Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN
Transcript
Page 1: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

cs

Accu-Chek Inform II Blood Glucose Monitoring

System.

Nurse Educator: Lois Sandford, RN

Page 2: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csTraining Objectives

By the end of this session you will be able to:

• Explain why training is required.

• Identify contraindications/test interferences.

• State Why and When you have to quality control your meter.

• Perform a Quality control test as per hospital policy.

• Identify Point of Care contact on site and where you get your meter supplies.

Page 3: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

cs

Why Training?

• “The use of blood glucose meters by untrained staff, without adequate management supervision of the equipment and without the use of quality control procedures, can lead to misleading results, adversely affecting the treatment of patients”

DH

• Medicine and Healthcare Products Regulatory Agency

• Nursing & Midwifery Council

• Local Trust

• RCN

• Department of Health

Page 4: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

cs

Contraindications for use of Blood Glucose System.

Due to the Hazard Notices being issued the MHRA recommend that therapeutic decisions should not be based on a capillary sample in these circumstances:

Venous Sample should Venous Sample should be sent to the Lab for be sent to the Lab for verification of blood verification of blood glucose levelglucose level

• Dehydration

• Hypotension

• Shock

• Peripheral Occulsive Disease

• Hyperosmolar Non-Ketotic Coma

• Diabetic Ketoacidosis

• Unconscious Patients

• Decompensated Heart Failure NYHA Class IV

Page 5: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

cs

LIMITATIONS.

Due to medical conditions:

• Galactose > 1.11 mmol/L (overestimation)

• Lipids > 16.8 mmol/L (overestimation)

• Sodium > 150 mmol/L (underestimation)

(Hypernatremia)

For further information refer to Test Strip Information Leaflet,

Workstation insert or Ward Manual

Page 6: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csInform II – Front View.

Test Strip Port

Touch screen

On/off button

Base unit

Page 7: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csInform II – Back View.

Bar code scanner

Battery Pack

Wireless Card Reader

Page 8: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csAdvanced Performa Test Strip.

0.6μL sample size

Robust End Dosing Strip

Sample volume detection

Fast 5 Second Result

Page 9: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csAccu-Chek Inform II

Power up screens

Page 10: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csStart Up – User Identification.

1 Press and release the On/Off button . The system is now on.

2 The Power Up screen appears.

•Bar code scan of operator ID

•User ID unique to user and

compulsory for all activities

Page 11: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csMain Menu Screen Shot.

Page 12: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

cs

Quality Control Testing

• Why Quality Control?

• How to Quality Control Inform II Meter

• When to Quality Control?

• Quality Control Lockout.

• EQA

Page 13: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

cs

Quality Control Screens:

Page 14: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csControl and Strip Confirmation

Screens

Page 15: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csPerforming a Quality Control

Test.

Page 16: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csResult Screen:

Page 17: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csQuality Control Test:

• Checks accuracy of test strips and Inform Meter

• Level 1 & 2 controls MUST be analysed every 25 hours – otherwise the user will be locked out of analysing patient samples.

• In an emergency – 1 stat test is allowed.

Page 18: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csSTAT TEST :

Page 19: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csFailed Quality Control Test

• Level 1 & 2 solutions must pass QC test.

• If test displays FAIL: Add Comment and repeat test.

• Troubleshoot:

• Check expiry dates, mix/change QC solutions & test strips – repeat test.

• Contact relevant department in hospital report fault and obtain replacement meter

If meter continues to fail DO NOT use it!If meter continues to fail DO NOT use it!

Page 20: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csPatient Preparation.

•Gain consent

•Infection Control

•Handwashing

•Sites for lancing

Page 21: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

cs

The Case For Using a Finger-Pricking Device!!!

Page 22: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csPatient Testing.

•Select patient test from menu

•Select Glucose Test

•Scan in Patient hospital number.

•Meter will prompt you to confirm patient details

•Scan in Test strips

Page 23: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csCheck test strip code and insert

strip.

~ Check Patients details at top of screen

Page 24: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csApply patient blood when

prompted.

Always perform a visual inspection of the yellow test area.

Page 25: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

cs

Results.

• Normal Range = 3.5 m/mols – 7.0 m/mols.

• Critical Range = < 3.0 m/mols - >20.0 m/mols.

• If Critical Range ……….. prompts….

• If HI………………………prompts…..

• If LO………………………prompts……

• N.B. Comments…………………………

• Hypo Boxes on the ward.

Page 26: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

cs

Review Results.

• From main menu – select review result – Glucose Test

• All results are displayed on screen – Patient & QC.

• Select patient – to view patient results - scan ID

• Select QC – to view QC results.

Page 27: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

cs

Cleaning the Meter.

.

• A lightly damp cloth before and after patient use and mild soapy water may be used.

• Acti-max Clothes – dry off after use.

• Follow hospital policy/local Guidelines for decontamination.

(Sani-Cloth wipes)

Page 28: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

csTraining Objectives

By the end of this session you will be able to:

• Explain why training is required.

• Identify contraindications/test interferences.

• State Why and When you have to quality control your meter.

• Perform a Quality control test as per hospital policy.

• Identify Point of Care contact on site and where you get your meter supplies.

Page 29: Diagnostics Accu-Chek Inform II Blood Glucose Monitoring System. Nurse Educator: Lois Sandford, RN.

Dia

gn

osti

cs

Thank youfor listening…

Any further questions?