Effect of phase-encoding scheme on the relationshipbetween contrast agent concentration and signal intensityon inversion recovery turbo fast low-angle shot T1-weightedimages
Mahmood Nazarpoor
Received: 13 June 2013 / Revised: 20 January 2014 / Accepted: 21 January 2014
� Japanese Society of Radiological Technology and Japan Society of Medical Physics 2014
Abstract Previous studies have shown that the image
parameters such as the repetition time and inversion time
(TI) and image sequences can have an effect on the max-
imum linearity between signal intensity (SI) and concen-
tration. Our aim in this study was to investigate the effect
of a phase-encoding scheme (linear phase-encoding and
center out phase-encoding) at different TIs on the maxi-
mum linear relationship between contrast agent concen-
tration and SI using inversion recovery TurboFLASH
(turbo fast low angle shot) T1-weighted images in MRI. A
phantom was designed to hold 25 vials which contained
either different (between 0 and 19.77 mmol/L) or constant
(1.20 mmol/L) concentrations of contrast agent. The vials
of constant concentration were used for measurement of
coil non-uniformity. The vials of different concentrations
were used for the measurement of the maximum linear
relationship between SI and concentration and the effect of
a phase-encoding scheme on the linearity, where the
squared correlations (R2) between the corrected SI and the
concentration were equal to 0.95 and 0.99. The results
showed that the linear relationship between SI and con-
centration was dependent on the phase-encoding scheme
and on the TIs. The results also showed that the maximum
linearity with the linear phase-encoding acquisition was
higher than that with the center out phase-encoding
acquisition at the same R2. This study shows that the phase-
encoding scheme is an important parameter when the SI is
measured. These schemes can have an effect on the max-
imum linearity between SI and concentration.
Keywords T1-weighted image � Inversion recovery �Signal intensity � Concentration of contrast agent �Center out phase-encoding � Linear phase-encoding
1 Introduction
Gd-based MR contrast agents play an indirect role in the
MR signal by reducing the relaxation times of the sur-
rounding nuclear spins [1]. MRI is not able to directly
measure the concentration of the contrast agent. The con-
centration is measured indirectly from the signal intensity
(SI). If the relationship between SI and concentration can
be shown to be linear over a range of concentrations, then
relative changes in SI can be used for finding the
concentration.
Therefore, the SI–time curve can be used instead of the
concentration–time curve for the measurement of hemo-
dynamic parameters, such as organ blood flow and organ
blood volume [2].
Different publications suggested different values for the
maximum linearity between SI and concentration in the
stationary state and steady state [3–5]. For this reason, the
relationship between SI and concentration should be
investigated more thoroughly for determining the maxi-
mum concentration that yields a linear response.
Our previous studies showed that image parameters such
as the repetition time (TR) and inversion time (TI) and
image sequences can have an effect on the maximum lin-
earity between SI and concentration [6–8].
Our aim in this study was to investigate the effect of a
phase-encoding scheme (center out phase-encoding and
linear phase-encoding) at different TIs on the maximum
linear relationship between the contrast agent concentration
and SI after correction for coil non-uniformity. The squared
M. Nazarpoor (&)
Department of Radiology, Faculty of Paramedicine, Tabriz
University of Medical Sciences, Tabriz, Iran
e-mail: [email protected]; [email protected]
Radiol Phys Technol
DOI 10.1007/s12194-014-0260-7
correlation (R2) between the corrected SI and the concen-
tration was set to 0.95 and 0.99 by use of inversion
recovery TurboFLASH (turbo fast low-angle shot) T1-
weighted images.
2 Materials and methods
2.1 Phantom
For assessment of the relationship between SI and con-
centration in stationary vials, a phantom was designed to
hold vials which contained either different or constant
concentrations of a contrast agent. The shape of the
phantom was approximately cubic; it was made of Perspex.
Its length, width and height were 20, 18, and 20 cm,
respectively [6]. Figure 1 shows a coronal image of the
phantom, which contains the vials (glass tube, inner
diameter approximately 15 mm) with different contrast
agent concentrations. The position of the different con-
centrations inside the vials can be seen from the figure.
