashort-TE MRI at 3T. American Journal of Roentgenology,
192(3):W117–W124.
Freeman, R. and Hill, H. (1971). Fourier transform study of NMR spin–lattice relaxation by “pro-
gressive saturation”. The Journal of Chemical Physics, 54(8):3367–3377.
Gatehouse, P. D., Thomas, R. W., Robson, M. D., Hamilton, G., Herlihy, A. H., and Bydder, G. M.
(2004). Magnetic resonance imaging of the knee with ultrashort TE pulse sequences. Magnetic
Resonance Imaging, 22(8):1061–1067.
Gold, G. E., Thedens, D. R., Pauly, J. M., Fechner, K., Bergman, G., Beaulieu, C. F., and Macovski,
A. (1998). MR imaging of articular cartilage of the knee: new methods using ultrashort TEs.
American Journal of Roentgenology, 170(5):1223–1226.
Graeser, M., Knopp, T., Gruttner, M., Sattel, T. F., and Buzug, T. M. (2013). Analog receive signal
processing for magnetic particle imaging. Medical Physics, 40(4).
Griswold, M. A., Jakob, P. M., Heidemann, R. M., Nittka, M., Jellus, V., Wang, J., Kiefer, B., and
Haase, A. (2002). Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magnetic
Resonance in Medicine, 47(6):1202–1210.
Grodzki, D. M., Jakob, P. M., and Heismann, B. (2012). Ultrashort echo time imaging using
pointwise encoding time reduction with radial acquisition (PETRA). Magnetic Resonance in
Medicine, 67(2):510–518.
Gupta, R. K., Ferretti, J. A., and Becker, E. D. (1974). Rapid scan Fourier transform NMR
spectroscopy. Journal of Magnetic Resonance (1969), 13(3):275–290.
Haacke, E. M., Brown, R. W., Thompson, M. R., Venkatesan, R., et al. (1999). Magnetic Resonance
Imaging: Physical principles and sequence design, volume 82.
Haase, A., Frahm, J., Hanicke, W., and Matthaei, D. (1985). 1H NMR chemical shift selective
(CHESS) imaging. Physics in Medicine and Biology, 30(4):341.
Haase, A., Frahm, J., Matthaei, D., Hanicke, W., and Merboldt, K.-D. (1986). FLASH imaging.
rapid NMR imaging using low flip-angle pulses. Journal of Magnetic Resonance (1969), 67(2):258–
266.
Hafner, S. (1994). Fast imaging in liquids and solids with the back-projection low angle shot (BLAST)
technique. Magnetic Resonance Imaging, 12(7):1047–1051.
Hahn, E. L. (1950). Spin echoes. Physical Review, 80(4):580.
Hargreaves, B. (2012). Rapid gradient-echo imaging. Journal of Magnetic Resonance Imaging,
36(6):1300–1313.
Hartmann, S. and Hahn, E. (1962). Nuclear double resonance in the rotating frame. Physical Review,
128(5):2042.
136 BIBLIOGRAPHY
Hennig, J., Nauerth, A., and Friedburg, H. (1986). RARE imaging: a fast imaging method for
clinical MR. Magnetic Resonance in Medicine, 3(6):823–833.
Hennig, J., Welz, A. M., Schultz, G., Korvink, J., Liu, Z., Speck, O., and Zaitsev, M. (2008).
Parallel imaging in non-bijective, curvilinear magnetic field gradients: a concept study. Magnetic
Resonance Materials in Physics, Biology and Medicine, 21(1-2):5.
Heule, R., Ganter, C., and Bieri, O. (2014). Triple echo steady-state (TESS) relaxometry. Magnetic
Resonance in Medicine, 71(1):230–237.
Homer, J. and Beevers, M. S. (1985). Driven-equilibrium single-pulse observation of T1 relaxation. a
reevaluation of a rapid “new” method for determining NMR spin-lattice relaxation times. Journal
of Magnetic Resonance (1969), 63(2):287–297.
Hornak, J. P., Szumowski, J., and Bryant, R. G. (1988). Magnetic field mapping. Magnetic Reso-
nance in Medicine, 6(2):158–163.
