Danfoss Analytical - Complexity made simple Dissolved oxygen meter Manual EVITA ® OXY 521B0270 Analytical
Transcript
Page 1: Evita Oxy Manual

Danfoss Analytical - Complexity made simple

Dissolved oxygen meterManual EVITA® OXY

521B0270

Analytical

Page 2: Evita Oxy Manual

1

Introduction...............................................

Mechanical Installation............................

Electrical Installation................................

Initial Start U

p.........................................

Program

ming U

SC

5000/6000/7000.......

Calibration...............................................

Maintenance...........................................

Other S

ettings.........................................

Troubleshooting......................................

Technical D

ata.........................................

HA

RT

®....................................................

Appendices I, IIa, IIb.................................

Contacts................................................. 2414203032344048525458 18

Contents

Page 3: Evita Oxy Manual

2Introduction

EVITA® OXY sensors are used for measuring theconcentration of dissolved oxygen in water –especially in controlling and monitoring aerationprocesses in wastewater treatment plants, but alsoin applications such as fish farms.

EVITA® OXY consists of an OXY 1100 sensor, anOXY 4100 transmitter and a USC Universal SignalConverter.

There are two versions of the transmitter – a ballfloat version, OXY 4100, and a probe version, OXY4150/OXY 3150. The operating principle is identicalfor both versions.

There are three versions of the USC - a USC 5000basic version, and a USC 6000 with addedfeatures, both point to point. A USC 7000 withHART® multidrop functionality enabling up to 15transmitters on only two wires.

The EVITA® OXY 1100 sensor is the heart of theoxygen meter. The transmitter converts measurementstaken by the sensor into a current signal and at thesame time also performs the necessary calculations/corrections. The signal converter is the display andprogramming unit and offers additional outputs.

OXY 4150/3150 transmitter

OXY 1100 sensor

OXY 4100 transmitter

USC 6000/7000 19”

USC 6000/7000 IP 67 (NEMA 4X)

USC 5000 IP 67 (NEMA 4X)

Page 4: Evita Oxy Manual

3

The EVITA® OXY system offers several advantages:

2-wire transmitter– simple and fast installation

Easy maintenance

– simple automatic calibration in atmospheric airinitiated by using the TILTCAL® or the USC 5000/6000/7000

– self-cleaning ball float

– simple sensor replacement after 2-3 years

High reliability

– self-diagnosing

– diaphragm leakage monitoring

– fault indication on transmitter current output

Flexible communication

– HART® communication as standard gives addedfeatures§ setting of measuring range and units from a

distance§ display of dissolved oxygen and temperature,

remaining OXY 1100 sensor life and specificevent codes

– Option module can be fitted in the USC 6000/7000 without the need for tools

Page 5: Evita Oxy Manual

4Mechanical installation

Mounting on handrailThe mounting bracket for OXY 4100 and 4150 canbe mounted directly on handrail using the suppliedhose clips . Retighten the hose clips after a fewdays use to ensure a tight mounting of the mountingbracket.

vmin.= 0.05 m/s (2”/sec).

Fig. 1

(max. 14’)

Transmitter

Fig. 2

In case of strong sideways forces, the PVC pipeshould be mounted with supporting wires. Thistakes up some of the sideways mechanical forceson the pipe and avoids cracking of the mountingbracket.

Page 6: Evita Oxy Manual

5

Mounting of OXY 4150 transmitter usingcable bracketThe strength of the transmitter cable is such thatthe transmitter can be allowed to hang from it.

Bracket is available as accessory supplied byDanfoss.

Fig. 4

Mounting on concrete wallThe mounting bracket for OXY 4100 and 4150 can bemounted directly on concrete wall using 4 screws(diameter max. 8 mm or 5/16”).

or 5/16”

Fig. 3

Page 7: Evita Oxy Manual

6

A: OXY 4100; 1 1/4” pipe thread

B: is supplied with transmitter

C: 45° PVC or ABS elbow; inside diameter: 50 mmor 1 1/2”

D: PVC or ABS tube; 50 mm or 1 1/2”

E: 90° PVC or ABS elbow; inside diameter: 50 mmor 1 1/2”

Q: PVC or ABS adhesive

Ball float assemblyMount the single parts as shown in fig. 5.

Fig. 5

Assembly

Page 8: Evita Oxy Manual

7

Fig. 6

Assembly (For 1½” tubes only!)The orange plastic part supplied must be insertedin the bracket with pivot function, if using 1½” PVCor ABS tubes only.

Page 9: Evita Oxy Manual

8

Mounting and replacement of OXY 1100sensorMount the sensor on the ball float as shown in fig. 7.At mark 5: Turn until a “click” is heard.

Fig. 7

OXY 4100

Page 10: Evita Oxy Manual

9

Mounting and replacement of OXY 1100sensorMount the sensor on the probe as shown in fig. 8.At mark 5: Turn until a “click” is heard.

Fig. 8

OXY 4150/3150

Page 11: Evita Oxy Manual

10

Fig. 9

Mounting on concrete wallMount the bracket on a wall using four screws(diameter max. 8 mm or 5/16”).

Mounting on handrailMount the bracket on a vertical or horizontal pipeusing the two stainless steel hose clips.

Fig. 11 Fig. 12

Fig. 10

USC 5000/6000/7000

Page 12: Evita Oxy Manual

11

Mounting in panelFirst mount the terminal board in the front panel IP65 (NEMA 4X) enclosure, figs. 13 and 14, or theback of the panel bracket, fig. 15. Plug in the 19”USC 6000/7000 signal converter in the enclosureor bracket using the four small screws to tigthen it.

Fig. 13 Fig. 14 Fig. 15

USC 6000/7000 19” version

Page 13: Evita Oxy Manual

12

Press the add-on module forward as far aspossible. See fig. 17. Avoid touching the pc boardand the sockets.

The add-on module has now been installed andthe signal converter is ready to be installed on theterminal box. Communication to the operator menuand electrical inputs and outputs are automaticallyestablished by power on.

Unpack the add-on module and locate it in thebottom of the signal converter as shown in fig. 16.

Fig. 16 Fig. 17 Fig. 18

Optional add-on module USC 6000/7000

Page 14: Evita Oxy Manual

13

Unpack the add-on module and locate it in thebottom of the signal converter as shown in fig. 19.

Press the add-on module forward as far as possible.See fig. 20. Avoid touching the pc board and thesockets.

The add-on module has now been installed and thesignal converter is ready to be installed on theterminal box. Communication to the operator menuand electrical inputs and outputs are automaticallyestablished by power on.

