Ethylene Oxide ResidualsAn Alternative Testing
Method
Presented to:U.S. Technical Advisory Group for ISO/TC 198
April 24, 2007
0
.01
.02
.03
.04
.05
4000 3500 3000 2500 2000 1500 1000
Drivers for Alternative MethodsDrivers for Alternative Methods
Meet new standard requirements for Meet new standard requirements for
1099310993--7 & 109937 & 10993--1818
Research alternative methodologies Research alternative methodologies
for Chemical Characterizationfor Chemical Characterization
Determine compliant, cost effective Determine compliant, cost effective
“One Shot Analysis”“One Shot Analysis”
Seek method that minimizes product Seek method that minimizes product
consumptionconsumption
ReviewReviewEquipment is QualifiedEquipment is Qualified
Validating test method for the analysis of Validating test method for the analysis of
EO/ECH/EGEO/ECH/EG
Publishing in scientific journal this yearPublishing in scientific journal this year
Evaluate & develop other applicationsEvaluate & develop other applications
�� Currently validating a method for the Currently validating a method for the
analysis of water content in polymers to analysis of water content in polymers to
assist with establishing manufacturing assist with establishing manufacturing
specificationsspecifications
Absorption of Infrared RadiationAbsorption of Infrared Radiation
FTIR TechnologiesFTIR Technologies
Absorbance / Wavenumber (cm-1) Y-Zoom CURSOR
File # 2 : 71_4H2O 3/23/2007 10:44 AM Res= 4.000
Water in G.5 N2, Cosameter 74.57 ppm -3.16 ppm BG= 71.41 ppm, 10m,0.981 atm, 121C,Avg files 34-36 BAL
0
.05
.1
3000 2500 2000 1500 1000
Gas Phase FTIR
Typical Bench Top FTIR
High Resolution vs. Low ResolutionHigh Resolution vs. Low Resolution
Absorbance / Wavenumber (cm-1) Y-Zoom CURSOR
File # 2 : 71_4H2O
0
.02
.04
.06
.08
2000 1900 1800
High Resolution (0.5 cm-1)
Low Resolution (4 cm-1)
Fourier Transform Fourier Transform
(Time Domain to Frequency)(Time Domain to Frequency)
Interferogram (Time Domain Spectrum)Interferogram (Time Domain Spectrum)
-20000
-10000
0
10000
500 400 300 200 100
Transformed InterferogramTransformed InterferogramFrequency DomainFrequency Domain
Photon intensity at the detector (YPhoton intensity at the detector (Y--axis) at each wavenumber cmaxis) at each wavenumber cm--1 (X1 (X--axis)axis)
�� Single beam spectrumSingle beam spectrum
Arbitrary Y / Wavenumber (cm-1) Y-Zoom CURSOR
10000
20000
30000
40000
3500 3000 2500 2000 1500 1000
Water Absorbance (specific frequencies)
Arbitrary Y / Wavenumber (cm-1) Y-Zoom CURSOR
10000
20000
30000
40000
3500 3000 2500 2000 1500 1000
Single Beam (SBBKG)
Single Beam (SBSample)
Transmission SpectrumTransmission Spectrum
Transmission Spectrum of Water Transmission Spectrum of Water �� Simple intensity ratio (SBSimple intensity ratio (SBsamplesample/SB/SBbg)bg)
�� Scales logarithmically with concentrationScales logarithmically with concentration
Transmittance / Wavenumber (cm-1) Y-Zoom CURSOR
70
80
90
100
3500 3000 2500 2000 1500 1000
Transmittance:Transmittance:
oI
IT =
Absorbance SpectrumAbsorbance Spectrum
Absorbance of Spectrum of Water Absorbance of Spectrum of Water
�� Absorbance = Log 10 1/TransmissionAbsorbance = Log 10 1/Transmission
�� Scales linearly with concentrationScales linearly with concentration
Absorbance / Wavenumber (cm-1) Y-Zoom CURSOR
File # 5 : 23_9H2O 11/11/2006 8:45 PM Res=None
0
.02
.04
.06
3500 3000 2500 2000 1500 1000
Absorbance:Absorbance:
I
IlogA
o=
Beer’s Law SummaryBeer’s Law Summary
Ai = ai * b * ci
Ai = Absorption at a given frequency o the ith sample component
ai = Absorption coefficient (absorptivity) of the ith sample component
b = Pathlength of cellci = Concentration of ith sample
A = log10 (1/T) = -log10 TA = AbsorbanceT= Transmittance
Absorption CoefficientAbsorption Coefficient
& Pathlength& Pathlength
Certified gases are used to generate single Certified gases are used to generate single
component reference standardscomponent reference standards
The The absorption coefficientabsorption coefficient is a property of a is a property of a
material and it defines the extent to which a material and it defines the extent to which a
material absorbs energy material absorbs energy
�� Affected by TemperatureAffected by Temperature
�� Affected by PressureAffected by Pressure
Path length is fixed and calibratedPath length is fixed and calibrated
Innovative New Alternative toInnovative New Alternative to
Existing Technologies ….Existing Technologies ….
