Date post: | 20-Jan-2016 |
Category: |
Documents |
Upload: | tyrone-pearson |
View: | 222 times |
Download: | 5 times |
Enhancement of Single Stretched Wire Measurements
of LHC Short Straight Sections
Guy Deferne, Nikolay Smirnov, CERN
Joe DiMarco, FNAL
14th International Magnetic Measurement Workshop26-29 September 2005, Geneva, Switzerland
28.09.2005 IMMW14 2
Content
Introduction – The Single Stretched Wire (SSW) at CERN Part I - Roll Angle Offset Calibration
Introduction Case of a centered magnet Case of a non-centered magnet Example of calibration
Part II - Integrated Strength (Gdl) Measurements For more details, please see the presentation of N. Smirnov et al, « Focusing strength measurements of the main quadrupoles for the LHC », submitted at the MT19 Conference, 2005)
Introduction Shape of the wire Gdl Measurement procedure Magnetic properties of the wire System performance
Conclusion
28.09.2005 IMMW14 3
• The Single Stretched Wire (SSW) are the reference systems used for the integrated strength and field direction measurements
– Three systems operational– All three are similar to the
ones used at DESY and at FNAL
One of the two stages
System installed on a cold bench
Introduction
The SSW features used at CERN:
• Warm and cold integrated axis of MQ and associated correctors• Integrated strength of MQ and MB at cold• Warm and cold integrated roll angle of MQ, correctors and MB
Nearly 500 Short Straight Section (SSS) for the LHC. All of them are an assembly of different magnets 10 % of those must be magnetically measured
28.09.2005 IMMW14 4
Roll angle offset calibration (1) Introduction
The field direction is one of the essential parameters of the measurement campaign
The SSW systems must be periodically calibrated in sense of roll angle in order to perform fast measurements that fulfill the requirements
Two different type of calibration can be used, depending if the measured magnet is centered or not in between the stages
Part I – Roll angle offset calibration
28.09.2005 IMMW14 5
aField (normal offset) (swapped offset)
2
Roll angle offset calibration (2)
Stages are aligned w/r to gravity with 5µrad
resolution Wyler inclinometers
etSystemOffsaswappednormal
measured
2
)(
zA
zB
offsetstageA
offsetstageB
αgravity
αgravity
αfield αgravity
Stage A
Magnet
Stage B
If zA=zB AND αfield=αgravity (reference magnet), then
Swapping stages
• Introducing the system offset in the calculation allows to perform absolute angle measurement without swapping the stages
Case of a centered magnet
If zA=zB, then
Absolute field direction w/r to gravity:
System offset:
28.09.2005 IMMW14 6
On cold benches, the magnet is NOT centered in between the stages
• Inside an SSS, the different magnets are located at different longitudinal positions
Layout of the magnets inside an SSS
Stage AStage B
MQ
MSCB
MO/MQT
2.94m
4.975m
6.849m
11.68m
Roll angle offset calibration (3) Case of a non-centered magnet (1)
28.09.2005 IMMW14 7
Roll angle offset calibration (4)
• Simply swapping the two stages as for a regular absolute offset calibration is not sufficient; the variation of the offset along the longitudinal position must be taken into account
• During the calibration, one of the stage is placed closer to the magnet
A stage
B stage
Reference quadrupole
SSW stages installed on the calibration bench
Case of a non-centered magnet (2)
If zA≠zB, then
where: z is the distance from stage A to the magnet center and
l is the distance between the stages
Offset(z) normal (1z
l) swapped
z
l
28.09.2005 IMMW14 8
Roll angle offset calibration (5)
System offset vs longitudinal position
-0.4
-0.2
0
0.20.4
0.6
0.8
1
1.2
2 3 4 5 6 7
Z distance from stage A [m]
Ro
ll a
ng
le o
ffs
et
[mra
d]
SSW1_09.2004
SSW1_11.2004
SSW2_09.2004
SSW2_01.2005
• The roll angle offset parameter is updated for each type of magnet, in each SSW system.
Example of calibration
28.09.2005 IMMW14 9
Integrated Strength (Gdl) (1)
• A second essential parameter of the measurement campaign is the integrated gradient (Gdl) of quadrupoles
• This is one of the most challenging magnetic parameters of the LHC magnets
– Required absolute error (Magnetic measurement): 5 [units]– Required repeatability (Magnetic measurement): < 5 [units]
• For the Stretched Wire Systems, this means an error on the wire positioning less than 2.5µm over 12m!
Introduction
Only a proper method and good properties of the wire can give results inside specifications
Part II – Integrated Strength (Gdl) measurement
28.09.2005 IMMW14 10
Integrated Strength (Gdl) (2) Shape of the wire (1)
The position of the wire at the stages is known within 1 µm,
but inside the magnet, the wire is deflected by magnetic forces and gravity
Therefore, the weight of the wire and its magnetic properties must be taken into account.
Magnetic forces
Gravity
Stretched wire
)()(
)()(
2
2
2
2
zFz
zXT
zFwgz
zYT
magx
magy
Vertical position of the wire:
Horizontal position of the wire:
• T is the wire tension
• w is the mass per unit of length
• Fmag is the magnetic force.
