Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1
Transcript
Page 1: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

1

Dual-sided readout studies

Chiara CasellaMatthieu Heller

Oliver Holme

Page 2: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

2

Dual-sided readout studies• Difference in Left and Right light yield is used to estimate

gamma interaction position along z axis

• Original AXPET concept was based on dual-sided readout– Initial studies showed insufficient z-axis resolution (~cm)– New results from A. Goertzen indicated better z resolution (~3.5mm)

• Investigation started to find best achievable resolution– Following A. Goertzen method– Using digital SiPM for a compact axial module

L R

z-axis 0

Page 3: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

3

Laboratory setup• Philips digital SiPM tiles – L, R and Tagger

– L and R tiles are cooled to ~14˚C to reduce dark count

• Only events with 3 tile coincidence are stored– Acts to collimate a beam of gamma photons

• Automated data taking with Philips and LabVIEW software– Scans through several position along z-axis

Page 4: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

4

Crystal scanning• 3x3x100mm LYSO crystals• Crystal aligned with one SiPM pixel– All other cells were inhibited

• Typical scan properties– 11 data taking runs at different positions along z axis

• Typical run properties– ~15 minutes (with 3.7 MBq source)– Up to 9000 events stored to file– Up to 1300 events after photopeak cuts

Page 5: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

5

Data processing for each z position• Data from SiPMs is converted to ROOT files• Light yield is processed:

– Cut on photopeaks (all three tiles)– Correct for SiPM cell saturation– Correct for SiPM intra-tile cross talk– Obtain photopeak central values

Raw dataGaussian fitCorrected data

Page 6: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

6

Data processing for complete scan• z position estimator calculated as for each event• Mean for each z position is plotted against z• Fit function is applied to plot

– Fit gives a relationship between any value and z position

Error bars show σ of (R-L)/(R+L)

(R-L)/(R+L)

(R-L

)/(R

+L)

z position [mm]

Page 7: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

Data processing for complete scan• z axis resolution

– FWHM ofvalues for each z position mapped to FWHM of z

• Beam width is subtracted to obtain final resolution

– Beam size calculated as 1.7mm

7Rz FWHM

Error bars show σ of (R-L)/(R+L)(R

-L)/

(R+L

)

Position [mm]

FWHM

Page 8: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

8

Crystals and surface treatments• Coatings/wrappings:

– Bare– White TiO2 paint– Teflon (2 layers)– Enhance Specular Reflector (ESR)

• Surface treatments (one or multiple faces):– Polished (from manufacturer)– Depolished (with ~10 µm powder)– Machined strips (with CNC diamond tool)

Page 9: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

9

Machined strips• Two patterns of machined strips– Every 5mm, aligned on four faces

– Every 5mm, staggered by 1.25mm on four faces

Machined strips are ~0.7mm wide

Page 10: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

10

Results

Page 11: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

11

Non-exponential LY versus z• Crystals with less attenuation– LY versus z is not exponential

• Crystals with greater attenuation– LY versus z is more similar to exponential

Polished, teflon wrapped Two faces depolished, teflon wrapped

Page 12: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

12

Inter-tile cross talk• SiPMs have cross talk between cells in a tile

– Intra-cell cross talk– Caused by avalanches producing secondary photons

• The secondary photons are emitted isotropically– They can also enter and propagate through the crystal– Inter-tile cross talk

• Inter-tile cross talk reduces L & R difference– Degrades z resolution

• This degradation can be measured by capturing L and R tiles independently, with the other tile off– Not possible in real applications– Physically filtering the cross talk photons might be possible

Page 13: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

13

Photopeak shifting

LEFT RIGHT SUM

Machined strips, 5mm (aligned), teflon wrapped

z-ax

is

Page 14: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

14

Light yields and energy resolutionMachined strips, 5mm (aligned), teflon wrapped

Page 15: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

15

z resolutionMachined strips, 5mm (aligned), teflon wrapped

Page 16: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

16

Different surface treatments• Resolution down to ~3mm has been achieved– Treatments give different energy resolutions due to

different LY sum

Polished + TiO2 paint Machined strips (four faces, aligned) + ESR Depolishing (two faces) + ESR

Energy resolution

Ener

gy re

solu

tion

[%]

Page 17: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

17

Machined strips• Fine scans reveal discrete structure

– Gives stepped LY pattern– Seen in LY sum and energy resolution too

Aligned strips, light yield - RIGHT Staggered strips, light yield - RIGHT

Fitt

ed p

hoto

peak

pos

ition

[pe]

Fitt

ed p

hoto

peak

pos

ition

[pe]

Position [mm]Position [mm]

Staggered strips give a pseudo-continuous behaviour

Page 18: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

18

Depolishing

One face depolished + ESR Two faces depolished + ESR

Page 19: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

19

WrappingsFour faced machined strips (aligned)

ESR wrapped λeff ≈ 70 mm

Teflon wrapped λeff ≈ 40 mm

Unwrapped λeff ≈ 28mm

Light yield - LEFT Light yield - RIGHT

Light yield - SUM

z [mm] z [mm]

z [mm]

ESR gives almost constant LY SUM

Page 20: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

20

WrappingsFour faced machined strips (aligned)

Teflon and ESR give similar resolutionESR gives more linear z position estimatorESR gives more light to improve resolution further

Page 21: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

21

Results summaryCrystal surface Wrapping/

CoatingRz FWHM at

mid-point (mm)Energy resolution at

mid-point (%)Polished Bare 52 8Polished Teflon 22 8Polished ESR 37 7Polished TiO2 paint 3.1 15Depolished (1 face) Bare 5.8 18Depolished (1 face) Teflon 4.4 12Depolished (1 face) ESR 4.9 8Depolished (2 faces) Teflon 3.4 16Depolished (2 faces) ESR 3.7 11Machined strips (aligned) Bare 3.6 21Machined strips (aligned) Teflon 2.6 12Machined strips (aligned) ESR 3.2 8Machined strips (staggered) Teflon 3.1 12Machined strips (staggered) ESR 3.8 7

A crystal with better Rz at z=50mm is not necessarily better at other z positions

Page 22: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

22

Next steps and outlook• Improve results for 100mm crystals– Extend machined strips to full crystal length– Depolish more faces

• Shorter 60mm crystals– Could achieve resolution down to 2mm– Attenuation per mm would need to be increased

• Photodetectors with higher PDE ?– Greater light yield would enable better resolution– Current SiPM PDE is about 40%

Page 23: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

23

Page 24: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

24

Toy Monte Carlo results• Measured results are consistent with Toy MC– Estimated ~3mm as best achievable resolution– Can indicate future possibilities

λeff (mm) Crystal Length (mm)

Max LY Rz at mid-point FWHM (mm)

33 100 1500 3.03

33 100 3000 2.14

33 60 1500 2.27

20 60 1500 1.85

10 60 1500 1.99

20 60 3000 1.32

7 20 1500 0.70

7 20 3000 0.5320mm crystal figures match other published studies

Increased LY improves resolution

60 mm crystals would enable improvementdown to 2mm with current light yields

Pure Poisson distributionNoise

Not representative

Page 25: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

25

Machined strips width

Page 26: Dual-sided readout studies Chiara Casella Matthieu Heller Oliver Holme 1.

26

Depolished (one face)


Recommended