ELECTRON SPIN ELECTRON SPIN RESONANCE RESONANCE
Transcript
Page 1: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

ELECTRON SPIN ELECTRON SPIN RESONANCERESONANCE

Page 2: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

DefinitionDefinitionDefinitionDefinition

It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR).

Is a branch of absorption spectroscopy in which radiation of microwave frequency is absorbed by paramagnetic substances.

It was invented by Zavoisky in 1944.

It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR).

Is a branch of absorption spectroscopy in which radiation of microwave frequency is absorbed by paramagnetic substances.

It was invented by Zavoisky in 1944.

Page 3: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

ESR is concerned with magnetic behaviour of spinning electrons.

ESR spectrum is obtained by transition from one spin state to other state of an electron, transition is induced by radiation of microwave frequency.

ESR uses reagents called electron spin reagents, which are stable containing odd electrons have the capacity to react with aminoacids.

ESR of compounds give information about structure, viscosity, polarity, phase transformation and chemical reactivity.

Page 4: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

ESR NMR

Chemical shift is absent.

The transition b/w electrons occurs upon the absorption of a quantum of radiation in microwave region .

The frequency required is 8000MCS-1 .

Limited application due to systems containing unpaired electrons.

Chemical shift is present .

The transition occurs upon the application of radiofrequency.

The frequency required is 40MCS-1

Page 5: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

THEORYTHEORY

The energy levels are produced by the interaction of the magnetic moment of an unpaired electron in a molecule ion with an applied magnetic field.

ESR spectrum is due to transition b/w these energy levels by absorbing radiations of microwave frequency.

The energy levels are produced by the interaction of the magnetic moment of an unpaired electron in a molecule ion with an applied magnetic field.

ESR spectrum is due to transition b/w these energy levels by absorbing radiations of microwave frequency.

Page 6: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

Like a proton, an electron has a spin, which gives it a magnetic property known as a magnetic moment.

When an external magnetic field is supplied, the paramagnetic electrons can either orient in a direction parallel or antiparallel to the direction of the magnetic field

This creates two distinct energy levels for the unpaired electrons.

Page 7: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.
Page 8: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.
Page 9: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

For an electron of spin s=1/2, the spin angular momentum have values of ms= +1/2 or -1/2

• In the absence of magnetic field, the two values give rise to doubly degenerate energy levels(same energy).

Page 10: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

If a magnetic field is applied, this leads to two non-degenerate energy levels. The low energy state will have the spin magnetic moment alligned with the field (ms= -1/2), and

high energy state - opposite to field (ms= +1/2).

Page 11: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.
Page 12: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.
Page 13: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

•These two states possess energies split up from original state with no applied magnetic field by +µe H (low energy state) &

-µeH (high energy state)

Where, H – magnetic field µe – magnetic moment

In ESR, transition b/w different energy levels take place by absorbing quantum of radiaton of ferquency in microwave region

Page 14: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

h Planck’s constant 6.626196 x 10-27 erg.sec frequency (GHz or MHz)g g-factor (approximately 2.0)β Bohr magneton (9.2741 x 10-21erg.Gauss-1)B magnetic field (Gauss or mT)

∆E=hv=gβB

Page 15: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

Thus ESR spectrum of free electron consist of single peak corresponding to transition b/w two levels.

When absorption occurs2 µeH = hv

Where , v - frequency of absorbed radiation (cycles/sec)

The energy of transition is given by ΔE = hv = gβHWhere, β – Bohr’s magneton ( a factor for converting

angular momentum into magnetic moment) g – proportionality factor(a function of electron

environment). Also called spectroscopic splitting factor OR Lande’s splitting factor. The value is not constant

Page 16: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

g is proportionality factor, function of electron’s enviornment,also called as spectroscopic spliting factor or Lande’s spliting factor

g is not constant.

For a free electron , value is 2.0023.In free radicals, value is same(electron character is same).

g is proportionality factor, function of electron’s enviornment,also called as spectroscopic spliting factor or Lande’s spliting factor

g is not constant.

For a free electron , value is 2.0023.In free radicals, value is same(electron character is same).

Page 17: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

ọ In some crystals, g vary from 0.2 to 0.8.The reason is that unpaired electron is localised in a paricular orbital about the atom and the orbital angular momentum couples with spin angular momentum give rise to a low value of g in ionic crystals.

ọ g value depend upon the orientation of the molecule having the unpaired electron w.r.t applied magnetic field.

ọ In a crystal the value of g along x,y,z axes denoted as gx,gy,gz.

ọ In some crystals, g vary from 0.2 to 0.8.The reason is that unpaired electron is localised in a paricular orbital about the atom and the orbital angular momentum couples with spin angular momentum give rise to a low value of g in ionic crystals.

