CPC 2000 Temporary Operation Instruction GB For Low-Pressure Controller CPC 2000
Transcript

CPC 2000

Temporary Operation Instruction GB

For Low-Pressure Controller CPC 2000

Low- Pressure Controller CPC 2000

2 WIKA Operation Instruction Low Pressure Controller Version 0.1

Information This symbol provides you with information, notes and tips. Warning! This Symbol warns you against actions that can cause injury to people or damage to the instrument.

Low- Pressure Controller CPC 2000

Version 0.1 WIKA Operation Instruction Low Pressure Controller 3

Content

1. General ...................................................................................................................... 4 1.1. Safety Instruction ................................................................................................... 5 1.2. Specific safety instructions ..................................................................................... 6 1.3. Shipping, assembly, electrical connection and startup ........................................... 6

2. Instrument description ................................................................................................ 7 2.1. Functional description ............................................................................................ 7 2.2. Rear view connections and elements ..................................................................... 7 2.3. Power Supply ......................................................................................................... 7 2.4. Front view connections and elements .................................................................... 8 2.5. Pressure ports ....................................................................................................... 8 2.6. Button ON/OFF ...................................................................................................... 8 2.7. Button MENU ......................................................................................................... 8 2.8. The Menu............................................................................................................... 9 2.8.1. Range .................................................................................................................... 9 2.8.2. Unit ........................................................................................................................ 9 2.8.3. Steps ..................................................................................................................... 9 2.8.4. Operation Mode ..................................................................................................... 9 2.8.5. Language ............................................................................................................... 9 2.8.6. Settings .................................................................................................................. 9 2.9. Button LEAK .........................................................................................................10 2.10. Button ZERO ........................................................................................................10 2.11. Overpressure protection ........................................................................................10

3. Operation and operation mode .................................................................................11 3.1. Control (CTRL)......................................................................................................11 3.2. Measure (MEAS) ..................................................................................................12 3.3. Automatic (AUTO) .................................................................................................13 3.4. Leakage (LEAK)....................................................................................................14

4. Digital interface .........................................................................................................15 4.1. USB ......................................................................................................................15 4.2. RS-232 .................................................................................................................15 4.3. Interface configuration...........................................................................................15 4.4. Commands for the serial interface ........................................................................15 4.5. Drivers ..................................................................................................................17

5. Troubleshooting ........................................................................................................18 5.1. Troubleshooting measures ....................................................................................18

6. Re-calibrating and servicing ......................................................................................19

7. Dismantling of the system .........................................................................................20

8. Transport of the system ............................................................................................20

9. Storage of the system ...............................................................................................21

10. Placing out of service ................................................................................................21

11. Specifications ...........................................................................................................22 11.1. Dimensions in mm ................................................................................................23

12. Accessories ..............................................................................................................24

13. Notes ........................................................................................................................25

Low- Pressure Controller CPC 2000

4 WIKA Operation Instruction Low Pressure Controller Version 0.1

1. General

In the following chapters detailed information on the Low-Pressure Controller CPC2000 and its proper use can be found. Should you require further information, or should there be problems which are not dealt within detail in the operating instructions, please contact the following address:

WIKA Alexander Wiegand SE & Co. KG Alexander Wiegand Strasse D-63911 Klingenberg Fon. +49 - (0) 93 72 / 132-9986 Fax. +49 - (0) 93 72 / 132-217 E- Mail: [email protected]

Factory calibration of the instrument is according to relevant international standards. The warranty period for the Low-Pressure Controller CPC2000 is 24 months according to the general terms of supply of ZVEI. The warranty period is void if the appliance is put to improper use or if the operating instructions are not observed or if an attempt is made to open the appliance. We also point out that the content of these operating instructions neither forms part of an earlier or existing agreement, assurance or legal relationship nor is meant to change these. All obligations of WIKA Alexander Wiegand SE & Co. KG result from the respective sales contract and the general business terms of WIKA Alexander Wiegand SE & Co. KG. WIKA is a registered trade mark of WIKA Alexander Wiegand SE & Co. KG. Names of companies or products mentioned in this handbook are registered trade marks of the manufacturer. We reserve the right to effect reasonable changes on the basis of technical improvements. Any reproduction of this manual or parts thereof by any means is prohibited. Version key regarding firmware and respective manual. Manual Firmware 0.1 2.5 © 2006 Copyright WIKA Alexander Wiegand SE & Co. KG