Fig. 19 Fig. 20 Fig. 21

Optional add-on module USC 6000/7000 19” version

Page 15: Evita Oxy Manual

14

The transmitter is connected using the two-wire,shielded cable. The two leads carry supply voltage,a 4-20 mA current signal, and HART® communication.

Examples of coupling to loop-powered display andPLC/SCADA system are shown in fig. 22, fig. 23and fig. 24, respectively.

+ –

ShieldBlack

Red

Fig. 22EVITA® OXY connected to loop-powered display.

Electrical installationOXY 4100/4150/3150 stand alone transmitter

If the cable is extended, the total length of the cablemust not exceed 1000 m (3000’). Always use two-wire, shielded cable for the extension (min. 2 x 0.2mm2 (24 AWG)).

Page 16: Evita Oxy Manual

15

EVITA® OXY supplied from PLC/SCADAsystem.

Fig. 23

Red

Black Shield

EVITA® OXY connected to a PLC/SCADAsystem with external power supply.

Fig. 24

Black

Red Shield

Page 17: Evita Oxy Manual

16

The transmitter is connected using the two-wire,shielded cable. The two leads carry supply voltage,a 4-20 mA current signal and HART® communication.

Fig. 25

Depending on the USC version the power supplyshould be connected to terminals L, N and PE whenusing 110/230 V a.c., or to terminals 1 and 2 whenusing 24 V a.c./d.c., see fig. 25 and fig. 26.

If the cable is extended, the total length of the cablemust not exceed 1000 m (3000’). Always use two-wire, shielded cable for the extension (min. 2 x 0.2mm2 (24 AWG)).

Fig. 26

USC 5000/6000 and OXY 4100/4150 point to point

USC 5000/6000 - IP 67 version USC 6000 - 19” version

Shield

Black

Red

Relay 3*

Relay 2*

Relay 1*I out 1

I out 2*

*Only USC 6000

I out 1

Relay 3Shield

BlackRed

I out 2

Relay 2

Relay 1

24V24VPE

Digital input

Digital input

Page 18: Evita Oxy Manual

17

The USC 7000 is able to communicate with up to15 EVITA® OXY transmitters using HART® multidropprotocol on only two wires.

Each transmitter should be connected in paralleloutside the USC connection box and only two wireswith shield mounted on the terminals in the USC7000.

Digital input

Relay 1

Relay 2

Relay 3

Relay 4

Relay 5

I out 1

I out 2

I out 3

I out 4

USC 7000 and OXY 4100/4150 multidrop

Fig. 27

24V24VPE

Shield

BlackRed

Page 19: Evita Oxy Manual

18

1. Apply voltage with the OXY 1100 sensor pointingupwards in free air for 2 minutes. This will zerothe sensor life counter. The life counter can beread via the signal converter (USC) or HART®

communication.

2. Place the transmitter so that the OXY 1100sensor points downwards, still in free air. Leaveit in this position for 1 hour to allow the sensor tostabilise.

3. Calibrate in accordance with the instructions inthe “Calibration” section p. 30.

Fig. 28 Fig. 29

Initial start upOXY 4100/4150/3150 stand alone transmitter

Page 20: Evita Oxy Manual

19

1. Place the transmitter in free air so that the OXY1100 sensor points downwards, see fig. 30.

2. Turn on the power to the signal converter USC5000/6000/7000.

3. Leave the transmitter in this position for at least1 hour to allow the sensor to stabilise beforecalibrating.

Fig. 30

4. Check factory settings, see p. 21. Change ifnecessary.

5. Reset the lifetime counter, see fig. 32, p. 22.

6. Calibrate in accordance with the instructions inthe “Calibration” section, p. 30.

7. After calibration is completed, place thetransmitter in the measuring media.

System with USC 5000/6000/7000

Page 21: Evita Oxy Manual

20

The keypad is used to set the EVITA® OXY and tostep through the menus. The function of the keysare as follows:

TOP UP KEY(ESC)

FORWARD KEY

BACKWARD KEY

CHANGE KEY

SELECT KEY

LOCK/

This key (hold for 2 sec.) is used to switch betweenthe operator menu and setup menu. A short presswill cause a return to the overlying menu.

This key is used to step forward through the menusin the setup menu.

This key is used to step backward through themenus.

This key changes the settings or numerical valuesin the setup menu. In the operator menu it is used tostep through the menues.

This key selects which digits to be changed.

This key allows the operator to change settings andgives access to submenus.

Ready for change

Value locked

Access to submenu

Confirmation of choice

Fig. 31

UNLOCK KEY(ENTER)

Various display symbols

Programming USC 5000/6000/7000Keypad and display layout

Page 22: Evita Oxy Manual

21

The EVITA® OXY system is supplied with thefollowing factory settings:

If factory settings are satisfactory, proceed to p. 30.

Factory settings

Parameter Factory settings Options

Oxygen unit mg/l mg/l; ppm; %

Current output 1 0-10 mg/l Range: min. 0 mg/l or ppm; 0%(System packages 1 and 2) max. 50 mg/l or ppm; 500%

Span: min. 1 mg/l or ppm; 10%

Temperature unit °C °C; °F

Current output 2 0-40 °C Range: min. -10°C (14°F)(System package 1) max. 70°C (158°F)

Span: min. 1°C (1.8°F)

Current outputs OFF OFFUSC 5000/6000/7000 ON

Relays 1, 2 and 3 (USC 6000) OFF Alarm; Warming; Limit; Timer; OFF

Relays 1, 2, 3, 4 and 5 (USC 7000) OFF Alarm; Warming; Limit; Timer; OFF

TILTCAL® Enabled Enabled; Disabled

Password 1000 1000-9999

Page 23: Evita Oxy Manual

22

The menu of the signal converter is built up in twoparts. An operator menu and a setup menu.

Fig. 32

Operator menuThe operator menu is for daily use. The signalconverter always starts up in the operator menushowing the actual concentration of dissolvedoxygen.

Oxygen

Transmitter 0

12.31 mg/lTemperature

Transmitter 0

Rem. Lifetime

Transmitter 0

Calibration

Transmitter 0

50.%No. of alarms

Transmitter 0

OxygenTransmitter 1

7.09 mg/lTemperatureTransmitter 1

Rem. LifetimeTransmitter 1

CalibrationTransmitter 1

85.%No. of alarmsTransmitter 1

OxygenTransmitter 2

2.51 mg/lTemperatureTransmitter 2

Rem. LifetimeTransmitter 2

CalibrationTransmitter 2

70.%No. of alarmsTransmitter 2

Oxygen

Transmitter 15

6.22 mg/lTemperatureTransmitter 15

Rem. LifetimeTransmitter 15

CalibrationTransmitter 15

95.%No. of alarmsTransmitter 15

0

Scan HART bus

081R

9301

.02.

