CRUISE REPORT: NS02 (Updated FEB 2013) Highlights Cruise Summary Information WOCE Section Designation NS02 Expedition designation (ExpoCodes) 316N20020530 Alias NSEAS, 316N166_11, 316N200206, Nordic 02, NS2 Chief Scientists Dr. William M. Smethie, Jr. / LDEO Dates 2002 May 30 - 2002 Jul 1 Ship R/V Knorr Ports of call Reykjavik, Iceland - Glasgow, Scotland Geographic Boundaries 78° 48.48' N 18° 31.61' W 20° 2.96' E 59° 35.91' N Stations 159 Floats and drifters deployed 0 Moorings deployed or recovered 0 Contact Information: Dr. William M. Smethie, Jr. Lamont-Doherty Earth Observatory Tel: (845) 365-8566 129 Comer Fax: (845) 365-8176 61 Route 9W - PO Box 1000 [email protected] Palisades NY 10964-8000 US
Transcript
Page 1: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

CRUISE REPORT: NS02 (Updated FEB 2013)

Highlights

Cruise Summary Information

WOCE Section Designation NS02 Expedition designation (ExpoCodes) 316N20020530

Alias NSEAS, 316N166_11, 316N200206, Nordic 02, NS2 Chief Scientists Dr. William M. Smethie, Jr. / LDEO

Dates 2002 May 30 - 2002 Jul 1 Ship R/V Knorr

Ports of call Reykjavik, Iceland - Glasgow, Scotland

Geographic Boundaries 78° 48.48' N

18° 31.61' W 20° 2.96' E 59° 35.91' N

Stations 159 Floats and drifters deployed 0

Moorings deployed or recovered 0

Contact Information:

Dr. William M. Smethie, Jr. Lamont-Doherty Earth Observatory Tel: (845) 365-8566 129 Comer Fax: (845) 365-8176 61 Route 9W - PO Box 1000 [email protected] Palisades NY 10964-8000 US

Page 2: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

Links To Select Topics Shaded sections are not relevant to this cruise or were not available when this report was compiled.

Cruise Summary Information Hydrographic Measurements Description of Scientific Program CTD Data:

Geographic Boundaries Acquisition Cruise Track (Figure): PI CCHDO Processing Description of Stations Calibration Description of Parameters Sampled Temperature Pressure Bottle Depth Distributions (Figure) Salinities Oxygens

Floats and Drifters Deployed Bottle Data Moorings Deployed or Recovered Salinity Oxygen Principal Investigators Nutrients Cruise Participants Carbon System Parameters CFCs Problems and Goals Not Achieved Helium / Tritium Other Incidents of Note Radiocarbon

Underway Data Information References Navigation Bathymetry Nutrients

Acoustic Doppler Current Profiler (ADCP) CFCs Thermosalinograph Carbon System Parameters XBT and/or XCTD Meteorological Observations Acknowledgments Atmospheric Chemistry Data

Report Data Processing Notes

Page 3: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

Nordic SeasR/V Knorr Voyage 166-1130 May 2002 - 1 July 2002

Reykjavik, Iceland - Glasgow, Scotland

Chief Scientist: Dr. William M. Smethie, Jr.Lamont-Doherty Earth Observatory

Columbia University

20˚W

20˚W

20˚E

20˚E

60˚N 60˚N

70˚N 70˚N

80˚N 80˚N

500

500

1000

1000

1000

1000

1500

2000

2500

3000

3000

3000

3500

3500

20˚W

20˚W

20˚E

20˚E

60˚N 60˚N

70˚N 70˚N

80˚N 80˚N

001

033 039

040

050

06706

8

095

096

124

125

136137

138

146

147

159

Reykjavik

Glasgow

Tromso

Nordic Seas Cruise Track

ODF Preliminary Cruise Report19 December 2002

Data Submitted by:

Oceanographic Data FacilityScripps Institution of Oceanography

La Jolla, Ca. 92093-0214

Page 4: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate
Page 5: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

1. Summary

A hydrographic survey consisting of CTD/rosette sections in the North Atlantic was carried out May to July 2002.The R/V Knorr departed Reykjavik, Iceland on 30 May 2002. 159 CTD/Rosette stations were occupied from 1-30June. 3286 bottles were tripped resulting in 3255 usable bottles. Water samples (up to 36) and CTDO data werecollected in most cases to within 10 meters of the bottom. Salinity, dissolved oxygen and nutrient samples wereanalyzed from every bottle sampled on the rosette. The cruise ended in Glasgow, Scotland on 1 July 2002.

2. Personnel

Name Affiliation Duties

Alfimov, Vassili Ion Physics Div.,Uppsala Univ. I129/SF6Anderson, George C. SIO/MPL Nutrients/RosetteBahr, Frank WHOI LADCP/RosetteBellerby, Richard G.J. *BCCR TCO2/PCO2/Alkalinity/pHCalderwood, John K. SIO/STS/SEG ET/O2/RosetteDachille, Anthony LDEO Helium/Tritium/O18Ghan, Ryan LDEO CFCsGorman, Eugene LDEO CFCsGorodetskaia, Irina LDEO CTD Console/LADCP/Sample CopHarris, Grant B. WHOI SSSG TechnicianHiller, Scott M. SIO/STS/SEG Watch Leader/ET/SaltsJohnson, Mary Carol SIO/STS/ODF CTD Data ProcessingLaird, Robert S. WHOI SSSG TechnicianMathieu, Guy LDEO/Retired CFCsMattson, Carl W. SIO/STS/SEG TIC/Watch Leader/ET/SaltsMessias, Marie-Jose Univ.of East Anglia,UK SF6Mignon, Benoit *BCCR TCO2/PCO2/Alkalinity/pHMuus, David A. SIO/PORD Bottle Data ProcessingNondal, Gisle Univ.of Bergen TCO2/PCO2/Alkalinity/pHPatrick, Ronald G. SIO/STS/ODF O2/RosetteQuiroz, Erik W. Univ.of So.Miss. NutrientsSearson, Sarah LDEO CFCsSequeira, Sandra *LODYC I129/SF6Smethie, William M.,Jr. LDEO Chief ScientistSmith, Helen B. Univ.of East Anglia,UK SF6Swift, James H. SIO/STS/ODF co-Chief Scientist

*BCCR Bjerknes Centre for Climate Research*LODYC Laboratoire d’Oceanographie Dynamique et de Climatologie

Table 2.0 Scientific Personnel Nordic Seas

3. Scientific Programs

Analysis Institution Principal Investigator

Basic Hydrography (Salinity,O2,Nutrients,CTD) SIO J.SwiftCFCs LDEO W.SmethieI129/SF6 LDEO W.SmethieHe/Tr/O18 LDEO P.SchlosserpCO2/TCO2/Alkalinity/PH BCCR R.BellerbyADCP and LADCP WHOI F.BahrUW Meteorology WHOI

Table 3.0 Principal Programs of Nordic Seas

Page 6: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-2-

The SIO ODF hydrographic measurements program is described in detail in this report.