Low- Pressure Controller CPC 2000

Version 0.1 WIKA Operation Instruction Low Pressure Controller 5

This instruction manual describes the features of the Low-Pressure Controller CPC2000 and provides guidelines for its use. Improper use of this instrument or failure to follow these instructions may cause injury or equipment damage. All individuals responsible for operating this instrument must therefore be properly trained and aware of the hazards, and must carefully follow these operating instructions and the safety precautions detailed within. Contact the manufacturer if you do not understand parts of this instruction manual. Handle this manual with care: It must be readily available throughout the lifecycle of the instrument. It must be provided to any individuals who assume responsibility for operating the instrument at a later date. It must include any supplementary materials provided by the manufacturer. Conformity This instrument corresponds to the state of the art and meets all legal requirements set forth in EC directives as evidenced by the CE label.

1.1. Safety Instruction

This device has been designed and tested in accordance with the relevant safety regulations for electronic devices. However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advice given in this manual is followed when using the device.

1. The system must only be operated by trained and authorized personnel who know

the manual and can work according to them! 2. Trouble-free operation and reliability of the device can only be guaranteed so long as

the device is not subjected to any climatic conditions other than those stated under ―11 Specifications‖.

3. The CPC2000 always has to be handled with the care required for an electronic precision instrument (protect from humidity, impacts, strong magnetic fields, static electricity and extreme temperatures, do not insert any objects into the instrument and its openings). The device and sensors must be handled with care (don‗t throw, hit, etc.). Protect plugs and sockets from contamination.

4. If the device is moved from a cold to a warm environment, condensation may cause the equipment to fail. You should therefore ensure the device temperature has adjusted to the ambient temperature before trying to switch it on.

5. If the instrument is to be connected to other devices (e.g. via serial interface) care must be taken when designing the equipment connections. It is possible that internal wiring within the external device (e.g. connection of GND to earth) may cause excessive voltages which could harm or destroy the instrument or other connected devices.

6. The mains plug of the CPC2000 must always be accessible when connected to a power outlet, i.e., you must be able to pull the plug from the power outlet without difficulty, at any time.

If the device is operated with a faulty mains power supply (e.g. short circuit from mains voltage to output voltage) this could result in dangerous voltages at the device (e.g. at the sensor socket or serial interface)!

7. Significant electromagnetic radiation can adversely affect the measuring signal of the

CPC2000 or even disrupt the display of the signal completely.

Low- Pressure Controller CPC 2000

6 WIKA Operation Instruction Low Pressure Controller Version 0.1

8. The display window behind the plastic film is made of glass (which can splinter). If there is a possibility that this glass incl. the plastic film might break during operation, anyone in the close vicinity of the equipment must wear eye protectors, both before and during operation.

9. Test and calibration assemblies must always be constructed and also dismantled in an unpressurised state (atmosphere).

10. If the CPC2000 comes in contact with oil (which is forbidden), then subsequent use with fuels or gases is strictly prohibited, since this could cause an explosion and risks harm to both people and machinery.

11. If the equipment is damaged and might no longer operate safely, then it should be taken out of use and securely marked in such a way so that it is not used again.

Operator safety may be at risk if:

There is visible damage to the device. The device is not working as specified. The device has been stored under unsuitable conditions for an extended period of

time. If there is any doubt, please return the device to the manufacturer for repair or maintenance.

12. Customers must not attempt to alter or repair the device themselves. Please return the device to the manufacturer for any repair or maintenance.