Detection of Acids, Bases, and VolatilesDetection of Acids, Bases, and Volatiles
Identification and QuantitationIdentification and Quantitation
Fast collection and analysis time Fast collection and analysis time –– Get Product Get Product
to Market Fasterto Market Faster
Provides Low Limits of Detection required by Provides Low Limits of Detection required by
GuidelinesGuidelines
Static Headspace FTIRStatic Headspace FTIR
FTIR Static Headspace (SHS):Outgassing Profile
FITR Spectra
Co
nce
ntr
atio
n
Ab
sorb
ance
Unit
s
Time
FTIR Static HS Gas Cell
Total Cumulative
Outgassing vs Time
SAMPLE
IR Source
IR Detector
FTIR Static Headspace (SHS):Outgassing Profile
FITR Spectra
Co
nce
ntr
atio
n
Ab
sorb
ance
Unit
s
Time
FTIR Static HS Gas Cell
Total Cumulative
Outgassing vs Time
SAMPLE
IR Source
IR Detector
FTIR Static Headspace (SHS):Outgassing Profile
Co
nce
ntr
atio
n
Ab
sorb
ance
Unit
s
Time
FTIR Static HS Gas Cell
Total Cumulative
Outgassing vs Time
IR Source
IR Detector
FTIR Static Headspace (SHS):Outgassing Profile
FITR Spectra
Co
nce
ntr
atio
n
Ab
sorb
ance
Unit
s
Time
FTIR Static HS Gas Cell
Total Cumulative
Outgassing vs Time
SAMPLE
IR Source
IR Detector
FTIR Static Headspace (SHS):Outgassing Profile
FTIR Static HS Gas Cell
Total Cumulative
Outgassing vs Time
FITR Spectra
Co
nce
ntr
atio
n
Ab
sorb
ance
Unit
s
Time
SAMPLE
IR Source
IR Detector
FTIR Static Headspace (SHS):Outgassing Profile
FTIR Static HS Gas Cell
Total Cumulative
Outgassing vs Time
FITR Spectra
Co
nce
ntr
atio
n
Ab
sorb
ance
Unit
s
Time
SAMPLE
IR Source
IR Detector
Multiple Regions for AnalysisMultiple Regions for Analysis
0
.02
.04
.06
.08
.1
980 960 940 920 900 880 860 840
0
.01
.02
.03
.04
.05
3100 3000 2900 2800
0
.01
.02
.03
.04
.05
4000 3500 3000 2500 2000 1500 1000
0
.005
.01
.015
1300 1200 1100
Ethylene Oxide Reference Spectrum
0
50E-05
.001
.0015
.002
.0025
4000 3500 3000 2500 2000 1500 1000
Paged Z-Zoom CURSOR
0
10E-05
20E-05
30E-05
40E-05
4000 3500 3000 2500 2000 1500 1000
Paged Z-Zoom CURSOR
0
10E-05
20E-05
30E-05
40E-05
4000 3500 3000 2500 2000 1500 1000
Pag ed Z-Zoom CU RSOR
Reference Spectra of Reference Spectra of EOEO//ECHECH//EGEG
EO Reference
Spectrum
ECH Reference
Spectrum
EG Reference
Spectrum
Calibration Reference SpectraCalibration Reference Spectra(Prediction Model)(Prediction Model)
-2.2720.0353.52654.7755EO_H
0026.2926.2926EO_H
-4.4550.01213.70114.3414EO_H
8.3160.0128.6768.018EO_H
2.1430.0125.3835.275EO_HSPC
Prediction %Error (ppm)
SEC (ppm)
Predicted Conc (ppm)
Actual Conc (ppm)Filename
CLS Prediction Model LinearityCLS Prediction Model Linearity
Linearity of EO at 121 oC
y = 0.9821x
R2 = 0.9993
0
10
20
30
40
50
60
0 10 20 30 40 50 60
Actual Concentration (PPMv)
Pre
dic
ted
Co
nc
en
tra
tio
n (
PP
Mv)
AnalysisAnalysisReference spectra Reference spectra
�� Prepared on the instrument used for analysis but can be transferPrepared on the instrument used for analysis but can be transferred to red to other instrumentsother instruments