Idiyatullin, D., Corum, C., Moeller, S., and Garwood, M. (2008). Gapped pulses for frequency-swept
MRI. Journal of Magnetic Resonance, 193(2):267–273.
Idiyatullin, D., Corum, C., Moeller, S., Prasad, H. S., Garwood, M., and Nixdorf, D. R. (2011). Den-
tal magnetic resonance imaging: making the invisible visible. Journal of Endodontics, 37(6):745–
752.
Idiyatullin, D., Corum, C., Park, J.-Y., and Garwood, M. (2006). Fast and quiet MRI using a swept
radiofrequency. Journal of Magnetic Resonance, 181:342–349.
Idiyatullin, D., Suddarth, S., Corum, C. A., Adriany, G., and Garwood, M. (2012). Continuous
SWIFT. Journal of Magnetic Resonance, 220:26–31.
Insko, E. and Bolinger, L. (1993). Mapping of the radiofrequency field. Journal of Magnetic Reso-
nance, Series A, 103(1):82–85.
Jesson, J., Meakin, P., and Kneissel, G. (1973). Homonuclear decoupling and peak elimination
in Fourier transform Nuclear Magnetic Resonance. Journal of the American Chemical Society,
95(2):618–620.
Jones, D. K. (2010). Diffusion MRI: Theory, Methods, and Applications. Oxford University Press.
Kanayamay, S., Kuhara, S., and Satoh, K. (1996). In vivo rapid magnetic field measurement
and shimming using single scan differential phase mapping. Magnetic Resonance in Medicine,
36(4):637–642.
Kendi, A. T. K., Khariwala, S. S., Zhang, J., Idiyatullin, D. S., Corum, C. A., Michaeli, S., Pambuc-
cian, S. E., Garwood, M., and Yueh, B. (2011). Transformation in mandibular imaging with sweep
imaging with fourier transform magnetic resonance imaging. Archives of Otolaryngology–Head &
Neck Surgery, 137(9):916–919.
Kobayashi, N., Idiyatullin, D., Corum, C., Weber, J., Garwood, M., and Sachdev, D. (2015). SWIFT
MRI enhances detection of breast cancer metastasis to the lung. Magnetic Resonance in Medicine,
73(5):1812–1819.
Korte, J. C., Layton, K. J., Tahayori, B., Farrell, P. M., Moore, S. M., and Johnston, L. A. (2014).
Encoding chemical shift with Rabi modulated continuous wave excitation. In Proceedings of the
22nd Annual Meeting of ISMRM, Milan, Italy.
Korte, J. C., Layton, K. J., Tahayori, B., Farrell, P. M., Moore, S. M., and Johnston, L. A. (2017a).
BIBLIOGRAPHY 137
NMR spectroscopy using Rabi modulated continuous wave excitation. Biomedical Signal Process-
ing and Control, 33:422–428.
Korte, J. C., Tahayori, B., Farrell, P. M., Moore, S. M., and Johnston, L. A. (2015). Gapped
measurement of spin system response to periodic continuous wave excitation. In The Australian
and New Zealand Magnetic Resonance Society, Bay of Islands, New Zealand.
Korte, J. C., Tahayori, B., Farrell, P. M., Moore, S. M., and Johnston, L. A. (2016). Rabi modulated
continuous wave imaging. In Proceedings of the 24th Annual Meeting of ISMRM, Singapore.
Korte, J. C., Tahayori, B., Farrell, P. M., Moore, S. M., and Johnston, L. A. (2017b). Relaxometry
via steady-state ring-locked trajectories. In Proceedings of the 25th Annual Meeting of ISMRM,
Honolulu, USA.
Kuethe, D. O., Caprihan, A., Lowe, I. J., Madio, D. P., and Gach, H. M. (1999). Transforming
NMR data despite missing points. Journal of Magnetic Resonance, 139(1):18–25.
Lauterbur, P. C. et al. (1973). Image Formation by Induced Local Interactions: Examples Employing
Nuclear Magnetic Resonance. Nature, 242(5394):190–191.