13. Any operation not included in the following instructions or outside the specifications must not be attempted.

1.2. Specific safety instructions

The Low-Pressure Controller CPC2000 is used for testing and calibrating pressure sensors. Always observe the operating requirements —particularly the permissible supply voltage— indicated on the product label and in the ―11 Specifications‖ section of this manual. The instrument may only be handled as indicated in this manual. The manufacturer is not liable for damages caused by improper use or failure to follow these instructions. Violations of this type render all warranty claims null and void.

1.3. Shipping, assembly, electrical connection and startup

Do not close the pressure ports when shipping, as changes in barometric pressure could damage instruments with low measuring ranges.

Only technical personnel who are appropriately trained and authorized by the operator of the facility may assemble the instrument and set up its electrical connections.

Pressurized air or breath is not to be used for performance tests, as this could damage instruments with low measurement ranges.

Measurement errors may occur if the instrument is not kept protected from sunlight. Specific safety precautions are given in the individual sections of this manual.

Low- Pressure Controller CPC 2000

Version 0.1 WIKA Operation Instruction Low Pressure Controller 7

2. Instrument description

2.1. Functional description

The CPC2000 microprocessor-controlled Low-Pressure Controller can be used for the following:

Easy generating of positive and negative reference pressures Measuring positive and negative pressures Measuring differential pressure Identifying leaks in a test assembly

2.2. Rear view connections and elements

2.3. Power Supply

The Low-Pressure Controller CPC2000 is designed at the factory to accommodate a supply voltage of 24VDC. The supply voltage connector is located on the rear of the instrument.

WARNING! Electric shock hazard! Failure to unplug the power supply cord before replacing the fuse may result in fatal injuries!

USB

RS-232

AC adapter connector

Low- Pressure Controller CPC 2000

8 WIKA Operation Instruction Low Pressure Controller Version 0.1

2.4. Front view connections and elements

2.5. Pressure ports

Differential pressure instruments will be connected to the + and – pressure ports Gauge pressure instruments will be connected to the + pressure port (- port is open to

the atmosphere) Negative gauge pressure instruments will be connected to the – pressure port (+ port

is open to the atmosphere)

2.6. Button ON/OFF

Button to turn the instrument ON or OFF

2.7. Button MENU

Via this key, the customer gets into the Setup- Menu and is able to do various adjustments to the different operation modes. After using the MENU button in the control mode the controller is in vent mode.

Pressing the SETUP button allows the operator to adjust the following 6 settings. A detailed description to the menu items can be found in section: ―2.8 The Menu‖

RANGE: Set the control pressure range (End value)

UNIT: Choosing the pressure unit (Pa, kPa, hPa, bar, mbar, psi, inH2O, inHg, mmHg, Torr)

STEPS: Alignment of the step size in %. Within the range 0 … 50% adjustable. MODE: Choosing the operation mode

o CONTROL (CTRL): Controls a setpoint o MEASURE (MEAS): Measuring of differential and gauge pressure o AUTOMATIC (AUTO): drives through a programmed sequence

LANGUAGE: Choosing the language SETTINGS: Further settings to the controller can be done in this menu item.

+ Pressure connector

-

Main switch

Navigation keys

Low- Pressure Controller CPC 2000

Version 0.1 WIKA Operation Instruction Low Pressure Controller 9

2.8. The Menu

Press the button MENU and use the keys UP / DOWN to get to the right menu item. You get into the sub menu via the key SELECT. The following sub menu items are available.

2.8.1. Range

It is possible to change the controller end value of the sensor range to protect device under test with smaller pressure ranges.

2.8.2. Unit

This feature allows the operator to select the units used for displaying pressures. Certain measurement ranges cannot be displayed meaningfully in some units, in which case the units in question are not available. The following units may be selected: Pa, kPa, hPa bar, mbar psi, inH2O, inHg, mmHg Torr

2.8.3. Steps

This feature allows the user to modify the percentage of the set value. The value is free adjustable within the range 0 … 50 %.

2.8.4. Operation Mode

This feature allows the user to modify the general operation mode of the controller. The following modes are available: MEASURE (MEAS): The controller measures the pressure in the selected pressure unit CONTROL (CTRL): The controller regulates to a set point. AUTOMATIC (AUTO): Runs through a programmed sequence.