01

Error pending

Reset lifetime

Transmitter 0

Reset lifetimeTransmitter 1

Reset lifetimes

Transmitter 2

Reset lifetime

Transmitter 15

18.5°C

18.5°C

18.5°C

18.5°C

0

0

0Last calibration

Transmitter 0

72. d

Last calibrationTransmitter 1

26. d

Last calibration

Transmitter 2

84. d

Last calibration

Transmitter 15

71. d

The USC 5000/6000 is point-to-point installationwhile the USC 7000 is for a multidrop system. Infig. 32 menus marked with grey are only visible inthe multidrop system. It is possible to hide some ofthe menus in the operator menu, see fig. 43, p. 34.

Menu structure

Page 24: Evita Oxy Manual

23

Setup menuThe setup menu is shown in an overview diagram(fig. 52) that is shown on the last page in thismanual and that can be unfolded duringprogramming of the USC.

The setup menu consists of two parts:

• USC setup menu - for USC settings• Transmitter setup menu - for transmitter settings

The setup menu is for commissioning and serviceand for changing the settings. Access to the setupmenu is gained by pressing the key for 2seconds. The setup menu will operate in twomodes:

• View mode is a read only mode. The preselectedsettings can only be scanned. The view mode isaccessed by pressing the key for 2 seconds.Instead of keying in the password, press thekey.

The factory setting of the password can be re-established as follows:

• Switch off the power supply

• Press the key and switch on the power supply

• Release the key after 10 seconds

The password is now reset to 1000.

• Setup mode is a read and write mode. The pre-selected settings can be scanned and changed.The setup menu is accessed by pressing thekey for 2 seconds and entering the password. Thefactory set password is 1000, but can be changedto any value between 1000 and 9999.

Set factory settings

Output when notmeasuringOutput during error

Time constant

Mfr. device type

Manufacturer Id.

Transmitter mode

TILTCAL

Event logReset lifetime

Start measuring

TILTCAL

After calibration

Calibr. type

Scan HART bus

Further info.

HW version

SW version

Serial no.

Code no.

Event log

Digital input

Relay 1-5Relay 1-5

Error pending

Scan HART bus

Password0000.

Settings

USCOutput

USCReset mode

USCService mode

USCProduct identity

USCOperator menusetup

USCPassword

USCLanguage

USCHART Settings

USC

Set defaultsetting

USCI Out 1-4USC

TypeUSC5000

USCChange HARTTrm. Mode

USCOXYUSC

Basic settings

TR. 0Advanced settings

TR. 0Send command

TR. 0Info

TR. 0Settings OXYSer.No. SSSSVVNUUÅ

TR. 0

Set transmitteraddress 00

TR. 0Oxygen unit

ppm

TR. 0CalibrationTR. 0

Error pending

Time since last

calibr.