4. Description of Measurement Techniques

4.1. Hydrographic Measurements Program

The basic hydrography program consisted of salinity, dissolved oxygen and nutrient (nitrite, nitrate, phosphate andsilicate) measurements made from bottles taken on CTD/rosette casts, plus pressure, temperature, salinity anddissolved oxygen from CTD profiles. 159 CTD/rosette casts were made, usually to within 10 meters of the bottom.The resulting data set met and in many cases exceeded WHP specifications. The distribution of samples isillustrated in figures 4.1.0-4.1.4.

4000

3000

2000

1000

0

DE

PTH

(M

ET

ER

S)

-100 100 300 500 700 900 1100 1300 1500 1700DISTANCE (Km)

Nordic Seas 2002 - KN166/XI R/V Knorr

001

67 0.10 N 14 59.90 W

005 008 011 014 018 021 024 027 030 033 037

Figure 4.1.0 Sample distribution, stations 1-39.

4000

3000

2000

1000

0

DE

PTH

(M

ET

ER

S)

2000 2200 2400 2600 2800 3000 3200DISTANCE (Km)

Nordic Seas 2002 - KN166/XI R/V Knorr

040 043 046 050 053 056 060 064

Figure 4.1.1 Sample distribution, stations 40-67.

Page 7: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-3-

4000

3000

2000

1000

0

DE

PTH

(M

ET

ER

S)

3300 3500 3700 3900 4100 4300 4500 4700DISTANCE (Km)

Nordic Seas 2002 - KN166/XI R/V Knorr

068 073 076 078 080 082 085 089 091 094

Figure 4.1.2 Sample distribution, stations 68-95.

4000

3000

2000

1000

0

DE

PTH

(M

ET

ER

S)

5100 5300 5500 5700 5900 6100 6300DISTANCE (Km)

Nordic Seas 2002 - KN166/XI R/V Knorr

096 100 104 107 109 111 114 118 121

Figure 4.1.3 Sample distribution, stations 96-124.

4000

3000

2000

1000

0

DE

PTH

(M

ET

ER

S)

6800 7000 7200 7400 7600 7800 8000 8200 8400DISTANCE (Km)

Nordic Seas 2002 - KN166/XI R/V Knorr

125 128 131 138 142 145 147 151 157

59 36.10 N 3 30.00 W

Figure 4.1.4 Sample distribution, stations 125-159.

Page 8: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-4-

4.2. Water Sampling Package

Hydrographic casts were performed with a rosette system consisting of a 36-bottle rosette frame (ODF), a 36-placepylon (SBE32) and 36 10-liter PVC bottles (ODF). Underwater electronic components consisted of an ODF-modified NBIS MKIII CTD (ODF #5) with dual conductivity and temperature sensors, Sea-Bird SBE43 oxygensensor with pump, RDI LADCP and Simrad altimeter.

Equipment used on stations 1-159:36-bottle rosette frame ODF s/n RSS1-94Bullister 10-liter bottles ODF 1-3636-place Water Sampler Sea-Bird SBE32 s/n 3213290-0113Digital Reversing Temperature Sensor Sea-Bird SBE35 s/n 3516590-0011ODF CTD #5 NBIS MKIIIB s/n 01-1070

Pressure Paine s/n 77017Temperature#1 Rosemount 171BJ s/n 13407Temperature#2 Rosemount 171BJ s/n 17534Conductivity#1 GO 09035-00151 s/n P41Conductivity#2 GO 09035-00151 s/n O24

Oxygen Sensor Sea-Bird SBE43 s/n 430072Pump for Oxygen Sensor Sea-Bird SBE5T s/n 052146Altimeter Simrad 1007 s/n 0201074Battery Pack for Pump and Altimeter ODFLADCP RDI CS-150KHZ BB s/n 2644

XTC Firmware v2.04CPU Firmware v5.59

LADCP Battery Pack WHOI

Additional equipment used only on Stations 138-146:Data Logger/Controller (for ODF CTD #5) ODFCTD Sea-Bird 9plus s/n 09P19275-0576

Pressure Paroscientific Digiquartz s/n 75639Temperature SBE3P s/n 4035Conductivity SBE4C s/n 2178

Table 4.2.0 Underwater sampling package.

CTD #5 was mounted horizontally along the bottom of the rosette frame, with the SBE43 dissolved oxygen andSBE35 PRT sensors deployed next to the CTD. The altimeter reported distance-above-bottom. The dissolvedoxygen sensor and altimeter were interfaced with the CTD, and their data were incorporated into the CTD datastream. The LADCP was vertically mounted to the frame inside the bottle rings. The rosette system was suspendedfrom a three-conductor 0.322" electro-mechanical cable. Power to the CTD and pylon was provided through thecable from the ship. The R/V Knorr’s port-side Markey CTD winch was used throughout the leg.

The deck watch prepared the rosette approximately 45 minutes prior to each cast. All valves, vents and lanyardswere checked for proper orientation. The bottles were cocked and all hardware and connections rechecked. Time,position and bottom depth were logged by the console operator at arrival on station. The rosette was moved intoposition under a projecting boom from the starboard ("forward") hangar using an air-powered cart on tracks. Tw ostabilizing tag lines were threaded through rings on the frame, and CTD sensor covers were removed. As directedby the watch leader, the winch operator raised the package, extended the boom over the side of the ship and quicklylowered the package into the water; then the tag lines were removed.

Each rosette cast was lowered to within 6-20 meters of the bottom. Bottles on the rosette were identified withunique serial numbers. These numbers corresponded to the pylon tripping sequence 1-36, the first trip closing bottle#1. No bottles were changed out during the leg, although parts of bottles may have been replaced or repaired.

Av erages of CTD data corresponding to the time of bottle closure were associated with the bottle data during a cast.Pressure, depth, temperature, salinity and density were immediately available to facilitate examination and qualitycontrol of the bottle data as the sampling and laboratory analyses progressed.

Page 9: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-5-

Recovering the package at the end of deployment was essentially the reverse of launching, with the additional use ofair-tuggers for added stabilization. The rosette was moved into the starboard hangar for sampling. The bottles androsette were examined before samples were taken, and anything unusual was noted on a sample log for each cast.

Routine CTD maintenance included soaking the conductivity and CTD O2 sensors in distilled water between casts tomaintain sensor stability. The rosette was stored in the starboard side hangar between casts to insure the CTD wasnot exposed to direct sunlight or wind, in order to maintain the internal CTD temperature near ambient airtemperature.

Rosette maintenance was performed on a regular basis. O-rings were changed as necessary and bottle maintenancewas performed each day to insure proper closure and sealing. Valves were inspected for leaks and repaired orreplaced as needed.

4.3. Underwater Electronics Packages

CTD data were collected with a modified NBIS MKIII CTD (ODF CTD #5). This instrument provided pressure,temperature, conductivity and dissolved O2 channels, and additionally measured a second PRT temperature andconductivity as a calibration check and backup. Other data channels included elapsed-time, altimeter, accelerometer,water-leak detector and several power supply voltages. CTD #5 supplied a non-standard 17-byte (NBIS-format + 2bytes) data stream at a data rate of 20 Hz. Modifications to the instrument included a revised pressure sensormounting, a four-channel 14-bit A/D converter, implementation of 8-bit and 16-bit multiplexer channels, an elapsed-time channel, instrument ID in the polarity byte and power supply voltages channels. The CTD sensor configurationis provided in Table 4.2.0 in the previous section.

