Design and construction of an quasi-adiabatic dissolutioncalorimeter with a novel dosing apparatus and a low heat capacity
Gordan Horvat • Josip Pozar • Zvonimir Dojnovic •
Dragutin Grgec • Sasa Blazeka
Received: 27 November 2013 / Accepted: 14 April 2014
� Akademiai Kiado, Budapest, Hungary 2014
Abstract A quasi-adiabatic calorimeter for determining
the molar solution enthalpies (DsolH) of non-volatile solids
was constructed. The design of the instrument was adjusted
to allow the determination of solution enthalpies of small
amounts of solids. For that purpose, the novel apparatus for
sample dosage with virtually negligible ‘‘blank heat’’ was
built. The rather low heat capacity of the calorimeter was
achieved by reducing the volume of the reaction cell
(20 mL), the dosing unit, and electric elements (the
thermistor and the heater). Good thermal isolation of the
reaction cell from the surroundings was accomplished by
placing the cell into an evacuated polypropylene vessel. A
computer program for processing the calorimetric data
according to modified Regnault–Pfaundler method was
written. The performance of the calorimeter was tested by
determining the heats of the reactions serving as thermo-
chemical standards at 25 �C (the dissolution of KCl and
NaCl in water and the dissolution of tris(hydroxymethyl)-
aminomethane in 0.1 mol dm-3 HCl(aq)). The obtained
data were in very good agreement with the literature
values.
Keywords Calorimetry � Quasi-adiabatic calorimeter �Molar solution enthalpy
Introduction
The interest in obtaining the molar solution enthalpies of
solids (DsolH) has been considerable, since many experi-
mentally unavailable reaction enthalpies can be obtained
from the cycles including the DsolH values [1, 2]. The
molar solution enthalpies also provide a deeper insight into
the energetics of the solute solvent interactions. For
instance, the thermodynamics of the solute solvation in
different solvents [3, 4] or the influence of the solvent on
the thermodynamic reaction state functions [5] can be
rationalized in terms of the so-called transfer functions of
reactants and products. The standard enthalpy of transfer
from one solvent to another can be easily obtained as a
difference between the corresponding DsolH values.
The molar solution enthalpies of solids are of use to the
researchers in the field of solid-state chemistry, particularly
those dealing with polymorphism. Based on the DsolH of
polymorphs in a certain solvent, the differences in the
corresponding lattice enthalpies can be calculated [6].
These values, combined with the knowledge regarding the
crystal structure of the forms involved, enable a deeper
thermodynamic insight into the structure–stability relation,
at least in terms of energetics. It should be also noted that
the accurate determination of molar solution enthalpies is
of great practical importance, particularly in chemical
technology [7–9] and in pharmaceutical industry [10–13].
In the past, the solution enthalpies of solid compounds
were predominantly determined by using the so-called
‘‘macro solution calorimeters’’ [14–16]. These instruments
were basically Dewar’s vessels with a relatively large
reaction cell (up to 100 mL) accommodating the thermo-
metric sensor, the calibrating device, and the dosing unit.
The heat capacity of the calorimeter was large, and the
consumption of the sample in the experiment was
This paper is dedicated to late Mr. Zvonimir Dojnovic.
Electronic supplementary material The online version of thisarticle (doi:10.1007/s10973-014-3829-9) contains supplementarymaterial, which is available to authorized users.
G. Horvat (&) � J. Pozar � Z. Dojnovic � D. Grgec � S. Blazeka
Division of Physical Chemistry, Department of Chemistry,
Faculty of Science, University of Zagreb, Horvatovac 102a,
10000 Zagreb, Croatia
e-mail: [email protected]
123
J Therm Anal Calorim
DOI 10.1007/s10973-014-3829-9
substantial. The sample addition (most commonly achieved
by crushing a glass ampoule containing the solid) was
usually accompanied with a measurable ‘‘blank heat,’’
which had to be determined in a separate experiment. The
calorimetric determinations of DsolH were hence limited to
solids of high solubility with relatively large absolute
values of molar solution enthalpies.
