EPI – Echo Planar Imaging. Joakim Rydell 2004-03-26. Echo Planar Imaging. Like ordinary MR, but faster Ordinary MR: approximately one minute EPI: less than one second The Nobel Prize in Medicine 2003 Lauterbur (MR) Mansfield (EPI). Repetition, k-space. Fourier transform of image - PowerPoint PPT Presentation
Popular Tags:
19
EPI – Echo Planar Imaging Joakim Rydell 2004-03-26
Transcript
Page 1: EPI – Echo Planar Imaging

EPI – Echo Planar Imaging

Joakim Rydell

2004-03-26

Page 2: EPI – Echo Planar Imaging

Echo Planar Imaging

Like ordinary MR, but fasterOrdinary MR: approximately one minuteEPI: less than one second

The Nobel Prize in Medicine 2003Lauterbur (MR)Mansfield (EPI)

Page 3: EPI – Echo Planar Imaging

Repetition, k-space

Fourier transform of image

Traversed using magnetic field gradients

Page 4: EPI – Echo Planar Imaging

Repetition, pulse sequence

Slice selection Phase encoding Frequency encoding

Page 5: EPI – Echo Planar Imaging

Introducing EPI

Acquire all rows in k-space with just one excitation

EPI

Ordinary MR

Page 6: EPI – Echo Planar Imaging

EPI – The Movie

Page 7: EPI – Echo Planar Imaging

Going even faster – Partial Fourier

Real image hermitian symmetric k-space Acquire half of k-space, compute the rest Not always as good as it seems...

Page 8: EPI – Echo Planar Imaging

Partial Fourier, drawbacks

Phase errorsSlowly varyingApproximately 65 % of k-space required

SNR decreaseFull k-space acquisition gives sqrt(2) SNR

improvement

Page 9: EPI – Echo Planar Imaging

Why use ordinary MR?

Signal decay Low SNR N/2 ghosting Geometric distortion dB/dt

Page 10: EPI – Echo Planar Imaging

Signal decay

Long readout period Weak signal near end

of acquisition Blurring in phase-

encoding direction

Page 11: EPI – Echo Planar Imaging

Low SNR

Fast acquisition Less averaging Noisy images

Page 12: EPI – Echo Planar Imaging

N/2 ghosting

Alternating acqusition directions

Shifted sampling positions

N/2 ghosting Reference scans

Page 13: EPI – Echo Planar Imaging

Geometric distortion

Low bandwidth/pixel Pixels shifted in

phase encoding direction

Field maps

Page 14: EPI – Echo Planar Imaging

dB/dt

Fast gradient switching induces eddy currents, causing field inhomogenity

Rapid changes in the magnetic field can cause peripheral nerve stimulation

Page 15: EPI – Echo Planar Imaging

k-space trajectories

Cartesian Blurring N/2 ghosting High dB/dt

Spiral Non-uniform sampling

Cartesian spiral

Page 16: EPI – Echo Planar Imaging

Ramp sampling

Sampling on gradient ramps

Non-uniform sampling In time In k-space

Page 17: EPI – Echo Planar Imaging

Multishot

Both EPI and ordinary imaging Collect several rows per excitation Alignment difficulties

Page 18: EPI – Echo Planar Imaging

Applications

Cardiac imaging fMRI ...

Page 19: EPI – Echo Planar Imaging

EVI – Echo Volumar Imaging

3D EPI Acquire a full volume

with just one excitation

Not quite possible yet