The CTD pressure sensor mounting had been modified to reduce the dynamic thermal effects on pressure. Thesensor was attached to a length of coiled, oil-filled stainless-steel tubing threaded into the end-cap pressure port.The transducer was also insulated. The NBIS temperature compensation circuit on the pressure interface wasdisabled; all thermal response characteristics were modeled and corrected in software.

The secondary CTD temperature and conductivity sensors, mounted in a single turret, could have been used tocalculate coherent salinities if the primary sensors failed. However, they were only used as a secondary temperaturecalibration reference or to occasionally verify unusual T/S structures observed in the primary sensors.

An SBE35 laboratory-grade reference PRT was employed as an additional temperature calibration check. This is aninternally recording device triggered by the SBE32 pylon confirmation signal, providing a calibration point for eachbottle trip. The SBE35 was used on stations 1-136.

An SBE43 dissolved O2 sensor was located near the primary CTD temperature sensor. The sensor was pumpedusing an SBE5T running at ∼3000rpm with a flow rate of ∼25ml/s. The SBE43 signal was interfaced directly to theMKIII CTD and digitized using one of the 4 channels of the 14-bit A/D converter.

An SBE32 36-place carousel was the water sampler control unit on the rosette. The SBE32 has the advantage ofrequiring a single sea cable conductor for power and signals, and providing for the use of the SBE35 reference PRT.

The CTD system was configured for single-conductor operation by combining together the 3 sea cable conductors.An SBE33 deck box located in the Knorr’s Main Lab supplied power and telemetry control for the rosette watersampler and MKIII CTD. Data from the MKIII CTD were routed to the SBE32 underwater unit and sent up thewire to the ship. The SBE33 deck box decoupled the signal and sent it to the ODF CTD demodulator unit, whichconverted it to an RS232 9600-baud binary data stream. The binary data were fed into the main CTD acquisitioncomputer. Bottle-trip commands were sent from this computer to the SBE33 deck box, which transmitted thecommands down the cable to the SBE32 water sampler unit.

Both SBE33 deck boxes failed after station 137. An ODF data logger unit was installed on the rosette so that CTD#5 data could continue to be collected for data consistency. The logger internally stored CTD #5 data at the full-stream data rate during stations 138-146. The Knorr’s SBE9plus CTD was also installed on the rosette, then usedwith an SBE11plus deck unit and a Sea-Bird DOS-based acquisition system to control the SBE32 water sampler andto allow the operator to select and trip bottles. The altimeter was connected to the SBE9plus to allow detection ofbottom depth during the cast. The digitized CTD #5 data from the ODF data logger were dumped onto the dataprocessing computer after each cast and processed as usual. The SBE9plus CTD data were retained for bottle tripinformation only. The SBE33 deck boxes were repaired and reinstalled prior to station 147, and CTD operationscontinued as they were for the first 137 stations. The SBE9plus and data logger remained on the rosette but were not

Page 10: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-6-

used after station 146.

4.4. Navigation and Bathymetry Data Acquisition

P-code navigation data were acquired from the ship’s Trimble Tasman GPS receiver via RS-232. Data were loggedautomatically at 12-second intervals by the Linux computer. Underway bathymetry was logged every 2 seconds bythe ship’s computer system, recording a corrected SeaBeam 2112 center-beam depth. Unedited depth data weremerged with the navigation data to provide a time-series of underway position, course, speed and bathymetry data.These data were used for all station positions and for cast bottom depths.

4.5. CTD Data Acquisition and Real-Time Control System

The CTD data acquisition and real-time control system consisted of a generic PC workstation running RedHat 7.2Linux, ODF-built CTD and pylon deck units, CTD and pylon power supplies, and a VCR recorder for real-timeanalog backup recording of the sea-cable signal. The Linux system consisted of a color display with 3-button mouseand keyboard (the CTD console), 4 RS-232 ports, 40-GB disk and CD-R drive. Two other Linux systems werenetworked to the data acquisition system, as well as to the rest of the networked computers aboard the Knorr. Thesesystems were available for real-time CTD data display and provided for CTD and hydrographic data managementand backup. One HP 1600C color inkjet printer and various postscript printers on the Knorr provided hardcopyfrom any of the workstations.

The CTD FSK signal was demodulated and converted to a 9600 baud RS-232C binary data stream by the CTD deckunit. This data stream was fed to the CTD acquisition computer. The pylon deck unit was connected to the CTDacquisition computer through a bi-directional 300 baud serial line, allowing bottle trips to be initiated and confirmedby the data acquisition software. A bitmapped color display provided interactive graphical display and control of theCTD rosette sampling system, including real-time raw and processed CTD data, navigation, and rosette trip displays.

The CTD data acquisition, processing and control system was prepared by the console watch a few minutes beforeeach deployment. A console operations log was maintained for each deployment, containing a record of everyattempt to trip a bottle as well as any pertinent comments. Most CTD console control functions, including startingthe data acquisition, were initiated by pointing and clicking a mouse cursor on the display at icons representingfunctions to perform. The system then presented the operator with short dialog prompts with automatically-generated choices that could either be accepted as defaults or overridden. The operator was instructed to turn on theCTD and pylon power supplies, then to examine a real-time CTD data display on the screen for stable voltages fromthe underwater unit. Once this was accomplished, the data acquisition and processing were begun and a time andposition were automatically logged for the beginning of the cast. A backup analog recording of the CTD signal on aVCR tape was started at the same time as the data acquisition. A rosette trip display and pylon control windowpopped up, giving visual confirmation that the pylon was initializing properly. Various plots and displays wereinitiated. When all was ready, the console operator informed the deck watch by radio.

Once the deck watch had deployed the rosette and informed the console operator that the rosette was at the surface(also confirmed by the computer displays), the console operator or watch leader provided the winch operator with atarget depth (wire-out) and maximum lowering rate, normally 60 meters/minute for this package. The package thenbegan its descent, building up to the maximum rate during the first few hundred meters, then optimally continuing ata steady rate without any stops during the down cast.

The console operator examined the processed CTD data during descent via interactive plot windows on the display,which could also be run at other workstations on the network. Additionally, the operator decided where to tripbottles on the up cast, noting this on the console log. The altimeter signal was also monitored for bottom proximity.

Around 100-200 meters above the bottom, depending on bottom conditions, the altimeter typically began signaling abottom return on the console. The winch speed was usually slowed to ∼30 meters/minute during the final approach.The winch and altimeter displays allowed the watch leader to refine the target depth relayed to the winch operatorand safely approach to within 10-20 meters of the bottom.

Bottles were closed on the up cast by pointing the console mouse cursor at a graphic firing control and clicking abutton. The data acquisition system responded with the CTD rosette trip data and a pylon confirmation message in awindow. All tripping attempts were noted on the console log. The console operator then instructed the winchoperator to bring the rosette up to the next bottle depth. The console operator was also responsible for generatingthe sample log for the cast.

Page 11: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-7-

After the last bottle was tripped, the console operator directed the deck watch to bring the rosette on deck. Once therosette was on deck, the console operator terminated the data acquisition and turned off the CTD, pylon and VCRrecording. The VCR tape was filed. Usually the console operator also brought the sample log to the rosette roomand served as the sample cop.

