C z e c h M e t r o l o g y I n s t i t u t e
Okružní 31, 638 00
Brno
1. ------IND- 2014 0039 CZ- EN- ------ 20140203 --- --- PROJET
Executive summary for the EC (not part of this legislation)
Breath alcohol analysers are subject national metrological regulations and are placed on the market and
made operational in the Czech Republic following type approval and initial verification in accordance
with Act No 505/1990 on metrology, as amended. The subject of this notified regulation is to lay down
metrological and technical requirements for specified measuring instruments, including the testing
methods for the type approval and verification of specified measuring instruments.
This regulation sets forth the metrological level required for the certificates of type approval and
verification of measuring instruments issued abroad to be recognised.
The requirements to which these measuring instruments are subject are also based on European
standard EN 15964:2011 and take account of the draft recommendation OIML R 126:2012.
(End of executive summary.)
As the authority with substantive and territorial jurisdiction in the matter of laying down the
metrological and technical requirements for specified measuring instruments and laying down the testing
methods for type approval and verification of specified measuring instruments under § 14(1) of Act No
505/1990 on metrology, as amended (hereinafter referred to as the “Metrology Act”), and in accordance
with the provisions of § 172 et seq. of Act No 500/2004, the Administrative Procedure Code
(hereinafter referred to as the “APC”), the Czech Metrology Institute (hereinafter referred to as the
“CMI”) commenced ex officio proceedings on 22 November 2013 under § 46 of the APC and, based on
the supporting documents, issues the following:
I.
G E N E R A L M E A S U R E
number: 0111-OOP-C040-13
ref. no. 0313/011/13-Po,
laying down the metrological and technical requirements for specified measuring instruments,
including the testing methods for type approval and verification of the following specified
instruments:
“breath alcohol analysers”
General Measure No 0111-OOP-C040-13
2
1 Basic definitions
For the purposes of this General Measure, the terms and definitions according to VIM and VIML1) and
the definitions below shall apply.
1.1
breath alcohol (concentration) analyser
a measuring instrument designed to determine ethanol concentration in the air exhaled by the tested
person; the detection is based on an electrochemical method of measurement
1.2
portable breath alcohol analyser
a moveable measuring instrument intended for use outside or inside buildings, e.g. a handheld device
powered by an autonomous battery, which may also be equipped with a separate mobile printer
1.3
measurement of alcohol in breath
determination of the mass concentration of ethanol in exhaled air generated in the pulmonary alveoli of
the tested person, in mg/L
Under existing legislation in the Czech Republic, the results of this measurement method are converted
by the analyser from breath alcohol concentration to blood alcohol concentration and expressed in units
of % (per mille).
NOTE Mass concentration of alcohol in exhaled breath is the proportion by mass of ethanol in a volume of
exhaled air at the temperature of 34°C and pressure of 1013 hPa. Mass concentration of alcohol in the blood is
the proportion by mass of ethanol relative to the volume or mass of blood at the temperature of 20°C and
pressure of 1013 hPa.
According to Henry’s law, breath alcohol concentration (BrAC) depends on blood alcohol concentration
(BAC). In the Czech Republic, breath alcohol analysers must be adjusted to convert the value using the
BAC/BrAC ratio of 2100:1. This ratio is statistically the most commonly used in European countries for
determining the mass concentration of alcohol directly in units of BAC. This means that in the Czech Republic
analysers must be adjusted to use the % (per mille) / mg/L conversion ratio of 2.1:1.
1.4
alveolar air
the air exhaled by the tested person in a strong exhalation involving contractions of abdominal and
internal intercostal muscles; alveolar air is generally obtained in the last third of the period of exhalation
1.5
measurement of alcohol in the blood
determination of mass concentration of ethanol in the blood in g/kg or in % (per mille)
NOTE Devices for measuring breath alcohol concentration can express the mass concentration of ethanol
determined in mg/L or convert this value using the analyser into mass concentration of alcohol in the blood in
% (per mille), for which the conversion constant of 2.1 is used in the Czech Republic (see measuring device
adjustment).
1) International vocabulary of metrology – Basic and general concepts and associated terms (VIM) and
International Vocabulary of Terms in Legal Metrology (VIML) are part of the technical harmonisation
compendium “Terminology in the Area of Metrology”, which is publicly accessible at www.unmz.cz
General Measure No 0111-OOP-C040-13
3
1.6
active mode of the breath alcohol analyser
in this mode, the breath alcohol analyser provides results of measurements of breath alcohol
concentration in the form of a specific numerical value in the desired units
1.7
drift
a change in the measuring instrument indications which occurs during a stated period of time at a given
mass concentration of ethanol in air
1.8
memory effect
dependence of the value indicated by the measuring instrument on the value indicated by the same
measuring instrument for the preceding sample
2 Metrological requirements
The metrological requirements are based on the requirements under Recommendation OIML R 1262)
“Evidential breath analysers”. Measuring instruments type approved prior to the entry into force of this
regulation shall be subject to the metrological requirements applicable at the time they were put into
circulation.
NOTE If, for the sake of measurement accuracy, the number of measurements (analyses) or the period of time
after which the breath alcohol analyser needs to be adjusted to reference values are limited, the manufacturer
must specify this time limit in the documentation for the measuring instrument and/or ensure that the
measuring instrument directly indicates that this limit has been reached.
2.1. Operating conditions
2.1.1 Operating conditions for use of the measuring instrument
The operating conditions for the breath alcohol analyser shall be specified by the manufacturer for the
particular type of measuring instrument. If not specified, the requirements below shall apply.
The ambient operating temperature range must be at least –10°C to +40°C.
The relative humidity limits must be at least 10% to 85%.
Atmospheric pressure must be within the range of 860 hPa to 1 060 hPa.