The phantom of different concentrations consisted of 25
vials, filled with different concentrations of Gd-DTPA
(Magnevist, Schering Health Care Ltd, West Sussex, UK).
The concentration of Gd-DTPA varied between 0 and
19.77 mmol/L (0.00, 0.30, 0.45, 0.60, 0.75, 0.90, 1.20,
1.50, 1.80, 2.10, 2.39, 2.69, 2.99, 3.28, 3.58, 3.98, 4.96,
5.95, 7.93, 9.90, 13.85, and 19.77 mmol/L).
A clinical head and neck coil was used with the phan-
tom. The phantom was located inside the coil. The vials
were set vertically, and the axes of the vials were per-
pendicular to the image plane (coronal image).
Two experiments were performed, one using vials with
different concentrations for measurement of the relation-
ship between SI and concentration and the other using vials
of a constant concentration for measurement of coil non-
uniformity. The vials in the phantom with constant con-
centration (1.20 mmol/L) were placed in exactly the same
positions as the vials with different concentrations [6].
One of the major sources of image non-uniformity in the
high-field MR scanners is the radio-frequency (RF) coil
inhomogeneity [9–11]. Therefore, for measuring the accurate
SI in an image, the response of the RF coils should be uniform.
The non-uniformity of the coil will be calculated from vials
with constant concentration. If the coil is uniform, the SI of the
vials with the constant concentration (1.20 mmol/L) for the
different parts of the phantom should be the same. As the coil
is not uniform, the SI was different. We chose the SI of one vial
in the center of the coil as reference SI (i.e., the vial with a
concentration of 1.20 mmol/L). Therefore, the SI of the vials
with the constant concentration should be equal to that of SI.
After normalization of the mean SI from the vials with con-
stant concentration, the correction factor due to the non-uni-
formity of the coil can be calculated for different parts of the
vials with different concentrations.
For calculating the corrected SI for different concen-
trations, the SI of each vial was multiplied by its correction
factor [7, 8].
2.2 Theory
In MRI, the acquired data actually represent the Fourier
transformation of the imaged object rather than the object
itself. In the other words, MRI acquires data in k-space, and
Fig. 1 Coronal image of the
phantom. The positions of
different concentrations inside
the vials can be seen from this
figure
M. Nazarpoor
a Fourier transformation will convert k-space data to image
data. The center of k-space determines the majority of the
image contrast, and the edges of k-space determine high-
resolution structural information on the image [12, 13].
For T1-weighted monitoring of the contrast agent, a
number of T1-weighted sequences are available [14]. The
inversion recovery sequence first uses a 180� pulse, which
is then followed by a 90� excitation pulse.
Figure 2 is a schematic representation of the inversion
recovery gradient echo TurboFLASH sequences. TE is the
echo time.
The standard inversion recovery sequence, which is
dependent on TI and TR, is described by the following
equation [7, 8]:
SðtÞ ¼ S0 1� 2 exp �TICðtÞ
Kþ 1
T1Pr e
� �� ��
þ exp �TRCðtÞ
Kþ 1
T1Pr e
� �� ��; ð1Þ
where S(t) and S0 are the SI after administration of a
contrast agent and the observed SI in the absence of any
magnetization preparation pre-pulses or contrast agent,
respectively. T1Pre is the longitudinal relaxation time
before contrast application. C(t) is the concentration of the
contrast agent at time t. K is a constant that depends on the
contrast medium.
The transverse relaxation time (T2) will also be influ-
enced at higher concentrations of the contrast medium.
Therefore, Eq. (1) should be multiplied by exp � TE
T2
� �at
higher concentrations of contrast agents [15, 16].
Two common acquisition strategies in MRI are linear
phase-encoding and center out phase-encoding [17]. As
stated by Jivan et al. [18], after an inversion pulse, which
occurs at time zero, as the longitudinal magnetization
recovers, images can be acquired by the TurboFLASH
image sequence. Each image can be formed by application
of a set of a pulses, each of which gives one k-space line.
The signal will decrease after each a to reach a steady-state
saturation level. After this step, it will be constant during
image acquisition.