4.6. CTD Data Processing

ODF CTD processing software consists of over 30 programs running under the Linux operating system. The initialCTD processing program (ctdba) is used either in real-time or with existing raw data sets to:

• Convert raw CTD scans into scaled engineering units, and assign the data to logical channels• Filter various channels according to specified filtering criteria• Apply sensor- or instrument-specific response-correction models• Provide periodic averages of the channels corresponding to the output time-series interval• Store the output time-series in a CTD-independent format

Once the CTD data are reduced to a standard-format time-series, they can be manipulated in various ways.Channels can be additionally filtered. The time-series can be split up into shorter time-series or pasted together toform longer time-series. A time-series can be transformed into a pressure-series, or into a larger-interval time-series.The pressure calibration corrections are applied during reduction of the data to time-series. Temperature,conductivity and oxygen corrections to the series are maintained in separate files and are applied whenever the dataare accessed.

ODF data acquisition software acquired and processed the CTD data in real-time, providing calibrated, processeddata for interactive plotting and reporting during a cast. The 20 Hz data from the CTD were filtered, response-corrected and averaged to a 2 Hz (0.5-second) time-series. Sensor correction and calibration models were applied topressure, temperature, conductivity and O2. Rosette trip data were extracted from this time-series in response to tripinitiation and confirmation signals. The calibrated 2 Hz time-series data, as well as the 20 Hz raw data, were storedon disk and were available in real-time for reporting and graphical display. At the end of the cast, variousconsistency and calibration checks were performed, and a 2-db pressure-series of the down cast was generated andsubsequently used for reports and plots.

CTD plots generated automatically at the completion of deployment were checked daily for potential problems. Thetwo PRT temperature sensors were inter-calibrated and checked for sensor drift. The CTD conductivity sensor wasmonitored by comparing CTD values to check-sample conductivities, and by deep theta-salinity comparisonsbetween down and up casts as well as adjacent stations. The CTD O2 sensor was calibrated to check-sample data.

A few casts exhibited conductivity offsets or noise due to biological or particulate artifacts. Many casts were subjectto noise in the data stream caused by sea cable, slip-ring or deck box problems (especially stations 40-137); or bymoisture in interconnect cables between the CTD and external sensors (i.e. O2). Intermittent noisy data were filteredout of the 2 Hz data using a spike-removal filter. A least-squares polynomial of specified order was fit to fixed-length segments of data. Points exceeding a specified multiple of the residual standard deviation were replaced bythe polynomial value.

Density inversions can be induced in high-gradient regions by ship-generated vertical motion of the rosette.Detailed examination of the raw data shows significant mixing occurring in these areas because of "ship roll". Inorder to minimize density inversions, a ship-roll filter was applied to all casts during pressure-sequencing to disallowpressure reversals. The first few seconds of in-water data were excluded from the pressure-series data, since thesensors were still adjusting to the going-in-water transition.

Pressure intervals with no time-series data can optionally be filled by double-quadratic interpolation/extrapolation.The only pressure intervals missing/filled during this leg were at 0 db, caused by chopping off going-in-watertransition data during pressure-sequencing.

When the down-cast CTD data have excessive noise, gaps or offsets, the up-cast data are used instead. CTD datafrom down and up casts are not mixed together in the pressure-series data because they do not represent identicalwater columns (due to ship movement, wire angles, etc.). It was not necessary to use any up casts for Nordic SeasCTD data.

Page 12: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-8-

4.7. CTD Laboratory Calibration Procedures

Laboratory calibrations of the CTD pressure and temperature sensors were used to generate tables of correctionsapplied by the CTD data acquisition and processing software at sea.

Pressure and temperature calibrations were last performed on CTD #5 at the ODF Calibration Facility (La Jolla) inMarch 2002, prior to Nordic Seas.

The CTD pressure transducer (Paine 211-35-440-05 8850 psi, s/n #77017) was calibrated in a temperature-controlled water bath to a Ruska Model 2400 Piston Gauge pressure reference. Calibration curves were measured at5 temperatures from -1.31 to 25.65°C to three maximum loading pressures (1191, 4152 and 6080 db). Figure 4.7.0summarizes the laboratory pressure calibration performed in March 2002.

-10

0

DW

T-C

TD

(db

)

0 1000 2000 3000 4000 5000 6000Pressure (db)

ODF CTD #5 Mar ’02

= -1.31 = 2.84 = 4.94 = 17.59 = 25.65

Figure 4.7.0 Pressure Calibration for ODF CTD #5, March 2002.

CTD PRT temperatures were calibrated to a NBIS ATB-1250 resistance bridge and Rosemount standard PRT. Theprimary (Rosemount 171BJ, s/n #13407) and secondary (Rosemount 171BJ, s/n #17534) CTD temperatures wereoffset by 1.5°C to avoid the 0-point discontinuity inherent in the internal digitizing circuitry. Figure 4.7.1summarizes the laboratory temperature calibration performed on the primary PRT March, 2002.

-1.502

-1.501

-1.500

-1.499

-1.498

-1.497

-1.496

-1.495

-1.494

-1.493

-1.492

0 10 20 30

AT

B-P

RT

1 (d

eg C

)

ITS-90 PRT1 Temp (deg C)ODF CTD #5 Mar ’02

order= 2

1.600883e-05-4.125136e-04-1.496818e+00

r=0.9262752 p=1.0000000sd=0.0004081 n= 305

Figure 4.7.1 Primary PRT Temperature Calibration for ODF CTD #5, March 2002.

These calibration procedures are being repeated post-cruise at ODF.

4.8. CTD Shipboard Calibration Procedures

ODF CTD #5 was used for all Nordic Seas casts. A redundant PRT sensor was used on CTD #5 as a calibrationcheck while at sea. An SBE35 laboratory-grade reference PRT was deployed on the rosette as an additional check of

Page 13: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-9-

the primary and secondary PRT temperatures at bottle trip levels. CTD conductivity and dissolved O2 werecalibrated to in-situ check samples collected during each rosette cast.

4.8.1. CTD #5 Pressure

Pre-cruise pressure calibration data were applied to CTD #5 raw pressures during each cast. Down-cast surfaceoffsets were automatically adjusted to 0 as the CTD entered the water; any difference between this value and thecalibration value was automatically adjusted during the top 50 decibars.

Residual offsets at the end of each up-cast (the difference between the last pressure in-water and 0) were monitoredduring the cruise to check for shifts in the pressure calibration. Most residual differences were less than 1.2 decibar.Preliminary checks of the post-cruise calibrations indicate a maximum 0.5 decibar shift at any pressure/temperaturecombination, indicating there was likely no significant shift in pressure calibration during the cruise. CTD pressuredata will not be considered final until after the post-cruise laboratory calibrations have been completed and analyzed.

4.8.2. CTD #5 Temperature

Pre-cruise laboratory calibrations for the CTD #5 primary temperature sensor (PRT1) were applied to all shipboardCTD data.

A second Rosemount PRT sensor was deployed as the secondary temperature channel and compared with theprimary PRT channel on all casts to monitor for drift. The response times of the sensors were first matched, thenpreliminary corrected temperatures were compared for a series of standard depths from each CTD down-cast.

Comparison of the two CTD #5 PRTs at intermittent down-cast pressures deeper than 500 decibars showed thedifference increasing by about +0.001°C during the cruise. Figure 4.8.2.0 summarizes the shipboard comparisonbetween the primary and secondary PRT channels for CTD #5.