Nowadays, the ever growing number of organic com-
pounds with interesting chemical properties (receptors of
neutral species, ionophores, or pharmaceutically active
compounds) puts new demands on the design of dissolution
calorimeters [17–22]. This is due to the fact that these solids
are commonly scarcely soluble (particularly in water and in
protic solvents), and the absolute values of their molar
solution enthalpies are in most cases relatively low. As a
consequence, the ‘‘blank heats’’ in the calorimeters used for
the determination of the corresponding DsolH values should
be as close to zero as possible. In addition, the heat capacity
of the instrument must be low. It is therefore not surprising
that sensitive isothermal microcalorimeters have been pro-
posed as the instruments of choice for determining the
DsolH of such compounds [23]. However, the commercially
available isothermal microcalorimeters are still quite
expensive, and their manufacture requires far more skill and
knowledge from a calorimetrist than needed for the con-
struction of a quasi-adiabatic calorimeter. In the present
paper, we report on the construction of a simple, yet quite
sensitive quasi-adiabatic calorimeter with a moderate heat
capacity. This instrument could be an affordable alternative
to commercial dissolution microcalorimeters.
Theory
The equation which describes the heat measuring principle
of the quasi-adiabatic calorimeter reads as follows [24]:
DrHdn ¼ �CpdT þ dq; ð1Þ
where DrH denotes the reaction enthalpy, Cp is the isobaric
heat capacity of the calorimeter, dT is the temperature
change within the reaction cell, whereas the last term
denotes the heat exchanged between the calorimeter and
the surroundings (thermostat). By the appropriate con-
struction of the quasi-adiabatic calorimeter, the heat flow
between the system and the thermostat is minimized.
In a typical quasi-adiabatic dissolution, experiment
temperature is measured before, during and after the
reaction (Fig. 1). A pre-reaction period (between points A
and B in Fig. 1) is used for determining the rate of tem-
perature change in the calorimeter caused by non-reaction
heat effects, i.e., heat exchange with the surroundings, heat
production by stirring the reaction mixture, and the work of
the electrical current on the thermistor. In the short time
interval during this period, temperature change is usually
linear with time. The central part of temperature–time
curve, between points B and C, is the period in which the
studied process takes place. In the post-reaction period
(from C to D), the temperature is monitored for the same
reason as during the period A to B.
The integral form of Eq. (1) during the reaction time
period equals:
Zn
0
DrHdn ¼ �ZTC
TB
CpdT þZtC
tB
Cp
dT
dt
� �dt ð2Þ
If the expression (2) is divided by the isobaric heat
capacity of the calorimeter, which is presumed to be tem-
perature independent, and then integrated, the following
relation is obtained:
DTR ¼ TC � TB � DTadd ¼ DTtot � DTadd: ð3Þ
In derivation of this expression, the left term in Eq. (2)
was substituted with:
Zn
0
DrHdn ¼ �CpDTR: ð4Þ
In the Eq. (3), DTR is the temperature change caused by
the heat effect of the reaction, DTtot denotes the total
temperature change during the examined process, and
DTadd is the temperature change due to non-reaction heat
effects. The DTtot can be easily determined from the tem-
perature–time curve by subtracting the last value of the
temperature that belongs to the linear AB period from the
first temperature value that belongs to the linear CD period.
On the other hand, the DTadd has to be calculated by esti-
mating the magnitude of non-reaction heat effects during
bCDTC
tB tC
A
B
DC
Time
Tem
pera
ture
TBbAB
Fig. 1 Temperature–time curve for the passive quasi-adiabatic
calorimeter
G. Horvat et al.
123
the reaction. This can be done by using one of the several
different approaches [25–30]. In general, a function
describing the time dependence of the temperature change
caused by the non-reaction heat effects is needed:
DTadd ¼ZtC
tB
dT
dt
� �dt: ð5Þ
In the present paper, the slopes of the linear temperature
changes during periods AB and CD, namely bAB and bCD,
were used to obtain the dT/dt dependence during the
reaction period (further denoted by bBC(T)) by means of the
following two-parameter linear equation:
bBCðTÞ ¼bCD � bAB
�TCD � �TAB
ðTBCðtÞ � �TABÞ þ bAB; ð6Þ
where �TAB and �TCD denote average temperatures in the AB
and CD periods, respectively.
Temperature time dependence during reaction period
(TBCðtÞ) can be obtained by fitting the collected data with
an empirically found function. By combining the Eqs. (5)
and (6), the expression for DTadd is obtained:
DTadd ¼bCD � bAB
�TCD � �TAB
ZtC
tB
TBCðtÞdt � �TABðtC � tBÞ
0@
1A
þ bABðtC � tBÞ: ð7Þ
The integral in the Eq. (7) can be divided in several
integrals in order to adequately describe the temperature
time dependence during reaction period. In our data ana-
lysis, the reaction period was divided in two segments, and
time dependence of temperature in each of these intervals
was fitted with a polynomial of sixth order.