2.1.2 Reference operating conditions
The reference ambient temperature during a laboratory test must be within the range of 19°C to 27°C.
The relative humidity limits must be at least 20% to 80%.
Atmospheric pressure must be within the range of 860 hPa to 1 060 hPa.
Carbon dioxide mass concentration must be less than 10%.
The flow rate of the test gas (CRM for ethanol in nitrogen) must be within the range of 0.20 L/s to
0.35 L/s.
2) OIML R 126 “Evidential breath analyzers”, freely accessible at www.oiml.org
General Measure No 0111-OOP-C040-13
4
Back pressure must not exceed 25 hPa (at the flow rate of 0.25 L/s).
The volume of the test gas analysed in the breath alcohol analyser must be at least 1.2 L.
The time of exhalation into the breath alcohol analyser must be at least 4 seconds.
2.1.3 Temperature stabilisation period
Under reference conditions, the measuring instrument should be capable of accurate measurement:
– following temperature stabilisation as specified by the manufacturer - no more than 15
minutes after the measuring instrument has been switched on,
– in less than 5 minutes after it has been switched from the sleep (standby) mode to the
measuring mode.
If these requirements are not met, the relevant times must be clearly indicated on the measuring
instrument and stated in the manufacturer’s documentation.
2.2 Measuring range
The measuring range of the breath alcohol analyser must be from 0.00 mg/L to no less than 2.00 mg/L.
The upper limit of the measuring range shall be specified by the manufacturer and must not exceed
3.00 mg/L.
A breath alcohol analyser must be capable of indicating (including as a numerical value) that the upper
limit of the measuring range has been exceeded.
A breath alcohol analyser must be capable of displaying the results of measurement:
a) in mg/L (mg of ethanol per L of exhaled air);
b) in % (per mille corresponding to g/kg of exhaled air).
2.3 Maximum permissible error
Maximum permissible errors (positive or negative) for each measurement are given in Table 1.
Table 1 – Maximum permissible errors
Mass concentration of ethanol in exhaled air
(mg/L)
Maximum permissible error
(mg/L)
< 0.4 0.020
≥ 0.4 and ≤ 2.0 5%*)
> 2.0 (reference value/2) – 0.90
*) Values in percent are relative to the measured value of mass concentration of ethanol.
2.4 Repeatability of measurement
The repeatability of measurement is expressed as the standard deviation of a given number of
measurements, which must be less than the values given for each measuring range in Table 2.
TABLE 2 – Repeatability of measurement
Mass concentration of ethanol in exhaled air
(mg/L)
Maximum permissible standard deviation
(mg/L)
General Measure No 0111-OOP-C040-13
5
< 0.4 0.010
≥ 0.4 and ≤ 2.0 2.5%*)
> 2.0 [(reference value/2) – 0.90] / 3
*) Values in percent are relative to the measured value of mass concentration of ethanol.
2.5 Drift
Zero drift and short-term drift at 0.40 mg/L must be less than 0.007 mg/L in 4 hours.
Long-term drift at 0.40 mg/L must be less than 0.014 mg/L in two months.
2.6 Memory effect
The memory effect must be less than 0.010 mg/L.
2.7 Small changes in mass concentration of gas, residual effect
When alternately measuring two test gases, the measurement results for the test gas (CRM) with lower
mass concentration must not differ by more than 0.010 mg/L.
3 Technical requirements
The technical requirements are based on the requirements of Recommendation OIML R 1262)
“Evidential breath analysers”. Measuring instruments type approved prior to the entry into force of this
regulation shall be subject to the technical requirements applicable at the time they were put into
circulation.
3.1 General
A portable breath alcohol analyser shall be designed for use outside or inside buildings and allow
connection of a mobile printer.
If the breath alcohol analyser converts the breath alcohol concentration value into per mille, the
indicated value must represent a 2.1-multiple of the value expressed in mg/L. (This value is accepted in
European countries as the conversion factor for official purposes; the conversion error has been
determined by experimental measurements and does not exceed 0.20 % when converted into blood
alcohol concentration).
The exhaled volume must be at least 1.2 L and this value must correspond to an exhalation time of at
least 4 seconds. The effect of the volume of the passing air shall be tested in accordance with Article
5.3.5.
The breath alcohol analyser shall perform a measurement only if the sample taken is a sample of
alveolar air. The analyser shall in particular inhibit a measurement if the exhalation is discontinuous, or
if the exhaled air contains breath from the upper respiratory tract. In the case of discontinuous
exhalation or exhalation from the upper respiratory tract, the measurement result must not be indicated
by the breath alcohol analyser as a numerical value.
The breath alcohol analyser shall inhibit a measurement if air is inhaled instead of exhaled. The
measuring instrument must inhibit detection when air is “sucked” instead of “blown” - in such cases, the
measurement result must not be indicated by the breath alcohol analyser as a numerical value.
Before each measurement operation, the measuring instrument shall make an automatic adjustment and
verify automatically that it is capable of making a correct measurement. If this verification reveals that
General Measure No 0111-OOP-C040-13
6
not all the conditions for correct operation of the breath alcohol analyser have been fulfilled, any
measurement made by the breath alcohol analyser must be automatically inhibited.
The breath alcohol analyser must be used only with a mouthpiece specified by the manufacturer and in
compliance with the mouthpieces indicated for the particular approved type of measuring instrument.
3.2 Indicating device
The scale interval for normal operation of the measuring instrument shall be to the order of 0.01 mg/L.
The breath alcohol analyser may have resolution to the order of 0.001 mg/L only when state
metrological controls on the measuring instrument are being carried out.