-0.02

-0.01

0.00

0.01

0.02

0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160

PRT

1-PR

T2

(Deg

C)

Station Number

order= 1

7.864073e-06 2.849901e-04

r=0.3361281 p=1.0000000sd=0.0008885 n= 619

Figure 4.8.2.0 Shipboard comparison of CTD #5 dual PRTs, PRT1-PRT2, pressure>500db.

The comparison of CTD #5 PRT1 with the SBE35 reference PRT at up-cast bottle trips showed the differencedecreasing by about -0.001°C at pressures deeper than 500 decibars during the cruise. Figure 4.8.2.1 summarizesthe comparison between the CTD #5 primary PRT and the SBE35 reference PRT temperatures.

Page 14: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-10-

-0.02

-0.01

0.00

0.01

0.02

0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160

PRT

1-SB

E35

(de

gC)

Station #CTD#5-PRT1 minus SBE35 PRT Temperatures

order= 1

-3.799191e-06-8.933949e-04

r=0.1671862 p=1.0000000sd=0.0008757 n= 1326

Figure 4.8.2.1 Comparison between CTD #5 primary and SBE35 PRTs, PRT1-SBE35, pressure>500db.

A preliminary check of the post-cruise calibration indicates PRT1 shifted up to 0.001°C between calibrations attemperature ranges applicable to this cruise. CTD temperature data will not be considered final until after the post-cruise laboratory calibrations have been completed and analyzed.

4.8.3. CTD #5 Conductivity

Corrected CTD rosette trip pressures and temperatures were used with bottle salinities to calculate bottleconductivities. Differences between the bottle and CTD conductivities were then used to derive a conductivitycorrection. This correction is normally linear for the 3-cm conductivity cell used in the Mark III CTD.

Conductivity differences above and below the thermocline were fit to CTD conductivity for stations 1-111 todetermine a conductivity slope. A first-order fit was calculated, with outlying values (4,2 standard deviations)rejected. Figure 4.8.3.0 shows the data used to determine the Nordic Seas preliminary conductivity slope.

-1600

-1500

-1400

-1300

-1200

-1100

-1000

27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43

1000

*CT

D-B

ottle

Con

duct

ivity

CTD Conductivity (mS/cm) CTD #5, stations 1-111 prelim. conductivity slope, pressure < 15 db or > 250 db only, after 4,2sdev rej

order= 1

-3.677969e+01 8.264711e+00

r=0.9997019 p=1.0000000sd=1.5196348 n= 1605

Figure 4.8.3.0 Nordic Seas CTD #5 preliminary conductivity slope, using stations 1-111.

Once the preliminary conductivity slope was applied, residual CTD conductivity offset values were calculated forstations 1-111 using bottle conductivities deeper than 700 db. Data from all pressure levels were used to determinethe offsets for stations 46-67, which were all shallower than 700 db. Figure 4.8.3.1 illustrates the Nordic Seaspreliminary conductivity offset residual values.

Page 15: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-11-

-0.02

-0.01

0.00

0.01

0.02

0.03

0.04

0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160

Con

duct

ivity

Off

sets

(m

S/cm

)

Station Number CTD #5, stations 1-111 individual-cast conductivity offsets, pressure > 700 db only (all pressures stations 46-67)

order= 0

8.776361e-03

r=0.0000000 p=0.0000000sd=0.0014394 n= 98

Figure 4.8.3.1 Nordic Seas CTD #5 preliminary conductivity offsets by station number, stations 1-111.

Bottle data from over half of the first 15 casts were eliminated from the offset determinations because they were asmuch as 0.002 PSU low compared to nearby casts when the CTD deep theta-S data were consistent. A possiblecause of this was inadequate equilibration of the extremely cold bottle samples to the fairly warm lab and Autosalbath temperatures. The bath temperature was lowered by 3°C starting with station 32 samples, and later samplebottles were better equilibrated in a tub of room-temperature water prior to analysis.

Smoothed offsets were applied to each cast. Some offsets were manually adjusted to account for discontinuousshifts in the conductivity transducer response or bottle salinities, or to maintain deep theta-salinity consistency fromcast to cast. The stations 1-111 preliminary slope and stations 68-111 mean offset were applied through station 159.Cast-by-cast comparisons showed intermittent drifts up to 0.007 mS/cm, probably attributable to organic fouling ofthe sensor. These larger drifts were reversed during long transits between track lines, when sensors soaked indistilled water for longer periods than usual. There was an overall drift of 0.003 mS/cm in the conductivity sensoroffset and no apparent slope changes over the entire leg.

The final shipboard Nordic Seas conductivity slopes are summarized in Figure 4.8.3.2. Figure 4.8.3.3 summarizesthe final shipboard conductivity offsets.

-0.06

-0.05

-0.04

-0.03

-0.02

-0.01

0.00

0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160

Con

duct

ivity

Slo

pe (

mS/

cm)

Station Number CTD #5 final conductivity slopes

Figure 4.8.3.2 Nordic Seas CTD #5 conductivity slope corrections by station number.

Page 16: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-12-

-0.03

-0.02

-0.01

0.00

0.01

0.02

0.03

0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160

Con

duct

ivity

Off

set (

mS/

cm)

Station Number CTD #5 final conductivity offsets

Figure 4.8.3.3 Nordic Seas CTD #5 conductivity offsets by station number.

Summary of Residual Salinity Differences

Figures 4.8.3.4, 4.8.3.5 and 4.8.3.6 summarize the Nordic Seas residual differences between bottle and CTDsalinities after applying the conductivity corrections. Only CTD and bottle salinities with quality code 2(acceptable) were used to generate these figures and statistics. Residual differences exceeding ±0.025 PSU areincluded in the calculations for averages and standard deviations, even though they are not plotted.

-20

-10

0

10

20

0 1000 2000 3000 4000 5000

Salin

ity R

esid

ual (

PSU

x100

0)

Pressure (db) CTD #5, residual salt diffs, after correction, all pressures

order= 0

2.077649e-01

r=0.0000000 p=0.0000000sd=6.4909139 n= 3168 cl= 68.27%

Figure 4.8.3.4 Salinity residual differences vs pressure (after correction).

Page 17: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-13-

-20

-10

0

10

20

0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160

Salin

ity R

esid

ual (

PSU

x100

0)

Station Number CTD #5, residual salt diffs, after correction, all pressures

order= 0

2.077649e-01

r=0.0000000 p=0.0000000sd=6.4909139 n= 3168 cl= 68.27%

Figure 4.8.3.5 Salinity residual differences vs station # (after correction).

-20

-10

0

10

20

0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150 160

Salin

ity R

esid

ual (

PSU

x100

0)

Station Number CTD #5, residual salt diffs, after correction, pressures > 500db

order= 0

1.959242e-01

r=0.0000000 p=0.0000000sd=2.5813378 n= 1443 cl= 68.27%

Figure 4.8.3.6 Deep salinity residual differences vs station # (after correction).