2.8.5. Language

This feature allows the operator to select either German or English as the language used.

2.8.6. Settings

Zeroing: Turning the automatic zeroing after quitting the menu on or off. The customer is also able to the set a frequency in which the sensor automatically performs a zero-point adjustment.

RS232 / USB: Choosing the actual interface Display: Brightness adjustment Auto: Setting the pressure range, unit and

pressure steps for the automatic mode. Steps Up and Steps Down can be assigned different

Navigation in the MENU: SELECT: Serves to confirm/ select the respective settings Left/ right: Select the digit

Up/Down: Switch between the different menu items.

Low- Pressure Controller CPC 2000

10 WIKA Operation Instruction Low Pressure Controller Version 0.1

step size. Info: Basic information about the instrument and manufacturer.

2.9. Button LEAK

The test feature blocks both pressure ports, allowing the operator to measure a drop in pressure on the test object itself. (This feature is only in the control mode available)

2.10. Button ZERO

Button for manual zero adjustment of the integrated sensor. By external influences such as temperature, location, or ambient pressure the zero point of the device can move. During the zero-point adjustment, the unit automatically determines this shift, and expects them into the displayed pressure.

The zero adjustment requires a switch of the internal valves, which is not possible without a pressure drop. If this interferes with the use case of the automatic zero point adjustment can be switched off.

2.11. Overpressure protection

It is not allowed to override the permissible overpressure safety: CPC2000-1mbar / CPC2000-10mbar / CPC2000-100mbar have a 5 time overpressure safety; CPC2000-1000mbar has a 2 time overpressure safety.

Low- Pressure Controller CPC 2000

Version 0.1 WIKA Operation Instruction Low Pressure Controller 11

To change set value by x%:

Pressure change in % via / - key

Pressure change to 100% via - key

Pressure change back to 0% via -key

3. Operation and operation mode

3.1. Control (CTRL)

The control mode is used to calibrate pressure transducers, switches or other pressure instruments. In CTRL- Mode the pump is active and regulates the set point. The internal sensor provides the corresponding measuring value over the display. Settings:

Press the MENU button. Verify the menu item MODE with SELECT and switch between the MODE options until CTRL is shown in the display. Accept with SELECT.

Change the end value of the control range in the menu item RANGE. Accept with SELECT.

Choosing a pressure unit in the menu item UNIT and accept with SELECT. Quit the menu with the key MENU

Utilization: After quitting the menu the controller is in vent mode, which means the controller is vented to atmospheric pressure. The maximum control pressure is displayed in the upper middle of the display. The percent value can be read in the lower middle of the display. You can switch between the pressure set point and percent value via the SELECT button.

The percent value can be changed via the UP / DOWN (in the defined steps) and LEFT / RIGHT button.

The target pressure set point can be changed only with the keys UP / DOWN. With LEFT / RIGHT can be jumped to the appropriate place of the target digit.

As soon as the display of the Controller indicates a stable value, the data can be noted:

Set point in % of Span

Current measuring value

Unit

End Value / Span

Battery status

Operation mode

Current set point in unit

Low- Pressure Controller CPC 2000

12 WIKA Operation Instruction Low Pressure Controller Version 0.1

In CTRL- Mode it is possible to do a manual zero point adjustment to the controller and check connected devices for leaks.

Manuel zero point adjustment: Press ZERO and the reference sensor in the instrument will be synchronized.

Leak Test: By pressing the LEAK button, the control process will be stopped and the system will be closed. By pressing the LEAK button again the controller leaves this mode and the control mode will be activated.

3.2. Measure (MEAS)

The MEAS- mode is used to measure gauge or differential pressure. The pump is not active during measuring. The pressure is captured by the internal sensor. Settings:

Press the MENU button. Verify the menu item MODE with SELECT and switch between the MODE options until MEAS is shown in the display. Accept with SELECT.