If the breath alcohol analyser is capable of resolution to the order of 0.001 mg/L, then in normal
operation the measuring instrument must round the values obtained down to two decimal places, i.e. to
0.01 mg/L, e.g. 0.427 mg/L shall be rounded and displayed as 0.42 mg/L.
The name or symbol of the unit of measurement used must be shown on the display in close proximity
to the numerical result of the measurement.
3.3 Printer
The measurement results of the printer output must be identical to those shown on the display of the
measuring instrument, including the symbol of the measurement unit used. The printer must constitute a
separate part of the measuring instrument. A mobile printer must be connected to the measuring
instrument by a separate cable or wireless connection.
3.4 Software
Software critical for metrological characteristics must be identifiable by the manufacturer as a separate,
numbered version, which conforms to the approved measuring instrument type. Software identification
must be possible while the measuring instrument is being operated. The installed software must be
secured by the manufacturer against accidental or unauthorised outside interference (e.g. by means of a
service password). If the software needs to be reinstalled (as a maintenance intervention in the
measuring instrument), new verification of the measuring instrument’s metrological characteristics must
be conducted.
3.5 Physiological factors influencing measurement
If components of pharmaceuticals or products of abnormal metabolism contained in solvents or industrial
products, such as toluene vapours, acetaldehyde and other gases, are present in exhaled breath, these
substances may influence the measurement result. The extent of such influence must be negligible, especially
when assessed in relation to the limit below which the result of breath alcohol content test is considered
negative (i.e. it must not exceed 0.20% when converted into blood alcohol concentration). Breath alcohol
analysers must be based on a highly selective electrochemical principle of measurement. Identification of the
influence of interfering components forms part of the test of the measuring instrument under Article 5.3.6.
3.6 Resistance to external influences
External disturbances affecting the breath alcohol analyser referred to in Chapters 3.6.1 and 3.6.2 must
not lead to measurement errors which would exceed the maximum permissible error of the breath
alcohol analyser under Article 2.3.
If significant errors or significant faults occur, they must be detected and reported by the checking
facility contained in the breath alcohol analyser.
General Measure No 0111-OOP-C040-13
7
In such cases, the breath alcohol analyser must not indicate the measurement result as a numerical
value.
3.6.1 Resistance to mechanical influences
The design of the breath alcohol analyser and the materials used must provide sufficient strength,
stability and resistance to mechanical vibrations and shocks.
3.6.2 Resistance to climatic influences
In the off-mode, breath alcohol analysers must withstand, without damage, the limit temperatures of –
20°C and 70°C and after returning to the operating temperature range, they must operate within the
limits of the maximum permissible error.
Breath alcohol analysers must not be sensitive to relative ambient humidity, whether in operation or
while in storage.
3.6.3 Electromagnetic compatibility (EMC)
Breath alcohol analysers must not be affected by electrical or electromagnetic interference, or must
respond to it in certain specific manners, e.g. by reporting an error, inhibition of measurement etc. They
must not emit unwanted electromagnetic fields.
When tested for electromagnetic compatibility in a laboratory, the breath alcohol analyser must be
functioning normally. It is allowed to restart the measuring instrument as a response to interference.
3.7 Supply voltage
Breath alcohol analysers powered by direct current from batteries must be operated without difficulty,
at least within the range specified by the manufacturer (Umin to Umax). When outside of this specified
supply voltage range, the measuring instrument must switch itself off or switch to a mode in which
measurement is not possible.
3.8 Resistance to unauthorised tampering
The breath alcohol analyser must be designed such that any mechanical actions performed on this
measuring instrument that may influence measurement accuracy result in permanent visible damage to
the measuring instrument, or the imprint of official marks.
4 Markings
All of the information below must be indicated on the breath alcohol analyser in a manner ensuring that
it is indelible, non-removable and legible throughout the period of use.
4.1 Labelling of breath alcohol analysers
Each breath alcohol analyser shall indicate, at least, the information below:
– the manufacturer's insignia or name,
– type approval mark of the measuring instrument,
– serial number and year of manufacture,
– measuring range in mg/L,
– ambient operating temperature range in °C.
General Measure No 0111-OOP-C040-13
8
5 Type approval
5.1 General
5.1.1 Tests to be performed
The type approval of a breath alcohol analyser shall comprise the following:
a) external inspection,
b) functional tests of the breath alcohol analyser:
– accuracy and repeatability test,
– drift test, short- and long-term stability tests,
– memory effect test,
– residual effect test,
– test of volume influence (flow rate change),
– test of influence of interfering components and CO2,
c) tests of resistance to mechanical influence:
– test of impact resistance,
– free-fall test,
– test of influence of random mechanical vibrations,
d) tests of resistance to climatic conditions:
– cold test (in both off and on mode),
– dry heat test,
– damp heat test (in both off and on mode),
e) test of supply voltage influence,
f) tests for electromagnetic compatibility (EMC):
– electrostatic discharge immunity test,
– test of resistance to radiated high-frequency electromagnetic fields,
– test of resistance to TETRA signals.
5.1.2 Test equipment
The measuring instruments, certified reference materials and equipment specified below shall be used
for testing breath alcohol analysers:
– certified reference materials (CRM),
Gas mixtures of the required composition (see Table 3) shall be used; these shall be primary
reference materials for the relevant gas mixture mole (mass) fractions or secondary CRM for
ethanol in nitrogen, traceable to a suitable primary reference gas mixture. If necessary, prior to
type approval, the device shall be adjusted to the mass concentration recommended by the
manufacturer of the measuring instrument (e.g. 0.48 mg/L of ethanol in nitrogen, which
corresponds to around 1.008 g/kg or mass % of ethanol in the blood). The mole fractions of the
certified reference gases shall be converted to mass concentrations at 34°C at atmospheric
pressure.