Where:
Note: the sign of the force Fmag depends on the magnetic properties of the wire and its location in the field
)()()( zYzYzY maggravity )()( zXzX mag
)(2 wiregravity lzzT
wgY
)(2 0
22
zT
DGorXY magmag
with
• z is the longitudinal position
• w is the wire weight per unit of length
• G the gradient of the field
• D the wire displacement
• is the wire susceptibility
• T is the wire tension
Where:
Possible solutions of (1):
(1)
28.09.2005 IMMW14 11
Note:
Integrated Strength (Gdl) (3) Shape of the wire (2)
Sag of the wire:2
8wirel
T
wgSag
As tension measurement is affected by friction problems and gauge accuracy,
the SSW system measures the fundamental frequency of the wire,
which includes its mechanical properties
w
T
lf
wire2
1
Shape of stretched wire with / without magnetic forces
Z
Y p
os
itio
n
Magnetic Field
Mag Force
Mag Force
B stageA stage
28.09.2005 IMMW14 12
Integrated Strength (Gdl) (4) Gdl measurement procedure
-70
-68
-66
-64
-62
-60
-58
-56
-54
0 0.002 0.004 0.006 0.008
1/f**2 (s**-2)
Inte
gra
ted
Tra
sfe
r F
un
ctio
n [
T/k
A] TF @ 0.76kA (injection)
TF @ 5kA (intermediate)TF @ 11.85kA (nominal)
Dependence of transfer function as a function of wire tension with CuBe wire 0.13mm diam
• The Gdl obtained from a horizontal movement, can be calculated from the linear fit of different wire tensions.
• The Gdl obtained from a vertical movement must be calculated from the parabolic fit of different wire tensions.
The parabolic term of the polynomial expansion comes from the sag of the wire:
22 )( SagLGF stepmagy
,
To cancel the effect of the magnetic forces, the value of the strength is calculated from the extrapolation of Gdl = f(1/f2) at the interception of the axis, where 1/f2 0 (infinite tension)
1
f 2
s2
Note:
28.09.2005 IMMW14 13
Integrated Strength (Gdl) (5) Magnetic properties of the wire
Four different wires have been tested:
1. 0.1mm CuBe wire from California Fine Wire Co, USA
2. 0.1mm Mg wire from California Fine Wire Co, USA
3. 0.13mm CuBe wire from Goodfellow Co, UK
4. Carbon fiber strand from Toho Tenaz Europe gmbh, type HTA5241
(5). ( 0.078mm silicon carbide, type SCS-9A from Speciality Materials Co, USA)
(Note: type 5. could not be magnetically tested because of its high rigidity)
Wire 0.76kA 5kA 11.85kA χ
CuBe 0.1mm 30.4 2000 9480 >0
Mg 0.1mm 6.1 500 4977 >0
CuBe 0.13mm 2.3 50 474 <0
Multi filament Carbon strand
- - 380 <0
Slopes of strength for different types of wire in [T/s2]
CuBeCarbon fiber HTA5241
SCS-6 Silicon Carbide
Different type of tested wire
Note:
• if strength rises when tension increases, wire is diamagnetic
• If strength falls when tension increases, wire is paramagnetic
28.09.2005 IMMW14 14
0 1 2 3 4 5 6 7 8 9
1 0
0 2 0 0 4 0 0 6 0 0 8 0 0 I n t e g r a t e d F i e l d G r a d i e n t [ T ]
D(G
dl)
units
H y p e r b o l a E x p e r i m e n t a l r e s u l t s P a r a b o l a
The error on the extrapolation is proportional to the slope
The lower is the susceptibility, the smaller is the random error on Gdl
Integrated Strength (Gdl) (6) System performance (1)
Gdl
1/T
1. Random error
Typical random error vs Gdl
3)()/()( GdlGdlGdl snn
Gdl stdev at interception point
Total random error
Contribution of the slope (magnetization of the wire)
Qualification of two SSW systems
Parameter/SSWUnits SSW#1 SSW#2
δnoise units*T 210 400
ηslope units*T-3 2.5*10-9 5*10-9
Gdl lower acceptable limit (5units at one sigma)
T 42 80
As Gdl at injection is 44T, its measurement can be performed only with certain statistic
Inje
ctio
n
Nom
inal
Tolerance limit
Dev
(Gdl
) [u
nits
]
28.09.2005 IMMW14 15
Integrated Strength (Gdl) (7) System performance (2)
2. Systematic errors
Estimated contribution
Powering of the magnet: <0.1 [units]
Amplifier and integration: <0.1 [units]
Main field errors: <0.2 [units]
Stray field (on cold test benches): 2.3 [units]
Alignment of stages: <1 [units]
Wire susceptibility: <2 [units]
____
Total (rms) <3.2 [units]
A cross calibration with a system that uses a different method of Gdl measurement could give a valuable confirmation of the systematic error of the SSW systems performances
The twin shafts and the automated scanner (both are rotating coils systems) used in SM18 cannot guaranty so far better than 20 [units] of absolute error
Thus an estimation of the all possible and known sources of systematic error is used to qualify the SSW systems at CERN
28.09.2005 IMMW14 16
Conclusion
•A special method has been developed to measure the integrated strength (Gdl) of the LHC MQ with an accuracy within the specification (even though a statistical study is required at injection field)
•Gdl measurements:Absolute
RandomError specifications: 5 [units] 5 [units] Error obtained: est. <3.2 [units] <5 [units]
• Different wires have been tested for their magnetic and mechanical properties and one type, the 0.13mm diam. CuBe from Goodfellow Ltd, has been chosen for the LHC measurements
• As we have shown that not any wire can be used for Gdl measurements, the search for wires with a lower susceptibility will continue
•A procedure has been developed to calibrate the roll angle offset of the Single Stretched Wire systems as a function of the magnet longitudinal position.
•This allows to measure the field angle w/r to gravity of the main quadrupoles and correctors magnets installed in an Short Straight Section without any swapping of the system stages.
Roll Angle Calibration
Integrated Strength