+ All Categories
Home > Documents > Datalys® 780 Series Patient Monitors - lutechmedical.com · Respironics CO2 Mainstream (Capnostat...

Datalys® 780 Series Patient Monitors - lutechmedical.com · Respironics CO2 Mainstream (Capnostat...

Date post: 13-Jun-2018
Category:
Upload: trinhtram
View: 268 times
Download: 3 times
Share this document with a friend
2
Intensive Care Monitor Datalys® 780 Series Patient Monitors
Transcript
Page 1: Datalys® 780 Series Patient Monitors - lutechmedical.com · Respironics CO2 Mainstream (Capnostat 5): CO2 Range: CO2 Accuracy: ... Nurse call interface (760 and 780) SD card socket

Intensive Care Monitor

Datalys® 780 Series Patient Monitors

Page 2: Datalys® 780 Series Patient Monitors - lutechmedical.com · Respironics CO2 Mainstream (Capnostat 5): CO2 Range: CO2 Accuracy: ... Nurse call interface (760 and 780) SD card socket

LUTECH INDUSTRIES INC.

105 Remington Blvd. , Sui te C, Ronkonkoma, NY 11779 U.S.A.

Ph: 631-676-7432 Fx: 631-584-6699 Websi te: www.lutechmedical .com E-mai l : info@lutechmedical .com

Technical Speci f icat ionsRespirationMeasurement Method:RR Range:

RR Resoultion:RR Accuracy:Apnea Alarm:Alarm:

TemperatureTemp Units:Temp Range: Temp Resolution: Temp Accuracy:

Lutech SpO2Saturation Range: Saturation Resolution:Saturation Accuracy:

PR Range:PR Accuracy:

Alarm:

Nellcor SpO2 (Optional) Saturation Accuracy: PR Range: PR Accuracy:

Masimo SpO2 (Optional) Saturation Accuracy:

PR Range: PR Accuracy:

AcuTee™ NIBP Measurement Method: Measurement Modes:Auto Mode Times:Pressure Types:Pressure Units:Pressure Range: Range for Systolic Pressure:

Range for Diastolic Pressure:

Range for Mean Pressure:

Alarm Type:Pressure Accuracy: Pulse Rate Range:Over-Pressure Protection:

IBP (Optional)Measured Pressures:

Pressure Units:Pressure Range: Pressure Accuracy:

Cardiac Output (Optional)Measurement Method:C.O. Range:C.O. Resolution:C.O. Accuracy:Blood Temp (BT) Range:Injection Temp (IT)Range:BT/IT Resolution:BT/IT Accuracy:

Thoracic impedanceAdu/Ped: 0-120rpmNeo: 0-150rpm1rpm±1rpmYes, 10-40sUser-adjustable high and low limits

°C/°F selectable32-122°F (0-50°C)0.1°F (0.1°C)±0.3°F (±0.1°C)

1-100%1%±2% (70-100%, Adu/Ped)±3% (70-100%, Neo)Unspecified (0-69%)20-250bpm±3bpm (non-motion)±5bpm (motion)User-adjustable high, mid, low limit ranges

±3% (70-100%)Unspecified (0-69%)20-250bpm±3bpm

±2% (70-100%, Adu/Ped)±3% (70-100%, Neo)Unspecified (0-69%)25-240bpm±3bpm (non-motion)±5bpm (motion)

Automatic oscillometricManual, AutomaticAdjustable 1-480minSystolic, Diastolic, MeanmmHg/kPa selectable

Adu: 40-270mmHgPed: 40-200mmHgNeo: 40-135mmHg

Adu: 10-215mmHg Ped: 10-150mmHgNeo: 10-100mmHg

Adu: 20-235mmHgPed: 20-165mmHgNeo: 20-110mmHgSystolic, Diastolic, Mean±5mmHg40-240bpm

Adu: 300mmHgPed: 240mmHgNeo: 150mmHg

P1, P2, ART, PA, CVP, LAP, RAP, ICPmmHg/kPa selectable0-300mmHg±2% (100-300mmHg)±4% (0-100mmHg)

Thermodilution0.1-20L/min0.1L/min±5% or ±0.1L/min77-109.4°F (25-43°C)32-77°F (0-25°C)0.1°F (0.1°C)±0.3°F (0.1°C, no sensor)

CO2 (Optional)Measurement Method:

Respironics CO2 Mainstream (Capnostat 5): CO2 Range: CO2 Accuracy:

RR Range: Sidestream (LoFlo): CO2 Range: CO2 Accuracy:

RR Range:

PhaseIn CO2 Mainstream (IRMA) or Sidestream (ISA)CO2 Range:CO2 Accuracy:RR Range:

PhaseIn Anaesthetic Gas (Optional)Measurement Method:Mainstream (IRMA AX+): AG Range:

AG Accuracy:

RR Range:Sidestream (ISA AX+/OR+): AG Range:

AG Accuracy:

RR Range:

Impedance Cardiography (Optional)Measurement Method:C.O. Range:C.O. Accuracy:Stroke Volume Range:Stroke Volume Accuracy:HR Range:HR Accuracy:

Index of Consciousness (Optional)qCON Range:SQ I Range:EMG Range:ESR Range:Total Index Update Delay:

Non-dispersive infrared (NDIR)

0-150mmHg (at 760mmHg)0-19.7% (at 760mmHg)±2mmHg (0-40mmHg)±5% of reading (41-70mmHg)±8% of reading (71-100mmHg)±10% of reading (101-150mmHg)0-150rpm

0-150mmHg (at 760mmHg)0-19.7% (at 760mmHg)±2mmHg (0-40mmHg)±5% of reading (41-70mmHg)±8% of reading (71-100mmHg)±10% of reading (101-150mmHg)2-150rpm

0-15vol%±2mmHg (0-40mmHg)0-150rpm

Non-dispersive infrared (NDIR)

CO2: 0-10vol%N2O: 0-100vol%HAL: 0-8vol%ISO: 0-8vol%ENF: 0-8vol%SEV: 0-10vol%DES: 0-22vol%CO2: ±0.3vol%+2% of readingN2O: ±0.2vol%+2% of readingHAL: ±0.15vol%+5% of readingISO: ±0.15vol%+5% of readingENF: ±0.15vol%+5% of readingSEV: ±0.15vol%+5% of readingDES: ±0.15vol%+5% of reading0-150rpm

CO2: 0-10vol%N2O: 0-100vol%HAL: 0-8vol%ISO: 0-8vol%ENF: 0-8vol%SEV: 0-10vol%DES: 0-22vol%O2: 0-100% (for ISA OR+ only)CO2: ±0.2vol%+2% of readingN2O: ±0.2vol%+2% of readingHAL: ±0.15vol%+5% of readingISO: ±0.15vol%+5% of readingENF: ±0.15vol%+5% of readingSEV: ±0.15vol%+5% of readingDES: ±0.15vol%+5% of readingO2: ±1vol%+2% of reading (for ISA OR+ only)0-150rpm

Thoracic bio-impedance1.4-15L/minUnspecified5-250mLUnspecified40-250bpm±2bpm

0-990-100%0-100dB0-100%8s

SafetyISO 13485:2003 approvedCE marking according to MDD93/42/EECDegree of protection against harmful ingress of water: IP X 1

Dimension and WeightDimension: Datalys 780:

Datalys 780Plus:

Net Weight: Datalys 780: Datalys 780Plus: Gross Weight: Datalys 780: Datalys 780Plus:

Operation EnvironmentPower: Temperature: Humidity:

Patient RangePatients:

Performance SpecificationsDisplay Datalys 780: Datalys 780Plus:Resolution: Trace:

Sweep Speed:

Indicator:

Interface:

Battery:

Trend Time: Alarm:

Networking:

Recorder (Optional): Record Mode:

Record Speed:Paper Width:

ECGLead Selection:

Gain Selection:Sweep Speed:

Filter Bandwidth:

HR Range:

HR Resolution:HR Accuracy:HR Alarming Time:ST SegmentDetection Alarm:Arrhythmia Analysis:Alarm:

Anti-ElectricalInterference:

13.5in x 11.75in x 6.5in(343mm x 299mm x 165mm)13.57in x 11.46in x 6.5in(345mm x 291mm x 165mm)

9.7lbs (4.4kg)10.6lbs (4.8kg)

18lbs (8.2kg)22lbs (10kg)

AC100-250V, 50/60Hz32-104°F (0-40°C)15%-85%

Adult, Pediatric, Neonate

12.1” color TFT touch screen 15” color TFT touch screen800x6007 waveforms (750 & 760)12 waveforms (780/780Plus)6.25mm/s, 12.5mm/s, 25mm/s, 50mm/sPower, Battery, Alarm lights and QRS, Operating Key, Alarm soundsParameter cable interfaceAC power input socket Network interfaceUSB portVGA interface (760 and 780)Nurse call interface (760 and 780)SD card socket (760 and 780)DB9 interface (780 only)Lithium polymer plug & play rechargeable battery (4 hours of continuous working time)120 hoursUser-adjustable high and low limits for audible and visual alarmConnects to central monitoring system (wired and wireless)Built-in, thermal array, 2 channelsManual, Upon alarm, Continual, Time–defined25mm/s, 50mm/s1.97in (50mm)

3-lead (I, II, III)5-lead (I, II, III, V, aVR, aVL, aVF)12-lead (I, II, III, V1-V6, aVR, aVL, aVF) - for 780/780Plus onlyX1/4, X1/2, X1, X2, X4, Auto6.25mm/s, 12.5mm/s, 25mm/s, or 50mm/sMon: 0.5-40HzDIA: 0.05-130HzOPE: 1-20HzAdu: 15-300bpmPed/Neo: 15-350bpm1bpm±1% or ±2bpm<12s

-0.20mV-0.20mV (default)YesUser-adjustable high, mid, low limit ranges

Line frequency, electro-surgery and defibrillation protection, pacemaker detection