EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491 www.EdgeTech.com DISCOVER BATHYMETRIC USER SOFTWARE MANUAL 0014878_REV_B
Transcript
Page 1: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

EdgeTech

4 Little Brook Road

West Wareham, MA 02576

Tel: (508) 291-0057

Fax: (508) 291-2491

www.EdgeTech.com

DISCOVER BATHYMETRIC U S E R S O F T W A R E M A N U A L

0014878_REV_B

Page 2: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

ii

DISCOVER BATHYMETRIC 0014878_REV_B

The information, figures, and specifications in this manual are proprietary and are issued in strict

confidence on condition that they not be copied, reprinted, or disclosed to a third party, either wholly or

in part, without the prior, written consent of EdgeTech. Any reproduction of EdgeTech supplied software

or file sharing is strictly prohibited.

©Copyright 2014 by EdgeTech. All rights reserved.

Microsoft® and Windows® are registered trademarks of Microsoft Corporation.

HYPACK® and HYSWEEP® are registered trademarks of Hypack, Inc.

Page 3: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

iii

MANUAL PREFACE

We, the employees at EdgeTech, would like to thank you for purchasing the DISCOVER BATHYMETRIC

Acquisition software. At EdgeTech, it is our policy to provide high quality, cost effective products and

support services that meet or exceed your requirements. We also strive to deliver them on time and to

continuously look for ways to improve them. We take pride in the products we manufacture and want

you to be entirely satisfied with your equipment.

Purpose of this Manual

The purpose of this manual is to provide the user with information on the use of EdgeTech’s DISCOVER

BATHYMETRIC Acquisition software. For detailed information relating to hardware that came with this

software, refer to the appropriate user hardware manual.

Although this manual encompasses the latest operational features of DISCOVER, some features may be

periodically upgraded. Therefore the information in this manual is subject to change and should be used

for reference only.

Summary of Sections

This manual is divided into the following sections:

SECTION 1: OVERVIEW: Provides a general introduction to the software.

SECTION 2: DISCOVER TOOLBAR MENU: Explains the functionality of the toolbar.

SECTION 3: DISCOVER CONTROL MENU: Describes the control menu.

SECTION 4: DISCOVER TOP MENU: Discusses how to use the top menu.

SECTION 5: DISCOVER FILE TYPES: Details the file types used by DISCOVER.

SECTION 6: BATHYMETRY DISPLAY: Explains how to use the Bathymetry Displays.

SECTION 7: BASIC CONTROLS: Provides information on the simple controls.

SECTION 8: ADVANCED CONTROLS: Discusses the advanced controls.

SECTION 9: ENGINEERING CONTROLS: Explains the engineering controls for advanced users.

SECTION 10: BINNED DISPLAY: Details the function and use of the binned display.

SECTION 11: BATCH PROCESSING JSF FILES: Discusses Batch Processing of DISCOVER’s JSF files.

APPENDIX A: CONFIGURING THE COM PORTS: Instructs how to complete this procedure.

Page 4: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

iv

DISCOVER BATHYMETRIC 0014878_REV_B

Warnings, Cautions, and Notes

Where applicable, warnings, cautions, and notes are provided in this manual as follows:

WARNING!

Identifies a potential hazard that could cause personal injury or death to

yourself or to others.

CAUTION!

Identifies a potential hazard that could be damaging to equipment or

could result in the loss of data.

NOTE: Recommendations or general information that is particular to

the material being presented. It may also refer to another part of this

manual or to another manual.

Liability

EdgeTech has made every effort to document DISCOVER BATHYMETRIC in this manual accurately and

completely. However, EdgeTech assumes no liability for errors or for any damages that result from the

use of this manual or the software it documents. EdgeTech reserves the right to upgrade features of this

software and to make changes to this manual without notice at any time.

Page 5: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

v

SOFTWARE SERVICE OVERVIEW

EdgeTech provides software services free of charge. The Software Service Agreement (SSA) does not

address customer-specified modifications or enhancements. These services may be ordered separately.

Furthermore, EdgeTech software upgrades are meant for the sole use of EdgeTech customer. Any

reproduction of EdgeTech supplied software or file sharing is strictly prohibited.

Software Updates and Enhancements

EdgeTech customers can download new software releases with all modifications and enhancements along

with user's manual changes from the EdgeTech FTP site. Major software issues, should they occur, will be

reported directly to the customer. New software releases consist of the following:

Software enhancements that are not on the price list

Software fixes and changes

Product integration

Documentation updates to on-line help

Tests for compatibility with other modules

Software patches consist of software that has undergone the following:

Minor software enhancements

Software fixes and changes

Software Telephone, Facsimile, and E-mail Support

The SSA entitles EdgeTech customers to contact EdgeTech Customer Service by telephone, facsimile, or

e-mail to report a difficulty, to discuss a problem or to receive advice on the best way to perform a task.

When contacted, EdgeTech Customer Service will do the following:

Respond within 24 hours

Immediately attend to serious problems affecting operations

Attempt to find an immediate work-around

Page 6: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

vi

DISCOVER BATHYMETRIC 0014878_REV_B

CUSTOMER SERVICE

Customer service personnel at EdgeTech are always eager to hear from users of our products. Your

feedback is welcome, and is a valuable source of information which we use to continually improve these

products. Therefore we encourage you to contact EdgeTech Customer Service to offer any suggestions or

to request technical support:

NOTE: Please have your system Serial Number available when

contacting Customer Service.

E-mail: [email protected]

Mail: 4 Little Brook Road West Wareham, MA 02576

Telephone: (508) 291-0057

Facsimile: (508) 291-2491

24-Hour Emergency Technical Support Line: (508) 942-8043

For more information please go to www.EdgeTech.com.

Page 7: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

vii

COMPANY BACKGROUND

EdgeTech (formerly EG&G Marine Instruments) traces its history in underwater data acquisition and

processing back to 1966. EdgeTech has designed, developed, and manufactured products, instruments,

and systems for the acquisition of underwater data, including marine, estuarine, and coastal applications

for over 45 years.

The company has responded to the needs of the scientific, Naval and offshore communities by providing

equipment, such as sub-bottom profilers, side scan sonar, acoustic releases, USBL positioning systems,

and bathymetric systems that have become standards in the industry.

EdgeTech has also consistently anticipated and responded to future needs through an active research and

development program. Current efforts are focused on the application of cutting edge CHIRP and acoustic

technology.

Page 8: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

viii

DISCOVER BATHYMETRIC 0014878_REV_B

TABLE OF CONTENTS

MANUAL PREFACE ............................................................................................................................ iii

Purpose of this Manual ............................................................................................................................ iii

Summary of Sections ............................................................................................................................... iii

Warnings, Cautions, and Notes ................................................................................................................ iv

Liability ..................................................................................................................................................... iv

SOFTWARE SERVICE OVERVIEW .......................................................................................................... v

Software Updates and Enhancements...................................................................................................... v

CUSTOMER SERVICE .......................................................................................................................... vi

COMPANY BACKGROUND ................................................................................................................ vii

TABLE OF CONTENTS ....................................................................................................................... viii

LIST OF FIGURES ................................................................................................................................ xi

SECTION 1: OVERVIEW ..................................................................................................................1-1

SECTION 2: DISCOVER TOOLBAR MENU .........................................................................................2-1

SECTION 3: DISCOVER CONTROL MENU .........................................................................................3-1

3.1 Sonar Control .............................................................................................................................. 3-1

3.2 Video Gains ................................................................................................................................. 3-1

3.3 Display ......................................................................................................................................... 3-2

3.4 Disk .............................................................................................................................................. 3-2

3.5 Bottom Track ............................................................................................................................... 3-3

3.6 Grids ............................................................................................................................................ 3-4

3.7 Status ........................................................................................................................................... 3-5

SECTION 4: DISCOVER TOP MENU ..................................................................................................4-1

4.1 File Menu ..................................................................................................................................... 4-1

4.2 View Menu .................................................................................................................................. 4-1

4.3 Configuration Menu .................................................................................................................... 4-2

4.3.1 Display Configuration............................................................................................................ 4-2

4.3.2 Record Configuration ............................................................................................................ 4-3

4.3.3 External Features Configuration ........................................................................................... 4-4

4.3.4 Sonar Port Settings ............................................................................................................... 4-4

Page 9: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

ix

4.3.5 System Time Set .................................................................................................................... 4-5

4.3.6 Network Settings .................................................................................................................. 4-6

4.3.7 Frequency Selection.............................................................................................................. 4-6

4.3.8 Alerts ................................................................................................................................... 4-10

4.3.9 Trigger Settings ................................................................................................................... 4-10

4.3.10 Auxiliary Sensors ................................................................................................................. 4-11

4.4 Control Menu ............................................................................................................................ 4-12

4.4.1 Advanced Sonar Controls.................................................................................................... 4-12

4.4.2 Connection Information ..................................................................................................... 4-12

4.4.3 Diagnostic Information ....................................................................................................... 4-13

4.4.4 Sonar Diagnostics ................................................................................................................ 4-13

4.5 Bathymetry Menu ..................................................................................................................... 4-14

4.6 Help Menu ................................................................................................................................. 4-14

SECTION 5: DISCOVER FILE TYPES ...................................................................................................5-1

5.1 Stave File ..................................................................................................................................... 5-1

5.2 Processed File .............................................................................................................................. 5-1

5.3 Binned File ................................................................................................................................... 5-2

SECTION 6: BATHYMETRY DISPLAY ................................................................................................6-1

SECTION 7: BASIC CONTROLS .........................................................................................................7-1

7.1 Limits ........................................................................................................................................... 7-1

7.2 Gating .......................................................................................................................................... 7-3

7.3 Binning ......................................................................................................................................... 7-4

SECTION 8: ADVANCED CONTROLS ................................................................................................8-1

8.1 Filters ........................................................................................................................................... 8-1

8.2 Echo Strength Filter ..................................................................................................................... 8-2

8.3 Water Column Filter .................................................................................................................... 8-2

8.4 SNR Filter ..................................................................................................................................... 8-3

8.5 Quality Filter ................................................................................................................................ 8-3

SECTION 9: ENGINEERING CONTROLS ............................................................................................9-1

9.1 Auto Detect Altitude ................................................................................................................... 9-1

