+ All Categories
Home > Documents > EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample...

EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample...

Date post: 26-Dec-2015
Category:
Upload: miranda-owen
View: 218 times
Download: 4 times
Share this document with a friend
Popular Tags:
24
EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization measurements : Applications to paleo- and archeointensity determinations
Transcript
Page 1: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

EPSL, 2004, 229/1-2 pp. 31-43.

Maxime LE GOFF and Yves GALLET

A new three-axis vibrating sample magnetometer for continuous high-temperaturemagnetization measurements :

Applications to paleo- and archeointensity determinations

Page 2: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

Paleo- and archeointensity studiesare fastidious and often unproductive

due to poor data quality

It is possible to make them fasterby measuring the magnetization

at high temperatures

Page 3: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

TRIAXE SENSOR

Page 4: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.
Page 5: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

must take into account the thermal dependence of the spontaneous

magnetization

High-temperature version of the Thellier and Thellier method revised by Coe

Page 6: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

0

100

200

300

400

500

600

700

800

900

0 50 100 150 200 250 300 350 400 450 500

Temperature (°C)

Mto

tal (

10

-8 A

m²)

NRM

T1 = 148°C

H = 0

High-Temperature Magnetization Measurements

Heating Troom to T1 (H=0)

Page 7: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

0

100

200

300

400

500

600

700

800

900

0 50 100 150 200 250 300 350 400 450 500

Temperature (°C)

Mto

tal (

10

-8 A

m²)

H = 0NRM

T1 = 148°C

T2 = 448°CCurve #1

High-Temperature Magnetization Measurements

Heating T1 to T2 (H=0)

Page 8: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

0

100

200

300

400

500

600

700

800

900

0 50 100 150 200 250 300 350 400 450 500

Temperature (°C)

Mto

tal (

10

-8 A

m²)

H = 0NRM

T1 = 148°C

T2 = 448°C

Curve #1

Curve #2

Curve #3

High-Temperature Magnetization Measurements

Cooling to T1 andHeating T1 to T2 (H=0)

Page 9: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

0

100

200

300

400

500

600

700

800

900

0 50 100 150 200 250 300 350 400 450 500

Temperature (°C)

Mto

tal (

10

-8 A

m²)

H = 70µT

H = 70 µTNRM

T1 = 148°C

T2 = 448°CCurve #1

Curve #2

Curve #3

High-Temperature Magnetization Measurements

Applying Hlab = 70µT

Page 10: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

0

100

200

300

400

500

600

700

800

900

0 50 100 150 200 250 300 350 400 450 500

Temperature (°C)

Mto

tal (

10

-8 A

m²)

NRM

T1 = 148°C

T2 = 448°C

Curve #1

Curve #2

Curve #3

Curve #4

High-Temperature Magnetization Measurements

Cooling T2 to T1 (H = 70µT)

Page 11: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

0

100

200

300

400

500

600

700

800

900

0 50 100 150 200 250 300 350 400 450 500

Temperature (°C)

Mto

tal (

10

-8 A

m²)

H = 0NRM

T1 = 148°C

T2 = 448°C

Curve #1

Curve #2

Curve #3

Curve #4

High-Temperature Magnetization Measurements

Hlab = 0µT

Page 12: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

0

100

200

300

400

500

600

700

800

900

0 50 100 150 200 250 300 350 400 450 500

Temperature (°C)

Mto

tal (

10

-8 A

m²)

NRM

T1 = 148°C

T2 = 448°C

Curve #1

Curve #2

Curve #3

Curve #4Curve #5

High-Temperature Magnetization Measurements

Heating T1 to T2 (H=0)

Page 13: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

R(Ti) = Hlab * 1(Ti) / 5(Ti)

R’(Ti) = Hlab * 1’(Ti) / 5’(Ti)

Remaining NRM (1) and TRM (5) fractions between Ti and T2

R ratio

Lost NRM (1’) and TRM (5’) fractions between T1 and Ti

R’ ratio

0

50

100

150

200

250

300

350

400

100 150 200 250 300 350 400 450

TiT1 = 150°C

T2 = 450°C

Two possibilities to derive intensity data

NRM

TRM

Page 14: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

20

30

40

50

60

70

80

150 200 250 300 350 400 450

Temperature (°C)

R, R

' (µ

T)

R

R'

Test on a pseudo NRM acquired in a field of 50µT in the same conditions as the laboratory TRM (Hlab=50 µT)

Page 15: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

0

10

20

30

40

50

60

70

80

150 200 250 300 350 400 450

R

R'

Ancient NRM : R increasing, R’ much more constant

Ti, (°C)

R, R’ (µT)

Page 16: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

45

50

55

60

65

150 200 250 300 350 400 450

Temperature (°C)

R,

R' (

µT

)

NRM 0.5°/mn; TRM 25°/mn

NRM 0.5°/mn, TRM 6°/mn

NRM 0.5°/mn, TRM 2°/mn

Cooling rate effect

R’ must be considered

Page 17: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

Correction for the TRM anisotropy

Page 18: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

Comparison between archeointensity

data obtained from the Thellier and

Thellier method revised by Coe and

from the Triaxe

(Samples from Mesopotamia previously studied by Genevey et al., JGR, 2003)

Page 19: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

0

10

20

30

40

50

60

70

80

150 200 250 300 350 400 450

R

R'

0

50

100

150

200

250

150 250 350 450

Selection criteria

• A large NRM fraction must be involved

• Expected behaviour of R and R’ curves

Page 20: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

0

20

40

60

80

100

120

140

200 250 300 350 400 450 500

Temperature (°C)

R

R'

Example of rejected sample

Page 21: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

MR 15

10

20

30

40

50

60

70

150 200 250 300 350 400 450

Temperature (°C)

R' (

µT)

Archeointensity results from a site of baked bricks from Mesopotamia (Mari, Syria)

6 samples

Page 22: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

0

10

20

30

40

50

60

70

80

-4000 -3000 -2000 -1000 0

Thellier Thellier - Coe

Triaxe Same Site

Triaxe New Site

Comparison between archeointensity results obtained from the Thellier and Thellier and Triaxe methods

H (µT)

Age (BC)

Page 23: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

CONCLUSIONS

• The experimental procedure was successfully tested with archeological baked materials. We need now to work on volcanic samples.

• Other applications of the Triaxe :- Viscosity at high temperature.- Alteration.- Coupling between different magnetic phases.- etc.

Page 24: EPSL, 2004, 229/1-2 pp. 31-43. Maxime LE GOFF and Yves GALLET A new three-axis vibrating sample magnetometer for continuous high-temperature magnetization.

Recommended