It should be noted that as a result of non-instant heat
conduction through the shield enclosing the thermistor, the
temperature measured during an experiment slightly differs
from the temperature of the system studied. Temperature in
the reaction cell can be obtained from the (measured)
temperature of the thermistor, Tth:
TsysðtÞ ¼ TthðtÞ þ sth
dTthðtÞdt
� �: ð8Þ
where sth denotes the time constant of the thermistor which
is equal to Cp,th/kth. This constant is easily determined from
the integral form of the Eq. (8) by non-linear regression of
Tth(t) data measured by immersing the thermometric sensor
in the thermostated bath whose temperature differs from
the initial temperature of thermistor.
The expression for reaction enthalpy can be obtained by
integration of (4) where DrH is presumed to be independent
on the reaction extent:
qp ¼ DH ¼ DrHDn ¼ �CpDTR: ð9Þ
DrH is equal to the slope of the linear dependence of
enthalpy change caused by the reaction on Dn.
Experimental
Materials
The solids, KCl (Sigma-Aldrich, TRACESelect; C99.9995 %),
NaCl (Sigma-Aldrich, ACS; C99.95 %), and tris(hydroxy-
methyl)-aminomethane (Sigma-Aldrich, 99.9? %) (further
denoted as TRIS), were used without further purification.
The potassium chloride and TRIS were stored over calcium
chloride in a vacuum dessicator. Sodium chloride was dried
at 200 �C for 20 h and held in a vacuum dessicator after-
ward. Prior to dissolution or drying, NaCl and TRIS were
grounded to a fine powder in order to hasten the dissolution
process. The KCl and NaCl were dissolved in deionized
B
I
E D
FG
A
C
H
Fig. 2 Shematic drawing of the calorimeter. A reaction cell, B poly-
propylene shield, C polypropylene lid, D thermistor, E heater,
F stirrer, G dosing apparatus, H screw, I vacuum valve
Quasi-adiabatic dissolution calorimeter
123
water. For the dissolution of TRIS in 0.1 mol dm-3 HCl(aq),
a standard solution of HCl (Kemika, Titrival) was used.
Apparatus
The calorimeter (Figs. 2 and S1–S4, Supporting Informa-
tion) consisted of a cone-shaped stainless steel reaction cell
(V = 20 mL) (A), mounted onto a polypropylene vessel
(B, 5-cm thick), the polypropylene lid (C, 5-cm thick), the
thermistor (D), the heater (E), the stirrer (F), and the dosing
apparatus (G). The lid of vessel could be fitted to the
reaction cell by means of two screws (H) and two bolts.
The tubes accommodating the thermistor, heater, stirrer,
and the dosing apparatus (D–G) were welded to the inner
side of the lid. The connecting wires were cemented into
the lid by the two component epoxy resin glue. The
thermistor and the heater were coated with caps made of
gold-plated copper. The stirrer was made from stainless
steel.
The dosing apparatus (Figs. 3 and S4, Supporting
Information), also made from stainless steel, includes the
cylinder (a), the lid (b), and the piston (c). The piston and
the lid are connected by a rod (d), approximately 0.5-cm
thick. The tight fit of the lid to the cylinder is achieved by
means of spring with a large force constant (e) and a ring
made of synthetic rubber, mounted on the inner surface of
the lid (f). The spring is coiled around a second rod (g,
1-cm thick), filling the space between the rear of the piston
(h) and the cap of the apparatus (i). The operation principle
is quite similar to that of a spring ballpoint pen. At a
desired instant, the spring could be compressed by applying
the pressure on the cap, via a servo electromotor, thereby
causing the piston (c) and the lid (b) to move downward.
This, in turn, exposes the solid, placed in the space between
the piston and the lid (the sample compartment (j)), to the
solvent. The volume of the sample compartment inside the
dosing apparatus was 0.75 cm3. Blanks showed no
detectable heat effect due to this operation (Fig. S5).
The temperature sensor was a 5 kX thermistor (at
T & 298.15 K), whereas the heater consisted of coiled
copper wire (R = 15.2 X). The power of the heater could
be adjusted by varying the voltage across the feeding cir-
cuit. The resistance of the thermistor and the potential drop
across the heater were measured by means of the Agilent
34411A digital multimeter. The heating time was measured
by a stopwatch with a hundredth of a second precision. The
watch was coupled mechanically to the switch actuating
the heater. The stirrer (fitted with a propeller) was operated
by a synchronous motor driven by a 24 V power supply.