9.2 Example ....................................................................................................................................... 9-2

Page 10: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

x

DISCOVER BATHYMETRIC 0014878_REV_B

9.3 Multipath Suppression Level ....................................................................................................... 9-3

9.4 Maximum Across Track Average Size .......................................................................................... 9-4

9.5 Processing Refinements .............................................................................................................. 9-5

9.6 Time Varying Gain (TVG) ............................................................................................................. 9-5

9.6.1 Default .................................................................................................................................. 9-5

9.6.2 Display ................................................................................................................................... 9-6

9.6.3 Manual .................................................................................................................................. 9-6

SECTION 10: BINNED DISPLAY ........................................................................................................ 10-1

10.1 Seafloor Display ......................................................................................................................... 10-2

SECTION 11: BATCH PROCESSING JSF FILES .................................................................................... 11-1

APPENDIX A: CONFIGURING COM PORTS ...................................................................................... A-1

A.1 One Sensor ................................................................................................................................... A-1

A.2 Two Sensors ................................................................................................................................. A-9

A.3 Three Sensors ............................................................................................................................. A-12

Page 11: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

xi

LIST OF FIGURES

Figure 1-1: Activating the Bathymetry Displays ......................................................................................... 1-1

Figure 1-2: DISCOVER BATHYMETRIC Windows (see note on following page) ......................................... 1-2

Figure 1-3: Bathymetric Processor Module ............................................................................................... 1-3

Figure 2-1: DISCOVER BATHYMETRIC Toolbar Menu ................................................................................. 2-1

Figure 2-2: Display Configuration Dialog Box ............................................................................................. 2-2

Figure 2-3: Display Configuration Dialog Box ............................................................................................. 2-3

Figure 3-1: Sonar Control Tab .................................................................................................................... 3-1

Figure 3-2: Video Gains Tab ....................................................................................................................... 3-2

Figure 3-3: Display Tab ............................................................................................................................... 3-2

Figure 3-4: Disk Tab .................................................................................................................................... 3-3

Figure 3-5: Bottom Track Tab..................................................................................................................... 3-3

Figure 3-6: Grids Tab .................................................................................................................................. 3-4

Figure 3-7: Grids setup to Display Vertical Mark every 10 M .................................................................... 3-4

Figure 3-8: Status Tab ................................................................................................................................ 3-5

Figure 3-9: Supporting Information Display ............................................................................................... 3-5

Figure 4-1: File Menu ................................................................................................................................. 4-1

Figure 4-2: View Menu ............................................................................................................................... 4-1

Figure 4-3: Configuration Menu ................................................................................................................. 4-2

Figure 4-4: Display Configuration............................................................................................................... 4-3

Figure 4-5: Record Configuration ............................................................................................................... 4-3

Figure 4-6: External Features ..................................................................................................................... 4-4

Figure 4-7: Sonar Port Settings .................................................................................................................. 4-5

Figure 4-8: System Time Set Configuration Window ................................................................................. 4-6

Figure 4-9: Network Configuration ............................................................................................................ 4-6

Figure 4-10: Bathymetry Array Selection Dialog Box; Currently Bathymetry on the High Channel is selected

as the current configuration (520 kHz). ..................................................................................................... 4-6

Page 12: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

xii

DISCOVER BATHYMETRIC 0014878_REV_B

Figure 4-11: Array Designed to Accept Bathymetry on the 230 kHz Channel (Blue Dot) .......................... 4-7

Figure 4-12: Array Designed to Accept Bathymetry on the 520 kHz Channel (Red Dot) ........................... 4-7

Figure 4-13: Confirming the Configuration Change ................................................................................... 4-8

Figure 4-14: System Configuration Error in Discover ................................................................................. 4-8

Figure 4-15: Please Restart Discover Screen .............................................................................................. 4-9

Figure 4-16: Bathymetric Results Display .................................................................................................. 4-9

Figure 4-17: Alerts .................................................................................................................................... 4-10

Figure 4-18: Trigger Configuration Window ............................................................................................ 4-10

Figure 4-19: Auxiliary Sensor Messages ................................................................................................... 4-11

Figure 4-20: Control Menu ....................................................................................................................... 4-12

Figure 4-21: Advanced Sonar Controls .................................................................................................... 4-12

Figure 4-22: Connection Info ................................................................................................................... 4-13

Figure 4-23: Diagnostic Window .............................................................................................................. 4-13

Figure 4-24: Sonar Diagnostics ................................................................................................................. 4-14

Figure 4-25: Help Menu ........................................................................................................................... 4-14

Figure 6-1: Bathymetry Display and View Menu ....................................................................................... 6-1

Figure 6-2: Bathymetry Controls ................................................................................................................ 6-2

Figure 7-1: Basic Bathymetric Controls ...................................................................................................... 7-1

Figure 7-2: Specifying Window Size ........................................................................................................... 7-2

Figure 7-3: Specifying Blanking Range ....................................................................................................... 7-2

Figure 7-4: setting the blanking range due to keel interference ............................................................... 7-3

Figure 7-5: Swath to Depth Ratio Filter ..................................................................................................... 7-4

Figure 7-6: Distance Binning Parameters ................................................................................................... 7-5

Figure 7-7: Angle Binning Parameters ....................................................................................................... 7-5

Figure 8-1: Advanced Bathymetric Controls .............................................................................................. 8-1

Figure 8-2: Water Column Filtering Range ................................................................................................. 8-2

Figure 9-1: Engineering Controls................................................................................................................ 9-1

Page 13: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

xiii

Figure 9-2: Auto Detect Altitude Disabled ................................................................................................. 9-2

Figure 9-3: Example Extreme Circumstance - Surveying next to a reef bank ............................................ 9-3

Figure 9-4: Example Extreme Circumstance - Input Approximate Depth .................................................. 9-3

Figure 9-5: Advanced Processing Refinements .......................................................................................... 9-4

Figure 9-6: Processing Refinements - Advanced Controls ......................................................................... 9-5

Figure 9-7: DISCOVER Toolbar Menu for Low Frequency (left) and High Frequency (right) ..................... 9-6

Figure 9-8: Inputting a Manual TVG ........................................................................................................... 9-6

Figure 10-1: Binned Display, surveying next to a channel’s edge using distance binning. ...................... 10-1

Figure 10-2: Binned Display, surveying next to a dock wall using angle binning..................................... 10-1

Figure 10-3: Seafloor Display, surveying next to a channel’s edge using distance binning. .................... 10-2

Figure 10-4: Seafloor Display, surveying next to a dock wall using angle binning. ................................. 10-2

Figure 10-5: Gray Scale ............................................................................................................................ 10-3

Figure 10-6: Rainbow ............................................................................................................................... 10-3

Figure 11-1: DISCOVER BATHYMETRIC Windows .................................................................................... 11-1

Figure 11-2: Disk Tab ................................................................................................................................ 11-2

Figure 11-3: Accessing the Controls ......................................................................................................... 11-2

Figure 11-4: Basic Controls ...................................................................................................................... 11-3

Figure 11-5: Advanced Controls ............................................................................................................... 11-3

Figure 11-6: Opening a File ...................................................................................................................... 11-4

Figure 11-7: File Status ............................................................................................................................. 11-4

Figure 1-1: Activating the Bathymetry Displays ......................................................................................... 1-1

Figure 1-2: DISCOVER BATHYMETRIC Windows (see note on following page) ......................................... 1-2

Figure 1-3: Bathymetric Processor Module ............................................................................................... 1-3

Figure 2-1: DISCOVER BATHYMETRIC Toolbar Menu ................................................................................. 2-1

Figure 2-2: Display Configuration Dialog Box ............................................................................................. 2-2

Figure 2-3: Display Configuration Dialog Box ............................................................................................. 2-3

Figure 3-1: Sonar Control Tab .................................................................................................................... 3-1

Page 14: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

xiv

DISCOVER BATHYMETRIC 0014878_REV_B

Figure 3-2: Video Gains Tab ....................................................................................................................... 3-2

Figure 3-3: Display Tab ............................................................................................................................... 3-2

Figure 3-4: Disk Tab .................................................................................................................................... 3-3

Figure 3-5: Bottom Track Tab..................................................................................................................... 3-3

Figure 3-6: Grids Tab .................................................................................................................................. 3-4

Figure 3-7: Grids setup to Display Vertical Mark every 10 M .................................................................... 3-4

Figure 3-8: Status Tab ................................................................................................................................ 3-5

Figure 3-9: Supporting Information Display ............................................................................................... 3-5

Figure 4-1: File Menu ................................................................................................................................. 4-1

Figure 4-2: View Menu ............................................................................................................................... 4-1

Figure 4-3: Configuration Menu ................................................................................................................. 4-2

Figure 4-4: Display Configuration............................................................................................................... 4-3

Figure 4-5: Record Configuration ............................................................................................................... 4-3

Figure 4-6: External Features ..................................................................................................................... 4-4

Figure 4-7: Sonar Port Settings .................................................................................................................. 4-5

Figure 4-8: System Time Set Configuration Window ................................................................................. 4-6

Figure 4-9: Network Configuration ............................................................................................................ 4-6

Figure 4-10: Bathymetry Array Selection Dialog Box; Currently Bathymetry on the High Channel is selected

as the current configuration (520 kHz). ..................................................................................................... 4-6

Figure 4-11: Array Designed to Accept Bathymetry on the 230 kHz Channel (Blue Dot) .......................... 4-7

Figure 4-12: Array Designed to Accept Bathymetry on the 520 kHz Channel (Red Dot) ........................... 4-7

Figure 4-13: Confirming the Configuration Change ................................................................................... 4-8

Figure 4-14: System Configuration Error in Discover ................................................................................. 4-8

Figure 4-15: Please Restart Discover Screen .............................................................................................. 4-9

Figure 4-16: Bathymetric Results Display .................................................................................................. 4-9

Figure 4-17: Alerts .................................................................................................................................... 4-10

Figure 4-18: Trigger Configuration Window ............................................................................................ 4-10

Figure 4-19: Auxiliary Sensor Messages ................................................................................................... 4-11

Page 15: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

xv

Figure 4-20: Control Menu ....................................................................................................................... 4-12

Figure 4-21: Advanced Sonar Controls .................................................................................................... 4-12

