Referensdosimetri Crister Ceberg Medical Radiation Physics Lund University Sweden
Transcript
Page 1: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Referensdosimetri

Crister Ceberg

Medical Radiation Physics

Lund UniversitySweden

Page 2: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Reference dosimetry

• Determination of absorbed dose to water under reference conditions

• Not accounting for uncertainties related to non-reference conditions (for instance the patient...)

• High accuracy is needed (~ 1.5%)

• Requires a rigorous dosimetry protocol

Page 3: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Reference dosimetry protocols

• AAPM TG-51 (1999), North America

• DIN 6800-2 (1997, revised 2008), Germany

• NCS (1997), The Netherlands

• IPEMB (1996), UK

• IAEA TRS-398 (2000)

Page 4: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

IAEA TRS-398

• Published in 2000

• Replace previous protocol TRS-277

• Based on calibration of theinstruments in absorbed doseto water

Page 5: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

IAEA TRS-398

http://www-naweb.iaea.org/NAHU/DMRP/codeofpractice.html

Page 6: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

International measurement system

Page 7: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Traceability

• Bureau International des Poids et Mesures (BIPM) 1875– International Laboratory for SI units

• Primary Standard Dosimetry Laboratory (PSDL)– Widely acknowledged

– Highest metrological qualities

– Accepted without reference to other standards

– About 20 worldwide

• Secondary Standard Dosimetry Laboratory (SSDL)– Depends on calibration at a PSDL

• Clinical user at the hospital– Depends on calibration at SSDL or PDSL

Page 8: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Primary standard laboratory (PSDL)

• ISO definition of primary standard– “a standard that is designated or widely acknowledged as having

the highest metrological qualities and whose values is accepted without reference to other standards of the same quantity“

• This requires an absolute dosimeter,– ”that can be assembled and used to measure the absorbed dose

deposited in its own sensitive volume without calibration in a known field of radiation” (Attix)

Page 9: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Absolute dosimeters

• Ionization chamber (BIPM)

• Fricke dosimeter (PTB until 2006)

• Water calorimeter (NIST, NRC, PTB since 2006)

• Graphite calorimeter (NPL)

Page 10: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Ionization chamber

• Graphite ion chamber with well defined volume

• Designed to fulfill the Bragg-Gray conditions with very little perturbation

Page 11: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Ionization chamber

Boutillion and Perroche, PMB 38 (1993)

Page 12: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Ionization chamber

M/m

W

Boutillion and Perroche, PMB 38 (1993)

Page 13: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Fricke dosimeter

• Ferrous ions (Fe2+) in water

• Interaction with ionizing radiation leads radiolysis of water

• Iron ions are oxidized to ferric ions (Fe3+)

• The UV transmission spectrum is affected

• Response is determined by total absorption of electrons

Page 14: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Fricke dosimeter

Feist, PMB 27 (1982)

Page 15: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Calorimeter

• Measurable temperature increase in medium– Water

– Graphite

• Assumes that– All radiation energy is transfered to heat

– All heat comes from radiation energy

Page 16: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Calorimeter

• Measures directly the absorbed energy per unit mass

D=DTi·ci

Increasedtemperature [K]

Specific heat capacity[J kg-1 K-1]

Page 17: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Calorimeter

Ross and Klassen, PMB 41 (1996)

Page 18: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Calorimeter

Medin, Lund University

Page 19: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Calorimeter

B A

AC

Variable R

Thermistors

R

RVariable C

DU = UA-UB

Medin, Lund University

Page 20: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Calorimeter

0 60 120 180 240 300 360

-8

-6

-4

-2

0

2

4

6

DV

post-irradiation drift

pre-irradiation drift

Bri

dg

e o

utp

ut vo

lta

ge

(V

)

Time (s)

Medin, Lund University

Page 21: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Calorimeter

• Requires several correction factors

D=DTi·ci·kc·kp·kdd·(1-kHD)-1

kc: heat conductivity

kp: perturbation due to glass container

kdd: radiation field inhomogeneity

kHD: heat defect

Page 22: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Degree of equivalence

http://kcdb.bipm.org/

Page 23: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Uncertainty in Dw (1SD) at PSDL

• BIPM (ion chamber) 0.30%

• NIST (water calorimeter) 0.35%

• NRC (water calorimeter) 0.41%

• METAS (water calorimeter) 0.41%

• PTB (water calorimeter) 0.20%

• LSDG (water calorimeter) 0.66%

• NMi (graphite calorimeter) 0.40%

• ARPANSA (graphite calorimeter) 0.20%

• BEV (graphite calorimeter) 0.37%

• ENEA (graphite calorimeter) 0.44%

• LNHB (graphite calorimeter) 0.47%

Page 24: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

ND,W-based formalism

• Q0 denotes the reference beam quality at the standard laboratory

• DW,Q0 is known at the standard laboratory

• MQ0 is the dosimeter reading under reference conditions

• ND,W,Q0 is the dosimeter’s ”absorbed-dose-to-water” calibration coefficient

Dw,Q0 =MQ0 ND,w,Q0

Page 25: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Correction for radiation quality