The calorimeter and the solvents used were thermostated
in an air thermostat made of Plexiglas. Temperature within
the thermostat was regulated up to a tenth of degree pre-
cision (Fig. S6, Supporting Information) by means of a
heater and thermocouples used for cooling. Contact mer-
cury thermometer was used for temperature control.
The addition of solid was made when a sufficiently low
change in temperature of the calorimetric cell content with
time was observed (approximately 10 min after the
experiment startup).
Methods
The thermistor was calibrated in the temperature range
278.15 B T/K B 323.15. For that purpose, a constant
temperature bath (PolyScience 9101) with a thermometer
of 0.01 K temperature resolution was used. The accuracy
of its temperature scale was assured by calibration that can
be traced to NIST temperature standards. The time constant
of the temperature sensor was obtained by monitoring the
time needed to accomplish the thermal equilibrium
between the sensor and the thermostat. These experiments
were done in triplicate. The obtained time constant was
used for the deconvolution of temperature–time curves
according to Eq. (8).
Prior to determination of the reaction enthalpies, the
performance of the calorimeter was tested by electrically
supplying the variable amount of heat (1 \ q/J \ 6) to the
reaction cell containing 20 mL of deionised water.
The solution enthalpies of KCl and NaCl in water and of
TRIS in 0.1 mol dm-3 HCl(aq) were determined by pro-
cessing the time dependence of the thermistor resistance.
The calorimeter was in all cases calibrated after the dis-
solution of compounds. The obtained data were processed
using the aforementioned, modified Regnault–Pfaundler
method implemented in a home-made software, written in
Wolfram Mathematica 7 (Fig. S7, Supporting Information).
Molar solution enthalpies were determined by linear
regression from the dependence of enthalpy changes on the
reaction extent given by Eq. (9). Linear and non-linear
curve fitting and data plotting were performed using the
Origin 7.5 software.
a
j
c
d
e
f
g
i
b
h
Fig. 3 Schematic drawing of
the dosing apparatus. a cylinder,
b lid, c piston, d steel rod,
e spring, f ring made of
synthetic rubber, g steel rod,
h piston rear, i cap, j sample
compartment
G. Horvat et al.
123
Results and discussion
Thermistor calibration and the performance
of the calorimeter
The response of the thermometric sensor was examined by
its abrupt immersion from air with the temperature of
298.55 K to a thermostated bath held at 295.15 K (Fig. S8,
Supporting Information). Time dependence of temperature
was fitted according to integral form of Eq. (8), and a time
constant of thermistor was obtained, sth = (1.34 ± 0.01) s.
The results of thermistor calibration are shown in Fig. 4.
As can be seen, the agreement between the experimental
data and the values calculated according to relation [31]
R Tð Þ ¼ Aexp �B=Tð Þ þ C ð10Þ
was excellent. The parameters A = (5.53 9 10-2 ± 5 9
10-4) X, B = (3,383 ± 2) K, and C = (–130 ± 3) X were
therefore used in processing the calorimetric data.
The sensitivity of the calorimeter was checked by a series
of electrical calibration experiments in which a variable
energy input to a calorimeter containing 20 mL of deionised
water was applied. The obtained heat capacities are given in
Table 1. The temperature–time curves recorded at the lowest
and the highest heats supplied are shown in Fig. 5a, b,
respectively.
The data presented in Table 1 suggest that the obtained
heat capacity is virtually independent on the calibration
heat. In addition, the standard deviations of the Cp values,
determined by repetitive calibrations with a similar energy
input from the heater (±1 %), were quite low. The
instrument is therefore sufficiently precise to allow the
determination of rather small enthalpy changes.
Determinations of molar solution enthalpies
The calorimeter accuracy was tested by performing three
reactions with a well-known reaction enthalpies at
&298.15 K, the dissolution of KCl [32, 33] and NaCl in
water [34–36] and of TRIS in HCl (aq, c = 0.1 mol dm-3)
[14, 21, 26, 37, 38]. For the reactions examined, multiple
experiments with variable sample mass and consequently
with variable solution enthalpy changes were carried out.