TR. 0

I out 1-4USC

~~~~

081R

9418

.02.01

Value at 4 mA

Value at 20 mA

Salinity

Rel. humidity

Rem. lifetimeTransmitter 0

72. h

No. of alarms

Transmitter 0

12. d

CalibrationTransmitter 0

42. d

TemperatureTransmitter 0

OxygenTransmitter 0

10.64 mg/l

Reset lifetimeTransmitter 0

Last calibrationTransmitter 0

Rem. lifetimeTransmitter 0

72. h

No. of alarmsTransmitter 0

12. d

Calibration

Transmitter 0

42. d

TemperatureTransmitter 0

OxygenTransmitter 0

10.64 mg/l

Reset lifetimeTransmitter 0

Last calibrationTransmitter 0

Temp. unit

Univ. cmd. rev.

Transm. cmd. rev.

SW version

HW version

Device id.

Sensor type

Fig. 33

OVERVIEW SEE LAST PAGE

Page 25: Evita Oxy Manual

24

USC 7000 multidrop systemChange transmitter address from defaultTo allow the USC 7000 to recognise the transmitteron the HART® bus, the transmitter needs a uniqueaddress between 1 and 15. The EVITA® OXYtransmitter is factory delivered with the address 0,which is used for point-to-point installations (USC5000/6000).

To programme a transmitter address, all othertransmitters in the multidrop system need to bedisconnected from the HART® bus. To allow theUSC 7000 to locate the transmitter, press “ScanHART bus”.

Change the address to an address between 1 and15 that is not used by any other transmitter. Re-connect the other transmitters and press “ScanHART bus” to complete the installation.

Rem. lifetimeTransmitter 0

72. hNo. of alarmsTransmitter 0

12. dCalibrationTransmitter 0

42. dTemperatureTransmitter 0

OxygenTransmitter 0

10.64 mg/l

Error pending

Scan HART bus

Password0000.

Settings

USC

Basic settingsTR. 0

Advanced settingsTR. 0

Settings OXYSer.No. SSSSVVNUUÅ

TR. 0

Set transmitteraddress 00

TR. 0

Scan HART busAre you sure ?

USC

081R

9302

.02.

01

Reset lifetimeTransmitter 0

Last calibrationTransmitter 0

Fig. 34

Page 26: Evita Oxy Manual

25

If other settings are wanted, the factory settings canbe changed as shown below.

Fig. 35

Setting of the displays and current outputs units foroxygen and temperature can be done in “Basicsettings”. Setting of the transmitter’s measuringrange for oxygen can be done by changing thevalue at 4 mA and the value at 20 mA.

Setting of oxygen/temperature units and transmitter’s current outputHowever the setting of current outputs range foroxygen and temperature is done as shown in fig.36, p. 26 and fig. 37, p. 27.

Page 27: Evita Oxy Manual

26

Output

USC

I out 1USC

I out 2

USC

Function

Off

I OUT 1

On

Off

081R9412.02.01

ValueOxygen

I OUT 1

OxygenTemperature

Value at 4 mA+000.00 mg/l

I OUT 1

Range: 0-999

Value at 20 mA+999.00 mg/l

I OUT 1

Range: 0-999

Output when notmeasuring 20 mA

I OUT 1Time const.040. s

I OUT 1

030-300s

Value at 4 mA+000.00 °C

I OUT 1Value at 20 mA+999.00 °C

I OUT 1

Range: -10 - 999 Range: -10 - 999

20 mA

4 mAHoldNormal

Output duringerror Normal

I OUT 1

23 mA

3.8 mANormal

* *

* *

**

Setting of current output 1 on USC 5000 andcurrent outputs 1 and 2 on USC 6000.

Fig. 36

* Units selected in “Transmitter basic settings“, see fig. 35, p. 25.** Only USC 6000

USC 5000/6000 current output settings

Same as I out 1

Page 28: Evita Oxy Manual

27

USC 7000 current output settingsSetting of current outputs 1 to 4 on USC 7000.

Fig. 37

OutputUSC

I out 1USC

I out 2USC

I out 3USC

I out 4USC

Selecttransmitter 00

I OUT 1

00-15

FunctionOff

I OUT 1

OnOff

081R9413.02.01

ValueOxygen

I OUT 1

OxygenTemperature

Value at 4 mA+00.00 mg/l

I OUT 1

Range: 0-999

Value at 20 mA+999.00 mg/l

I OUT 1

Range: 0-999

Output when notmeasuring 20 mA

I OUT 1Time const.040. s

I OUT 1

030-300s

Value at 4 mA+000.00 °C

I OUT 1Value at 20 mA+999.00 °C

I OUT 1

Range: -10 - 999 Range: -10 - 999

20 mA4 mAHoldNormal

Output duringerror Normal

I OUT 1

23 mA3.8 mANormal

* *

* *

* Units selected in “Transmitter basic settings”, see fig. 35, p. 25.

Same as I out 1 Same as I out 1 Same as I out 1

Page 29: Evita Oxy Manual

28

USC 6000 relay output settingsSetting of relay outputs 1 to 3 on USC 6000. * Units selected in “Transmitter basic settings” fig. 35, p. 25

Same as Relay 1

Fig. 38

OutputUSC

I out 1USC

Timer interval000. h

RELAY 1

1-999 hours

Active time00 min

RELAY 1

1-59 min

Relay atactive Open

RELAY 1

Open/Closed

Relay atwarning Open

Open/Closed

RELAY 1

Relay aterror Open

RELAY 1

Open/Closed

RELAY 1

Relay 1

USCRelay 2USC

Relay 3

USC~~

081R

9414

.02.

01

FunctionOff

RELAY 1

OnTimerOff

FunctionError

RELAY 1

ErrorWarningHi LimitLo Limit

OxygenTemperature

ValueOxygen

Setpoint050.00 mg/l

RELAY 1

0 - 999.99

Hysteresis005.00 mg/l

RELAY 1

0 - 999.99

Setpoint035.00 ºC

RELAY 1Hysteresis005.00 ºC

RELAY 1

Relay abovelimit Open

RELAY 1

0 - 999.99 0 - 999.99

Open/Closed

Relay belowlimit Open

RELAY 1

Open/Closed

* *

* *

Same as Relay 1

Same as Hi Limit

Page 30: Evita Oxy Manual

29

Output

USC

I out 1

USC

Timer interval000. h

RELAY 1

1-999 hours

Active time00 min

RELAY 1

1-59 min

Relay atactive Open

RELAY 1

Open/Closed

Relay atwarning OpenOpen/Closed

FunctionError

RELAY 1

RELAY 1

Relay aterror Open

RELAY 1

Open/Closed

RELAY 1Setpoint050.00 mg/l

RELAY 1

0 - 999.99

Hysteresis005.00 mg/l

RELAY 1

0 - 999.99

Relay abovelimit Open

RELAY 1

Open/Closed

Relay 1

USCRelay 2

USCRelay 3

USC

~~

Relay belowlimit Open

RELAY 1

Open/Closed

081R

9415

.02.

01

Relay 4

USCRelay 5

USC

FunctionOff

RELAY 1

One transmitterAll ErrorsAll WarningsTimerOff

Relay atwarning OpenOpen/Closed

RELAY 1

Relay atError Open

RELAY 1

Open/Closed

Selecttransmitter no. 00

RELAY 1

00-15 ErrorWarningHi LimitLo Limit

OxygenTemperature

ValueOxygen

Setpoint035.00 ºC

RELAY 1Hysteresis005.00 ºC

RELAY 1

0 - 999.990 - 999.99

* *

* *

USC 7000 relay output settingsSetting of relay outputs 1 to 5 on USC 7000 * Units selected in “Transmitter basic setting”, fig. 35, p. 25.

Fig. 39

Same as Relay 1 Same as Relay 1 Same as Relay 1 Same as Relay 1

Same as Hi Limit

Page 31: Evita Oxy Manual

30Calibration

EVITA® OXY sensors must be calibrated inatmospheric air. Calibration every 6 months isrecommended. Calibration can be initiated from theUSC 5000/6000/7000 or by using the TILTCAL®

feature in the OXY transmitter. During calibration,the OXY 1100 sensor must not be exposed to directsunlight.

Using TILTCAL®

Hold the transmitter with the OXY 1100 sensorupwards to initiate calibration by activating the tiltswitch. After 5 min. turn the transmitter with the OXY1100 sensor downwards and the transmitter hascompleted the calibration routine. During calibration,the current output signal is factory set to 4 mA.Optional 20 mA or the last measured value (Hold)can be chosen via the USC 5000/6000/7000.

Fig. 40

Before calibration, the OXY 1100 sensor must becleaned, see section “Maintenance”. With fastcalibration in atmospheric air a 1% system accuracycan be obtained. Temperature and barometricpressure compensation is made automaticallyduring calibration via an internal temperature andpressure sensor.

Tilted Not tilted

TILTCAL®

Page 32: Evita Oxy Manual

31

If during the 5 minutes calibration period thetransmitter is turned with the OXY 1100 sensordownwards, calibration stops and measurementscontinue with the previous calibration value.Temperature compensation is made automaticallyduring calibration via an internal temperaturesensor. In addition, using factory settings,compensation is also made for air pressure,relative humidity and salinity. These factory-setvalues can be changed via the USC signalconverter or via HART® communication.