General Measure No 0111-OOP-C040-13
9
Table 3 - Certified reference materials
CRM no Mass concentration of ethanol in nitrogen (mg/L)
1 0.04
2 0.10
3 0.25
4 0.40
5 0.70
6 0.95
7 1.50
8 1.95
9 90% of the maximum range in mg/L
Expanded uncertainty values of the CRM for ethanol in nitrogen must not exceed 2% (with a
coverage factor of k = 2). The manufacturing tolerance for the mass concentration of ethanol in
the reference materials shall be 15%.
– flow meter with a measuring range of at least 0 L/s to 0.40 L/s,
– stopwatch with an accuracy of 0.1 s,
– thermo-hygro-barometer for monitoring the conditions in the laboratory,
– reducing valves with adjustable gas pressure on the output,
– test equipment for connecting and measuring the breath alcohol analyser.
5.2 External inspection
The purpose of external inspection of the breath alcohol analyser shall be to assess:
a) the completeness of the required technical documentation, including operating instructions,
b) the conformity of the metrological and technical characteristics specified by the manufacturer in
the documentation with the requirements under this regulation as referred to in Chapters 2, 3
and 4,
c) the completeness and condition of the breath alcohol analyser according to the prescribed
technical documentation,
d) conformity of the software version of the breath alcohol analyser with the version specified by
the manufacturer.
5.3 Functional tests of the breath alcohol analyser
Prior to the test, the breath alcohol analyser shall be prepared in accordance with the requirements of
Article 2.1.3. The functional tests shall be performed under the reference conditions specified in Article
2.1.2.
Gas mixtures of the specified composition shall be fed into the breath alcohol analyser from cylinders
with CRMs for gas mixtures, as per Table 3. Measurements shall be made sequentially from gas
mixtures with lower mass concentrations to mixtures with higher mass concentrations. At the beginning
and end of measurement, i.e. after the measurement for the highest mass concentration (90% of the max.
General Measure No 0111-OOP-C040-13
10
range in mg/L of ethanol in nitrogen) has been made, a “zero” check shall be performed using “zero”
gas, which is nitrogen in a cylinder with a purity of at least 4.0 (99.99%).
The actual operating conditions in the laboratory (temperature, humidity and pressure) shall be recorded
by the sensors of the measuring instrument being tested.
Evacuation of gases from the device connected to the cylinders with the CRMs for the gas mixtures (see
Table 3) and the “zero” gas needs to be ensured during operations.
5.3.1 Accuracy and repeatability test
This test shall involve 20 measurements shall be made for each of the nine mass concentrations within
the measuring range (nominal values of 0.04 mg/L, 0.10 mg/L, 0.25 mg/L, 0.40 mg/L, 0.70 mg/L, 0.95
mg/L, 1.50 mg/L, 1.95 mg/L, and 90% of the upper limit of the measuring range specified for the given
type of measuring instrument).
When measurements have been completed, the average mass concentration value shall be calculated
according to the relationship:
n
β
β
n
i
i
1
where
ßi are the individual values of mass concentration of ethanol measured by the breath alcohol
analyser when measuring gaseous CRM,
n is the number of measured values, i.e. 20.
The deviation of the measured value from the certified value of ethanol concentration in the gaseous
CRM must be less than or equal to the maximum permissible error according to Table 1 under Article
2.3, while the evaluation shall take into account the uncertainty of the measured value and uncertainty
of the certified mass concentration of ethanol in the gaseous CRM measured:
MPE))(2 2
RM
2
CRM u(ββuββ
where
is the arithmetic mean of the mass concentrations of ethanol obtained when measuring the gaseous
CRM,
)β(uc is the combined standard uncertainty,
ßRM is the mass concentration of ethanol in the gaseous CRM measured,
u(ßRM) is the standard uncertainty (derived from the expanded uncertainty with the expansion
coefficient k = 2),
MPE is the maximum permissible error of the breath alcohol analyser for the given measurement
conditions and level of mass concentration of ethanol in the CRM (see Table 1).
In order to assess repeatability, standard deviation SD shall be calculated, which must not be greater
than the maximum permissible standard deviation according to Table 2 under Article 2.4 for the given
mass concentration.
The following relationship holds for the calculation of standard deviation SD:
1
)(1
2
nSD
n
i
i
where
General Measure No 0111-OOP-C040-13
11
is the arithmetic mean of the mass concentrations of ethanol obtained when measuring the gaseous
CRM,
ßi is the value obtained.
Table 4 may be used for recording and evaluation of the test.
Table 4 – Accuracy and repeatability test
Reference mass
concentration
( RM)
0.04
mg/L
0.10
mg/L
0.25
mg/L
0.40
mg/L
0.70
mg/L
0.95
mg/L
1.50
mg/L
1.95
mg/L
90%
of the max.
range in
mg/L
Accuracy
Average mass
concentration
( )
Error detected -
difference
RMββ
U (k = 2)
Maximum
permissible error
(see Table 1)
< 0.020
mg/L
< 0.020
mg/L
< 0.020
mg/L 5% 5% 5% 5% 5%
(reference
value/2) –
– 0.90 mg/L
Repeatability
Standard
deviation (SD)
Maximum
permissible
standard
deviation (see
Table 2)
< 0.010
mg/L
< 0.010
mg/L
< 0.010
mg/L <2.5% <2.5% <2.5% <2.5% <2.5%
< [(reference
value/2) –
0.90] / 3
5.3.2 Drift test and short-term and long-term stability tests
The device stability test shall comprise three parts – drift, short-term stability and long-term stability.
Drift and short-term stability tests shall involve 10 measurements of test gas repeated after four hours.