Layton, K. J., Tahayori, B., Mareels, I. M., Farrell, P. M., and Johnston, L. A. (2014). Rabi
resonance in spin systems: Theory and experiment. Journal of Magnetic Resonance, 242:136 –
142.
Lehto, L. J., Sierra, A., Corum, C. A., Zhang, J., Idiyatullin, D., Pitkanen, A., Garwood, M., and
Grohn, O. (2012). Detection of calcifications in vivo and ex vivo after brain injury in rat using
SWIFT. NeuroImage, 61(4):761–772.
Leuraud, K., Richardson, D. B., Cardis, E., Daniels, R. D., Gillies, M., O’hagan, J. A., Hamra, G. B.,
Haylock, R., Laurier, D., Moissonnier, M., et al. (2015). Ionising radiation and risk of death from
leukaemia and lymphoma in radiation-monitored workers (INWORKS): an international cohort
study. The Lancet Haematology, 2(7):e276–e281.
Levitt, M. H. (2001). Spin Dynamics: Basics of Nuclear Magnetic Resonance. John Wiley & Sons.
Liang, Z., Lauterbur, P., in Medicine, I. E., and Society, B. (2000). Principles of Magnetic Resonance
Imaging: A Signal Processing Perspective. IEEE Press series in biomedical engineering. SPIE
Optical Engineering Press.
Ljunggren, S. (1983). A simple graphical representation of Fourier-based imaging methods. Journal
of Magnetic Resonance (1969), 54(2):338–343.
Luhach, I., Idiyatullin, D., Lynch, C. C., Corum, C., Martinez, G. V., Garwood, M., and Gillies,
R. J. (2014). Rapid ex vivo imaging of PAIII prostate to bone tumor with SWIFT-MRI. Magnetic
Resonance in Medicine, 72(3):858–863.
Lustig, M., Donoho, D., and Pauly, J. M. (2007). Sparse MRI: The application of compressed sensing
for rapid MR imaging. Magnetic Resonance in Medicine, 58(6):1182–1195.
Ma, D., Gulani, V., Seiberlich, N., Liu, K., Sunshine, J. L., Duerk, J. L., and Griswold, M. A.
(2013). Magnetic resonance fingerprinting. Nature, 495(7440):187.
Madio, D. P. and Lowe, I. J. (1995). Ultra-fast imaging using low flip angles and FIDs. Magnetic
Resonance in Medicine, 34(4):525–529.
Mansfield, P. (1977). Multi-planar image formation using NMR spin echoes. Journal of Physics C:
Solid State Physics, 10(3):L55.
138 BIBLIOGRAPHY
Mansfield, P. and Maudsley, A. (1977). Medical imaging by NMR. The British Journal of Radiology,
50(591):188–194.
Mansfield, P., Maudsley, A. A., and Bains, T. (1976). Fast scan proton density imaging by NMR.
Journal of Physics E: Scientific Instruments, 9(4):271.
Markley, J. L., Horsley, W. J., and Klein, M. P. (1971). Spin-lattice relaxation measurements in
slowly relaxing complex spectra. The Journal of Chemical Physics, 55(7):3604–3605.
Maudsley, A., Simon, H., and Hilal, S. (1984). Magnetic field measurement by NMR imaging.
Journal of Physics E: Scientific Instruments, 17(3):216.
Maudsley, A. A., Oppelt, A., and Ganssen, A. (1979). Rapid measurement of magnetic field distribu-
tions using nuclear magnetic resonance. Siemens Forschungs-und Entwicklungsberichte/Siemens
Research and Development Reports, 8(6):326–331.
McDonald, G. G. and Leigh, J. S. (1973). A new method for measuring longitudinal relaxation
times. Journal of Magnetic Resonance (1969), 9(3):358–362.
Meiboom, S. and Gill, D. (1958). Modified spin-echo method for measuring nuclear relaxation times.
Review of Scientific Instruments, 29(8):688–691.