The center out phase-encoding acquires the center of
k-space at time zero or the start of the RF pulses, and
most of the image contrast is determined by the central
views of k-space and a small number of flash lines of k-
space around the centre of k-space line. The linear
phase-encoding acquires the center of k-space at the time
from the inversion pulse to the mid-line of k-space when
the longitudinal magnetization recovery has reached a
steady-state level; this is usually called the effective TI
(see Fig. 3) [13].
TI in Eq. (1) is the TI for the center out phase-encoding
and the effective TI for the linear phase-encoding
acquisitions.
2.3 Image acquisition
The phantom was positioned within the head and neck coil.
All studies were carried out with a 1.5-T clinical MR
scanner (Vision, Siemens Medical, Erlangen, Germany).
We used T1-weighted TurboFLASH images to measure SI
TE
Centre of K space
180 degree inversion pulse
Effective TI
180 degree inversion pulse
Echo
α α α α α α1 20 n−3 n−2 n−1
Time
TI set
TR for one FLASH line
TurboFLASH
Inversion recovery TurboFLASH
TR for one TurboFLASH image
Fig. 2 Inversion recovery gradient echo TurboFLASH for one
image. This is an inversion recovery TurboFLASH with n a pulses,
where the n/2th line traverses the center of k-space (for linear phase-
encoding). ‘‘TI set’’ is the time between the 180� inversion pulse and
the first a excitation pulse. The effective TI is the time from the
inversion pulse to the mid-line of k-space n/2nd. TE and TR are the
echo time and repetition time, respectively. Imaging gradients are not
shown in this diagram
Effect of phase-encoding scheme
in the vials with different and constant concentrations. For
investigation of the effect of the phase-encoding scheme on
SI, the two acquisitions were preformed for the inversion
recovery (IR) sequence.
The TurboFLASH imaging parameters were as follows:
matrix size = 128 9 128, time for one FLASH
line = 8.5 ms, echo time (TE) = 4 ms, TR = 3 ms,
effective TI varied between 744 and 2944 ms (744, 844,
944, 1044,1144, 1244, 1344,1444, 1544, 1744, 1944, 2144,
2344, 2544, 2744, and 2944 ms) for the linear phase-
encoding, and TI varied between 700 and 1600 ms (700,
800, 900, 1000, 1200, 1400 and 1600 ms) for the center out
phase-encoding. Because one aim of this study was to
investigate the effect of the two phase-encoding schemes
on the maximum linearity between SI and contrast agent
concentration, the TI and effective TI values were differ-
ent. Each image was repeated 10 times (TR for one Tur-
boFLASH image) in the MR scanner with a slice thickness
of 10 mm, a pixel size of 2 9 2 mm, and a flip angle of
15�.
2.4 Image analysis
The image data were transferred from the MR scanner to a
Unix workstation. The image-processing software Interac-
tive Data Language (IDL, Research Systems, Inc., http://
www.rsinc.com) was used for processing.
Programs were written to calculate automatically:
– The mean image of 10 acquisitions, to improve the
signal to noise ratio.
M+
M−
Timea
TI
Effective TI
c
Before the start of image acquisition
End of image acquisition
Ideal longitudinal magnetization recovery
Real longitudinal magnetization recovery
128 α degree pulses
b
Fig. 3 Schematic representation of the magnetization during the T1-
weighted TurboFLASH sequence, showing the ideal and the real
shape of the longitudinal magnetization recovery. M- and M? are
the longitudinal magnetization values before and after the inversion
pulse. After applying an inversion pulse, which occurs at time zero, as
the longitudinal magnetization recovers, images can be acquired.
Each image is formed by application of radiofrequency (RF)
(a = 15�) pulses, each of which gives one k-space line. Point a
shows the start of the RF pulses. The amplitude of the signal after the
first a is higher than the other a. The time from the inversion pulse
(TI) to the mid-line of k-space is usually called the effective TI. The
effective TI (b) is placed at TI ? 64 9 8.5 ms for a typical
128 9 128 matrix size image and for one FLASH line (8.5 ms).