Measurements to determine drift and short-term stability shall be made using zero gas (with no ethanol
content) and test gas with mass concentration of 0.40 mg/L of ethanol in nitrogen (see Table 3, CRM no
4) according to the following procedure:
– 10 measurement using zero gas - 4-hour break - 10 measurements using zero gas,
– 10 measurement using test gas - 4-hour break - 10 measurements using test gas.
The drift obtained from both measurements must comply with the requirements under Article 2.5.
Long-term stability shall be monitored in the same way as short-term stability, but the subsequent
measurement of test gas shall be made after two months; the maximum permissible drift of 0.014 mg/L,
as required under Article 2.5, must be complied with. Table 5 may be used for recording and evaluation
of this test.
General Measure No 0111-OOP-C040-13
12
Table 5 – Drift and short- and long-term stability
Measurement no Zero gas Test gas 0.40 mg/L Test gas 0.40 mg/L
time of
measureme
nt
time of
measureme
nt + 4 hrs
time of
measureme
nt
time of
measureme
nt + 4 hrs
date of
measurement
date of
measurement
+ 2 months
1
…
10
Mean value
Drift
Max. permissible
drift 0.007 mg/L 0.007 mg/L 0.014 mg/L
5.3.3 Memory effect test
The memory effect test shall be performed using the following procedure:
Step 1: 10 measurements shall be made using gas with mass concentration of 0.10 mg/L of
ethanol in nitrogen (see Table 3, CRM no 2),
Step 2: alternating measurements using gas with mass concentration of 1.95 mg/L of ethanol in
nitrogen (see Table 3, CRM no 8) or, 1.50 mg/L of ethanol in nitrogen (see Table 3,
CRM no 7) for measuring instruments with a measuring range of up to 2 mg/L, and
measurements using gas with mass concentration of 0.10 mg/L of ethanol in nitrogen (see
Table 3, CRM no 2).
Step 2 shall be repeated ten times, thereby obtaining 10 values measured at high mass concentrations
and 10 values measured at low mass concentrations. The memory effect must comply with the
requirements under Article 2.6.
Each individual measurement must comply with the maximum permissible error for the given mass
concentration as referred to in Article 2.3. Table 6 may be used for recording and evaluation of this test.
Table 6 – Memory effect
Measurement no Pre-test - 0.10 mg/L
(Step 1)
Alternating measurements of two concentrations (Step
2)
1.95 mg/L or
(1.50 mg/L)
0.10 mg/L
1
…
10
Mean value 1 - - - 2
Difference ( 1 – 2)
General Measure No 0111-OOP-C040-13
13
5.3.4 Residual effect test
The residual effect test shall be performed using the following procedure:
Step 1: 10 measurements shall be made using gas with mass concentration of 0.25 mg/L of
ethanol in nitrogen (see Table 3, CRM no 3),
Step 2: alternating measurements using gas with mass concentration of 0.40 mg/L of ethanol in
nitrogen (see Table 3, CRM no 4) and measurements using gas with mass concentration
of 0.25 mg/L of ethanol in nitrogen (see Table 3, CRM no 3).
Step 2 shall be repeated ten times, thereby obtaining 10 values measured at the higher mass
concentration and 10 values measured at the lower mass concentration.
The residual effect must comply with the requirements under Article 2.7. Each individual
measurementmust comply with the maximum permissible error for the given mass concentration as
referred to in Article 2.3.
Table 7 may be used for recording and evaluation of this test.
Table 7 - Residual effect
Measurement no Pre-test - 0.25 mg/L Alternating measurements of two concentrations
0.40 mg/L 0.25 mg/L
1
…
10
Mean value 1 – 2
Difference ( 1 – 2)
5.3.5 Test of influence of volume (flow rate change)
The influence of the delivered volume of gas shall be tested using test gas with mass concentration of
0.40 mg/L of ethanol in nitrogen (see Table 3, CRM no 4) at volumes of 1.5 L and 3 L and gas flow
rates of 0.20 L/s to 0.35 L/s. The difference between measurements must not exceed 0.010 mg/L. Each
of these 10 measurements must comply with the maximum permissible error for the given mass
concentration as referred to in Article 2.3. Table 8A may be used for recording and evaluation of this
test.
Table 8A – Influence of volume (flow rate change)
Measurement no Gas volume - 1.5 L Gas volume - 3 L Difference
1
…
10
Max. difference
If the breath alcohol analyser checks the volume of exhaled air automatically, measurements at higher
volumes of exhaled air are irrelevant.
General Measure No 0111-OOP-C040-13
14
In such a case, measurements shall be made using test gas with mass concentration of 0.40 mg/L of
ethanol in nitrogen (see Table 3, CRM no 4) at a flow rate within the maximum range of 0.15 L/s to
0.40 L/s or at the minimum gas flow rate possible (0.15 to 0.20) L/s and two higher flow rates of the
reference gas (0.25 to 0.30) L/s and (0.35 to 0.40) L/s. Ten measurements shall be performed for each
chosen gas flow rate. Again, the difference between measurements must not exceed 0.010 mg/L. Each
of these 10 measurements must comply with the maximum permissible error for the given mass
concentration as referred to in Article 2.3. Table 8B may be used for recording and evaluation of this
test.
Table 8B – Influence of volume (flow rate change)
Measurement no gas flow rate
(0.15 to 0.20) L/s
gas flow rate
(0.25 to 0.30) L/s
gas flow rate
(0.35 to 0.40) L/s
Difference
1
…
10
Max. difference
5.3.6 Test of influence of interfering components and CO2
This test shall be performed to check compliance with Article 3.5 by means of ten measurements using
dry gas with mass concentration of 0.40 mg/L of ethanol in nitrogen (see Table 3, CRM no 4) without
the interfering component and with the interfering component. The maximum influence of a given
interfering component must not exceed the values indicated in Table 9 and the sum of all interfering
components and CO2 must not exceed 0.40 mg/L. Table 9 may be used for recording and evaluation of
this test.