The set values and the significance of deviationsfrom them are given in appendix I.A system accuracy of 0.5 % can be achieved bylifting the transmitter out of the medium and placingit with the OXY 1100 sensor downwards for 1 hourto make sure the sensor is completely stabilised.Calibration can then be initiated.

Using the USC 5000/6000/7000In operator menu use the key to select the“Calibration” menu and press the key to initiatea calibration.Compensations for relative humidity and salinitycan be made by entering these values using theUSC 5000/6000/7000, see fig. 48, p. 37.

Fig. 41

USC 5000/6000/7000calibration

* See “Advanced settings, fig. 48, p. 37.

Page 33: Evita Oxy Manual

32Maintenance

Under normal conditions, the OXY 1100 sensor willoperate for 2-3 years and can be replaced in a fewminutes.

General maintenance of the EVITA® OXY sensor islimited to cleaning about every 2-3 months andcalibration about every 6 months.

1. Lift the transmitter out of the medium.2. Clean the sensor with pure water, to which a

little cleaning agent/washing-up liquid can beadded.

3. Dry the sensor with a soft cloth.4. Immerse the transmitter into the medium to be

measured.

CleaningDuring cleaning, the sensor must not be allowed toremain pointing upwards for more than 3 minutes ata time if the TILTCAL® function is enabled,otherwise a calibration cycle will start. TheTILTCAL® function can be disabled according to fig.48, p. 37.

Replacement of sensor1. Remove the defective/worn sensor.2. Insert new sensor (see fig. 7, p.8 or fig. 8, p. 9).3. Reset the lifetime counter (see p. 18 or fig. 32, p.

22).4. Place the transmitter so that the OXY 1100

sensor points downwards. Keep it in thisposition for at least 1 hour to allow the newsensor to stabilise (see fig. 29, p. 18).

5. Perform calibration (see section “Calibration” p.30-31).

Fig. 42

Page 34: Evita Oxy Manual

33

Page 35: Evita Oxy Manual

34Other Settings

Fig. 43

• A in the operator menu setupmeans that this reading is shownin the operator menu.

• A in the operator menu setupmeans that this reading is notavailable when viewing theoperator menu.

In the “Operator menu setup” it is possible to hideall menus except “Oxygen” in the operator menu.

Page 36: Evita Oxy Manual

35

Transmitter information

USC information

Fig. 45

System information

In the menu “Info” it is possible to view variousinformation about the transmitter, e.g. time since lastcalibration, transmitter mode and software andhardware versions.

In the menu “Product identity” it is possibel to readthe USC type (5000/6000/7000) and the serialnumber.

Fig. 44

Page 37: Evita Oxy Manual

36

The factory settings can be re-established asshown in fig. 46.

Fig. 46

In the “Send command” menu it is possible to startcalibration, to start measuring after calibration, toreset lifetime and to set factory settings.

Fig. 47

See fig. 41, p. 31

Reset mode

Send command

Page 38: Evita Oxy Manual

37

In the “Advanced settings” menu it is possible tochoose which calibration type to be used and whichtype of start after calibration. Furthermore factors forcompensation for salinity and relative humidity canbe keyed in.

Fig. 48

Advanced settings

TR. 0

0-40 g/lManual start

Automatic start

2550125 um

EnabledDisabled

Salinity040. g/l

TR. 0After calibrationAutomatic start

TR. 0Sensor type50 um

TR. 0TILTCALEnabled

TR. 0

Manual

Automatic

Calibr. typeManual

TR. 0

00-15

Set transmitteraddress 00

10-300 s

Time const.040. s

TR. 0

High

LowNormal

HighLowHold

Output duringerror Low

TR. 0Output when notmeasuring Low

TR. 0

TR. 0

0-100 %

Rel. humidity060. %

TR. 0

081R9300.02.01

“Automatic calibration” means calibration inatmospheric air and “Manual calibration” meanscalibration in a standard solution where theconcentration of oxygen is keyed in, see fig. 41, p.31.

“Transmitter address” has to be 0 for current outputto function in a point-to-point installation usingUSC 5000 or 6000. When operating in multidropsystem using USC 7000, the transmitter must havea unique address between 1 and 15.

Advanced settings

Page 39: Evita Oxy Manual

38

In the “service mode” it is possible to control therelay and the current outputs directly. This may beuseful when e.g. a control loop is to be adjusted orif an alarm signal should be checked.

Service mode

USCI Out 2

USC I Out 3

USC I Out 4

USC

Relay 1USC

Relay 2USC

Relay 3USC

Relay 4USC

Relay 5USC

Digital inputHigh

USC

(High/Low)

ModeNormal

I OUT 1

NormalForced Forced value

00.0 mA

I OUT 1

00.0-24.0 mA

ModeNormal

RELAY 1

NormalOpenClosed

Event log

USC

Log xxxxdxxhxxxxxxxxxxxxxxxx

USC

I Out 1

USC

081R

9406

.02.

01

Fig. 49

Forced current and relay outputs

Service mode

Page 40: Evita Oxy Manual

39

Page 41: Evita Oxy Manual

40

The EVITA® OXY has an advanced self-diagnosticsystem. If there is an error or a warning, a flashingbell is displayed. To view the pending errorinformation, press the key in the operator menu– see below.

The USC and the transmitter each has an ‘eventlog’ that stores past errors, information andwarnings.

Fig. 50

TroubleshootingError system

Operator menu

To access the USC event log, see fig. 49, p. 38. Tosee the transmitter event log, see fig. 44, p. 35.

Page 42: Evita Oxy Manual

41

InformationThe system will continue to measure as usual.Relay and output signals will not be affected.

WarningThe system will continue to measure, but there hasbeen an event, which may cause malfunction of thesystem, and which may require an operator. Thereason for the warning may disappear on its own.Warning relay, if defined, will show warning inaccordance with the setup.

ErrorThe whole system or parts of it is malfunctioning,and the output signal is not reliable anymore. Theerrors require an operator.Error relays, if defined, will show error inaccordance with the setup.

The EVITA® OXY transmitter error pending listshows the 5 latest occured errors that are stillpending, and the event log shows the 25 latestevents, i.e. informations, warnings and errors.

The USC signal converter error pending list canshow up to 14 pending errors and warnings,displayed with the USC errors first and then thetransmitter errors. If there are more than 14pending errors, the message “Additional errorscannot be viewed” will show in the display. TheUSC event log shows the 9 latest events.

1. digit: CategoryI(Information);W(Warning);E(Error)

2. digit: Sensor typeU for USC;O for OXY

3. and 4. digit:Error number 00-99The numbers aredescribing thefollowing events:0-9 Information10-49 Warning50-99 Error

Each event has an unambiguous identificationconsisting of 4 digits:

EO55: The OXY 1100 diaphragm hasbeen damaged.

E O 5 5

Error system

Page 43: Evita Oxy Manual

42

Troubleshooting guide - OXY 4100/3150/4150 stand alone transmitter

Symptom Current signal Fault Remedy

Constant current signal I = 0 mA Incorrect wiring Check wiring and make the necessary corrections.(varying oxygen Undervoltage Check the supply voltage at transmitter.concentration) Supply voltage must be at least 12 V.

Initialisation fault 1. Switch off transmitter for min. 5 seconds.2. Switch on again while sensor points downwards.

Transmitter defective Replace transmitterI = 3.80 mA Undervoltage Check the supply voltage at transmitter.

Supply voltage must be at least 12 V.Initialisation fault 1. Switch off transmitter for min. 5 seconds.

2. Switch on again while sensor points downwards.Transmitter defective Replace transmitter

I = 3.85 mA Diaphragm puncture Replace transmitterMoisture behindOXY 1100 sensor Fully dry the terminals at transmitter.

A hair dryer might be of assitance.I = 3.90 mA Calibration not possible, Replace OXY 1100 sensor.

