Digital FLUORO

Summer 2008

Page 2: Digital fluoro

DIGITAL FLUORO

Page 3: Digital fluoro

• Digital fluoroscopy is currently most commonly configured as a conventional fluoroscopy system in which the analog video signal is converted to a digital format with an analog-to-digital converter (ADC). An in-depth discussion of digital detector technologies (eg, flat-panel "direct" detection of x rays and charge-coupled device technology) is beyond the scope of this article. After a review of several fundamental digital imaging concepts including binary numbers, pixels, and gray levels, emphasis will be placed on discussions of the digital imaging tools specific to digital fluoroscopy and digital subtraction angiography (DSA).

Page 4: Digital fluoro

• Digital fluoroscopy is currently most commonly configured as a conventional fluoroscopy system in which the analog video signal is converted to a digital format with an analog-to-digital converter (ADC).

• With discussions of the digital imaging tools specific to digital fluoroscopy and digital subtraction angiography (DSA).

Page 5: Digital fluoro

Early Fluoroscopy

Page 6: Digital fluoro

CONVENTIONAL FLUOROSCOPYINVENTED BY THOMAS EDISON

Page 7: Digital fluoro

Direct Fluoroscopy: obsolete

In older fluoroscopic examinations radiologist stands behind screen and view the pictureRadiologist receives high exposure; despite protective glass, lead shielding in stand, apron and perhaps goggles

Main source staff exposure is NOT the patient but direct beam

Page 8: Digital fluoro
Page 9: Digital fluoro
Page 10: Digital fluoro

pixels• the smallest element of a digital image.• A digital image is normally composed of a two-

dimensional (square) matrix of pixels.• The matrix size of an image is used to describe

the number of pixels in each row and column of the image.

Page 11: Digital fluoro

The size of a pixel determines the smallest detail visible in the image,

the number of pixels (of a given size) determines the field of view of the image. pixel value is assigned to a certain color or gray level for visualization. The imaged object is more faithfully reproduced as matrix size increases and pixel size decreases.

Page 14: Digital fluoro

Digital fluoroscopy• is most commonly configured as a conventional

fluoroscopy system (tube, table, image intensifier, video system) in which the analog video signal has been digitized with an ADC.

• Alternatively, digitization may be accomplished with a digital video camera (eg, a charge-coupled device) or via direct capture of x rays with a flat-panel detector.

• For digital fluoroscopy systems in which the analog video signal is digitized with an ADC, the resolution is limited by the resolution of the video camera, which is typically 1–2 line pairs per millimeter.

Page 15: Digital fluoro
Page 16: Digital fluoro
Page 17: Digital fluoro
Page 18: Digital fluoro
Page 19: Digital fluoro

Digital Image recording

• In newer fluoroscopic systems film recording replaced with digital image recording.

• Digital photospots acquired by recording a digitized video signal and storing it in computer memory.

• Operation fast, convenient.• Image quality can be enhanced by application

of various image processing techniques, including window-level, frame averaging, and edge enhancement.

• But, the spatial resolution of digital photospots is less than that of film images.

Page 20: Digital fluoro

Digital radiography principle

Clock

MemoryADCI

Iris

t

t

ANALOGUESIGNAL

DIGITALSIGNAL

See more in Lecture L20

Page 21: Digital fluoro

Digital spot film images

• and photospot images may be acquired by using the same digital fluoroscopy system.

• Individual frames from a digital fluoroscopy sequence can be stored digitally and can be used instead of conventional spot film and photospot images.

• Digital photospot images will have the same characteristics (eg, resolution) as digital fluoroscopic images.

Page 22: Digital fluoro

Digital format – print to film

Page 23: Digital fluoro

Remote – over the table tube

Page 24: Digital fluoro

Different fluoroscopy systems

• Remote control systems– Not requiring the presence of

medical specialists inside the X Ray room

• Mobile C-arms– Mostly used in surgical

theatres.

Page 25: Digital fluoro
Page 26: Digital fluoro
Page 27: Digital fluoro
Page 28: Digital fluoro
Page 29: Digital fluoro
Page 30: Digital fluoro

• The digital image data from digital fluoroscopy may be processed by using many useful image processing techniques.

• These techniques may serve to decrease radiation exposure to the patient and medical imaging staff or improve visualization of anatomy.

• Processing options include last image hold, gray-scale processing, temporal frame averaging, and edge enhancement.

• Additional processing is available when digital fluoroscopy data are used to perform DSA.