Mills, T. C., Ortendahl, D. A., and Hylton, N. M. (1986). Investigation of partial flip angle magnetic
resonance imaging. Technical report, Radiologic Imaging Lab., Univ. of California, San Francisco,
400 Grandview Drive, South San Francisco, CA 94080.
National Toxicology Program (2016). Report on Carcinogens, Fourteenth Edition. U.S. Department
of Health and Human Services, Public Health Service.
Nauerth, A. and Gewiese, B. (1993). SPI—Single point FID imaging. In Proceedings of the 12th
Annual Meeting of SMRM, page 1215.
Nelson, M. T., Benson, J. C., Prescott, T., Corum, C. A., Snyder, A., and Garwood, M. (2012).
Breast MRI using SWeep Imaging with Fourier Transform (SWIFT). European Journal of Radi-
ology, 81:S109.
Nielles-Vallespin, S., Weber, M.-A., Bock, M., Bongers, A., Speier, P., Combs, S. E., Wohrle, J.,
Lehmann-Horn, F., Essig, M., and Schad, L. R. (2007). 3D radial projection technique with
ultrashort echo times for sodium MRI: clinical applications in human brain and skeletal muscle.
Magnetic Resonance in Medicine, 57(1):74–81.
Oppelt, A., Graumann, R., Barfuss, H., Fischer, H., Hartl, W., Schajor, W., et al. (1986). FISP—a
new fast MRI sequence. Electromedica, 54(1):15–18.
Ouwerkerk, R., Bleich, K. B., Gillen, J. S., Pomper, M. G., and Bottomley, P. A. (2003). Tissue
sodium concentration in human brain tumors as measured with 23Na MR imaging. Radiology,
227(2):529–537.
Ozen, A., Jouda, M., Korvink, J., , and Bock, M. (2017a). Full digital cancellation-based NMR with
concurrent excitation and acquisition using a lock-in amplifier. In Proceedings of the 25th Annual
Meeting of ISMRM, Honolulu, USA.
Ozen, A., Korvink, J., Atalar, E., and Bock, M. (2017b). In vivo MRI with concurrent excitation
and acquisition using dynamic analog cancellation with real-time feedback. In Proceedings of the
25th Annual Meeting of ISMRM, Honolulu, USA.
Pauly, J., Conolly, S., Nishimura, D., and Macovski, A. (1989a). Slice-selective excitation for very
short T2 species. In Proceedings of SMRM, volume 28.
BIBLIOGRAPHY 139
Pauly, J., Nishimura, D., and Macovski, A. (1989b). A k-space analysis of small-tip-angle excitation.
Journal of Magnetic Resonance (1969), 81(1):43–56.
Pauly, J., Nishimura, D., and Macovski, A. (1989c). A linear class of large-tip-angle selective
excitation pulses. Journal of Magnetic Resonance (1969), 82(3):571–587.
Prammer, M., Haselgrove, J., Shinnar, M., and Leigh, J. (1988). A new approach to automatic
shimming. Journal of Magnetic Resonance, 77:40–52.
Pruessmann, K. P., Weiger, M., Scheidegger, M. B., Boesiger, P., et al. (1999). SENSE: sensitivity
encoding for fast MRI. Magnetic Resonance in Medicine, 42(5):952–962.
Purcell, E. M., Torrey, H. C., and Pound, R. V. (1946). Resonance absorption by nuclear magnetic
moments in a solid. Physical Review, 69:37–38.
Rahmer, J., Bornert, P., Groen, J., and Bos, C. (2006). Three-dimensional radial ultrashort echo-
time imaging with T2 adapted sampling. Magnetic Resonance in Medicine, 55(5):1075–1082.
Rautiainen, J., Lehto, L. J., Tiitu, V., Kiekara, O., Pulkkinen, H., Brunott, A., van Weeren, R.,
Brommer, H., Brama, P. A., Ellermann, J., et al. (2013). Osteochondral repair: evaluation with
sweep imaging with Fourier transform in an equine model. Radiology, 269(1):113–121.
Redfield, A. G. (1955). Nuclear Magnetic Resonance Saturation and Rotary Saturation in Solids.
Physical Review, 98(6):1787.