The CTD conductivity calibration represents a best estimate of the conductivity field throughout the water column.3σ from the mean residual in Figures 4.8.3.5 and 4.8.3.6, or ±0.0098 PSU for all salinities and ±0.0022 PSU fordeep salinities, represents the limit of repeatability of the bottle salinities (Autosal, rosette, operators and samplers).This limit agrees with station overlays of deep theta-salinity. Within most casts (a single salinometer run), theprecision of bottle salinities appears to be better than 0.001 PSU. The precision of the CTD salinities appears to bebetter than 0.0005 PSU.

Historical data comparisons and tabulation of temperature and conductivity correction coefficients will be includedin the final Nordic Seas report, after corrections are finalized.

4.8.4. CTD Dissolved Oxygen

A single pumped SBE43 dissolved O2 sensor was used for the entire cruise. This was the first time an SBE43 sensorwas used by ODF with the NBIS MKIII. There were numerous problems with oxygen pump power during the first39 casts. These were resolved by re-wiring one of the 3 conductors in the sea cable to use for power to the altimeterand oxygen pump; there were only two more casts with oxygen pump problems the rest of the cruise. A free-flow/no-pump experiment was conducted during station 108, which resulted in unusable CTD dissolved O2 data.

There were a number of problems with the response characteristics of the typical SensorMedics O2 sensor used withthe NBIS MKIII CTD, the major ones being a secondary thermal response and a sensitivity to profiling velocity.Stopping the rosette for as little as half a minute, or slowing down for a bottom approach, could cause shifts in theCTD O2 profile as oxygen became depleted in water near the sensor. This apparently was still a problem on Nordic

Page 18: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-14-

Seas, despite using the new Sea-Bird SBE43 pumped sensor.

Preliminary oxygen processing will be conducted post-cruise, and further details will be added to this section whenavailable.

4.9. Bottle Sampling

At the end of each rosette deployment water samples were drawn from the bottles in the following order:

• CFCs• 3 He• O2

• SF6

• pH• Total CO2

• Alkalinity• Tritium• 18O• Nutrients• 129 I• Salinity

Note that some properties were subsampled by cast or by station, so the actual sequence of samples drawn wasmodified accordingly. Fiv e stations also had 3-5 bottles tripped just for noble gas sampling.

The correspondence between individual sample containers and the rosette bottle from which the sample was drawnwas recorded on the sample log for the cast. This log also included any comments or anomalous conditions notedabout the rosette and bottles. One member of the sampling team was designated the sample cop, whose soleresponsibility was to maintain this log and insure that sampling progressed in the proper drawing order.

Normal sampling practice included opening the drain valve and then the air vent on the bottle, indicating an air leakif water escaped. This observation together with other diagnostic comments (e.g., "lanyard caught in lid", "valve leftopen") that might later prove useful in determining sample integrity were routinely noted on the sample log.

Once individual samples had been drawn and properly prepared, they were distributed to their respective laboratoriesfor analysis. Oxygen, nutrients and salinity analyses were performed on computer-assisted (PC) analyticalequipment networked to the data processing computer for centralized data analysis. The analysts for each specificproperty were responsible for insuring that their results were updated into the cruise database.

5. Bottle Data Processing

Bottle data processing began with sample drawing, and continued until the data were considered to be final. One ofthe most important pieces of information, the sample log sheet, was filled out during the drawing of the manydifferent samples. It was useful both as a sample inventory and as a guide for the technicians in carrying out theiranalyses. Any problems observed with the rosette before or during the sample drawing were noted on this form,including indications of bottle leaks, out-of-order drawing, etc. Additional clues regarding bottle tripping or leakproblems were found by individual analysts as the samples were analyzed and the resulting data were processed andchecked by those personnel.

The next stage of processing was accomplished after the individual parameter files were merged into a commonstation file, along with CTD-derived parameters (pressure, temperature, conductivity, etc.). The rosette cast andbottle numbers were the primary identification for all ODF-analyzed samples taken from the bottle, and were used tomerge the analytical results with the CTD data associated with the bottle. At this stage, bottle tripping problemswere usually resolved, sometimes resulting in changes to the pressure, temperature and other CTD propertiesassociated with the bottle. All CTD information from each bottle trip (confirmed or not) was retained in a file, soresolving bottle tripping problems consisted of correlating CTD trip data with the rosette bottles.

Diagnostic comments from the sample log, and notes from analysts and/or bottle data processors were entered into acomputer file associated with each station (the "quality" file) as part of the quality control procedure. Sample datafrom bottles suspected of leaking were checked to see if the properties were consistent with the profile for the cast,

Page 19: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-15-

with adjacent stations, and, where applicable, with the CTD data. Various property-property plots and verticalsections were examined for both consistency within a cast and consistency with adjacent stations by data processors,who advised analysts of possible errors or irregularities. The analysts reviewed and sometimes revised their data asadditional calibration or diagnostic results became available.

Based on the outcome of investigations of the various comments in the quality files, WHP water sample codes wereselected to indicate the reliability of the individual parameters affected by the comments. WHP bottle codes wereassigned where evidence showed the entire bottle was affected, as in the case of a leak, or a bottle trip at other thanthe intended depth.

WHP water bottle quality codes were assigned as defined in the WOCE Operations Manual [Joyc94] with thefollowing additional interpretations:

2 No problems noted.3 Leaking. An air leak large enough to produce an observable effect on a sample is

identified by a code of 3 on the bottle and a code of 4 on the oxygen. (Small air leaksmay have no observable effect, or may only affect gas samples.)

4 Did not trip correctly. Bottles tripped at other than the intended depth were assigned acode of 4. There may be no problems with the associated water sample data.

5 Not reported. No water sample data reported. This is a representative level derivedfrom the CTD data for reporting purposes. The sample number should be in the rangeof 80-99.

9 The samples were not drawn from this bottle.

WHP water sample quality flags were assigned using the following criteria:

1 The sample for this measurement was drawn from the water bottle, but the results ofthe analysis were not (yet) received.

2 Acceptable measurement.3 Questionable measurement. The data did not fit the station profile or adjacent station

comparisons (or possibly CTD data comparisons). No notes from the analyst indicateda problem. The data could be acceptable, but are open to interpretation.

4 Bad measurement. The data did not fit the station profile, adjacent stations or CTDdata. There were analytical notes indicating a problem, but data values were reported.Sampling and analytical errors were also coded as 4.

5 Not reported. There should always be a reason associated with a code of 5, usuallythat the sample was lost, contaminated or rendered unusable.

9 The sample for this measurement was not drawn.

WHP water sample quality flags were assigned to the CTDSAL (CTD salinity) parameter as follows:

2 Acceptable measurement.3 Questionable measurement. The data did not fit the bottle data, or there was a CTD

conductivity calibration shift during the up-cast.4 Bad measurement. The CTD up-cast data were determined to be unusable for

calculating a salinity.7 Despiked. The CTD data have been filtered to eliminate a spike or offset.

Page 20: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-16-

Table 5.0 shows the number of samples drawn and the number of times each WHP sample quality flag was assignedfor each basic hydrographic property. Nutrient data are temporarily omitted from this chart until they can beincorporated into the ODF bottle data files. CTD Oxygen statistics will also be included when preliminary fittinghas been completed.

Rosette Samples Stations 001-159Reported WHP Quality Codes

Levels 1 2 3 4 5 7 9

Bottle 3286 0 3236 17 2 0 0 31CTD Salt 3286 0 3145 0 0 0 141 0Salinity 3216 0 3168 44 4 0 0 70Oxygen 3179 0 2825 6 348 3 0 104

Table 5.0 Frequency of WHP quality flag assignments for Nordic Seas.