Figure 4-22: Connection Info ................................................................................................................... 4-13

Figure 4-23: Diagnostic Window .............................................................................................................. 4-13

Figure 4-24: Sonar Diagnostics ................................................................................................................. 4-14

Figure 4-25: Help Menu ........................................................................................................................... 4-14

Figure 6-1: Bathymetry Display and View Menu ....................................................................................... 6-1

Figure 6-2: Bathymetry Controls ................................................................................................................ 6-2

Figure 7-1: Basic Bathymetric Controls ...................................................................................................... 7-1

Figure 7-2: Specifying Window Size ........................................................................................................... 7-2

Figure 7-3: Specifying Blanking Range ....................................................................................................... 7-2

Figure 7-4: setting the blanking range due to keel interference ............................................................... 7-3

Figure 7-5: Swath to Depth Ratio Filter ..................................................................................................... 7-4

Figure 7-6: Distance Binning Parameters ................................................................................................... 7-5

Figure 7-7: Angle Binning Parameters ....................................................................................................... 7-5

Figure 8-1: Advanced Bathymetric Controls .............................................................................................. 8-1

Figure 8-2: Water Column Filtering Range ................................................................................................. 8-2

Figure 9-1: Engineering Controls................................................................................................................ 9-1

Figure 9-2: Auto Detect Altitude Disabled ................................................................................................. 9-2

Figure 9-3: Example Extreme Circumstance - Surveying next to a reef bank ............................................ 9-3

Figure 9-4: Example Extreme Circumstance - Input Approximate Depth .................................................. 9-3

Figure 9-5: Advanced Processing Refinements .......................................................................................... 9-4

Figure 9-6: Processing Refinements - Advanced Controls ......................................................................... 9-5

Figure 9-7: DISCOVER Toolbar Menu for Low Frequency (left) and High Frequency (right) ..................... 9-6

Figure 9-8: Inputting a Manual TVG ........................................................................................................... 9-6

Figure 10-1: Binned Display, surveying next to a channel’s edge using distance binning. ...................... 10-1

Figure 10-2: Binned Display, surveying next to a dock wall using angle binning..................................... 10-1

Page 16: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

xvi

DISCOVER BATHYMETRIC 0014878_REV_B

Figure 10-3: Seafloor Display, surveying next to a channel’s edge using distance binning. .................... 10-2

Figure 10-4: Seafloor Display, surveying next to a dock wall using angle binning. ................................. 10-2

Figure 10-5: Gray Scale ............................................................................................................................ 10-3

Figure 10-6: Rainbow ............................................................................................................................... 10-3

Figure 11-1: DISCOVER BATHYMETRIC Windows .................................................................................... 11-1

Figure 11-2: Disk Tab ................................................................................................................................ 11-2

Figure 11-3: Accessing the Controls ......................................................................................................... 11-2

Figure 11-4: Basic Controls ...................................................................................................................... 11-3

Figure 11-5: Advanced Controls ............................................................................................................... 11-3

Figure 11-6: Opening a File ...................................................................................................................... 11-4

Figure 11-7: File Status ............................................................................................................................. 11-4

Figure A-1: Sonar Remote Desktop Application, Main Screen .................................................................. A-2

Figure A-2: Sonar Remote Desktop, Shortcut to EdgeTech Folder ........................................................... A-3

Figure A-3: Finding the SonarSerial.ini File ............................................................................................... A-4

Figure A-4: SonarSerial.ini File .................................................................................................................. A-5

Figure A-5: Diagnostic Window, Reporting Ports Inactive Error ............................................................... A-5

Figure A-6: Rebooting Sonar from the Main Screen ................................................................................. A-6

Figure A-7: Checking Serial Port Information ........................................................................................... A-7

Figure A-8: Checking Sensor Status........................................................................................................... A-8

Figure A-9: Closing the System Status Window ........................................................................................ A-9

Figure A-10: SonarSerial.ini File .............................................................................................................. A-10

Figure A-11: Checking Sensors’ Usages ................................................................................................... A-11

Figure A-12: All Sensors Set .................................................................................................................... A-13

Figure A-13: Checking all COM Ports ...................................................................................................... A-15

Page 17: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

1-1

SECTION 1: OVERVIEW

EdgeTech’s DISCOVER BATHYMETRIC Acquisition Software package provides a way to control, store, and

display bathymetric and dual frequency side scan sonar data. These data are presented on a color

waterfall display and are stored in the binary EdgeTech JSF file format on the hard disk of the Topside

Processor. The JSF file format has been in use for 10+ years, and has new public extensions to support the

additional bathymetry data messages.

This manual describes the DISCOVER BATHYMETRIC Acquisition Software entirely and how to operate it

during a survey or post AUV missions.

DISCOVER BATHYMETRIC has four main windows as illustrated by Figure 1-2 and described below:

1. The Sonar Control window displays the side scan data on a color waterfall and it contains a Top

MenuA, a Toolbar MenuB, and a Control MenuC. These menus are illustrated in Figure 1-2 and

correspond to A, B, and C, respectively. A brief description of the shortcut keys in the DISCOVER

Toolbar Menu is provided in SECTION 4: DISCOVER TOP MENU, the DISCOVER Control Menu in

SECTION 3: DISCOVER CONTROL MENU, and the Top Menu in SECTION 4: DISCOVER TOP MENU.

2. The Bathymetry Display shows the raw interferometric data and filtering parameters.

3. The Binned Display shows the bathymetric data in a pseudo multibeam format as specified in the

binning parameters.

4. The Seafloor Display shows the binned data output in a three dimensional perspective and

provides a color scale by depth.

NOTE: If this is the first time DISCOVER BATHYMETRIC has been used,

these windows can be found under the Bathymetry menu at the top of

DISCOVER as shown below. These three bathymetry displays are

described over the next few sections.

Figure 1-1: Activating the Bathymetry Displays

Displays

Page 18: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

Figure 1-2: DISCOVER BATHYMETRIC Windows (see note on following page)

A - B -

C -

HF Port Display Window HF Stbd Display Window

LF Port Display Window LF Stbd Display Window

DIS

CO

VER

BA

TH

YM

ETR

IC

00

14

87

8_R

EV_B

1

-2

SE

CTI

ON

1:

OV

ERV

IEW

Page 19: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

1-3

NOTE: Figure 1-2 windows shown from left to right: (1) Main Sonar

Control window and Side Scan Waterfall, (2) Bathymetry Display, (3)

Binned Data Display, and (4) Seafloor Display.

DISCOVER BATHYMETRIC also has a built in module that is responsible for computing the seafloor

solutions called the Bathymetric Processor. The Bathymetric Processor displays the number of processed

pings, the last altitude found, and the number of angle sets used in the seafloor calculations.

Figure 1-3: Bathymetric Processor Module

NOTE: The Bathymetric Processor will restart if you press the “x” up at

the top right corner. In doing so, the Bathymetric Processor will delete all

current settings and revert back to its factory default configuration.

Please refrain from pressing the “x” on the Bathymetric Processor

window.

Page 20: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …
Page 21: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

2-1

SECTION 2: DISCOVER TOOLBAR MENU

The DISCOVER Toolbar Menu permits the user to have easy access to certain features in the Top Menu

and Control Menu. The Toolbar Menu (list B in Figure 1-2) is demonstrated in Figure 2-1 and an explanation

for each is provided in this section.

Figure 2-1: DISCOVER BATHYMETRIC Toolbar Menu

i. LF Sonar On – Toggles the system’s low frequency channel On/Off. This feature can also be found

under the DISCOVER Control Menu on the Sonar Control Tab.

ii. LF Normalize Data – Normalizes the low frequency side scan data with respect to the highest

intensity. This feature can also be found under DISCOVER Control Menu on the Video Gains Tab.

iii. LF Current Gain Value – Shows the current gain value applied to the low frequency channel with

respect to the highest intensity. The plus and minus symbols next to the LF Current Gain Value

adjusts the current low frequency channel’s gain by increments of 1 dB per click. This feature can

also be found under the DISCOVER Control Menu on the Video Gains Tab.

iv. LF Auto Set TVG – Automatically sets the low frequency channel’s Time Varying Gain (TVG). This

control compensates for spherical spreading and attenuation losses. This feature can also be

found under the DISCOVER Control Menu on the Video Gains Tab and is called Norm TVG.

v. LF Current TVG Value – Shows the current TVG value applied to the low frequency side scan data.

The plus and minus symbols next to the LF Auto Set TVG adjusts the current TVG applied to the

low frequency side scan data by increments of 1 dB click. This feature can also be found under the

DISCOVER Control Menu on the Video Gains Tab.

vi. LF Toggle Speed Correction – Toggles speed correction on/off for the low frequency side scan

data. This function adjusts the display settings to make the low frequency image look more

proportioned and is also located in the Display Configuration dialog box under View > Display.

i ii iii iv vi vii viii ix x xi xii xiii xiv xv xvi v xvii xviii

Page 22: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

2-2 SECTION 2: DISCOVER TOOLBAR MENU

DISCOVER BATHYMETRIC 0014878_REV_B

Figure 2-2: Display Configuration Dialog Box

vii. Playback – Replays a previously collected bathymetric and side scan file. This feature can also be

accessed under the DISCOVER Control Menu on the Disk Tab.

viii. Pause Playback – Pauses a previously collected side scan file. This feature can also be found under

the DISCOVER Control Menu on the Disk Tab.

ix. Stop Playback – Stops a previously collected side scan file. This feature can also be found under

the DISCOVER Control Menu on the Disk Tab.

x. Fast Forward – Fast forwards a previously collected side scan file. This feature can also be found

under the DISCOVER Control Menu on the Disk Tab.

xi. Mark – Places a mark in the recorded data where something might be of interest (this mark can

be seen during playback also). This feature is only located in the DISCOVER Toolbar Menu.

xii. HF Sonar On – Toggles the system’s high frequency channel On/Off. This feature can also be found

under the DISCOVER Control Menu on the Sonar Control Tab.

xiii. HF Normalize Data – Normalizes the high frequency side scan data with respect to the highest

intensity. This feature can also be found under DISCOVER Control Menu on the Video Gains Tab.

xiv. HF Current Gain Value – Shows the current gain value applied to the high frequency channel with

respect to the highest intensity. The plus and minus symbols next to the HF Current Gain Value

adjusts the current high frequency channel’s gain by increments of 1 dB per click. This feature can

also be found under the DISCOVER Control Menu on the Video Gains Tab.

xv. HF Auto Set TVG – Automatically sets the high frequency channel’s Time Varying Gain (TVG). This

control compensates for spherical spreading and attenuation losses. This feature can also be

found under the DISCOVER Control Menu on the Video Gains Tab and is called Norm TVG.

xvi. HF Current TVG Value – Shows the current TVG value applied to the high frequency side scan data.