ọ g value depend upon the orientation of the molecule having the unpaired electron w.r.t applied magnetic field.

ọ In a crystal the value of g along x,y,z axes denoted as gx,gy,gz.

Page 18: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

INSTRUMENTATIONINSTRUMENTATION

Page 19: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

Most EPR spectrometers are similar to the original instrumentation used for NMR before the development of pulsed Fourier Transform spectroscopy.

Most also use electromagnets rather than permanent magnets or cryomagnets.

Page 20: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.
Page 21: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.
Page 22: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

1. SOURCE

a) Klystron :• most common source operating in MW band

region of 3 cm λ• Vacuum pump, can produce MW oscillations

centered on small range of frequency• Frequency on monochromatic radiation

determined by applied voltage to Klystron• Operated at 9500 Mc/sec.

Page 23: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

b. Isolator• A non-reciprocal device, a strip of ferrite material• Minimises vibrations in frequency of MW produced by

Klystron oscillator• Variation in frequency due to backward reflection in

region b/w Klystron & circulator

c. Wavemeter• b/w isolator & attenuator to know frequency of MW

produced by Klystron oscillator• Calibrated in frequency units

d. Attenuator• b/w wavemeter & circulator, adjusts the level of MV

power incident upon sample• Possesses an absorption element

Page 24: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

2. CIRCULATOR / MAGIC T• MW radiations enter circulator through a wave

guide by a loop of wire that couples with oscillating magnetic field & sets up a corresponding field in wave guide

• Made up of hollow rectangular copper or brass tubing with silver or gold plating inside to produce a highly conducting flat surface

Page 25: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

3. SAMPLE CAVITY

• Cavity system is constructed in such a way as to maximise applied magnetic field along the sample dimension

• 0.15 – 0.5 ml sample can be used (less DEC)

• Flat cells (0.25 mm thickness) used for samples (high DEC) – 0.05 ml sample

• Most ESR spectrometers use dual sample cavities for sample and reference

Page 26: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

4. MAGNET SYSTEM

•A static magnetic field is provided by an electromagnet with a current- regulated power supply.

•A homogeneous and stable field is required for best results.

•The resonant cavity is placed b/w pole pieces of electromagnet

•A Hall probe, driven from a stable constant-current power system, with a digital multimeter (DMM) reading the Hall voltage, is used to measure the value of the magnetic field between the poles of the magnet

Page 27: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

5. CRYSTAL DETECTORS

Silicon crystal is most commonly used detectorAct as a microwave rectifierConvert MW power into direct current output

6. AMPLIFIER AND PHASE SENSITIVE DETECTORThe signal from crystal detector undergo narrow-band amplification by auto amplifierNoise present in the amplified signal is removed by phase sensitive detector

7. OSCILLOSCOPE AND PEN RECORDER

Page 28: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

WORKING• Sample is kept in resonant cavity• The cavity is a long path-length cell in which wavas

are reflected to & fro thousand of times• Klystron oscillator produce MW that pass through

isolator, wave meter and attenuator to reach circulator through arm 1

• MW power divide b/w arms 2 & 3 (sample cavity)• Arm 2 have balancing resistance

• When total resistance on 2 & 3 are same, MW power is absorbed completely

• When R on arm 3 changes due to some ESR resonance absorption by the sample, then some MW power enter into detector through arm 4

Page 29: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

• detector convert the MW power into DC

• If the magnetic field around the sample is changed to the value required for resonance, absorption takes place by the sample

• The current is recorded by the recorder and show an absorbance peak

• If main magnetic field is swept slowly for several minutes, recorder shows the derivative of the absorption spectrum

• generally absorption spectra are recorded as first derivative spectra

Page 30: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.
Page 31: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

PRESENTATION OF ESR SPECTRUM

•ESR spectrum is obtained by plotting intensity against the strength of magnetic field•ESR spectrum is better represented by derivative curve•First derivative (slope) of absorption curve is plotted against strength of magnetic field•Each negative slope in derv. curve represents a peak or shoulder in absorption spectrum•Every crossing of the derv. axis with a negative slope indicates a true maximum and crossing with positive slope indicates minimum•Thus no. of peaks/shoulders in absorption curve can be determined from no. of minima or maxima in the derivative curve.

Page 32: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

• Area covered by either the absorption or derivative curve is proportional to no of unpaired electrons in the sample

• To calculate no of electrons in the sample, comparison is made with standard sample having known no of unpaired electrons and possessing the same line shape

• Most widely used std is 1,1-diphenyl-picrylhydrazyl (DPPH) free radical

Page 33: ELECTRON SPIN RESONANCE. DefinitionDefinition It is also called electron paramagnetic resonance (EPR) or electron magnetic resonance (EMR). Is a branch.

DPPH is chemically stableSplitting factor g = 2.00361,53 * 1021unpaired electrons per gm