Menu items RANGE and STEPS are not needed for the MEAS- Mode Choosing a pressure unit in the menu item UNIT and accept with SELECT. Quit the menu with the key MENU. Sub menu item Settings:

o Go to the menu item Settings via the UP / DOWN buttons and accept with SELECT:

o Go to the sub item zeroing and accept with SELECT. o Under the sub item Auto – Zero you can turn the auto-zero point adjustment

during measuring on or off. Utilization: Depending on the settings the controller does a zero point adjustment after quitting the menu. Afterwards the controller starts measuring the pressure.

Zero Point adjustment can be done any time manually by pressing the Zero button.

Crosses the measured pressure the maximum permissible load (125%) the internal sensor will be blocked by a solenoid valve. Error is indicated in the display. You can unlock and de- allocate the instrument with SELECT key

Current measuring value

Battery status

Operation mode

End Value / Span

Unit

Deviation in pressure unit

Low- Pressure Controller CPC 2000

Version 0.1 WIKA Operation Instruction Low Pressure Controller 13

3.3. Automatic (AUTO)

In Automatic- mode a pressure sequence can be programmed. This is useful if several pressure transducers or pressure switched will be calibrated with the same pressure values. Additionally the mode makes it possible to drive off a fixed profile several times successively. Settings

Press the MENU button. Verify the menu item MODE with SELECT and switch between the MODE options until AUTO is shown in the display. Accept with SELECT.

The submenu items Range, Unit and Steps are not relevant for the Automatic- Mode. Go to the menu item Settings via the UP / DOWN buttons and accept with SELECT.

The item Auto- Zero is not relevant. Go to the item Automatic- Mode via the UP /

DOWN buttons and accept with SELECT. o RANGE: Sets the max control range

(End value) o UNIT: Sets the pressure unit o STEPS: Number of upward and

downward cycle steps. The step size will be calculated trough the controller.

In the submenu item CONFIGURATION 1 you can specify further attitudes for AUTO- Mode.

Cycles: Here you specify the number of cycles, which are to be driven off in the AUTO mode.

T Start: Waiting period until venting the instrument.

T Stop: Waiting period after reaching the specified pressure

T Hold: Hold Time until controlling to a new set point.

T Pause: Waiting period between two cycles Auto Zero: Auto Zero adjustment after each

cycle.

Press MENU to leave submenu. Utilization After leaving the menu the controller is on hold. All information for the automatic mode are shown at the display.

Current Phase: Phase 1: Venting the system

Operation mode

Span

Unit

Current pressure value in percent of span

Current Cycle

Low- Pressure Controller CPC 2000

14 WIKA Operation Instruction Low Pressure Controller Version 0.1

The illustration shows the flow chart for the AUTO mode. According to the following table you

can recognize which cycle gone through and the respective meaning. Stage Meaning

1 Waiting fort he start command via SELECT key 2 Waiting period which can be set in the Configuration 4 Time for Zero- Point Adjustment 7 Hold time at the zero point 9 Time to control to the set point 10 Hold Time 13 Hold time at the maximum pressure 15 Time to control to the set point 16 Hold time at the zero point 18 Break time which can be set in the Configuration Between phase 10 and 16 the data of the controller can be noted and evaluated if the value is stable. You can leave the AUTO mode with the MENU key. Vent will indicated in the upper left of the display.

3.4. Leakage (LEAK)

Connected device under test can be tested of leaks in CTRL- Mode. Settings To check a device under test for leaks the instrument has to be in CTRL- Mode.

Enter the desired pressure via UP / DOWN or LEFT / RIGHT keys. As soon as the pressure is stable press the LEAK button. The internal Pump is

switched off and the pressure stays in the system Utilization

Current measuring value

Leakage measuring time

Operation mode

Start value

Unit

Deviation in pressure unit

Deviation in %

Low- Pressure Controller CPC 2000

Version 0.1 WIKA Operation Instruction Low Pressure Controller 15

4. Digital interface

You have the possibility to adjust or observe the controller with a PC over digital interface (USB or RS-232). In control, measure and automatic mode it is possible to release the current instrument status switching on or off as a cycle. Release cycle is 1 second.

4.1. USB

The USB interface makes a virtual ComPort available to the PC. The instrument control is not different then controlling a RS-232 interface.