Robson, M., Gatehouse, P., So, P.-W., Bell, J. D., and Bydder, G. M. (2004). Contrast enhancement
of short T2 tissues using ultrashort TE (UTE) pulse sequences. Clinical Radiology, 59(8):720–726.
Robson, M. D. and Bydder, G. M. (2006). Clinical ultrashort echo time imaging of bone and other
connective tissues. NMR in Biomedicine, 19(7):765–780.
Sacolick, L. I., Wiesinger, F., Hancu, I., and Vogel, M. W. (2010). B1 mapping by Bloch-Siegert
shift. Magnetic Resonance in Medicine, 63(5):1315–1322.
Saiko, A. P., Fedaruk, R., and Markevich, S. A. (2015). Multi-photon transitions and Rabi resonance
in continuous wave EPR. Journal of Magnetic Resonance, 259:47–55.
Santyr, G. E., Fairbanks, E. J., Kelcz, F., and Sorenson, J. A. (1994). Off-resonance spin locking
for MR imaging. Magnetic Resonance in Medicine, 32(1):43–51.
Santyr, G. E., Henkelman, R. M., and Bronskill, M. J. (1989). Spin locking for magnetic resonance
imaging with application to human breast. Magnetic Resonance in Medicine, 12(1):25–37.
Scheffler, K. and Hennig, J. (2001). T1 quantification with inversion recovery TrueFISP. Magnetic
Resonance in Medicine, 45(4):720–723.
Schmitt, P., Griswold, M. A., Jakob, P. M., Kotas, M., Gulani, V., Flentje, M., and Haase, A. (2004).
Inversion recovery TrueFISP: quantification of T1, T2, and spin density. Magnetic Resonance in
Medicine, 51(4):661–667.
Schneider, E. and Glover, G. (1991). Rapid in vivo proton shimming. Magnetic Resonance in
Medicine, 18(2):335–347.
Sekihara, K., Matsui, S., and Kohno, H. (1985). A new method of measuring static field distribution
using modified Fourier NMR imaging. Journal of Physics E: Scientific Instruments, 18(3):224.
Sepponen, R. E., Pohjonen, J. A., Sipponen, J. T., and Tanttu, J. I. (1985). A method for T1p
imaging. Journal of Computer Assisted Tomography, 9(6):1007–1011.
140 BIBLIOGRAPHY
Slichter, C. P. (1978). Principles of Magnetic Resonance. Springer series in solid-state sciences:
Volume 1. Berlin ; New York : Springer-Verlag.
Stockmann, J. P., Ciris, P. A., Galiana, G., Tam, L., and Constable, R. T. (2010). O-space imaging:
Highly efficient parallel imaging using second-order nonlinear fields as encoding gradients with no
phase encoding. Magnetic Resonance in Medicine, 64(2):447–456.
Stollberger, R. and Wach, P. (1996). Imaging of the active B1 field in vivo. Magnetic Resonance in
Medicine, 35(2):246–251.
Stollberger, R., Wach, P., McKinnon, G., Justich, E., and Ebner, F. (1988). RF-field mapping in
vivo. In Proceedings of the 7th Annual Meeting of SMRM, page 106.
Sumpf, T. J., Uecker, M., Boretius, S., and Frahm, J. (2011). Model-based nonlinear inverse re-
construction for T2 mapping using highly undersampled spin-echo MRI. Journal of Magnetic
Resonance Imaging, 34(2):420–428.
Tahayori, B., Johnston, L., Layton, K., Farrell, P., and Mareels, I. (2015). Solving the Bloch equation
with periodic excitation using harmonic balancing: Application to Rabi modulated excitation.
IEEE Transactions on Medical Imaging, 34(10):2118–2130.
Tahayori, B., Johnston, L., Mareels, I., and Farrell, P. (2008). Challenging the optimality of the
pulse excitation in magnetic resonance imaging. In Proceedings of the 17th IFAC World Congress,
pages 10301–10306.
Tahayori, B., Johnston, L. A., Mareels, I. M. Y., and Farrell, P. M. (2009). Revisiting excitation pat-
tern design for magnetic resonance imaging through optimisation of the signal contrast efficiency.
