+ All Categories
Home > Documents > Derivation of Calibration Factors and Dial Settings Michaela … · Fidelis Calibration factors and...

Derivation of Calibration Factors and Dial Settings Michaela … · Fidelis Calibration factors and...

Date post: 11-Apr-2018
Category:
Upload: lyquynh
View: 223 times
Download: 3 times
Share this document with a friend
19
Derivation of Calibration Factors and Dial Settings Michaela Baker 29 November 2012
Transcript

Derivation of Calibration Factors and Dial Settings

Michaela Baker

29 November 2012

NPL Secondary standard ion chamber

Direct calibration with NPL primary standards: - Photon emitting radionuclides

- High energy beta emitting isotopes

Excellent stability over several decades (stdev: 0.1%)

- Accurate weighing

- Dilution

- Same reference time

NPL Calibration factors (pA/BMq):

Defined for specific: - Radionuclide (Individually derived from NPL primary standards)

- Dedicated holder

- Glass containers (vials, ampoules)

- Volume of liquid

Direct correlation between: Sample Activity (MBq) Current output (pA)

NPL Secondary standard ion chamber

Volume correction Factors: Radionuclide and Container dependent

- Minimal volume (mass) of active solution + Activity Assay

- Gradual top up with inactive carrier + Activity assay at each stage

Calibrator Response vs. Sample Volume: - graph (normalise to nominal volume)

- estimate % difference

Volume correction: I0 / Im = a2(m- m0)2 + a1(m- m0) + 1

where: I0 = current expected at the nominal mass of “m0”

Im = measured current at an individual mass “m”

NPL Secondary standard ion chamber

11.5

11.55

11.6

11.65

11.7

11.75

11.8

11.85

11.9

1.4 2.4 3.4 4.4 5.4 6.4 7.4 8.4

I m (

pA

)

Mass (g)

I123 10R Schott vial

y = -0.000299x2 + 0.004444x + 1.000026 R² = 0.998322

0.984

0.989

0.994

0.999

1.004

1.009

1.014

1.019

-2.5 -2 -1.5 -1 -0.5 0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5

Io/Im

(m-4)

I123 10R Schott vial

Fidelis Secondary standard radionuclide calibrator:

(Vinten, Isocal IV, NPL-CRC)

NPL-designed ion chamber: - same as the NPL SS ion chamber

- tested by NPL

Electrometer and user interface unit

Fidelis Calibration factors and volume correction factors:

- Transferable from the NPL SS ion chamber

- Calibrator independent

- Direct traceability to NPL standards

- Continuously updated: www.npl.co.uk/fidelis

Fidelis Calibration Factors

Calibrated directly using NIST standards of 60Co and 57Co

Expressed as: Calibration setting number (Dial factor)

Defined for specific: - Radionuclide: Relative to 60Co dial factor

- Container: “glass ampoule with 0.6mm wall thickness”

good approximation to syringes - corrections given

- Volume of liquid: 5 ml

Linear relation: Sample Activity (MBq) Dial factor <60Co Dial Factor <ICh response

Capintec Dial Setting numbers

123I UK comparison:

Container type/size:

0.8

0.9

1

1.1

1.2

1.3

1.4

1.5

1.6

1.7

0% 30% 60% 90% 120%

Sample volume as % of total container volume

Ca

pin

tec

ac

tiv

ity/N

PL

ac

tiv

ity

5 ml syringe (A)

1 ml syringe

5 ml syringe (B)

10 ml syringe

2 ml BS ampoule

5 ml BS ampoule

P6 vial

5 ml NBS ampoule0

5

1015

20

25

3035

40

45

5055

60

65

0.5 0.6 0.7 0.8 0.9 1.0 1.1 1.2 1.3 1.4 1.5 1.6

Reported value / NPL value

Number of

resultsReported P6 vial value/NPL value

Reported syringe value/NPL value

Why recalibrate?

10R Schott vial

(1.0 mm)

P6 Vial

( 1.2mm)

Difference

(0.2mm)

123I 1.721 pA/MBq 1.685 pA/MBq 2 %

90Y 0.0734 pA/MBq 0.0682 pA/MBq 7 %

99mTc 1.240 pA/MBq 1.227 pA/MBq 1 %

Confirm the accuracy of the calibration factors

Recalibrate/derive new factors for the preferred measurement format:

- radionuclide

- container type: vials, syringes

- sample volume

- shielding

- attenuating holders (99Mo breakthrough kit, 123I and 90Y Copper insert)

Why recalibrate?

