8/13/2019 DS_PE_TTF_1_en_us_16483
http://slidepdf.com/reader/full/dspettf1enus16483 1/2
Electronic
Pressure Measurement
Datasheet TTF-
Page 1 of Datasheet TTF-1 1/2011
Special Purpose Thin-Film OEM-Pressure TransducerType TTF-1
Possible versions of the TTF-1 Pressure Transducer
Applications
n Applications with limited installation space
n Embedded pressure sensors
n Design-in solutions
n Hydraulic pressure monitoring
Special Features
n Thin-lm technology
n Pressure ranges from 0 ... 10 bar to 0 ... 1000 bar
(150 to 15,000 PSI)
n Stainless steel wetted parts
n Media temperature range: -40 °F ... +212 °F
(-40 °C ... +100 °C)
n Integrated temperature compensation
Description
Flexible installation
The compact size the TTF-1 pressure transducer allows it be
used in applications where mounting space is very limited.
Rugged sensor element
The TTF-1 sensor is machined from a single piece of stain-less steel. The thin lm Wheatstone bridge is applied to the
diaphragm using a physical vapor deposition process called
“sputtering”. This process forms a tight molecular bond with
the sensor diaphragm. This eliminates the risk of “sensor
creep” that may occur over time with bonded foil strain gauge
designs that use an adhesive to bond the sensor to the
diaphragm.
The high grade stainless steel design of the sensor element
provides a stainless steel barrier between the sensor element
and the media. This completely welded, dry measuring cell
does not require the use of any internal pressure transmis-
sion uid. TTF-1 sensors constructed of Elgiloy are available
for special requirements.
The sensor element is welded directly to the pressure port.
This eliminates the use of soft sealing materials that may
deteriorate and leak over time. It also eliminates weak points
that sometimes occur when o-rings or adhesive joints seals
are used.
Excellent all-around performance
The TTF-1 provides is temperature compensated from -40 °F
... +212 °F (-40 °C ... +100 °C), so that in most applications
no additional temperature compensation is required.
The pressure transducer oers high overpressure safety
and is resistant to pressure spikes and dynamic pressure
changes. In addition the TTF-1 is an excellent sensor for
applications where high accuracy, zero point stability and low
temperature error are essential requirements.
Individual customer designs
Based on many years of production experience, WIKA can
provide customer-specic designs for quantities above 1000
pieces.
8/13/2019 DS_PE_TTF_1_en_us_16483
http://slidepdf.com/reader/full/dspettf1enus16483 2/2
WIKA Instrument Corporation
1000 Wiegand Boulevard
Lawrenceville, GA 30043
1-888-WIKA-USA /770-513-8200 (in GA)
Fax 770-338-5118
[email protected] www.wika.com
RSpecifications and dimensions given in this data sheet represent the state of engineering at the time of printing.Modifications may take place and materials specified may be replaced by others without prior notice.
Specifcation
1)
Type TTF-1
Pressure ranges 10 bar 16 1)bar 25 bar 40 .5 bar 60 bar 100 bar 160 bar 250 bar 400 bar 600 bar 1,000 2) bar
Over pressure safety 20 bar 32 bar 50 bar 80 bar 120 bar 200 bar 320 bar 500 bar 800 bar 1,200 bar 1,500 bar
Burst pressure 100 bar 160 bar 250 bar 400 bar 550 bar 800 bar 1,000 bar 1,200 bar 1,700 bar 2,400 bar 3,000 bar
(1 bar = 14.504 psi) 1) Availability depends on the specic design
2) Higher pressure ranges available on request
Pressure connection On request
Materials
Wetted parts Stainless steel
Power supply UB DC V Typical 6 (6 ... 10 recommended) {other on request}
Dielectric strength AC V 500
Insulation resistance MOhm > 300
Bridge resistance RB kOhm 6.5 + 1.3 (between + US and – US)
Span at nominal pressure mV/V 2
Zero signal mV/V Typical 0 + 0.2 (maximum 0 + 0.5)
Response time (10 ... 90%) ms < 0.1
Linearity (Limit point calibration)
■ With pressure range bar 10 16 25 40 60 100 160 250 400 600 1000
■ Typical % of span +0.5 +0.4 +0.25 +0.35 +0.30 +0.25 +0.22 +0.20 +0.18 +0.15 +0.12
Span tolerance % of span +30 +25 +15 +20 +15 +15 +15 +15 +15 +15 +15
1-year stability % of span 0.2 (at reference conditions) Permissible temperature of
■ Medium 3) °C -40 ... +212 °F / -40 ... +100 °C
■ Ambient 3) °C -22 ... +176 °F / -40 ... + 80 °C {extended temperature range on request}
■ Storage 3) °C -22 ... +176 °F / -40 ... + 80 °C {extended temperature range on request}
3) Also complies with EN 50178, Tab. 7, Operation (C) 4K4H, Storage (D) 1K4, Transport (E) 2K3
Compensated temperature range °C -40 ... +212 °F / -40 ... +100 °C
Temperature coecients within
compensated temperature range
■ Mean TC of zero % of span Typical + 0.1/10K
■ Mean TC of range % of span Typical + 0.1/10K
{ } Items in curved brackets are optional extras for additional price.
Dimensions in inches (mm) (1mm = 0.039”) Circuit diagram
black
red
Typical design
example white
blue
Page 2 of 2 Datasheet TTF-1 1/2011