Table 9 - Influence of interfering components and CO2
Interfering component ( 5%) Acetone
0.50 mg/L
Methanol
0.10 mg/L
Isopropanol
0.10 mg/L
CO
0.20 mg/L
Toluene
0.20 mg/L
Measurement using pure gas
Measurement with the
interfering component
Influence
Max. influence 0.05 mg/L 0.10 mg/L 0.10 mg/L 0.08 mg/L 0.08 mg/L
(continued)
Table 9 - Influence of interfering components and CO2 (continued)
Interfering component ( 5%) Methane
0.30 mg/L
Acetaldehyde
0.15 mg/
CO2
100 mg/L L Total
Measurement using pure gas –
Measurement with the
interfering component
–
Influence
Max. influence 0.08 mg/L 0.10 mg/L 0.05 mg/L 0.40 mg/L
General Measure No 0111-OOP-C040-13
15
5.4 Tests of resistance to mechanical influences
5.4.1 Test of resistance to impact
This test, which is applied to verify compliance with the requirements under Article 3.6.1, must be
performed with the breath alcohol analyser without a carrying case under the following conditions:
– every impact must have the intensity of 15 gn, where acceleration 1 gn = 10 m/s2,
– duration: 11 ms,
– number of impacts: 1 for each axis perpendicular to the sample.
When at least two hours have elapsed after completion of the test, the measuring instrument must
comply with the requirements referred to in Article 2.3 when tested within the reference temperature
range.
5.4.2 Free-fall test
The free fall test, which is applied to verify compliance with the requirements under Article 3.6.1, the
breath alcohol analyser under test shall be allowed to fall freely from the height of 500 mm onto a rigid
test surface. The measuring instrument shall be allowed to fall on 3 sides, specifically the back, right
and left sides. The test shall not be performed for the front side (where the display is located). The test
shall be repeated two times for each side of the measuring instrument, i.e. a total of six free falls.
Immediately after the free falls, the measuring instrument shall be checked for changes in appearance.
No changes in indication may occur during the test. When at least one hour has elapsed after completion
of the test, the measuring instrument must comply with the requirements referred to in Article 2.3 in an
accuracy test at reference temperature.
5.4.3 Test of influence of random mechanical vibrations
In the test of compliance with the requirements under Article 3.6.1, the measuring instrument must be
exposed to broadband vibrations in three perpendicular axes under the following conditions:
– frequency range from 10 Hz to 150 Hz,
– spectral density of acceleration from 10 Hz to 20 Hz: 0.02 gn/Hz,
– spectral density of acceleration from 20 Hz to 150 Hz:–3 dB per octave,
– duration of 1 h for each axis.
When at least one hour has elapsed after completion of the test, the measuring instrument must comply
with the requirements referred to in Article 2.3 in an accuracy test at reference temperature.
5.5 Tests of resistance to climatic conditions
5.5.1 Cold test
a) The breath alcohol analyser in off-mode shall be placed in a temperature chamber at the temperature
of
–20°C for the period of 2 hrs.
When one hour has elapsed after completion of the test, the measuring instrument, stabilised at 20°C,
must comply with the requirements under Article 2.3 in an accuracy test using dry gas with mass
concentration of 0.40 mg/L of ethanol in nitrogen (see Table 3, CRM no 4) at reference temperature.
General Measure No 0111-OOP-C040-13
16
2) The breath alcohol analyser in on-mode shall be placed in a temperature chamber at the temperature
of –10°C. An accuracy test using dry gas with mass concentration of 0.40 mg/L of ethanol in nitrogen
(see Table 3, CRM no 4) shall be performed ten times under these climatic conditions. Each of these 10
measurements must comply with the maximum permissible error for the given mass concentration as
referred to in Article 2.3.
5.5.2 Dry heat test
This test involves placing the breath alcohol analyser in off-mode in a temperature chamber at the
temperature of +70°C for the period of 6 hrs.
Immediately after completion of the test, it shall be checked for changes in appearance. The appearance
of the measuring instrument must not change and the material and surface must not be cracked, swollen
or changed in colour.
When one hour has elapsed after completion of the test, the measuring instrument, stabilised at 20°C,
must comply with the requirements under Article 2.3 in an accuracy test using dry gas with mass
concentration of 0.40 mg/L of ethanol in nitrogen (see Table 3, CRM no 4) at reference temperature.
5.5.3 Damp heat test
a) A cyclic damp heat test (A cycle of 12 + 12 Hrs) shall be performed using two cycles within the
range of 25°C at a relative humidity of over 95% to 55°C at a relative humidity of 93%. The measuring
instrument shall be switched off during the test.
Immediately after the test, the measuring instrument shall be checked for changes in appearance.
When one hour has elapsed after completion of the test, the measuring instrument, stabilised at 20°C,
must comply with the requirements under Article 2.3 in an accuracy test using dry gas with mass
concentration of 0.40 mg/L of ethanol in nitrogen (see Table 3, CRM no 4) at reference temperature.
b) The damp heat test shall be performed at the temperature of 20°C at a relative humidity of 85%.
The measuring instrument shall be switched on during the test. An accuracy test using dry gas with
mass concentration of 0.40 mg/L of ethanol in nitrogen (see Table 3, CRM no 4) shall be performed ten
times under these climatic conditions. Each of these 10 measurements must comply with the maximum
permissible error for the given mass concentration as referred to in Article 2.3.
