Evaluation of a commercial fluorometer probe for rapid assessment of water quality trends using benthic chlorophyll Brad Rogers, PhD student, Environmental Science Dan Storm, Professor, Biosystems Engineering Derek West, Undergraduate Student, Biosystems Engineering Andy Dzialowski, Associate Professor, Zoology Bill Henley, Professor, Botany Oklahoma State University Stillwater, Oklahoma Support Provided by: Oklahoma Conservation Commission US Environmental Protection Agency, Region VI OSU Environmental Sciences Graduate Program OSU Biosystems and Agricultural Engineering
Transcript
Page 1: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

Evaluation of a commercial fluorometer probe for rapid assessment of water

quality trends using benthic chlorophyll Brad Rogers, PhD student, Environmental Science

Dan Storm, Professor, Biosystems Engineering

Derek West, Undergraduate Student, Biosystems Engineering

Andy Dzialowski, Associate Professor, Zoology

Bill Henley, Professor, Botany

Oklahoma State University

Stillwater, Oklahoma

Support Provided by:

Oklahoma Conservation Commission

US Environmental Protection Agency, Region VI

OSU Environmental Sciences Graduate Program

OSU Biosystems and Agricultural Engineering

Page 2: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

Background • Environmental monitoring

• EMAP (environmental monitoring and assessment program)

• EPA wanted more comprehensive monitoring

• Statistical comparisons

• Long term trends

• Prediction

• Data storage

• Nonpoint source management program (Sec 319)

• Water quality • Stream habitat • Aquatic communities

Page 3: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

Biologic Monitoring

• Assess ecologic conditions • Stream Habitat and aquatic communities

• Includes adjacent land use/land cover

Tar Creek (Miami OK)

Page 4: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

Assess Ecologic Condition

• Aquatic communities • Fish

• Macroinvertbrates

• Periphyton

Glover River (near Broken Bow)

Page 5: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

Periphyton Sampling Traditional Method

Page 6: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

• Commercial Fluorometer • In situ and in vivo

• Provides results in 20 seconds

• Stores files for later upload

• Measure total biomass (by chlorophyll fluorescence)

• Internal proprietary algorithm for relative abundance for three periphyton divisions

• Used in monitoring and research across the world

• Only one published comparison with traditional method

Periphyton Sampling BenthoTorch®

Page 7: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

BenthoTorch® Based on Principal of Fluorescence

• LED wave lengths (nm): 470, 525, 610, 700

• Light directed at algae in pulses (PAM)

• Chlorophyll emits light at longer wavelength (Stokes Shift)

• 700 nm used to compensate for background reflection

BenthoTorch®® Algal Class Fingerprint

Page 8: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

Objective • Assess the Accuracy of

the Benthotorch® in Estimating Total Benthic Algal Biomass Across Major Stream Types and Conditions Throughout Oklahoma

Sampling Sites by Ecoregion

Page 9: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

Results: Field 2014

• BenthoTorch®: no significant relationship with Taxonomist

• Using light adjustment & continuous measurements correlated better with laboratory results

• Variance within BenthoTorch® readings increased with increasing chlorophyll a

• Improved accuracy with non-filamentous

9080706050403020100

40

30

20

10

0

LAB Chlorophyll (ug/cm^2)

Be

nth

oto

rch

5 m

in. av

g. (u

g/c

m^

2

Y = 0.24x

R2 = 0.39

9080706050403020100

40

30

20

10

0

LAB Chlorophyll (ug/cm^2)

Be

nth

oto

rch

Sin

gle

(u

g/c

m^

2)

y = 0.17x

R2 =0.42

R2 = 0.39

y = 0.24x R2 = 0.42

y = 0.24x

Page 10: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

Conclusions: Field 2014

•BenthoTorch® may not provide useful information

• BenthoTorch® data not directly comparable to traditional methods

• Hypothesized reasons for variance in methods • Irregular substrates • 3-D structure of periphyton • Spatial variability of periphyton • Environmental conditions

Page 11: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

Laboratory Methods Concrete Culvert

Sample Collection 10 gallon Aquariums Ceramic Tile

BenthoTorch Reading on Tile

Black Calibration Plate

Spectrophotometer

Page 12: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

Results: Laboratory 2015-2016 BenthoTorch® Comparison with Lab Extracted Chlorophyll

• Significant (α=0.05) and reasonable regression equations

• Mean BenthoTorch® vs lab chlorophyll a not significantly different (paired t-test, α=0.05)

In Situ Tile Black Calibration Plate

3.02.52.01.51.00.50.0

3.0

2.5

2.0

1.5

1.0

0.5

0.0

Lab chlorophyll (µg/cm²)

Ben

tho

To

rch

BP

ad

juste

d

(µg

/cm

²)

3.02.52.01.51.00.50.0

3.0

2.5

2.0

1.5

1.0

0.5

0.0

Lab chlorophyll (µg/cm²)

Ben

tho

To

rch

on

Tile (

µg

/cm

²) R2 = 0.64

y = 0.86x

R2 = 0.77

y = 0.87x

Page 13: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

Conclusions 1. BenthoTorch® In Situ Laboratory Tiles

• Compares favorably in controlled environment with low chlorophyll a concentrations

2. Modified Black Calibration Plate method looks promising for field conditions

• Needs additional testing at sites with higher periphyton density

3. BenthoTorch® likely a good tool to detect trends in periphyton density

• Based on lab results, minimum 2 readings for each spot sampled

• Stream reach characterization methods must be developed

Page 14: Evaluation of a commercial fluorometer probe for rapid ...Oklahoma Conservation Commission US Environmental Protection Agency, Region VI ... In Situ Tile Black Calibration Plate 0.0

Questions?