4.2. RS-232

RxD, TxD and GND cable will be required for connection. The connection is carried out with a straight RS-232 cable. (male to female)

4.3. Interface configuration

Press the MENU button to get into the main menu. Use the buttons UP / DOWN until you get to the item Settings and accept with SELECT. Go to the submenu item RS232 / USB via the keys UP / DOWN and accept with SELECT. The following settings can be adjusted:

Active: Choice of used digital interface and activate the interface communication (USB, RS-232)

Baud: Choosing the baud rate (1200, 2400, 4800, 9600, 14400, 19200, 28800, 38400, 56000, 57600)

Data- Bits: fixed value (8) Stop- Bits: fixed value (1) Parity- Bits: fixed value (N)

4.4. Commands for the serial interface

All interface commands are introduced with a colon and locked with carriage return (CR).

Commands and parameters are to be separated by a blank. By adding a question mark to

the command the corresponding parameter can be read out. Received commands are

acknowledged with „SELECT ―, not existing or wrong instructions with „ERROR ―.

Command Function / Response

:saaz <0 | 1> Auto Zero before each cycle (Phase 4) 0 – Off 1 - On

:acy <1…100> Number of cylces 1…100

Low- Pressure Controller CPC 2000

16 WIKA Operation Instruction Low Pressure Controller Version 0.1

:asd <1...100> Steps Down 1…100

:asu <1…100> Steps Up 1…100

:ate <0…10000> Waiting period at the end value (Phase 13) 1…10000 --> Time in seconds

:ath <1…10000> Hold time (Phase 10 and 16) 1…10000 --> Time in seconds

:atp <1…10000> Break time (Phase 18) 1…10000 --> Time in seconds

:atr <1…10000> Tolerance band 1…10000 --> Tolerance in 0,01% Fullscale. Hold time starts counting when controll over 1 second in this tolerance band.

:ats <1…10000> Start delay time (Phase 2) 1…10000 --> Time in seconds

:o <0 | 1> Output status information via interface 0 --> Off 1 --> On

:pa <-110…110> Increase desired pressure by x percent -110…110 --> Change set point in %

:pd Decrease current pressure value in % (Step DOWN).

:pr <-1100…11000> Adjustment of active measunring/ control range -1100…11000 --> new range in 0,01% FS

:ps <-10…110> Set pressure in percent -10…110 --> Desired pressure in %

:pu Increase current pressure value in % (Step UP).

:saz<0 | 1> Set Auto Zero (in Measure or Control Mode) 0 --> Off 1 --> ON

:sbr<0…1> Set Baudrate RS232 0 --> 1200 1 --> 2400 2 --> 4800 3 --> 9600 4 --> 14400 5 --> 19200 6 --> 28800 7 --> 38400 8 --> 56000 9 --> 57600

:sbu<0…1> Set Baudrate USB 0 --> 1200 1 --> 2400 2 --> 4800 3 --> 9600 4 --> 14400 5 --> 19200 6 --> 28800 7 --> 38400 8 --> 56000 9 --> 57600

Low- Pressure Controller CPC 2000

Version 0.1 WIKA Operation Instruction Low Pressure Controller 17

:sci<n | u | r> Set active Interface n --> digital interface off u --> USB aktiv r --> RS232 aktiv

sdb <0…100> Display brightness 0…100 --> Brightness in %

:spu <0…9> Pressure Unit 0 --> Pa 1 --> hPa 2 --> ka 3 --> mbar 4 --> bar 5 --> Torr 6 --> mmHg 7 --> inHg 8 --> psi 9 --> mmH2O 10 --> inH2O

:ssl <d | e> System language d = German e = English

:ssw <1…100> Step size 1…100 --> Steps size of the buttons UP /DOWN in manual mode in %

:swm <a , c , m , z , l , v , s> Set operation mode of the controller a --> Start Automatic Mode c --> Start Control Mode m --> Start Measure Mode (Only in Control or Measure Mode available) z --> ZERO (only in Control Mode available) l --> LEAK- Test l --> back to Control Mode (Like LEAK button) v --> VENT (Vents the whole system) (Only in Automatic Mode available)

:szc <0,1> Auto Zero Adjustment before entering Control Mode 0 Off 1 On

:szi <1…60> Auto Zero adjustment period 1…60 --> Zeit in Minuten

4.5. Drivers

Current interface driver will be available for download on Future Technology Devices International Limited (FTDI) homepage for all established OS (Windows, Linux, Mac Os).