worn OXY 1100 sensorI = 3.95 mA Oxygen under set measuring range Set the lower measuring limit to a lower value.I = 4.00 mA Initialisation fault 1. Switch off transmitter for min. 5 seconds.

2. Switch on again while sensor points downwards.I = 22.00 mA Oxygen level higher than set Set the upper measuring limit to a higher value.

measuring rangeError in measuring value Variable Calibration value incorrect 1. Check FSO in atmospheric air. Hold the transmitter

up in the air with the sensor downwards. Checkvalue against DS/EN 25814 (see appendix IIa)

2. Check zero point (see appendix IIb)3. In the case of large reading error, recalibrate.

Page 44: Evita Oxy Manual

43

Symptom Current output Event Cause Remedysignal code

Error in measuring value Variable Calibration value incorrect 1. Check FSO in atmospheric air. Hold thetransmitter up in the air with the sensordownwards. Check the value against DS/EN25814 (see appendix Iia)

2. Check zero point (see appendix IIb)3. In the case of large reading error, recalibrate

Display switching between End of calibration phase Go to “Send command” menu and start measuring,“Oxygen” and see fig. 47, p. 36“End of calibration”

Flashing bell in display Depending on WO10 The measured value has exceeded the Change upper range in “Basic settings”,the settings transmitter output setting see fig. 35, p. 25

WO11 The measured value is below the Change lower range in “Basic settings”,transmitter output setting see fig. 35, p. 25

WO15 Life time expired. OXY 1100 should be Replace OXY 1100replaced as soon as possible

WO20 The temperature of OXY 1100 is below Accuracy reduced0°C (32°F). The accuracy might be reduced.

WO21 The temperature of OXY 1100 is above Accuracy reduced40°C (104°F). The accuracy might be reduced.

EO50 Calibration not possible. Replace OXY 1100OXY 1100 is worn or defective

EO51 Output signal from OXY 1100 is too low. Replace OXY 1100OXY 1100 is worn or defective

EO55 The OXY 1100 diaphragm has been damaged Replace OXY 1100

EO60 Error in EEPROM Replace transmitter

EO61 Error in communication between the m-controller Replace transmitterand the A/D converter or error in the A/D converter

EO62 Error in the input circuit Replace transmitter

EO63 Error in the temperature circuit Replace transmitter

EO64 The internal power supply is too low Replace transmitter

EO65 Error in diaphragm Replace transmitter

Troubleshooting guide - OXY 4100/3150/4150 transmitter

Page 45: Evita Oxy Manual

44

Symptom Current output Event code Cause Remedy

Empty display 1. Supply voltage 1. Check supply voltage2. USC 5000/6000/7000 defective 2. Replace USC 5000/6000/7000

Dots in upper line of (I out 1) EU61 See EU61 Same as EU61display

(I out 2) EU62 See EU62

(I out 3) EU63 See EU63

(I out 4) EU64 See EU64

Flashing bell in display Undefined WU10 I out 1: Value at 4 mA >= Value at 20 mA Check I out 1 setup

WU11 I out 2: Value at 4 mA >= Value at 20 mA Check I out 2 setup

WU12 I out 3: Value at 4 mA >= Value at 20 mA Check I out 3 setup

WU13 I out 4: Value at 4 mA >= Value at 20 mA Check I out 4 setup

WU14 The measured value has exceeded the Change value at 20 mA on I out 1current output 1 setting

WU15 The measured value has exceeded the Change value at 20 mA on I out 2current output 2 setting

WU16 The measured value has exceeded the Change value at 20 mA on I out 3current output 3 setting

WU17 The measured value has exceeded the Change value at 20 mA on I out 4current output 4 setting

WU18 The measured value is below the current output 1 setting Change value at 4 mA on I out 1

WU19 The measured value is below the current output 2 setting Change value at 4 mA on I out 2

WU20 The measured value is below the current output 3 setting Change value at 4 mA on I out 3

WU21 The measured value is below the current output 4 setting Change value at 4 mA on I out 4

Troubleshooting guide - USC

Page 46: Evita Oxy Manual

45

Symptom Current output Event code Cause Remedy

Flashing bell in display Undefined WU39 HART communication error Check cables

WU40 TMS communication error Replace TMS

WU41 A parameter is out of range. Fixed at power-upIt could not be replaced by its default value

WU43 Too many errors at the same time. Power-up USCSome errors are not registered correctly

WU49 Internal error Power-up USC

One or more EU60 CAN bus application does not respond. 1. Power-up USCcurrent outputs This error is shown when the maximum number of 2. Replace USC= 3.8 mA or 24 mA communication attempts is exceeded.

Undefined EU61 USC unable to read data required for I OUT 11. No reply from transmitter 1. Check shielding and cable connections2. Transmitter has got new address 2. Power-up USC3. Another master conflicts on HART® comm. 3. Disconnect the other master or set USC

to another master, see fig. 51, p. 524. Supply voltage for transmitter too low 4. Check supply voltage at transmitter,

supply voltage must be at least 20 V5. Too much noise on current loop 5. Check shielding and cable connections6. Transmitter defective 6. Replace transmitter7. USC defective 7. Replace USC

EU62 USC unable to read data required for I OUT 2 Same as EU611 - 7: Same as EU61

EU63 USC unable to read data required for I OUT 3 Same as EU611 - 7: Same as EU61

EU64 USC unable to read data required for I OUT 4 Same as EU611 - 7: Same as EU61

Page 47: Evita Oxy Manual

46

Symptom Current output Event code Cause Remedy

Flashing bell in display Undefined EU65 USC unable to read data required for relay 1 Same as EU611 - 7: Same as EU61

EU66 USC unable to read data required for relay 2 Same as EU611 - 7: Same as EU61

EU67 USC unable to read data required for relay 3 Same as EU611 - 7: Same as EU61

EU68 USC unable to read data required for relay 4 Same as EU611 - 7: Same as EU61

EU69 USC unable to read data required for relay 5 Same as EU611 - 7: Same as EU61

Page 48: Evita Oxy Manual

47

Page 49: Evita Oxy Manual

48

Measurement Clark’s principle

Transmitter cable Two-wire, screened cable 2 x 0.75 mm2 (18 AWG) cable, length 10 m (33’)

Supply voltage 12-30 Vd.c.HART®: max. ripple (47 Hz - 125 Hz): 0.2 V p-p

max. noise (500 Hz - 10 kHz): 1.2 mV rms

Power consumption Max. 720 mVA

Current output 3.80-24 mA

Time constant 10-300 s (default 40 s)

Digital communication HART® communication superposed on current output

Alarm signal On current output (see section “Trouble shooting”)

Measuring range Transmitter:(can be configured) OXY 3150: 4 - 50 ppm or mg/l; 40 - 500 %

OXY 4100/4150: 0 - 50 ppm or mg/l; 0 - 500 %Min. span: 1 ppm or mg/l; 10 %

Sensor (recommended values to maintain operating life):OXY 1100, 25 µm: 0.002 - 2 ppmOXY 1100, 50 µm: 0.1 - 10 ppmOXY 1100, 125 µm: 2 - 50 ppm

Temperature range Transmitter:Operation: -40°C - +60°C (-40°F to 140°F)Storage/transport: -40°C - +70°C (-40°F to 158°F)

Sensor:Operation: 0°C - +50°C (32°F to 122°F)Storage/transport: 0°C - +70°C (32°F to 158°F)

Measured media: 0°C - +40°C (32°F to 104°F)

Technical DataTransmitter

Page 50: Evita Oxy Manual

49

System accuracy Digital output: ±0.1 % of FSO at calibrating temperatureAnalogue output: ±0.2 % of FSO at calibrating temperatureDigital/analogue output: ±0.5 % of FSO at 0 - 40°C (32°F to 104°F)

Reaction time OXY 1100 25 µm: τ = 7 s50 µm: τ = 22 s

125 µm: τ = 110 sDiaphragm leakage monitoring Measurement between anode and a terminal in contact with the measured medium

Enclosure OXY 4100:IEC 529: IP 68 (1 m); NEMA 6P (3’)OXY 3150/4150:IEC 529: IP 68 (10 m); NEMA 6P (30’)

Dimensions and weight OXY 4100: d = 240 mm (9.6”); 2 kg (4.4 lb)OXY 3150/4150:d = 50 mm; l = 180 mm (d = 2.0”; l = 7.2”); 1 kg (2.2 lb)