Biomedical Signal Processing and Control, 4:317–328.
Twieg, D. B. (1983). The k-trajectory formulation of the NMR imaging process with applications
in analysis and synthesis of imaging methods. Medical Physics, 10(5):610–621.
Vold, R., Waugh, J., Klein, M. P., and Phelps, D. (1968). Measurement of spin relaxation in complex
systems. The Journal of Chemical Physics, 48(8):3831–3832.
Waldman, A., Rees, J., Brock, C., Robson, M., Gatehouse, P., and Bydder, G. (2003). MRI of the
brain with ultra-short echo-time pulse sequences. Neuroradiology, 45(12):887–892.
Wang, J., Qiu, M., Yang, Q., Smith, M., and Constable, R. (2005). Measurement and correction
of transmitter and receiver induced nonuniformities in vivo. Magnetic Resonance in Medicine,
53:408–417.
Wang, L., Tang, W., Zhen, Z., Chen, H., Xie, J., and Zhao, Q. (2014). Improving detection specificity
of iron oxide nanoparticles (IONPs) using the SWIFT sequence with long T2 suppression. Magnetic
Resonance Imaging, 32(6):671–678.
Wang, L., Wu, Y., Chang, G., Oesingmann, N., Schweitzer, M. E., Jerschow, A., and Regatte, R. R.
(2009). Rapid isotropic 3D-sodium MRI of the knee joint in vivo at 7T. Journal of Magnetic
Resonance Imaging, 30(3):606–614.
Weiger, M., Brunner, D. O., Dietrich, B. E., Muller, C. F., and Pruessmann, K. P. (2013a). ZTE
imaging in humans. Magnetic Resonance in Medicine, 70(2):328–332.
Weiger, M., Pruessmann, K. P., Bracher, A.-K., Kohler, S., Lehmann, V., Wolfram, U., Hennel,
F., and Rasche, V. (2012). High-resolution ZTE imaging of human teeth. NMR in Biomedicine,
25(10):1144–1151.
BIBLIOGRAPHY 141
Weiger, M., Pruessmann, K. P., and Hennel, F. (2011). MRI with zero echo time: Hard versus sweep
pulse excitation. Magnetic Resonance in Medicine, 66(2):379–89.
Weiger, M., Stampanoni, M., and Pruessmann, K. P. (2013b). Direct depiction of bone microstruc-
ture using MRI with zero echo time. Bone, 54(1):44–47.
Welsch, G. H., Scheffler, K., Mamisch, T. C., Hughes, T., Millington, S., Deimling, M., and Trattnig,
S. (2009). Rapid estimation of cartilage T2 based on double echo at steady state (DESS) with 3
Tesla. Magnetic Resonance in Medicine, 62(2):544–549.
Wu, Y., Ackerman, J. L., Chesler, D. A., Graham, L., Wang, Y., and Glimcher, M. J. (2003).
Density of organic matrix of native mineralized bone measured by water-and fat-suppressed proton
projection MRI. Magnetic Resonance in Medicine, 50(1):59–68.
Yarnykh, V. L. (2007). Actual flip-angle imaging in the pulsed steady state: a method for rapid three-
dimensional mapping of the transmitted radiofrequency field. Magnetic Resonance in Medicine,
57(1):192–200.
Zhang, J., Chamberlain, R., Etheridge, M., Idiyatullin, D., Corum, C., Bischof, J., and Garwood, M.
(2014). Quantifying iron-oxide nanoparticles at high concentration based on longitudinal relax-
ation using a three-dimensional SWIFT look-locker sequence. Magnetic Resonance in Medicine,
71(6):1982–1988.
Zhang, J., Idiyatullin, D., Corum, C. A., Kobayashi, N., and Garwood, M. (2016). Gradient-
modulated SWIFT. Magnetic Resonance in Medicine, 75(2):537–546.
Zheng, B., Goodwill, P., and Conolly, S. (2011). Transmit filter design methods for magnetic particle
imaging. In Proc. SPIE Medical Imaging, volume 7965.