�� Recorded at the experimental temperature / conditionsRecorded at the experimental temperature / conditions
�� Concentration is measured using second technique or certified Concentration is measured using second technique or certified standardstandard
�� Multiple concentrations are prepared to bracket the experiment rMultiple concentrations are prepared to bracket the experiment range ange
Reference spectra are implemented into a Classical Least SquaresReference spectra are implemented into a Classical Least Squares (CLS) (CLS) fit routinefit routine
Sample Spectra are quantified by comparison to Single Component Sample Spectra are quantified by comparison to Single Component Reference spectra using CLSReference spectra using CLS
Multiple Species can be quantified simultaneouslyMultiple Species can be quantified simultaneously
�� Analysis regions can be chosen to avoid interferences Analysis regions can be chosen to avoid interferences
�� Multiple regions can be chosen to circumvent saturated absorbancMultiple regions can be chosen to circumvent saturated absorbance e bands that may deviate from Beer’s Law bands that may deviate from Beer’s Law
IDEAL GAS LAWIDEAL GAS LAW
PV = nRTPV = nRT
Cell Volume = 5.7 LCell Volume = 5.7 L
Pressure = 1 atmPressure = 1 atm
Molecular Weight = 44 g/molMolecular Weight = 44 g/mol
Cell Temperature = 121 Cell Temperature = 121 ooCC
Mass(g) Sample))CoTemp( Cell(273
Kmol
atmL 0.08206
ol)Weight(g/mMolecular tm)Pressure(aVolume(L) Cellion(PPMv)Concentrat
g
g
••
•••
+
=
µ
•
•
Recovery of EO SpikesRecovery of EO Spikes
7.4% RSD
100.6Average Recovery
92.68.7
116.711.2
96.69.8
99.110.6
103.029.5
103.030.3
102.029.9
101.131.3
100.020.9
99.719.3
107.319.1
86.115.7
% RecoveryMeasured Concentration
(PPMv)
OutgassingOutgassing rates @ rates @
121121ooC (Exhaustive) and C (Exhaustive) and
37C (Simulated use)37C (Simulated use)
EO Release from Medical Device 121oC
0
5
10
15
20
25
0 5 10 15 20
Elapsed Time (Hours)
PP
Mv
121C EO
37C EO
Exhaustive Outgassing of Exhaustive Outgassing of
1 Medical Device1 Medical Device
Outgassing of EO @ 121C
0
5
10
15
20
25
0.0 5.0 10.0 15.0 20.0 25.0 30.0
Elapsed Time (Hours)
EO
PP
Mv
Nitrogen purge
Max PPMv @ ~ 13 hours
1.5 hrs static after purge max PPMv = 0.6 ppmv
EO Spike in Biodegradable EO Spike in Biodegradable
Polymer MatrixPolymer Matrix
0
.1
.2
.3
.4
.5
.6
3500 3000 2500 2000 1500 1000
Absorbance / Wavenumber (cm-1)
Water
absorbance
Carbon Dioxide
absorbance
CH4 Stretch of Polymer
Fingerprint absorbance of
Polymer
Spectrum of Biodegradable Polymer OutgassingSpectrum of Biodegradable Polymer Outgassing
EO Spike in Biodegradable EO Spike in Biodegradable
Polymer MatrixPolymer Matrix
Polymer outgassing spiked with EO Polymer outgassing spiked with EO
0
.01
.02
.03
.04
.05
3500 3000 2500 2000 1500 1000
Absorbance / Wavenumber (cm-1)
EO Spike in Biodegradable EO Spike in Biodegradable