Materials PBT/PC

EMC Emission: EN 50081-1Immunity: EN 50082-1

UL General protection

EX EEx ia IIB T4 (CENELEC)

Transmitter

The worn OXY 1100 sensor must be disposed as ordinary industrial waste, and the worn OXY 4100/4150 transmitter mustbe disposed as electronic waste. Before disposing the OXY 4100/4150, the cable must be cut off and disposed as metallicwaste.

Page 51: Evita Oxy Manual

50

The worn USC 5000/6000/7000 must be disposed as ordinary electronic waste.

USC

Type USC 5000 USC 6000 USC 7000

Measuring Dissolved oxygen: 0 to 10 – 500 % Dissolved oxygen: 0 to 10 – 500 % Dissolved oxygen: 0 to 10 – 500 %range 0 to 0.1 – 50mg/l or ppm 0 to 0.1 – 50mg/l or ppm 0 to 0.1 – 50mg/l or ppm

Temperature: 0 – 70°C (32°F – 158°F) Temperature: 0 – 70°C (32°F – 158°F) Temperature: 0 – 70°C (32°F – 158°F)

Measuring Oxygen: ± 0.5 % Oxygen: ± 0.5 % Oxygen: ± 0.5 %uncertainty Temperature: 0.1°C (0.18°C) Temperature: 0.1°C (0.18°C) Temperature: 0.1°C (0.18°C)

Current outputs One 4-20 mA (scaleable), galvanic isolated Two 4-20 mA (scaleable), galvanic isolated Four 4-20 mA (scaleable), galvanic isolatedMax. load: 800 ohm Max. load: 800 ohm Max. load: 800 ohm

Relay outputs None 3 relay outputs SPST (max. 48 Va.c., 4 A) 5 relay outputs SPST (max. 48 Va.c., 4 A)

Display Back-lit alphanumeric LCD display Back-lit alphanumeric LCD display Back-lit alphanumeric LCD display

Enclosure IP 67 according to IEC 529; NEMA 4X IP 67, IP 20 according to IEC 529; NEMA 4X IP 67, IP 20 according to IEC 529; NEMA 4X

Ambient Storage: -40°C to +70°C (-40°F to 158°F) Storage: -40°C to +70°C (-40°F to 158°F) Storage: -40°C to +70°C (-40°F to 158°F)temperature Operation: -40°C to +60°C (-40°F to 140°F) Operation: -40°C to +60°C (-40°F to 140°F) Operation: -40°C to +50°C (-40°F to 122°F)

Power supply 85-264 Va.c., 50/60 Hz, 5-12 VA 85-264 Va.c., 50/60 Hz, 5-12 VA 85-264 Va.c., 50/60 Hz, 5-12 VA11-30 Va.c./d.c. 5-12 W 11-30 Va.c./d.c. 5-12 W

Automatic Compensating for temperature, barometric Compensating for temperature, barometric Compensating for temperature, barometriccalibration pressure, humidity and salinity pressure, humidity and salinity pressure, humidity and salinity

Approvals CE, C-tick and UL approved CE, C-tick and UL approved CE, C-tick and UL approvedEmission: EN 50081 Emission: EN 50081 Emission: EN 50081Immunity: EN 50082 Immunity: EN 50082 Immunity: EN 50082

Page 52: Evita Oxy Manual

51

Page 53: Evita Oxy Manual

52HART®

The OXY 4100/4150/3150 offers additional featureswhich can be accessed through the USC 5000/6000/7000 signal converter or HART® communication.To use HART® communication it is necessary touse a PC or PLC with a HART® modem andassociated software.

All Universal Commands are fully implemented, asare the necessary Common Practice Commands,but to ensure complete utilisation of the transmitter,special Specific Device Commands are used.

HART® communication makes it possible to readoff:

· oxygen concentration· temperature· remaining lifetime of the OXY 1100 sensor· alarms· indication of completed calibration

When calibrating/replacing an OXY 1100 sensorthe following can be performed with HART®

communication:

- initiation of calibration

- keying in of values for relative humidity, airpressure and salinity for compensationpurposes

- zero setting of life counter

The following programming is possible withHART® communication:

– setting of measuring range

– deactivation of tilt-switch for initiating anautomatic calibration

– setting of current output signal during calibrationfor 4 mA, 20 mA or last measuring value

– deactivation of alarms on current output

– measurement start after completed calibration

– change OXY 1100 sensor type

The USC 5000/6000/7000 can act as either a primarymaster or a secondary master. This can be set asshown in fig. 51.

Fig. 51

USC

HART SettingsUSC

Change HARTTrm. Mode

USC

USC

USC

USC

081R

9403

.02.

01

Scan HART busUSC

Scan HART busAre you sure ?

USC

Set mode toSecondary master

Curr. Trm. modePrimary master

Curr. Trm. modeSecondary Master

Set mode toPrimary Master

Page 54: Evita Oxy Manual

53

Page 55: Evita Oxy Manual

54Appendix I

Factory set values and the significance of deviations from them

Parameter Factory setting Deviation Significance (% of actual value)

Atmospheric air pressure 1013.25 mbar 10 mbar approx. 1

Relative humidity 100% 10% approx 0.3

Salinity 0 g/l 1 g/l approx. 1

Page 56: Evita Oxy Manual

55

0 / 32.0 14.62 1 / 33.8 14.22 2 / 35.6 13.83 3 / 37.4 13.46 4 / 39.2 13.11 5 / 41.0 12.77 6 / 42.8 12.45 7 / 44.6 12.14 8 / 46.4 11.84 9 / 48.2 11.5610 / 50.0 11.2911 / 51.8 11.0312 / 53.6 10.7813 / 55.4 10.5414 / 57.2 10.3115 / 59.0 10.0816 / 60.8 9.8717 / 62.6 9.6618 / 64.4 9.4719 / 66.2 9.28

°C/°F mg/l or ppm

Appendix IIa

21 / 69.8 8.9122 / 71.6 8.7423 / 73.4 8.5824 / 75.2 8.4225 / 77.0 8.2626 / 78.8 8.1127 / 80.6 7.9728 / 82.4 7.8329 / 84.2 7.6930 / 86.0 7.56

Solubility of oxygen in water inequilibrium with air (≈100%)

at 101.325 kPa

Source: DS/EN 25814

Temperature Temperature Solubility of oxygen in water inequilibrium with air (≈100%)

at 101.325 kPa

°C/°F

20 / 68.0 9.09

Page 57: Evita Oxy Manual

56Appendix IIb

Zero point check

1 g of sodium sulphite (Na2SO3) and1 mg cobalt (II) salt (Cobalt(II)chloride-hexahydrate;CoCl2, 6H2O) is added to 1 litre of water to removeoxygen from the water. The water is then stirredand oxygen concentration measured in order tocheck the zero point.

Source: DS/EN 25814

Page 58: Evita Oxy Manual

57

Page 59: Evita Oxy Manual

58Contacts

Great BritainDanfoss LimitedPerivale Industrial ParkHorsenden Lane SouthGreenford, Middx. UB6 7QETel: 020 8 991 7000Fax: 020 8 991 7171Homepage: www.danfoss.co.uk

IrelandDanfoss (Ireland) Ldt.Unit A-1Nangor Road Business ParkDublin 12Telephone: (01) 626 81 11/626 93 32Telefax: (01) 626 93 34E-mail: [email protected]

New ZealandDanfoss (New Zealand) Limited8 George Bourke DriveP.O. Box 12-422, PenroseMt Wellington, AucklandTel: (09) 270 2110Fax: (09) 270 2112

SingaporeDanfoss Industries Pte. Ltd.6 Jalan PesawatSingapore 619364Tel: (65) 2614088Fax: (65) 2610488

USADanfoss Water& Wastewater Division8800 W. Bradley RoadMilwaukee, WI 53224-9541Phone: 1 414 355 8800Fax: 1 414 355 6117Homepage: www.us.water.danfoss.com

AustraliaDanfoss (Australia) Pty. Ltd.Melbourne (Head Office)1 Ricketts RoadPrivate Bag No. 20, Mt. Waverley,Victoria, 3149Tel: (3) 95431033Fax: (3) 9543 4027

CanadaDavis Controls2200 Bristol Circle :Oakville, OntarioL6H 5R3Tel: (905) 829-2000Fax: (905) 829-2630Homepage: www.daviscontrols.comE-mail: daviscontrols.on.ca

Please find below the extract of Danfosscompanies. Other company locations:www.danfoss.com

Page 60: Evita Oxy Manual

Set factory settings

Output when notmeasuringOutput during error

Time constant

Mfr. device type

Manufacturer Id.

Transmitter mode

TILTCAL

Event logReset lifetime

Start measuring

TILTCAL

After calibration

Calibr. type

Scan HART bus

Further info.

HW version

SW version

Serial no.

Code no.

Event log

Digital input

Relay 1-5Relay 1-5

Error pending

Scan HART bus

Password0000.

SettingsUSC

OutputUSC

Reset mode

USCService mode

USCProduct identity

USCOperator menusetup

USCPasswordUSC

LanguageUSC

HART SettingsUSC

Set defaultsetting

USCI Out 1-4

USCTypeUSC5000

USCChange HARTTrm. Mode

USCOXY

USC

Basic settingsTR. 0

Advanced settingsTR. 0

Send commandTR. 0

InfoTR. 0

Settings OXYSer.No. SSSSVVNUUÅ

TR. 0

Set transmitteraddress 00

TR. 0Oxygen unitppm

TR. 0Calibration

TR. 0

Error pending

Time since lastcalibr.

TR. 0

I out 1-4

USC