5.6 Test of influence of supply voltage
The test of influence of supply voltage shall be performed for breath alcohol analysers powered by
batteries with the supply voltage adjusted to Umax = Ubat.max and to Umin = Ubat.min, where Ubat.min is the
lowest operating battery voltage as specified by the supplier of the measuring instrument for ambient
temperature of 20°C, and Ubat.max is the no-load voltage of a new battery.
The measuring instrument must be functioning normally within the limits of the maximum permissible
error referred to in Article 2.3 in an accuracy test using dry gas with mass concentration of 0.40 mg/L
of ethanol in nitrogen (see Table 3, CRM no 4).
5.7 Tests for electromagnetic compatibility (EMC)
5.7.1 Electrostatic discharge immunity test
Electrostatic discharge immunity shall be tested with the measuring instrument switched on, preferably
using a 6 kV contact discharge or an 8 kV air discharge, where contact discharges cannot be applied.
The discharges shall be applied to the enclosure of the measuring instrument or a coupling plane
adjacent to the breath alcohol analyser.
General Measure No 0111-OOP-C040-13
17
During the test, 10 measurements for each discharge polarity shall be made sing measuring gas. One
discharge shall be applied during each measurement. The delay between the discharges should be at
least 10 seconds.
The difference between the values obtained under the conditions of interference and with no interference
must be less than 0.040 mg/l when dry gas with mass concentration of 0.40 mg/L of ethanol in nitrogen
(see Table 3, CRM no 4) is used.
After this test, the measuring instrument must be functioning normally within the limits of the maximum
permissible error referred to in Article 2.3 in an accuracy test with no interference performed using dry
gas with mass concentration of 0.40 mg/L of ethanol in nitrogen (see Table 3, CRM no 4).
5.7.2 Test of immunity to radiated high-frequency electromagnetic fields
Immunity to radiated high-frequency electromagnetic fields is tested with the measuring instrument
switched on within the frequency range of 26 MHz to 3000 MHz at test field intensity of 10 V/m when
measuring without modulation.
The test field is amplitude modulated to a depth of 80% and the modulation signal has a sinusoidal
waveform with modulation frequency of 1 kHz.
The measuring instrument under test is irradiated vertically and horizontally with a polarised field from
4 directions that are perpendicular to each other.
The tests shall be performed at the following frequencies: 26, 40, 60, 80, 100, 120, 144, 150, 160, 180,
200, 250, 350, 400, 435, 500, 600, 700, 800, 934, 960, 1 000, 1 200, 1 400, 1 700, 1 800, 1 900,
2 000, 2 400, 2 700, and 3 000 MHz. One measurement shall be made at each frequency using
measuring gas. If the measuring instrument is found to be influenced at any of the frequencies, testing in
the region of this frequency shall be performed to locate the point of maximum influence with a
frequency resolution of approx. 1%.
The difference between the values obtained under the conditions of interference and with no interference
must be less than 0.040 mg/l when dry gas with mass concentration of 0.40 mg/L of ethanol in nitrogen
(see Table 3, CRM no 4) is used.
After this test, the measuring instrument must be functioning normally within the limits of the maximum
permissible error referred to in Article 2.3 in an accuracy test with no interference performed using dry
gas with mass concentration of 0.40 mg/L of ethanol in nitrogen (see Table 3, CRM no 4).
5.7.3 Test of immunity to TETRA signals
Immunity to TETRA signals shall be tested with the measuring instrument switched on by irradiation
with a vertically and horizontally polarised field, gradually from four directions that are perpendicular
to each other.
The breath alcohol analyser shall be tested at the test frequencies of (380, 385, 390, 395, 400, 405, 410,
415, and 420) MHz ±0.1 MHz.
At each frequency, the of the test field level shall be gradually increased in 3 dB steps from 12 dB below
the specified test limit until the test limit is reached. The level at which any influence is observed to
begin to develop shall be recorded and included in the test report.
The test limit is expressed as the peak value of the modulated signal measured with a peak value
detector calibrated according to an equivalent effective value of the sinusoidal signal causing the same
deflection. The test limit for devices that are not used inside vehicles shall be 65 V/m.
Amplitude modulation by a square-wave signal with the frequency of 18 kHz and modulation depth of
> 98% further keyed with the frequency of 17 Hz must be used for tests of immunity to TETRA signals.
The keying duty cycle shall be 50%.
General Measure No 0111-OOP-C040-13
18
The difference between the values obtained under the conditions of interference and with no interference
must be less than 0.040 mg/l when dry gas with mass concentration of 0.40 mg/L of ethanol in nitrogen
(see Table 3, CRM no 4) is used.
After this test, the measuring instrument must be functioning normally within the limits of the maximum
permissible error referred to in Article 2.3 in an accuracy test with no interference performed using dry
gas with mass concentration of 0.40 mg/L of ethanol in nitrogen (see Table 3, CRM no 4).
6 Initial verification
6.1 General
6.1.1 Tests to be performed
The process of initial verification of a breath alcohol analyser shall comprise the following tests:
a) visual inspection,
b) accuracy test.
6.1.2 Test equipment
Initial verification tests of breath alcohol analysers shall be performed using the same measuring
instruments and equipment as those used for type approval in accordance with Article 5.1.2. In addition,
secondary CRM for ethanol in nitrogen traceable to a suitable primary reference material for the
relevant gas mixture mole (mass) fractions according to Table 10 shall be used.
Table 10 – Certified reference materials
CRM no Mass concentration of ethanol in
nitrogen (mg/L)
Number of
measurements using
the given gas
1 0.14 10
2 0.48 10
3 0.90 5
4 1.40 5
6.2 Visual inspection
The purpose of the visual inspection shall be to check that:
a) the measuring instrument submitted for verification conforms to the approved type,
b) the measuring instrument is complete and undamaged,
c) markings and inscriptions are complete, accurate and legible in conformity with the approved
type of the measuring instrument,
d) the software version installed is identical to that specified by the manufacturer and conforms to
the approved type of the measuring instrument.
