+ All Categories
Home > Documents > Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

Date post: 21-Jan-2016
Category:
Upload: marsha-charles
View: 219 times
Download: 0 times
Share this document with a friend
21
Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012
Transcript
Page 1: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

Dose Measurements

S. Roesler(DGS-RP)

Collimation WG – IR3 passive absorbers4 June 2012

Page 2: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

2

Outline

Collimation WG – IR3 passive absorbers4 June 2012

- Measurements 1) Ambient dose equivalent

• manual radiation survey (AD6 survey instrument), TS1 25/04/12• PMI readings

2) Absorbed dose• high level dosimetry (RPL dosimeters), 2011 (preliminary)

Thanks to: F.Jaquenod, C.Tromel, J.Tummer, H.Vincke

- Comparison IR3 vs IR7

- Scaling to nominal operational parameters and comparison to simulations for IR7

Page 3: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

3Collimation WG – IR3 passive absorbers4 June 2012

IR3 – Ambient dose equivalent (survey 25/04/12)

Survey: 25/04/12Survey: 25/04/12

Page 4: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

4Collimation WG – IR3 passive absorbers4 June 2012Survey: 25/04/12Survey: 25/04/12

IR3 – Ambient dose equivalent (survey 25/04/12)

Page 5: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

5Collimation WG – IR3 passive absorbers4 June 2012

12 µSv/h 6.5 µSv/h6.3 µSv/h 5.1 µSv/h

IR3 – Ambient dose equivalent (PMI readings 31/01/12)

Page 6: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

6Collimation WG – IR3 passive absorbers4 June 2012

IR3 – Absorbed dose (2011)

4.2 kGy (LSXL3-19) 5.2 kGy (LSXR3-19)

Page 7: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

7Collimation WG – IR3 passive absorbers4 June 2012

IR3 – Absorbed dose (2011)

1.8 kGy (LSXL3-18) 1.8 kGy (LSXR3-18)

Page 8: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

8Collimation WG – IR3 passive absorbers4 June 2012

IR3 – Absorbed dose (2011)

300 Gy(LSXL3-16)

Page 9: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

9Collimation WG – IR3 passive absorbers4 June 2012

IR3 – Absorbed dose (2011)

5.2 kGy (LSXL3-13) 1.9 kGy (LSXR3-14)

Page 10: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

10Collimation WG – IR3 passive absorbers4 June 2012

Survey: 25/04/12Survey: 25/04/12

IR7 – Ambient dose equivalent (survey 25/04/12)

Page 11: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

11Collimation WG – IR3 passive absorbers4 June 2012

Survey: 25/04/12Survey: 25/04/12

IR7 – Ambient dose equivalent (survey 25/04/12)

Page 12: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

12Collimation WG – IR3 passive absorbers4 June 2012

Survey: 25/04/12Survey: 25/04/12

IR7 – Ambient dose equivalent (survey 25/04/12)

Page 13: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

13Collimation WG – IR3 passive absorbers4 June 2012

Survey: 25/04/12Survey: 25/04/12

IR7 – Ambient dose equivalent (survey 25/04/12)

Page 14: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

14Collimation WG – IR3 passive absorbers4 June 2012

IR7 – Absorbed dose (2011)

10.0 kGy (LSXL7-24) 9.2 kGy (LSXR7-22)

Page 15: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

15Collimation WG – IR3 passive absorbers4 June 2012

IR7 – Absorbed dose (2011)

6.6 kGy (LSXL7-18) 5.6 kGy (LSXR7-18)

Page 16: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

16Collimation WG – IR3 passive absorbers4 June 2012

IR3 vs. IR7 – Absorbed dose D3 (2011)

10.0 kGy (LSXL7-24) 9.2 kGy (LSXR7-22)

4.2 kGy (LSXL3-19) 5.2 kGy (LSXR3-19)

IR3

IR7

Page 17: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

17Collimation WG – IR3 passive absorbers4 June 2012

IR3 vs. IR7 – Absorbed dose Q5 (2011)