~~~~

081R

9418

.02.

01

Value at 4 mA

Value at 20 mA

Salinity

Rel. humidity

Rem. lifetimeTransmitter 0

72. h

No. of alarmsTransmitter 0

12. dCalibrationTransmitter 0

42. dTemperatureTransmitter 0

OxygenTransmitter 0

10.64 mg/l

Reset lifetimeTransmitter 0

Last calibrationTransmitter 0

Rem. lifetimeTransmitter 0

72. h

No. of alarmsTransmitter 0

12. dCalibrationTransmitter 0

42. d

TemperatureTransmitter 0

OxygenTransmitter 0

10.64 mg/l

Reset lifetimeTransmitter 0

Last calibrationTransmitter 0

Temp. unit

Univ. cmd. rev.

Transm. cmd. rev.

SW version

HW version

Device id.

Sensor type

Overview

2 sec.

Fig. 52

Menus marked with grey are only visible in USC 7000* Add-on module

**

Page 61: Evita Oxy Manual

Danfoss can accept no responsibility for possible errors in catalogues, brochures and other printed material. Danfoss reserves the right to alter its products without notice. This also appliesto products already on order provided that such alterations can be made without subsequential changes being necessary in specifications already agreed.All trademarks in this material are property of the respective companies. Danfoss and the Danfoss logotype are trademarks of Danfoss A/S. All rights reserved.

Partnership with commitmentPurchasing Danfoss products gives you not onlyhigh quality and unsurpassed reliability, but also alasting partnership that will always provide you withthe technology of tomorrow.

Danfoss is well-known for accurate mechanicaland electronic systems, and is perhaps even morerenowed for its intelligent, mechatronic products.Water treatment processes always have been a vitalpart of our business, and today our commitment tothe water and wastewater industry is stronger thanever.

Our wastewater product range also containsflowmeters,water valves, pressure transmitters andfrequency converters.

Flowmeters Water valves VLT ® frequency converters

DKCFE.PS.202.B2.02/521B0270 EVITA® OXY is a trademark of Danfoss A/S