6.3 Accuracy and repeatability test
The initial verification test of the breath alcohol analyser shall be performed using CRMs in accordance
with the requirements under Articles 5.3 and 5.3.1; the numbers of measurements for each gas are given
in Table 10 in Article 6.1.2.
General Measure No 0111-OOP-C040-13
19
The combined standard uncertainty of this arithmetic mean concentration comprises standard
uncertainty evaluated using method A - )(Au and standard uncertainty evaluated using method B -
)(Bu :
22 )()()( BAC uuu
Standard uncertainty evaluated using method A is calculated from the standard deviation of the
arithmetic mean multiplied by a coefficient which depends on the number of measurements:
AlA sku )(
where
kl coefficient which depends on the number of measurements n:
n 2 3 4 5 6 7 8 9 10 and
more
k l 7.0 2.3 1.7 1.4 1.3 1.3 1.2 1.2 1.0
sA standard deviation of the arithmetic mean,
)1(
)(1
2
nns
n
i
i
A
where
is the arithmetic mean of the mass concentrations of ethanol obtained when measuring the gaseous
CRM,
ßi the value obtained,
n number of measurements.
NOTE Type B uncertainty of the measured values is determined by the uncertainty of the certified content of
the CRM used for adjustment of the breath alcohol analyser, the uncertainty caused by deviation from linearity,
uncertainty of indication and/or uncertainty of the influence of deviation of temperature of the exhaled breath
from 34°C (detected in type approval of the measuring instrument), influence of sensitivities of other gas
components, influence of changes in barometric pressure (if the measuring instrument does not provide
automatic correction), influence of changes in the flow rate of the calibration gas, influence of other factors
related to the different principles of the measuring instruments used, etc.
For the purposes of verification and to ensure unambiguous and correct assessment of the breath alcohol
analyser, type B uncertainty shall be estimates only from the standard uncertainty of the mass
concentration of ethanol in the measured gaseous CRM u(ßRM) and the standard uncertainty of display
resolution uind:
2)( indB uu 2
RM )u(β
where
u(ßRM) is the standard uncertainty of the mass concentration of ethanol in the measured gaseous CRM
(derived from, e.g., the expanded uncertainty with coverage factor k = 2),
uind is the standard uncertainty of the display of the measuring instrument (rounding):
32
indind
ru
where
General Measure No 0111-OOP-C040-13
20
rind is the resolution of the values read from the display of the breath alcohol analyser.
Other uncertainties of the measured values that contribute to standard uncertainty evaluated using
method B does not need to be determined in the verification.
6.4 Correct adjustment of the measuring instrument
An analyser delivered for verification by the customer shall be correctly adjusted, with the minimum
standard uncertainty of the measured values. Otherwise, adjustment of the breath alcohol analyser to
reference values needs to be performed prior to the verification of the measuring instrument.
7 Subsequent verification
Subsequent verification shall follow a procedure identical to that for initial verification provided under
Chapter 6.
8 Notified standards
For the purposes of specifying the metrological and technical requirements for measuring instruments
and specifying the testing methods for their type approval and verification arising from this General
Measure, the CMI shall provide notification of the Czech technical standards, other technical standards
or technical documents of international or foreign organisations, or other technical documents
containing more detailed technical requirements (hereinafter referred to as “notified standards”). The
CMI shall publish a list of these notified standards attached to the relevant measures, together with the
general measure, in a manner accessible to the public (on the www.cmi.cz website).
Notified standards or parts thereof shall be considered complied with within the scope of and under the
conditions laid down in this General Measure, if the requirements laid down in this measure, to which
these standards or parts thereof apply, are complied with.
II.
G R O U N D S
The CMI has issued this General Measure laying down metrological and technical requirements for
specified measuring instruments and the testing methods for the verification of these specified measuring
instruments so as to implement § 24c of the Metrology Act.
Under item 7.4.2 of the Annex “List of the Types of Specified Measuring Instruments” to Decree No
345/2002 Coll. specifying the measuring instruments whose verification is mandatory and the
measuring instruments subject to type approval, as amended, breath alcohol analyser are classified as
measuring instruments subject to type approval and verification.
As such, the CMI has issued this General Measure laying down metrological and technical requirements
for breath alcohol analysers and the testing methods for type approval and verification of these specified
measuring instruments in order to implement § 24c of the Metrology Act for this particular type of
measuring instrument - breath alcohol analysers.
This regulation (General Measure) has been notified in accordance with Directive 98/34/EC of the
European Parliament and of the Council of 22 June 1998 laying down a procedure for the provision of
General Measure No 0111-OOP-C040-13
21
information in the field of technical standards and regulations and of rules on Information Society
services, as amended.
III.
I N S T R U C T I O N S
In accordance with § 173(2) of the APC, no appeals may be made regarding general measures.
In accordance with the provisions of § 172(5) of the APC, decisions on objections are final and appeals
may not be made against them.
Conformity of a general measure with legislation may be assessed in review proceedings pursuant to §§
94 to 96 of the APC. A party to the proceedings may initiate review proceedings to be conducted by the
administrative authority which issued the general measure. If the administrative authority finds no
reason to commence the review proceedings, it shall communicate and provide grounds for this within
thirty days. Pursuant to § 174(2) of the APC, a ruling on the commencement of review proceedings may
be issued within three years of the entry into force of the general measure.
IV.
E N T R Y I N T O F O R C E This General Measure shall enter into force on the fifteenth day of the date of publication (§ 24d of the
Metrology Act).
...…………………………..............
RNDr. Pavel Klenovský
Director General