Homepage URL: http://www.ftdichip.com Download URL: http://www.ftdichip.com/Drivers/VCP.htm (FTDI Homepage -->

Drivers --> VCPDrivers)

Low- Pressure Controller CPC 2000

18 WIKA Operation Instruction Low Pressure Controller Version 0.1

Installation guide for each OS: URL: http://www.ftdichip.com/Documents/InstallGuides.htm (FTDI Homepage --> Documents --> Installation Guide)

5. Troubleshooting

The individual responsible for the electrical connections must be notified immediately if the instrument is damaged or if errors occur that cannot be corrected as indicated in section 5.1. This individual must take the instrument out of service until the error has been corrected and ensure that it cannot be used unintentionally.

Only the manufacturer may perform repairs that require the housing to be opened.

Always unplug the power cord before opening the instrument!

If faults cannot be repaired, the system must be put out of operation immediately and this information is to be given to the manufacturer.

Repairs must only be carried out by the manufacturer. Interventions and changes on the appliance are not allowed.

In case of faults caused by defects of the electrical or pneumatic/hydraulic equipment the operators must inform their superiors immediately and call in the qualified and authorised technical staff for maintenance.

5.1. Troubleshooting measures

Problem Cause Correction Action

Instrument does not work, Display does not light up

No power Check to see if the electrical cord is plugged in properly

Battery is empty and has to bee recharged

Instrument does not reach set pressure; pump runs continuously

Leaks in system, Tubing diameter too large

Secure tubing properly; eliminate any leaks

Maximum tubing diameter 5 mm

Further help can be found through WIKA‗s Test & Calibration Technology Department.

Low- Pressure Controller CPC 2000

Version 0.1 WIKA Operation Instruction Low Pressure Controller 19

6. Re-calibrating and servicing

We recommend having the system re-calibrated by the manufacturer at regular intervals of approx. 12 months. Every re-calibration at the factory also includes a comprehensive and free checking of all system parameters. During the first re-calibration a service file is started automatically in which every re-calibration and all extra services are recorded. The instrument requires almost no maintenance, because all parts which are moveable are extremely robust. There are no parts which have to be serviced by the user. During each re-calibration the functioning of all components is checked and changed if necessary. To clean the membrane keypad and the display, use only customary plastic or glass cleaning agents in compliance with the guidelines of the manufacturer. Use cloths which are not prone to generate fluffs.

Before cleaning the surface of the instrument, it has to be made sure that it is not pressurized, the power is switched off and the power supply is disconnected.

Low- Pressure Controller CPC 2000

20 WIKA Operation Instruction Low Pressure Controller Version 0.1

7. Dismantling of the system

Work on electrical or pneumatic/hydraulic equipment must only be carried out by qualified and authorized technical staff, observing the corresponding safety regulations.

When dismantling the system please proceed as follows: 1. Make sure that there is no positive or negative overpressure on the system and that

all parts of the appliance are at room temperature. 2. Switch off the system from the main switch ON/OFF at the front. 3. First pull out the mains cable from the mains socket and then from the power supply

input socket of the system. 4. Disconnect the pressure connections. 5. Make sure that the system is free of any medium.

8. Transport of the system

Before the system is shipped all remains of any medium stuck to it must be removed. This is particularly important if the medium is a health hazard such as a corrosive, toxic, carcinogenic, radioactive, etc.

Before returning any equipment, please fill out the appropriate Goods-Return Form, found on the WIKA Web site (www.wika.de see: Service / Product return), and a reference number will be issued to you. Then send the instrument with this reference number attached to the delivery address indicated. Please follow the following instructions to prevent damage.