The plus and minus symbols next to the HF Auto Set TVG adjusts the current TVG applied to the

high frequency side scan data by increments of 1 dB per click. This feature can also be found under

the DISCOVER Control Menu on the Video Gains Tab.

Page 23: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

2-3

xvii. HF Toggle Speed Correction – Toggles speed correction on/off for the high frequency side scan

data. This function adjusts the display settings to make the high frequency image look more

proportioned and is also located in the Display Configuration dialog box under View > Display.

Figure 2-3: Display Configuration Dialog Box

xviii. Reset Displays – Resets the screen to the default display. This feature is only located in the

DISCOVER Toolbar Menu.

NOTE: The gains are strictly used for viewing purposes and do not affect

the processing of the bathymetry or side scan data.

Page 24: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …
Page 25: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

3-1

SECTION 3: DISCOVER CONTROL MENU

The DISCOVER Control Menu allows the user to turn the sonar on/off, adjust the range and gains, playback

or record a data file, and displays the sonar status. These functions are displayed here and are described

over the next few pages.

3.1 Sonar Control

Navigate to the Sonar Control Tab to turn the High Frequency and/or Low Frequency Sonar On/Off by

checking the Sonar On box or by clicking on the ON button from the Shortcut Menu. The range (m) and

ping rate (%) can also be adjusted from this control menu. The range is specified per side for both

frequencies, and the ping rate (%) slider is only available for the bathymetry frequency. In the example

below, the total swath for the high frequency is 70m, and for the low frequency and bathymetry it is set

to 120 m.

Figure 3-1: Sonar Control Tab

CAUTION!

Never turn on the sonar if it is in air. Please ensure the system is

submerged in water before turning it on. Transducer failure may result.

3.2 Video Gains

The user can adjust the gain levels for both the low and high frequency independently in DISCOVER

BATHYMETRIC to facilitate the viewing of the side scan waterfall. These levels can be regulated one of two

ways:

Video Gains Tab (Figure 3-2)

Shortcut Menu (Figure 2-1, buttons ii. through v. for LF, and xiii. through xvi. for HF)

The user can also adjust the Gain and TVG manually by either typing in the desired value or

increasing/decreasing the value using the arrow buttons.

Page 26: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

3-2 SECTION 3: DISCOVER CONTROL MENU

DISCOVER BATHYMETRIC 0014878_REV_B

Figure 3-2: Video Gains Tab

NOTE: The gains are strictly used for viewing purposes and do not affect

the processing of the bathymetry or side scan data.

3.3 Display

The user can adjust the zoom and color palette by using the Display Tab. These features are used solely

on the customer’s preferences.

Figure 3-3: Display Tab

3.4 Disk

To set up a recording directory, navigate to the Disk Tab and click Browse (red arrow). Usually the

recording directory is located on the E:\Data Disk on the Topside Computer. When ready, press the Record

button on the right hand side of the Disk Tab. The name of the file will appear in the Record File: box.

If the user does not specify a specific file name, DISCOVER BATHYMETRIC will automatically name every

file according to the date and time (UTC) the user hits record along with its appropriate extension. For

example, if the user hits record at 13:18 (UTC) on July 12, 2012 the file name will be recorded as

20120712131822_Processed.jsf.

NOTE: DISCOVER BATHYMETRIC can output several file types: a stave

file, a processed file, and a binned file. Please refer to the next section for

more information.

Page 27: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

3-3

Figure 3-4: Disk Tab

For basic playback, navigate to the Disk Tab (Figure 3-4) and select Browse (green arrow). Locate the

desired JSF file and use the function buttons to play, stop, fast forward, etc. These playback buttons are

also conveniently located on the Shortcut Menu.

NOTE: If a stave file was recorded in DISCOVER BATHYMETRIC while

data was being recorded in the 3rd party acquisition software, the file can

be re-processed for bathymetric estimation using different parameters

post-survey. This recorded DISCOVER BATHYMETRIC file is very useful for

debugging and training purposes, but is not necessary for normal

operations.

3.5 Bottom Track

In the case where the system is purely operated as a dual frequency side scan system, a side scan bottom

tracker is available under the Bottom Track tab (Figure 3-5).

Figure 3-5: Bottom Track Tab

NOTE: If the system is running as a bathymetry system, it is not

recommended to run the side scan bottom tracker. This may cause

confusing results since the side scan bottom tracker is not as accurate as

the bathymetric solutions computer by the Bathymetric Processor.

Page 28: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

3-4 SECTION 3: DISCOVER CONTROL MENU

DISCOVER BATHYMETRIC 0014878_REV_B

3.6 Grids

DISCOVER BATHYMETRIC also has the option to place a grid on the screen. The grid may facilitate

measuring the size of a target or estimating its range from the sonar. The Grids Tab allows the user to

place horizontal or vertical grids over the side scan data and to adjust the size and scale of these grids.

See Figure 3-6 for the Grid Tab and Figure 3-7 for an example of grids in the waterfall display.

Figure 3-6: Grids Tab

Figure 3-7: Grids setup to Display Vertical Mark every 10 M

Page 29: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

3-5

3.7 Status

The Status Tab conveniently displays all of the important information for the system, such as navigation

and sensor data.

Figure 3-8: Status Tab

The information is also conveniently displayed at all times at the bottom of the DISCOVER BATHYMETRIC

Sonar Control window.

Figure 3-9: Supporting Information Display

NOTE: Depth, in this context, relates to the actual depth of the sonar

below the waterline, and would only be reported if the sonar had a depth

sensor input in PSI. For a pole mounted system, this depth is usually known

and is provided to the 3rd party processor as an installation offset,

therefore the Depth field in the status bar is populated with N/A.

Page 30: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …
Page 31: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-1

SECTION 4: DISCOVER TOP MENU

The DISCOVER Top Menu consists of the File, View, Configuration, Control, Bathymetry, and Help sub-

menus. A description for each is provided in this section.

4.1 File Menu

The File menu allows the user to load or save the current DISCOVER BATHYMETRIC default settings and

configurations.

Figure 4-1: File Menu

4.2 View Menu

The View menu provides a way to reset all the active windows, view the tool bar and control dialogs, and

easily access several data displays, such as altitude, depth, pitch and roll, and water depth.

Figure 4-2: View Menu

Page 32: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-2 SECTION 4: DISCOVER TOP MENU

DISCOVER BATHYMETRIC 0014878_REV_B

4.3 Configuration Menu

The Configuration menu provides a way to adjust several settings such as the display, recording

parameters, external features, sonar serial port settings, system time, network, frequency selection (for

field exchangeable configurations only), trigger, alerts, and auxiliary sensors configurations. Each one of

these configurations is described here for completeness.

Figure 4-3: Configuration Menu

4.3.1 Display Configuration

The Display Configuration window allows the user to set up certain display parameters for both high and

low frequency channels such as Slant Range Correction, Speed Correction, Auto Normalize, Auto TVG, and

Subsample Method.

NOTE: the Speed Correction has a shortcut icon located in the DISCOVER

Toolbar Menu .

Page 33: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-3

Figure 4-4: Display Configuration

Also, the Display Configuration allows the user to input several manual parameters such as sound speed,

altitude, and speed. It is not recommended to set a manual altitude, as this will conflict with the estimates

from the Bathymetric Processor. It is imperative, however, to ensure the system is receiving sound

velocity, so if the sonar’s sound velocity sensor should fail, and then the user would use this window to

input a manual sound velocity into the system.

4.3.2 Record Configuration

The Record Configuration window provides the user with a way to specify a maximum file size in

megabytes (MB) and a free space warning in MB. It also allows the user to select whether or not they

would like to record the raw unprocessed stave data and have DISCOVER name the batch files the same

as the raw data files. It is only recommended to log unprocessed stave data for troubleshooting or

customer support related issues and by default it is set to off (unchecked). The File Name Matching

however, is recommended to be set to on (checked).

Figure 4-5: Record Configuration

Page 34: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-4 SECTION 4: DISCOVER TOP MENU

DISCOVER BATHYMETRIC 0014878_REV_B

4.3.3 External Features Configuration

The External Features window presents a way to enable the Target Logger and Coverage Mapper.

Figure 4-6: External Features

Typically, these features are not used with the EdgeTech 6205 Swath Bathymetry and Side Scan Sonar.

4.3.4 Sonar Port Settings

The Sonar Port Settings dialog box (Figure 4-7) allows the user to configure the RS-232 serial ports in the

6205 Sonar Head using the Discover Acquisition Software. Here the user can configure any one of the

three available RS-232 serial ports, while the fourth is a direct connection from the SV to the Sonar Head

and is displayed in this screen for viewing purposes only. To configure the ports from this menu, the user

must enable/disable the port, specify the baud rate, and choose a parser before Discover can recognize

the incoming data strings.

These serial ports are provided to intake the attitude (roll, pitch, and heave), heading and navigation data

from the supporting sensors. Attitude and heading data must be connected to Ports 1 and 2, since they

are configured for high speed and high accuracy (10Mbps). Port 3 is configured for standard RS-232 serial

connections and can be used to intake the positioning data if three serial ports are required. Most

common setups use only Ports 1 and 2 to ingest all the necessary information while the third is provided

as a spare.

For more information about the sonar serial ports and how to configure them depending on your setup,

please refer to Section 3.2.2 in the EdgeTech 6205 Hardware Manual.

Page 35: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-5

Figure 4-7: Sonar Port Settings

NOTE: This dialog box is only available in Discover Bathymetric versions

34.0.1.110 or higher. To configure the Sonar Port Settings for older

versions of Discover Bathymetric, please refer to APPENDIX A:

4.3.5 System Time Set

The System Time Set window enables/disables DISCOVER BATHYMETRIC to set the topside’s computer

time. If the system is being operated primarily as a bathymetry system, then this box should ALWAYS be

checked. If this parameter is disabled, then there is a potential for mismatched time stamps within the

data.