IR3

IR7

6.6 kGy (LSXL7-18) 5.6 kGy (LSXR7-18)

5.2 kGy (LSXL3-13) 1.9 kGy (LSXR3-14)

Page 18: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

18

IR7-Right

January 2011 January 2012 January 2013(Jan.2011 x fac.15)

January 2017(Jan.2011 x fac.54)

Element Contact AisleContact

B1/B2Aisle(1m)

Contact Aisle Contact Aisle

TCP.D6 10.0 1.2 31/76 6 150.0 18.0 540.0 65.0

TCP.C6 18.0 2.5 275/345 16 270.0 38.0 970.0 135.0

TCP.B6 31.0 3.1 300/560 40 465.0 47.0 1670.0 170.0

TCAPA 70.0 3.0 1100/860 35 1050.0 45.0 3780.0 160.0

TCAPB 13.0 1.2 145/203 13 195.0 18.0 700.0 65.0

TCSG.A6 8.0 1.5 450/450 28 120.0 23.0 430.0 80.0

TCAPC 65.0 2.5 870/840 16 975.0 38.0 3510.0 135.0

Dose equivalent rates (μSv/h)

Measurements Extrapolations

Collimation WG – IR3 passive absorbers4 June 2012

IR7 – Scaling to nominal parameters

Page 19: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

19

Dose equivalent rates (μSv/h) (about two months cooling)

IR7-Right January 2013 Nominal(Jan.2013 x factor 3.6)

Element Contact Aisle Contact Aisle

TCP.D6 150.0 18.0 540.0 65.0

TCP.C6 270.0 38.0 970.0 135.0

TCP.B6 465.0 47.0 1670.0 170.0

TCAPA 1050.0 45.0 3780.0 160.0

TCAPB 195.0 18.0 700.0 65.0

TCSG.A6 120.0 23.0 430.0 80.0

TCAPC 975.0 38.0 3510.0 135.0

S.Roesler et al., EDMS 863919

• reasonable agreement for passive absorber • dose rate somewhat lower than predicted for first secondary collimator (due to initial settings?)

Collimation WG – IR3 passive absorbers4 June 2012

IR7 – Scaling to nominal parameters

Page 20: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

20

Dose equivalent rates (μSv/h)

IR3-Right January 2011 January 2012January 2013

(Jan.2011 x fac.15)January 2017

(Jan.2011 x fac.54)

Element Contact AisleContactB1/B2

Aisle(1m)

Contact Aisle Contact Aisle

TCP 13.0 0.3 180/88 6 195.0 4.5 702.0 16.0

TCAPA 24.0 0.7 440/360 8 360.0 11.0 1300.0 38.0

TCSG.5 7.5 0.2 195/172 12 113.0 3.0 405.0 11.0

MQWA.C 9.0 270/176 5 135.0 490.0

Measurements Extrapolations

Collimation WG – IR3 passive absorbers4 June 2012

IR3 – Scaling to nominal parameters

Page 21: Dose Measurements S. Roesler (DGS-RP) Collimation WG – IR3 passive absorbers 4 June 2012.

21

Summary

Collimation WG – IR3 passive absorbers4 June 2012

• measurement results exist from 1) PMI ionization chambers (ambient dose equivalent) 2) dedicated radiation surveys (ambient dose equivalent) 3) HLD (absorbed dose in air)• direct comparison with simulation results for the operational conditions of 2011 not (yet) available, scaling of measured residual dose rates to nominal operational parameters indicated rough agreement (within factor 2-5) with simulations for IR7

• measurements show asymmetry between beam 1 and 2 for both IR3 and 7, doses higher by about a factor of two for beam 1• residual dose rates higher in IR7 as compared to IR3 (factor of 10 for passive absorber in in front of D3), ratio between both probably different for number of lost protons due to different layouts (e.g., number and location of collimators and absorbers)• absorbed doses upstream of D3 and Q5 yokes (at the locations of the dosimeters!) higher for IR7 (factor of 2-5), conclusions on peak dose requires simulations


Recommended