Methods of calibration:

Standardised source

NPL calibration of hospital-supplied sources (comparison exercise)

Transfer of calibration factors/dial settings to other containers

Theoretically derived factors from photon energy response curves

(needs validated by other means)

- Standard MBq

Assay standard source: - Measured MBq

Fidelis: Increase/decrease calibration factor by % difference between

Standard MBq and Measured MBq => NEW Calibration Factor (pA/MBq)

Capintec: Gradually increase/decrease Dial setting until

Measured MBq = Standard MBq => NEW Dial Factor

Issues?

Limited availability – (Half life)

High cost

Calibration with standard source

NPL calibration of hospital-supplied sources

Advantages:

Cost effective

Direct traceability

Prior to calibration: Measured MBq

From Calibration: NPL MBq % difference: Measured MBq and NPL MBq

Post calibration:

Same matrix source Activity assay: Measured MBq

Fidelis: Increase/decrease Calibration factor by % difference from NPL calibration

=> NEW Calibration Factor (pA/MBq)

Capintec: Adjust Dial factor until Measured MBq is increased/decreased by

% difference from NPL calibration => NEW Dial Factor

Or: Prior to calibration: Measured Activities over a range of Dial factors

Post calibration: Select Dial factor for which Measured MBq = NPL MBq

NPL calibration of hospital-supplied sources

Calibration transfer to other containers

EXTENSION OF CALIBRATION FACTORS

TO OTHER CONTAINERS

Stock Solution

Ionisation Ionisation

Chamber Chamber precalibrated uncalibrated

geometry geometry

MBq/g pA/g

Calibration Figure

pA/MBq (well-defined geometry)

Accurate

weighing

Alternative method without weighing

Calibration transfer to other containers

Vial active

solution full: (known calibration factor)

Measure: Vial MBq

Transfer

active

solution

Vial residue

active solution: (refill with carrier)

Measure: Vial res MBq

Syringe activity = Vial full activity – Vial residue activity

Advantages:

•No weighing involved

•Not dependent of the type of syringe/needle or volume of solution in the syringe

Limitations:

- The accuracy of the “vial residue activity” measurement (beta emitters)

Syringe

active solution

Why?

- Correct Activity Assay prior to administration

- Improved diagnostic and treatment

- Waste disposal (123I)

Where from?

- Impurity check – gamma spectrometry

- Information available from the supplier

Impurities correction

How?

- Relative decay rate of impurities to that of the main radionuclide and/or

- Relative response of impurities to that of the main radionuclide

NPL ion chamber

Half life

(days)

Cal. Factor

(pA/MBq)

123I 0.55098 1.685

121Te 19.16 5.932

125I 59.388 0.3706

NPL ion chamber

Half life

(days)

Cal. Factor

(pA/MBq)

99mTc 0.25028 1.227

99Mo 2.7479 2.700

Impurities correction

89Sr 85Sr Ro/Ri % Impurity Corr

Half-lives: 50.57 days 64.850 days

NPL Chamber (P6): 0.0279 pA/MBq 5.258 pA/MBq 186 18 %

Capintec: 40.0 5 %

Xi (at measurement time) 0.12 %

89Sr correction

for 85Sr impurity

GPG93

89Sr Activity = Corr. factor A * Indicated Activity

“NPL Report DQL-RN 012

Comparison of Strontium-89

Solution Sources in UK Hospitals, 2003”

Available from:

www.npl.co.uk/rcuf see “publications”

Impurities correction

99Mo breakthrough kit

99Mo kit – dimensions

Lead walls thickness: 0.7 cm

Lead Density: 11.34 g cm-3

Height: 9.4 cm

Inner diameter: 4 cm

Calibrator

type

99mTc

Attenuation factor

99Mo

Attenuation factor

Lead wall thickness

effect

NPL

ion chamber

Measured: 4.5 GBq -100%

Theoretical: 4 GBq – 100%

Measured: 4.43

Theoretical: 4.5

± 0.05 cm:

± 6% response variation

Capintec 4.5 GBq -100% 8.04

Thank you

Title of Presentation

Name of Speaker

Date

The National Measurement System is the UK’s national infrastructure of measurement

Laboratories, which deliver world-class measurement science and technology through four

National Measurement Institutes (NMIs): LGC, NPL the National Physical Laboratory, TUV NEL

The former National Engineering Laboratory, and the National Measurement Office (NMO).

The National Measurement System delivers world-class

measurement science & technology through these organisations


Recommended