Page 36: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-6 SECTION 4: DISCOVER TOP MENU

DISCOVER BATHYMETRIC 0014878_REV_B

Figure 4-8: System Time Set Configuration Window

4.3.6 Network Settings

The Network Settings window provides the user with a way to configure their Sonar IP Address and Port

Number. DO NOT change this parameter unless a corresponding change has been made to the sonar

subsystem. This should only be done in truly exceptional circumstances, by expert users, or system

administrators.

Figure 4-9: Network Configuration

4.3.7 Frequency Selection

The Frequency Selection option (Figure 4-10) is only available for field exchangeable systems, which is

available on the EdgeTech 6205 230 kHz/520 kHz models only. It provides a way for the user to switch the

bathymetry from the low frequency channel to the high frequency channel and vice versa. From this dialog

box the user has the capability to choose the receiving frequency of the bathymetric data as long as the

hardware matches the configuration.

Figure 4-10: Bathymetry Array Selection Dialog Box; Currently Bathymetry on the High Channel is

selected as the current configuration (520 kHz).

Page 37: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-7

To accept bathymetry on the 230 kHz channel (low frequency) the arrays with a blue insert must be used,

and to accept bathymetry on the 520 kHz channel (high frequency) then the arrays with a red insert must

be used (as shown in Figure 4-11 and Figure 4-12).

Figure 4-11: Array Designed to Accept Bathymetry on the 230 kHz Channel (Blue Dot)

Figure 4-12: Array Designed to Accept Bathymetry on the 520 kHz Channel (Red Dot)

Page 38: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-8 SECTION 4: DISCOVER TOP MENU

DISCOVER BATHYMETRIC 0014878_REV_B

Once the user makes a selection from the drop down list the software requires confirmation on the chosen

configuration and provides a reminder that the hardware must match the configuration chosen. This

secondary dialog box is shown in Figure 4-13.

Figure 4-13: Confirming the Configuration Change

Once the user confirms to Switch Sonar Arrays the sonar will restart with the new configuration. Please

allow 10-15 seconds for the Sonar to reboot. Once Discover reconnects to the Sonar it will recognize the

different configuration and will require a reboot as well. Here the user is recommended to Switch System

Types so that Discover will exit and the user must restart using the new configurations. Please allow for

3-5 seconds for the Discover software to exit before rebooting the application.

Figure 4-14: System Configuration Error in Discover

Page 39: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-9

Figure 4-15: Please Restart Discover Screen

NOTE: This dialog box is only available in Discover Bathymetric versions

34.0.1.110 or higher.

CAUTION!

If the incorrect hardware (i.e. bathymetry array) is used with the improper

frequency selection specified in DISCOVER, incorrect bathymetry will be

observed and/or acquired. This incompatibility will appear in the

Bathymetric Results display as illustrated in Figure 4-16. Please pay

special attention to the frequency selection when exchanging arrays and

/ or configuring the software.

Figure 4-16: Bathymetric Results Display

Page 40: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-10 SECTION 4: DISCOVER TOP MENU

DISCOVER BATHYMETRIC 0014878_REV_B

4.3.8 Alerts

The Alerts settings, when checked, will sound out alarms for Altitude, Depth, Roll, and Pitch. If it is desired

to use these alerts, please ensure your topside computer can provide sound and has the volume set to a

reasonable value.

Figure 4-17: Alerts

4.3.9 Trigger Settings

The Trigger Configuration window allows the user to set the master trigger to either the low or high

frequency side scan channel. By default, this trigger is set to the bathymetry frequency.

Figure 4-18: Trigger Configuration Window

NOTE: It is not recommended to change the master trigger to anything

other than the bathymetry system, unless it is in a truly exceptional

circumstance. Setting this parameter to anything other than the

bathymetry sub system will result in decreased data density along track.

This should only be changed by expert users or system administrators.

Page 41: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-11

4.3.10 Auxiliary Sensors

The Auxiliary Sensors menu allows the user to configure the serial port messages. Please ensure there are

no redundant messages applied to the system. This will cause the system to bounce back and forth

between the redundant messages and may cause time stamping issues within the data.

Figure 4-19: Auxiliary Sensor Messages

Page 42: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-12 SECTION 4: DISCOVER TOP MENU

DISCOVER BATHYMETRIC 0014878_REV_B

4.4 Control Menu

The Control Menu provides a way for the user to adjust the advanced sonar controls, show the connection

and diagnostic information, and access the sonar diagnostics. The control menu is shown in Figure 4-20.

Figure 4-20: Control Menu

4.4.1 Advanced Sonar Controls

The Advances Sonar Controls dialog box is used to turn the sonar’s transmitters on/off and view or

override the pulses for both high and low frequency channels. DO NOT change these parameters unless a

corresponding change has been made to the sonar subsystem. This should only be done in truly

exceptional circumstances, by expert users, or system administrators.

Figure 4-21: Advanced Sonar Controls

4.4.2 Connection Information

Under Connection Information, the user will find the Sonar Control, Sonar Data, Hypack Navigation

Output, and Hypack RTK Output connections status window (Error! Reference source not found.). This

indow displays the connection status between DISCOVER BATHYMETRIC and the installed External

Topside Interface. To turn off this window, simply click Done.

Page 43: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-13

Figure 4-22: Connection Info

4.4.3 Diagnostic Information

The Diagnostic Window provides valuable information about the sonar system. In the example presented

in Error! Reference source not found., the Sonar Head is on and connected to DISCOVER BATHYMETRIC

hile Ports 1 and 2 have been disconnected and Port 4 is connected and receiving data (active).

Figure 4-23: Diagnostic Window

4.4.4 Sonar Diagnostics

The Sonar Diagnostics window allows the user to perform certain diagnostic checks on the sonar. This

should be done by expert users or online with customer service. These diagnostics are usually not part of

an everyday survey check.

Sonar Head is ON and

CONNECTED

SVP is ON and

CONNECTED

GPS/IMU is DISCONNECTED

IMU/GPS is DISCONNECTED

Page 44: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

4-14 SECTION 4: DISCOVER TOP MENU

DISCOVER BATHYMETRIC 0014878_REV_B

Figure 4-24: Sonar Diagnostics

4.5 Bathymetry Menu

The Bathymetry Menu allows the user to access the bathymetry display and control windows. Please see

the next few sections on how to use these.

4.6 Help Menu

The Help Menu shows the information pertaining to what version of the software you have installed on

your Topside Processor.

Figure 4-25: Help Menu

Page 45: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

5-1

SECTION 5: DISCOVER FILE TYPES

This section describes each of the three different output files DISCOVER BATHYMETRIC can produce. These

file types include a stave, a processed, and a binned file.

NOTE: All file types include all of the supporting information necessary

to correct the bathymetry (time, position, heading, attitude, sound

velocity, and if available, GPS Height).

5.1 Stave File

A stave file contains all of the raw stave, or individual receiver, data that is used to compute the solutions

of the seafloor. This file type does not contain bathymetry and is designated by the keyword “_Stave.”

This keyword is appended to the end of every stave file. An example file name used to designate a stave

file is 20130829150420_Stave.jsf.

NOTE: For legacy systems, a stave file may be designate by the word

“Stave” without the underscore. In the AUV case, all files downloaded

from the vehicle are all stave files and must be post-processed with

DISCOVER BATHYMETRIC before any other work can begin, unless

otherwise noted.

5.2 Processed File

A processed file contains bathymetry in the traditional interferometry format and contains up to 6000

points per ping per side (port/starboard). In this form, the data has to be decimated or thinned in order

to plot the results. A processed file can be identified by the keyword “_Processed”. This keyword is

appended to the end of every processed file. An example file name used to designate a processed file is

20130829150420_Processed.jsf.

NOTE: For legacy systems, a processed file may be identified by the file

name only, with no extensions (i.e. 20130829150420.jsf).

Page 46: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

5-2 SECTION 5: DISCOVER FILE TYPES

DISCOVER BATHYMETRIC 0014878_REV_B

5.3 Binned File

A binned file contains bathymetry in which the interferometric data has been binned (or decimated) based

on a user defined number of beams (or bins) and total swath. This format closely resembles the output

format of a traditional dual headed multibeam.

A binned file can be identified by the keyword “_Binned” and is the standard output format of DISCOVER

BATHYMETRIC. This keyword is appended to the end of every binned file. An example file name used to

designate a binned file is 20130829150420_Binned.jsf.

Page 47: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

6-1

SECTION 6: BATHYMETRY DISPLAY

The Bathymetry Display, shown Figure 6-1, displays the raw interferometric data, filtering gates,

processed pings, detected altitudes, and angle sets. This display also uses colors to help the user

understand the effects of the processing parameters applied to the data. These colors are described here

for completeness.

NOTE: This window is typically used for troubleshooting or customer

support related issues and is not part of the standard survey operations.

Green shows the unfiltered points deemed as good (passed the

filtering criteria). These points are passed on to the final data set.

White illustrates the filtered points deemed as bad (failed the filtering

criteria). These points are not used to produce the final data set.

Yellow demonstrates the points that failed a specific filter of interest

and can be displayed by activating it from the View menu. This can

also be described as the highlighted filter and only one can be shown

at a time. The example presented in Figure 6-1 shows that the Echo

Strength Filter is the current highlighted filter.

Purple shows the points filtered out by the defined horizontal and

angle gates for both port and starboard heads and by the swath to

depth ratio filter. The gates and blanking range have also been color

coded as purple dashed lines.

Figure 6-1: Bathymetry Display and View Menu

Page 48: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

6-2 Error! Reference source not found.Error!

ference source not found.

DISCOVER BATHYMETRIC 0014878_REV_B

NOTE: The number of processed pings, detected altitudes, and angle

sets are mostly used for troubleshooting purposes and must be considered

during real time acquisition if there are no solutions in the Profile and

Seafloor windows. If these numbers are NOT increasing, it is a clear

indication that either Bathymetric Processing is not turned on, filtering

parameters are set too high, or there’s a problem with the built in bottom

tracker.