Page 31: Digital fluoro
Page 32: Digital fluoro
Page 33: Digital fluoro
Page 34: Digital fluoro
Page 35: Digital fluoro
Page 36: Digital fluoro
Page 37: Digital fluoro
Page 40: Digital fluoro
Page 41: Digital fluoro

RADIATION PROTECTION

STAT CH 7 & 8

DHS-RHB

RAD PROT/FLUORO SYLLABUS

Page 42: Digital fluoro

Radiation Safety and Fluoroscopy

Time Distance Shielding

Also see Intro to Fluoro LectureFrom RT 124 Spring 2007

Page 43: Digital fluoro

Patient Protection

• Tabletop exposure rate– Maximum 10 R/min– Typically 1 – 3 R/min

Page 44: Digital fluoro

Patient Protection

• Minimum source-to-skin distance– 12” for mobile equipment– 15” for stationary systems

• Audible alarm at 5 mins.

• Same rules for collimation

Page 45: Digital fluoro

Patient Protection

• Typical exposure rates– Cinefluorography

• 7.2 R/min

– Cassettes• 30 mR/exposure

– 105 mm film• 10 mR/exposure

Page 46: Digital fluoro

Protection of Radiographer and Radiologist

• Lead apron– 0.25 mm Pb/eq

• Highest energy scatter– 90o angle to the incident beam– Same level as radiographer /radiologist’s

gonads

Page 47: Digital fluoro

The Patient & Scatter

Page 48: Digital fluoro

WHATS WRONG?

Page 49: Digital fluoro
Page 50: Digital fluoro

Remote – over the table tube

Page 51: Digital fluoro

Under table tube vs Over table tube (remote units)

Need moving shield for Remote rooms

Page 52: Digital fluoro

IsoExposure CurvesWhere is it SAFE??

Page 53: Digital fluoro

ISOEXPOSURE CURVES

Page 54: Digital fluoro

Patient entrance skin exposure (ESE) is higher when the

fluoroscopic x-ray tube is too close to the tabletop.

Page 55: Digital fluoro

KEEP I.I. CLOSE TO PATIENT

Page 56: Digital fluoro
Page 57: Digital fluoro

PATIENT EXPOSURE

• REDUCE DISTANCE OF IMAGE INTESIFIER

• INCREASE DISTANCE FROM THE TUBE

Page 58: Digital fluoro
Page 59: Digital fluoro

PATIENT EXPOSURE

• REDUCE SIZE OF

COLLIMATED BEAM

WHEN POSSIBLE

Page 60: Digital fluoro

ESE FOR FLUORO

• TLD PLACED AT SKIN ENTRACE POINT

• 1 – 5 R/MINUTE AVE IS 4 R/MIN

• INTERGRAL DOSE –• 100 ERGS OF TISSUE = 1 RAD EXPOSURE• OR 1 GM RAD = 100 ERGS

Page 61: Digital fluoro

1 RAD -100 ERGS – 1 GRAM

1 rad (of ionizing radiation) deposits 100 ergs of energy per gram

Page 62: Digital fluoro

DOSE REGULATIONS

• BEFORE 1974 - AT TABLETOP

• 5R/MIN (WITHOUT AEC)

• 5R/MIN (WITHOUT AEC) – BOOST MODE

• After 1974 with AEC

• 10 R/MIN 20R/MIN BOOST

Page 63: Digital fluoro

SSD – TUBE TO SKIN DISTANCE

• FIXED UNITS

• 18” PREFERRED

• 15 “ MINIMUM

MOBILE UNITS ( C-ARMS)

• 12’ MINIMUM

Page 64: Digital fluoro

Conventional “FIXED” Fluoro

Page 65: Digital fluoro
Page 66: Digital fluoro

Mobile – C- arm fluoro

Page 67: Digital fluoro
Page 68: Digital fluoro
Page 69: Digital fluoro
Page 70: Digital fluoro
Page 71: Digital fluoro
Page 72: Digital fluoro
Page 73: Digital fluoro
Page 74: Digital fluoro

FLUOROSCOPY –

ABOVE AND BELOW THE TABLE TUBE EXPOSURE TO TECH

Page 75: Digital fluoro
Page 76: Digital fluoro
Page 77: Digital fluoro

Pulsed Fluoro

• Some fluoroscopic equipment is designed for pulsed-mode operation. With the pulsed mode, it can be set to produce less than the conventional 25 or 30 images per second. This reduces the exposure rate.

• Collimation of the X ray beam to the smallest practical size and keeping the distance between the patient and image receptor as short as possible contribute to good exposure management.

Page 79: Digital fluoro

Dose rate to patients

Page 80: Digital fluoro
Page 81: Digital fluoro
Page 82: Digital fluoro

Mini c-arm

Page 83: Digital fluoro
Page 84: Digital fluoro
Page 85: Digital fluoro
Page 86: Digital fluoro
Page 87: Digital fluoro