5.1. Bottle Pressure and Temperature

All pressures and temperatures for the bottle data tabulations on the rosette casts were obtained by averaging CTDdata for a brief interval at the time the bottle was closed on the rosette, then correcting the data based on CTDlaboratory calibrations.

The temperatures are reported using the International Temperature Scale of 1990.

5.2. Salinity Analysis

Equipment and Techniques

A single Guildline Autosal Model 8400A salinometer (s/n 55-654) was used for measuring salinity on all stations.Salinometer 46-992 was the backup salinometer and was not used. The salinometers were modified by ODF andcontained interfaces for computer-aided measurement. The water bath temperature was set and maintained at avalue near the laboratory air temperature. It was set at 24°C for stations 1-31 and stations 109-159, and at 21°C forstations 32-108.

The salinity analyses were performed when samples had equilibrated to laboratory temperature, usually within 8-30hours after collection. Equilibration time was sometimes accelerated by immersing sample bottles in a tub of roomtemperature water because of the extreme differences between in situ sample temperatures and the lab temperature.The salinometer was standardized for each group of analyses (usually one or two casts, up to ∼50 samples) using atleast one fresh vial of standard seawater per group. Some groups included up to 80 samples and 8 casts betweenstandardizations, usually for a series of shallow stations. A computer (PC) prompted the analyst for controlfunctions such as changing sample, flushing, or switching to "read" mode. The salinometer cell was flushed andresults were logged by the computer until two successive measurements met software criteria for consistency. Thesevalues were then averaged for a final result.

Sampling and Data Processing

Salinity samples were drawn into 200 ml Kimax high-alumina borosilicate bottles, which were rinsed three timeswith sample prior to filling. The bottles were sealed with custom-made plastic insert thimbles and Nalgene screwcaps. This assembly provides very low container dissolution and sample evaporation. Prior to collecting eachsample, inserts were inspected for proper fit and loose inserts were replaced to insure an airtight seal. The draw timeand equilibration time were logged for all casts. Laboratory temperatures were logged at the beginning and end ofeach run.

PSS-78 salinity [UNES81] was calculated for each sample from the measured conductivity ratios. The difference (ifany) between the initial vial of standard water and one run at the end as an unknown was applied linearly to the datato account for any drift. The data were added to the cruise database. 3216 salinity measurements were made and170 vials of standard water were used. The estimated accuracy of bottle salinities run at sea is usually better than0.002 PSU relative to the particular standard seawater batch used.

Page 21: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-17-

Laboratory Temperature

The temperature in the salinometer laboratory varied from 19 to 25°C during the cruise. The air temperature changeduring a run of samples ranged from less than 0.5°C to more than 4°C. The laboratory temperature was 4°C lowerto 3°C higher than the Autosal bath temperature.

Standards

IAPSO Standard Seawater (SSW) Batch P-140 was used to standardize the salinometer.

5.3. Oxygen Analysis

Equipment and Techniques

Dissolved oxygen analyses were performed with an ODF-designed automated oxygen titrator using photometricend-point detection based on the absorption of 365nm wav elength ultra-violet light. The titration of the samples andthe data logging were controlled by PC software. Thiosulfate was dispensed by a Dosimat 665 buret driver fittedwith a 1.0 ml buret. ODF used a whole-bottle modified-Winkler titration following the technique of Carpenter[Carp65] with modifications by Culberson et al. [Culb91], but with higher concentrations of potassium iodatestandard (∼0.012N) and thiosulfate solution (∼65 gm/l). Standard solutions prepared from pre-weighed potassiumiodate crystals were run at the beginning of each session of analyses, which typically included from 1 to 3 stations.Pre-made ODF liquid potassium iodate standards were also used. Several standards were made up during the cruiseand compared to assure that the results were reproducible, and to preclude the possibility of a weighing or dilutionerror. Reagent/distilled water blanks were determined, to account for presence of oxidizing or reducing materials.The auto-titrator generally performed very well.

Sampling and Data Processing

Samples were collected for dissolved oxygen analyses soon after the rosette sampler was brought on board. Using aTygon drawing tube, nominal 125ml volume-calibrated iodine flasks were rinsed 2-3 times with minimal agitation,then filled and allowed to overflow for at least 3 flask volumes. Reagents were added to fix the oxygen beforestoppering. The flasks were shaken twice to assure thorough dispersion of the precipitate, once immediately afterdrawing, and then again after about 20 minutes.

Drawing oxygen samples usually included taking the sample draw temperature from a small platinum-resistancethermometer embedded in the drawing tube. However, the thermometers normally used were in a shipment thatnever arrived. Sampling the temperatures with substitute equipment was deemed too awkward, and the practice wasabandoned after 9 stations.

The samples were analyzed within 1-15 hours of collection, and then the data were merged into the cruise database.

Thiosulfate normalities were calculated from each standardization and corrected to 20°C. The 20°C normalities andthe blanks were plotted versus time and were reviewed for possible problems. New thiosulfate normalities will berecalculated after the blanks have been smoothed as a function of time, if warranted. These new normalities willthen be smoothed, and the oxygen data recalculated.

As samples warmed up to room temperature they would often degas which would cause an occasional noisyendpoint due to gas bubbles in the light path. 3179 oxygen measurements were made. In addition, 14 replicatesamples were analyzed to compare reagent batches, and 57 oxygen samples (3 per rosette bottle) were analyzedfrom noble gas sampling.

Volumetric Calibration

Oxygen flask volumes were determined gravimetrically with degassed deionized water to determine flask volumes atODF’s chemistry laboratory. This is done once before using flasks for the first time and periodically thereafter whena suspect bottle volume is detected. The volumetric flasks used in preparing standards were volume-calibrated bythe same method, as was the 10 ml Dosimat buret used to dispense standard iodate solution.

Page 22: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-18-

Standards

Potassium iodate standards, nominally 0.44 gram, were pre-weighed in ODF’s chemistry laboratory to ±0.0001grams. The exact normality was calculated at sea after the volumetric flask volume and dilution temperature wereknown. Liquid potassium iodate standards were also used on the cruise. The normality of the liquid standard wasdetermined at ODF by calculation from weight and by comparison to other iodate standards. Potassium iodate wasobtained from GFS Chemical Co. and was reported by the supplier to be >99.4% pure. All other reagents were"reagent grade" and were tested for levels of oxidizing and reducing impurities prior to use.

One lot number of NaI reagent gav e a high negative blank which was not measurable with normal methods. It wasnecessary to use this "bad" reagent for stations 40 to 67 since this was the only NaI remaining on board. Afterstation 67, the ship ran in north of Tromso/ , Norway to get shipments of good reagents from Swedish and ODFsources, delivered by a pilot boat. Comparisons were done between "good" and "bad" reagents to calculate acorrection to apply to stations run with the "bad" reagents. No other reagent problems occured during the remainderof the cruise.

5.4. Nutrient Analysis

Equipment and Techniques

Nutrient analyses (phosphate, silicate, nitrate and nitrite) were performed on an ODF-modified 4-channel TechniconAutoAnalyzer II, generally within one hour after sample collection. Occasionally samples were refrigerated up to 4hours at ∼4°C. All samples were brought to room temperature prior to analysis.