1. Wrap the system in antistatic plastic foil. 2. Using the insulating material, place the system in the box in such a way that there is

about the same amount of insulating material on all sides of the transport box. 3. If possible, add a bag of desiccant to the box 4. Make sure that the shipment is marked as transport of a highly sensitive measuring

instrument.

Low- Pressure Controller CPC 2000

Version 0.1 WIKA Operation Instruction Low Pressure Controller 21

9. Storage of the system

Before the system is stored all remains of any medium stuck to it must be removed. This is particularly important if the medium is a health hazard such as a corrosive, toxic, carcinogenic, radioactive, etc. The storage place must satisfy the following conditions:

Ambient temperature: -10 to +70 °C Humidity: 30 to 80 % relative humidity without condensation

Avoid the following influences:

Direct sunlight or vicinity to hot objects Mechanical vibration Soot, steam, dust and corrosive gasses Explosion-hazard environment, inflammable atmosphere

The system should be stored in its original transport box, in a place that meets the conditions listed above. Please follow the following instructions to avoid damage.

1. Wrap the system in antistatic plastic foil. 2. Using the insulating material, place the system in the box in such a way that there is

about the same amount of insulating material on all sides of the transport box. 3. If the system is stored for a longer time (more than 30 days) add a bag with desiccant

to the box.

10. Placing out of service

Before the system is shipped all remains of any medium stuck to it must be removed. This is particularly important if the medium is a health hazard such as a corrosive, toxic, carcinogenic, radioactive, etc.

The electronic components of the instrument contain environmentally hazardous materials and materials that can be reused. For this reason the instrument must be recycled in accordance with the environmental guidelines of the relevant jurisdiction once it has been taken permanently out of service. When placing the system out of service, please dismantle it according to the instructions in the manual under section: 5. „Dismantling of the system―. WIKA provides for appropriate disposal of used calibration instruments. For this purpose, simply return the Low-Pressure Controller CPC2000 to WIKA carriage prepaid.

Low- Pressure Controller CPC 2000

22 WIKA Operation Instruction Low Pressure Controller Version 0.1

11. Specifications

Specifications Model CPC 2000

Measuring ranges mbar 1 10 100 1000 Overpressure safety % FS 125 Accuracy % FS 0,3 0,1 0,1 0,1

Pressure typ Positive or negative gauge pressure and/or differential pressure

Unit Pa, kPa, hPa, bar, mbar, psi, inH2O, inHg, mmHg, Torr

Languages German, English

Zero automatically (in definable time gaps) manual (Zero-Taste)

Control- Step size 0 … 50% individual adjustable or 100% Control speed sec. ca. 5 (depends on the test volume) Pressure connections mm 6,6 x 11 (tube diameter d=6mm)

Material of wetted parts Ni, Al, CuBe, PU Instrument Version Bench top with handle Display Lighted multilane graphic display Resolution % FS 0,01 Response Time sec. 1 Keypad membrane keypad Pressure generation internal, electric pump

Digital interface RS-232 and USB Power supply 24VDC/1A Power consumption VA 24 Battery running time h ca. 8 Battery typ Li-Ion- Cells

Ingress protection IP 20

Permissible Media non- aggressive gases Operating temperature °C +10 … +40 Storage temperature °C -10 … +70 Humidity (relativ) % 30…80

Mass kg 4,6 Dimensions mm 102,6 (H) x 257 (B) x 271 (T) without handle

CE- Mark certificate of conformity

Calibration** Incl. 3.1 calibration report according to DIN EN 10 204

Low- Pressure Controller CPC 2000

Version 0.1 WIKA Operation Instruction Low Pressure Controller 23

11.1. Dimensions in mm

Front view

Top view

Low- Pressure Controller CPC 2000

24 WIKA Operation Instruction Low Pressure Controller Version 0.1

12. Accessories

Scope of supply Portable low-pressure controller CPC2000 Power lead 1.5 m with mains plug Operating instructions in English and German language 3.1 calibration report according to DIN EN 10 204

Options

DKD calibration certificate RS-232 digital interface cable