The addition of potassium chloride in water resulted in a
prolonged decrease of temperature of the calorimeter
content (approximately 50 s for higher amounts of solid
added, Figs. 6 and S10, Supporting Information), which
indicated that the dissolution of this solid was a relatively
slow process. Once the reaction came to a halt, a linear
change of temperature with time was again observed. The
calorimeter was then calibrated by means of electric heater.
From the temperature change in the reaction period and the
isobaric heat capacity of the calorimeter, heats associated
with the KCl dissolution in water were determined (Fig. 7;
Table S1).
As can be seen in Fig. 7, the obtained DH versus ndependence was linear. This result indicated that the
corresponding slope can be considered as the specific
dissolution enthalpy of potassium chloride in the examined
concentration range. The determined value DsolH =
(17.74 ± 0.11) kJ mol-1 is in very good agreement with
the values which can found in the literature, namely
17.584 kJ mol-1 determined by Kilday [32] and
17.587 kJ mol-1 measured by Gunn [33].
As in dissolution experiments with potassium chloride,
the addition of sodium chloride to water resulted with
relatively slow temperature change during the reaction
period (the decrease in temperature of the reaction mixture
once the solid was exposed to the solvent could be
observed for approximately 50 s, Figs. 6 and S11, Sup-
porting Information). The calorimeter was again calibrated
when a linear dependence of temperature with time was
observed (approximately 100 s after the sample addition).
Heats associated with the dissolution of NaCl are given in
Fig. 7 and in Table S2.
The obtained DH versus n dependence was linear
(Fig. 7). Owing to low concentrations of NaCl, the
obtained value should be close to the standard molar
enthalpy of solution of NaCl (DsolH8). The agreement
280 285 290 295 300 305 310 315 320 325
2
3
4
5
6
7
8
9
10
Res
ista
nce/
kΩ
Temperature/K
Fig. 4 Thermistor calibration. Filled square measured values, line
calculated values
Table 1 Isobaric heat capacity of the calorimeter obtained by elec-
trical calibration
�qp/J (Cp ± SE)/J K-1
0.98 112.0 ± 2.2
1.53 112.2 ± 1.2
3.04 113.6 ± 1.1
5.89 112.7 ± 0.7
SE standard error of the mean (N = 16–23)
Quasi-adiabatic dissolution calorimeter
123
between the determined value DsolH8 = (4.258 ± 0.022)
kJ mol-1 and those found in the literature, 4.213 kJ mol-1
measured by Archer and Kirklin [35], and 4.192 kJ mol-1
determined by Lehto et al. [22] is very good.
Contrary to the slow dissolution of sodium chloride,
the dissolution of TRIS in hydrochloric acid (c =
0.1 mol dm-3) was rather rapid (Figs. 6, S12, Supporting
Information). Hence, after the addition of TRIS sample, a
temperature rise lasting 10–30 s was observed. As in the
cases of KCl and NaCl, the DsolH was independent on the
amount of solid dissolved (Fig. 7; Table S3). The deter-
mined DsolH = (-29.72 ± 0.22) kJ mol-1 was again in
quite good agreement with the literature data, ranging from
-29.63 kJ mol-1 to -29.75 kJ mol-1 [14, 21, 26, 37, 38].
Conclusions
The constructed isoperibolic calorimeter possesses rela-
tively low thermal conductivity achieved by placing the steel
reaction cell in the vacuum jacket made of polypropylene.
The novel design of the sample dosing apparatus enabled the
addition of the solid accompanied with virtually zero ‘‘blank
heat’’. Moreover, the solid can be loaded into the sample
compartment as a fine powder, without a mechanical pre-
treatment (for instance pressing into a tablet), required in
certain types of instruments [19]. Because of the small size of
the reaction cell (20 mL), the heat capacity of the calorimeter
is rather low. The described instrument hence fulfills the
requirements needed for the determination of solution
enthalpies of small amounts of solids. Its main advantage, in
comparison with the instruments of similar performance, is
the simple and inexpensive construction (the value of the
parts needed does not exceed a sum of 200 Euros).
0 10 20 30 40 50 60
0
2
4
6
8
10
(a) (b)
ΔTem
pera
ture
/mK
Time/s0 10 20 30 40 50 60
0
10
20
30
40
50
Time/s
ΔTem
pera
ture
/mK
Fig. 5 Electrical calibration of the calorimeter filled with 20 mL of water. a �qp ¼ 0:99 J, T0 = 296.12 K. b �qp ¼ 5:87 J, T0 = 297.06 K.