• Q denotes the quality at the user

• DW,Q is the absorbed dose in the user’s beam

• MQ is the dosimeter reading at the user

• kQ,Q0 corrects for the effects of the difference between the reference beam quality and the beam quality at the user

Dw,Q =MQ ND,w,Q=MQ ND,w,Q0 kQ,Q0

Page 26: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

The beam quality correction factor

• Most often, the reference beam quality is 60Co,and then kQ,Q0=kQ

ND,w,Q Dw,Q/MQ

ND,w,Q0 Dw,Q0/MQ0kQ,Q0= =

Page 27: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Experimental determination of kQ,Q0

• Ideally, the kQ,Q0 factor should be measured for each chamber at the desired beam quality

• However this requires– Standard laboratories with clinical beam qualities

– Independent dosimetry (e.g. calorimeters) operating at these beam qualities

Page 28: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Theoretical determination of kQ,Q0

• Instead, theoretically determined kQ,Q0 factors are tabulated in the IAEA TRS 398 protocol– for different chamber types

– for different beam qualities

Page 29: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

0.2%

Page 30: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Beam quality

• For photons TPR20,10

– TPR20,10=TPR(20)/TPR(10)

– TPR20,10=1.2661 PDD20,10 – 0.0595

• For electrons R50

– Depth where absorbed dose is 50% of maximum

– R50=1.029 R50,ion - 0.06 (R50,ion ≤ 10 cm)

– R50=1.059 R50,ion - 0.37 (R50,ion > 10 cm)6 MeV elektoner

0

10

20

30

40

50

60

70

80

90

100

0 5 10 15 20 25 30 35 40 45 50

Djup i vatten (mm)

Rela

tiv d

os

Page 31: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Practical details before use

• Store instruments in a safe place (dry, normal room temperature, clean, etc.)

• Inspect the instrument before use

• Make an x-ray image of the instrument at the time of purchase, or if problems arise

Page 32: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Stability check

• Check and document long term stability(both detector and electrometer)

• Perform stability check before and after the detectoris sent for calibration at the standards laboratory

Page 33: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Practical details during measurement

• Allow time to reach thermal equilibrium

• Turn on the electrometer at least 1-2 h before use

• Always collect several measurements (5-10)

• Pay attention to trends (can be sign of equipment failure)

• Leakage should be < 0.1% of MQ

Page 34: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Pre-irradiation effects

• Pre-irradiate the chamber to reach charge equilibrium in the exposed materials

• After change of voltage or polarity, be careful with re-stabilisation (> 20 min), consider using normalisationto an external monitor chamber

Page 35: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Pre-irradiation effects

• Normalized response for an NE 2571 to 60Co irradiation

McCaffrey et al. PMB 2005

Page 36: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Correction of MQ for influence quantities

• Uncorrected reading: M

• Temperature and pressure: kTP

• Electrometer correction factor: kelec

• Polarity effect: kpol

• Recombination effect: ks

• Humidity: kh (rarely used)

MQ=M kTP kelec kpol ks

Page 37: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Correction of MQ for influence quantities

• Uncorrected reading: M

• Temperature and pressure: kTP

• Electrometer correction factor: kelec

• Polarity effect: kpol

• Recombination effect: ks

• Humidity: kh (rarely used)

MQ=M kTP kelec kpol ks

Given by the SSDLtogether with ND,w

Not needed if calibratedat 50% humidity (20-80%)

Page 38: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Temperature and pressure

• The temperature and pressure correction converts the cavity air mass to the reference conditions:

where the reference conditions generally are– T0=20 C

– P0=101.3 kPa

(273.2+T) P0

(273.2+T0) PkTP=

Page 39: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Polarity

• Polarity correction for asymmetric charge collection(mean value of readings at both polarities):

where M+ and M- are the readings at positive and negative polarity, respectively

• If not performed at the standards laboratory,a correction for the relative effect can be done

|M+|+|M-|2M

kpol=

Page 40: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Recombination

• The recombination factor corrects for incomplete charge collection using the ”two-voltage” method:

where M1 and M2 are the readings at two different operating voltages V1 and V2, such that V1 ≥ 3 V2,and a0 - a2 are constants found in the TRS 398 protocol

• If not performed at the standards laboratory,a correction for the relative effect can be done

M1

M2

ks=a0+a1 + a2( ) M1

M2( )

2

Page 41: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Reference dosimetry in the user beam

• The absorbed dose to water is then given by

where MQ is corrected for influence quantities,and kQ,Q0 is determined based on the user’s beam quality

Dw,Q =MQ ND,w,Q0 kQ,Q0

Page 42: Crister Ceberg Medical Radiation Physics Lund … University Sweden Reference dosimetry • Determination of absorbed dose to water under reference conditions • Not accounting for

Reference dosimetry in the user beam

• Take care that the linac is operating properly!

• If not, accurate reference dosimetrycan be a source of errors


Recommended