The methods used are described by Gordon et al. [Gord92]. The analog outputs from each of the four colorimeterchannels were digitized and logged automatically by computer (PC) at 2-second intervals.

Silicate was analyzed using the technique of Armstrong et al. [Arms67]. An acidic solution of ammoniummolybdate was added to a seawater sample to produce silicomolybdic acid which was then reduced tosilicomolybdous acid (a blue compound) following the addition of stannous chloride. Tartaric acid was also addedto impede PO4 color development. The sample was passed through a 15mm flowcell and the absorbance measuredat 660nm.

A modification of the Armstrong et al. [Arms67] procedure was used for the analysis of nitrate and nitrite. For thenitrate analysis, the seawater sample was passed through a cadmium reduction column where nitrate wasquantitatively reduced to nitrite. Sulfanilamide was introduced to the sample stream followed byN-(1-naphthyl)ethylenediamine dihydrochloride which coupled to form a red azo dye. The stream was then passedthrough a 15mm flowcell and the absorbance measured at 540nm. The same technique was employed for nitriteanalysis, except the cadmium column was bypassed, and a 50mm flowcell was used for measurement.

Phosphate was analyzed using a modification of the Bernhardt and Wilhelms [Bern67] technique. An acidicsolution of ammonium molybdate was added to the sample to produce phosphomolybdic acid, then reduced tophosphomolybdous acid (a blue compound) following the addition of dihydrazine sulfate. The reaction product washeated to ∼55°C to enhance color development, then passed through a 50mm flowcell and the absorbance measuredat 820nm.

Sampling and Data Processing

Nutrient samples were drawn into 45 ml polypropylene, screw-capped "oak-ridge type" centrifuge tubes. The tubeswere cleaned with 10% HCl and rinsed with sample 2-3 times before filling. Standardizations were performed at thebeginning and end of each group of analyses (typically one cast, up to 36 samples) with an intermediateconcentration mixed nutrient standard prepared prior to each run from a secondary standard in a low-nutrientseawater matrix. The secondary standards were prepared aboard ship by dilution from primary standard solutions.Dry standards were pre-weighed at the laboratory at ODF, and transported to the vessel for dilution to the primarystandard. Sets of 6-7 different standard concentrations were analyzed periodically to determine any deviation fromlinearity as a function of concentration for each nutrient analysis. A correction for non-linearity was applied to thefinal nutrient concentrations when necessary. In addition, a "deep seawater" high nutrient concentration checksample was run with each station as an additional check on data quality.

After each group of samples was analyzed, the raw data file was processed to produce another file of responsefactors, baseline values, and absorbances. Computer-produced absorbance readings were checked for accuracy

Page 23: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

-19-

against values taken from a strip chart recording. The data were then added to the cruise database.

Nutrients, reported in micromoles per kilogram, were converted from micromoles per liter by dividing by sampledensity calculated at 1 atm pressure (0 db), in situ salinity, and an assumed laboratory temperature of 25°C.

3149 nutrient samples were analyzed. The pump tubing was changed 3 times.

Standards

Na2SiF6, the silicate primary standard, was obtained from Aesar Chemical Company and was reported by thesuppliers to be >98% pure. Primary standards for nitrate (KNO3), nitrite (NaNO2), and phosphate (KH2PO4) wereobtained from Johnson Matthey Chemical Co.; the supplier reported purities of 99.999%, 97%, and 99.999%,respectively. The efficiency of the cadmium column used for nitrate was monitored throughout the cruise andranged from 99-100%.

No major problems were encountered with the measurements. The temperature of the laboratory used for theanalyses ranged from 21°C to 28°C, but was relatively constant during any one station (±1.5°C).

References

Arms67.Armstrong, F. A. J., Stearns, C. R., and Strickland, J. D. H., “The measurement of upwelling and subsequentbiological processes by means of the Technicon Autoanalyzer and associated equipment,” Deep-Sea Research,14, pp. 381-389 (1967).

Bern67.Bernhardt, H. and Wilhelms, A., “The continuous determination of low lev el iron, soluble phosphate and totalphosphate with the AutoAnalyzer,” Technicon Symposia, I, pp. 385-389 (1967).

Carp65.Carpenter, J. H., “The Chesapeake Bay Institute technique for the Winkler dissolved oxygen method,”Limnology and Oceanography, 10, pp. 141-143 (1965).

Culb91.Culberson, C. H., Knapp, G., Stalcup, M., Williams, R. T., and Zemlyak, F., “A comparison of methods for thedetermination of dissolved oxygen in seawater,” Report WHPO 91-2, WOCE Hydrographic ProgrammeOffice (Aug 1991).

Gord92.Gordon, L. I., Jennings, J. C., Jr., Ross, A. A., and Krest, J. M., “A suggested Protocol for Continuous FlowAutomated Analysis of Seawater Nutrients in the WOCE Hydrographic Program and the Joint Global OceanFluxes Study,” Grp. Tech Rpt 92-1, OSU College of Oceanography Descr. Chem Oc. (1992).

Joyc94.Joyce, T., ed. and Corry, C., ed., “Requirements for WOCE Hydrographic Programme Data Reporting,”Report WHPO 90-1, WOCE Report No. 67/91, pp. 52-55, WOCE Hydrographic Programme Office, WoodsHole, MA, USA (May 1994, Rev. 2). UNPUBLISHED MANUSCRIPT.

UNES81.UNESCO, “Background papers and supporting data on the Practical Salinity Scale, 1978,” UNESCOTechnical Papers in Marine Science, No. 37, p. 144 (1981).

Page 24: CRUISE REPORT: NS02 - University of California, San Diego · Smith, Helen B. Univ.of East Anglia,UK SF6 Swift, James H. SIO/STS/ODF co-Chief Scientist *BCCR Bjerknes Centre for Climate

CCHDO Data Processing Notes

Date Person Data Type Action Summary 2010-04-19 Dave Muus BTL Website Update Data Onlline Detailed Notes

Notes on NS02 bottle data. April 19, 2010 D. Muus EXPOCODE 316N20020530

1. Merged CARINA exchange file "316N20020530_hy1.csv"(20081218PRINUNIVRMK) TCARBN, ALKALI and SF6 into

ODF exchange file "ns02_hy1.csv" (20050729WHPSIOdm). CARINA exchange file taken from CARINA project section of CCHDO website.

Originally named "316N20020530.exc.csv". ODF exchange file was prepared for Jim Swift, July, 2005, from ODF cruise file.

2. Changed SF6 data format from 8 decimal places to 4 decimal places.

3. Modified Summary file ns02su.txt (20040311WJPOSIODM): Expocode "316N166.11" changed to "316N20020530". SECT_ID "NSeas" changed to "NS02".

4. Made new WOCE format bottle file from new Exchange file (20100406CCHDOSIODM)

2013-02-01 Jerry Kappa CrsRpt Website Update Final PDF version online I've placed 1 new version of the cruise report:

ns02_316N20020530do.pdf

into the directory: co2clivar/arctic/ns02/

It includes summary pages and CCHDO data processing notes as well as a linked Table of Contents and links to figures, tables and appendices.

It will be available on the cchdo website following the next update script run.


Recommended