Now, the user can specify these processing parameters by using the Basic or Advanced Bathymetric

Controls. These control dialog boxes are described in the following sections and only one of these can be

displayed at a time.

Notice the Advanced Bathymetric Controls include all of the basic controls with the added filter section at

the bottom. For most cases, the basic controls will suffice, as these filters have been optimized.

To access the Bathymetric Controls, select Bathymetry from the Top Menu and then one of the three

options displayed below.

Figure 6-2: Bathymetry Controls

A description for each control menu has been provided in the sections that follow. These include Basic

Controls, Advanced Controls, and Engineering Controls.

Controls

Page 49: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

7-1

SECTION 7: BASIC CONTROLS

The Basic Bathymetric Controls are shown in Figure 7-1 and provide the user with the ability to turn

Bathymetric Processing on/off, use a third party’s external limits, use a set of user defined limits and gates,

and choose a binning option. Each one of these options has been defined in detail within this section.

Figure 7-1: Basic Bathymetric Controls

The Bathymetric Processing check box is used to turn on or off the Bathymetric Processor Module

responsible for computing the seafloor estimates. This setting should always be checked during

acquisition or batch processing, and unchecked during playback of a processed or binned file.

The Use External Limits check box tells the DISCOVER BATHYMETRIC to use or not use a set of limits

defined by a third party acquisition interface such as Hypack, QINSy, Chesapeake, etc. It is highly

recommended that this box is checked if any third party interfaces are used for acquisition. Please

remember, these parameters are based on the operating range and depth of the sonar.

7.1 Limits

The Limits section provides a way to specify the minimum and maximum processing depths, blanking

range, and maximum swath. The minimum and maximum depth, along with the maximum swath, defines

the Bathymetry Display window sizes as shown by Figure 7-2.

The blanking range is used to filter out excess noise around the Sonar Head. It is defined in meters and is

illustrated in Figure 7-3 using a purple circle. Here, the range has been specified as 5m. Typically this

parameter is not adjusted unless the user is surveying within constant deep water (greater than 50 m). If

this is the case, the user may adjust this value to a much larger slant range (i.e. 10 – 25 m).

Page 50: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

7-2 SECTION 7: BASIC CONTROLS

DISCOVER BATHYMETRIC 0014878_REV_B

Figure 7-2: Specifying Window Size

Figure 7-3: Specifying Blanking Range

The blanking range should be adjusted if the user is operating with a side mount since there can be

potential issues with keel interference. The blanking range should be set to a value slightly greater than

the distance from the sonar to the keel. In the example below, the blanking range would be adjusted to

5m to ensure the hull echoes do not interfere with the automatic power control.

Min Depth

Max Depth

Max Swath

Blanking

Range = 5m

Page 51: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

7-3

Figure 7-4: setting the blanking range due to keel interference

All of these limits can be set manually by typing into the field box or adjusting the slider bars next to their

corresponding field.

7.2 Gating

The Gating section provides a way for the user to filter out excess noise on the outer ends of the swath.

These are active during real time acquisition and batch processing and have immediate effect. These gates

can be controlled by specifying them numerically within the Basic and Advanced Bathymetric Controls or

interactively setting these on the Bathymetry Display screen.

The interactive controls consist of a click and drag motion, or double left click to set the horizontal gates

in the desired location, or double right click to set the angle gates in the desired location. Again, these

take immediate effect in the processing.

The Gating section also allows the user to use a swath to depth ratio filter. This filter is shown in the

Bathymetry Display by purple tick marks at the top of the range scale as shown in Figure 7-5. The points

that are filtered out by the max swath to depth ratio filter are also designated by the color purple.

Page 52: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

7-4 SECTION 7: BASIC CONTROLS

DISCOVER BATHYMETRIC 0014878_REV_B

Figure 7-5: Swath to Depth Ratio Filter

7.3 Binning

The basic controls also provide a way for the user to decimate the interferometric data down into a more

common multibeam like format (or pseudo multibeam). The possible options are No Binning, Distance

Binning, and Angle Binning. If no binning is chosen, then what is displayed in green in the Bathymetry

Display (raw range/angle data) will be logged to a file with its appropriate extension (i.e.

filename_Processed.jsf). If either distance binning or angle binning is selected, then what is displayed in

the Binning and Seafloor Displays is what gets logged to the file with its appropriate extension (i.e.

filename_Binned.jsf).

NOTE: The dual frequency side scan data is always logged to the

bathymetry file as long as its sub system is on and pinging.

The Distance Binning method is the default mode and breaks up the seafloor into equi-distant boxes (or

bins), so that the number of points is uniform across the entire swath. It is defined by specifying the total

number of bins, bin size in meters, and desired total swath in meters.

Page 53: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

7-5

Figure 7-6: Distance Binning Parameters

The user only has to populate two of the three fields and the third is automatically calculated. The user

should aim to achieve a 350-500 bins. The bin size should be based upon water depth, thereby the deeper

the water, the larger the bin size. The default values are shown in Figure 7-6 and are set to 400 bins, 0.35m

and 140m swath.

The Angle Binning method breaks up the sonar’s field of view into equi-angular sectors (or beams/bins)

and is primarily used to map vertical objects such as dock or sea walls, pilings, etc. This binning option is

defined by specifying the total number of bins and bin size in degrees.

Figure 7-7: Angle Binning Parameters

The user must only populate one of the fields since the total field of view of the sonar is hard coded to

200°. The user should aim to achieve between 350-500 samples and adjust the bin size based on water

depth; the deeper the water, the larger the bin size.

Page 54: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …
Page 55: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

8-1

SECTION 8: ADVANCED CONTROLS

The Advanced Bathymetric Controls allows the user to access and tweak the built in filters for the system.

Please note this control window includes all of the basic controls, along with the added filter section at

the bottom. Since this is the case, only the Filters section is described here. For all others, please refer to

the Basic Controls section.

Figure 8-1: Advanced Bathymetric Controls

8.1 Filters

The Filters section allows the user to access and tweak several system filters. These include the Echo

Strength, Water Column, SNR, and Quality filters. A description of each filter has been provided below.

NOTE: If No Binning is selected, a filter output control box appears down

in the right hand corner of the dialog. By default, these are trimmed (if

checked) from DISCOVER’s output. If a third party interface was

connected it would not receive any of these bad points, unless these boxes

were unchecked.

Page 56: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

8-2 SECTION 8: ADVANCED CONTROLS

DISCOVER BATHYMETRIC 0014878_REV_B

8.2 Echo Strength Filter

The Echo Strength Filter is a fundamental attribute that is used primarily to exclude or trim weak echo

points from the bathymetric data (i.e. water column and very weak backscatter amplitudes). The data

points have an amplitude that can span 0 to 32,767. Typical sea floor echoes are above 300 to 1800,

depending on the bottom type.

The Echo Strength filter by default is set to automatic, whereby the system measures the intensity of the

echoes returned from the sea floor +/- 30 degrees from Nadir and compares it to the intensity of the water

column noise and weaker echoes surrounding the sea floor estimates. If they differ significantly, then it

flags these points as bad. The automatic Echo Strength filter has a sensitivity slider to adjust how

aggressive or how conservative the filter will be. A default value of 1.5 works well under most conditions.

In the case where the automatic setting fails (too many points at nadir and surrounding region are

trimmed), then the user has the ability to turn off the automatic setting (uncheck the box) and set the

parameter manually. The values that can be manually inputted to the system are based on 1/10% of this

span, meaning that an echo strength of 3200 is realistically 320/32767 x 100% = 9.80, and so the value

that would be inputted to the system would equal to 10.

8.3 Water Column Filter

The Water Column filter only has an on/off option and by default is active (checked box). This filter uses

the bottom tracker to find the depth to the seafloor and then subtracts an offset that is pre-determined

by depth to set the filtering range. This offset is smaller in deeper water, than it is for shallow.

This filter can potentially be an issue when surveying close to a steep bank, as it may filter out part of the

rising seafloor as shown in Figure 8-2.

Figure 8-2: Water Column Filtering Range

Water Column

Filtering Range

Page 57: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

8-3

If this is the case, it is suggested to either turn off the filter by unchecking the box, or deviating away from

the bank until this effect is reduced.

8.4 SNR Filter

The SNR (or Signal to Noise Ratio) filter is a valuable statistic used to trim bad data points near the outer

swath, where the angle noise is high due to multipath effects. SNR values greater than 20 dB are excellent

in terms of angle estimation noise, and less than 5 dB is very poor. Useful thresholds are between 10 – 20

dB, depending on the desire to have maximum swath (more noise) or lower noise and less swath. Typically

this filter is set between 12 and 15 dB.

8.5 Quality Filter

In practice, measured interstave phases may either agree quite well (+/- 5° to 10°) or not at all (up to +/-

90°). The quality filter metric is a measure of how good the interstave measurements are and is defined

as 100% = phase differences equal to 0 (i.e., not possible in practice; setting the quality factor to 100% will

exclude all data points).

80% = has a +/- 20° tolerance

70% = has a +/- 30° tolerance

60% = has a +/- 40° tolerance

50% = has a +/- 50° tolerance

25% = has a +/- 90° tolerance

0% = allows all data points to pass through.

For most cases any data with a quality factor less than 50% should be discarded. By default, this setting is

set to 65%.

Page 58: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …
Page 59: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

9-1

SECTION 9: ENGINEERING CONTROLS

The Engineering Controls allows the user to disable/enable Auto Detect Altitude, adjust the Multipath

Suppression Levels and the Maximum Across Track Average Size, and refine the Processing Parameters.

Figure 9-1: Engineering Controls

NOTE: The parameters under the Engineering Controls must not be

changed unless it is directed by a member of the EdgeTech customer

support staff or expert user.

9.1 Auto Detect Altitude

The Auto Detect Altitude function tells the Bathymetric Processor to use its advanced algorithms to track

the bottom. By default, the Auto Detect Altitude is on and should not be turned off, unless the bottom

tracker fails repeatedly. If the Auto Detect Altitude box is unchecked, then the user has the opportunity

to input a manual value in which they think the bottom resides (see Figure 9-2).