The center out phase-encoding acquires the center of k space at point
a and the linear phase-encoding acquires the center of k space at point
b. c is the end of image acquisition
M. Nazarpoor
– The mean SI of the 9 innermost pixels of each vial with
different and constant concentrations to avoid partial
volume effects.
– The correction factors with respect to coil non-unifor-
mity. The SI of the vials with different concentrations
was then multiplied by these factors.
– The best-fit curve of concentration versus SI curve
based on Eq. (1).
– The maximum concentration, where the R2 values
between the corrected SI and the concentration were
equal to 0.95 or 0.99 from the best-fit curve for all
experiments.
The R2 gives the strength of the linear relationship
between SI and concentration. When R2 is 0.95, this indi-
cates that 95 % of the variation in SI is explained by the
variation in the concentration [19]. A computer program
calculated R2 from the first n points in the fitted data of SI
versus concentration. As n was increased, including more
data points, the value of R2 decreased. The value of n, and
hence the concentration, that gave an R2 value of 0.95 or
0.99 was found. This concentration is known as the max-
imum value that gives a linear relationship with the SI.
These programs could be run from either a UNIX
workstation or a personal computer.
3 Results
3.1 Center out phase-encoding acquisition
Figure 4 shows a typical result from the corrected SI (the
non-uniformity of the coil) versus concentration by use of
the center out phase-encoding acquisition. The corrected SI
was measured from mean SI of the 9 innermost pixels of
each vial. The mean non-uniformity coil correction factors
were calculated as 0.90, 0.93, 1.00, 0.95, 0.89, 0.94, 1.00,
0.96, 0.86, 0.97, 0.88, 1.00, 0.93, 0.90, 0.97, 1.03, 0.94,
0.96, 0.95, 0.99, 1.02, and 0.97 for the vials with concen-
trations of 0.00, 0.30, 0.45, 0.60, 0.75, 0.90, 1.20, 1.50,
1.80 (top left), 2.10, 2.39, 2.69, 2.99, 3.28, 3.58, 3.98, 4.96,
5.95, 7.93, 9.90, 13.85, and 19.77 mmol/L, respectively.
The error bars show the standard deviation for each vial.
The dashed lines show the curve fit of the data. The figure
shows that SI decreased at high concentrations due to the
T2-shortening effect on the SI.
The maximum linear relationship between concentra-
tions and corrected SI for the center out phase-encoding
that gave R2 values equal to 0.95 and 0.99 was 2.02 and
1.25 mmol/L, respectively, at TI = 1000 ms.
Figure 5 shows the maximum concentration versus TI
by use of the center out phase-encoding acquisition with 7
different TIs.
The linear relationship between SI and concentration
was up to 3.03 and 1.23 mmol/L for the short TI (700 ms)
and long TI (1600 ms), respectively, where R2 = 0.95. In
Fig. 4 Mean corrected SI from the 9 innermost pixels of the vials
versus the concentration of contrast agent for the center out phase-
encoding acquisition. The maximum linear relationship between
concentrations and corrected SI that gave squared correlations (R2)
equal to 0.95 and 0.99 was 2.02 and 1.25 mmol/L, respectively, at
TI = 1000 ms. The error bars show the standard deviation for each
vial
Fig. 5 The maximum concentration that gives R2 = 0.95 and 0.99
versus TI with the center out phase-encoding acquisition. The linear
relationship between SI and concentration is up to 3.03 and
1.23 mmol/L for the short TI (700 ms) and long TI (1600 ms),
respectively, where R2 = 0.95. In addition, these values reduce to
1.65 and 0.71 mmol/L for these inversion times where R2 = 0.99
Effect of phase-encoding scheme
addition, these values reduce to 1.65 and 0.71 mmol/L for
these TIs when R2 = 0.99. The R2 was measured from the
best-fit curve for each TI, similar to that obtained from
Fig. 5.
As can be seen from Fig. 5, an increase in TI is asso-
ciated with a decrease in the maximum concentration.
3.2 Linear phase-encoding acquisition
Figure 6 shows a typical result for the corrected SI versus
concentration for the linear phase-encoding acquisition.