Polymer MatrixPolymer Matrix
ResidualResidual Spectrum after Subtraction of Polymer Spectrum after Subtraction of Polymer
absorbanceabsorbance
-.1
-.05
0
.05
3500 3000 2500 2000 1500 1000
Absorbance / Wavenumber (cm-1)
EO absorbance bands
suitable for quantitation
EO Spike in Biodegradable EO Spike in Biodegradable
Polymer MatrixPolymer MatrixFingerprint overlay of Fingerprint overlay of Residual Spiked Polymer Residual Spiked Polymer spectrumspectrum and and EO Reference spectrumEO Reference spectrum
0
.002
.004
.006
1000 950 900 850 800
Absorbance / Wavenumber (cm-1)
EO Spike in Biodegradable EO Spike in Biodegradable
Polymer MatrixPolymer MatrixResidual spectrumResidual spectrum after subtraction of after subtraction of Reference EOReference EO
Absorbance / Wavenumber (cm-1)
0
.002
.004
.006
.008
950 900 850 800
Spiked EO Concentration (barometrically delivered) Spiked EO Concentration (barometrically delivered) -- 10.95 10.95 ppmvppmv
FTIR Indicated Concentration (CLS analysis) FTIR Indicated Concentration (CLS analysis) -- 11.46 11.46 ppmvppmv
105 % Recovery105 % Recovery
EO Matrix Spike on Outgassing Medical EO Matrix Spike on Outgassing Medical
DeviceDevice
Simulated Use
EO Release @ 37oC
0.0
5.0
10.0
15.0
20.0
25.0
0.00 20.00 40.00 60.00 80.00 100.00
Elapsed Time (Hours)
PP
Mv
EO
EOEO Spike 103% Recovery
EO Spike 114% Recovery
24 Hour outgas
= 4.8 ppmv
= 38 ug EO
Recovery of EO Matrix SpikesRecovery of EO Matrix Spikes
at 121 at 121 ooCC
3 %Percent RSD
107Average Spike
1062.862.708
1035.555.387
1104.444.036
1094.434.055
1094.594.034
11014.8313.423
10511.4610.952
1017.797.711
Average Recovery (%)
Average Indicated PPM
Average Prepared PPM
Spike
Method AdvantagesMethod Advantages
RealReal--time data collection allows time data collection allows
calculation of release rates, calculation of release rates,
formation of compounds, & formation of compounds, &
outgassing endpoints outgassing endpoints
FTIR method allows simultaneous FTIR method allows simultaneous
data collection for materials data collection for materials
characterization (ISO 10993characterization (ISO 10993--18 )18 )
Method AdvantagesMethod Advantages
FTIR is additive, interferences can FTIR is additive, interferences can
be subtractedbe subtracted
Multiple compounds can be Multiple compounds can be
detected in a single test using fewer detected in a single test using fewer
devices for testingdevices for testing
Limits of Detection can be lower Limits of Detection can be lower
than GC methodsthan GC methods
Method AdvantagesMethod AdvantagesBroad range of selectivity for Broad range of selectivity for
detection of compounds (organic detection of compounds (organic
and inorganic)and inorganic)
Multiple spectral regions can be Multiple spectral regions can be
used to quantify compoundsused to quantify compounds
Recent ASTM, NIOSH & EPA Recent ASTM, NIOSH & EPA
approvalsapprovals
--Methods have been validatedMethods have been validated