Page 60: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

9-2 SECTION 9: ENGINEERING CONTROLS

DISCOVER BATHYMETRIC 0014878_REV_B

Figure 9-2: Auto Detect Altitude Disabled

9.2 Example

See Figure 9-3. This is a complex environment for the Bathymetric Processor whereby the user is

surveying next to a large coral reef. The port array sees the reef bank and the starboard array looks out

to an open ocean. The seafloor on the starboard side is also too deep for the sensor and so no acoustic

data is received (hence the similar resemblance to water column data).

In such a case, the Bathymetric Processor would consistently fail to determine the altitude and the user

would disable the Auto Detect Altitude. Then the user would determine the range to the reef bank using

the side scan waterfall and input this estimate into the Manual Altitude box (Figure 9-4). For the example

below, the user would input 32 m (Figure 9-4). The Bathymetric Processor will then start looking for the

bottom around 32m until it locks onto the reef’s edge.

The Bathymetric Processor would then use this override to plot the seafloor correctly. The user would

then continue to monitor the Bathymetric Processor to ensure it does not fail again. If another failure

occurs, the user would adjust the Manual Altitude as needed.

Page 61: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

9-3

Figure 9-3: Example Extreme Circumstance - Surveying next to a reef bank

Figure 9-4: Example Extreme Circumstance - Input Approximate Depth

9.3 Multipath Suppression Level

The multipath suppression level is a parameter that is used to specify how the beam forming is calculated.

The data from the 10 half wavelength spaced elements can be combined prior to phase estimation

processing to combat the effects of multipath signal propagation.

Seafloor echoes arriving at the array from single or multiple bounces off the seafloor and the sea surface

are the primary cause of poor angle estimation and across track range limitations in a shallow water

deployed system.

Page 62: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

9-4 SECTION 9: ENGINEERING CONTROLS

DISCOVER BATHYMETRIC 0014878_REV_B

In deep water deployments, as may be seen on ROV or AUV mounted systems, this effect is much reduced,

if not eliminated. This parameter can take values between 1 and 3 and increasing this value combines

more of the elements to combat the effects of multipath echoes.

If this is set to 1, then no multipath suppression is applied and all 10 staves are used in the angle

estimations. In the absence of multipath errors, this will result in good data.

If this is set to 2, then adjacent element pairs are combined to help reject weak multipath echoes,

resulting in 9 synthesized half wavelength spaced elements.

If this is set to 3, then 3 adjacent elements are combined to help reject strong multipath echoes,

resulting in 8 synthesized half wavelength spaced elements.

In normal shallow water deployed systems, using a value of 3 gives the best result. This parameter is

adjusted in Figure 9-5. It is not recommended to change this parameter unless it is directed by a member

of the EdgeTech customer support staff or expert user.

Figure 9-5: Advanced Processing Refinements

9.4 Maximum Across Track Average Size

In estimating the angles across track, multiple adjacent time samples may be combined to improve the

signal to noise ratio prior to phase estimation. This is also a form of spatial averaging, which can be applied

to the data prior to the angle estimation processing. This parameter determines the maximum across

track region size that will be used in forming these averages, and is specified in meters (Figure 9-5).

For output data with very high across track resolution, this value should be kept small, for example 0.05

to 0.5m (5 to 50 cm). For less resolution and more averaging, region sizes up to 1m can be used to produce

reasonable effects in some cases (deep water).

Page 63: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

9-5

9.5 Processing Refinements

These parameters are explained in detail in order of appearance:

Figure 9-6: Processing Refinements - Advanced Controls

The default setting for the Bottom Track Data Source is Product. This means the Bathymetric Processor

takes the product of the information coming from both arrays to estimate the seafloor depth. This setting

seems to work well in most scenarios, especially for heavy boat traffic areas where boat wake is

prominent, but two other parameters have been added for those special cases a user may face. These

three other settings are Port and Starboard.

The Port and Starboard settings state that the Bathymetric Processor should only use the port or

starboard side only to compute the seafloor depth. A case where Port or Starboard may be used is if one

of the arrays should fail during a survey (highly unlikely), or the seafloor is beyond the limits of the sonar

(as discussed in Figure 9-3). This way the user will still be able to carry out a survey, even though one side

of the system does not work, or does not see the seafloor.

9.6 Time Varying Gain (TVG)

The TVG option allows the user to apply Time Varying Gain in dB per 100m to the bathymetry solutions.

EdgeTech has incorporated this feature to increase the bathymetric range in soft bottom conditions.

This parameter can have one of three options as described below:

9.6.1 Default

The Default option adjusts the TVG according to the operating frequency and requires no input from the

user.

Page 64: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

9-6 SECTION 9: ENGINEERING CONTROLS

DISCOVER BATHYMETRIC 0014878_REV_B

9.6.2 Display

The Display option applies the TVG set in the display settings as shown in Figure 9-7. This option requires

constant monitoring of the side scan imagery and the TVG must be normalized as the composition of the

seafloor changes. This can be done by clicking on the T button for the bathymetry frequency, or it can be

manually adjusted by using the + or – buttons (see Figure 9-7).

Figure 9-7: DISCOVER Toolbar Menu for Low Frequency (left) and High Frequency (right)

9.6.3 Manual

The Manual option allows the user to input a manual TVG value. This number is specified in the box to the

right of the drop down menu. When this option is selected, the user can also specify a maximum TVG

value in the Max TVG (dB) field (see Figure 9-8). This option also requires constant monitoring of the side

scan imagery so that the correct TVG value is applied to the bathymetric solutions.

Figure 9-8: Inputting a Manual TVG

NOTE: If TVG mode is activated, then the Echo Strength Filter in the

Advanced Bathymetry Controls can be increased (if auto is not applied) to

reduce noise at the extremities of the bathymetry solutions. Monitoring

the results in the Bathymetry Display will help the user determine what

TVG value is best for the current conditions.

Page 65: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

10-1

SECTION 10: BINNED DISPLAY

The Binned Display shows the across track profile after decimation; this decimation is calculated either

by breaking up the swath into equi-distant horizontal bins or equi-angular sectors. For Distance Binning

(edge of a channel shown in Figure 10-1), the across track profile should have uniformly spaced points

across the swath. Sometimes there may be missing points within a bin due to noise or shadows (as shown

by the blue arrow in Figure 10-1).

Figure 10-1: Binned Display, surveying next to a channel’s edge using distance binning.

Angle binning will produce higher density data at nadir and decrease as the swath increases. Angle binning

is primarily used to map vertical objects like a dock wall as shown in Figure 10-2.

Figure 10-2: Binned Display, surveying next to a dock wall using angle binning.

Page 66: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

10-2 SECTION 10: BINNED DISPLAY

DISCOVER BATHYMETRIC 0014878_REV_B

10.1 Seafloor Display

The Seafloor Display plots each decimated ping sequentially onto the screen to produce a three

dimensional view of the seafloor. The color bar is adjusted according to the Minimum and Maximum

Depth settings under the Basic and Advanced Controls. Figure 10-3 and Figure 10-4 illustrate the results

from a seafloor binned by distance and a seafloor binned by angle, respectively.

Figure 10-3: Seafloor Display, surveying next to a channel’s edge using distance binning.

Figure 10-4: Seafloor Display, surveying next to a dock wall using angle binning.

Page 67: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

10-3

The Seafloor Display also offers three color palettes for easier viewing: Rainbow, Gray, and Spectrum.

These can be accessed using the View menu at the top of the display. The Spectrum color scale is

presented in Figure 10-3 and Figure 10-4, while Gray and Rainbow are shown in Figure 10-5 and Figure

10-6. The View menu also allows the user to clear or reset the display.

Figure 10-5: Gray Scale

Figure 10-6: Rainbow

Page 68: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …
Page 69: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

11-1

SECTION 11: BATCH PROCESSING JSF FILES

To Batch Process JSF files within DISCOVER:

1. Launch DISCOVER BATHYMETRIC. The following dialog boxes will appear:

Figure 11-1: DISCOVER BATHYMETRIC Windows

NOTE: If this is the first time DISCOVER BATHYMETRIC has been used,

these windows can be found under the Bathymetry menu at the top of

DISCOVER as shown below.

Page 70: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

11-2 SECTION 11: BATCH PROCESSING JSF FILES

DISCOVER BATHYMETRIC 0014878_REV_B

2. Navigate to the Disk tab and set up a recording directory by selecting Browse next to the Record

File box and navigating to the desired folder.

Figure 11-2: Disk Tab

3. Select the first file of the batch you would like to process and use the Basic/Advanced Controls to

set up your processing parameters. These controls may be accessed by selecting the Bathymetry

menu from the top of DISCOVER.

Figure 11-3: Accessing the Controls

These control dialog boxes are illustrated below and only one of these can be displayed at a time.

Notice the Advanced Bathymetric Controls include all of the Basic Controls with the added filter

section at the bottom. For most cases, the basic controls will suffice for batch processing as these

filters have been optimized.

Displays

Controls

Record Functions

Page 71: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

11-3

NOTES: See sections titled Basic and Advanced Controls for more

information.

Figure 11-4: Basic Controls

Figure 11-5: Advanced Controls

4. Once you have set up your controls, select all the files you would like to process, either by holding

down the Ctrl button and selecting multiple individual files, or highlighting the entire data set.

Then click on Open.

Page 72: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

11-4 SECTION 11: BATCH PROCESSING JSF FILES

DISCOVER BATHYMETRIC 0014878_REV_B

Figure 11-6: Opening a File

5. Press Play and hit Record (big red button). Processed files will now be logged to the directory that

was set up previously.

The batch processing speed may be increased to a speed of 15-30 by clicking on the Fast Forward

button. The maximum speed is governed by the processing capabilities of the PC running

DISCOVER.

DISCOVER also shows its progress by stating which file in the batch it is processing and how much

of that file has been processed. This can be seen on the Disk tab, next to the Playback controls.

Figure 11-7: File Status

6. Once all the files have been processed, hit Stop to close and save all the files.

Page 73: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

11-5

NOTE: DISCOVER can produce three types of files. A raw stave file, a

processed file, and a pseudo multibeam binned file. See section titled

DISCOVER File Types for more information.

For more information about the EdgeTech’s DISCOVER BATHYMETRIC Acquisition software, contact

EdgeTech Customer Service.

Page 74: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …
Page 75: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-1

A P P E N D I X A : C O N F I G U R I N G C O M P O R T S

There are 3 available COM ports on the Interface Box to provide the necessary information to the 6205.