The corrected SI was measured from the mean SI of the 9
innermost pixels of each vial, which was obtained after a
non-uniformity correction. The mean non-uniformity coil
correction factors were calculated as 0.89, 0.93, 0.99, 0.96,
0.89, 0.93, 1.00, 0.95, 0.90, 0.93, 0.90, 1.01, 0.94, 0.91,
0.98, 1.05, 0.97, 0.98, 0.95, 0.99, 1.06, and 0.99 for the vials
with concentrations of 0.00, 0.30, 0.45, 0.60, 0.75, 0.90,
1.20, 1.50, 1.80 (top left), 2.10, 2.39, 2.69, 2.99, 3.28, 3.58,
3.98, 4.96, 5.95, 7.93, 9.90, 13.85, and 19.77 mmol/L,
respectively. The error bars show the standard deviation for
each vial. The dashed lines show the curve fit of the data.
The figure shows that the T1-shortening effect is dominant
at low concentrations of Gd-DTPA, based on Eq. (1), and
the T2-shortening effect is dominant at high concentrations
and leads to a decrease in the SI.
The maximum linear relationships between concentra-
tions and corrected SI that gave R2 values equal to 0.95 and
0.99 were 2.14 and 0.93 mmol/L, respectively, at TI set at
400 ms (effective TI = TI ? 64 9 8.5 = 944 ms).
Figure 7 shows the maximum concentration that gives
R2 values equal to 0.95 and 0.99 versus effective TI for the
linear phase-encoding acquisition at 16 different TI values.
The linear relationship between SI and concentration
was up to 2.64 and 0.89 mmol/L, for short effective TI
(744 ms) and long effective TI (2944 ms), respectively
(R2 = 0.95). In addition, these values were reduced to 1.15
and 0.38 mmol/L for these effective TIs where R2 = 0.99.
Figure 7 indicates that an increase in the effective TI is
associated with a decrease in the maximum concentration.
The R2 was measured from the best-fit curve for each TI,
similar to that obtained from Fig. 5.
4 Discussion
MRI is unable to measure the concentration of contrast
within the region of interest of an organ; it is measured
indirectly from the SI. At high concentrations, both T1 and
T2 can be affected as the SI response becomes nonlinear,
with an unsteady plateau. Therefore, to calculate the con-
centration from the SI, one should measure the maximum
concentration for linearity between SI and the
concentration.
Fig. 6 Mean corrected SI from the 9 innermost pixels of the vials
versus concentration of contrast agent for the linear phase-encoding
acquisition. The maximum linear relationship between concentrations
and corrected SI that gave R2 equal to 0.95 and 0.99 was 2.14 and
0.93 mmol/L, respectively, at TI set at 400 ms (effective inversion
time = TI ? 64 9 8.5 = 944 ms). The error bars show the standard
deviation for each vial
Fig. 7 Maximum concentration that gives R2 = 0.95 and 0.99 versus
effective TI. The linear relationship between SI and concentration is
up to 2.64 and 0.89 mmol/L for short effective TI (744 ms) and long
effective TI (2944 ms), respectively (R2 = 0.95). In addition, these
values reduce to 1.15 and 0.38 mmol/L for these effective inversion
times when R2 = 0.99 with the linear phase-encoding acquisition
M. Nazarpoor
The correlation between MR SI values on T1-weighted
TurboFLASH and the concentration of Gd-DOTA (Gado-
terate, 0.5 mol/mL) was reported by Canet et al. [4]. They
showed that the SI increase on T1-weighted images was
linearly proportional to the Gd chelate concentration at low
concentrations (B0.8 mmol/L) with TI = 300 ms (effec-
tive TI = 716 ms); however, the SI response becomes
nonlinear at higher concentrations.
In this study, the SI was corrected by the correction
factor of the non-uniformity of the coil.