DTemperature = (Tsys(t) - T0)
0 25 50 75 100 125 150
–60
–40
–20
0
20
40
60
80
100
120
KCl
NaCl
ΔTem
pera
ture
/mK
Time/s
TRIS
Fig. 6 Temperature–time curves for the dissolution of KCl
(m = 32.22 mg, T0 = 298.106 K) and NaCl (m = 37.43 mg,
T0 = 298.127 K) in water (V = 20 mL) and for the dissolution of
TRIS (m = 20.40 mg, T0 = 298.104 K) in hydrochloric acid
(V = 20 mL, c = 0.1 mol dm-3) and for the subsequent electrical
calibration of the calorimeter. DTemperature = (Tsys(t) - T0)
0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4
–8
–6
–4
–2
0
2
4
6
8
10
KCl
NaCl
Hea
t/J
103 Δ /mol
TRIS
ξ
Fig. 7 Heat of dissolution of KCl, NaCl, and TRIS in water as a
function of the compound mass
G. Horvat et al.
123
Acknowledgements This work was supported by the Ministry of
Science, Education and Sports of the Republic of Croatia (Projects
119-1191342-2960 and 119-1191342-2961).
References
1. Hunenberger P, Reif M. Single-ion solvation: experimental and
theoretical approaches to elusive thermodynamic quantities. 1st
ed. Cambridge: Royal Society of Chemistry; 2011. p. 325–98.
2. Danil de Namor AF. Thermodynamics of Calixarene-Ion Inter-
actions. In: Asfari Z, Bohmer V, Harrowfield J, Vicens J, editors.
Calixarenes 2001. Dordrech: Kluwer Academic Publishers; 2001.
p. 346–64.
3. Cox B, Hedwig G, Parker A, Watts D. Solvation of ions. XIX.
Thermodynamic properties for transfer of single ions between
protic and dipolar aprotic solvents. Aust J Chem. 1974;27:477–501.
4. Piekarski H. Calorimetry: an important tool in solution chemistry.
Thermochim Acta. 2004;420:13–8.
5. Danil de Namor AF, Cleverley RM, Zapata-Ormachea ML.
Thermodynamics of calixarene chemistry. Chem Rev. 1998;98:
2495–526.
6. Salvetti G, Tognoni E, Tombari E, Johari GP. Excess energy of
polymorphic states or glass over the crystal state by heat of
solution measurement. Thermochim Acta. 1996;285:243–52.
7. Yagoubi M, Bertaud Y, Castanet R. Etude calorimetrique de
l’enthalpie de dissolution d’alumines industrielles dans des bains
cryolithiques. J Therm Anal. 1990;36:589–97.
8. Ayed F, Sorrentino F, Castanet R. Determination par calorimetrie
de dissolution des enthalpies de formation de quelques silicates,
aluminates et alumino-silicates de calcium. J Therm Anal.
1994;41:755–66.
9. Gailhanou H, Blanc P, Rogez J, Mikaelian G, Kawaji H, Olives J,
et al. Thermodynamic properties of illite, smectite and beidellite
by calorimetric methods: enthalpies of formation, heat capacities,
entropies and Gibbs free energies of formation. Geochim Cos-
mochim Acta. 2012;89:279–301.
10. Vippagunta SR, Brittain HG, Grant DJW. Crystalline solids. Adv
Drug Delivery Rev. 2001;48:3–26.
11. Yu L, Reutzel SM, Stephenson GA. Physical characterization of
polymorphic drugs: an integrated characterization strategy.
Pharm Sci Technol Today. 1998;1:118–27.
12. Royall P, Gaisford S. Application of solution calorimetry in
pharmaceutical and biopharmaceutical research. Curr Pharm
Biotechnol. 2005;6:215–22.
13. Souillac PO, Dave P, Rytting JH. The use of solution calorimetry
with micellar solvent systems for the detection of polymorphism.
Int J Pharm. 2002;231:185–96.
14. Hill JO, Ojelund G, Wadso I. Thermochemical results for ‘‘tris’’
as a test substance in solution calorimetry. J Chem Thermodyn.
1969;1:111–6.
15. Simeon V, Ivicic N, Tkalcec M. Construction of a simple reaction
calorimeter and heats of some standard reactions at 20 �C*.
Z Phys Chem. 1972;78:1–12.