This information includes position (latitude/longitude), heading, attitude (roll, pitch, and heave), and time.

This section explains how to configure the three COM ports depending on what type of sensors are used.

A.1 One Sensor

One sensor, such as the Applanix PosMV, may be used to supply position (latitude/longitude), heading,

attitude, and a time source. For this type of configuration, only COM1 will be utilized.

To configure the COM ports for one sensor:

1. Turn on the Topside PC and connect the 6205 via the deck cable to the 6205 Topside Interface

Box.

2. Turn on the 6205 Sonar Head and run the Remote Desktop Application, named Sonar. The main

screen as in Figure A-1 should appear.

Here, the user will see the Sonar Application running (designated by the blue arrow). This is the

firmware that controls the system.

3. Click on the 'X' at the top right to close the firmware.

4. Now, double click on the Shortcut to EdgeTech icon (red arrow, Figure A-2) and open the

SSSonarBathy folder, then the Sonar folder.

5. Look for the SonarSerial.ini file (Figure A-3) and open it using WordPad or Notepad (Figure A-4).

6. Under [Serial2] and [Serial3] set Create = 0, select File > Save, and close the file.

CAUTION!

Do not alter the file in any way other than what has been specified.

This will deactivate COM ports 2 and 3. By deactivating these ports, Sonar will report to DISCOVER

that only one sensor is to be used on COM1 to parse the necessary sensor information. Therefore,

DISCOVER will not throw up any errors if a sensor is not connected to COM2 or COM3.

If steps 3 through 6 were not carried out, then these errors would be reported in the Diagnostic

window (Figure A-5).

Page 76: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-2 SECTION 11: BATCH PROCESSING JSF FILES

DISCOVER BATHYMETRIC 0014878_REV_B

7. Close the SonarSerial.ini file and reboot the Sonar Firmware by double clicking on the Sonar icon

in the main Sonar Remote Desktop Screen (designated by the green arrow in Figure A-6).

Proceed to Step 8.

Figure A-1: Sonar Remote Desktop Application, Main Screen

Page 77: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-3

Figure A-2: Sonar Remote Desktop, Shortcut to EdgeTech Folder

Page 78: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-4 SECTION 11: BATCH PROCESSING JSF FILES

DISCOVER BATHYMETRIC 0014878_REV_B

Figure A-3: Finding the SonarSerial.ini File

Page 79: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-5

Figure A-4: SonarSerial.ini File

Figure A-5: Diagnostic Window, Reporting Ports Inactive Error

Page 80: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-6 SECTION 11: BATCH PROCESSING JSF FILES

DISCOVER BATHYMETRIC 0014878_REV_B

Figure A-6: Rebooting Sonar from the Main Screen

8. Now, plug the device into COM 1.

9. Click on the Misc. at the top of the Sonar screen (as shown in Figure A-7), then Serial Port

Information.

10. Check COM1 and set the correct baud rate for that device by using the drop down menu shown

in blue (Figure A-7). When the correct baud rate is set, the State should say OK and Errors should

not be increasing (as circled in green, Figure A-7). The correct data string should also be scrolling

in the lower part of the window (green arrow, Figure A-7).

11. Now click on File > Show Status in the sonar window (circled in red, Figure A-8).

12. On the right hand side of the window, click on the Config bullet under the Update button and

then on the drop down menu named Config Item (in green).

13. Select Serial and then click on the Update button several times (blue arrow).

14. Check to make sure the sensor's usage percentage is well below 80% (circled in purple). If it is not,

increase the baud rate of the sensor, change the baud rate under Serial Port Information as done

before in step 10, and check the Status again, otherwise the sonar will report errors. Proceed to

step 15.

Page 81: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-7

Figure A-7: Checking Serial Port Information

Page 82: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-8 SECTION 11: BATCH PROCESSING JSF FILES

DISCOVER BATHYMETRIC 0014878_REV_B

Figure A-8: Checking Sensor Status

15. When satisfied with the incoming data, press OK in the System Status window (Figure A-9).

Page 83: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-9

Figure A-9: Closing the System Status Window

16. Finally, close the Sonar - 192.9.0.101 - Remote Desktop Connection window. If this window is left

open during data acquisition, the user may experience a very slow computer. This is because the

remote desktop application uses a large amount of the link capacity to stay open. So it is always

a good idea to close the remote desktop application when finished with the configurations.

A.2 Two Sensors

Two sensors, such as a dual headed GPS (heading, position, and time source) and a Motion Reference Unit

(roll, pitch, and heave) may be used to supply the necessary information. For this type of configuration,

only COM1 and COM2 will be utilized.

To configure the COM ports for the two sensors:

17. Follow steps 1 through 6 as stated previously under sub-section A.1.

18. Under [Serial3] set Create = 0, select File > Save, and close the file (Figure A-10).

Page 84: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-10 SECTION 11: BATCH PROCESSING JSF FILES

DISCOVER BATHYMETRIC 0014878_REV_B

CAUTION!

Do not touch anything else or the system’s configuration files could be

seriously compromised and functionality may suffer.

Figure A-10: SonarSerial.ini File

This will deactivate COM3. By deactivating this port, Sonar will report to DISCOVER that only two

sensors are to be used on COM1 and COM 2 to parse the necessary sensor information. Therefore,

DISCOVER will not throw up any errors if a sensor is not connected to COM3.

19. Reboot the Sonar Firmware by double clicking on the Sonar icon in the main Sonar Remote

Desktop Screen (designated by the green arrow in Figure A-6).

Page 85: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-11

20. Now, plug in the devices into COM 1 and COM 2. Which device is connected to which COM port

does not matter, as long as their respective ports and baud rates are set in Sonar.

21. Click on the Misc. at the top of the Sonar screen (as shown in Figure A-7), then Serial Port

Information.

22. Check COM1 and set the correct baud rate for that device by using the drop down menu shown

in blue (Figure A-7). When the correct baud rate is set, the State should say OK and Errors should

not be increasing (as circled in green, Figure A-7). The correct data string should also be scrolling

in the lower part of the window (green arrow, Figure A-7).

23. Now Check COM2 and set the correct baud rate for the other device as done before in Step 6.

24. Now click on File > Show Status in the sonar window (circled in red, Figure A-11).

25. On the right hand side of the window, click on the Config bullet under the Update button and

then on the drop down menu named Config Item (in green, Figure A-11).

26. Select Serial and then click on the Update button several times (blue arrow, Figure A-11).

27. Check to make sure the sensors' usage percentages are well below 80% (circled in purple, Figure

A-11). If they are not, increase the baud rate of the sensor, change the baud rate under Serial Port

Information as done before in steps 6 and 7, and check the Status again, otherwise the sonar will

report errors.

Figure A-11: Checking Sensors’ Usages

Page 86: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-12 SECTION 11: BATCH PROCESSING JSF FILES

DISCOVER BATHYMETRIC 0014878_REV_B

28. When satisfied with the incoming data, press OK in the System Status window (Figure A-11).

29. Finally, close the Sonar - 192.9.0.101 - Remote Desktop Connection window.

If this window is left open during data acquisition, the user may experience a very slow computer.

This is because the remote desktop application uses a large amount of the link capacity to stay

open. So it is always a good idea to close the remote desktop application when finished with the

configurations.

A.3 Three Sensors

Three sensors, such as a gyro (heading), a GPS (position, and time source) and a Motion Reference Unit

(roll, pitch, and heave) may be used to supply the necessary information. For this type of configuration,

all available COM ports will be utilized.

To configure the COM ports for the three sensors, the following steps should be carried out.

30. Follow steps (1.) through (6.) as stated previously under Section A.1.

31. Make sure Create=1 under each [Serial#]. (Figure A-12)

Page 87: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-13

Figure A-12: All Sensors Set

32. Then select File > Save, and close the file.

CAUTION!

Do not touch anything else or the system’s configuration files could be

seriously compromised and functionality may suffer.

33. Close the SonarSerial.ini file and reboot the Sonar Firmware by double clicking on the Sonar icon

in the main Sonar Remote Desktop Screen (designated by the green arrow in Figure A-6).

34. Now, plug in the devices into COM 1, COM 2, and COM 3. The Gyro should be connected to COM1,

the MRU COM2, and the GPS to COM3. The devices connected to COM1 or COM2 may be

interchanged but the GPS must be connected to COM3.

Page 88: DIS OVER ATHYMETRI - EdgeTech€¦ · EdgeTech 4 Little Brook Road West Wareham, MA 02576 Tel: (508) 291-0057 Fax: (508) 291-2491  DIS OVER ATHYMETRI U S E …

A-14 SECTION 11: BATCH PROCESSING JSF FILES

DISCOVER BATHYMETRIC 0014878_REV_B

35. Click on the Misc. at the top of the Sonar screen (as shown in Figure A-7), then Serial Port

Information.

36. Check COM1 and set the correct baud rate for that device by using the drop down menu shown

in blue (Figure A-7). When the correct baud rate is set, the State should say OK and Errors should

not be increasing (as circled in green, Figure A-7). The correct data string should also be scrolling

in the lower part of the window (green arrow, Figure A-7).

37. Now check COM2 and set the correct baud rate for the other device as done before in Step 6.

38. Also, remember to check COM3 as before and set the correct baud rate for GPS.

39. Now click on File > Show Status in the sonar window (circled in red, Figure A-8).

40. On the right hand side of the window, click on the Config bullet under the Update button and

then on the drop down menu named Config Item (in green, Figure A-8).

41. Select Serial and then click on the Update button several times (blue arrow, Figure A-13).

42. Check to make sure the sensors' usage percentages are well below 80% (circled in purple, Figure

A-13). If they are not, increase the baud rate of the sensor, change the baud rate under Serial Port

Information as done before in steps 6 and 7, and check the Status again, otherwise the sonar will

report errors.

43. When satisfied with the incoming data, press OK in the System Status window (Figure A-13).

44. Finally, close the Sonar - 192.9.0.101 - Remote Desktop Connection window.

NOTE: If this window is left open during data acquisition, the user may

experience a very slow computer. This is because the remote desktop

application uses a large amount of the link capacity to stay open. So it is

always a good idea to close the remote desktop application when finished

with the configurations.