The present results, which used the inversion recovery
TurboFLASH sequence (linear phase-encoding), from the
vials, indicated that an increase in TI was associated with a
decrease in the maximum linear concentration when the R2
was equal to 0.95 or 0.99. The result (see Fig. 7) also
shows that, at long TI (2944 ms), the maximum linear
relationship between SI and concentration was up to
0.89 mmol/L when R2 = 0.95, or up to 0.38 mmol/L when
R2 = 0.99. These values can increase at short TI. In
addition, our result (see also Fig. 7) when we used the Gd-
DTPA contrast agent under Canet’s conditions was about
2.64 (R2 = 0.95) or 1.15 mmol/L (R2 = 0.99). It should be
noted that the two contrast agents have roughly the same
relaxivity [20].
The non-uniformity of the coils can affect the SI, and this
contributes a large error to the measured SI [11]. Because
Canet did not mention any correction of the non-uniformity
of the coil, the difference between the present study and that
of Canet et al.’s. report may be due to the non-uniformity of
the coil. Another reason may be the use of a different TR by
Canet, or a different value of R2 was used for finding the
maximum linear relationship between SI and concentration.
Neither of these values was cited by Canet.
Fritz-Hansen et al. [21] found linearity between signal
changes and tracer concentrations up to 1.0 mmol/L
(R2 = 0.999), and an effective TI of 720 ms when using
the IR TurboFLASH sequence (linear phase-encoding)
(TR, time for one FLASH line = 6.5 ms, TE = 3.0 ms,
flip angle = 12�, effective TI varying between 170 and
2000 ms). They did not mention the actual TR. Our image
parameters are slightly different from Fritz-Hansen’s
image parameters. The present study (see Fig. 7) showed
that the maximum linearity was about 1.20 at R2 = 0.99, at
an effective TI of 720. The difference between the present
study and that of Fritz-Hansen et al. may be due to the use
of different TE, TR, and flip angle, or it may be due to the
effect of non-uniformity of the coil [11].
A correlation between the SI and the concentration of
Gd-DTPA on T1-weighted imaging with use of the satu-
ration recovery sequence was also reported by Takeda et al.
[5] and Vallee et al. [22]. Because the image sequence was
different, it is not possible to compare Takeda’s and Val-
lee’s results with ours.
In addition, Dean et al. [23] and Unger et al. [20]
reported that the linearity between SI and concentration
was 1 mmol/L, but they did not mention in detail how they
found this value for comparison with the results of the
present study.
As mentioned above, different papers have reported
different values for the maximum linear relationship
between SI and the concentration of the contrast agent in
T1-weighted imaging. These values were varied between
0.8 and 1 mmol/L concentration for the IR sequence [4, 5,
20, 23].
The linear relationship between SI and the concentration
of the contrast agent was investigated in our previous study
(center out phase-encoding, TR = 2 s) [7]. That study
showed that the linear relationship between SI and con-
centration extended up to 5.34 and 2.46 mmol/L for the
short TI (400 ms) and long TI (800 ms), respectively,
where R2 = 0.95. In addition, these values were reduced to
2.63 and 1.57 mmol/L for these TIs, where R2 = 0.99. The
slightly difference between the present study (see Fig. 5)
and the previous study is due to the use of a different TR.
For finding this linearity, the value of R2 which gives the
strength of the linear relationship between SI and concen-
tration is important. The use of R2 for measuring the lin-
earity is mentioned with different values in different
papers. Takeda et al. [5] accepted the value of 0.76 for the
linearity. Bourke et al. [24] and Fritz-Hansen et al. [21]
also stated that R2 = 0.74 and R2 = 0.999 show a rea-
sonably strong association between SI and concentration,
respectively. In this study, the maximum linearity was
calculated for both R2 = 0.95 and 0.99, which indicates a
far higher degree of linearity.
In summary, our results (see Figs. 5, 7) show that there
is a difference between the maximum linearity of the
concentration and the SI with the different phase-encoding
schemes. The figures show that the maximum linearity with
the linear phase-encoding acquisition was higher than with
the center out phase-encoding acquisition at the same R2.
The difference in values of the maximum linearity between
the two phase-encoding schemes is due to the different
methods of sampling k-space. In particular, the time at
which the center of the k-space is acquired, the linear
phase-encoding acquisition will have experienced multiple
FLASH lines before the center of k-space is reached,
making the image more sensitive to T1 effects compared to
the center out phase-encoding acquisition [18].