16. Yu H, Liu Y, Tan Z, Dong J, Zou T, Huang X-M, et al. A
solution-reaction isoperibol calorimeter and standard molar
enthalpies of formation of Ln(hq)2Ac (Ln = La, Pr). Thermo-
chim Acta. 2003;401:217–24.
17. Nilsson S-O, Wadso I. A flow-microcalorimetric vessel for
solution of slightly soluble solids. J Chem Thermodyn.
1986;18:1125–33.
18. Hallen D, Nilsson S-O, Wadso I. A new flow-microcalorimetric
vessel for dissolution of small quantities of easily or slightly
soluble liquids. J Chem Thermodyn. 1989;21:529–37.
19. Bastos M, Bai G, Qvarnstrom E, Wadso I. A new dissolution
microcalorimeter: calibration and test. Thermochim Acta.
2003;405:21–30.
20. Moreno-Pirajan JC, Giraldo-Gutierrez L. Design, calibration, and
test of a new dissolution isoperibol microcalorimeter. Instrum Sci
Technol. 2007;35:453–63.
21. Vargas EF, Moreno JC, Forero J, Parra DF. A versatile and high-
precision solution-reaction isoperibol calorimeter. J Therm Anal
Calorim. 2008;91:659–62.
22. Lehto V-P, Tenho M, Hamalainen O-P, Salonen J. Calorimetric
determination of dissolution enthalpy with a novel flow-through
method. J Pharm Biomed Anal. 2010;53:821–5.
23. Russell DJ, Thomas D, Hansen LD. Batch calorimetry with sol-
ids, liquids and gases in less than 1 mL total volume. Thermo-
chim Acta. 2006;446:161–7.
24. Blandamer MJ, Cullis PM, Gleeson PT. Three important calori-
metric applications of a classic thermodynamic equation. Chem
Soc Rev. 2003;32:264–7.
25. Gunn SR. On the calculation of the corrected temperature rise in
isoperibol calorimetry. Modifications of the Dickinson extrapo-
lation method and treatment of thermistor–thermometer resis-
tance values. J Chem Thermodyn. 1971;3:19–34.
26. Yff BTS, Royall PG, Brown MB, Martin GP. An investigation of
calibration methods for solution calorimetry. Int J Pharm.
2004;269:361–72.
27. Santos LMNBF, Silva MT, Schroder B, Gomes L. Labtermo:
methodologies for the calculation of the corrected temperature rise
in isoperibol calorimetry. J Therm Anal Calorim. 2007;89:175–80.
28. Haloua F, Hay B, Foulon E. Uncertainty analysis of theoretical
methods for adiabatic temperature rise determination in calo-
rimetry. J Therm Anal Calorim. 2013;111:985–94.
29. King A, Grover H. Temperature correction methods in calorim-
etry. J Appl Phys. 1941;12:557–68.
30. West ED. A two-body model for calorimeters with constant-
temperature environment. J Appl Phys. 1968;39:4206–15.
31. White N. Traceable temperatures: an introduction to temperature
measurement and calibration. 2nd ed. Chichester: Wiley; 2001.
32. Kilday MV. Enthalpy of solution of SRM 1655 (KCl) in H2O.
J Res Nat Bur Stand. 1980;85:467–81.
33. Gunn S. Comparison standards for solution calorimetry. J Phys
Chem. 1965;9:2902–13.
34. Hubert N, Solimando R, Pere A, Schuffenecker L. Dissolution
enthalpy of NaCl in water at 25 �C, 45 �C and 60 �C. Determi-
nation of the Pitzer’s parameters of the H2O–NaCl system and the
molar dissolution enthalpy at infinite dilution of NaCl in water
between 25 �C and 100 �C. Thermochim Acta. 1997;294:157–63.
35. Archer D, Kirklin D. NIST and standards for calorimetry. Ther-
mochim Acta. 2000;347:21–30.
36. Archer DG, Kirklin DR. Enthalpies of solution of sodium chloride and
potassium sulfate in water. Thermodynamic properties of the potas-
sium sulfate ? water system. J Chem Eng Data. 2002;47:33–46.
37. Gunn S. A comparison of standard reactions for solution and
combustion calorimetry I. Solution-calorimeter measurements.
J Chem Thermodyn. 1970;2:535–47.
38. Ivicic N, Simeon V. Enthalpies of ligation of Co2? and Cu2? to
epimeric threonines and isoleucines. Thermochim Acta. 1978;25:
299–305.
Quasi-adiabatic dissolution calorimeter
123