In other words, before the initial a pulse in the IR gra-
dient echo TurboFLASH sequence, the protons of the
liquid inside the slice are unsaturated and emitted a very
strong signal. After the initial a, the slice contains saturated
protons and gives a signal lower than that before the initial
a. After the second a, the slice contains more saturated
protons than after the first a and gives a signal weaker than
Effect of phase-encoding scheme
that from the first a. The unsaturated protons give a
stronger signal than the saturated protons. Therefore, the
center out phase-encoding acquisition started from before
the initial a pulse and emits a very strong signal. In addi-
tion, after the n/2th a pulses (n = 128 and a = 15� for this
study), the amplitude of the signal should be constant (the
linear phase-encoding acquisition) and gives a signal
weaker than that from before the first a (the center out
phase-encoding acquisition) [13].
In spite of phase acquisition and R2 = 0.95 or 0.99, at a
typical effective TI = 800 ms, which is normally used for
in vivo perfusion, as at this time the blood has not signal at
1.5 T, the maximum linearity is about twice that previously
reported (i.e., 0.8 mmol/L) for measuring the perfusion
parameters on T1-weighted imaging [2, 25, 26].
Our previous study showed that the results of this study
can be used for in vivo study if all image parameters are the
same as in the in vitro study [27].
5 Conclusion
In conclusion, our previous studies showed that the image
parameters such as TR and TI and image sequences can
have an effect on the maximum linearity between SI and
concentration. This study shows that the phase-encoding
scheme is an important parameter when the SI is measured.
These schemes can have an effect on the maximum
linearity.
Conflict of interest The authors declare that there is no conflict of
interest.
References
1. Hacklander T, Reichenbach JR, Hofer M, Modder U. Measure-
ment of cerebral blood volume via the relaxing effect of low-dose
gadopentetate dimeglumine during bolus transit. Am J Neurora-
diol. 1996;17:821–30.
2. Nazarpoor M. Organ blood flow measurement with T1 and T2*-
weighted MRI techniques (book). Saarbrucken: LAP LAMBERT
Academic Publishing GmbH & Co. KG; 2012. ISBN 978-3-8484-
3648-4.
3. Fritz-Hansen T, Rostrup E, Ring PB, Larsson HBW. Quantifi-
cation of gadolinium-DTPA concentration for different inversion
times using an IR-turbo flash pulse sequence: a study on opti-
mizing multislice perfusion imaging. Magn Reson Imaging.
1998;16(8):893–9.
4. Canet E, Douek P, Janier M, Bendid K, Amaya J, Millet P, Revel
D. Influence of bolus volume and dose of gadolinium chelate for
first-pass myocardial perfusion MR imaging studies. J Magn
Reson Imaging. 1995;4:411–5.
5. Takeda M, Katayama Y, Tsutsui T, Komeyama T, Mizusawa T.
Dose gadolinium-diethylene triamine pentaacetic acid enhanced
MRI of kidney represent tissue concentration of contrast media in
the kidney? In vivo and in vitro study. Magn Reson Imaging.
1994;12:421–7.
6. Nazarpoor M. The effect of repetition time on the maximum
linear relationship between contrast agent concentration and
signal intensity on T1 weighted image using inversion recovery
(IR) sequence. IJR. 2009;6(4):247–52.
7. Nazarpoor M. Effects of inversion and saturation times on rela-
tionships between contrast agent concentrations and signal
intensities of T1-weighted magnetic resonance images. Radiol
Phys Technol. 2010;3:120–6.
8. Nazarpoor M, Poureisa M, Daghighi MH. Comparison of maxi-
mum signal intensity of contrast agent on T1 weighted images
using spin echo, fast spin echo and inversion recovery sequences.
Iran J Radiol. 2013;10(1):27–32.
9. Condon BR, Patterson J, Wyper D, Jenkins A, Hadley DM. Image
non-uniformity in magnetic resonance imaging: its magnitude
and methods for its correction. BJR. 1987;60:83–7.
10. Simmons A, Tofts PS, Barker GJ, Arridge SA. Sources of
intensity non uniformity in spin echo images at 1.5 t. Magn Reson
Med. 1994;32:121–8.
11. Nazarpoor M. Non uniformity of different coils in MRI. In: 8th
Iranian congress of medical physics, Shahid Beheshty University,
Iran, p 105, 28–29 May 2008.
12. Westbrook C, Kaut C. MRI in practice. 2nd ed. Oxford: Black-
well Science Ltd, Oseney Meal; 2000.
13. Nazarpoor M. Inflow effect of signal intensity for the center out
phase-encoding and linear phase-encoding acquisitions on
inversion recovery T1-weighted TurboFLASH images. J Cardio-
vasc Thorac Res. 2009;1(4):29–37.
14. Bernstein MA, King KF, Zhou XJ. Handbook of MRI pulse
sequences. London: Academic Press; 2004.
15. Rohrer M, Bauer H, Mintorovitch J, Requardt M, Weinmann HJ.
Comparison of magnetic properties of MRI contrast media
solution at different magnetic field strengths. Invest Radiol.
2005;40:715–24.
16. McRobbie DW, Moore EA, Graves MJ, Prince MR. MRI from
protons to pictures. Cambridge: Cambridge University Press;
2006. p. 69.
17. NessAiver M, editor. All you really need to know about MRI
physics. 1st ed. Baltimore: University of Maryland Medical
Center; 1997 (chapters 6 and 7).
18. Jivan A, Horsfield MA, Moody AR, Cherryman GR. Dynamic T1
measurement using snapshot-FLASHMRI. J Magn Reson.
1997;127:65–72.
19. Kleinbaum DG, Kupper LL, Muller KE, editors. Applied
regression analysis and other multivariable methods. Belmont:
Duxbury Press/Wadsworth Publishing Company, Inc edition;
1998 (chapter 6).
20. Unger EC, Ugurbil K, Latchaw RE. Contrast agent for cerebral
perfusion MR imaging. J Magn Reson Imaging. 1994;4:
235–42.
21. Fritz-Hansen T, Rostrup E, Ring PB, Larsson HBW. Quantifi-
cation of gadolinium-DTPA concentrations for different inver-
sion times using an IR-turbo FLASH pulse sequence: a study on
optimizing multislice perfusion imaging. Magn Reson Imaging.
1998;8:893–9.
22. Vallee JP, Lazeyras F, Kasuboski L, Chatelain P, Howarth N,
Righetti A, Didier D. Quantification of myocardial perfusion with
FAST sequence and Gd bolus in patients with normal cardiac
function. J Magn Reson Imaging. 1999;19:197–203.
23. Dean BL, Lee C, Kirshch JE, Runge VM, Dempsey RM, Petti-
grew LC. Cerebral hemodynamics and cerebral blood volume:
MR assessment using gadolinium contrast agents and T1-
weighted turbo-FLASH imaging. Am J Neuroradiol.
1992;13:39–48.
24. Bourke GJ, Daly LE, Gilvray JM, editors. Interpretation and uses
of medical statistics. 3rd ed. Oxford: Blackwell Scientific Pub-
lications; 1985.
M. Nazarpoor
25. Martel AL, Moody AR (1998) Assessment of brain perfusion
using parametric and factor images extracted from dynamic
contrast-enhanced MR images. In: Book of abstract (part of SPIE
conference on physiology and function from multidimensional
image), vol 3337, San Diego
26. Moody AR, Martel A, Kenton A, Allder S, Horsfield MA, Delay
G, Morgan P. Contrast-reduced imaging of tissue concentration
and arterial level (CRITICAL) for assessment of cerebral
hemodynamics in acute stroke by magnetic resonance. Invest
Radiol. 2000;35:401–11.
27. Nazarpoor M, Poureisa M, Daghighi MH. Investigations of
optimal dose of contrast agent concentration from routine dose
using spin echo and inversion recovery T1-weighted sequences in
MRI. Med J Tabriz Univ Med Sci Health Serv. 2013;34(5):74–8
(Farsi).
Effect of phase-encoding scheme